DIAGNOSTIC ULTRASOUND SYSTEM

Product Data
No. MPDUS0073EAB

Aplio Platinum Series


INTRODUCTION
AplioTM 300 is a high-performance system that is based on
the same platform as the other systems in Canon Medical
Systems' updated flagship Aplio series.
Aplio 300 achieves high sensitivity and high quality by using
state-of-the-art digital technology in the T/R section, which
is the core of a diagnostic ultrasound system.
Aplio 300 features advanced algorithms that run on circuits
that incorporate the latest circuit technology, semiconductor
technology, and surface-mount technology.
Aplio 300 is designed to support a full range of applications
and can be used as a general-purpose system or a special-
ized system, depending on the installed software.

Full-digital ultrasound beam transmission and


reception
Aplio 300 employs full-digital transmission and reception cir-
cuits. The higher definition ultrasound beams and data pro-
cessing technology available with full-digital systems allow
higher sensitivity and image quality to be achieved simulta-
neously. 19-inch monitor model

Enhanced diagnostic capabilities


The spatial resolution, contrast resolution, and temporal res-
olution have been improved through new technologies,
resulting in enhanced diagnostic capabilities.

Transducers supporting a wide range of


frequencies
Echoes over a wide range of frequencies can be
obtained using a single transducer, allowing the 21.5-inch Wide monitor model
optimal sensitivity and quality to be achieved for each
region examined. This function permits a single
transducer to be utilized for a wide range of Ergonomics
applications, greatly improving the throughput and Aplio 300 employs a non-interlace high-definition
price-to-performance ratio. monitor with excellent viewability. This feature
decreases operator fatigue in long examinations.
Intelligent panel and software The ergonomic design of the system ensures
The intelligent panel and software facilitate operation and comfortable and efficient examinations for operators,
contribute to a higher throughput. physicians, and patients.

Upgradability Operability
Aplio 300 can be periodically upgraded to the latest System operability is optimized for the overall clinical
version. workflow in hospitals.
SYSTEM MATRIX OF TUS-A300
Unit Model name Remarks
TUS-A300 19-inch or 21.5-inch LCD monitor, DVD/CD drive, ApliPureTM, Quick Scan, TDI, QSP, HPRF, and DICOM Media
Main unit
Aplio 300 Storage are included.
<Options for main unit>
CW unit UICW-A500A For cardiovascular examinations, add Continuous Doppler capability to sector and pencil transducers.
For cardiovascular examinations (for regions other than the USA)
UJUR-A500A
(ECG (Electrocardiogram), Respiration, Auxiliary input (Aux.), ECG gating, Heart rate)
Reference signal unit
For cardiovascular examinations (for the USA)
UJUR-A501A
(ECG (Electrocardiogram), Respiration, Auxiliary input (Aux.), ECG gating, Heart rate)
Reference signal sensor unit UJUR-772A PCG and Pulse sensor.
Rack for mounting a DVD recorder and/or a color printer on two levels. This kit cannot be mounted togeth-
Mounting kit for peripheral UZRI-A500A er with the UZPH-A300A (transducer holder on the right side of the operating panel).
units Not necessary when UIFR-A500A is used.
UZRI-A501A Rack for mounting a B/W printer
Footswitch UZFS-A500A Switch used for freezing, printing, and some other operations by foot.
Gel warmer UZGW-007A This unit warms the ultrasound gel to a suitable temperature.
UAEH-770A TEE transducer hanger for the PET-510MA, PET-510MB, and PET-508MA.
M-TEE hanger
UAEH-002A TEE transducer hanger for the PET-511BTM, PET-512MA, PET-512MC and PET-512MD.
This unit is required for using the 4D transducer or the motor-driven TEE transducer
4D unit UIMV-A500A
(PET-511BTM/PET-512MC/PET-512MD).
Battery kit UEBT-A500A For standby mode.
Adds a transducer holder on the right side of the operating panel. Dedicated option for Aplio300.
Transducer Holder (Right) kit UZPH-A300A This kit cannot be mounted together with the UZRI-A500A (rack for mounting a color printer and a video
recorder).
Adds three transducer holders on the left side of the operating panel. Dedicated option for Aplio300 plati-
UZPH-A301A
num series. (for V1.0 to V4.0)
Transducer Holder (Left) kit
Adds three transducer holders on the left side of the operating panel. Dedicated option for Aplio300 plati-
UZPH-A302A
num series. (for V5.0 or later)
Transducer cable hanger kit UZMK-A500A Long hanger on which the transducer cable is hooked.
CV kit UACV-A500A This kit consists of preset data suitable for cardiovascular examinations, a CV sticker, and a startup screen.
HV power kit UIHV-A400A For the PVL-715RST.
This kit consists of the magnetic sensor required for Smart Fusion and the kit that mounts the magnetic
sensor to the system main unit and to the transducer PVT-375BT or PVT-375SC. The USFN-A500A is required
Mounting kit for Fusion unit UIFR-A500A separately. Mounting kit for Fusion sensor is required separately. Either a DVD recorder or a color printer
can be installed. It is also possible to install both. In this case, however, there may be some interference with
vertical movement of the operating panel.
Sensor kit for Fusion unit UIFR-A501A Magnetic sensor for the Smart Fusion function.
This pole cart allows the magnetic field transmitter included in the system main unit to be positioned
Fusion pole cart UZWT-A500A
independently.
UAFS-001A For PVT-382BT The UIFR-A500A is required separately.
UAFS-002A For PVT-350BTP The UIFR-A500A is required separately.
UAFS-003A For PVT-781VT The UIFR-A500A is required separately.
Mounting kit for fusion sensor UAFS-004A For PLT-1005BT The UIFR-A500A is required separately.
UAFS-005A For PVL-715RST The UIFR-A500A is required separately.
UAFS-006A For PVT-781VTE The UIFR-A500A is required separately.
UAFS-007A For PVT-475BT The UIFR-A500A is required separately.
Option kit for connecting to a DICOM network via a wireless LAN instead of the previously used LAN cable
Wireless LAN kit UIWL-A500A USDI-A500A is required. Complied with the Radio Law of Japan and applicable laws and regulations of USA,
EU member states, Iceland, Norway, Liechtenstein, and Switzerland.
USDI-A500A Verification, Storage, Print, Storage Commitment, MULTI FRAME (Network Transfer), MWM (Modality Worklist
DICOM kit
USDI-A500A/EL Management), Query/Retrieve, MPPS (Modality Performed Procedure Step), Structured Reporting.
USSE-A500A Adds the cardiac stress examination function to the system.
Stress Echo kit
USSE-A500A/EL The UJUR-A500A or UJUR-A501A is required (available separately).
USWT-A500A Adds the cardiac wall motion analysis function to the system.
2D Wall Motion Tracking kit
USWT-A500A/EL The UJUR-A500A or UJUR-A501A is required.

