Professional Documents
Culture Documents
Unit-1
1. Which one is not an auxiliary Element in Medical Instrument?
a). Calibration signal
b) Control and Feedback
c) Transmission of data to remote location
d) Receive of data from remote location
Answer: d
2. Match the following
i) calibration signal - (1) intensive care monitors
ii) control and feedback - (2) nurses’ stations
iii) Alarms - (3) known voltage
iv) transmission of data to remote locations - (4) insulin delivery
a) i) – (3), ii) – (4), iii) – (1), iv) – (2)
b) i) – (2), ii) – (3), iii) – (4), iv) – (1)
c) i) – (3), ii) – (1), iii) – (4), iv) – (2)
d) i) – (4), ii) – (3), iii) – (1), iv) – (2)
Answer: a
3. What is the function of generating sensor?
a) Produces output from energy in measurement
b) Chang in flow of energy from an external source through the sensor
c) Measure directly interfaces with sensor
d) known relation to desired measurand
Answer: a
4. what is the measurand the instrument designed to isolate?
a) Desired input
b) Interfering inputs
c) Modifying inputs
d) Indirect measured
Answer: a
5. Relation to desired measurand known as
a) Indirect measured
b) Direct measured
c) Desired input
d) Interfering inputs
Answer: a
6. Find the correct statement of Modifying inputs measurements.
c)
d)
Answer: b
15. From given figure which one is low accuracy and low precision of measurement
a)
b)
c)
d)
Answer: d
21. How much skin impedance are reduced for 1 cm2 to apply the frequency of 1 Hz to
1MHz using electrode skin interface
a) 200KΩ to 200Ω
b) 200Ω to 20KΩ
c) 100KΩ to 100Ω
d) 100Ω to 100KΩ
Answer: a
24. Chose the odd one out of, Haff – cell potential polarization?
a) Metal involved
b) Concentration of its ion in solution
c) Temperature
d) Electrolyte
Answer: d
24. Metal dose not connect the subject directly but via electrolyte bridge also called
as____________________
a) Electrode
b) Junction Electrode
c) Liquid Junction Electrode
d) Electrolyte
Answer: c
27. Find odd one out of, given the recording problems
a) Thermal nose
b) Noise at the electrolyte
c) Electric and magnetic field interference
d) Surface artifact
Answer: d
29. when you’re measuring the bio signal. if signal is distortion, then you’re using Ag-AgCl
electrode use shielded the cable to reduce ______________
a) Interference
b) Ohmic noise
c) Movement artifact
d) Distortion
Answer: a
30. Equal Half cell potential and high impedance will minimize the artifacts is known as
_______________
a) Interference
b) Ohmic noise
c) Movement artifact
d) Distortion
Answer: c
31. Which of the following statements is not true for biosensors?
a) Biosensors convert a biological signal into an electrical signal
b) Biosensors are used to determine the concentration of substances and other
parameters of biological interest even where they do not utilize a biological system
directly
c) Biosensors utilize the different biological systems such as enzymes, whole cell
metabolism, ligand binding and the antibody-antigen reaction
d) Biosensor consists of a vessel, or series of vessels, used to perform a desired
conversion by enzymic means
Answer: d
a)
35. If the physical change accompanying the reaction is heat output, the biosensors are
referred to as _________________
a) Potentiometric biosensors
b) Optical biosensors
c) Calorimetric biosensors
d) Amperometric biosensors
Answer: c
36. When the physical change produced in a biosensor is due to the movement of electrons
produced in a redox reaction, the biosensor is referred to as ____________
a) calorimetric biosensor
b) potentiometric biosensor
c) piezo-electric biosensor
d) amperometric biosensor
Answer: d
38. Biosensors measure glucose concentrations between which of the following ranges?
a) 10-1 to 10-2 M
b) 10-2 to 10-4 M
c) 10-1 to 10-4 M
d) 10-1 to 10-7 M
Answer: d
39. Given below is the diagram of biosensor. Identify the unmarked component.
a) Microprocessor
b) Filter
c) Transducer
d) A/D converter
Answer: c
40. Transducers employed in the bulk of enzyme electrodes use which of the following
principles?
