PACSPLUS PPW

PACSPLUS Workstation
User Manual Ver 5.1

Please read this instruction before use.

V 5.0 1
PACSPLUS PPW

Caution
(1) This manual is subject to change without prior notice.
(2) If you find any error or have any difficulty to understand in this manual, please contact to our company or sales
agents.
(3) In case of misuse other than described in this manual , we disclaim all the responsibility for damage and loss.
(4) Technical support and/or upgrade service (chargeable or non-chargeable) are available only for the customer
who completed user registration.
(5) It is refundable for the claim for the defect of software within one month from the date of purchasing. If the
claim is due to the problem of computer hardware or network devices, we are not liable to any responsibility.
Please be careful on this point.

※ For the execution, it may require dongle depends on software type.

For the first time users

Thank you for your procurement of PACSPLUS PPW software.


For technical support, it requires User Registration. When you install this software, read <Software End-user
License Agreement> carefully before. You can complete < User Registration > by the instructions in the rear
part of this manual for technical support.
PACSPLUS PPW is running in the environment of Windows XP, Windows 2000 and Windows Server 2003.
( Do not use in Windows 98 and ME environment)

We recommend you backup your data by DVD or external hard disc driver for preventing loss of your data.
We are not liable to any loss of data as a result that customer does not backup their data. In addition, we are
not liable to any loss or damage as a result of any forced operation, any misuse, or any arbitrary use without
our permission.
Please contact our sales agent in the case of any malfunction in PPW software, not in the case of hardware, OS
software nor third party’s product.

Copyright
Copyright © 2006 Medical Standard Co, Ltd. All rights reserved.

Trademark
PACSPLUS is a registered trademark of Medical Standard Co., Ltd

V 5.0 2
PACSPLUS PPW

Index

IV. Application 47
I. Introduce of PACSPLUS 4
Personal Use 48
What is PACSPLUS Workstation?5
- JPEG image display 49
Key Features 5
- DICOM file import 50
Hardware Specification 7
- Export 50
Terminology 8
- Slide 50
Symbol explanation 9
- Educational File MGMT 50
- Send to Folder 51
II. Installation/Execution 10
Modification of exam info 52
Installation 11
Report 56
Execution 13
DICOM Query/Retrive 58
Terminology of each part 14
DICOM CD-R Open 59
CD Backup 60
III. How To Use 15
Long bone 61
Terminology 16
Image subtraction 62
Worklist Features 18
Send DICOM image 63
Worklist Icon 19
Rearrange image 64
Image display window 20
Auto Close 65
Toolbar 21
Chest CT View 66
View mode 22
Menu (overview) 23
V. Setting (configuration) 67
File Menu 24
Worklist Configuration 68
-Print(Paper) 25
Preference (Option) 79
-Film Print(DICOM) 25
- Network Setting 79
Edit Menu 26
- Style Setting 70
-Clipboard to Copy 27
- Environment Setting 71
View Menu 28
- Viewer Setting 72
-Scout view 29
- Display 74
-Show/Find Scout lines 30
- External Program 76
Image Process Menu 32
- Hanging Protocol 78
- Magnify Glass 33
Toolbar Configuration 80
-Real size feature 34
DICOM Network Setting 81
-Arbitrary Rotate 35
- ROI(Region of Interest) 35
VI. Trouble Shooting
Overlay Menu 36
82
Tool Menu 39
Monitor Display 83
W/L preset Menu 40
Error Message 84
LUT preset Menu 41
Sharpness preset Menu 43
Option Menu 44
-Conference folder 44
Hanging Protocol Menu 45
Macro Menu 46

V 5.0 3
PACSPLUS PPW

I. Introduce of PACSPLUS Workstation

V 5.0 4
PACSPLUS PPW

What is PPW?

・ PACSPLUS Viewer PPW is a software utilized in patient treatment, education


activity, research activity and conference. It inclusive functions which
manages the search of medical images, display, inquiry and saving efficiently
in the database.
・ PPW provides functions such as DICOM storage, DICOM query and retrieve and
DICOM print by the DICOM standards. Accordingly, integrate with 3rd vendor’s
PACS product or modality is available.
・ When PPW interface with HIS (Hospital Information System) or EMR (Electronic
Medical Record), more efficient management of patients data and medical
image is available.

Key Features

・Create Local database and conference folder.


・DICOM image display, Non-DICOM image converts to DICOM (JPEG,
TIFF,PCT,BMP).
・DICOM storage, query & retrieve, print, DICOM DIR
・Support multiple monitors (2 high resolution black & white monitor and 1 color monitor
can be used simultaneously).
・Support TWAIN driver.
・CD, DVD backup, auto-execution module.
・Various image processing and image editing.
・Long Bone Stitching .
・CT ratio.
・Measuring function of Cobbs angle for bent spine.
・Mammo display (Mammo data).
・Auto log Off.
・Scout view of MR and CT image.
・Automated QC using Look-up Table (LUT).
・Image hanging protocol .
・Interface with 3D program.
・Interface with HIS/RIS.
・Interface with External program.
・DICOM JPEG 2000.

V 5.0 5
PACSPLUS PPW

1) Image Receive
2) Image Routing
3) Image Display
4) Image Backup Backup
CD/DVD

Send
PPW
Viewer AE

Database
CR Save

Send DICOM Print Film


Router
Router AE imager

DICOM Store PACS


Server

PPW

V 5.0 6
PACSPLUS PPW

Hardware Specification

1. Minimum requirement
・CPU: Pentium IV, 2GHz or higher
・Memory: 1GB or higher
・HDD: 80GB or higher
・Ethernet network card: 100Mb NIC
・OS: Windows 2000 Pro/XP Professional
・Monitor: Resolution 1280X1024 or higher, 32bit Color 17“ or higher

2. Recommended requirement
・CPU: Pentium IV, 3 GHz or higher
・Memory: 2GB or higher
・HDD: 80GB or higher (SCSI)
・Ethernet network card: 100Mb or Gigabit NIC
・OS: Windows 2000 Pro/XP Professional
・Monitor: Resolution 1280X1024 or higher, 32bit Color17“ or higher

* Diagnostic Monitor

1)1 Monitor
・ Diagnostic 3M pixel/5M pixel high resolution monochrome monitor:
Ex) Trupix, Barco, EIZO, Siemens, Totoku, etc
・ Graphic card: Single head high resolution card 1set
Ex) Astro, DOME, Real Vision or Barco card

2)2 Monitors
・ Diagnostic 3M pixel/5M pixel high resolution monochrome monitor 2sets
Ex) Barco, EIZO, Totoku, Trupix, etc
・ Graphic card: Dual head high resolution card 1set
Ex) Astro, DOME, Real Vision or Barco card.