2
MPDUS0073EAB

Unit Model name Remarks


USHI-A500A
CHI kit Adds the Contrast Imaging function to the system.
USHI-A500A/EL
USCQ-A501A
CHI-Q kit Adds the Time Curve Analysis (TCA) function to the system. The USHI-A500A is required.
USCQ-A501A/EL
USCQ-A502A
Fitting curve kit Function for calculating characteristic value parameters by Fitting curve. The USCQ-A501A is required.
USCQ-A502A/EL
USPP-A300A
ApliPureTM Plus kit Increase the temporal resolution and enables speckle reduction in imaging with ApliPure+. (Note1)
USPPA300A/EL
USDT-A300A
Differential THI kit Enables Differential THI.
USDT-A300A/EL
USPI-A300A
Precision Imaging kit Adds the Precision Imaging and Tissue Enhancement functions to the system. (Note1)
USPI-A300A/EL
Tissue Specific Optimization USSO-A400A Increases the lateral resolution by adjusting parameters based on the speed of sound in tissues.
kit USSO-A400A/EL For linear and convex transducers.
USST-A500A This kit makes it possible to reconstruct and display dynamic images of the fetal heart using a 4D transducer. The
4D STIC Imaging kit
USST-A500A/EL UIMV-A500A and a 4D transducer are required.
USFN-A500A CT/MRI volume data is loaded, and a CT/MRI planar image and an ultrasound image at the same position
Smart Fusion kit
USFN-A500A/EL are displayed together. The UIFR-A500A is required.
USLM-A500A Image processing technology that makes 3D/4D images of fetuses and anatomical structures appear more realistic.
Luminance kit
USLM-A500A/EL The UIMV-A500A and a 4D transducer are required.
USFR-A500A
Smart 3D kit This kit enables free-hand 3D which is available without 4D transducers.
USFR-A500A/EL
USPA-A500A
Protocol Assistant kit A sequence of operations is registered, and each operation is executed by single switch operation.
USPA-A500A/EL
Superb Micro vascular Imaging USMI-A300A
Visualizes low-velocity blood flow at a high frame rate.
kit USMI-A300A/EL
USTS-A300A
Trapezoid Scan kit Expands the linear field of view. (Note1)
USTS-A300A/EL
USDF-A300A
Dynamic FlowTM kit Adds Advanced Dynamic Flow (ADF, high-resolution flow imaging function) to the system. (Note1)
USDF-A300A/EL
USMP-A500A
MicroPureTM kit This kit enables MicroPure, which supports visualization of small structures.
USMP-A500A/EL
USEL-A301A
Elastography-FLR kit This kit enables Elastography (with FLR measurement) with linear and convex transducers. (*1)
USEL-A301A/EL
USEL-A300A
Elastography kit The kit enables Elastography without FLR measurement. (*2)
USEL-A300A/EL
USFL-A500A
FLR kit This kit adds FLR measurement to USEL-A500A. (*1)
USFL-A500A/EL
USSW-A500A
Shear wave kit This kit allows tissue stiffness to be visualized by generating images that show shear wave propagation.
USSW-A500A/EL
USPV-A500A
Panoramic View kit B/W images can be obtained with a wider field of view by moving the transducer in the lateral direction.
USPV-A500A/EL
USVI-A500A This software calculates the display area and ratio of the Power Doppler image.
Vascularity Index kit
USVI-A500A/EL This kit is applicable to the linear transducers.
USXM-A500A
FLEX-M kit This kit displays an M-mode image for an arbitrarily specified plane on a 2D-mode image.
USXM-A500A/EL
USAN-A500A
Auto NT kit This kit enables Auto-Nuchal Translucency measurement.
USAN-A500A/EL
USPS-A300A
Probe 4port kit Provides a fourth transducer connector.
USPS-A300A/EL
This kit provides software for security management of the system.
Security Management kit USSM-A500A
This kit is already installed in systems shipped with software V6.80 *R103 and V7.00 *R004 or later.
Hard copies of the Operation Manual Applications Volume and Operation Manual Measurements Volume for
regions where the standard configuration includes these manuals only as PDF files on CDs.
Operation manuals UOPM-A300A
(Applicable regions: Japan, Europe, USA, Canada, Australia, New Zealand, Turkey, Indonesia, Philippines, Pakistan,
Egypt, and Costa Rica.)

3
Unit Model name Remarks
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AJ
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A. Only for Japan.
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AA
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A.
Mounting kit for UPS unit UZUI-A502A/MA
Line voltage: 110V~120V, Line frequency: 60Hz region
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AE
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A.
UZUI-A502A/ME
Line voltage: 220V~240V, Line frequency: 50Hz/60Hz region
UZKF-A500A
(French)

UZKG-A500A
(German)

UZKI-A500A
(Italian)
Local Language This kit is intended for changing the key tops of the full keyboard to support specific
Key-Top kit UZKS-A500A languages.
(Spanish)
UZKR-A500A
(Russian)
UZKW-A500A
(Danish, Norwegian,
Swedish, Finnish)

Note: 1. Provided in the standard configuration of systems with V3.0 or later in which the 19-inch and 21.5-inch monitor is used.
2. "/EL" is a supplemental model name indicating options supplied by electronic license, only available in Europe, USA, and Canada.
(*1) Not available in the USA. (*2) Only available in the USA.

4
MPDUS0073EAB

BLOCK CHART SYSTEMS

Main unit
DICOM kit
TUS-A300
USDI-A500A
Reference signal unit
Wireless LAN kit
UJUR-A500A/-A501A
UIWL-A500A
Reference signal sensor unit
UJUR-772A
ApliPure plus kit
USPP-A300A (Note)
Stress Echo
USSE-A500A
Differental THI kit
USDT-A300A
2D Wall Motion Tracking kit
USWT-A500A
Precision imaging kit
USPI-A300A (Note)
CW unit
UICW-A500A Tissue Specific Optimization kit
USSO-A400A
CV kit
UACV-A500A Endocavitary transducer
PVL-715RST
FLEX-M kit HV power kit
USXM-A500A UIHV-A400A

Footswitch 4D unit
UZFS-A500A UIMV-A500A

Transducer cable hanger kit Luminance kit


USLM-A500A
UZMK-A500A

Battery kit 4D STIC Imaging kit


USST-A500A
UEBT-A500A

M-TEE hanger Smart 3D kit


UAEH-770A (PET-510MA/MB, PET-508MA) USFR-A500A
UAEH-002A (PET-511BTM, PET-512MC/512MA/512MD)
Protocol Assistant kit
Gel warmer USPA-A500A
UZGW-007A
SMI kit
CHI kit USMI-A300A
USHI-A500A
Trapezoid scan kit
CHI-Q kit
USTS-A300A (Note)
USCQ-A501A

Fitting Curve kit Dynamic Flow kit


USCQ-A502A USDF-A300A (Note)

MicroPure kit
USMP-A500A

A B

5
B

Smart Fusion kit


USFN-A500A

Fusion Pole Cart


UZWT-A500A

Mounting kit for Fusion unit


UIFR-A500A

PVT-382BT PVT-350BTP PVT-781VT PLT-1005BT PVL-715RST PVT-781VTE PVT-475BT


PVT-375BT
Mounting kit for Mounting kit for Mounting kit for Mounting kit for Mounting kit for Mounting kit for Mounting kit for
PVT-375SC Fusion sensor Fusion sensor Fusion sensor Fusion sensor Fusion sensor Fusion sensor Fusion sensor
UAFS-001A UAFS-002A UAFS-003A UAFS-004A UAFS-005A UAFS-006A UAFS-007A

Sensor kit for Fusion unit


UIFR-A501A

C
Elastography-FLR kit Elastography kit
USEL-A301A USEL-A300A
Mounting kit for
peripheral unit Mounting kit for
FLR kit (DVD recorder and UPS unit
USEL-A500A color printer) UZUI-A502A
UZRI-A500A
UPS unit cannot be
used together with
Shear wave kit Mounting kit for UZRI-A500A and UIFR-A500A.
USSW-A500A peripheral unit These items cannot
(B/W printer) be mounted
Panoramic View kit UZRI-A501A at the same time.
USPV-A500A
Transducer holder
(right) kit
Vascularity Index kit
UZPH-A300A
USVI-A500A

Transducer holder
Auto NT kit
(left) kit
USAN-A500A
UZPH-A301A

Probe 4 port kit


Transducer holder
USPS-A300A (left) kit
UZPH-A302A

Note: Provided in the standard configuration of system with V3.0 or


later in which the 19-inch and 21.5-inch monitor is used.

