9100c NXT
The anesthesia workstation
that gives you peace of mind
Precise Versatile Dependable
Enables you to effectively Scalable across a wide Based on GE/Datex Ohmeda’s
deliver anesthesia and range of patient groups legacy of 100+ years of
foster seamless recovery and surgical procedures innovation and trust
3
8
1 15
9
16
2 11 18
10 17
4 5 7
20
DOC 1973444
12
6
19
14
13
1 Flowhead assembly 8 Ventilator display for 2 waveforms 15 Auxiliary power and switch
2 Pipeline & cylinder pressure gauge 9 USB for SW update + RS 232 (15 pin) 16 Pipeline connections
3 Task light 10 Selectatec manifold & vaporizers 17 Cylinder yoke option
4 Breathing circuit with CO2 bypass 11 Ergonomic handle 18 Hose hooks
5 System switch 12 Oxygen flush 19 Scavenging system
6 ACGO port and switch 13 Wheel caster & brake 20 Flip-up shelf
7 PAW gauge 14 Storage space
gehealthcare.com
Incremental settings: PEEP high: Gas supply
Oxygen cell storage:
20 to 100 mL (increments of 5 mL) 10 to 99 cm H2O (increments of 1 cm H2O)
Pipeline input range: -15° to 50°C/5° to 122°F 10 to 95% relative
100 to 300 mL (increments of 10 mL)
280 kPa to 600 kPa/41 psi to 87 psi humidity 500 to 800 mmHg
300 to settings
Alarm 1,000 mL (increments of 25 mL)
1000 to 1,500 mL (increments of 50 mL) Sustained airway pressure:
Paw > PEEP + 10 cm H2O for 15+1 seconds Pipeline connections: Electromagnetic compatibility
Minute volume range: DISS - Male; S90- 116 (French Air Liquide);
0 to 60 L/min BSPP 1/4, BSPP 3/8 (Scandinavian) or Immunity:
Sub atmospheric pressure:
NIST (ISO 5359). All fittings available for O2, Complies with all requirements of EN/ IEC
Paw < -10 cm H2O
Pressure (PInspired) range: Air, and N2O
5 to 50 cm H2O (increments of 1 cm H2O) Alarm silence Emissions:
Cylinder input3: CISPR 11 group I class B
Pressure (Pmax) range: Mute duration: Pin indexed in accordance with CGA-V-1;
110 seconds Approvals:
10 to 99 cm H2O (increments of 1 cm H2O) contains input filter and check valve EN/IEC 60601-1, EN/IEC 60601-1-2,
Rate: Primary regulator diaphragm minimum ISO 80601-2-13.
Ventilator components
4 to 99 bpm (increments of 1 bpm) burst pressure:
2 to 60 bpm (increments of 1 bpm) 2,758 kPa/400 psig Breathing circuit specifications
Flow sensor
(SIMV, PSVProTM)
Type: Primary regulator nominal output: Operational modes
Inspiratory/expiratory ratio: Variable orifice flow sensor Pin indexed: The primary regulator is set to Breathing circuit is circle mode only
2:1 to 1:8 (increments of 0.5) pressure less than 345 kPa (50 psi).
Dimensions: Carbon dioxide absorbent canister
Inspiratory pause: 22 mm OD and 15 mm ID/22 mm ID O2 controls
Off, 5% to 60% with increments of 5% Absorbent capacity: 1200mL
Oxygen Sensor Method:
Proportionate decrease of N2O with reduction Integrated expiratory limb water reservoir
Trigger window: Type: in O2 Pressure
5% to 80% or 4 seconds, whichever is less, Galvanic fuel cell Ports and connectors
increments of 5% Supply failure alarm: Exhalation: 22 mm OD ISO 15 mm ID taper
Life Cycle:
Range: 230 kPa to 250 kPa/ 33 psig to 36 psig
Flow trigger: Approximately 12 months (Dependent on usage) Inhalation: 22 mm OD ISO 15 mm ID taper
Sounds at maximum volume every 10 seconds
0.2 to 10 L/min with increments of Bag port: 22 mm OD
0.2 L/min for volumes < 1 L/min; and Anaesthetic agent delivery O2 flush range:
0.5 L/min for volumes ≥ 1 L/min. Pressure gauge
25 to 75 L/min
Delivery
Positive End Expiratory Pressure (PEEP) Scale range: -2 to 10 kPa/–20 to 100 cm H20
Vaporizers: Tec 7+ Flowmeters
Type: Integrated, electronically controlled Bag-to-Ventilator switch
Number of positions: 2 O2 ranges:
Range: 0.1 to 1.0 L/min and Type: Bi-stable
OFF, 4 to 25 cm H2O (increments of 1 cm H2O) Mounting: 1.0 to 10.0 L/min
Tool-free installation Selectatec® Control:
Ventilator performance manifold interlocks and isolates vaporizers Controls ventilator and direction of breathing
N2O ranges:
gas within the circuit
Pressure range at inlet: 0.1 to 1.0 L/min and
280 kPa to 600 kPa/ 41 psig to 87 psig 1.0 to 10.0 L/min Integrated Adjustable Pressure Limiting
Type: Bi-stable
(APL) valve
Peak gas flow: Air range:
120 L/min + fresh gas flow 0.1 to 10.0 L/min Range: 0.5 to 70 cm H2O
Flow valve range: Hypoxic guard system Tactile knob indication at:
0 to 102 L/min 30 cm H2O and above
Fresh gas flow compensation Type:
Adjustment range of rotation:
Mechanical gear™
0 to 30 cm H2O (0 to 230°)
Ventilator monitoring
30 to 70 cm H2O (230 to 330°)
Range:
Expiratory minute volume range:
Provides a nominal minimum 21% concentration Materials
0 to 60L/min (increments of 0.1L/min)
of oxygen in O2/N2O mixture
DOC 1973444