3) 2 Monitors
・ Diagnostic 3M pixel/5M pixel high resolution monochrome monitor 1set
・ Display HIS/RIS on color monitor 1set
・ Graphic card: VGA card 1set and single head high resolution card 1set

4) 3 Monitors
・ Diagnostic 3M pixel/5M pixel high resolution monochrome monitor 2sets
・ Display HIS/RIS on color monitor 1set
・ Graphic card: VGA card 1set and dual head high
resolution card 1set

V 5.0 7
PACSPLUS PPW

Terminology

Term Explanation

PPW PPW 20

Viewer PPW 20

Menu Image display window on pull-down menu bar

Worklist Worklist window

Image Display Window Image display window

ToolBar Tool bar

Exam List Exam list on worklist

Exam Examination ex) Brain MR

Series Series ex) T1-weighted axial of brain MR

Image Images ex) Images of T1-weighted axial

Local Host Host name of PPW

Remote Host Remote viewer, server, modality host name

Mouse right Click Mouse right click

Mouse left-click and Drag Mouse left click and drag

Mouse Scroll Mouse Scroll

Modlity Modality

W/L Window Level

V 5.0 8
PACSPLUS PPW

Symbol explanation

Symbol Explanation

On/Off On/Off by Check / Uncheck

memo Memo or hint

Ref

Caution

[Order →] Click order. ex)[order Menu File → Exit]

< > Click “< >”

Bold Click bold

Italic Type in Italic

manual Refer to the manual

* Extra explanation

1 Extra explanation of the image

① Display the steps to operate

【Ctrl+A】 Ctrl key + A key at the same time

【Shift+A】 Shift key + A key at the same time

Move to previous page on the exam

Move to next page on the exam

Move to previous Exam / Series

Move to next Exam / Series

{ }\ Path ex) {image PATH}\{STUDYDATE}\{PATIENTID}\

V 5.0 9
PACSPLUS PPW

II.PPW Software
Installation / Execution

V 5.0 10
PACSPLUS PPW

Installation


① ③

④ Select number of monitors to be used

One Monitor
1 Worklist + 1 Viewer
④ 1 Worklist & Viewer

Two Monitor
1 Worklist + 2 Viewers
1 Worklist & 1 Viewer + 1 Viewer


⑦ ⑧

V 5.0 11
PACSPLUS PPW

PACSPLUS Viewer Installation

V 5.0 12
PACSPLUS PPW

Execution

It pops-up the message for the very first time.

Please click ‘ok’.

Log on
①User ID, Password.
ex)Type in pacsplus for user ID, Type in pacsplus for password.
② Select Local from DB server.
Log out(PPW Exit)
[Order Menu, File → Exit]

Log off
Log on with another ID and password.
[order Menu, File → LogOff]

V 5.0 13
PACSPLUS PPW

Terminology of each part

1. Worklist (Exam search)

Tab

Search
condition

Exam List

Old Exam Report

Exam info
2. Exam display window Menu
Image
Close
Layout
Exam
Exam Open

Tool Bar

Exam list

V 5.0 14
PACSPLUS PPW

III. How To Use

V 5.0 15
PACSPLUS PPW

Terminology

Icon Function Short cut Menu

Open Worklist Ctrl+Tab File → Open Worklist

Program Exit F3 File → Exit

Image close File → Close

Arrow
Image Process→ Arrow
(Return the features)

Open next exam Ctrl +


File → Next Exam
Right Arrow

Image Layout

Select all images Ctrl+A Edit → Select All (A)

Cancel selection Ctrl+E Edit → Deselect All (E)

Copy to clipboard Ctrl+C Edit → Copy to Clipboard

Use window level on


W Image Process → Window Level
Selected image

Use window level on Image Process → All Window Level


Every images of series

Zoom in / out Z Image Process → Zoom In/Out

Panning P Image Process → Panning

Magnifying a certain
G Image Process → Magnify glass
area
Initiation of processed
Image Process → Display All Original
image

Initiation of selected
Image Process → Display Original
iamages

Report R Tool → Report

Print Ctrl+P File → Print

V 5.0 16
PACSPLUS PPW

Icon Function Short cut Menu

Measurement of length Annotation → Measure

Measurement of angle Annotion → Angle

Text Annotion → Annotion

Picking Annotion → Picking

Display Pixel value Alt + L Annotion → Pixel View

V 5.0 17
PACSPLUS PPW

Worklist Features

① ② ④

⑧ ⑨

① Database Tab.
② CD Backup Tab.
③ Search condition setting.(Type in or select arrow)
④ Click Search.
⑤ Exam List. (search exam info that transmitted from modality)
Multi exams display. (With shift, CTRL key, select multi exams)
⑥ Sort the exam list by click. Please select the image to be opened and double-click the image.
⑦ Click Exam Open.
⑧ Old exam list on the exam list.
⑨ Display report of exam on selected exam on the exam list.
⑩ Search on the condition of searched, such as number of exam, series and image.

V 5.0 18
PACSPLUS PPW

Worklist Icon

Application Application
Modify Exam Macro Manager)

Setting

Merge Exam Worklist Configuration

Merge Series Search Filter Clear

Split Exam

Split Series Report View

Delete Exam Search

Application

DICOM Send SCU) Exam Open

Local Send to Folder)

V 5.0 19
PACSPLUS PPW

Image Display Window 3 4


5
2

1. Exam list window.


2. Display searched exam.
3. Movement of Image.
4. View window exit.
5. Search another exam and
image layout adjustment.

Memo
① ②

① Left-click of mouse ②【Left click of mouse】


Drag for the layout to be
desired

V 5.0 20
PACSPLUS PPW

Toolbar

Open Worklist Close Exam Open Next Exam

2 Series View
Pointer 1 Series View
(Vertical)

Report Cine Mode 2 Exam View

Set Layout
Reset All Images Compare Mode
(manual)

Reset Selected W/L Selected


Images Images W/L All images

Measurement Flip HORZ. Flip VERT.