6
MPDUS0073EAB

TRANSDUCER OPTIONS/OPERATION MODES


Transducer Freq. B-mode Freq. THI-mode Freq. D-THI Freq.
Scan type Range
name (MHz) (MHz) (MHz) (MHz)
PST-25BT Sector 5S1 2.5 4.2 3.0 1.8 4.0 3.0 2.0
PST-30BT Sector 5S2 3.0 4.8 3.4 2.0 4.4 3.6 2.8
PST-50BT Sector 6S3 5.0 6.0 4.2 3.0 6.2 5.4 4.4
PST-65AT Sector 9S4 6.5 8.5 6.2 4.2 9.0 7.0 5.0 9.0 7.0
PVT-350BTP Convex 6CP1 3.5 6.0 4.0 1.9 6.0 5.0 4.0 3.0 5.0
PVT-375BT Convex 6C1 3.5 6.0 4.0 1.9 6.0 5.0 4.0 3.0 5.0
PVT-375MV Convex 4D 6CV1 3.5 6.0 4.0 1.9 6.0 5.0 3.0 5.0
PVT-375SC Convex 6Cs1 3.5 6.0 4.0 1.9 6.0 5.0 4.0 3.0 5.0 3.0
PVT-382BT Convex 6MC1 3.5 5.5 3.7 1.8 5.0 3.8 2.8 5.0
PVT-382MV Convex 4D 6MCV1 3.5 5.5 3.7 1.8 5.0 3.8 2.8 5.0
PVT-475BT Convex 8C1 4.0 6.0 4.0 1.9 6.0 5.0 4.0 3.0 6.0 5.0 4.0
PVT-661VT Convex 9C3 6.0 8.8 7.3 4.7 8.0 7.2 6.0 7.0 6.0
PVT-674BT Convex 10C3 6.0 9.2 6.0 3.6 8.0 7.0 6.0 8.0 6.0 5.0
PVT-675MV Convex 4D 8CV2 6.0 7.2 5.0 2.8 7.0 5.5 4.0 6.0 5.0
PVT-675MVL Conve x4D 9CV2 6.0 7.2 5.0 2.8 8.0 6.0 4.0 6.0 5.0
PVT-681MV Convex 4D 9CV3 6.0 8.8 7.3 4.7 8.0 7.2 6.0 7.0 6.0
PVT-681MVL Convex 4D 11CV3 6.0 9.0 7.0 4.0 10.0 8.0 6.0 11.0 9.0 7.0
PVT-712BT Convex 11MC4 7.0 10.2 7.5 4.2 10.0 8.0 6.0 11.0 9.0 8.0
PVT-745BTF Convex 11CI4 7.0 11.0 8.0 4.0 9.0 7.6 5.8 10.0 8.0
PVT-745BTH Convex 11CI4 7.0 11.0 8.0 4.0 9.0 7.6 5.8 10.0 8.0
PVT-745BTV Convex 11CI4 7.0 11.0 8.0 4.0 9.0 7.6 5.8 10.0
PVT-770RT Convex/Convex 10C5 7.0 10.0 7.3 5.0 9.2 8.4 7.0 10.0 8.0
PVT-781VT Convex 11C3 7.0 9.0 7.0 4.0 10.0 8.0 6.0 11.0 9.0 7.0
PVT-781VTE Convex 11C3 7.0 9.0 7.0 4.0 10.0 8.0 6.0 11.0 9.0 7.0
Convex
PVL-715RST 11CL4 7.5 10.0 7.3 5.0 9.2 8.4 7.0 10.0 8.0
Linear
PLT-308P Linear 6LP3 3.75 5.7 4.2 3.1 5.5 4.4 3.6
PLT-604AT Linear 10L4 6.0 9.2 6.7 4.0 7.6 6.6 5.8 8.0 6.0
PLT-704AT Linear 11L5 7.5 11.0 8.6 5.0 8.4 7.2 6.2 9.0 8.0
PLT-704SBT Linear 11L4 7.5 11.0 8.6 4.8 8.4 7.2 6.2 9.0 8.0
PLT-705BT Linear 11L3 7.0 11.0 8.6 4.8 8.4 7.2 6.2 9.0 8.0
PLT-705BTF Linear 11LI4 7.0 11.0 8.0 4.0 8.4 6.6 5.0 9.0 8.0
PLT-705BTH Linear 11LI4 7.0 11.0 8.0 4.0 8.4 6.6 5.0 9.0 8.0
PLT-805AT Linear 12L5 8.0 12.0 10.0 6.2 9.0 7.6 6.6 9.0 8.0
PLT-1005BT Linear 14L5 10.0 12.0 10.0 7.0 11.0 9.0 7.0 14.0 10.0
PLT-1202BT Linear 17LH7 12.0 14.0 12.0 7.0 14.0 12.0 8.0 17.0 14.0 13.0
PLT-1202S Linear 14L7 12.0 14.0 12.0 7.0 14.0 12.0 8.0 18.0 14.0 13.0
PLT-1204BT Linear 18L7 12.0 14.0 12.0 7.2 14.0 12.0 8.0 18.0 14.0 13.0 *1: Optional software is required.
*2: TE (Tissue Enhancement)
PLT-1204MV Linear 4D 14LV7 12.0 14.0 12.0 7.2 14.0 12.0 8.0 14.0 13.0
*3: Depends on the preset.
PET-508MA Sector, TEE 7S3 5.0 6.5 5.0 3.0 6.6 5.6 4.4 *4: Not available in the USA.
PET-510MA Sector, TEE 7S3 5.0 6.5 5.0 3.0 6.6 5.6 4.4 *5: Not available in the USA and
Canada.
PET-510MB Sector, TEE 7S3 5.0 6.5 5.0 3.0 6.6 5.6 4.4
*6: Only for the USA.
PET-511BTM Sector, TEE 8SM2 5.0 7.5 5.0 2.5 7.0 5.6 3.5 *7: Optional software USPP-A300A is
PET-512MA Sector, TEE 8S2 5.0 7.5 5.0 2.5 7.0 5.6 3.5 required.
*8: Optional unit UIHV-A400A is
PET-512MC Sector, TEE 8SM2 5.0 7.5 5.0 2.5 7.0 5.6 3.5
required.
PET-512MD Sector, TEE 8SM2 5.0 7.5 5.0 2.5 7.0 5.6 3.5 *9: Optional unit UIMV-A500A is
PET-805LA Linear, LAPA 12LI4 8.0 11.2 8.0 4.0 8.4 6.6 5.0 9.0 8.0 required.
*10: Optional unit UICW-A500A is
PC-20M Pencil P2 2.0
required.
PC-50M Pencil P5 5.0

7
Precision Dynamic
Transducer name 2D M PWD CWD*10 CDI THI D-THI*1 ApliPure Boost
Imaging*1/TE Flow*1
PST-25BT       *2 *3
PST-30BT       *2 *3
*2
PST-50BT        *3
*3 *2
PST-65AT         *3
PVT-350BTP         
PVT-375BT         
PVT-375MV         
PVT-375SC         
PVT-382BT         
PVT-382MV         
PVT-475BT         
PVT-661VT         
PVT-674BT          
PVT-675MV         
PVT-675MVL         
PVT-681MV         
PVT-681MVL         
PVT-712BT         
PVT-745BTF         
PVT-745BTH         
PVT-745BTV         
PVT-770RT         
PVT-781VT         
PVT-781VTE         
Convex         
PVL-715RST
Linear         
PLT-308P        
PLT-604AT         
PLT-704AT         
PLT-704SBT          
PLT-705BT         
PLT-705BTF         
PLT-705BTH         
PLT-805AT         
PLT-1005BT          
PLT-1202BT         
PLT-1202S         
*1: Optional software is required.
PLT-1204BT         
*2: TE (Tissue Enhancement)
PLT-1204MV          *3: Depends on the preset.
PET-508MA       *2 *4: Not available in the USA.
*5: Not available in the USA and
PET-510MA       *2
Canada.
PET-510MB       *2 *6: Only for the USA.
PET-511BTM       *2 *7: Optional software USPP-A300A is
required.
PET-512MA       *2
*8: Optional unit UIHV-A400A is
PET-512MC       *2 required.
PET-512MD       *2 *9: Optional unit UIMV-A500A is
required.
PET-805LA         
*10: Optional unit UICW-A500A is
PC-20M  required.
PC-50M 

8
MPDUS0073EAB

TRANSDUCER OPTIONS/OPERATION MODES


CHI
Volume
Transducer name Power TDI Dynamic Contrast 4D*9 STIC*1, 9
2D MFI VRI Color*9
Flow SMI*1
PST-25BT *3   
PST-30BT *3   
PST-50BT *3 
PST-65AT *3  *4 *4
PVT-350BTP     
PVT-375BT       
*4 *4 *4 *4
PVT-375MV        *3
PVT-375SC       
PVT-382BT     
PVT-382MV  *4 *4 *4 *4  
PVT-475BT       
PVT-661VT   *4 *4 *4 *4
PVT-674BT   *4 *4 *4
PVT-675MV    *3
PVT-675MVL    *3
PVT-681MV      
*4 *4 *4
PVT-681MVL       
PVT-712BT 
PVT-745BTF  *4 *4
PVT-745BTH  *4 *4
PVT-745BTV 
PVT-770RT 
PVT-781VT   *4 *4 *4
*4 *4
PVT-781VTE     *4
*4 *4
Convex    
PVL-715RST
Linear 
PLT-308P 
PLT-604AT  *4 *4 *4 *4
*4 *4 *4
PLT-704AT    
PLT-704SBT  *4 *4 *4 *4
PLT-705BT    
PLT-705BTF  *4 *4
PLT-705BTH  *4 *4
*4 *4
PLT-805AT     *4 *4
PLT-1005BT       
PLT-1202BT   *4 *4
PLT-1202S   *4 *4 *1: Optional software is required.
*4 *4 *4 *2: TE (Tissue Enhancement)
PLT-1204BT     
*3: Depends on the preset.
PLT-1204MV  *4 *4   *4: Not available in the USA.
PET-508MA  *5: Not available in the USA and
Canada.
PET-510MA 
*6: Only for the USA.
PET-510MB  *7: Optional software USPP-A300A is
PET-511BTM  required.
PET-512MA  *8: Optional unit UIHV-A400A is
required.
PET-512MC  *9: Optional unit UIMV-A500A is
PET-512MD  required.
PET-805LA    *10: Optional unit UICW-A500A is
required.
PC-20M
PC-50M

