Professional Documents
Culture Documents
TM
MOSAIC HP
Fractional Technology
Operator’s Manual
CAUTION: Users must read this manual carefully and thoroughly before
using this product.
No part of this manual may be re-manufactured or copied in any form without the
written consent of Lutronic Corporation.
IMPORTANT
踊 Lutronic Corporation
June 2012
P/N: 4100110311
Revision 1.1
Lutronic Inc.
850 Auburn Court Tel: 888-588-7644
Fremont, CA 94538 E-mail: officeusa@lutronic.com
USA
Obelis s.a Tel: + (32) 2. 732.59.54
Bd. Général Wahis 53 Fax: + (32) 2.732.60.03
1030 Brussels, BELGIUM E-Mail : mail@obelis.net
Lutronic Corporation
Room 403-1, 2, 3, 4, 5, 404 Tel: +82-31-908-3440
Ilsan Technotown 1141-1 Fax: +82-31-907-3440
Baekseok-dong, Ilsandong- Web site: www.lutronic.com
gu, E-mail: office@lutronic.com
Goyang-si, Gyeonggi-do,
Korea
Table of Contents
2.1 Overview 2 - 1
2.2 Safety for Treatment Room 2 - 3
2.3 General Caution for Physicians, Staff Members and Patients 2 - 5
2.3.1 Safety Checks on the Handpiece Prior to Use 2 - 5
2.3.2 Cautionary Items for Moving the Device in Treatment Room 2 - 5
2.4 Electrical Safety 2 - 6
2.5 Protecting the Eyes 2 - 7
2.6 Fire hazards 2 - 8
2.7 Safety Features of the MOSAIC HP Laser System 2 - 9
2.7.1 Key Switch 2 - 9
2.7.2 Emergency Stop Switch 2-10
2.7.3 Audible and Visual Warnings during Operation 2-11
2.7.4 Dual Safety Switches 2-11
2.7.5 Remote Interlock 2-12
2.8 The MOSAIC HP Laser System Complies with following Standards 2-13
2.9 Safety Labeling on the MOSAIC HP Laser System 2-15
2.9.1 Emergency Stop Label 2-17
2.9.2 Danger Label of Laser Radiation 2-17
2.9.3 Laser Radiation Label 2-17
2.9.4 Warning Label for Electrical Shock 2-18
2.9.5 Danger Label of Exposure to Laser Radiation 2-18
2.9.6 Label for the Optical Fiber Applicator 2-18
2.9.7 Nameplate Label 2-19
3.1 Overview 3- 1
3.2 System Specification 3- 1
3.3 Components of the MOSAIC HP Laser System 3- 2
3.3.1 System Console 3- 3
3.3.2 Touch LCD Screen 3- 3
3.3.3 Handpiece 3- 4
3.3.4 Handpiece tips 3- 6
3.3.5 Footswitch and Main Power Area 3- 8
3.4 System Software 3- 9
Chapter 4. Installation
4.1 Overview 4- 1
4.2 Installation Component List 4- 1
4.3 Conditions for Installation 4- 3
4.3.1 Space Requirements 4- 3
4.3.2 Electrical Requirements 4- 4
4.3.3 Environmental Requirements 4- 5
4.4 Installation of the Device 4- 6
4.4.1 STEP 1: Installing the Stand 4- 6
4.4.2 STEP 2: Installing the Handpiece 4- 8
4.4.3 STEP 3: Final Connections 4- 11
4.4.4 STEP 4: Final Checks 4- 12
4.5 Mounting the Handpiece Tip and Starting the System 4- 13
4.5.1 STEP 1: Combining the Handpiece Tip 4- 13
4.5.2 STEP 2: Applying the Main Power 4- 14
4.6 Moving the Device 4- 15
4.6.1 Moving the Device in the Treatment Room 4- 15
Chapter 5. Operation
5.1 Overview 5 - 1
5.2 Pre-operational checklist 5 - 1
5.3 Operation of the MOSAIC HP Laser System 5 - 2
5.3.1 STEP 1: Starting the System 5 - 2
5.3.2 STEP 2: Operation Mode 5 - 3
5.3.3 STEP 3: Thermal Mode 5 - 7
5.3.4 STEP 4: Hair Mode 5 - 9
5.3.5 Use of the Handpiece 5- 11
5.3.6 Turning off the System 5- 12
5.4 Descriptions of the buttons in the Operation Mode Screen 5- 14
5.4.1 Basic buttons of Operation Mode 5- 14
5.4.2 SELECT TREATMENT REGION 5- 18
5.4.3 SELECT AREA 5- 19
5.4.4 Measuring the Area by the GRID MASK 5- 21
5.4.5 FACIAL TREATMENT Mode: F-MODE / P-MODE 5- 22
5.5 Descriptions of SETUP Mode Buttons 5- 24
5.6 Calibration Mode 5- 30
5.6.1 User Calibration Process Procedure 5- 30
6.1 Overview 6- 1
6.2 Cleaning the touch LCD Screen and the System Console 6- 1
6.3 Cleaning the Handpiece 6- 2
6.4 Cleaning the Handpiece Tip 6- 3
6.5 Troubleshooting 6- 5
6.5.1 When the power does not turn the device ON, please check the 6- 5
following items.
6.5.2 When the device is powered on but no laser beam is being 6- 5
radiated, please check the following items.
6.5.3 When the laser beam radiates with insufficient output power, 6- 6
please check the following items.
