You are on page 1of 50

GE Healthcare

gehealthcare.com

Technical
Publication

5161957-100
Revision 4

CT/e, ProSpeed AI/FI Series


CT/e Dual, ProSpeed AII/FII Series
O2 PC TO LINUX PC865 UPGRADE IN-
STRUCTION

Copyright © 2007-2010 by General Electric Company


CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

REV 4 5161957-100
• 本维修手册仅提供英文版本。
警告 • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
(ZH-CN)
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。

VÝSTRAHA • TENTO PROVOZNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.


(CS) • V PŘÍPADĚ, ŽE EXTERNÍ SLUŽBA ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V
JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA
ÚKOLEM ZÁKAZNÍKA.
• NESNAŽTE SE O ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI
TENTO PROVOZNÍ NÁVOD A POCHOPILI JEHO OBSAH.
• V PŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT K PORANĚNÍ
PRACOVNÍKA PRODEJNÍHO SERVISU, OBSLUŽNÉHO PERSONÁLU NEBO
PACIENTŮ VLIVEM ELEKTRICKÉHOP PROUDU, RESPEKTIVE VLIVEM
MECHANICKÝCH ČI JINÝCH RIZIK.

• DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.


ADVARSEL
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
(DA)
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL HAR VÆRET KONSULTERET OG ER FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE
SKADE PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATØREN ELLER PATIENTEN.

• DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS


WAARSCHUWING VERKRIJGBAAR.
(NL)
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS
DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.

a
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
(EN) OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO
PROVIDE TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE
SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

• KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.


HOIATUS
• KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST
(ET) ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT
KÄESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU
SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE
VÕI MUU OHU TAGAJÄRJEL.

• TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.


VAROITUS
• JOS ASIAKKAAN HUOLTOHENKILÖSTÖ VAATII MUUTA KUIN
(FI) ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
HUOLTOHENKILÖSTÖN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN
VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.

• CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.


ATTENTION
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE
(FR)
AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE
LE FAIRE TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE
LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ
LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES
DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

b
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

• DIESE SERVICEANLEITUNG EXISTIERT NUR IN ENGLISCHER SPRACHE.


WARNUNG
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST
(DE) ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU
SORGEN.
• VERSUCHEN SIE NICHT DIESE ANLAGE ZU WARTEN,
OHNE DIESE SERVICEANLEITUNG GELESEN UND VERSTANDEN ZU HABEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH STROMSCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN
KOMMEN.

• ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ.


ΠΡΟΕΙΔΟΠΟΙΗΣΗ
• ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ
(EL) ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.

• EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ


FIGYELMEZTETÉS EL.
(HU) • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.

• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.


AÐVÖRUN
(IS) • EF AÐ ÞJÓNUSTUVEITANDI VIÐSKIPTAMANNS ÞARFNAST ANNAS
TUNGUMÁLS EN ENSKU, ER ÞAÐ SKYLDA VIÐSKIPTAMANNS AÐ SKAFFA
TUNGUMÁLAÞJÓNUSTU.
• REYNIÐ EKKI AÐ AFGREIÐA TÆKIÐ NEMA AÐ ÞESSI ÞJÓNUSTUHANDBÓK
HEFUR VERIÐ SKOÐUÐ OG SKILIN.
• BROT Á SINNA ÞESSARI AÐVÖRUN GETUR LEITT TIL MEIÐSLA Á
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS FRÁ RAFLOSTI,
VÉLRÆNU EÐA ÖÐRUM ÁHÆTTUM.

c
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

• IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO


AVVERTENZA IN INGLESE.
(IT)
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO
DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE
COMPRESO IL CONTENUTO.
• IL NON RISPETTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO
ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

• このサービスマニュアルには英語版しかありません。

• サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業

(JA) 者の責任で行うものとさせていただきます。

• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。

• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、
感電や機械的又はその他の危険により負傷する可能性があります。

• 본 서비스 지침서는 영어로만 이용하실 수 있습니다.


경고 • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것
(KO) 은 고객의 책임입니다.

• 본 서비스 지침서를 참고했고 이해하지 않는 한은 해당 장비를 수리하려고 시도하지


마십시오.

• 이 경고에 유의하지 않으면 전기 쇼크, 기계상의 혹은 다른 위험으로부터 서비스 제공


자, 운영자 혹은 환자에게 위해를 가할 수 있습니다.

d
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

• ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.


BRĪDINĀJUMS
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
(LV) VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT
TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
• ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.

ĮSPĖJIMAS • ŠIS EKSPLOATAVIMO VADOVAS YRA PRIEINAMAS TIK ANGLŲ KALBA.


(LT) • JEI KLIENTO PASLAUGŲ TIEKĖJAS REIKALAUJA VADOVO KITA KALBA – NE
ANGLŲ, NUMATYTI VERTIMO PASLAUGAS YRA KLIENTO ATSAKOMYBĖ.
• NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS, NEBENT
ATSIŽVELGĖTE Į ŠĮ EKSPLOATAVIMO VADOVĄ IR JĮ SUPRATOTE.
• JEI NEATKREIPSITE DĖMESIO Į ŠĮ PERSPĖJIMĄ, GALIMI SUŽALOJIMAI DĖL
ELEKTROS ŠOKO.
• MECHANINIŲ AR KITŲ PAVOJŲ PASLAUGŲ TIEKĖJUI, OPERATORIUI AR
PACIENTUI.

ADVARSEL • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.


(NO) • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER
DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.

• NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU


OSTRZEŻENIE ANGIELSKIM.
(PL) • JEŚLI DOSTAWCA USŁUG KLIENTA WYMAGA JĘZYKA INNEGO NIŻ
ANGIELSKI, ZAPEWNIENIE USŁUGI TŁUMACZENIA JEST OBOWIĄZKIEM
KLIENTA.
• NIE PRÓBOWAĆ SERWISOWAĆ WYPOSAŻENIA BEZ ZAPOZNANIA SIĘ I
ZROZUMIENIA NINIEJSZEGO PODRĘCZNIKA SERWISOWEGO.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOŻE SPOWODOWAĆ
URAZY DOSTAWCY USŁUG, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA ELEKTRYCZNEGO, ZAGROŻENIA MECHANICZNEGO BĄDŹ
INNEGO.

e
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

• ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL


ATENÇÃO EM INGLÊS.
(PT) • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE
TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTÉNCIA TÉCNICA
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA
DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS,
MECÂNICOS OU OUTROS.

• ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.


ATENŢIE • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ
(RO) LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ
FURNIZEZE O TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.

• ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДЛАГАЕТСЯ ТОЛЬКО


ОСТОРОЖНО! НА АНГЛИЙСКОМ ЯЗЫКЕ.
(RU) • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО
РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ, А НА КАКОМ-ТО ДРУГОМ ЯЗЫКЕ,
КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО ОБЕСПЕЧИТЬ ПЕРЕВОД.
• ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО
ОБРАТИТЕСЬ К ДАННОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В
НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ ТРЕБОВАНИЙ ДАННОГО ПРЕДУПРЕЖДЕНИЯ МОЖЕТ
ПРИВЕСТИ К ТОМУ, ЧТО СПЕЦИАЛИСТ ПО ОБСЛУЖИВАНИЮ, ОПЕРАТОР
ИЛИ ПАЦИЕНТ ПОЛУЧАТ УДАР ЭЛЕКТРИЧЕСКИМ ТОКОМ,
МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ ДРУГОЕ ПОВРЕЖДЕНИЕ.

• TENTO NÁVOD NA OBSLUHU JE K DISPOZÍCII LEN V ANGLIČTINE.


UPOZORNENIE
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
(SK) ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA O OBSLUHU ZARIADENIA SKÔR, AKO SI NEPREČÍTATE
NÁVOD NA OBLUHU A NEPOROZUMIETE MU.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA
ELEKTRICKÝM PRÚDOM, DO MECHANICKÉHO ALEBO INÉHO
NEBEZPEČENSTVA.

f
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

ATENCION • ESTE MANUAL DE SERVICIO SOLO EXISTE EN INGLES.

(ES) • SI ALGUN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA


QUE NO SEA EL INGLES, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCION.
• NO SE DEBERA DAR SERVICIO TECNICO AL EQUIPO,
SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE
OTRA NATURALEZA.

• DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.


VARNING
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
(SV)
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA
I SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA
FAROR.

DİKKAT • BU SERVİS KILAVUZUNUN SADECE İNGİLİZCESİ MEVCUTTUR.


(TR) • EĞER MÜŞTERİ TEKNİSYENİ BU KILAVUZU İNGİLİZCE DIŞINDA BİR BAŞKA
LİSANDAN TALEP EDERSE, BUNU TERCÜME ETTİRMEK MÜŞTERİYE DÜŞER.
• SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE
ETMEYİNİZ.
• BU UYARIYA UYULMAMASI, ELEKTRİK, MEKANİK VEYA DİĞER
TEHLİKELERDEN DOLAYI TEKNİSYEN, OPERATÖR VEYA HASTANIN
YARALANMASINA YOL AÇABİLİR.

ПРЕДУПРЕЖДЕНИЕ • ТОВА УПЪТВАНЕ ЗА РАБОТА Е НАЛИЧНО САМО НА АНГЛИЙСКИ ЕЗИК.


(BG) • АКО ДОСТАВЧИКЪТ НА УСЛУГАТА НА КЛИЕНТА ИЗИСКА ЕЗИК,
РАЗЛИЧЕН
ОТ АНГЛИЙСКИ, ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА ОСИГУРИ ПРЕВОД.
• НЕ ИЗПОЛЗВАЙТЕ ОБОРУДВАНЕТО ПРЕДИ ДА СТЕ СЕ КОНСУЛТИРАЛИ
И РАЗБРАЛИ УПЪТВАНЕТО ЗА РАБОТА.
• НЕСПАЗВАНЕТО НА ТОВА ПРЕДУПРЕЖДЕНИЕ МОЖЕ ДА ДОВЕДЕ ДО
НАРАНЯВАНЕ НА ДОСТАВЧИКА НА УСЛУГАТА, ОПЕРАТОРА ИЛИ
ПАЦИЕНТ
В РЕЗУЛТАТ НА ТОКОВ УДАР ИЛИ МЕХАНИЧНА ИЛИ ДРУГА ОПАСНОСТ.

g
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

警告 • 本服務手冊僅提供英文版本。
(ZH-HK) • 倘若客戶的服務供應商需要英文以外之服務手冊,客戶有責任提供翻譯服務。
• 除非已參閱本服務手冊及明白其內容,否則切勿嘗試維修設備。
• 不遵從本警告或會令服務供應商、網絡供應商或病人受到觸電、機械性或其他
的危險。

警告 • 本維修手冊僅有英文版。
(ZH-TW) • 若客戶的維修廠商需要英文版以外的語言,應由客戶自行提供翻譯服務。
• 請勿試圖維修本設備,除非 您已查閱並瞭解本維修手冊。
• 若未留意本警告,可能導致維修廠商、操作員或病患因觸電、機械或其他危險
而受傷。

UPOZORENJE • Ovaj servisni priručnik dostupan je na engleskom jeziku.


(HR) • Ako davatelj usluge klijenta treba neki drugi jezik, klijent je dužan osigurati
prijevod.
• Ne pokušavajte servisirati opremu ako niste u potpunosti pročitali i razumjeli
ovaj servisni priručnik.
• Zanemarite li ovo upozorenje, može doći do ozljede davatelja usluge,
operatera ili pacijenta uslijed strujnog udara, mehaničkih ili drugih rizika.