9
STIC Smart Micro- Elasto- 2D Shear
Transducer name SMI*1 BEAM*7 Fusion*1 Remarks
Color*1, 9 3D*1 Pure*1 graphy*1 WMT*1 wave*1
PST-25BT 
PST-30BT 
PST-50BT 
PST-65AT 
PVT-350BTP   *5
PVT-375BT     
PVT-375MV *3 *9
PVT-375SC      *11
PVT-382BT  
PVT-382MV *9
PVT-475BT      *11
PVT-661VT 
PVT-674BT   
PVT-675MV *3 *9
*3
PVT-675MVL  *9
PVT-681MV  *9
PVT-681MVL   *9
PVT-712BT 
PVT-745BTF 
PVT-745BTH 
PVT-745BTV 
PVT-770RT
PVT-781VT   *4 
*4
PVT-781VTE    
Convex   
PVL-715RST *8
Linear  
PLT-308P  *5
PLT-604AT  
PLT-704AT   
PLT-704SBT    
PLT-705BT   
PLT-705BTF 
PLT-705BTH 
PLT-805AT    
PLT-1005BT       
PLT-1202BT   
PLT-1202S  
PLT-1204BT     
*1: Optional software is required.
PLT-1204MV *9 *2: TE (Tissue Enhancement)
PET-508MA  *3: Depends on the preset.
*4: Not available in the USA.
PET-510MA  *4
*5: Not available in the USA and
PET-510MB  *6 Canada.
PET-511BTM  *5, 9 *6: Only for the USA.
*7: Optional software USPP-A300A is
PET-512MA  *5 required.
PET-512MC  *9 *8: Optional unit UIHV-A400A is
required.
PET-512MD  *9
*9: Optional unit UIMV-A500A is
PET-805LA required.
PC-20M *10 *10: Optional unit UICW-A500A is
required.
PC-50M *10
*11: Using Single Crystal technology.
10
MPDUS0073EAB

4D OPERATION MODES
The optional unit UIMV-A500A is required for this operation.
4D Live Single Sweep
Transducer Pulse Pulse Pulse Pulse Volume 4D Max sweep range
STIC*1 STIC color*1 Luminance*1
name Fund. Subtract. Subtract. Fund. Subtract. Subtract. color Biopsy (deg)
ON OFF ON OFF
PVT-375MV     *2 *3 *3    75
*2
PVT-382MV         80
PVT-675MV     *2 *3 *3   75
*2 *3 *3
PVT-675MVL          90
PVT-681MV    *2    90
*2
PVT-681MVL        150
PLT-1204MV    *2    30

*1: Optional software is required.


*2: Differential THI
*3: Depends on the preset.

3D OPERATION MODES
The optional software USFR-A500A is required for this operation.
2D CHI*1
Transducer Pulse Pulse Dynamic
CDI Power Contrast SMI*1 Shear wave*1 Remarks
name Fund. Subtract. Subtract. Flow*1 2D
SMI*1
ON OFF
PVT-350BTP   *2     *4
*2
PVT-375BT          
PVT-375SC   *2       
*2
PVT-382BT       
PVT-475BT   *2       
*2 *3 *3
PVT-674BT         
*2
PVT-712BT      
PVT-745BTF   *2    *3
PVT-745BTH   *2    *3
*2
PVT-745BTV      
PLT-308P     
PLT-604AT   *2    *3
PLT-704AT   *2    *3
PLT-704SBT   *2    *3 *3 
*2
PLT-705BT         
PLT-705BTF   *2    *3
PLT-705BTH   *2    *3
PLT-805AT   *2    *3
*2
PLT-1005BT          
*2 *3
PLT-1202BT        
PLT-1202S   *2    *3
PLT-1204BT   *2    *3 *3 

*1: Optional software is required.


*2: Differential THI
*3: Not available in the USA.
*4: Not available in the USA and Canada.

11
SPECIFICATIONS · Line density
· Parallel signal processing
System
– Parallel signal processing is available for up to two direc-
• Scan methods: Linear scan (some transducers
tions.
can perform oblique scanning)
• Scan Angle and Scan Width
Sector scan
– Adjustment of the field width (scan width, scan angle) is
Convex scan
possible.
Trapezoid scan (USTS-A300A is
– Adjustment of beam steering (scanning position) is pos-
required.)
sible.
• Monitor:
– Adjustment of linear beam steering is possible.
High-definition 19-inch LCD monitor
• PAN/EXPAND
– Resolution: 1280 × 1024 (SXGA)
Real-time PAN/EXPAND
– High-definition 21.5-inch LCD monitor
– Scale enlargement/reduction using the encoder is possi-
– Resolution: 1600 × 900 (WXGA++)
ble.
• Presets
– Movement to the desired section using the trackball is
– System preset: 1 type
possible.
– Application preset: 20 types
– The transmission focus is optimized in steps above.
– The specified range of an image can be magnified. (Spot
Compatible Peripheral Devices
Zoom)
• Black-and-white digital printer
• Transmission Focus
– UP-D897 (100 V to 240 V, SONY)
Transmission conditions: A maximum of 8 steps
– UP-D898MD (100 V to 240 V, SONY)
• Transmission Frequency
– P95DW (100 V to 240 V, MITSUBISHI)
Multi frequency
• Color digital printer
– 2D mode
– UP-D25MD (100 V to 240 V, SONY)
5 types (at maximum)
– CP30DW (120 V, 220V to 240 V, MITSUBISHI)
– THI mode (second harmonic)
• DVD video recorder
5 types (at maximum)
– HVO-550MD (100 V to 240 V, NTSC/PAL, SONY)
• Pulser
– DVO-1000MD (100 V to 240 V, NTSC/PAL, SONY)
Dual Pulser
– BD-X201M (100 V to 240 V, NTSC, JVC)
• Reception Conditions
* For regions other than Europe
Continuous focus
– BD-X201ME (100 V to 240 V, PAL, JVC)
• GAIN
* For Europe
– The display brightness for 2D can be changed.
• USB flash drive
(also available when the image is frozen)
• Barcode reader
– The display brightness for 2D and M can be changed
simultaneously.
2D mode (B mode)
• STC
• Viewing Depth
8-step slide controls (common for 2D and M)
The viewing depth depends on the transducer used.
• Acoustic Output
– Convex
Adjustment is possible to 100%.
· Minimum depth: 2 cm
· Maximum depth: 40 cm • Adjusting the 2D Image Quality
– Linear – Dynamic range (also available when the image is frozen)
· Minimum depth: 1 cm – Edge enhancement
· Maximum depth: 14 cm – Time-smoothing (persistence)
– Sector – Map (also available when the image is frozen)
· Minimum depth: 1 cm – Auto gain control
· Maximum depth: 28 cm – Lateral gain
• Line density – Gamma (also available when the image is frozen)
– The line density differs depending on the transducer – P-SEL (frame rate)
used. – ApliPure
– The line density can be changed. – Precision (USPI-A300A is required.)
• Ultrasound Frame Rate – Tissue Enhancement
– The ultrasound frame rate can be adjusted by using the – Boost
following in combination.
12
MPDUS0073EAB