6.6 Understanding Error and Warning Messages 6- 7
6.7 Periodic servicing 6- 8
Revision 1.1 / 4100110311 6/11
MOSAIC HP
Technology for Beauty and Health
7. 1 Overview 7- 1
Education and Training 7- 1
7. 2 Clinical Guide 7- 2
Introduction of the MOSAIC HP Laser System 7- 2
Fractional Laser Interaction with Skin Tissue 7- 4
What’s New in the MOSAIC HP Fractional Technology? 7- 7
Treatment Parameter 7- 11
7. 3 Clinical Applications 7- 13
7. 4 Treatment Protocol 7- 14
Pre-treatment Preparation 7- 15
Step 1. Consultation 7- 15
Step 2. Preparing the Treatment Area of Skin 7- 15
Step 3. Applying Topical Anesthetic 7- 15
Step 4. Measuring the Size of Treatment Area(in Dynamic Mode) 7- 16
Step 5. Recording and Encoding the Size of the Treatment Area 7- 17
Step 6. Removal of Topical Anesthetic 7- 18
Step 7. Applying Gliding Oil Designated for the MOSAIC HP 7- 18
Laser System
Treatment Procedure except alopecia treatment 7- 19
Step 1. Treatment Parameter Setting 7- 19
Step 2. Operation Technique 7- 19
Caution for the MOSAIC HP Treatment 7- 21
Efficacy Expectations 7- 23
Incidence of Pain 7- 23
Skin Reaction Post Treatment 7- 24
Post-treatment Care 7- 25
Caution and Routine Skin care after Treatment 7- 26
Treatment Interval and Treatment Sessions 7- 27
7. 5 Contraindications 7- 28
7. 6 Treatment Parameter for Each Type of Lesion 7- 29
Treatment Parameter Setting
Chapter 8. Warranty
8.1 Overview 8- 1
8.2 Unpaid Services 8- 1
8.3 Exclusion to unpaid service within the period of the warranty 8- 2
List of Figures
Figure 2.1 Warning for Laser Radiation 2 - 3
Figure 2.2 Laser protection goggles for the operator(left) 2 - 4
/ for the patient (right)
Figure 2.3 Safety features (1) 2 - 9
Figure 2.4 Emergency Stop Switch and Key Switch 2- 10
Figure 2.5 Safety Features (2) 2- 11
Figure 2.6 Remote Interlock 2- 12
Figure 2.7 Locations of Labels on the Front of the Device 2- 14
Figure 2.8 Locations of Labels on the Back of the Device 2- 15
Figure 2.9 Emergency Stop Label 2- 16
Figure 2.10 Danger Label Of Laser Radiation 2- 16
Figure 2.11 Laser Radiation Label 2- 16
Figure 2.12 Warning Label For Electrical Shock 2- 17
Figure 2.13 Danger Labels Of Exposure To Laser Radiation 2- 17
Figure 2.14 Fiber Optic Applicator Label 2- 17
Figure 2.15 Name Plate Label 2- 18
Figure 2.16 Locations Of Labels On The Rear Of The Device 2- 18
A flash GUI is applied to the LCD touch screen so that users can
easily adjust parameters to their optimal settings. The contained
software provides device information and clinical information to
the users.
Symbols Description
CE Marking
Manufacturer
CAUTION
If you adjust or manipulate the device in a way that does not
comply with the instructions given in this manual, then you
may be exposed to dangerous laser beams that may cause
you severe injury or loss of sight.
2.1 Overview
This chapter provides basic safety information for laser generating devices
and cautionary items, covering electrical safety and is intended
characteristics of the laser beams, for users who want to operate this
product. A physician performing a surgical operation using the MOSAIC
HP laser system, an operator using the MOSAIC HP laser system or a
patient or staff member who receives medical service from it must be made
fully aware of the potential danger and safety guidelines associated with the
use of this laser-generating device. In addition, this device should only be
handled by operators or staff members who have attended a certified
training session. Before a surgical operation, the patient should be informed
of the possible danger and cautionary items. Every operator who uses this
device is responsible for the safety issues that may arise while performing a
surgical operation.
1. This device protects itself from possible over current and voltage
problems by utilizing a slow-blow fuse with 250VAC/3A,
125VAC/6A. Safety Extra Low Voltage (SELV) method is used
in the design of this system. The fuse installed in the inner area of
MOSAIC HP laser system successfully blocks off unwanted
power supply.
3. The LED on the handpiece and the tip of the handpiece visually
displays the status of laser-beam radiation that is achieved when
the handpiece is in contact with the patient’s skin. In addition,
while the laser beam is being radiated, an audible continual
‘humming’ sound can be heard and signifies radiation is
occurring.
DANGER
The MOSAIC HP laser system is in compliance with the US
Federal Regulation (21. CFR 1040.10 and 1040.11) of the CDRH
(the Center for Devices and Radiological Health), which is
governed by the FDA (the Food and Drug Administration). In
terms of the CDRH standards, MOSAIC HP laser system belongs
to CLASS IV. This is the most powerful class of lasers used for
medical applications and requires that extra precautions be
taken when using this device.
Figure 2.2 Laser protection goggles for the operator (left) for the patient (right)
DANGER
The laser beam of the MOSAIC laser system is in the form of
infrared rays which are invisible to the naked eye. All persons
present in the operating area must wear protective goggles at
all times. Failure to do so may result in eye exposure to laser-
beam radiation causing severe if not permanent damage and/or
injury. Everyone participating in an operation should wear
goggles with protection from 1550nm wavelength laser beam or
glasses with side protection pursuant to the relevant safety
standard. Prolonged or direct exposure to the radiation produced
by this machine may result in serious injury or even blindness
even if laser safety goggles are worn.
Never allow any liquids to enter the console of the MOSAIC HP laser
system. This may lead to the danger of electrical shock.
The user should not clean the handpiece or repair this instrument while
electrical power is being supplied to this instrument.
WARNING
If persons other than service persons authorized by Lutronic
Corp. attempts to remove the external cover or disassemble
this device, then they may get exposed to the dangerous
laser beam or high voltage or current generated by this
device. Such actions are excluded from the coverage of the
warranty of this product.
DANGER
Never look directly into the laser aperture or at the end of the
handpiece when power is applied. Serious eye injury and/or
blindness could result, even if laser safety goggles are worn.
Never direct the rays of the laser beam onto any object other
than necessary targets. A laser beam reflected from another
surface (regardless of its color) into the eye may cause
injury, permanent damage or even blindness.
Fire retardant fabrics should be used in the surgery to minimize the risk
of fire. Avoid using materials saturated with oxygen.
Before using the laser, make sure all solvent, adhesives and cleansing
liquids have completely evaporated.
DANGER
You should be especially careful when using the laser beam
near flammable material. For instance, if the laser beam hits
an anesthetic, gaseous material such as nitric acids or
oxygen or flammable material such as alcohol or cotton, it
may cause a fire.
DANGER
Never direct the rays of a laser beam at unnecessary targets.
It may damage or burn the object.
2.7.3 Speaker
2.7.3 Handpiece
The key switch is composed of three steps. The key switch’s LED
displays the status of the switch corresponding to each step.
Shifting the key from the 2nd step to the 3rd step is quickly followed by
an automatic return to the 2nd step position; the motion is like that of a
car ignition.
Note: To remove the key, you must turn the system off.
st
Key Switch 1 step
Key Switch LED:
No light on & Power is OFF
nd
Key Switch 2 step
Key Switch LED:
Blue light on / Red blinking
rd
Key Switch 3 step
Key Switch LED:
Red/Blue blinking
CAUTION
Only authorized and trained persons should operate this
device. To prevent unauthorized persons from using this
device, remove the key from the key switch after performing
an operation and store it in a safe place.
To restore operation, rotate the button clockwise (in the direction of the
arrows) until it pops out again and follow the normal startup sequence.
See Chapter 5 for startup procedure.
If you want to start the device again, you should turn the emergency
switch clockwise (direction of arrow) to let the button pop out. After
that, turn the key switch to turn off the device and turn it on again.
A message saying “LASING” will blink in the upper part of the LCD
screen to notify the user that a laser beam is being produced.
