Product Data
No. MPDUS0073EAB
Upgradability Operability
Aplio 300 can be periodically upgraded to the latest System operability is optimized for the overall clinical
version. workflow in hospitals.
SYSTEM MATRIX OF TUS-A300
Unit Model name Remarks
TUS-A300 19-inch or 21.5-inch LCD monitor, DVD/CD drive, ApliPureTM, Quick Scan, TDI, QSP, HPRF, and DICOM Media
Main unit
Aplio 300 Storage are included.
<Options for main unit>
CW unit UICW-A500A For cardiovascular examinations, add Continuous Doppler capability to sector and pencil transducers.
For cardiovascular examinations (for regions other than the USA)
UJUR-A500A
(ECG (Electrocardiogram), Respiration, Auxiliary input (Aux.), ECG gating, Heart rate)
Reference signal unit
For cardiovascular examinations (for the USA)
UJUR-A501A
(ECG (Electrocardiogram), Respiration, Auxiliary input (Aux.), ECG gating, Heart rate)
Reference signal sensor unit UJUR-772A PCG and Pulse sensor.
Rack for mounting a DVD recorder and/or a color printer on two levels. This kit cannot be mounted togeth-
Mounting kit for peripheral UZRI-A500A er with the UZPH-A300A (transducer holder on the right side of the operating panel).
units Not necessary when UIFR-A500A is used.
UZRI-A501A Rack for mounting a B/W printer
Footswitch UZFS-A500A Switch used for freezing, printing, and some other operations by foot.
Gel warmer UZGW-007A This unit warms the ultrasound gel to a suitable temperature.
UAEH-770A TEE transducer hanger for the PET-510MA, PET-510MB, and PET-508MA.
M-TEE hanger
UAEH-002A TEE transducer hanger for the PET-511BTM, PET-512MA, PET-512MC and PET-512MD.
This unit is required for using the 4D transducer or the motor-driven TEE transducer
4D unit UIMV-A500A
(PET-511BTM/PET-512MC/PET-512MD).
Battery kit UEBT-A500A For standby mode.
Adds a transducer holder on the right side of the operating panel. Dedicated option for Aplio300.
Transducer Holder (Right) kit UZPH-A300A This kit cannot be mounted together with the UZRI-A500A (rack for mounting a color printer and a video
recorder).
Adds three transducer holders on the left side of the operating panel. Dedicated option for Aplio300 plati-
UZPH-A301A
num series. (for V1.0 to V4.0)
Transducer Holder (Left) kit
Adds three transducer holders on the left side of the operating panel. Dedicated option for Aplio300 plati-
UZPH-A302A
num series. (for V5.0 or later)
Transducer cable hanger kit UZMK-A500A Long hanger on which the transducer cable is hooked.
CV kit UACV-A500A This kit consists of preset data suitable for cardiovascular examinations, a CV sticker, and a startup screen.
HV power kit UIHV-A400A For the PVL-715RST.
This kit consists of the magnetic sensor required for Smart Fusion and the kit that mounts the magnetic
sensor to the system main unit and to the transducer PVT-375BT or PVT-375SC. The USFN-A500A is required
Mounting kit for Fusion unit UIFR-A500A separately. Mounting kit for Fusion sensor is required separately. Either a DVD recorder or a color printer
can be installed. It is also possible to install both. In this case, however, there may be some interference with
vertical movement of the operating panel.
Sensor kit for Fusion unit UIFR-A501A Magnetic sensor for the Smart Fusion function.
This pole cart allows the magnetic field transmitter included in the system main unit to be positioned
Fusion pole cart UZWT-A500A
independently.
UAFS-001A For PVT-382BT The UIFR-A500A is required separately.
UAFS-002A For PVT-350BTP The UIFR-A500A is required separately.
UAFS-003A For PVT-781VT The UIFR-A500A is required separately.
Mounting kit for fusion sensor UAFS-004A For PLT-1005BT The UIFR-A500A is required separately.
UAFS-005A For PVL-715RST The UIFR-A500A is required separately.
UAFS-006A For PVT-781VTE The UIFR-A500A is required separately.
UAFS-007A For PVT-475BT The UIFR-A500A is required separately.
Option kit for connecting to a DICOM network via a wireless LAN instead of the previously used LAN cable
Wireless LAN kit UIWL-A500A USDI-A500A is required. Complied with the Radio Law of Japan and applicable laws and regulations of USA,
EU member states, Iceland, Norway, Liechtenstein, and Switzerland.