• 2D Map • M Map
– The grayscale pattern can be changed and virtual color M image virtual color setting is possible. The setting can
setting for the 2D image is possible. be changed even after the image has been frozen.
– Settings can also be changed when the image is frozen. • THI (Tissue Harmonic Imaging)
• 2D Quick Scan 2D mode and THI mode are linked, and M images can be
The gain and STC can be adjusted automatically. displayed in THI mode.
• THI (Tissue Harmonic Imaging) – Pulse subtraction method
THI signal processing methods – Filtering method
– Pulse subtractionTM method – Differential method (USDT-A300A is required.)
– Filtering method • M Mark
– Differential method (USDT-A300A is required.) – The M cursor can be displayed on 2D or C images.
– The M cursor displayed position can be adjusted.
• Display Orientation
• Flex-M (USXM-A500A is required.)
– Top/bottom reversal is possible.
Any desired plane can be set on the 2D-mode image and
– Left/right reversal is possible.
the M-mode image for the set plane can be reconstructed.
• ApliPureTM
– ApliPure Doppler (Spectrum Doppler)
This function reduces ultrasound wave interference with- • Doppler Mode
in tissues, which appear as speckle patterns or speckle – PWD (pulsed-wave Doppler)
noise on 2D images. – HPRF PWD (can be switched to HPRF mode by preset-
– ApliPure+ (USPP-A300A is required.) ting)
This function can display the boundaries between tissues – CWD (continuous-wave Doppler)
more clearly and reduce speckle noise and acoustic (UICW-A500A is required.)
shadows. – Pencil CWD (pencil-type transducer)
• MicroPureTM (USMP-A500A is required.) (UICW-A500A is required.)
– Small structures can be extracted by performing filtering • Doppler Pulse Repetition Frequency (PRF)
for 2D-mode images. – PWD: 0.3 kHz to 52.1 kHz
– Visualization of very small calcifications and other – CWD: 1.4 kHz to 52.1 kHz
extremely small lesions can be improved. • Doppler Scan
– 2D/D simultaneous scan
• Precision Imaging (USPI-A300A is required.)
– D only scan
Structures in 2D-mode images can be displayed more
• Doppler Sampling Volume
clearly and the background can be displayed more
The Doppler range gate width can be changed.
smoothly.
• Doppler Sampling Shift
• TSO (Tissue Specific Optimization) 0 cm to the maximum depth
(USSO-A400A is required.) • Doppler Filter
– Reception focus compensation can be performed. The Doppler filter cutoff can be changed.
– Automatic reception focus compensation can be • Doppler Gain
performed. (Auto TSO) The display brightness for Doppler can be changed.
• BEAM (Biopsy Enhancement Auto Mode) • Doppler Quick Scan
– Display of the needle can be enhanced in the image. The Doppler scale and baseline shift can be adjusted auto-
– The enhancement level can be adjusted. matically.
* The optional USPP-A300A is required in order to use • Doppler Frequency Analysis and Image Processing
this mode. – Method: FFT
– No. of data items: 256 (maximum)
M mode • Indication of Doppler Spectrum Direction
• M Transmission Frequency Reverse display of the velocity spectrum is possible.
Multi frequency: 5 types (at maximum) • Doppler Baseline Shift (Zero Shift)
• M Sweep Speed – The velocity baseline of Doppler images can be shifted.
The Sweep Speed can be changed in M mode. – The baseline shift setting can also be adjusted when
• M Gain images that were frozen are displayed.
M gain can be corrected for 2D gain. • Doppler Audio
• M Image Processing Parameters Stereo output (blood flow toward and away from the trans-
– M dynamic range ducer)
– M edge enhancement • Doppler Map
– M auto gain control The brightness conversion table and the virtual color for
– M gamma (can be changed even after the image has Doppler images can be set.
been frozen). • Display of Doppler Scale
2 types (velocity, Doppler shift frequency)
13
• Doppler Focus • C Baseline (Zero Shift)
Automatically follows the sample position. – The velocity baseline of color Doppler images can be
• Doppler Angle Mark shifted.
This mark is displayed for measuring the angle between – The baseline shift setting can also be adjusted when
the direction of the velocity and the direction of the ultra- images that were frozen or images in the image memory
sound beam. are displayed.
• Doppler Oblique Scan (PWD Steering) • Reverse C Display
– Oblique scans are possible using a specific linear trans- – Coloring is reversed.
ducer. – Changes can also be made when the image is frozen.
– Auto Invert function • Black and White/Color Balance
• Doppler Multifrequency – By comparing the color Doppler images and B/W imag-
The PWD transmission frequency can be changed. es, color weighting to B/W can be set.
• Doppler Sweep Speed – Changes can also be made when the image is frozen.
The Sweep Speed can be changed in Doppler mode. • C Parallel Processing
• Doppler Display Dynamic Range Beam reception from up to four directions for each trans-
The display dynamic range of the Doppler image can be mission beam.
changed. • C Gain
• Doppler Auto Trace (measurement performed after freez- The display brightness of color Doppler images can be
ing the image) changed.
Measurement of peak velocity and mean velocity is possi- • C Multifrequency
ble by automatic velocity tracing. The transmission frequency for color Doppler image acqui-
The following Doppler waveform trace is possible. sition can be changed.
– Trace style: Waveform Peak, Mean, Peak + • C Line density
Mean The color Doppler image line density can be changed.
– Trace area specified: Forward, Reverse, Full, Auto • C ROI (Region of Interest)
– Measurement item: Max, Min, Mean, PI, RI, etc. Position, size, and steering adjustment is possible for color
Doppler ROIs.
Color Doppler • C Transmit Focus
Color Doppler 2D Mode (BCD/CDF Mode) Automatically follows the color Doppler ROI position.
• Display mode • C Filter
– CDI mode – The C filter can be changed.
· Flow velocity – FIO filter
· Flow velocity/variance • Variance Curve
· Power – The display of the color variance component can be
– Power Angio mode adjusted.
Direction display is possible.
– TDI mode Color Doppler M mode (MDF Mode)
– Dynamic Flow mode (USDF-A300A is required.) • Display mode
Direction display is possible. – MCDI mode
– TwinViewTM · Velocity display
Simultaneous dual-screen display with 2D mode is avail- · Velocity/variance display
able. · Power display
– Superb Micro vascular Imaging mode – M-TDI mode
(USMI-A300A is required.) • M Color Doppler Map (CDI MAP)
· Clutter suppressed Color Doppler map can be selected for each mode.
· Blood flow enhanced • M Color Doppler Velocity Range Selection (C Scale)
• C Map The velocity range can be selected.
– C map can be selected for each color Doppler mode. • M Color Doppler Baseline (C Baseline)
– Changes can also be made when the image is frozen. – The zero-velocity line on the M Color Doppler image can
• C Scale (Switching the Velocity Range) be shifted.
The velocity range can be changed. – The baseline shift setting can also be adjusted when
• C Time-Smoothing (Persistence) images that were frozen are displayed or when the
The result of temporal correlation processing between the image in the image memory is played back.
previous image and current image can be displayed. • Color Reverse Display
• C Frame Interpolation – The colors can be reversed.
Automatic setting according to the transducer information – Changes can also be made when the image is frozen.
and selected study conditions.
14
MPDUS0073EAB

• Black and White/Color Balance – The following functions can be used.