OPOZORILO • Ta servisni priročnik je na voljo samo v angleškem jeziku.


(SL) • Če ponudnik storitve stranke potrebuje priročnik v drugem jeziku, mora
stranka zagotoviti prevod.
• Ne poskušajte servisirati opreme, če tega priročnika niste v celoti prebrali in
razumeli.
• Če tega opozorila ne upoštevate, se lahko zaradi električnega udara,
mehanskih ali drugih nevarnosti poškoduje ponudnik storitev, operater ali bolnik.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.

h
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

REV 4 5161957-100

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should be thoroughly
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Medical Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no practical
design of equipment can provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents, and
representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

LITHIUM BATTERY CAUTIONARY STATEMENTS

CAUTION
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent
type recommended by the manufacturer. Discard used batteries according to the manufacturer’s
instructions.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this publication.

i
CT/e, ProSpeed AI/FI Series
CT/e, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

j
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

LIST OF EFFECTIVE PAGES

Tab - Section Page REV 1 (Introduction) 1-1 to 1-2 2

Title page 4 2 (Material List and Wiring


- 2-1 to 2-6 4
Title page rear blank Diagram)
(Warning and other 3 (Upgrade Procedures) 3-1 to 3-22 4
a to i 4
important information)
4 (Final Check) 4-1 to 4-2 4
(Revision Information) A 4
- Blank/Rear cover -
(Table of Contents) i 4

REVISION HISTORY
REV Date Primary Reason for Change
1 01/18/07 Initial release.

2 11/26/07 Added multi-language warning information.


Added BIOS setting information.

3 04/14/08 Added a note for software V/R2.00 is only for CTe systems (NPDAS).

4 03/03/10 Added four Languagewarnings for HongKong, Taiwan, Croatian and Slovenian.
Updated material list for Twin Linux. (Sec. 2)
Updated wiring diagram for Twin Linux PC865. (Sec. 2)
Updated upgrade procedure for the Host Video Cable replacement, DBPCI II Board replacement and BIOS setting. (Sec. 3)
Added information for rating plate attachment and product locator card delivery. (Sec. 4)

A
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
blank

B
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 STEERING GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

SECTION 2 - MATERIAL LIST AND WIRING DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

SECTION 3 - UPGRADE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 POWER OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-3 HOST PC REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-4 PERIPHERAL REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-5 NEST REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3-6 REAR FAN ASSY REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3-7 RESTORE OC APPEARANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3-8 BIOS SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3-9 INSTALL SOFTWARE AND OPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
3-10 CONFIGURE SYSTEM & PROTOCOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
3-11 FULL CALIBRATION SYSTEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
3-12 FINALIZATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22

SECTION 4 - FINAL CHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4-1 ATTACHING THE RATING PLATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 SENDING THE PRODUCT LOCATOR CARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

i
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

ii
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 2 5161957-100

SECTION 1 - INTRODUCTION

1-1 PURPOSE
The O2 to Linux OC Upgrade Kit is intended to be used on the installed base of CT/e, ProSpeed AI/FI Series (O2
system) and CT/e Dual, ProSpeed AII/FII Series (O2 system) scanner. According to the following procedures the
O2 console can be upgraded to Linux console.
A steering guide is shown in Section 1-2.
The materials of the upgrade kit are listed in Section 2.

1-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 2 5161957-100

1-2 STEERING GUIDE


The following flowchart shows the main flow of the upgrade procedures. Print this page and check after finishing
each step.

Save all the Patient Images


Write down all the configuration,
Option and Protocol Information

Upgrade Procedures

BIOS Setup
Install Software and Options
Configure System and Protocols
Full Calibrate System

Perform Test Scan


Evaluate image Quality

1-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

SECTION 2 - MATERIAL LIST AND WIRING DIAGRAM

Table 2-1 Material List of Upgrade Kit for Single Slice (5166686)
ITEM PART NO. DESCRIPTION QTY.

1 2369471-2 Host PC Support 1

2 2369472 PC Holder A 2

3 2369473 PC Holder B 2

4 2369474 Strap 2

5 N9408HR Screw 4

6 5189991 E-SATA Host computer SB61G2 V4 1

7 5133747 160G SATA Hard Disk 2

8 5189897 External SATA Hard Disk Cable 1

10 5142669 HDD Power Cable Outside PC 1

11 2369354 Multi-port Serial Card 1

13 5115864 Creative Audio Card 1

14 5159503 4 Port Serial Card Cable 1

15 2318421 Audio In Cable 1

16 2318422 Audio Out Cable 1

17 2248114-1 Fan Assy 2

18 5172692 CT/e SCSI DVD-RAM Driver 1

19 5189532 Shield Board 1 1

20 2146513 U0011AA Screw 8

21 5161957-100 O2 PC to Linux 865PC Upgrade instruction 1

22 5144980 OS SW HP8200 1
ProSpeed, CT/e Xtreme Series SW V/
23 5196209* 1
R2.00
*: Software V/R2.00 (5196209) is ONLY for CTe Systems (NPDAS) where the O2 to Linux Computer Upgrade
has been performed. Do not upgrade software version for Be1 systems (CTe systems with MDAS and GDAS)
into this upgrade system.

2-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

Table 3: Material List of Upgrade Kit for Dual Slice (5372961)


ITEM PART NO. DESCRIPTION QTY.