USDI-A500A Verification, Storage, Print, Storage Commitment, MULTI FRAME (Network Transfer), MWM (Modality Worklist
DICOM kit
USDI-A500A/EL Management), Query/Retrieve, MPPS (Modality Performed Procedure Step), Structured Reporting.
USSE-A500A Adds the cardiac stress examination function to the system.
Stress Echo kit
USSE-A500A/EL The UJUR-A500A or UJUR-A501A is required (available separately).
USWT-A500A Adds the cardiac wall motion analysis function to the system.
2D Wall Motion Tracking kit
USWT-A500A/EL The UJUR-A500A or UJUR-A501A is required.
2
MPDUS0073EAB
3
Unit Model name Remarks
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AJ
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A. Only for Japan.
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AA
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A.
Mounting kit for UPS unit UZUI-A502A/MA
Line voltage: 110V~120V, Line frequency: 60Hz region
This kit includes the UPS unit and the mounting bracket.
UZUI-A502A/AE
This UPS unit cannot be used together with UZRI-A500A and UIFR-A500A.
UZUI-A502A/ME
Line voltage: 220V~240V, Line frequency: 50Hz/60Hz region
UZKF-A500A
(French)
UZKG-A500A
(German)
UZKI-A500A
(Italian)
Local Language This kit is intended for changing the key tops of the full keyboard to support specific
Key-Top kit UZKS-A500A languages.
(Spanish)
UZKR-A500A
(Russian)
UZKW-A500A
(Danish, Norwegian,
Swedish, Finnish)
Note: 1. Provided in the standard configuration of systems with V3.0 or later in which the 19-inch and 21.5-inch monitor is used.
2. "/EL" is a supplemental model name indicating options supplied by electronic license, only available in Europe, USA, and Canada.
(*1) Not available in the USA. (*2) Only available in the USA.
4
MPDUS0073EAB
Main unit
DICOM kit
TUS-A300
USDI-A500A
Reference signal unit
Wireless LAN kit
UJUR-A500A/-A501A
UIWL-A500A
Reference signal sensor unit
UJUR-772A
ApliPure plus kit
USPP-A300A (Note)
Stress Echo
USSE-A500A
Differental THI kit
USDT-A300A
2D Wall Motion Tracking kit
USWT-A500A
Precision imaging kit
USPI-A300A (Note)
CW unit
UICW-A500A Tissue Specific Optimization kit
USSO-A400A
CV kit
UACV-A500A Endocavitary transducer
PVL-715RST
FLEX-M kit HV power kit
USXM-A500A UIHV-A400A
Footswitch 4D unit
UZFS-A500A UIMV-A500A
MicroPure kit
USMP-A500A
A B
5
B
C
Elastography-FLR kit Elastography kit
USEL-A301A USEL-A300A
Mounting kit for
peripheral unit Mounting kit for
FLR kit (DVD recorder and UPS unit
USEL-A500A color printer) UZUI-A502A
UZRI-A500A
UPS unit cannot be
used together with
Shear wave kit Mounting kit for UZRI-A500A and UIFR-A500A.
USSW-A500A peripheral unit These items cannot
(B/W printer) be mounted
Panoramic View kit UZRI-A501A at the same time.
USPV-A500A
Transducer holder
(right) kit
Vascularity Index kit
UZPH-A300A
USVI-A500A
Transducer holder
Auto NT kit
(left) kit
USAN-A500A
UZPH-A301A
6
MPDUS0073EAB
7
Precision Dynamic
Transducer name 2D M PWD CWD*10 CDI THI D-THI*1 ApliPure Boost
Imaging*1/TE Flow*1
PST-25BT *2 *3
PST-30BT *2 *3
*2
PST-50BT *3
*3 *2
PST-65AT *3
PVT-350BTP
PVT-375BT
PVT-375MV
PVT-375SC
PVT-382BT
PVT-382MV
PVT-475BT
PVT-661VT
PVT-674BT
PVT-675MV
PVT-675MVL
PVT-681MV
PVT-681MVL
PVT-712BT
PVT-745BTF
PVT-745BTH
PVT-745BTV
PVT-770RT
PVT-781VT
PVT-781VTE
Convex
PVL-715RST
Linear
PLT-308P
PLT-604AT
PLT-704AT
PLT-704SBT
PLT-705BT
PLT-705BTF
PLT-705BTH
PLT-805AT
PLT-1005BT
PLT-1202BT
PLT-1202S
*1: Optional software is required.
PLT-1204BT
*2: TE (Tissue Enhancement)
PLT-1204MV *3: Depends on the preset.
PET-508MA *2 *4: Not available in the USA.