External Grounding
Terminal
Standard/
Title Ratification
Regulation
93/42/EEC
As amended by Medical Device Directive 2007
2007/47/EC
1990;A1;1993;A
EN 60601-1 Medical electrical equipment
2;1995;
IEC 60601-1 Part 1: General requirements for safety
A13;1996
Medical electrical equipment
EN 60601-1-2 Part1-2: General requirements for safety
2007
IEC 60601-1-2 –Collateral standard: Electromagnetic
compatibility - Requirements and Tests
Medical electrical equipment
EN 60601-1-4 Part 1-4: General requirements for safety
1996;A1;1999
IEC 60601-1-4 - Collateral Standard: Programmable electrical
Medical systems
Medical electrical equipment – Part 1-6
EN 60601-1-6 General requirements for basic safety and essential 2007
IEC 60601-1-6 performance 2006
- Collateral Standard: Usability
LASER APERTURE
2.9.3 Laser-Beam
Radiation Label
This label warns users of the risk of exposure to direct or scattered laser
radiation if the cover is removed or opened.
3.1 Overview
This chapter provides a general description of the MOSAIC HP Laser
System, including its main components, system control, and technical
specifications.
Component Specification
Laser wavelength 1550nm
Weight 53kg
System console
Touch Screen LCD
Handpiece
Foot switch
Handpiece Hook
Optic Fiber Cable
Stand
Touch Screen LCD
Calibration Port
Handpiece Ball
Casters
The system console contains the control systems and main laser generation
systems. There are no user serviceable parts inside the system console, and it
should never be opened by unauthorized personnel. The main parts of the system
are:
Computer and System Control Module
DC Power Supply
LASER Module
Scanner Module Energy Calibration Port
Handpiece Hook
Handle and Castors
The user interface of the MOSAIC HP Laser System includes a touch LCD
screen system with a Flash Graphical User Interface(GUI)
The size of the LCD is 15 inches (measured diagonally) and provides users
with a clear uncluttered interface and access to all relevant controls and
settings.
The LCD displays information about the current status of the system and
the value of each parameter. The function of each button is explained in
detail in Chapter 5.
Operators may select one of four modes from the initial selection screen.
-CALIBRATION / OPERATION / HAIR / SETUP
The system is designed in such a way that the LCD screen can be folded
away for protection when the device is not being used.
CAUTION
Do NOT touch the Handpiece LCD while moving the device.
Doing so may damage the device. To properly transport the
system, hold the handle and slowly pull the device to the
desired position.
Handpiece tip
CAUTION
Never use the handpiece to move the machine or attempt to
move the equipment by pulling on the handpiece. Doing so may
damage the equipment and void warranty. To move the system
use the installed handle and pull the device into position
slowly.
6X6
• When pulse energy is below
Pulse Rate : 25mJ, pulse density can be
1/2 Speed, 1/3 Speed
adjusted between 50 and 400
5 X 10
Not used. When loaded
system automatically
switches to Static Mode • When pulse energy is above
and makes Dynamic 26mJ, pulse density can be
Mode unavailable. adjusted between 50 and 300
1
10 x 10mm
Non-Contact Tip
Pulse Rate :
Normal, 3/4 Speed,
1/2 Speed,1/3 Speed
12 x 2mm
Hair Tip
Not used. It is available to use in
N/A
HAIR Mode
求
求
求
求
CAUTION
Handpiece tips require constant and careful maintenance for
optimal performance and safety. If the window of the handpiece
tip becomes damaged or rendered unsuitable for operation,
replace it with a new tip immediately. The handpiece tip is no
longer fit to exert optimal lasing and may affect clinical outcome.
External grounding
terminal
The foot switch is designed with a metal cover for durability and to guard
against accidentally system initiation as required by international standards
(IEC60601-2-22).
Before the system will operate, the foot switch cable must be connected to
the foot switch socket located on the main power panel on the back of the
system.
As you step on the foot switch, the touch LCD screen will display the
“READY” message and enters into the standby mode for lasing. Lasing
actually starts only if the handpiece tip touches the treatment area while the
foot switch is pressed.
CAUTION
When moving the MOSAIC HP laser system, the foot switch
must be disconnected from the device in order to prevent
damage to the connector or cable. Never pull on the cable or
subject it to unnecessary stress.
The operator may generate the correct amount of laser radiation by varying
output of the laser beam for each different purpose.
Chapter 4. Installation
4.1 Overview
This chapter describes the installation of process and optimum operational
environment for the MOSAIC HP laser system. Only qualified personnel
and/or people trained by Lutronic Corporation should install, operate or
move the device.
NOTE
Use only original, replacement parts and accessories provided
by Lutronic Corporation, which can be obtained from Lutronic
Corporation or an authorized Lutronic Corporation distributor.
The use of unapproved parts and accessories will void
warranty.
1580mm
583mm
400mm
WARNING
Always check that the correct power supply is available for the
equipment. The connection of a power supply of the wrong
voltage or frequency may damage the equipment and/or result
in injury.
Keep the system away from dust; air-borne dust and fine dust
particles can damage sensitive electrical equipment and/or interfere
with the operation of the device.
1. First ensure that the system is disconnected from the AC power supply.
To locate the optical cable support, open the rear door by inserting the
key into the lock on the rear door and turning it approximately 90
degrees. The optical cable support stand is mounted inside the door.
2. Loosen the red capped bolt as shown in figure 4.3 and undo the
VelcroTM binding straps holding the stand in place.
3. Remove the stand from its mounting, and completely unfold the stand
before tightening the red capped bolt again.
Joint Bolt
4. Remove the plug from the stand mounting bracket and, as shown in
Figure 4.4, align the marks on the outside of the stand base and the
stand mounting bracket while carefully sliding the post at the bottom of
the stand into the socket on the mounting bracket.
Groove on
stand socket
Figure 4.5 Opening the Touch Screen LCD and the Inserted Stand
CAUTION
Never use excessive force as it may result in damage to the
system.
Figure 4.6 Taking out the Handpiece and the Optical Fiber
2. Drape the armored cable containing the optical fibers into the groove of
the stand as shown in Figure 4.7 taking care not to let the handpiece fall.
CAUTION
Never coil or bend the armored cable containing the optical
fibers too sharply or pull or twist them. Such actions can
seriously damage the optical fibers and lead to degradation in
performance or failure. Improper handling will void the warranty.
3. As shown in Figure 4.8, hook the handpiece fixing ball into the
handpiece hook or the stand hook.
4. As shown in Figure 4.9, slot the armored cable into the provided recess
on the top cover and firmly tighten the locator rings.
Optical Fiber Stopper
5. Slot the provided clips onto the vertical section of the stand to keep the
armored cable safely in place as shown in Figure 4.10.
6. Check that the cable is securely located in the support, that there is no
snagging of the cables and that the arm and handpiece can be moved
freely. Close and lock the rear door and remove the key.
1. Remote Interlock
Insert the remote interlock plug into its socket on the main power panel
located on the back of the system. The remote interlock plug has only
two prongs and can only be located in the correct corresponding socket.
