Date: 06/08/2016 MT‐PIPCAL‐1.3
Revision Date: 06/07/2016
Metals Group
Revision Author(s): M . Thompson Pipette Calibration Check
INSTRUCTIONS:
METALS GROUP PIPETTE CALIBRATION CHECK SPREADSHEET Corrective Action Required for Failing
Enter the water temperature to the nearest
Pipets
Analyst: Balance: 0.5 deg. and the weight of the 5 replicates
Temp.°C Thermometer: of water dispensed by the pipette.
See Corrective Action Tab for
Date: The density is automatically entered
Resolution
Water Density (in g/mL) upon entering the temperature.
Measurement 1 Measurement 2 Measurement 3 Measurement 4 Measurement 5 Adjustable Pipettes; If one of the bracketed
Pipet Mean Vol. RSD of Vol. % Difference % RSD weights exceeds tolerance, all weight for that
Pipet ID # Vol. in uL % Difference
Location Weight (g) Vol. (uL) Weight (g) Vol. (uL) Weight (g) Vol. (uL) Weight (g) Vol. (uL) Weight (g) Vol. (uL) (uL) (uL) PASS/FAIL PASS/FAIL pipette will need corrective action.
Fixed volume Pipette [ 100uL - 2500uL ]
% Difference for Fixed volume Pipette
100uL - 2500uL 0.60%
Adjustable volume Pipette [ 10uL - 100uL ] % Difference for Adjustable volume Pipette
10uL - 100uL
10 3.00%
50 1.00%
100 0.80%
Adjustable volume Pipette [ 100uL - 1000uL ] % Difference for Adjustable volume Pipette
100uL - 1000uL
100 3.00%
500 1.00%
1000 0.60%
Adjustable volume Pipette [ 1000uL - 10000uL ] % Difference for Adjustable volume Pipette
1000uL - 10000uL
1000 3.00%
2000 3.00%
5000 1.00%
10000 0.60%