*5: Not available in the USA and
PET-510MA *2
Canada.
PET-510MB *2 *6: Only for the USA.
PET-511BTM *2 *7: Optional software USPP-A300A is
required.
PET-512MA *2
*8: Optional unit UIHV-A400A is
PET-512MC *2 required.
PET-512MD *2 *9: Optional unit UIMV-A500A is
required.
PET-805LA
*10: Optional unit UICW-A500A is
PC-20M required.
PC-50M
8
MPDUS0073EAB
9
STIC Smart Micro- Elasto- 2D Shear
Transducer name SMI*1 BEAM*7 Fusion*1 Remarks
Color*1, 9 3D*1 Pure*1 graphy*1 WMT*1 wave*1
PST-25BT
PST-30BT
PST-50BT
PST-65AT
PVT-350BTP *5
PVT-375BT
PVT-375MV *3 *9
PVT-375SC *11
PVT-382BT
PVT-382MV *9
PVT-475BT *11
PVT-661VT
PVT-674BT
PVT-675MV *3 *9
*3
PVT-675MVL *9
PVT-681MV *9
PVT-681MVL *9
PVT-712BT
PVT-745BTF
PVT-745BTH
PVT-745BTV
PVT-770RT
PVT-781VT *4
*4
PVT-781VTE
Convex
PVL-715RST *8
Linear
PLT-308P *5
PLT-604AT
PLT-704AT
PLT-704SBT
PLT-705BT
PLT-705BTF
PLT-705BTH
PLT-805AT
PLT-1005BT
PLT-1202BT
PLT-1202S
PLT-1204BT
*1: Optional software is required.
PLT-1204MV *9 *2: TE (Tissue Enhancement)
PET-508MA *3: Depends on the preset.
*4: Not available in the USA.
PET-510MA *4
*5: Not available in the USA and
PET-510MB *6 Canada.
PET-511BTM *5, 9 *6: Only for the USA.
*7: Optional software USPP-A300A is
PET-512MA *5 required.
PET-512MC *9 *8: Optional unit UIHV-A400A is
required.
PET-512MD *9
*9: Optional unit UIMV-A500A is
PET-805LA required.
PC-20M *10 *10: Optional unit UICW-A500A is
required.
PC-50M *10
*11: Using Single Crystal technology.
10
MPDUS0073EAB
4D OPERATION MODES
The optional unit UIMV-A500A is required for this operation.
4D Live Single Sweep
Transducer Pulse Pulse Pulse Pulse Volume 4D Max sweep range
STIC*1 STIC color*1 Luminance*1
name Fund. Subtract. Subtract. Fund. Subtract. Subtract. color Biopsy (deg)
ON OFF ON OFF
PVT-375MV *2 *3 *3 75
*2
PVT-382MV 80
PVT-675MV *2 *3 *3 75
*2 *3 *3
PVT-675MVL 90
PVT-681MV *2 90
*2
PVT-681MVL 150
PLT-1204MV *2 30
3D OPERATION MODES
The optional software USFR-A500A is required for this operation.
2D CHI*1
Transducer Pulse Pulse Dynamic
CDI Power Contrast SMI*1 Shear wave*1 Remarks
name Fund. Subtract. Subtract. Flow*1 2D
SMI*1
ON OFF
PVT-350BTP *2 *4
*2
PVT-375BT
PVT-375SC *2
*2
PVT-382BT
PVT-475BT *2
*2 *3 *3
PVT-674BT
*2
PVT-712BT
PVT-745BTF *2 *3
PVT-745BTH *2 *3
*2
PVT-745BTV
PLT-308P
PLT-604AT *2 *3
PLT-704AT *2 *3
PLT-704SBT *2 *3 *3
*2
PLT-705BT
PLT-705BTF *2 *3
PLT-705BTH *2 *3
PLT-805AT *2 *3
*2
PLT-1005BT
*2 *3
PLT-1202BT
PLT-1202S *2 *3
PLT-1204BT *2 *3 *3
11
SPECIFICATIONS · Line density
· Parallel signal processing
System
– Parallel signal processing is available for up to two direc-
• Scan methods: Linear scan (some transducers
tions.
can perform oblique scanning)
• Scan Angle and Scan Width
Sector scan
– Adjustment of the field width (scan width, scan angle) is
Convex scan
possible.
Trapezoid scan (USTS-A300A is
– Adjustment of beam steering (scanning position) is pos-
required.)
sible.
• Monitor:
– Adjustment of linear beam steering is possible.