It will not fit into the adjacent three-pronged socket which is designed for
the footswitch control cable. To install the remote interlock plug, locate
the two prongs over their sockets and allow the locking ring to snap into
place. Once in place, check that the plug is secure by gently pulling on
the rubber cover of the plug. Do not pull back on the locking ring as this
will release the locking mechanism which allows one to remove the plug.
2. Footswitch
Insert the footswitch connector into the three-pronged socket on the main
power panel as you did remote interlock plug.
2. Lock the casters on the front of the device by pushing firmly down on the
levers.
CAUTION
Before moving the device, secure the handpiece, the optical
fiber cable, the power cord and the foot switch. Use the handle
on the front of the machine to slowly and gently locate the
device in the desired location. After having finished moving the
unit, lock the front and/or rear casters.
After inserting
handpiece tip
Insert
carefully
WARNING
Always wear protective goggles. The handpiece should always
be held face downward. Never look down into the handpiece
when power is being supplied to the system. Avoid holding the
handpiece face-upwards or leaving it unprotected to prevent
the ingress of foreign matter.
Check that the power supply is compatible with the system and
connect the power cord to the outlet on the wall.
CAUTION
When the device is not in use, make sure that the main power
switch is (O) off. If the device is to be left unused for a long
period of time, disconnect the system from the power and
consider removing the remote interlock plug and putting it in a
safe place as this will prevent accidental starting of the system.
While holding the handle on the front of the unit, carefully move the
device to the new location while taking note of the general conditions
for installation previously explained in this chapter.
CAUTION
When moving the device to a different location, we encourage
you to contact Lutronic Corporation and request that a service
representative be sent to uninstall and re-install the system.
A fee may be charged for this service.
Chapter 5. Operation
5.1 Overview
This chapter contains important information about operating the MOSAIC
HP Laser System in a clinical environment and should be read in full before
using the system.
The casters for the device are properly fixed in order to prevent the
device from moving?
Any objects nearby that can reflect light, such as a mirror, should be
covered if not removed from the treatment room.
Power plug is connected to the outlet only after making sure that the
local power supply in the treatment room is 100~240VAC with
proper grounding.
DANGER
Before connecting the power plug of the device to the local
power supply outlet, insert the handpiece tip on the handpiece
to be used. Never look directly into the laser aperture or at the
end of the handpiece when power is applied. Serious eye injury
and/or blindness could result.
When the LCD display is activated, release the key which will
automatically return to the middle position. With the key in the
middle position, the system is operational and the key cannot be
removed. The system will perform various Power On Self Test
(POST) procedures before presenting the main welcome screen.
1 2 3 4
1. CALIBRATION
This is the device to measure the output energy of the system whenever it
is necessary to do so.
2. OPERATION
The operator may press this button to enter into the OPERATION mode.
3. HAIR
The operator may press this button to enter into the HAIR mode.
4. SETUP
The operator may press this button to set user-defined functions such as
simulation mode or buzzer sounds.
1
7 8
11
2 9
5 10 12
3. Press button 7 to select the treatment area (Body/Face) desired Ref 5.4.2
4. Press button 8 to enter to set the size of the treatment area. Ref 5.4.3
5. Select button 11 (Pulse Energy) and 12 (Pulse Rate) to adjust the amount
and speed of laser radiation that is suitable for the treatment area.
6. Step on the footswitch. At this moment, status message (#6) will shift
from “STANDBY” to “READY”.
7. If contacting the treatment area with the handpiece, the status message
(#6) will display “Lasing.” The laser beam will be radiated while the
buzzer sounds.
8. To terminate a session of treatment operation, lift foot off the footswitch
and hook the handpiece on its hanger.
NOTE
Before contacting the treatment area with handpiece,
the operator must first press the foot switch to start the
laser beam radiation.
DANGER
In DYNAMIC mode, the laser beam radiates continuously as the
handpiece contacts the treatment area. Thus it is imperative
that the operator continuously moves the handpiece in
accordance with the treatment area.
STATIC Mode
1. In the initial screen, select the “OPERATION” mode.
2. Press button 4 to select STATIC mode as shown in Figure 5.3.
1
7
11
2 9
5 14 13
3. Press button 7 to select the treatment area (body / face) desired Ref 5.4.2
4. Select buttons 11(Pulse Energy) and 14(Density) to adjust the energy and
density of laser beam radiation.
5. Depress the footswitch. At this time, the status message (#6) menu will
shift from “STANDBY” to “READY”.
6. Contacting the handpiece to the treatment area will cause the status
message (#6) to display “Lasing.” The laser beam of the energy and
density set by the operator will be radiated accompanying buzzer sound.
Removing the handpiece from contact in the skin will shift to “Ready”.
7. Remove the handpiece from the treatment area and make contact with
the treatment area, and then the laser beam will be radiated once again as
shown in step 6. For treatment operation, keep repeating this step.
8. To terminate a treatment session, take foot off the footswitch and hook
the handpiece on its hook.
NOTE
Before putting the handpiece in contact with the patient’s
skin, depress the footswitch. The MOSAIC HP laser system
is designed to initiate radiation only in this order of
operation.
Thermal mode
NOTE
If the handpiece is not contacting the treatment area or if
the operator takes foot off the foot switch, the laser beam
radiation will stop immediately.
HAIR mode
6
1
11
2
4
13
5
4. Push the status message (#6). At this time, the status message (#6)
will shift from “STANDBY” to “READY”.
6. Remove the handpiece from the treatment area and make contact
with the treatment area, and then the laser beam will be radiated
once again as shown in step 4 and 5. For treatment operation, keep
repeating this step.
NOTE
DANGER
It is required the exclusive hair tip is used properly for treating
hair only.
Right
Right Wrong
3. Once the message “It is now safe to turn off the system” shows up, turn
the key switch counterclockwise until it stops, remove the key and keep
it in a safe place to prevent unauthorized and untrained persons from
operating with the device.
1. The operator should only use the emergency stop switch in case of
emergency. Press the emergency stop switch and the device will stop
running immediately.
2. To restore the system, rotate the emergency stop switch clockwise until
the switch clicks into position.
WARNING
Because the latest set parameters are recalled, even though the
device is restarted, the operator must first reset the parameters
before carrying out a new treatment.
NOTE
When turning off the device, the operator must press the
“END” button. Once the message “It is now safe to turn off
the system” is displayed, the operator can turn off the
device. This ensures device safety and stability.
The function of each button in Operation Mode (Refer to Figure 5.11.) and
provided the descriptions of multi-functional buttons separately are
provided below.
1. MODE SELECTION
2. TIP LIFE
The unit for tip life is kJ (Maximum 9999kJ) and the tip life of a new tip is
indicated as “0kJ”. The tip lives are indicated according to each tip
distinguished from the others. Each tip life is derived as the total energy
accumulated as many as the number of times each tip is used.
3. TIP INFORMATION
Indicates whether the handpiece tip is inserted to the handpiece or not. The
information on the size of the tip is also displayed. If the tip is not inserted
in the handpiece properly, then a warning sign will be initiated.