– Color weighting for B/W images can be set by compar- · Volume Color
ing the M Color Doppler images and B/W images. · Multi View
– Changes can also be made when the image is frozen. · VolPure
• Color Gain · Volume View
The display brightness of the M Color Doppler image can · STIC/STIC Color (USST-A500A is required.)
be changed. · 4D Biopsy
• M Color Doppler Multi-Frequency · Luminance (USLM-A500A is required.)
Doppler transmission frequency can be selected in M – The following measurements can be performed.
Color Doppler image acquisition. · MPR
• M Color Doppler Filter · Volume
– Doppler low-cut filter can be changed. • Stress Echo (USSE-A500A is required.)
– FIO filter function Exercise and pharmacological stress echo examinations
can be performed.
Reference Signals – Data acquisition mode
(UJUR-A500A or UJUR-A501A is required.) This mode is intended for selecting and creating protocols.
• Type – Review mode
– Electrocardiogram (ECG) This mode provides playback function/data output func-
· Lead I is the standard connection. tion/Wall Motion Scoring (WMS) function.
· External input is possible. • Panoramic View (USPV-A500A is required.)
· Top/bottom inversion is possible. – A continuous image can be acquired by moving the
· Lead switching transducer horizontally on the body surface.
· Pacemaker – Measurement using Panoramic View can be performed.
· INST • Elastography (USEL-A301A or USEL-A300A is required.)
– Phonocardiogram (PCG) (UJUR-772A is required.) – Tissue stiffness can be visualized based on the changes
· PCG microphone: Acceleration type in velocity resulting from physical compression and
· Filter: Switching is possible. decompression of the target region.
– Pulse (UJUR-772A is required.) – FLR measurement can be performed to calculate the
Pulse transducer: Air conduction type strain within the set ROI. (Not available in the USA.)
– Respiration • VI (Vascularity Index) (USVI-A500A is required.)
Impedance method using the reference signal cable The number of pixels in the Power/ SMI (*) image and
– Auxiliary within the ROI, the area of the Power/ SMI (*) image and of
The connected device must comply with IEC 60601-1. the ROI, and the ratio of the number of pixels in the Power/
• Heart Rate SMI (*) image to the number of pixels within the ROI can
The heart mark blinks in synchronization with the heart be displayed for an image acquired in Power/ SMI (*)
beat detected by the ECG. mode.
The heart rate is displayed. • Fusion (Smart Fusion) (USFN-A500A etc. is required.)
Synchronization of ultrasound scanning with CT/MRI imag-
Other Diagnostic Images (option)
es display adjusted according to the examination position
• CHI (Contrast Harmonic Imaging) (USHI-A500A is required.)
determined using a magnetic sensor attached to the trans-
The second-harmonic wave signals from the microbubbles
ducer can be performed.
in the contrast medium can be effectively visualized.
• Protocol Assistant (USPA-A500A is required.)
– The following image modes can be selected.
A series of operations (a protocol) that has been created
· PS (Pulse Subtraction) -Low, PS-Low2
for the intended examination can be executed automati-
· VRI (Vascular Recognition Imaging)
cally.
· Fundamental
Protocols can be created and edited.
· ADF (CHI)
• Shear wave (USSW-A500A is required.)
· SMI (Superb Micro vascular Imaging) (*)
Images representing the speed of propagation of tissue
– The following functions can be selected.
displacement (shear wave speed) can be visualized (shear
· 2D TwinView
wave scan) by locally displacing tissues by transmitting a
· Micro Flow Imaging (MFI)
burst wave with high acoustic pressure.
· Image Stabilizer
– The following Shear wave display modes are available.
· MI Constant function
· Speed: shear wave speed display (m/s)
• 4D (UIMV-A500A is required.)
· Elasticity: elasticity display (kPa)
Three-dimensional image data (volume data) can be gen-
· Propagation: propagation display
erated and displayed by using image data acquired for
– Shear wave measurement can be performed.
three-dimensional image reconstruction.

15
• Smart 3D (USFR-A500A is required.) – Thumbnail area
3D image can be generated from the 2D image and any · Image data acquired during the current examination is
input volume shape. displayed.
· Information from a previous examination of the patient
(*) USMI-A300A is required. currently being examined is displayed.
– Information message display area
Display-Related Features An operation guide and other messages are displayed.
• Display Method – Status area
– Images on the main unit: 60 Hz non-interlaced display The following system statuses can be displayed.
– Images from external · Saving dynamic/still image
playback devices: 60 Hz non-interlaced display · Used space on HDD
• Monitor Display/Character Display · DICOM printer status/peripheral device status
– ID area · DVD/CD write status
· Patient ID · Trackball assignment status
· Patient name • Annotation
· Hospital name – Manual input using the keyboard is possible.
· Date: Selected from among the formats – Auto annotation (previously specified text) is possible.
shown below. • Pictograms
YYYY/MM/DD Body icons and transducer mark
MM/DD/YYYY • Biopsy Guide Mark
DD/MM/YYYY
Biopsy guide mark display is possible.
YYYY: Western calendar year
• Language
MM: Month
DD: Day The following languages are supported for the display of
· Time: Selected from among the formats some screens and keyboard entry.
shown below. Supported Screen Keyboard
hh:mm:ss: AM (PM) language display entry
hh:mm:ss: 24-hour representation
hh: Hour English  
mm: Minute

ss: Second English (UK) 
(same as"English")
· VCR counter
· Age German  
· Sex French  
· Heart rate display (heart-shaped mark/heart rate)
· Name of the Imaging Preset Italian  
· Name of the operator Spanish  
· Gestational age Danish  
– Acoustic power display area
Dutch  
· Acoustic power value (%)
· TI value Norwegian  
– Auto data Swedish  
Frame rate
Finnish  
Acoustic power index = MI indication
Transducer frequency Portuguese  
Depth Icelandic  
Dynamic range
Russian  
GAIN
CDI filter : Applicable : Not applicable
PRF
Doppler filter
Doppler angle
Doppler gate size

16
MPDUS0073EAB

Measurement Functions – OB (obstetrics) measurements


Basic Measurement Functions · The data for determining fetal growth based on the
• 2D-mode measurements measured fetal size is displayed.
– Distance · The list of measured data or a graph of the measured
· Distance value development (fetal growth conditions) is displayed.
· Trace Length · Week function (gestational age)
· Mean-IMT · Measurement data saving is possible.
– Area · Maximum quadruplet can be measured.
– Angle · Doppler mode measurement is possible.
· Angle · Auto NT measurement (USAN-A500A is required.)
· Joint – Vascular measurement
– Volume · CCA (Common Carotid Artery) measurement
– Stenosis ratio · ECA (External Carotid Artery) measurement
· %Stenosis (Distance) · ICA (Internal Carotid Artery) measurement
· %Stenosis (Area) · Vert A (Vertebral Artery) measurement
• 4D-mode measurements (UIMV-A500A is required.) · Subclav A (Subclavian Artery) measurement
– Distance · Auto-IMT measurement
– Area and Circumference · IMT-C10 measurement
– Angle
Advanced Measurement Functions
– CTAR
• 2D wall motion tracking (USWT-A500A is required) Wall
– Volume
motion can be analyzed by semi-automatically extracting
• M-mode measurements
the left ventricular myocardium from the image data
– Slope
acquired by the system and displaying it for the evaluation
– Distance
of myocardial motion.
– Time
– Wall motion tracking in 2D dynamic images
– Heart rate
– Wall motion information display
• PW/CW Doppler measurements
– Polar map display
– Velocity
– Local/whole myocardial wall motion parameter curve
– Acceleration
display
– Time
– Parameter setting display of various parameters
– Heart rate
– Analytical data output to a file
– PI
• TCA (Time Curve Analysis) (USHI-A500A and USCQ-A501A
– RI
– S/D are required.)
– Flow volume Quantitative analysis can be performed using a graph
– Doppler trace showing the changes in intensity over time of images
acquired in CHI mode.
Application Measurement Functions – Generation of a graph from the changes in intensity over
• Cardiac measurements time
– 2D-mode measurements – Motion Tracking function
· LV (left ventricular function) measurements – Function for calculating characteristic value parameters
· LA (left atrial volume) measurements by Curve Fitting (USCQ-A502A is required.)
· AV (aortic valve) measurements – Analytical data output to a file
· MV (mitral valve) measurements • MPI (Myocardial Performance Index)
· PV (pulmonary valve) measurements (Not available in the USA and Canada.)
· LV MASS measurements The MPI value can be calculated from the time change
– M-mode measurements curve in TDI.
· LV (left ventricular function) measurements – Display of the changes in velocity over time in TDI as a
· AV (aortic valve) measurements graph
· MV (mitral valve) measurements – Function for calculating the MPI value
– Doppler measurements – Analytical data output to a file
· Trans-Aortic valve flow measurement
Measurement of Stored Image Data
· Trans-Mitral valve flow measurement
The following measurements can be performed for the
· Trans-Pulmonary vein flow measurement
DICOM data stored in HDD.
· Trans-Tricuspid valve flow measurement
• Basic measurement
· Trans-Pulmonary valve flow measurement
• Application measurement
· Blood flow waveform auto measurements
· Coronary measurements
· PISA measurements
17
Report Function DICOM *
• Worksheet functions • DICOM Media Storage
– The measurement and calculation items can be dis- • DICOM Verification
played for each application measurement. • DICOM Storage
– Data editing is possible (except for some items). • DICOM Print
– Display of the following values can be set to ON or OFF. • DICOM Storage Commitment
Mean value, latest value, maximum value, minimum • DICOM MULTI FRAME (Network Transfer)
value. • DICOM MWM (Modality Worklist Management)
– Trend graphs can be displayed (OB measurement work- • DICOM Query/Retrieve
sheet). • DICOM MPPS (Modality Performed Procedure Step)
– Comment entry is possible. • DICOM Structured Reporting
• Report function (On Board Report) * USDI-A500A is required for all items other than DICOM
– Reports can be created on the system. Media Storage, which is provided in the standard config-
– The created reports can be printed. uration.
– The created reports can be output as PDF files.
– The report template can be edited. Signal I/O
– Comment entry is possible. • Transducer Connectors
– Transducer connectors: 3
Cine Memory (large-capacity image memory) – Pencil transducer connector: 1
• Memory Capacity • VCR Input/Output Signals
960 MB – S-VHS output
• Record/Playback Mode – S-VHS input
– Loop playback is possible. – Audio output: L, R
– Frame advance playback is possible. – Audio input: L, R
– Cine playback is possible in Doppler or M mode. – DVD Recorder control signal: RS-232C
– Live images can be recorded. (Clips) • External Video Output Signals
– S-Video (S-VHS)
Video Recording – RGB SXGA signals
The following DVD remote control operations are possible: – RGB composite signals (NTSC/PAL timing)
Record, stop, play, fast-forward, rewind, forward search, * RGB SXGA signals and RGB composite signals cannot
reverse search, and freeze (pause). be output simultaneously.
• Printer Signal
Recording Function USB: 2 ch
• Printers (option) • RS-232C
– Black-and-white printer: USB connection For fusion transmitter (option)
– Color printer: USB connection • External USB
• Video Recording Units (Option) 5 USB ports (2 on the rear of the main unit, 2 on the side of
– DVD Video the main unit, and 1 on the side of the panel)
• Electronic Filing • Ethernet
– Hard disk drive 10BASE-T/100BASE-TX/Gigabit Ethernet: 1 ch
– DVD/CD drive • SATA
– USB flash drive – For connecting the built-in HDD
– Network: DICOM (option) – For DVD drive
• Footswitch
Image Format to Export (UZFS-A500A is required.)
– Still: BMP/ JPEG 2-switch footswitch
– Movie: MPEG-4/ cinepack/ WMV9 • UPS Mode
(UZUI-A502A is required.)
Network – This function enables the system to be shut down safely
• Ethernet: 10BASE-T/100BASE-TX/Gigabit if there is a power failure.
Ethernet – The system can be operated in battery mode if the
• Network client system power supply from the outlet is interrupted due to power
failure etc. (Only for the USA and Canada)