High-definition 19-inch LCD monitor
• PAN/EXPAND
– Resolution: 1280 × 1024 (SXGA)
Real-time PAN/EXPAND
– High-definition 21.5-inch LCD monitor
– Scale enlargement/reduction using the encoder is possi-
– Resolution: 1600 × 900 (WXGA++)
ble.
• Presets
– Movement to the desired section using the trackball is
– System preset: 1 type
possible.
– Application preset: 20 types
– The transmission focus is optimized in steps above.
– The specified range of an image can be magnified. (Spot
Compatible Peripheral Devices
Zoom)
• Black-and-white digital printer
• Transmission Focus
– UP-D897 (100 V to 240 V, SONY)
Transmission conditions: A maximum of 8 steps
– UP-D898MD (100 V to 240 V, SONY)
• Transmission Frequency
– P95DW (100 V to 240 V, MITSUBISHI)
Multi frequency
• Color digital printer
– 2D mode
– UP-D25MD (100 V to 240 V, SONY)
5 types (at maximum)
– CP30DW (120 V, 220V to 240 V, MITSUBISHI)
– THI mode (second harmonic)
• DVD video recorder
5 types (at maximum)
– HVO-550MD (100 V to 240 V, NTSC/PAL, SONY)
• Pulser
– DVO-1000MD (100 V to 240 V, NTSC/PAL, SONY)
Dual Pulser
– BD-X201M (100 V to 240 V, NTSC, JVC)
• Reception Conditions
* For regions other than Europe
Continuous focus
– BD-X201ME (100 V to 240 V, PAL, JVC)
• GAIN
* For Europe
– The display brightness for 2D can be changed.
• USB flash drive
(also available when the image is frozen)
• Barcode reader
– The display brightness for 2D and M can be changed
simultaneously.
2D mode (B mode)
• STC
• Viewing Depth
8-step slide controls (common for 2D and M)
The viewing depth depends on the transducer used.
• Acoustic Output
– Convex
Adjustment is possible to 100%.
· Minimum depth: 2 cm
· Maximum depth: 40 cm • Adjusting the 2D Image Quality
– Linear – Dynamic range (also available when the image is frozen)
· Minimum depth: 1 cm – Edge enhancement
· Maximum depth: 14 cm – Time-smoothing (persistence)
– Sector – Map (also available when the image is frozen)
· Minimum depth: 1 cm – Auto gain control
· Maximum depth: 28 cm – Lateral gain
• Line density – Gamma (also available when the image is frozen)
– The line density differs depending on the transducer – P-SEL (frame rate)
used. – ApliPure
– The line density can be changed. – Precision (USPI-A300A is required.)
• Ultrasound Frame Rate – Tissue Enhancement
– The ultrasound frame rate can be adjusted by using the – Boost
following in combination.
12
MPDUS0073EAB
• 2D Map • M Map
– The grayscale pattern can be changed and virtual color M image virtual color setting is possible. The setting can
setting for the 2D image is possible. be changed even after the image has been frozen.
– Settings can also be changed when the image is frozen. • THI (Tissue Harmonic Imaging)
• 2D Quick Scan 2D mode and THI mode are linked, and M images can be
The gain and STC can be adjusted automatically. displayed in THI mode.
• THI (Tissue Harmonic Imaging) – Pulse subtraction method
THI signal processing methods – Filtering method
– Pulse subtractionTM method – Differential method (USDT-A300A is required.)
– Filtering method • M Mark
– Differential method (USDT-A300A is required.) – The M cursor can be displayed on 2D or C images.
– The M cursor displayed position can be adjusted.
• Display Orientation
• Flex-M (USXM-A500A is required.)
– Top/bottom reversal is possible.
Any desired plane can be set on the 2D-mode image and
– Left/right reversal is possible.
the M-mode image for the set plane can be reconstructed.
• ApliPureTM
– ApliPure Doppler (Spectrum Doppler)
This function reduces ultrasound wave interference with- • Doppler Mode
in tissues, which appear as speckle patterns or speckle – PWD (pulsed-wave Doppler)
noise on 2D images. – HPRF PWD (can be switched to HPRF mode by preset-
– ApliPure+ (USPP-A300A is required.) ting)
This function can display the boundaries between tissues – CWD (continuous-wave Doppler)
more clearly and reduce speckle noise and acoustic (UICW-A500A is required.)
shadows. – Pencil CWD (pencil-type transducer)
• MicroPureTM (USMP-A500A is required.) (UICW-A500A is required.)