Rotation Rotation Counter


Angle
Clockwise Clockwise

Text Arrow Inverse

Auto W/L
On/Off
Pixel View Zoom In/Out

Scout line Auto Close


Refresh
On/Off On/Off

Please select Option → Toolbar Configuration to display other icons.

2 Series View Measurement


Print Auto Select
(Horizontal) Comparisom

Comment 4 Series View Circle Polygon

Free Curve DICOM Header Magnifying glass Rectanglar

Chest CT Ratio Free Hand

V 5.0 21
PACSPLUS PPW

View Mode

Exam Series #1 Image #1


Image #2
Series #2
Image #n
Series #n

1 Series
Display one series window without split

2 Series(Left/Right)
Left/Right 2 split

2 X 2Split
2X2 (Up/Down・Left/Right 4 Split)

Series Layout (Scout View)


Adjustment of necessary series layout. (Max 4 X 4)

Same Image comparison


Able to compare same image with different W/L value .
[Order Menu, Option → Preference → Viewer → Chest CT View W/L value setting]

Exam Comparison Mode


With one patient, compare CT, MR series and able to compare different patient image.
Left bottom
Select exam on the exam list that image display window. Drag & drop.
Cine view (Cine, Tile, Stack)

Speed adjustment Range adjustment

:Move to next right : (Play) : Move to all image with fast t


image : (Reverse play) rack
:Move to next left image : (Pause) : Move to next image with fast
: Move to start point track
: Move to last point : Play repeat
: Pause play / stop

V 5.0 22
PACSPLUS PPW

Menu

1 2 3 4 5 6 7 8 9 10 11 12

1 2 3 4

5 6 9 11

10 12
7

V 5.0 23
PACSPLUS PPW

File Menu

1. Open Worklist【Ctrl+O】
1 Worklist windows open.
2 2. Save 【Ctrl+S】
3 Save.
4
3. Close
5
Close image.
6
4. Next Exam
Application
Close current exam and open next exam.
5. Reserve Exam (withhold )
Keep to withhold image search.
Application <Auto Close>, display exam list on the left bottom.
6. Close All
Close all searched exams.
7. Print.
Please refer to the next page.
7 8. Log Off.

Refer to the next page 9. Exit.

8 Exit PPW.
9

V 5.0 24
PACSPLUS PPW

l Print Feature

1. Paper Print

[ Order of Menu, File →


Print → Paper]

2. Film Print (DICOM)

[Order of Menu, File →


Print → DICOM]

Before use of film Print (DICOM), setting of DICOM printer is necessary.


Memo
Please contact agent of DICOM printer for setting.

V 5.0 25
PACSPLUS PPW

Edit Menu

1
or
2
3
4 1.Delete selected 2.Delete unselected 3.Restore deleted
image. image. image
5

Refer to next page

6
7
4.Hide selected 5. Display hide image
image

6.Select all 7.Cancel selection


【Ctrl+A】 【Ctrl+E】

V 5.0 26
PACSPLUS PPW

l Copy to clipboard
(Create slide, edit documents)

1) How to copy all Images


a. Select Image.
b. [On Menu, Edit → copy to clipboard] 【Ctrl+C】
(Open Word or powerpoint and , paste with 【Ctrl+V】)

2) How to copy a part of Image


a. Select Image.
b. With ROI function, appoint ROI.
c. Right-click on the boarder of selected region.
d. Select Copy to Clipboard.

* Property : Property of ROI

V 5.0 27
PACSPLUS PPW

View Menu

1 On/Off 1. Auto close


If Auto close is on and open a exam of next patient, current
2 opened image will be closed automatically.

2. Refresh
If Image not found is display or the number of image
display is incorrect, refresh the image again.
3
3. DICOM overlay, ROI/Annotation, Ruler Show/Hide
Caption
On/Off

Refer to next page

On/Off

4. Toolbar, Caption, Status Bar


On/Off

V 5.0 28
PACSPLUS PPW

1. Scout View (Series Layout)

①Click series layout of Toolbar,


②Setting layout of view window.
③Adjust layout of series unit.

5 5

④Scroll the mouse of selected series image,


⑤Scout line will be moving of another series.

V 5.0 29
PACSPLUS PPW

2. Show Scout lines

1) From selected series, right-click of the mouse → Show Scout lines.


2) Image from selected series, every scout line will be displayed.
3) Click a certain point of part (Arrow), those certain image and scout line will be displayed.

3. Find Scout Image

1) Display 1 series mode of MR image,


2) On Menu, View → Add Scout Image.
3) Display Scout Image Add Dialog (①).
4) Display image of clicked series (②).

V 5.0 30
PACSPLUS PPW

Open very first exam


Open previous exam open
Open next exam
Open last exam

① On the top of combo box of Add scout Image


Dialog, select scout image.
②Click Add Image.
③It’ll be inserted into fore part of the series.

* Refer to 4. Scout Image

4. Scout Image
1) On feature of Add scout image,
2 2) Check Unselected line draw,
3) Scout window will be displayed(①).
4) If you click on a specific slice line, related image
will be displayed (②).
1

V 5.0 31
PACSPLUS PPW

Image Process Menu

1. Arrow ESC
1
Able to return to basic mode which uses Arrow.
2
3
2. Zoom In/Out Z
4
5 After click Zoom In/Out, appoint the area to be zoomed with
arrow. Select all image【Ctrl+A】, all image will be applied. For
6 zoom in, arrow will be centering around and operates.
7 3. Magnify Glass G
8 1) 【Shift+Mouse Drag】
Magnify glass range.
9 2) 【Ctrl+Mouse Drag】
10 Adjust zoom in/out on the range.
3) 【Alt+Mouse Drag】
Adjust W/L on the ragne.

Refer to next page 4. Panning P


5. Window Level W
Drag and W/L value will be modified.
Adjust W/L on selected image.

6. All Window Level


Adjust every W/L value with drag and right-click of mouse.
Refer:
Refer to next page Shift (x5 Speed)
Ctrl (x10 Speed)
Alt (W/L ROI window size)

7. Auto Leveling On/Off


Auto W/L setting.

8. Inverse
Positive/negative image inverse.

9. Display Original
Return to selected original image.

10. Display All Original


Return to all searched original images.

