AFTER SALES TECHNICAL BULLETIN P-080 14.07.2008

Product: Originator: Issued by:


Planmeca Proline XC Timo Müller Mark Niemi

SOFTWARE VERSIONS AND


COMPABILITY INFORMATION OF
PLANMECA PROLINE XC
DIGITAL X-RAY UNIT
CAUTION
The CPU PCB and software chips are sensitive to static electric discharges.

The recommended software combinations are presented in Table 1 and Table 5. The software version of the
graphical user interface (color GUI) is reviewed in Table 4.

The spare parts order codes for chips with latest available software versions are:
Generator Processor PCB: 10010376
Keyboard Processor PCB: 10010375

NOTE! PK, PG and GUI software should be updated at the same time
NOTE! See also PLANMECA PROLINE XC X-RAY UNITGUI SOFTWARE 1.4.3.0.R

TABLE 1: RECOMMENDED SOFTWARE COMBINATION

KEYBOARD GENERATOR GUI


Planmeca Proline XC digital X-ray unit
PROCESSOR PROCESSOR
Keyboard Processor PCB (code 10009419) and
PK 5.62 PG 5.60 1.4.3
Generator Processor PCB (code 10009814)

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
1
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

TABLE 2: THE LATEST VERSION OF THE KEYBOARD PROCESSOR SOFTWARE (PK)

SOFTWARE DATE OF ADDITION AND MODIFIATION


VERSION RELEASE
PK 5.62 14.7.2008 • New PA Cross-section imaging program
• Requires Dimaxis 4.4.0
• Separate program key
• Separate patient positioning tool
• Fixed Parameter History
• Fixed DAP values
• Exposure quick settings updated
PK 5.60 3.4.2007 • New imaging program for Quality Assurance (QA - manual
10016248) together with new Al-phantom NOTE: Requires
Didapi 4.1.7 or later
• Ceph ON/OFF selection in service mode corrected so that the
selection can be made easier when error 24 appears.
• Patientpositioning lights are activated for 2 minutes (before it was
0,5 min)
• If the battery on Keyboard PCB is empty the clock will not blink
anymore. Programrecongition during startup and other usability
functions on GUI has thereby been corrected
• NOTE: Requires PG 5.60
PK 5.58 10.10.2005 • DIDAPI Error 33/ Flashing ready light correction
o 3s delay in sending ready state information to the PC.
o 3s delay in switching off the sensor head after ready
state.
• NOTE: Press exposure switch only after “waiting for
exposure” is displayed on Dimaxis program.
• NOTE: Requires PG5.54 or later
PK 5.57 6.10.2005 • DIDAPI Error 33/ Flashing ready light correction
o Added a 2s delay in sending ready information to PC.
• NOTE: Requires PG5.54
PK 5.56 27.9.2005 • Use of child slot of primary collimator in panoramic mode of a
Pan/Ceph X-ray unit.
• Added True Profile TMJ imaging mode (requires a True Profile
TMJ activation key and GUI SW 1.1.0)
• ER 60 internal diagnostics improved
• NOTE: Requires PG5.54
PK 5.55 1.6.2005 • The exposure switch type can no longer be chosen in the Service
Mode i.e. the green Ready indicator is on, when the X-ray unit is
ready for the exposure.
• Requires PG5.53.
PK 5.54 3.5.2005 • Supports the color GUI.
• Different jaw sizes and shapes can be selected.
• Segmented imaging can be selected from 5 different segments.

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
2
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

TABLE 3: THE LATEST VERSION OF THE GENERATOR PROCESSOR SOFTWARE (PG)

SOFTWARE DATE OF ADDITION AND MODIFIATION


VERSION RELEASE
PG 5.60 3.4.2007 • New imaging program for Quality Assurance (QA) together with
new Al-phantom NOTE: Requires Didapi 4.2.0 or later
• NOTE: Requires PG 5.60
PG 5.56 10.1.2006 • Tube head cooling fan output (FAN-connector, 24V) is always on.
• NOTE: Requires PK5.58
PG 5.55 15.12.2005 • Correction in image symmetry
• NOTE: Requires PK5.58
PG 5.54 27.9.2005 • Use of child slot of primary collimator in panoramic mode of a
Pan/Ceph X-ray unit.
• Added True Profile TMJ imaging mode (requires a True Profile
TMJ activation key and GUI SW 1.1.0)
• ER 60 internal diagnostics improved
• NOTE: Requires PK5.56
PG 5.53 3.5.2005 • Supports the 2-sensor-system. The panoramic sensor has
manually to be displaced, when taking a cephalometric image.
• Requires PK5.55.