1 2369471-2 Host PC Support 1

2 2369472 PC Holder A 2

3 2369473 PC Holder B 2

4 2369474 Strap 2

5 N9408HR Screw 4

6 5189991 E-SATA Host computer SB61G2 V4 1

7 5133747 160G SATA Hard Disk 2

8 5189897 External SATA Hard Disk Cable 1

9 5142669 HDD Power Cable Outside PC 1

11 2369354 Multi-port Serial Card 1

13 5115864 Creative Audio Card 1

14 5159503 4 Port Serial Card Cable 1

15 2318421 Audio In Cable 1

16 2318422 Audio Out Cable 1

17 2248114-1 Fan Assy 2

18 5172692 CT/e SCSI DVD-RAM Driver 1

19 5189532 Shield Board 1 1

20 5126711 CTT 1.8.12 ,Prospeed II, cte dual Xtreme Series Core OS 1

21 5126713 Prospeed II, CTe Dual Xtreme Series Application Software VR4.00 1

22 5128444 ProSpeed II Xtreme Patch V4.01 1

23 5161319 ProSpeed II, CTE Dual Xtreme Series Patch V4.03 1

24 5193587 Twin Linux Patch Software V/R4.05 1

25 2146513 U0011AA SCREW 8

26 5161957-100 O2 PC to Linux 865PC Upgrade Instruction 1

28 2304737-2 Cable, DVI Video Cable - OCTANE II 1

29 5215584 DBPCI2 PWA with NP BRKT 2

30 5257021 CT/e Dual Patch Software V/R4.06 1

31 5372901 Twin Linux Patch Software V/R4.07 1

32 46-303268P1 Product Locator Card form for Rating Plate 1

33 5178663 Twin Linux Patch V/R4.04 1

2-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

Illustration 2-1 OC Wiring Diagram (2245647SCH) of Single Slice O2 System

2-3
CT/e, ProSpeed AI/FI Series

5161957-100
O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
CT/e Dual, ProSpeed AII/FII Series

       
THERMO CABLE1
0RQLWRU $*3*UDSKLFV

149 ,$+267 3RZHURQRIIUHVHW 176


63($.(5
6HULDO$7$ 
6HULDO$7$ 
THERMO CABLE2 &20387(5
$ 46
42
$
PCI HOST

150
CARD

- 131 &1 NAA1 &1


-
(WKHUQHW
6HULDO
.H\ERDUG
0RXVH

.H\ERDUG -

86%
- 133 3($.(5Z3$1(/
-
REAR CN1 w/PANEL
OC Wiring Diagram of Single Slice PC865 Linux System

- PERIPHERAL 40
3'8 - 134 135 136
-
J1
-
7*3 WR*DQWU\ - FRONT PNL
-
% ([WVSHDNHU %
- -
ZLWK$PS 138
-
HDD
- SATA0
SATA
SATA0 SATA SYS/IMG
BRACKET
SATA1 SATA1
- SATA HDD
143
205 RAWDATA
-
Audio
SCSI-IDE
47 Adapter E-IDE
SCSI DVD-RAM
REAR CN2 w/PANEL

140 J13
& 0RXVH - - NPSC &





142 J14
(WKHUQHW - - 139
$8;6HULDO - - 147
152
$8;6HULDO - -

2-4
Security Key
$8;6HULDO - -
$8;6HULDO - -

$XGLRRXW
$XGLRLQ
6&6,
-
-
-
-
48 141
' '

BP CONTROLLER

Serial port expander


ethernet card
PANEL

SCSI CARD

Sound Card
6ODYH0RQLWRU - -
VSPL

DBPCI

CARD
49
0RQLWRU -
45
w/PANEL

DASIFN2
145
'$6 -
- 1 2 3 4 5 6 7
-
( PCI BACKPLANE (
1(67
9
&RQQHFWRU3DQHO
ᷛ䆄5(9 47< ᬍऩো)=1R ㅒᄫ0$'(%< ᮹ᳳ'$7( ᦣ 䗄'(6&5,37,21
ৡ ⿄
*(+:0(',&$/6<67(06 7,7/(
2&7DEOH$66<
Illustration 2-2

䆒䅵'(6,*1(5 ↨՟6&$/(
) ᅵḌ&+(&.(5
᭛ӊҷো'2&12 )
GE HEALTHCARE

໡ᅵ5(&+(&.(5
ᷛᅵ67$1'$5' 6&+
ᡍ‫ޚ‬$33529$/ 6KHHWRI
       
REV 4
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

Illustration 2-3 OC Wiring Diagram (2320208SCH) of Dual Slice O2 System

2-5
CT/e, ProSpeed AI/FI Series

5161957-100
O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
CT/e Dual, ProSpeed AII/FII Series

       
THERMO CABLE1
0RQLWRU $*3*UDSKLFV

149 ,$+267 3RZHURQRIIUHVHW 176


63($.(5
6HULDO$7$ 
6HULDO$7$ 
THERMO CABLE2 &20387(5 42
$ 46 $
PCI HOST

150
CARD

- 131 &1 NAA1 &1


-
(WKHUQHW
6HULDO
.H\ERDUG
0RXVH

.H\ERDUG -

86%
- 133 3($.(5Z3$1(/
-
REAR CN1 w/PANEL

- PERIPHERAL 40
3'8 -
OC Wiring Diagram of Dual Slice PC865 Linux System

134 135 136


-
J1
-
7*3 WR*DQWU\ - FRONT PNL
-
% ([WVSHDNHU - - %
ZLWK$PS 138
-
HDD
- SATA0
SATA
SATA0 SATA SYS/IMG
BRACKET
SATA1 SATA1
- SATA HDD
143
205 RAWDATA
-
Audio
SCSI-IDE
47 Adapter E-IDE
SCSI DVD-RAM
REAR CN2 w/PANEL