V 5.0 32
PACSPLUS PPW

l Magnify Glass

Area【Shift+Mouse Drag】

l Rectangle (Oval) ROI Image Process


(ROI내의 W/L, 확대, Sharpness)

Zoom【Ctrl+Mouse Drag】 【Alt+Mouse Drag】

V 5.0 33
PACSPLUS PPW

lReal size feature

Display real size of image and measure a certain area

1. Types of Real Size Display


1) Default Size Display:If there’s a pixel space info in DICOM header, automatically it will display as Cm with data.
2) Manual Real Size Display: if there’s no pixel information in image, it is not able to measure to display real size of
image. For this case, value of pixel will be displayed.
User sets monitors setting value by manually. (real size, resolution).

2. Manual Real Size Setting.


Ex) There’s no pixel value on DICOM header information of captured image

[On the menu, Image Process → Manual Real Size Setting → Monitor’s specifications]

3. Manual Real Size Mode


Setup manually and it applies other images.
(check / uncheck)

4. Display Manual Real Size


On the selected image, use the specification which came from monitor.

V 5.0 34
PACSPLUS PPW

l Arbitrary Rotate


①Move a needle of window left / right to adjust angle.
②Save.

Remark: Please use this function to change DICOM image and it is not able to return to original.

l Rectangle (Oval) ROI Image Process


(W/L on ROI, Zooming, Sharpness)

Select rectangle or circle of ROI for image processing and adjust w/l with mouse drag.

1) W/L change: Drag mouse with right-click on a certain area where was appointed.
2) Zooming:Drag mouse on a certain area.
3) Sharpness change:Right click to select sharpness.

V 5.0 35
PACSPLUS PPW

Annotation (Overlay) Menu

1. ROI (Region of Interest)

1
2
3

4 [ On the boundary line, right-click à ROI à area, location, length, etc


information will be displayed]
(For CT image, hounsfield unit value is Pixel)
2. Measure (length measurement) M
3. Compare Measure
4. Angle measurement A

* Angle measurement between two lines which are not cross each othe
r:
[Draw two lines (Measure) → press shift and select another line → M
ake angle information message will be displayed → Yes → display angl
e of two lines]

1. ROI

(외각) (내각)

< Rect > < Circle> < Polygon>

V 5.0 36
PACSPLUS PPW

< Auto Contour> < Freehand Draw >

3. Compare measurement 4.Angle measurement

V 5.0 37
PACSPLUS PPW

5. Annotation T
6. Arrow Line
7. Freehand Line
8. Pseudo-color
9. Picking K

5 10. Profile
6 11. Pixel View L
7 Display Pixel value.
8
12. Chest CT Ratio (CTR)
9
With shifting lines, display CT ratio (%)
10
11 13. Bent spine angle of cobbs measurement
12

13
8. Pseudo-color 9. Picking 10. Profile

11. Pixel View 12. Chest CT ratio 13. Scoliosis RT/LT

V 5.0 38
PACSPLUS PPW

Tool Menu

1. Report R
1 Report creation.
2
Application

Auto Font: Print font adjustment.


Save & Close: After Save, close report window.

2. Comment
Display comment on worklist.

V 5.0 39
PACSPLUS PPW

W/L Preset Menu

W/L Preset

Adjust W/L with saved short-Cut key.


Adjust W/L by Modality and select Short-Cut key then type in W/L name to be saved.
After that, W/L name will be displayed on W/L preset. To open the exam next time,
adjusted W/L will be applied.

Caution It applies to series unit only.

V 5.0 40
PACSPLUS PPW

LUT Preset Menu

Adjust LUT curve by user, it is able to adjust gray scale range of image. (10 bit or higher gray-level image)
1) With two pointer on the line, adjust curve of S and check the change of image processing.
2) Adjust W/L with right click on image window.
3) Modality, exam part, series, exam unit, it is able to setup LUT value and W/L value.
(It depends on privileges by user.)

Lookup Table Tool

Drag two points, adjust curve


Adjust selected shape minutely
Add point (Add)
Delete point (Delete)
Reset: Return to original value
Vert: Top/Bottom curve change
Horz: Left/Right curve change

Caution

Based on modality or body part information, it applies with DICOM information. User type in the contents and
modality information and it operates automatically (auto matching).

V 5.0 41
PACSPLUS PPW

l LUT preset


1) LUT preset check of Menu.


2) After image processing (LUT curve adjustment, Window Level adjustment),
check preset of LUT (①).
3) Click Add(②),
Station name and body part of image will be displayed automatically.
4) Type in LUTName (③),
5) <OK> click (④) .
If there’s Series Description, acquisition device of body part on image.

l How to setup auto LUT

[On menu, Option → Preference → Environ


ment Setting]
1) Server : it uses LUT value when image is c
alled from server.
2) Local: it uses LUT value when image is ca
lled from local.

V 5.0 42
PACSPLUS PPW

Sharpness Preset Menu

1. Sharpness Preset
1
sharpness value applies to level 17 which already arranged.
2
2. No Sharpen
Sharpness display will not be in use.
3
3. Grade 1-30
Type of definition.


③ Add Click window

Update Click window

4. Add 5. Update
1) [On menu, Sharpen Preset → Sharpen 1) [On menu, Sharpen Preset → Sharpen
Preset] Preset]
2) Adjust Sharpen value to be desired.(①) 2) Adjust Sharpen value to be desired.(①)
3) Add Click. 3) Appoint update list → Update Click.
4) Type in preset name on add info → OK.(②) 4) Type in preset name in update information → OK.(②)
6) Preset will be added which was registered on the list. * When it updates, preset name is not necessary to type
in again.
6) Appointed list will be updated.

V 5.0 43
PACSPLUS PPW

Option Menu

Conference Folder
Setting

Setting

1 2

1. conference folder Add


1) [On menu, Option → Conference Folder → Folder Add]
2) Type in FolderName → Next.(①)
3) Input the path of folder → Next.
4) In add folder info, writer, folder detail, folder name is required to fill in → Next.
5) OK.
6) New conference folder will be created on worklist. (②)

2. Conference Folder Delete


1)[On menu, Option → Conference Folder → Folder Delete]
2)Please select folder to be deleted → Next → OK.