4. OPERATION MODE
This function allows users to choose operation mode. This button converts
STATIC to READY, and vice versa.
STATIC: In this mode, the laser beam with the input density set by the
user will be radiated once each time the handpiece contacts the
treatment area. Ref 5.3.2
DYNAMIC: In this mode, the laser beam will radiate continuously
while the handpiece is in contact with the treatment area. Ref 5.3.2
5. TOTAL ENERGY
This is the total amount of laser-beam radiation and it is the value obtained
by multiplying the total number of spots in the laser-beam and its pulse
energy.
6. STATUS MESSAGE
Displays the status of the device and represents the following information.
Lasing : Status of the laser radiation
Ready : Status of preparation for the laser radiation
Standby: Status of standby for the laser radiation
8. AREA
Enter the size of each treatment area targeted for operation.
Ref 5.4.4 & 5.4.5
9. FACIAL REGION
This screen appears when selecting FACE in the 7. SELECT REGION
menu.
The region(s) display the density of the laser radiation in the treatment
area selected.
Display :
13. DENSITY
In STATIC mode, it is possible to set the density of the laser-beam by using
the Jog Shuttle or (+)/ (-) button. Ref 3.3.4
CAUTION
Since, in DYNAMIC mode, three different modes for the speed
of the laser beam radiation, namely, “Normal, 3/4 Speed, 1/2
Speed, 1/3 Speed” are the values set with respect to the pulse
rate.
Each time button 7 is pressed, the icon and the screen will change.
7 7
NOTE
When restarting the system, all the settings regarding
previously used parameters will be restored but the treatment
area will always remain in Face mode.
Figure 5.13 Entering the region of the treatment area FACE mode
4. Use the (1) Jog Shuttle or press the (+)/ (-) button to enter the
region of the treatment area.
5. Follow the above steps for Left Cheek, Right Cheek, Chin and
Upper Lip and Nose and enter proper sizes of the treatment areas.
Figure 5.14 Input Dialogue for the Size of the Treatment Area in
BODY Mode
2. Enter the size of the treatment area by using the (1) Jog Shuttle or
(+)/ (-) button.
NOTE
When restarting the system, all the settings regarding
previously used parameters will be restored but the treatment
area will always remain in Face mode.
Fore Head
Block
Nose
1. Place the GRID MASK onto the facial area of the patient and then
firmly attach it.
2. The basic size of each treatment area and the number of blocks
located in the outer region should be summed up.
3. With the SELECT AREA function, operator may enter the size of
each treatment area. Ref 5.4.3
NOTE
IT is recommended that the measured size of the treatment area
be recorded in a separate record log, as it could be very useful
for future treatment operation(s). For accuracy, the treatment
area may be marked using the provided treatment pen.
F-MODE
1. F-MODE is the basic mode with the default setting where the
entire facial area is selected as a treatment area.
Total Density
of the area
excluded from
surgery
Total Density of
the area selected
for surgery
WARNING
If no treatment area is selected in P-MODE, then the warning
sign will be displayed on the screen and no laser-beam will be
radiated.
2 5
3 6
4 7
1. MODE SELECTION
CALIBRATION
Energy calibration mode button
MODE
OPERATION
Operation mode button
MODE
2. TREATMENT
P-MODE Setting
In this mode only a certain area of the patient’s face can be selected.
Manipulate the touch screen in the sequence as shown in Figure 5.21
F-MODE Setting
In this mode the entire area of the patient's face as a treatment area can be
selected. Manipulate the touch screen in the sequence as shown in Figure
5.22
1
2
3. SIMULATION
Simulation mode helps operators to become familiarized with the
device in the earlier stages of use.
2
1
NOTE
Once the simulation mode is set, the laser-beam will not be
radiated. But all the treatment parameters and information will
be displayed on the screen as if a real treatment operation is
taking place.
4. INFO
Touch the INFO button to view the location and address of Lutronic
Corporation headquarters as shown in Figure 5.25.
5. HELP
Help button allows users to learn how to operate the device.
6. SOUND
EFFECT SOUND Setting
The Effect Sound function allows users to set the sound-effect ON/OFF.
Manipulate the touch screen in the sequence as shown in Figure 5.26.
2
1
The Laser Sound function allows users to set the laser sound ON/OFF.
Manipulate the touch screen in the sequence as shown in Figure 5.27.
1
2
DANGER
According to IEC60601-2-22, the sound effect when lasing
might be set “OFF” only the simulation mode that real laser
never emits, and it is required that the sound effect when lasing
is set “ON” in the operation mode that real laser emits.
7. SERVICE
This service mode is available for updating the software of the device and
repairing the device. Only persons authorized by Lutronic may access this
mode.
WARNING
If an unauthorized service person uses this service mode, it
may cause malfunction of the device which in turn could result
in severe physical injury. If such an action occurs, it will void
the warranty of the product.
求
1 2 3
1. After removal of the handpiece tip, please insert the handpiece into the
Calibration port located on the device.
NOTE
If the result of calibration displays as CHECK or Error
message, please contact Lutronic or your local Lutronic
agent.
6.1 Overview
Chapter 6 describes the maintenance guidelines, including the periodic
inspection needed to maintain the performance of the device and the
information on error codes to be used in self-testing devices for the device.
CAUTION
When users wish to inspect the device on their own, they must
first turn off the power of the device and then disconnect the
power cord. Attempting to inspect the device while the electrical
power is on may cause severe damage to the device or to the
user.
6.2 Cleaning the Touch LCD Screen and the System Console
Soak a soft pad in non-corrosive cleansing liquid such as isopropyl
alcohol or 90% ethanol and gently wipe the surface areas.
Wipe down the device once again with a clean, dry pad or let it
naturally air dry.
CAUTION
Do not apply cleansing liquid directly to the system console or
the LCD touch screen as it may damage or cause the system to
malfunction.
拠居 Use a soft, lint-free dry cloth. Soak it in alcohol and clean the entire
area of the handpiece except its window.
挙居 When alien substances are found on the lens of the handpiece, blow
them away with an air blower carefully so that the window is not
damaged.
渠居 If the above procedure does not eliminate all the alien substances,
then as shown in Figure 6.1, soak a cotton bud in 90% alcohol and
wipe out the area thoroughly.
Window
虚居 After cleaning, shine a light onto the tip of the handpiece and
examine it.
NOTE
If you are unable to eliminate alien substances using the above
protocol, then:
Soak a lens paper or cotton swab in ethanol. Allow one drop
of ethanol to fall onto the top of the lens. Soak alien
substances in ethanol for 3 to 5 seconds.
Once the alien substance is thoroughly soaked, use a
cotton swab to remove it.
1. Use a soft, lint-free dry cloth and soak one end of it in alcohol.
3. If the above procedure does not eliminate all of the alien substances,
then as shown in Figure 6.2, soak a cotton swab in 90% alcohol and
wipe the area thoroughly.