140 J13
0RXVH - - NPSC





& 142 154 J14 &
(WKHUQHW - - 139
$8;6HULDO - - 147
152
$8;6HULDO - -

2-6
Security Key
$8;6HULDO - -
$8;6HULDO - -

$XGLRRXW
$XGLRLQ
6&6,
-
-
-
-
-

-

-
-
-
48 141

BP CONTROLLER
' '

Serial port expander


(1)
DBPCI2card
PANEL

SCSI CARD
NPRS
NPRS
NPRS
NPRS

DBPCI2 (0)

Sound Card
6ODYH0RQLWRU - -
VSPL

NPRIF

CARD
49

ethernet
0RQLWRU -

NPRM
5
6
7
8
45

NPRS
NPRS
NPRS
NPRS
w/PANEL

DASIFN2
145
'$6 -
- 1 2 3 4 5 6 7

4
3
2
1
-
PCI BACKPLANE
( 153 (
1(67
9
&RQQHFWRU3DQHO
ᷛ䆄5(9 47< ᬍऩো)=1R ㅒᄫ0$'(%< ᮹ᳳ'$7( ᦣ 䗄'(6&5,37,21
Illustration 2-4

ৡ ⿄
*(+:0(',&$/6<67(06 7,7/(
2&7DEOH$66<
䆒䅵'(6,*1(5 ↨՟6&$/(
GE HEALTHCARE

) ᅵḌ&+(&.(5
᭛ӊҷো'2&12 )
໡ᅵ5(&+(&.(5 6&+
ᷛᅵ67$1'$5'
ᡍ‫ޚ‬$33529$/ 6KHHWRI
       
REV 4
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

SECTION 3 - UPGRADE PROCEDURE

NOTICE
Since Irix system state can not be restored to Linux system, so all the related information
(such as: options, configurations, protocols, etc...) should be written down in advance for
later reconfig.
All the Irix option keys should be prepared by Linux option keys, Irix 3D option has been re-
placed by Linux Volume Analysis option, for other detailed information, please contact ap-
plication specialist and sales people.
Entire upgrade time needs 8 hours, this is up to different engineers.

The following procedures can be performed refer to Section 2 Wiring Diagram.

3-1 PREPARATION
1. Verify that the contents of this upgrade kit are correct according to the Material List (see Section 2).
2. Save Patient Images.
3. Record camera data (for the system with camera only).
4. Record Options Information.
a. On the desktop menu, click on List Option button located on the upper right of the screen.
Option list will appear.

b. Click on the current Irix system “Installed option” and check if the corresponding Linux option keys are ready
on hand.
c. Click Quit to return to the desktop menu.
5. Write down all the configuration and protocols information for later use.
6. Click on Shutdown on the desktop menu.
7. Few seconds later, the message “Okay to power off the system now. Press any key to restart.” appears.
8. Press the Power button on Front of OC to power off.

3-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-2 POWER OFF

WARNING
THE ELECTRICAL SHOCK HAZARD!
IF THE SYSTEM HAS UPS ASSY, MAKE SURE YOU TURN OFF THE UPS POWER BEFORE
YOU ACCESS THE OPERATOR CONSOLE.

CAUTION
FOLLOW ALL ELECTRO-STATIC DISCHARGE PROCEDURES. ALWAYS USE ESD WRIST
STRAP AND GROUNDING CORDS. THE USE OF A GROUND MONITOR IS SUGGESTED.

Understand and follow the Safety Guidelines Manual.


1. Turn OFF the OC main power.
2. For system with UPS, turn OFF the UPS power.
3. Cut off the power from the Hospital Power Distribution Box and perform LOTO procedures.
4. Remove front, rear, left and right covers from OC.
5. Turn off CB1, CB2 at right side of OC.

3-2
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-3 HOST PC REPLACEMENT


1. Remove the six (6) screws at the front of the SCSI Bay and the two (2) screws at the rear of the SCSI Bay that
fasten the SCSI Bay (see Illustration 3-1).
2. Remove the SCSI Bay right dust cover by unscrewing its four(4) screws.
3. Remove the OC Host PC Front Cover by unscrewing its six (6) screws (Illustration 3-1).

Illustration 3-1 Host Computer Assy Removal

SCSI Bay

ple
xam
E SCSI Dust Cover

SCSI Bay

Host CPU Assy

Loosen the two screws that fasten


the lower part of the SCSI Bay

OC Host Cover
OC Host Cover Screws (x6)

3-3
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-3 HOST PC REPLACEMENT (CONTINUED)


4. Disconnect the remaining connectors from the Host CPU Assy (Host Video, SCSI Cable, Ethernet, Serial 1,
Serial 2, Keyboard, Mouse, Bit3 Cable (HDD connector), Power Cable).
5. Disconnect the Audio connectors located on the right side of the Host CPU. (See Illustration 3-3)
6. Remove the two (2) screws that fasten the Host CPU Assy. (See Illustration 3-2)

Illustration 3-2 CPU Assy (Front View)

Screws (x2)

Host CPU Assy

SCSI
Cable

Host
Video Power Cable

Ethernet
Serial
(x2)

Bit3 Cable

3-4
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-3 HOST PC REPLACEMENT (CONTINUED)


7. Slide out the Host CPU and the Host PC Support, avoiding that the cables and connectors obstruct the Host
CPU Assy removal. (See Illustration 3-3).

Illustration 3-3 CPU Assy Removal


Host CPU Assy

Audio
Connectors

Host PC Support

Screws (x2)

Host PC
Support

8. Install four PC holders and two straps to the new Host PC Support (2369471-2). (Illustration 3-4)
9. Place the Host PC865 (5189991) on the Host PC Support, fasten the four PC holders and tighten the two straps
to fix the Host PC865.