V 5.0 44
PACSPLUS PPW

Hanging Protocol Setting Menu

1 Series of Exam 1 2 Series of Exam 1

S1 S1 S2
S1: Series 1
S2: Series 2
S3: Series 3
S4: Series 4

E1: Exam 1
E1: Exam 1

3 Series of Exam 1 4 Series of Exam 1 2 Exams

S1
S1 S2

S2 E1 E2

S3 S4
S3

Refer Please refer to setting option of hanging protocol in 2 Monitor system

Memo
Exam Series #1 Image #1
Image #2
Series #2
Image #n
Series #n

V 5.0 45
PACSPLUS PPW

Macro Menu

1. Macro
Viewer Macro
1



Save the value of image processing, it applies to the image which is


opened currently at anytime.
③ 1) Type in MacroName, Macro detail (①) → Record(②) ex) Chest
3) Fill in add macro name and description then click OK.(③)
4) After image processing, click stop button.(④)
5) To apply macro feature, select added list to macro menu.(⑤)

V 5.0 46
PACSPLUS PPW

IV. Application

Educational purpose of M.D or conference, individual database can be created.


Conference Folder provides various features.

Please read this instruction before use.

V 5.0 47
PACSPLUS PPW

Personal Use

1. JPEG image display


Using the function of import non-DICOM file of PPFOCUS, you can open JPEG
image (or TIFF, BMP) and then convert them to DICOM files and save them to
local database. Using export function, you can convert DICOM files to non-
DICOM files like JPEG, TIFF, BMP, etc.
(Refer to V. Advanced features > Open Non-DICOM file)

2. Slide or document making


Using the function of Copy to Clipboard (with ROI function), you can copy a
DICOM image or its part to Word, PowerPoint, etc.
(Refer to IV. How to use > Copy to clipboard)

3. Educational purpose
Using Reference image view function, you can review reference images during
diagnoses, patient care and education, etc.
(Refer to V. Advanced features > Reference image view)

4.CD backup
Using the function of CD backup, you can copy DICOM images and database
of PPFOCUS to CD/DVD for media backup, educational purpose, CD copy
when the transfer patients to other hospitals.
(Refer to V. Advanced features > CD backup)

V 5.0 48
PACSPLUS PPW

1. JPEG Image search


(ImportNon-DICOM file )
Import Non-DICOM file
From JPEG file, save in local database.
1) [On menu File → Non-DICOM File Open]
2) Select image.
3) [On menu File → Import]
4) From import Into Local DB Window, type in patient information and exam information.
5) Click OK and save in local database. (On exam list of Worklist, image search)

Applied Range:
Selected Image:import selected image.
All Images:import all images.

V 5.0 49
PACSPLUS PPW

2. DICOM file import


(Open DICOM file and save into local database에 save)

1) [On menu, File → Open DICOM File]


2) Select DICOM file in Directory Window.
3) [On menu, File → Import]
(refer to previous page)

Refer
Keep original : Keep current compression mode.
Uncompression : Cancel every compression.
Lossless JPEG : JPEG compression without lossless.
Lossy-Minimal : Minimal loss compression.
Lossy-Mild : Mild compression.
Lossy-moderate : Moderate compression.
3. Export
Lossy-Severe : Severe compression.
Lossy-Very severe : Very severe compression.

From exam display window, select DICOM image and appoint file name then save.
(Select one of type of DICOM, JPEG, BMP, TIFF)
1) Among displayed image, select image to be saved.
2) [On menu, File → Export]
3) From export window, select file type(DICOM, JPEG, BMP, TIFF) → select compression type.
4) Appoint location to be saved.
5) Save.

4. Slide (document edit)


Reference III. Copy To Clipboard

5. Educational File Management


With conference folder, education, seminar database management.(CD backup)

Reference III. How to use>Option Menu>Conference Folder

V 5.0 50
PACSPLUS PPW

6. Send to Folder

Transmit exam information to local or conference folder.

[To send to folder, select exam to be transmitted → right-click of mouse → Send


To Folder → select destination database (local or conference Folder) → Send]

caution To transmit to the conference folder, write conference folder in


advance.

V 5.0 51
PACSPLUS PPW

Modification of exam information


1 2

1 2

1. Merge Exam 2. Merge Series


Exam merges to another exam. Merge multiple series to one.

1) Same patient information will be displayed in exa 1) Appoint key series from series title.
m list. a. Select key series in series list.
a. Select exam → Exam merge. b. click <>> (<<> is cancel to appoint) → merge.
b. Type in Patient ID → Search → On database, pati
ent information to be modified will be displayed on r
ight side. 2) Appoint key series to be merged.
c. Click patient → Merge. a. Select series in series list.
b. click <>> (<<> is cancel to appoint) → merge.
2) Merge to another patient exam
a. Type in patient ID to be modified → Search.
b. Select exam on displayed list → merge.

V 5.0 52
PACSPLUS PPW

3 4 5 6 7

② ①
① ③


③ ④

3. Split Exam 4. Split Series


Split exam of series. Split image as series.

1) Select current exam to be split from series list. 1) Select series to be split in series list.
2) Split current selected exam. 2) Split current selected series.
3) Click <>>. (<<> cancel to appoint, (Shift or Ctrl to sel Select new series to be split from image list.
ect multiple,<>>>/<<< > select all / cancel to select all) 3) Click <>>. (<<> cancel to appoint, (Shift or Ctrl to select
4) Split series. multiple,<>>>/<<< > select all / cancel to select all)
4) Split series.

5. Delete Exam

6. DICOM Send SCU

Reference 62p. Refer to Network Setting

7. Send to Local Folder

Reference 47p. Refer to send to Folder

V 5.0 53
PACSPLUS PPW

8. Macro
Worklist macro



It is able to search exam list by defined user’s search condition.


<ex> For general chest exam, click search to display exam list.
1) Type in search condition, click Macro.
2) Type in MacroName, Macro (①) → Add(②).ex) CR-Chest
3) Added information will be displayed (③).
4) Close(④),
5) Click macro button that created on worklist display. (Button name is same as macro name.)

* After selection of macro condition to be deleted, click Delete.

V 5.0 54
PACSPLUS PPW

Report

1.One patient and multiple diagnostic reports

①Check same report to be diagnosis on reading selected exams.


②Save.

Save & Next macro feature is that save current diagnosis and acquires exam to be
Reference displayed from worklist.

V 5.0 55
PACSPLUS PPW

2. Different patients and multiple diagnosis

① Click Shift or CTRL, select exam to be diagnosed.


② Create a Report.
③Save.