Window
4. After cleaning it, shine a light onto the tip and examine it.
5. Store the tip in the provided case until its next use.
NOTE
If you are unable to eliminate alien substances using the above
protocol, then:
CAUTION
The handpiece tips should be sterilized before applying it
to the surgical area of a patient. Improper handling of the
handpiece tip may incur damage as a result.
NOTE
When the user irradiates with the laser beam, irradiation is not
only possible through the surface of the tip window but also on
the out sides of the tip window. This is visible as black spots
will appear on the surface of the tip window. In such cases,
please contact the Lutronic service representative immediately
for advice.
6.5 Troubleshooting
Chapter 6 describes the maintenance guidelines, including the periodic
inspections, needed to maintain quality performance of the device, and the
information on error codes to be used in self-testing devices for the device.
6.5.1 When the power does not turn the device ON, please check the
following items.
Check that the power cord is correctly inserted into the main power
outlet.
Check to see if the handpiece tip is making good contact with the
surface area.
Check to see if the remote interlock is properly inserted into the socket.
CAUTION
If the laser beam is not being radiated for a long period of time,
turn off the device first and then turn it on again after two
minutes. Even if the problem persists, do not disassemble the
device or take inadequate actions on your own. Always contact
your authorized distributor or contact Lutronic for assistance.
6.5.3 When the laser beam radiates with insufficient output power,
please check the following items.
Check to see whether the handpiece window is damaged or dirty.
If alien substances reside on it, clean the window first before using
it again.
If the problem persists after taking the above actions, contact your
authorized distributor or contact Lutronic for assistance.
The system continuously monitors its condition and notifies the user of any
problem by displaying error messages such as that shown in Figure 6.3
listed in the following table.
CAUTION
You are advised to take proper actions for the nature of each
error message. If the error persists after taking proper
actions, contact your authorized distributor or Lutronic for
assistance. If a person, not properly authorized by Lutronic,
opens the cover of the device or takes an improper action,
then the purchaser will be ineligible for any unpaid services
guaranteed by the product warranty and may even lead to
fatal damage to the device or even cause severe physical
injury.
7.1 Overview
The MOSAIC HP laser system allows users to accurately and precisely adjust
number of micro beams from 300-2,000 per unit area (described as spots/cm2) of
the skin and penetrate laser beams from the epidermis to the deep dermis. The
non-treated skin tissue surrounding the MNCs is stimulated by micro necrotic
columns promoting rapid recovery of the entire skin by biostimulation reaction of
normal tissue.
Because less than 20% of the skin surface area is treated per each treatment,
which is a fraction of the target area, more than 80% of the adjacent viable tissue
is left intact, the downtime is minimized and severe side effects hardly occur.
Therefore, the treatment itself is much less destructive and promotes regeneration
of both epidermis and dermis, showing excellent total skin regeneration effects
compared to that of ablative skin rejuvenation method.
CO2
(10,600nm)
Er: YAG
(2940nm)
Er:Glass
(1550nm)
Although the MOSAIC HP laser system causes micro thermal damage in the skin,
that is necrosis and coagulation, skin renewal is stimulated via the rapid
remodeling response from surrounding normal tissues. MOSAIC HP does not
cause any ablation of skin due to vaporization of epidermal tissue during
treatment. After the treatment, stratum basale multiplies new cells pushing
epidermal debris formed from damaged epidermis toward stratum corneum.
During the healing process, the epidermal debris safely protects against invasive
damage and infection from outside until it naturally peels off, so that rapid wound
healing is achieved. Therefore, within 24 hours after laser was irradiated, re-
epithelialization on the damaged epidermal skin can be achieved through rapid
proliferation and movement of cells from stratum basale and patients won’t feel
any inconvenience for normal daily living. At the same time, the micro beams that
have penetrated into the deep dermis will induce synthesizing and remodeling of
collagen, while simultaneously replacing micro necrotic collagen and initiating
photobiostimulation to the surrounding normal fibroblasts. As a result, we may
∫term skin rejuvenation effects. (Refer to Figure 7.2)
expect long∫
The MOSAIC HP laser system has shown promotable effects in the various types
of alopecia - Androgenic alopecia, female pattern alopecia and alopecia areata.
(Figure 7.3) Followings are strong points of depilatory treatment with Fractional
Laser MOSAIC:
Firstly, hair re-growth would be expected by the mechanical stimulation with
depth control of micro beam.
Secondly, it does not interrupt to normal hormones inside the body. Therefore, it
would be a safe treatment for women who are contraindicated to systemic
medications for hair re-growth.
Thirdly, a synergy effect can be expected with the combination therapy with other
treatments for hair re-growth.
Lastly, this is a safe treatment which does not cause damages to normal hair and
normal hairs are not cut by the treatment.
In Dynamic Mode, the MOSAIC HP Laser System can accurately calculate and
automatically display the Total Density which is computed as dividing total
3. Thermal Mode
Though it has been impossible to change pulse width in existing Er:Glass
fractional lasers, MOSAIC HP laser system supplies thermal mode which has
wider pulse width compared to the normal mode. Pulse width at Thermal
Mode is double that of Normal Mode and it will maximize efficacy of Dermal
Heating. MNCs (Micro Necrotic Columns) are formed as thinner and deeper
shapes in the thermal Mode than in the normal mode. Lateral damaged areas
in the thermal mode are wider and shallower than that of the normal mode.
(Figure 7.4) On the basis of indications, normal or thermal mode would be
selected. Thermal Mode which has more efficient in Dermal Heating effect is
recommended for Skin Rejuvenation treatment. And normal mode is
recommended for treating dermal scars. Coverage percentage will be increased
relatively in the thermal mode that it is recommended to reduce the total density
of thermal mode to half of the normal mode considering the PIH occurrence in
Asian skin.
12x2
12x2 mm Hair tip 求10x10 mm Non-
Non-contact tip
Treatment Parameter
The most important treatment parameters used in the MOSAIC HP laser system
are Pulse Energy(mJ) that determines the penetration depth and diameter of the
Micro Necrotic Columns (MNCs), and Total Density(spots/cm2) that determines
the density of the micro beam radiated to each unit area of a patient’s skin. As the
Pulse Energy increases, the laser beam can penetrate more deeply through the skin
tissue and the diameter of the MNC increases as well. Therefore, operator should
avoid excessive thermal damage by relatively decreasing Total Density whenever
use high Pulse Energy in each treatment session, targeting the same treatment area
which is 5-20% in general). Since both Pulse Energy and Total Density are
important clinical parameters, they should be adjusted delicately to achieve
optimal clinical result while minimizing side effects.
Even when treating same kinds of lesions, users should apply different pulse
energies according to its particular location on the body such as face, neck, and
extremities, and should also consider different approach according to skin types
and different age groups. Generally, body area recovers slower than facial area,
and tends to involve post treatment complications easily. Therefore choice of
appropriate Pulse Energy should be very critical. The younger patients, the
more sensitive to pain induced by laser.
level, higher Total Density yields more dramatic effects due to larger area
stimulated by laser beams per session. However, since higher Total Density also
means dense spread of micro laser beams, there is higher risk to incur burning
to tissue due to overlapping of thermal damage (especially in skin type V and VI).