Illustration 3-4 Host PC Support Installation


PC Holder (x4)

Strap (x2)

3-5
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-3 HOST PC REPLACEMENT (CONTINUED)


10. Place the Host PC865 and the Host PC Support to the slide-rail of the OC, connect the cable connectors to the
rear panel of the Host PC (Host Video, Bit3 Cable (HDD Connector), Ethernet, keyboard, Mouse, Power Supply
Cable, Serial 2 (COM1)), then slide it backward in the OC.

Illustration 3-5 Host PC865 Rear Panel

Outside Control Cable

Power Supply Cable Host Video


Bit 3 Cable (DVI)

Serial 2 Ethernet
Mouse
Host Video
Keyboard (VGA)

NOTICE
For Twin Linux System: Replace the Video Cable (2156297-2, Host PC VGA <-> VSPL) with
the new Video Cable (2304737-2, Host PC DVI <-> VSPL).

Illustration 3-6 New Video Cable Connection

New Video Cable


(2304737-2)

3-6
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-3 HOST PC REPLACEMENT (CONTINUED)


11. Connect cable connectors shown as below table to the front panel of the Host PC. (see Illustration 3-7)
Table 3-1 Cable Connection
PN# Cable Name From and To
5142669 Power Cable Host PC front - HDD rear
5189897 External SATA Hard Disk Cables Host PC front - HDD rear
Note
Insert SATA Cable connector with reed, “click” is heard when it is well plugged in. Press the reed
on the SATA Cable connector to pull it out, otherwise the connector may be destroyed. (See Illus-
tration 3-7)

Illustration 3-7 Host PC865 Front View

SATA HDD Cable

Power Cable

Note: Connect the SATA Cable and Power Cable connectors to the front
panel of the Host Computer by fastening tightly screws.

3-7
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-4 PERIPHERAL REPLACEMENT


1. Slide out the SCSI Bay.
2. Disconnect the connectors (signal and power cable connectors) from the rear of the HDD(s), 3.5” MOD and the
CD-ROM.
3. Unscrew the HDD mounting screws, remove the HDD(s) and CD-ROM front covers by unscrewing screws and
slide the HDDs, 3.5” MOD and the CD-ROM forward and out of the OC.

Illustration 3-8 Remove the HDD(s) and the CD-ROM


CD-ROM

To NPSC board J16

Signal Cable

Blank connector

Power Cable 3.5" HDD


Connector (One HDD for single slice)
(Two HDDs for dual slices)
Mounting
Screws
4. Disconnect the cable connector from the SCSI Card (2202402).
5. Disconnect the cable connector from the NPSC Assy (J16), then remove the SCSI Cable (2249079).
6. Disconnect the cable conenctor from the NPSC Assy (J17), then remove the HDD Cable (2242275).

Illustration 3-9 NPSC Assy

NPSC SCSI Cable (from Host PC)


TH5

J18 J19 J20


C6
F10

C5 C7
F1
J1
J

F2
F 8

J 2
13

F3
14

J 3
F4 TH3
J 4
F5
J 5
F6
J
J

J 6
F 9
F 7

17
16

J 7
Signal Cable
(to Peripheral Assy J 8
J9 J10
, J13)
TH4

HDD Cable (to Raw HDD SCSI Cable (to SCSI Card, 2249079_J16)
, 2242275_J17)

3-8
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-4 PERIPHERAL REPLACEMENT (CONTINUED)


7. Connect the SCSI Cable (2156306) between the UW SCSI Card (J1) and NPSC (J14).

Illustration 3-10 SCSI Card

SCSI Card

J5

J2
J1

8. Check the jumper setting of the SCSI DVD-RAM Drive if ID set as 1. (Illustration 3-11)
9. Respectively connect the two Power Cable connectors to the DVD-RAM Drive and the Adapter.

Illustration 3-11 Jumper Setting of the SCSI DVD-RAM Drive

JP2 (disable)
(remove the block)

Power Cable
for Adapter
ID=1

Power Cable
for DVD-RAM Drive

Power Cable for SCSI DVD-RAM Drive

10. Install the SCSI DVD-RAM Drive to the Peripheral Assy of the OC by screwing the DVD-RAM mounting screws.
11. Connect the SCSI Signal Cable and Power Cable to the SCSI DVD-RAM Drive.
12. Install the two new HDDs to the Peripheral Assy of the OC by screwing the HDD mounting screws.
13. Connect the Signal Cables and Power Cables to the HDDs according to Table 3-1.

IMPORTANT NOTE: - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
When connect the SATA Cables, be sure that the upper HDD is connected to the upper con-
nector (SATA 0). Otherwise System Disk and Image Disk may be mixed up during next HDD
replacement and software loading will result in losing image.
------------------------------------------------------------------------

3-9
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT

Table 3-2 NEST List


Single Dual
Card Item Card Name O2 System Linux 865 System O2 System Linux 865 System
1 (up) Audio Card v v
2
3 Serial Card v v v v
4 SCSI Card v v v v
5 BP Card v v v v
DBPCI v v
6

7 DBPCI v v v v
8 (down) NPRIF v v v

For Single Slice


1. (For Single Slice with Serial Panel Assy):
a. Disconnect the cable connectors from the Serial Card (2243265) and the Serial Panel Bracket (2243962),
then remove the Serial Expander Cable (2242196). (Illustration 3-12)
b. Remove the Serial Panel Cable Assy (2249698) according to below table.