– Small structures can be extracted by performing filtering • Doppler Pulse Repetition Frequency (PRF)
for 2D-mode images. – PWD: 0.3 kHz to 52.1 kHz
– Visualization of very small calcifications and other – CWD: 1.4 kHz to 52.1 kHz
extremely small lesions can be improved. • Doppler Scan
– 2D/D simultaneous scan
• Precision Imaging (USPI-A300A is required.)
– D only scan
Structures in 2D-mode images can be displayed more
• Doppler Sampling Volume
clearly and the background can be displayed more
The Doppler range gate width can be changed.
smoothly.
• Doppler Sampling Shift
• TSO (Tissue Specific Optimization) 0 cm to the maximum depth
(USSO-A400A is required.) • Doppler Filter
– Reception focus compensation can be performed. The Doppler filter cutoff can be changed.
– Automatic reception focus compensation can be • Doppler Gain
performed. (Auto TSO) The display brightness for Doppler can be changed.
• BEAM (Biopsy Enhancement Auto Mode) • Doppler Quick Scan
– Display of the needle can be enhanced in the image. The Doppler scale and baseline shift can be adjusted auto-
– The enhancement level can be adjusted. matically.
* The optional USPP-A300A is required in order to use • Doppler Frequency Analysis and Image Processing
this mode. – Method: FFT
– No. of data items: 256 (maximum)
M mode • Indication of Doppler Spectrum Direction
• M Transmission Frequency Reverse display of the velocity spectrum is possible.
Multi frequency: 5 types (at maximum) • Doppler Baseline Shift (Zero Shift)
• M Sweep Speed – The velocity baseline of Doppler images can be shifted.
The Sweep Speed can be changed in M mode. – The baseline shift setting can also be adjusted when
• M Gain images that were frozen are displayed.
M gain can be corrected for 2D gain. • Doppler Audio
• M Image Processing Parameters Stereo output (blood flow toward and away from the trans-
– M dynamic range ducer)
– M edge enhancement • Doppler Map
– M auto gain control The brightness conversion table and the virtual color for
– M gamma (can be changed even after the image has Doppler images can be set.
been frozen). • Display of Doppler Scale
2 types (velocity, Doppler shift frequency)
13
• Doppler Focus • C Baseline (Zero Shift)
Automatically follows the sample position. – The velocity baseline of color Doppler images can be
• Doppler Angle Mark shifted.
This mark is displayed for measuring the angle between – The baseline shift setting can also be adjusted when
the direction of the velocity and the direction of the ultra- images that were frozen or images in the image memory
sound beam. are displayed.
• Doppler Oblique Scan (PWD Steering) • Reverse C Display
– Oblique scans are possible using a specific linear trans- – Coloring is reversed.
ducer. – Changes can also be made when the image is frozen.
– Auto Invert function • Black and White/Color Balance
• Doppler Multifrequency – By comparing the color Doppler images and B/W imag-
The PWD transmission frequency can be changed. es, color weighting to B/W can be set.
• Doppler Sweep Speed – Changes can also be made when the image is frozen.
The Sweep Speed can be changed in Doppler mode. • C Parallel Processing
• Doppler Display Dynamic Range Beam reception from up to four directions for each trans-
The display dynamic range of the Doppler image can be mission beam.
changed. • C Gain
• Doppler Auto Trace (measurement performed after freez- The display brightness of color Doppler images can be
ing the image) changed.
Measurement of peak velocity and mean velocity is possi- • C Multifrequency
ble by automatic velocity tracing. The transmission frequency for color Doppler image acqui-
The following Doppler waveform trace is possible. sition can be changed.
– Trace style: Waveform Peak, Mean, Peak + • C Line density
Mean The color Doppler image line density can be changed.
– Trace area specified: Forward, Reverse, Full, Auto • C ROI (Region of Interest)
– Measurement item: Max, Min, Mean, PI, RI, etc. Position, size, and steering adjustment is possible for color
Doppler ROIs.
Color Doppler • C Transmit Focus
Color Doppler 2D Mode (BCD/CDF Mode) Automatically follows the color Doppler ROI position.
• Display mode • C Filter
– CDI mode – The C filter can be changed.
· Flow velocity – FIO filter
· Flow velocity/variance • Variance Curve
· Power – The display of the color variance component can be
– Power Angio mode adjusted.
Direction display is possible.
– TDI mode Color Doppler M mode (MDF Mode)
– Dynamic Flow mode (USDF-A300A is required.) • Display mode
Direction display is possible. – MCDI mode
– TwinViewTM · Velocity display
Simultaneous dual-screen display with 2D mode is avail- · Velocity/variance display
able. · Power display
– Superb Micro vascular Imaging mode – M-TDI mode
(USMI-A300A is required.) • M Color Doppler Map (CDI MAP)
· Clutter suppressed Color Doppler map can be selected for each mode.