To prevent this side effect, it is advisable to avoid high Total Density when
applying high energy since higher energy involves wider thermal damage to the
area.
Generally, lower energy level is set in combination with higher Total Density
and vice versa. In the instance of achieving high Total Density at high Pulse
Energy, it is advisable to decrease Unit Density per pass (in Dynamic Mode) or
per stamp (in Static Mode), and repeat multiple passes or stamps instead in
order to achieve the same high Total Density. Intermittent cooling in between
each successive pass also minimizes thermal damage.
The MOSAIC HP Laser System is basic device for various scars and total skin
rejuvenation, and it can be applied to any of the following applications;
Various types of scars: acne scars, surgical scars, traumatic scars, burn
scars, keloid and striae distensae
Wrinkles: periorbital, periorbital wrinkles
Photo-damaged skin: rough and uneven skin tone and texture, large pores,
excessive secretion of sebum
Melasma
Pigmented lesions such as dyschromia, dyspigmentation, seborrheic
keratosis, etc.
Redness and telangiectasia
Inflammatory acne
Eyelid laxity
Total skin rejuvenation of face and off-face area (hand, neck, chest)
Hair Re-growth (New!)
The treatment procedure using the MOSAIC HP Laser System includes the
followings; Recording a patient’s clinical history, Accurate diagnosis of the lesions,
Notifying the patients of information on laser treatment, Informing patient of
expected outcomes and potential post-treatment side effects, informing patient of
pre-treatment cautions and general laser treatment procedure including anesthesia,
post-treatment care; prescribing post-treatment medication; notification of
guidelines to be followed after treatment; and finally follow-up for monitoring the
recovery process.
Pre-treatment Preparation
Step 1. Consultation
Before performing The MOSAIC HP Laser Therapy, every doctor should
first check presence of contraindication to laser therapy by reviewing the
patient’s symptoms, clinical history, body condition, history of drug abuse
in order to determine laser therapy is appropriate or not. Patients should
be informed of potential complications and expected outcomes of the
therapy.
Number of
1 Block=1cm2
blocks in pre-
counted shade
area of each
cosmetic unit
To enter the size of the treatment area, select and touch a certain cosmetic
unit on figure of face on the screen. The selected area will appear in orange
and the color of the input field corresponding to the selected area will also
change. To adjust treatment area size, press the + or – key or use the
∫shaped ring on the right.
circle∫
Before starting the treatment, you should set the treatment parameters
based on location of treatment area, size and characteristics of lesions.
First of all, select the Operation Mode in the main operation screen. Next,
in dynamic mode, select Pulse Rate (Hz) which means micro beams
radiation speed (Normal, 3/4 speed, 1/2 Speed, 1/3 Speed) and in static
mode, select the density (50, 75, 100, 200, 300, 400, 500 spots/cm2) which
means numbers of beams per square centimeter. Then adjust the Pulse
Energy based on the depth of each lesion. After that, start the treatment.
Figure 7.8
Multiple Moving Patterns and Facial Areas Needed Special Cautions
When treating a relatively small area in dynamic mode, the Total Density
count displayed increases quickly. Thus use small sized tip (6X6mm) and
set the speed of the laser beam radiation to "1/2 Speed" or "1/3 speed" to
reduce any unexpected thermal damage due to thermal overlapping.
Repeat pass on the entire treatment area until the Total Density per unit
area reaches the desired value. Frequently monitor skin reaction of the
patient and sensitivity to pain throughout treatment and adjust the Total
Density accordingly. Sensitive areas should be exposed to less Total Density
compared to other areas. Carefully observe skin reaction at the end of each
pass, then determine necessity of additional pass. Apply cooling frequently
in between each pass.
It is convenient to use static mode when treating sparsely-distributed scars.
To avoid unnatural looking of demarcation between treated areas and
untreated areas in static mode, finally apply 2-3 passes at the same energy
setting in dynamic mode to create smooth and natural looking. For broad
distributed lesions such as acne scars, dynamic mode is more effective than
static mode.
In general, treat specific lesions such as scars with high Pulse Energy at
first, and then treat entire treatment area with skin rejuvenation
parameter.
For detailed information on Pulse Energy and Total Density for each
indication, please refer to our the MOSAIC HP Clinical Parameter
Guidelines
Cautions for the MOSAIC HP Treatment
The treatment parameters for each lesion should be adjusted properly based on
the skin reaction of each patient. When performing the first session, you should
use relatively conservative treatment parameters. From second session, adjust
those parameters gradually based on the skin reaction (persistent time of
erythema and edema) and clinical effects.
When treating the perioral or periorbital area, avoid very high Pulse
Energy and Total Density. In particular, the perioral area requires special
attention as herpes simplex or PIH is likely to occur. Be sure to carefully
examine skin reaction of the treated area for 1-3 days post treatment. In
general, for the same level of Pulse Energy, the Total Density should be
around 1/3 of other areas.
When treating small areas such as nasal or periorbital area in dynamic
mode, use 6*6 handpiece tip at 1/2 or 1/3 speed.
According to the location of treatment area(i.e. face, neck) and the
characteristics of the lesions, the Pulse Energy should be properly adjusted.
In addition, based on the observed skin reaction (i.e. erythema or edema),
Total Density should also be properly adjusted.
When increasing Pulse Energy upto 25mJ or above in static mode, set
Total Density at 200spots/cm2, then reach desired Total Density by
repeating stamping. Always examine skin reaction after lasing two
successive shots over the same area, and apply cooling for about one
minute.
When treating acne scars accompanied by active inflammatory acne, active
acne may improve. However occasional aggravation and worsening may
also occur. Therefore special caution is required.
For high pulse energy treatment, Total Density need to be devided into 2
treatments with 50 spots/cm2 per each. It is recommended not to exceed
Total Density of 100 spots/cm2. After the first pass, clinician should check
the patient’s skin reaction carefully and decide to continue or cease the
procedure. For avoiding overlapping, lased area may be slightly changed
in the second pass.
Although the energy is high, density is very low in high pulsed energy
treatment. Following the guidance of treatment protocol, blisters would not
be occurred in high pulse energy treatments. Due to the heat in the skin,
a sufficient cooling after the MOSAIC HP treatment is required.
Efficacy Expectations
Incidence of Pain
The MOSAIC HP laser system creates micro laser beams that penetrate into the
deep dermis and some pain is expected during treatment. However, the level of
sensitivity to pain differs from one patient to another. The deeper the location of the
lesion, the higher the Pulse Energy should be used and thus patients may feel more
pain. And Dynamic Mode induces more pain than Static Mode. The pain is often
described as if a fine needle is irritating the skin with a slight burning sensation.