Table 3-3 Serial Panel Cable Assy Connection


From To
Serial Panel Assy, 1~4 Rear CN2, J3~J6 (Illustration 3-16)
Serial Panel Assy, 5 series cable Security Key

Illustration 3-12 SP Assy


SP Assy

Left View

1 2 3 4

Screw (x2)

c. Remove the Serial Panel Assy (2265429) by unscrewing its two screws. (See Illustration 3-12)

3-10
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)


(For the Single Slice with ST1800):
Refer to Illustration 2-1 (2245647SCH) remove the ST1800.

a. Remove cables according to figure:


Item Part # Part Name Connection
144 2252080 ST-1800 Cable Assy ST1800, 1~4 to RearCN2, J3~J6
146 2249079 F&W SCSI Cable Assy UW SCSI Card to NPSC
147 2156306 SCSI Cable Assy Host CPU, Ultra/Wide SCSI to ST1800
148 2252074 SCSI Cable 3 Assy NPSC Assy, J14 to ST1800

b. Remove cable between ST1800, 5 series cable and Security Key.


c. Remove the ST1800.
2. Disconnect the two NPR Cable from the NPRIF (2177243-2).
3. Disconnect the Power Cable and the Ground Cable from the NPRM.

Illustration 3-13 Cables Connection

Power Cable

NPRIF NPRM
NPR Cable (X2) Ground Cable

4. Remove the Serial Card by unscrewing its screw.


5. Remove the NPRIF, NPRSs and NPRM by unscrewing their screws.

3-11
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)


6. Install the Audio Card (5115864, refer to Illustration 3-15) and the 4-Port Serial Card (2369354) that respec-
tively fasten them. (Refer to the order of Table 3-2)

Illustration 3-14 Multi-port Serial Card Position

Audio Card

Multi-port
Serial Card

SCSI Card

BP Controller
Card

DBPCI

Illustration 3-15 Audio Card


Note: Slowly pushing the board, until its reaches the backplane.
If the board is blocked during inserting, remove the cable
connection screw shown as below.

Remove this Cable Connection Screw

3-12
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)


7. Unscrew the two screws on the Rear CN1 Panel, and slide the Rear CN1 Panel Assy out of the Connector
Panel Assy. (Illustration 3-16)

Illustration 3-16 Rear of OC

Rear Fan
(x2)

RearCN1
DASIFN
RearCN2

Illustration 3-17 RearCN1 and RearCN2


RearCN1 RearCN2
J11 J10
JP8 JP9

FRONT PANEL
Speaker

J13

J5

J12
Serial 4

Serial #3
J12

J6
Serial 2
J9

J4
TGP
J14

Serial 3

Serial #2
J8

J11

J5
Serial 1
OUT
J7

J3
J6

Serial 2
IN

Serial #1
J10

J4
PDU
J15

Serial 1

Serial #0
PWR
J2

J9

J3
FOOT-SW

J16

Ethernet

Ethernet
J8

J2
KEYBOARD

DBPCI
J1
J17

Mouse

Mouse
J7

J1

3-13
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)


8. Disconnect the Audio Cables (In/Out) from the J6 and J7 of the Rear CN1.
9. Remove the Audio In Cable (2240366) and the Audio Out Cable (2240367).
10. Connect the new Audio Cables to the Audio Card.
Cable No. Cable Name Connect From Connect to
2318421 Audio IN Cable Audio Card Mic IN (Red) Rear CN1, J6
2318422 Audio OUT Cable Audio Card Speaker Out (Green) Rear CN1, J7

Illustration 3-18 Connect the Audio Cables

Out
In

11. Connect the Multi-port Serial Cable (5159503) according to the labels on the cable connectors.
4-Ports Serial Cable Connect to
5159503 Multi-port Serial Card
S1 to RearCN2 S1 modem RearCN2 J3
S2 to RearCN1 J4 RearCN1 J4, TGP
S3 to Front Panel Empty
S4 to skey Service Key

12. Install the Rear CN1 Panel Assy by screwing its two screws.

3-14
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)

For Dual Slice


1. Disconnect the cable connectors from the Serial Card (2243265) and the Serial Panel Bracket (2243962), then
remove the Serial Expander Cable (2242196). (Illustration 3-12)
2. Remove the Serial Panel Cable Assy (2249698) according to Table 3-3.
3. Remove the Serial Panel Assy (2265429) by unscrewing its two screws.
4. Remove the Serial Card by unscrewing its screw.
5. Install the Audio Card (5115864, refer to Illustration 3-15) and the 4-Port Serial Card (2369354) that respec-
tively fasten them. (Refer to the order of Table 3-2)

Illustration 3-19 Multi-port Serial Card Position

Audio Card

Serial Card

SCSI Card

BP Card
DBPCI

NPRIF

NPRM

6. Disconnect the cable connectors from two DBPCIs (J2 and J3).
7. Unscrew the screws on the DBPCI board bracket.
8. Remove the DBPCI Assy from the PCI backplane.
9. Insert the new DBPCI II Assy to the PCI backplane, slowly pushing the board, until it reaches the backplane.
10. Secure the DBPCI II Assy by using its screws.
11. Unscrew the two screws on the Rear CN1 Panel, and slide the Rear CN1 Panel Assy out of the Connector
Panel Assy. (Illustration 3-16)
12. Disconnect the Audio Cables (In/Out) from the J6 and J7 of the Rear CN1.