V 5.0 56
PACSPLUS PPW

DICOM Query/Retrive

① ②

④ ⑥

⑤ ⑦

Through 3rd vendor’s DICOM Q/R server of PACS, it is able to query and search DICOM image.
① Select 3rd vendor’s server → Click Connect.
② Select move AE to be retrieved.(for move type)
③Search. (or, type in patient ID and patient name to be searched → Search)
④ From ③, searched data will be listed.
⑤ From ④, study information of selected patient will be listed.
⑥ From ⑤, study series information will be listed.
⑦ From ⑥, image information of selected series will be listed.
⑧ On each list, selected images will be added on the list of Retrieve → Delete-All, Select All → Retrieve
click.
⑨ Image count and stop button

V 5.0 57
PACSPLUS PPW

DICOM CD-R Open



③ ⑤

Search DICOM image of CD. (Also there’s DVD that is a DICOM compatible)

① Click DICOM CD-R Open tab on Worklist.


② Click <…> → select drive that CD is in it.
③ Search. (or, type in patient ID and patient name to be searched → Search)
④ Select patient on the left patient list.
⑤Open.

V 5.0 58
PACSPLUS PPW

CD Backup

Caution Saved data in hard-disk of PPW must be backed up on CD/DVD periodically.

DICOM CD(or DVD) backup


Memo 1) For transferring patient with image of CD backup.
2) Research or educational purposes to build personal database.
3) Mini PACS, off-line backup.



③ ④

⑦ ⑧

①Select CD Backup in worklist.


②Select Database (able to select conference folder)
③Select Study Date,
④ Click Search.
⑤Select exam on the exam list (Shift or Ctrl key for multiple exam selection)
⑥Click Backup.
⑦Complete back up.
⑧Burn a CD/DVD.

V 5.0 59
PACSPLUS PPW

Long Bone

① ②

③ ④

Memo Search images to be stitched from Worklist and open to do long bone image stitching.

①[On menu, click Tool → Long Bone Image]


② Select first, second, third images to display message on each image.
③[On menu, click Tool → Long Bone Image Stitching]
④[On menu, File → Save]

V 5.0 60
PACSPLUS PPW

Image Subtraction

① Source Image ② Destination Image

③ ④

①[On menu, click Tool → Select source/Destination image → select source image]
② Select mask image.
③[On menu, click Tool → Subtraction]
④ Display image subtraction to show blood vessel easily.

V 5.0 61
PACSPLUS PPW

Send DICOM image

Send DICOM to folder, image send image by e-mail.


[On menu, select File → Send to Folder]
[On menu, select File → Send to E-mail]

Remark:
Select image to be transmitted for sending only for selected images.
(CTRL or SHIFT key to select multiple image)

E-Mail sending
By outlook express, transmit image. Such as DICOM, JPEG, Bitmap, TIFF types of image.

1) Select image to be e-mailed.


2) Attach images to outlook and send. [On menu, File → Send To Email]

Reference
[Type of image, please refer to option → preference → environment setting → email image format]

V 5.0 62
PACSPLUS PPW

Rearrange Image

(Rearrange images location)

1. Rearrange Image(Manual)
[On menu, Tool → Rearrange Image (Manual)]
2

1) Click the image to be moved. (blue border will become green ①)


2) Move your mouse where to be moved, (the image border will be red②)
3) Click gray part of red border image(③) and move.

Caution To release rearrange function, click Arrow(ESC key)

2. Rearrange Image
[On menu, Tool → Rearrange Image (R)]

Rearrange type: Image date/time/image number.


Sort type : Ascending, descending.
Rearrange Image Series
Series rearrange manually by thumbnail window.(1 series unit)
1) Open image.
2) Move mouse into the bottom of image display window, and thumbnail window will be displayed.
3) Rearrange series with drag & drop of thumbnail image, it is able to display image
by drag of mouse

V 5.0 63
PACSPLUS PPW

Auto Close

It keeps to open same patient’s exam on the other hand, previous


opened exam will be closed automatically.

① <Menu – View – Auto Close>.

<Figure 1.Close auto close feature and open various images> <Figure 2.Activate auto close feature, open various images>

V 5.0 64
PACSPLUS PPW

Chest CT View

Comparison mode of same image, setup Window Level value and


One image can be displayed with different window level value.

[On menu, Option → Preference → Viewer]



Right/Left, setup Window/Level.

② <menu – View – Chest CT View>.

< Image that applied Chest CT View>

V 5.0 65
PACSPLUS PPW

V. Setting
(Configuration)

V 5.0 66
PACSPLUS PPW

Worklist Configuration

③ ①

④ ⑤ ⑥

① Select Worklist Configuration.


② Window layout on the left side will be modified.
③ Select the layout of exam list.
④ Move to up and down by drag & drop to change display sequence.
⑤ Reset column to very first condition.
⑥ Display searching result by study date as ascending or descending. (Default is descending)
⑦OK.

Reference Basically Worklist Layout is E+OE+R and it is able to change.

E+S+OS : Compose with Exam/Series/Old Series.


E+OE+R : Compose with Exam/Old Exam/Report.
E+OE : Compose with Exam/Old Exam.

V 5.0 67
PACSPLUS PPW

Preference (Option)

[On menu, Option → Preference → click the pertinent tab]


There are network, local DB backup, style, environment, viewer, display, external program, database,
hanging protocol to be selected. It is able to modify setting value with setting function. To avoid error
which occurs error by incorrect setting, please review the documents carefully.

1.Network Setting
(How to setup server of DICOM local communication)

1.Local
DICOM network information of
viewer.
1 1) Type in AE Title, IP, Port.
(Re-run program)
Check location of image.

2 2.Remote
Type in DICOM information into
remote.
①Add.
②Type in AE Title, Host Name, P
ort of remote.
① ③Add.

Reference

< DICOM Communication Time Out >

Setup time to be determined that occurs


error of DICOM communication.

V 5.0 68
PACSPLUS PPW

2. Style Setting

3
4

1. DICOM Information
Designate font of DICOM information.

2. Overlay
Designate thickness font of line, color of text, overlay.

3. Histogram
Designate thickness of line, color of line of histogram.

4. Image
Operating Image: Thickness of line and image boarder of operating image.
Select Image: Thickness of line or color of boarder of selected image.

V 5.0 69
PACSPLUS PPW

3. Environment Setting

1. LUT Auto matching


When image display which was setup, saved LUT value that saved in server will be used automatically.

2. Term of validity of Cache Database image


When user searches viewer image, cache database is in use. When image is searched, cache database
setting option designates cumulative period of cache image. If user setup cumulative period longer, local
drive of PC may be in full.