18
MPDUS0073EAB

Operating Conditions Conformance Standards


Power Supply Requirements • Canada
• Line voltage – General: CAN/CSA-C22.2 No. 601.1-M90
– Japan 100 VAC ±10% – Collateral: CAN/CSA-C22.2 No. 60601-1-1-02 (R06)
– USA, Canada 120 VAC±10% CAN/CSA-C22.2 No. 60601-1-4-02 (R2006)
– Europe 220 to 240 VAC ±10% – Particular: CAN/CSA-C22.2 No. 60601-2-37-03 + A1
– Other 1 110 to 120 VAC ±10% (R2007)
– Other 2 220 to 240 VAC ±10% • EU and other regions requiring compliance with European
• Line frequency Directive 93/42/EEC and subsequent amendments
50/60 Hz ± 1 Hz – General: EN 60601-1:2006 + A1:2013
• Power capacity – Collateral: EN 60601-1-2:2015
– Japan 1500 VA – Particular: EN 60601-2-37:2008 + A1:2015
– USA, Canada 1440 VA • USA
– Europe 1500 VA – General: UL 60601-1:2003 R4.06
– Other 1 1500 VA – Collateral: IEC 60601-1-1:2000
– Other 2 1500 VA IEC 60601-1-2:2001 + A1:2004
IEC 60601-1-4:2000
Environmental Conditions – Particular: IEC 60601-2-37:2001 + A1:2004 + A2:2005
• Operating conditions • Other regions requiring compliance with IEC 60601-1 Ed. 2
– Ambient temperature: 10°C to 35°C (20°C to 35°C when a – General: IEC 60601-1:1988 + A1:1991 + A2:1995
4D transducer is used) – Collateral: IEC 60601-1-1:2000
– Relative humidity: 35% to 80% (no condensation) IEC 60601-1-2:2001 + A1:2004
– Atmospheric pressure: 700 hPa to 1060 hPa IEC 60601-1-4:1996 + A1:1999
• Storage conditions – Particular: IEC 60601-2-37:2001 + A1:2004 + A2:2005
– Ambient temperature: -10°C to 50°C • Other regions requiring compliance with IEC 60601-1 Ed. 3
– Relative humidity: 30% to 90% (no condensation) – General: IEC 60601-1:2005
– Atmospheric pressure: 700 hPa to 1060 hPa – Collateral: IEC 60601-1-2:2007
Safety Classification – Particular: IEC 60601-2-37:2007
– According to the type of protection against electric
shock: CLASS I or Internally Powered Equipment * The above standards are applicable to the ultrasound sys-
– According to the degree of protection against electric tem at the time of purchase.
shock: These standards continue to remain applicable even if the
EQUIPMENT WITH TYPE-BF APPLIED PARTS (Transducer, system configuration is changed as a result of using
ECG electrodes, PCG and Pulse) options in combination.
– According to the degree of protection against harmful
ingress of water:
IPX0 (enclosed EQUIPMENT without protection against
ingress of water)
However, the footswitch is IPX8 and the transducers are
IPX7 (excluding the connector part).
– According to the degree of safety of application in the
presence of a FLAMMABLE ANESTHETIC MIXTURE WITH
AIR or WITH OXYGEN OR NITROUS OXIDE:
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE
– According to the mode of operation:
CONTINUOUS OPERATION
– Sterilization method
· System main unit: Not suitable for sterilization.
· Transducers: Sterilization methods are specified
in the relevant operation manuals.

19
DIMENSIONS, MASS, AND POWER CONSUMPTION
External dimensions Mass
mm (in) Power consumption
Unit Model name kg (lb)
(approx.)
Width Height Depth (approx.)
1,371 (54.0)
TUS-A300 19-inch 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(standard panel height) 50 VA*2
1,768 (69.6)
1,345 (53.0)
TUS-A300 21.5-inch 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(standard panel height) 50 VA*2
1,742 (68.6)
Main unit*1
1,437 (56.6)
TUS-A300 19-inch for USA 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(elevated panel height) 50 VA*2
1,834 (72.2)
1,411 (55.6)
TUS-A300 21.5-inch for USA 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(elevated panel height) 50 VA*2
1,808 (71.2)
287.7 (11.3)
Sony HVO-550MD (including
212 (8.4) 105.5 (4.2) 3.2 (7.1) 43.2 W
[NTSC/PAL] the projection
section)
DVD video recorder
Sony DVO-1000MD
212 (8.4) 128.5 (5.1) 382 (15) 6 (13.2) 56 W
[NTSC/PAL]
JVC BD-X201M [NTSC]
212 (8.4) 88 (3.5) 352 (13.9) 4.8 (10.6) 100 VA to 108 VA
JVC BD-X201ME [PAL]
Sony UP-D897 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
B/W digital printer Sony UP-D898MD 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
Mitsubishi P95DW 154 (6.1) 84.5 (3.3) 239 (9.4) 2.6 (5.7) 190 VA (printing)
Mitsubishi CP30DW 212 (8.3) 125 (4.9) 425 (16.7) 7.3 (16.1) 180 VA (printing)
Color digital printer
Sony UP-D25MD 212 (8.3) 98 (3.9) 398 (15.7) 5.5 (12.1) 240 VA (printing)
*1: Two types of systems (systems with a standard panel height and systems with an elevated panel height) are available.
Availability differs depending on the region.
*2: In Standby mode with internal battery UEBT-A500A