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
3
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com
Customer Service Information

PLANMECA PROLINE X-RAY UNIT: SOFTWARE AND PCB COMPABILITY INFORMATION AND ORDER CODES

Publication Nr. 10013274 GENERATOR PROCESSOR PCB (Software PG X.XX)


Updated Jul 14, 2008 Order Code 00665069 Order Code 00665180 Order Code 00665181 Order Code 10009814
PCB 105-10-03 PCB 105-10-03 PCB 105-10-03 with PCB 105-10-28
Version –B6 or older Version –C or later Adapter PCB 105-10-14
Version –C or later
Order Code 00665070 PM 2002 CC / PM 2002 Proline EC/CC: No compatible PCBs No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK0 1.67, order code: 00471010
Version B2- or later - PK1 1.67, order code: 00471011
- PG1.69, order code: 00471088
KEYBOARD PROCESSOR PCB (Software PK X.XX)

Order Code 00665070 PM 2002 CC: PM 2002 CC: Proscan (with stepper motor No compatible PCBs
PCB 105-10-06 - PK2.51 order code: 00471087 - PK2.51, order code: 00471087 type ceph lock):
Version –D or later - PG1.69, order code: 00471088 - PG2.07, order code: 00471064 - PK4.54, order code: 00471031
PM 2002 Proline EC/CC: PM 2002 Proline EC/CC Film: - PG4.54, order code: 00471030
- PK3.51, order code: 00471087 - PK3.75, order code: 00471087 Proscan (with solenoid type
- PG1.69. order code: 00471088 - PG2.25, order code: 00471064 ceph lock):
PM 2002 Proline EC/CC Dimax1: - PK4.12, order code: 00471031
- PK3.48, order code: 00471087 - PG4.13, order code: 00471030
- PG3.12, order code: 00471064
Order Code 10002568 No compatible PCBs PM2002 / Proline/ EC/CC/ Film: No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK6.00, order code: 10012715
Version –L or later - PG6.00, order code: 10012716
PM2002/ Proline/ EC/CC/Dimax
2/3:
- PK5.16, order code: 10004422
- PG5.15, order code: 10004421
Order Code 10009419 No compatible PCBs PM2002/ Proline /EC/CC/ Film: No compatible PCBs Proline XC/EC Film:
PCB 105-10-27 - PK6.00, order code: 10012715 - PK6.61, order code: 10011478
- PG6.00, order code: 10012716 - PG6.53, order code: 10011479
PM2002 / Proline /EC/CC/ Dimax Proline XC/EC Dimax 2/3:
2/3: - PK5.62, order code: 10010375
- PK5.16, order code: 10004422 - PG5.60, order code: 10010376
- PG5.15, order code: 10004421
• NOTE! Concerning codes 00471087 and 00471064. When ordering software chips for units equipped with keyboard processor PCB code 105-10-06 Version K or earlier, the required
software version must be given. If the required software is not known then the equipment must be described according to the above table. The spare parts department must be informed
about the X-ray unit model and type (film/digital), sensor type (dimax1/ dimax2 or 3), keyboard processor type and version, generator processor type and version.
• NOTE! Table shows all the possible PCB/ software/machine type combinations. Combinations that are not mentioned in the table are not possible. Planmeca does not guarantee a proper
operation if older software version than the one mentioned in the table is used.

PLANMECA Oy Phone: +358-20-7795 500


Asentajankatu 6 Fax: +358-20-7795 555
4
FIN-00880 Helsinki, Finland E-mail: aftersales@planmeca.com

You might also like