Memo We recommend to use 2 days for cumulative period of cache database.

3. E-Mail Image Format


Designate image type to e-mail with image attachment.

V 5.0 70
PACSPLUS PPW

5. Viewer Setting

1 2

3
4
Patient
Information
5
6

Series
information

1. Close exam
When image displays, displayed patient image closes automatically.
Always do not save exam: save without displayed image with modification and close.
Always save exam: Save with displayed image with modifications and close.
Confirm saving exam: Save displayed image with modifications and check to close and close.
(A message will be displayed when it will save or not)

V 5.0 71
PACSPLUS PPW

2. Chest CT view W/L


Designate left and right W/L value of chest CT view mode.

3. Confirm Opening Every Exam.


If user checks, before image open from worklist, check patient information and series information to be used.

4. Annotation continue mode.


If user checks for annotation, it is able to use annotation continuously once user selects .

5. W/L ROI window size


If user adjusts W/L, adjust ROI on display.

Memo Please re-execute viewer.

6. Interpolation (How to correct)


If user searches image of modality, designate method of correctness.
Setup CR or Non–CR separately.

No: Not use function of correctness.


Linear: Delicate correctness of image.
Bicubic: Better delicate correctness of image.

7. Auto W/L
Setup middle value of W/L to apply to image automatically.

Reference There are Use pixel medial value and Use pixel mean value,
it may takes longer to display Use pixel meal value

8. Exam list
Setup exam list which shows on display window.

Opened Exam List – Display current opened exam.


Old Exam List – Display previous exam which opens currently of the patient.
Both Exam List – Display all of previous exam and present exam.

V 5.0 72
PACSPLUS PPW

6. Display Setting

3 4

1.Monitor configuration

Resolution: Select monitor resolution which is used for image display.


Screen Font: Setup font to be used.
Viewer Font Size: Setup fond size for exam open display.
Worklist Font Size: Setup font size which is used in worklist.
Display Type: Select either Landscape or portrait.
Landscape : Display toolbar on the left.
Portrait: Display toolbar on the top.

[ Landscape type ] [ Portrait type ]

V 5.0 73
PACSPLUS PPW

2. Color monitor for viewer


When user use monitor environment, option for monitor of image efficiently.

Viewer Font Size: Adjust font size of color monitor.


Display Type: Setup display type of color monitor.
Modality : display image of designated modality on color monitor.

3. Thumbnail Window :

1) Display thumbnail when the mouse moves to the bottom of exam open window.
2) Display image of exam of yellow color that display on thumbnail window.
3) Drag & drop image to be designated image from thumbnail window.
4) When the mouse moves up on thumbnail window, thumbnail window will be disappeared.

Caution When thumbnail window is in use, the speed may be slow

4. Worklist size, save status


Select size of worklist and save.

5. Screen color style


It optimizes for LCD or CRT monitor environment and it is able to setup by each user.

V 5.0 74
PACSPLUS PPW

7. External Program Setting

2
3 ①
4

1. External program
It is able to use with 3D, MS office, HIS such as other company’s program. On this option, it is necessary to appoint
external program to be designated to use.

2. Description : Name of other company’s program or explanation of short-cut.

3. Path : Location of exe file, click <…> button to register the program.

4. Add Option: click <<<> button and display pop-up window then designate other company’s program of argume
nt.(①).

5. Save on electronic chart


For interface with electronic chart, please contact local distributor.
Memo

V 5.0 75
PACSPLUS PPW

l Argument list

Ui user ID Initialize registration factor

Ui user name PPW user ID

Pi patient ID PPW user name

Pn patient name Patient ID of exam to be searched

Ps patient gender Patient name of exam to be searched

An accession number Accession number of exam to be searched

Sd study date Study date of exam to be searched

Mo modality Modality of exam to be searched

De department Department of exam to be searched

Si study instance UID Study instance UID of exam to be searched

Bp examed body part Patient body part of exam to be searched

Ie user file Provides user file for every exam are searching

Is user file Provides user file for every series are searching

Ii user file Provides user file for every image are searching

Ik user file Provides user file for every key image are searching

V 5.0 76
PACSPLUS PPW

8. Hanging Protocol Setting ( CX Version )

3 4

How to setup Hanging protocol

When exam , series, image is displayed, display image as registered order. (It is similar with hanging on vi
ew box) by exam window, series window, image window.(For each modality, use series window, image
window)

Exam Window: Display exam. For example, if a user use hanging protocol, one modality will be configur
ed as one exam window.
Series Window: Display exam with each series.
Image Window: Display image is included on series.

Memo If you use multiple monitors, please contact local distributor.

V 5.0 77
PACSPLUS PPW

1. Display one series of one exam


Select this item to display 1 exam 1 series mode of image and click ok to display series mode.

2. Display one series of two exam

No: For 1 series layout mode, all series mode will not be in use.
Both Exam: For 1 series mode of 2 exams, select all series mode.
Left Exam : Left of exam open window, 1 series mode of image will be displayed.
Right Exam : Right of exam open window, 1 series mode of image will be displayed.

3. Default layout

Layout of setting of Hanging protocol


By modality, designate series layout of image. Display series layout of image which viewer as designated. Open
exam on worklist or drag & drop series as default layout.

1) When user setup hanging protocol by modality, select one from layout list.

2) Series layout change:


a. Click one of the box, window displays on the bottom.
b. Move mouse in window.
c. Drag layout to be selected.

3) Image layout change:


a. Click one of the box, window displays on the bottom.
b. Move mouse in window.
c. Drag layout to be selected.
d. OK.

Layout of left / right exam


Setup left / right exam layout, designate layout of monitor to be displayed.

4. Exam layout
Designate layout by modality.

V 5.0 78
PACSPLUS PPW

Toolbar Configuration

[On menu, Option → Toolbar Configuration]

1 Add icon that registered in toolbar.

1.ToolTab

2. Select option
<>>: Add selected icon
<<>: Delete selected icon
<>>>: Add all icons
<<<>: Delete all icons

Memo

Register toolbar on/off


1) Select tool menu and click on/off.
2) Select icon from menu which registered on program icon external program and drag icon to move.(After select external
program icon, click <>> button)
3) OK.
4) Register icon of external program on toolbar.

How to use registered external program


Method 1) Select external program name from tool of menu.
Method 2) Click registered one from display of toolbar.