· Blood flow enhanced • M Color Doppler Velocity Range Selection (C Scale)
• C Map The velocity range can be selected.
– C map can be selected for each color Doppler mode. • M Color Doppler Baseline (C Baseline)
– Changes can also be made when the image is frozen. – The zero-velocity line on the M Color Doppler image can
• C Scale (Switching the Velocity Range) be shifted.
The velocity range can be changed. – The baseline shift setting can also be adjusted when
• C Time-Smoothing (Persistence) images that were frozen are displayed or when the
The result of temporal correlation processing between the image in the image memory is played back.
previous image and current image can be displayed. • Color Reverse Display
• C Frame Interpolation – The colors can be reversed.
Automatic setting according to the transducer information – Changes can also be made when the image is frozen.
and selected study conditions.
14
MPDUS0073EAB
15
• Smart 3D (USFR-A500A is required.) – Thumbnail area
3D image can be generated from the 2D image and any · Image data acquired during the current examination is
input volume shape. displayed.
· Information from a previous examination of the patient
(*) USMI-A300A is required. currently being examined is displayed.
– Information message display area
Display-Related Features An operation guide and other messages are displayed.
• Display Method – Status area
– Images on the main unit: 60 Hz non-interlaced display The following system statuses can be displayed.
– Images from external · Saving dynamic/still image
playback devices: 60 Hz non-interlaced display · Used space on HDD
• Monitor Display/Character Display · DICOM printer status/peripheral device status
– ID area · DVD/CD write status
· Patient ID · Trackball assignment status
· Patient name • Annotation
· Hospital name – Manual input using the keyboard is possible.
· Date: Selected from among the formats – Auto annotation (previously specified text) is possible.
shown below. • Pictograms
YYYY/MM/DD Body icons and transducer mark
MM/DD/YYYY • Biopsy Guide Mark
DD/MM/YYYY
Biopsy guide mark display is possible.
YYYY: Western calendar year
• Language
MM: Month
DD: Day The following languages are supported for the display of
· Time: Selected from among the formats some screens and keyboard entry.
shown below. Supported Screen Keyboard
hh:mm:ss: AM (PM) language display entry
hh:mm:ss: 24-hour representation
hh: Hour English
mm: Minute
ss: Second English (UK)
(same as"English")
· VCR counter
· Age German
· Sex French
· Heart rate display (heart-shaped mark/heart rate)
· Name of the Imaging Preset Italian
· Name of the operator Spanish
· Gestational age Danish
– Acoustic power display area
Dutch
· Acoustic power value (%)
· TI value Norwegian
– Auto data Swedish
Frame rate
Finnish
Acoustic power index = MI indication
Transducer frequency Portuguese
Depth Icelandic
Dynamic range
Russian
GAIN
CDI filter : Applicable : Not applicable
PRF
Doppler filter
Doppler angle
Doppler gate size
16
MPDUS0073EAB
18
MPDUS0073EAB
19
DIMENSIONS, MASS, AND POWER CONSUMPTION
External dimensions Mass
mm (in) Power consumption
Unit Model name kg (lb)
(approx.)
Width Height Depth (approx.)
1,371 (54.0)
TUS-A300 19-inch 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(standard panel height) 50 VA*2
1,768 (69.6)
1,345 (53.0)
TUS-A300 21.5-inch 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(standard panel height) 50 VA*2
1,742 (68.6)
Main unit*1
1,437 (56.6)
TUS-A300 19-inch for USA 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(elevated panel height) 50 VA*2
1,834 (72.2)
1,411 (55.6)
TUS-A300 21.5-inch for USA 430 VA
518 (20.4) to to 879 (34.6) 125 (275.6)
(elevated panel height) 50 VA*2
1,808 (71.2)
287.7 (11.3)
Sony HVO-550MD (including
212 (8.4) 105.5 (4.2) 3.2 (7.1) 43.2 W
[NTSC/PAL] the projection
section)
DVD video recorder
Sony DVO-1000MD
212 (8.4) 128.5 (5.1) 382 (15) 6 (13.2) 56 W
[NTSC/PAL]
JVC BD-X201M [NTSC]
212 (8.4) 88 (3.5) 352 (13.9) 4.8 (10.6) 100 VA to 108 VA
JVC BD-X201ME [PAL]
Sony UP-D897 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
B/W digital printer Sony UP-D898MD 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
Mitsubishi P95DW 154 (6.1) 84.5 (3.3) 239 (9.4) 2.6 (5.7) 190 VA (printing)
Mitsubishi CP30DW 212 (8.3) 125 (4.9) 425 (16.7) 7.3 (16.1) 180 VA (printing)
Color digital printer
Sony UP-D25MD 212 (8.3) 98 (3.9) 398 (15.7) 5.5 (12.1) 240 VA (printing)
*1: Two types of systems (systems with a standard panel height and systems with an elevated panel height) are available.