a) Amperometric
b) Optical
c) Magnetic
d) Colorimetric
Answer: a
41. Which of these biosensors use the principle of heat released or absorbed by a reaction?
a) Potentiometric biosensor
b) Optical biosensors
c) Piezo-electric biosensors
d) Calorimetric biosensors
Answer: d
42. For constructing the glucose sensor, which of the following is used as a gel?
a) Urea
b) Urease
c) Acrylamide
d) Polyacrylamide
Answer: d
Unit-II
4. ECG identified by the PR interval tends to become longer with every succeeding ECG
complex until there is a P wave not followed by a QRS is observed in
a) Third-Degree Atrioventricular Block
b) Second-Degree Atrioventricular Block, Type II
c) Second-Degree Atrioventricular Block, Type I
d) First-Degree Atrioventricular Block, Type II
Answer: c
5. P wave indicates
a) Depolarization of right ventricle
b) Depolarization of left ventricle
c) Depolarization of both atria
d) Atria to ventricular conduction time
Answer: c
6. Ventricular muscle depolarization is indicated by
a) PR interval
b) P wave
c) U wave
d) The QRS complex
Answer: d
7. The characteristics – slurring of the initial QRS deflection, shortened PR interval, and
prolonged QRS duration are of this condition
a) Atrial tachycardia
b) Left bundle branch block
c) WPW (Wolff-Parkinson-White) syndrome
d) Myocardial ischemia
Answer: c
11. In which of these conditions can widen QRS and Tall-tented T waves be observed?
a) Hyponatremia
b) Hyperkalemia
c) Hyperglycemia
d) Hyperphosphatemia
Answer: b
13. Which of the four limbs acts like a ground and thus is not a part of the ECG test?
a) Right Arm
b) Left Arm
c) Right Leg
d) Left Leg
Answer: c
14. The most common placement of the leads from the Einthoven’s triangle is __________
a) Positive Left Arm and Negative Right Arm
b) Positive Left Leg and Negative Right Arm
c) Positive Left Leg and Negative Left Arm
d) Positive Right Arm and Negative Left Leg
Answer: b
15. The amplitudes of the segments of the wave are ____________
16. In the standard 12-leads ECG, how many leads are unipolar?
a) 3
b) 6
c) 9
d) 12
Answer: c
17. According to the international 10/20 system to measure EEG, even number denotes
which side of the brain?
a) left
b) top
c) bottom
d) right
Answer: d
23. When intramuscular EMG is required to look into the electrical activities of deeper or
overlaid muscles, ________ electrodes are used.
a) plate shape electrodes
b) surface electrodes
c) thin thread electrodes
d) fine wire electrodes
Answer: d
24. In voluntary contraction of the skeletal muscles, the muscle potential ranges from
________
a) 50 uV – 5 mV
b) 50 mV – 5 V
c) 0.05 uV – 2 mV
d) 50 mV – 500 mV
Answer: a
26. Which electrode can be used to pick up signals from individual fibers of muscle tissues?
a) biopolar needle electrode
b) concentric core needle electrode
c) multi-element needle electrode
d) monopolar needle electrode
Answer: c
27. he contraction of the skeletal muscles results in the generation of action potential in the
individual muscle fibers. Record of this action potential is called ______________
a) ECG
b) EMG
c) EEG
d) EKG
Answer: b
29. Common mode rejection ratio value of Bio amplifier should be grater then
____________ to reduce the interference from common mode signal
a) 75 dB
b) 78 dB
c) 80 dB
d) 82 dB
Answer: c
31. What is the signal bandwidth of Capacitive Isolation type Bio amplifier?
a) 68 kHz
b) 70kHz
c) 72kHz
d) 74kHz
Answer: b
32. Which one is not a false statement of Capacitive Isolation type Bio amplifier?
a) Ripple noise are not removed
b) It avoids device noise, radiated noise and conducted noise.
c) High immunity to magnetic noise.
d) High gain stability and linearity.
Answer: a
35. Chopper stabilized dc amplifiers are complex amplifiers having ________ amplifiers
incorporated in the module.
a) 1
b) 2
c) 3
d) 4
Answer: c
36. Which of the following amplifier is employed with resistive transducers which require
an external source of excitation?
a) differential amplifier
b) ac coupled amplifier
c) carrier amplifier
d) dc bridge amplifier
Answer: d
38. Which amplifier will reject any common mode signal that appears simultaneously at
both amplifier input terminal and amplifies only the voltage difference that appears
across its input terminals?
a) ac coupled amplifiers
b) differential amplifiers
c) carrier amplifiers
d) dc amplifiers
Answer: b