V 5.0 79
PACSPLUS PPW

DICOM Network Setting

PPW DICOM Engine must be on for DICOM image transmission.


There’s a short cut icon which the PPW is installed correctly.

Please double click the icon, DICOM engine will be running on window system tray.
Please refer to the figure as follow.

Please double click the icon to check the status of DICOM network.
The window will be displayed as follow.

If the network is not running, please click start to run the service.

Once PPW Dicom Engine is on, network status will be ready automatically.

Listen IP or Port, etc can be configured in PPW.

V 5.0 80
PACSPLUS PPW

VI. Troubleshooting

For preference
Memo
[On menu, click Option → Preference]
* Please check Network, database , style, environment settings.

Memo 1) To setup multi monitor, please setup graphic board.


2) If user wish to see HIS and viewer together, 3 monitors is required.
3) Setup two high resolution monitors(monochrome)for diagnostic workstation.
4) Please contact local distributor for multi monitor setting.

V 5.0 81
PACSPLUS PPW

Monitor Display (reaction time slows)

Close image that already read for efficient use of memory.


Use the viewer with small number of opened images.
[On menu, Option → Preference → Viewer]
1) Check “NO” of CR or None-CR for interpolation checkbox.
2) Delete unnecessary software such as game.
3) Please check setting of viewer for interpolation, cache, etc.
4) Scan anti-virus software periodically.
5) Upgrade cpu, memory or HDD.

V 5.0 82
PACSPLUS PPW

Error Message

1. Listening server error


When IP address of workstation is changed where viewer was installed, listening server error message will
be displayed in DICOM network display. When IP address is designated , the message will be displayed to
logon.
* Check IP address of workstation
[On MS Windows, Start→ designate File Name→run → type in CMD → OK → Type in IPCONFIG]
* When IP address can’t be searched, please contact local distributor.

1) [On menu, Setting → Network].


2) Check listen IP and port.
3) [Start→ Hide]

1) [On menu, Option → Preference → Network]


2) Check IP address and port.
3) OK.

2. Language error
For viewer installation, please select same language of OS. If user select different language of OS, language error
message will be displayed.

3. Insufficient space
Requires minimum of 70M byte HDD for viewer.

V 5.0 83
PACSPLUS PPW

4. DB Connection Error
When DB connection Error message pops up, it is problem of connection to PPS server.
1) Check cable connectivity physically.
2) Check ODBC setting and server IP address of PPS database.

5.File Server Error


It occurs in network connection of PPS server.
1) Try to access to FTP.
2) Do pint test for network connection to PPS file server.
(Start à Run à Type in CMD à Type in PING IP Address of network or server)

6. Local DB Backup
There are two database of Master.mdb and PacsPlus.mdb in PPW.
In conference Folder, there’s a mdb file.

1) When master.mdb broke down


Click PPW to run, the error message will be displayed as follows.

could not find file '(unknown)' driver's SQLSet connectAttr failed


State:S1000,nATIVE:-1811,Origin:[Microsoft]]ODBC Microsoft Access dirver]
State:IM006,Native:0,Origin:[Microsoft][ODBC driver Manager]

It means that system couldn’t read the database info in master.mdb as OptionàPreferenceà
Database tab Database and restore master.mdb
2) PacsPlus.mdb
It contains image of local folder and setting values.
ex) study, series, image ...
- When pacsplus.mdb is broken
The same message will be popped up with error message of master.mdb.
-How to backup local database
(Please refer to Option –Preference of manual)
- How to restore local database(same as the way of restoring master.mdb)
1) When error message pops up, click Cancel
2) Click OK

Caution When local database was not backed up, it is able to attach same mdb file of
different PPW workstation to the path of local database.
Local DB path
C:\MedicalStandard\PACSPLUS\PACSPPLUS Viewer\Database

V 5.0 84
PACSPLUS PPW

7. Not Found Exam

1) When exam is searching on worklist, double click compulsory


<Not Found Exam> error message may pops up.
Double click smoothly the image on exam list.
2) When network error occurs, the error message may pops up. Please check network
status.

8. Image Not Found

When specific image opens from worklist, <Image Not Found


> error message may pops up.

1. For standalone PPW system


1) Check path of image storing.
2) Check C\Medical Standard\PACSPLUS\PACSPLUS Vie
wer\Image where the images is saved..

* How to resolve
1) If there’s any image in the path of C\Medical
Standard\PACSPLUS\PACSPLUS Viewer\Image, please
use batch import.

Refer Please refer to Batch Import

2. Client system of PACSPLUS server (PPS)


1) Before image data of database was not transmitted and try to
open the exam
2) When image transmits from server, for full of database capacity
and new exam can’t be transmitted to server.

* How to resolve
1) Check the availability of disk of PC where PPW is installed.
2) Check FTP service status by DOS command.
3) Save image database and check drive path where the image is sa
ved.

V 5.0 85
PACSPLUS PPW

9. Crash Detected

1) This error message is hard to reproduce, mostly it is network or pc problem and there are various reasons and it is
not defined as a bug.
2) For example, when user use keys of PPW constraint or failure of network connection. And hardware, network
device, problem of computer operation, crash with other software, shortage of memory capacity. This error message
displays unexpectedly and hard to find out the reason.
3) When this message occurs, last edited work may not be saved, and system error or loss of data will not be applied.

How to resolve
1) Primarily, check whether PPW runs correctly. (If the PPW runs correctly, there’s no problem of PPW)
2) When PPW is not running correctly, please check the status of network.
3) Please contact local distributor when user did all the steps and didn’t not work.

V 5.0 86
PACSPLUS PPW

PACSPLUS PPW Software Registration

Thank you for purchasing PACSPLUS viewer software.

Please register your software now for technical support and product updates. We only provide technical supports
and services to registered users only. By returning this card, you confirm that you have read the End-User License
Agreement.
All items and fields on the registration form are required. We cannot proceed incomplete registrations.

Registration Instructions.

Please choose form these methods :

●Register by fax :

Please complete the registration card on the next page and fax it to the number below.
Fax Number : +82-2-2282-6690

●Register Online :

http://medicalstandard.com
Please go to the web site above and complete user information by pressing Agreement button.

Benefits for user registered.

Registered users agreed to the terms of End-User License Agreement receive user benefits. It provides
following benefits :
1. Information of the upgrade of software.
2. on-line technical support to users questions.
3. Notice of new software of the company.

V 5.0 87