Availability differs depending on the region.
*2: In Standby mode with internal battery UEBT-A500A
20
MPDUS0073EAB
MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit USPI-A300A Precision Imaging kit 0.1 (0.2)
TUS-A300 Aplio 300 125 (275.6) USSO-A400A Tissue Specific Optimization kit 0.1 (0.2)
Options/Accessories for main unit USST-A500A 4D STIC Imaging kit 0.1 (0.2)
UICW-A500A CW unit 2.0 (4.4) USLM-A500A Luminance kit 0.1 (0.2)
UJUR-A500A Reference signal unit 3.0 (6.6) USFR-A500A Smart 3D kit 0.1 (0.2)
UJUR-A501A Reference signal unit 3.0 (6.6) USPA-A500A Protocol Assistant kit 0.1 (0.2)
UJUR-772A Reference signal sensor unit 3.0 (6.6) USFN-A500A Smart Fusion kit 0.1 (0.2)
UZRI-A500A Mounting kit for peripheral unit 5.6 (12.4) Superb Micro vascular Imaging
USMI-A300A 0.1 (0.2)
UZRI-A501A Mounting kit for peripheral unit 0.8 (1.8) kit
UZFS-A500A Footswitch 1.6 (3.5) USTS-A300A Trapezoid Scan kit 0.1 (0.2)
UZGW-007A Gel warmer 1 (2.2) USDF-A300A Dynamic Flow kit 0.1 (0.2)
UAEH-770A M-TEE hanger kit 1.2 (2.6) USMP-A500A MicroPure kit 0.1 (0.2)
UAEH-002A Motor drive M-TEE hanger kit 1.3 (2.9) USEL-A301A Elastography-FLR kit 0.1 (0.2)
UIMV-A500A 4D unit 2.0 (4.4) USEL-A300A Elastography kit 0.1 (0.2)
UEBT-A500A Battery kit 1.5 (3.3) USFL-A500A FLR kit 0.1 (0.2)
UZPH-A300A Transducer Holder (Right) kit 0.8 (1.8) USSW-A500A Shear wave kit 0.1 (0.2)
UZPH-A301A Transducer Holder (Left) kit 0.3 (0.7) USPV-A500A Panoramic View kit 0.1 (0.2)
UZPH-A302A Transducer Holder (Left) kit 0.3 (0.7) USVI-A500A Vascularity Index kit 0.1 (0.2)
UZMK-A500A Transducer cable hanger kit 0.9 (2.0) USXM-A500A FLEX-M kit 0.1 (0.2)
UACV-A500A CV kit 1.0 (2.2) USAN-A500A Auto NT kit 0.1 (0.2)
UIHV-A400A HV power kit 0.4 (0.9) USPS-A300A Probe 4port kit 0.1 (0.2)
UIFR-A500A Mounting kit for Fusion unit 15 (33.1) UOPM-A300A Operation manuals 3.0 (6.6)
UIFR-A501A Sensor kit for Fusion unit 0.1 (0.2) UZKF-A500A Local Language Key-Top kit 0.5 (1.1)
UZWT-A500A Fusion pole cart 26 (57.3) UZKG-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-001A Mounting kit for fusion sensor 0.1 (0.2) UZKI-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-002A Mounting kit for fusion sensor 0.1 (0.2) UZKS-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-003A Mounting kit for fusion sensor 0.1 (0.2) UZKR-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-004A Mounting kit for fusion sensor 0.1 (0.2) UZKW-A500A Local Language Key-Top kit 0.5 (1.1)
UAFS-005A Mounting kit for fusion sensor 0.1 (0.2) UZUI-A502A Mounting kit for UPS unit 40 (88.2)
UAFS-006A Mounting kit for fusion sensor 0.1 (0.2) UP-D897 B/W printer 2.6 (5.7)
UAFS-007A Mounting kit for fusion sensor 0.1 (0.2) UP-D898MD B/W printer 2.6 (5.7)
UIWL-A500A Wireless LAN kit 0.5 (1.1) P95DW B/W printer 2.6 (5.7)
USDI-A500A DICOM kit 0.1 (0.2) UP-D25MD Color printer 5.5 (12.1)
USSE-A500A Stress Echo kit 0.1 (0.2) CP30DW Color printer 7.3 (16.1)
USWT-A500A 2D Wall Motion Tracking kit 0.1 (0.2) HVO-550MD DVD video recorder 3.2 (7.1)
USHI-A500A CHI kit 0.1 (0.2) DVO-1000MD DVD video recorder 6 (13.2)