20
MPDUS0073EAB

MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit USPI-A300A Precision Imaging kit 0.1 (0.2)
TUS-A300 Aplio 300 125 (275.6) USSO-A400A Tissue Specific Optimization kit 0.1 (0.2)
Options/Accessories for main unit USST-A500A 4D STIC Imaging kit 0.1 (0.2)
UICW-A500A CW unit 2.0 (4.4) USLM-A500A Luminance kit 0.1 (0.2)
UJUR-A500A Reference signal unit 3.0 (6.6) USFR-A500A Smart 3D kit 0.1 (0.2)
UJUR-A501A Reference signal unit 3.0 (6.6) USPA-A500A Protocol Assistant kit 0.1 (0.2)
UJUR-772A Reference signal sensor unit 3.0 (6.6) USFN-A500A Smart Fusion kit 0.1 (0.2)
UZRI-A500A Mounting kit for peripheral unit 5.6 (12.4) Superb Micro vascular Imaging
USMI-A300A 0.1 (0.2)
UZRI-A501A Mounting kit for peripheral unit 0.8 (1.8) kit
UZFS-A500A Footswitch 1.6 (3.5) USTS-A300A Trapezoid Scan kit 0.1 (0.2)
UZGW-007A Gel warmer 1 (2.2) USDF-A300A Dynamic Flow kit 0.1 (0.2)
UAEH-770A M-TEE hanger kit 1.2 (2.6) USMP-A500A MicroPure kit 0.1 (0.2)
UAEH-002A Motor drive M-TEE hanger kit 1.3 (2.9) USEL-A301A Elastography-FLR kit 0.1 (0.2)
UIMV-A500A 4D unit 2.0 (4.4) USEL-A300A Elastography kit 0.1 (0.2)
UEBT-A500A Battery kit 1.5 (3.3) USFL-A500A FLR kit 0.1 (0.2)
UZPH-A300A Transducer Holder (Right) kit 0.8 (1.8) USSW-A500A Shear wave kit 0.1 (0.2)
UZPH-A301A Transducer Holder (Left) kit 0.3 (0.7) USPV-A500A Panoramic View kit 0.1 (0.2)
UZPH-A302A Transducer Holder (Left) kit 0.3 (0.7) USVI-A500A Vascularity Index kit 0.1 (0.2)
UZMK-A500A Transducer cable hanger kit 0.9 (2.0) USXM-A500A FLEX-M kit 0.1 (0.2)
UACV-A500A CV kit 1.0 (2.2) USAN-A500A Auto NT kit 0.1 (0.2)
UIHV-A400A HV power kit 0.4 (0.9) USPS-A300A Probe 4port kit 0.1 (0.2)
UIFR-A500A Mounting kit for Fusion unit 15 (33.1) UOPM-A300A Operation manuals 3.0 (6.6)
UIFR-A501A Sensor kit for Fusion unit 0.1 (0.2) UZKF-A500A Local Language Key-Top kit 0.5 (1.1)
UZWT-A500A Fusion pole cart 26 (57.3) UZKG-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-001A Mounting kit for fusion sensor 0.1 (0.2) UZKI-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-002A Mounting kit for fusion sensor 0.1 (0.2) UZKS-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-003A Mounting kit for fusion sensor 0.1 (0.2) UZKR-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-004A Mounting kit for fusion sensor 0.1 (0.2) UZKW-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-005A Mounting kit for fusion sensor 0.1 (0.2) UZUI-A502A Mounting kit for UPS unit 40 (88.2)
UAFS-006A Mounting kit for fusion sensor 0.1 (0.2) UP-D897 B/W printer 2.6 (5.7)
UAFS-007A Mounting kit for fusion sensor 0.1 (0.2) UP-D898MD B/W printer 2.6 (5.7)
UIWL-A500A Wireless LAN kit 0.5 (1.1) P95DW B/W printer 2.6 (5.7)
USDI-A500A DICOM kit 0.1 (0.2) UP-D25MD Color printer 5.5 (12.1)
USSE-A500A Stress Echo kit 0.1 (0.2) CP30DW Color printer 7.3 (16.1)
USWT-A500A 2D Wall Motion Tracking kit 0.1 (0.2) HVO-550MD DVD video recorder 3.2 (7.1)
USHI-A500A CHI kit 0.1 (0.2) DVO-1000MD DVD video recorder 6 (13.2)
USCQ-A501A CHI-Q kit 0.1 (0.2) BD-X201M DVD video recorder 4.8 (10.6)
USCQ-A502A Fitting curve kit 0.1 (0.2) BD-X201ME DVD video recorder 4.8 (10.6)
USPP-A300A ApliPure Plus kit 0.1 (0.2)
USDT-A300A Differential THI kit 0.1 (0.2)

21
MPDUS0073EAB

MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
Transducers PLT-604AT Linear array transducer 0.9 (2.0)
PST-25BT Phased array transducer 0.8 (1.8) PLT-704AT Linear array transducer 0.85 (1.9)
PST-30BT Phased array transducer 0.8 (1.8) PLT-704SBT Linear array transducer 0.9 (2.0)
PST-50BT Phased array transducer 0.8 (1.8) PLT-705BT Linear array transducer 0.85 (1.9)
PST-65AT Phased array transducer 0.7 (1.5) PLT-705BTF Linear array transducer 0.78 (1.7)
Convex array biopsy PLT-705BTH Linear array transducer 0.78 (1.7)
PVT-350BTP 0.95 (2.1)
transducer PLT-805AT Linear array transducer 0.85 (1.9)
PVT-375BT Convex array transducer 0.95 (2.1) PLT-1005BT Linear array transducer 0.85 (1.9)
PVT-375MV Convex array transducer 1.1 (2.4) PLT-1202BT Linear array transducer 0.75 (1.65)
PVT-375SC Convex array transducer 0.95 (2.1) PLT-1202S Linear array transducer 0.85 (1.9)
PVT-382BT Convex array transducer 0.8 (1.8) PLT-1204BT Linear array transducer 0.85 (1.9)
PVT-382MV Convex array transducer 1.04 (2.3) PLT-1204MV Linear array transducer 1.23 (2.7)
PVF-475BT Convex array transducer 0.8 (1.8) Multi-plane transesophageal
PET-508MA 1.2 (2.6)
PVT-661VT Endocavitary transducer 1 (2.2) transducer
PVT-674BT Convex array transducer 0.9 (2.0) Multi-plane transesophageal
PET-510MA 1.25 (2.8)
PVT-675MV Convex array transducer 1.1 (2.4) transducer
PVT-675MVL Convex array transducer 1.1 (2.4) Multi-plane transesophageal
PET-510MB 1.25 (2.8)
transducer
PVT-681MV Endocavitary transducer 1.15 (2.5)
Multi-plane transesophageal
PVT-681MVL Endocavitary transducer 1.15 (2.5) PET-511BTM 1.5 (3.3)
transducer
PVT-712BT Convex array transducer 0.8 (1.8)
Multi-plane transesophageal
PVT-745BTF Convex array transducer 0.8 (1.8) PET-512MA 1.32 (2.9)
transducer
PVT-745BTH Convex array transducer 0.8 (1.8) Multi-plane transesophageal
PET-512MC 1.6 (3.5)
PVT-745BTV Convex array transducer 0.78 (1.7) transducer
PVT-770RT Endocavitary ttransducer 2 (4.4) Multi-plane transesophageal
PET-512MD 1.5 (3.3)
PVT-781VT Endocavitary transducer 0.97 (2.1) transducer
PVT-781VTE Endocavitary transducer 0.97 (2.1) PET-805LA Linear array transducer 1.17 (2.58)
PVL-715RST Endocavitary transducer 1.04 (2.3) PC-20M CW Doppler pencil transducer 0.085 (0.2)
PLT-308P Linear array biopsy transducer 0.85 (1.8) PC-50M CW Doppler pencil transducer 0.08 (0.2)

1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan

https://global.medical.canon
©Canon Medical Systems Corporation 2017-2018. All rights reserved.
Design and specifications are subject to change without notice.
Model number: TUS-A300 MPDUS0073EAB V7.0 2018-09 CMSC/Produced in Japan

Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.

Aplio, Dynamic Flow, ApliPure, MicroPure, Pulse Subtraction, TwinView and Made for Life are trademarks of Canon Medical Systems
Corporation.

Other company names appearing in this document may be trademarks or registered trademarks of their respective holders.