Lowering Pulse Rate from “Normal” to “3/4” , “1/2” or “1/3” speed helps to relieve
such irritation. In general, topical anesthesia is applied for more comfortable
treatment for patient. To minimize post-operative flushing and pain, use an air
cooler to blow cool air onto the treatment area or apply an ice pack to cool down the
skin.
For approximately one week post treatment, the epidermal debris on the stratum
corneum gives the treated skin area a bronze and rough appearance. After that the
epidermal debris naturally exfoliates as the newly proliferated epidermal tissues
replace the dead tissue. The skin naturally peels off, the skin color will become
brighter.
Pain
Largely reduced via cooling mask post treatment.
Edema
Disappears along with erythema after 1-3 days.
specific treatment areas, and induce burning. Pay special attention to such
relatively small treatment areas.
Herpes Simplex
Perioral areas are easily vulnerable to herpes simplex after laser therapy.
Thus patients should be asked to indicate any history of herpes simplex in
perioral area. If relatively high Pulse Energy and Total Density are used in
one session, please prescribe oral anti-virus medication for approximately 3
days post treatment to prevent herpes simplex.
Post-treatment Care
Sooth treated area during and after the surgery by applying an air cooler
or ice pack.
Use a moist cooling mask pack to calm down, erythema, flushing and pain
of the treated skin.
After calming down the skin, apply some moisture cream and sun-block
cream and allow patient to go home.
Right after treatment, the burning and stinging feeling may persist for
about one to two hours.
Immediately after treatment with the MOSAIC HP Laser System, the
patients can wash their faces, make-up or even shave.
Treatment Interval
The treatment interval should be properly adjusted depending on the type of
lesion, the treatment area (face, neck, arms and legs, etc.), and the relative
location within the skin (epidermis or dermis), and the percentage of entire
skin area the Total Density occupies at each session. As a lesion is located in
epidermis and superficially, or as treatment goes with low Total Density, a
shorter treatment interval is possible.
7.5 Contraindications
If simple herpes has occurred on any location in the treatment area within
the past one month.
Photo-sensitive skin.
∫treatment
Patients who are unable or unwilling to follow the post∫
instruction.
Acne Scar
Treat cheek, chin and nose areas which contain in acne scars
For more information about how to operate, please refer to 7.4 MOSAIC
HP Treatment Protocol.
Hypertrophic Scar
Pulse Energy: < 10mJ; Total Density: 1000~2000 spots/cm2
Treatment interval : 3-4 weeks
The number of sessions required: min 3-5 sessions
Treatment Method:
For fresh scars less than 3 moths, combination therapy with Pulse
Dye Laser is recommended.
Keloid
Pulse Energy: 7-12mJ; Total Density: 500~1000 spots/cm2
Treatment interval : 3-4 weeks
The number of sessions required: min 3-5 sessions
Striae Distensae
Pulse Energy: 18-24J; Total Density: 300~500 spots/cm2
Treatment interval : 4-6 weeks
The number of sessions required: min 3-5 sessions
Treat cheek, chin and nasal areas which contain deep wrinkle
and sagged skin
In order to improve skin tone, large pores and wrinkles, treat with
lower pulse energy (18-25mJ) and lower total density
(300spots/cm2) than in the static mode
In order to achieve dramatic clinical results, treatment with higher pulse energy
and lower total density is recommended. Decrease in Total Density allows users to
minimize the patient’s pain and erythema, or edema would be reduced after the
treatment.
Oily skin treated in high energy may not cause any problem. For thin skin type,
it would be better to start from lower energy and increase it according to the
patient’s skin reaction. In case of patients who have thin and bright skins, there
would be pin point bleeding.
Treat with 1~2 mm space between stamp by stamp because edema would
be occurred around stamps. Select the most conservative parameter for the
first session, and then gradually increase it for the following sessions
according to the patient’s skin reaction. For high pulse energy treatment,
Total density need to be devided into 2 treatments with 50 spots/cm2 per
each. It is recommended not to exceed Total Density of 100 spots/cm2
Treat with 2~3 mm spaces among Treat like filling in the space which was
stamps in the 1st pass. not treated in 1st pass.
Figure 7.10 High Pulse Energy Tx in Static Mode
求
< 2nd step : Dynamic Mode >
求
Treat generally whole face including treated area at the 1st step. In the
second step, the energy should be controlled by skin types. It is
recommended to use less than 18mJ for patients who have thin and dry
skins. For patients who have oily and wrinkly skins, 22-25mJ in dynamic
mode are recommended. Since skins have been stimulated and have some
edema and erythema, use reduced pulse rate at 1/2 or 1/3 speed. In order
to reduce patient’s inconvenience, do not exceed Total density of 300
spots/cm2 at the 2nd step.
求求
Recommended Total Density in the normal mode
求
Recommended Total Density in the thermal mode
求
求
Hair Re-growth
Pulse Energy: 4-6mJ
Density: 300~500 spots/cm2 (3-4 passes without overlapping)
Total Density: 900-2000 spots/cm2
Treatment interval and the number of sessions required
: 1-2 weeks, 3 months standard course
Press “standby” button and change into “ready” state. For lasing, press
foot switch and contact handpiece tip to the scalp.
Treatment with hair tip will be available regardless of skin contact.
Handpiece tip should always remain perpendicular to the skin surface, and
emit the beams in one direction to avoid lasing to normal hairs. Emit
beams directly to the scalp for minimizing energy loss.
Chapter 8. Warranty
8.1 Overview
Chapter 8 describes unpaid services available for those who use the
instrument properly and exclusions from such unpaid services are provided
by the warranty of this product.
WARNING
In order to maintain the coverage of unpaid services, only
service persons authorized by Lutronic should be allowed to
configure, modify or inspect the instrument. If the instrument is
used for purposes other than originally intended or not in
compliance with the instructions given in this manual, then the
user shall be ineligible for any type of unpaid services. Our
customers are strongly advised to be fully aware of all the
details of this manual. Lutronic reserves all rights and
responsibilities for judging the nature of damages to the
product and the possible causes for such damages. Such a
judgment made by Lutronic shall be deemed to be finalized and
cannot be overridden.
If you pull out the optical fiber or its handpiece with excessive
force to move the instrument, which results in damages to the
optical fiber, then you shall not be eligible for unpaid service
covered by this letter of warranty.
Consumable parts like the handpiece tip, which are necessary for
consistent maintenance of the instrument, are not covered by
unpaid services.
This letter of warranty can replace any form of implicit or explicit warranty
agreed by interested parties. However, the sales agent shall not provide any kind of
warranty for the condition of the laser or its marketability for any purpose.
Doctor: ____________________________
Address: _____________________________________________________________
E-mail: _____________________________
Model: __________________
Date of purchase:________________
Important! To validate this warranty, please completely fill out the above
items and email it to the address or number below within fifteen days of
the installation.
Lutronic Corporation.
E-mail: office@lutronic.com