Software Upgrade SOP


——FinecareTM FIA Meter Ⅲ Plus

Racing For Life


1 Prepare USB flash desk

CONTENTS 2 Change System Mode

3 Upgrade software
4

01 Prepare
USB flash desk
Prepare USB flash desk 5

1. Prepare an USB flash desk.


Prepare USB flash desk 6

2. Insert the USB Flash Desk into the computer.


Prepare USB flash desk 7

3. Create a new folder under the root directory in the USB Flash Desk,
named“MultiChannelUpgrade”.
Notice: The name must be exactly the same.
Prepare USB flash desk 8

4. Put the software files into the“MultiChannelUpgrade”. After that the Flash Desk
can be used.
9

02 Change System
Mode
Change System Mode 10

1. Here are two ways to Change System Mode. IF at“User login”page, input
Username“Fac”, Password “123456”, and then click“log in”.
Change System Mode 11

2. If at the homepage, choose “Settings”-“Mode”.


Change System Mode 12

3. Choose “Factory”, then click “Confirm”.


Change System Mode 13

4. Enter the User name:“Fac”, and the Password:“123456”, then click “Confirm”.
14
Record value
5. Before upgrade,we enter 【Function】-【Maintain】 at Fac mode
15

Record value
6.At 【Para Config】-【Control unit】-【Control unit】 Record 【Optical
sample start posi】 value ,here is 290.
16
Record value
At 【Other Set】,record 【Fluoresence coefficient】 value,here is 1
17

03 Upgrade Software
Upgrade Software 18

1. Insert the USB Flash Desk into the instrument.


Upgrade Software 19

2. Click“Function”-“System”-“Upgrade”.
Upgrade Software 20

3. Click “Confirm”.
Upgrade Software 21

4. Click “Confirm”.
Upgrade Software 22

5. Please wait for the progress bar to reach the end, don’t click any button or power
off.
Upgrade Software 23

6. When the progress bar reach the end, click “Confirm”and restart the instrument.
Upgrade Software 24

7. After restart, please upgard the software once again, repeat from step 2 to step 6.

Notice: The software should be upgraded twice to ensure the upgrade is successful.
Upgrade Software 25

8. After restart again, click“Function”-“System”-“About”.


Upgrade Software 26

9. Check“Complete Version”.
27

Check value
10.Recheck the value 【Optical sample start posi】,here is still 290,right. If
not same as before, need rewrite and save the value
28

Check value
11.Recheck the【Fluoresence coefficient】value ,here is 1 same as before ,If
not same as before, need rewrite and save the value
THANK YOU!