USCQ-A501A CHI-Q kit 0.1 (0.2) BD-X201M DVD video recorder 4.8 (10.6)
USCQ-A502A Fitting curve kit 0.1 (0.2) BD-X201ME DVD video recorder 4.8 (10.6)
USPP-A300A ApliPure Plus kit 0.1 (0.2)
USDT-A300A Differential THI kit 0.1 (0.2)
21
MPDUS0073EAB
MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
Transducers PLT-604AT Linear array transducer 0.9 (2.0)
PST-25BT Phased array transducer 0.8 (1.8) PLT-704AT Linear array transducer 0.85 (1.9)
PST-30BT Phased array transducer 0.8 (1.8) PLT-704SBT Linear array transducer 0.9 (2.0)
PST-50BT Phased array transducer 0.8 (1.8) PLT-705BT Linear array transducer 0.85 (1.9)
PST-65AT Phased array transducer 0.7 (1.5) PLT-705BTF Linear array transducer 0.78 (1.7)
Convex array biopsy PLT-705BTH Linear array transducer 0.78 (1.7)
PVT-350BTP 0.95 (2.1)
transducer PLT-805AT Linear array transducer 0.85 (1.9)
PVT-375BT Convex array transducer 0.95 (2.1) PLT-1005BT Linear array transducer 0.85 (1.9)
PVT-375MV Convex array transducer 1.1 (2.4) PLT-1202BT Linear array transducer 0.75 (1.65)
PVT-375SC Convex array transducer 0.95 (2.1) PLT-1202S Linear array transducer 0.85 (1.9)
PVT-382BT Convex array transducer 0.8 (1.8) PLT-1204BT Linear array transducer 0.85 (1.9)
PVT-382MV Convex array transducer 1.04 (2.3) PLT-1204MV Linear array transducer 1.23 (2.7)
PVF-475BT Convex array transducer 0.8 (1.8) Multi-plane transesophageal
PET-508MA 1.2 (2.6)
PVT-661VT Endocavitary transducer 1 (2.2) transducer
PVT-674BT Convex array transducer 0.9 (2.0) Multi-plane transesophageal
PET-510MA 1.25 (2.8)
PVT-675MV Convex array transducer 1.1 (2.4) transducer
PVT-675MVL Convex array transducer 1.1 (2.4) Multi-plane transesophageal
PET-510MB 1.25 (2.8)
transducer
PVT-681MV Endocavitary transducer 1.15 (2.5)
Multi-plane transesophageal
PVT-681MVL Endocavitary transducer 1.15 (2.5) PET-511BTM 1.5 (3.3)
transducer
PVT-712BT Convex array transducer 0.8 (1.8)
Multi-plane transesophageal
PVT-745BTF Convex array transducer 0.8 (1.8) PET-512MA 1.32 (2.9)
transducer
PVT-745BTH Convex array transducer 0.8 (1.8) Multi-plane transesophageal
PET-512MC 1.6 (3.5)
PVT-745BTV Convex array transducer 0.78 (1.7) transducer
PVT-770RT Endocavitary ttransducer 2 (4.4) Multi-plane transesophageal
PET-512MD 1.5 (3.3)
PVT-781VT Endocavitary transducer 0.97 (2.1) transducer
PVT-781VTE Endocavitary transducer 0.97 (2.1) PET-805LA Linear array transducer 1.17 (2.58)
PVL-715RST Endocavitary transducer 1.04 (2.3) PC-20M CW Doppler pencil transducer 0.085 (0.2)
PLT-308P Linear array biopsy transducer 0.85 (1.8) PC-50M CW Doppler pencil transducer 0.08 (0.2)
https://global.medical.canon
©Canon Medical Systems Corporation 2017-2018. All rights reserved.
Design and specifications are subject to change without notice.
Model number: TUS-A300 MPDUS0073EAB V7.0 2018-09 CMSC/Produced in Japan
Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.
Aplio, Dynamic Flow, ApliPure, MicroPure, Pulse Subtraction, TwinView and Made for Life are trademarks of Canon Medical Systems
Corporation.
Other company names appearing in this document may be trademarks or registered trademarks of their respective holders.