3-15
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-5 NEST REPLACEMENT (CONTINUED)


13. Remove the Audio In Cable (2240366) and the Audio Out Cable (2240367).
14. Connect the new Audio Cables to the Audio Card. (refer to Illustration 3-18)
Cable No. Cable Name Connect From Connect to
2318421 Audio IN Cable Audio Card Mic IN (Red) Rear CN1, J6
2318422 Audio OUT Cable Audio Card Speaker Out (Green) Rear CN1, J7

15. Connect the Multi-port Serial Cable (5159503) according to the labels on the cable connectors.
4-Ports Serial Cable Connect to
5159503 Multi-port Serial Card
S1 to RearCN2 S1 modem RearCN2 J3
S2 to RearCN1 J4 RearCN1 J4, TGP
S3 to Front Panel Empty
S4 to skey Service Key

16. Reconnect the cable connectors to the DBPCI II.


17. Install the Rear CN1 Panel Assy by screwing its two screws.

NOTICE
Please pay attention to the connection position of the cables. (Refer to Illustration 3-20)

Illustration 3-20 Nest

O2 System Linux System

DBPCI2 1

DBPCI 1 DBPCI2 0

DBPCI 0

J2
J3
DBPCI2 #1, J2 DASIFN, J2 (2249713)
J2 DBPCI2 #1, J3 RearCN1, J1 (2156308-2)
J3 DBPCI2 #0, J2 DASIFN, J9 (2319300)

3-16
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-6 REAR FAN ASSY REPLACEMENT


1. Remove the right rear fan assy by unscrewing its screws from the back of the OC. (See Illustration 3-21)
2. Disconnect the power cable.
3. Remove the fan from the panel by unscrewing the two(2) screws.
4. Install the new fan.
5. Repeat Steps 1 to 4 to replace the left rear fan assy.
Note
After the Rear Fan Assy replacement, confirm the air flow direction is outward.

Illustration 3-21 Air Flow Direction

AIR FLOW DIRECTION

Front Right Side Rear


Fan (x2)

AIR FLOW

AIR FLOW

3-17
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-7 RESTORE OC APPEARANCE


1. Install the two (2) screws to fasten the Host PC.
2. Install the OC Host PC Front Cover by screwing its six (6) screws.
3. Install the SCSI Bay right dust cover by screwing its four (4) screws.
4. Install the eight (8) screws to fasten the SCSI Bay.
5. Turn on CB1, CB2 at right side of the OC.
6. Install the Shield Board to the OC Front Cover.

Illustration 3-22 Install the Shield Board

Front Riew Rear View

For 5” MOD

For DVD-RAM

Shield Board

7. Install front, rear, left and right OC Covers.

3-8 BIOS SETUP


1. Power ON the OC, press Meas Dist to reach BIOS Set screen when the Pentium Logo displayed on the
screen.
2. Enter password "linux".
3. Select Set User Password, click "Enter" three times to cancel original password. Then "Save & Exit Setup".
Reboot the system, enter BIOS again.
The detailed BIOS setting information shown as below.

3-18
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION

LEVEL 1 LEVEL 2 LEVEL 3 Config Info


Application V/R3.1x or V/ Application V/R4.0x
R2.00 * (OS4.5.1)_Be1 (OS1.8.12)_Twin Linux
Standard CMOS IDE Channel 0 master
Features
IDE Channel 0 slave
Display detected hardware info
IDE Channel 1 master
IDE Channel 1 slave
Drive A None
Video EGA/VGA
Halt on No errors
Advanced BIOS CPU Feature Delay Prior to 16 Min
Features Thermal
Hard Disk Boot Priority 1. Bootable Add-in Cards
2. Ch0
3. Ch1
BIOS Write Protect Disabled
Virus Warning Disabled
CPU L1 & L2 Cache Enabled
Quick Power On Self Test Enabled
First Boot Device CD-ROM USB-CDROM
Second Boot Device Hard Disk CD-ROM
Third Boot Device Disable Hard Disk
Boot Other Device Disabled
Boot up Floppy Seek Disabled
Boot Up NumLock Status Off
Gate A20 Option Fast
Typematic Rate Setting Disabled
Security Option Setup
APIC Mode Enabled
MPS Version Control For OS 1.4
OS Select For DRAM > 64MDB Non-OS2
Report No FDD For WIN 95 No
Small Logo(EPA) Show Disabled
Advanced chipset DRAM Timing Selectable By SPD
Features
Memory Frequency For Auto
System BIOS Cacheable Enabled
Video BIOS Cacheable Disabled
Memory Hole A 15M-16M Disabled
AGP Aperture Size (MB) 64
Init Display First Onboard/AGP
On-Chip VGA Disabled
Turbo Mode Auto

3-19
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

LEVEL 1 LEVEL 2 LEVEL 3 Config Info


Application V/R3.1x or V/ Application V/R4.0x
R2.00 * (OS4.5.1)_Be1 (OS1.8.12)_Twin Linux
Integrated Peripherals On-Chip IDE Device IDE HDD Block Mode Enabled
IDE DMA Transfer Access Enabled
On-Chip Primary PCI IDE Disable Enabled
IDE Primary Master PIO Auto
IDE Primary Slave PIO Auto
IDE Primary Master UDMA Auto
IDE Primary Slave UDMA Auto
On-Chip Secondary PCI IDE Disable Enabled
IDE Secondary Master PIO Auto
IDE Secondary Slave PIO Auto
IDE Secondary Master Auto
UDMA
IDE Secondary Slave UDMA Auto
On-Chip Serial ATA Enhanced Mode SATA Only
Serial ATA Port0 Mode SATA0 master Primary Master
On Board Device USB Controller Enabled
USB 2.0 Controller Enabled
USB Keyboard Support Disabled
AC97 Audio Disabled
Super IO Device Onboard FDC Contoller Enabled
Onboard Serial Port 1 3F8/IRQ4
Onboard Infrared Port Disabled
Onboard Parallel Port 378/IRQ7 Disabled
Parallel Port Mode SPP
Onboard Lan Boot Disabled
ROM
Power Management ACPI Suspend Type S1(POS)
Setup
Power Management User Define
Video Off Method DPMS
Video Off In Suspend Yes
Suspend Type Stop Grant
MODEM Use IRQ 3
Suspend Mode Disabled
HDD Power Down Disabled
Soft-Off by PWR- Instant-Off
BTTN
Wake-Up by PCI card Disabled
Resume by Alarm Disabled
Primary IDE 0 Disabled
Primary IDE 1 Disabled

3-20
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

LEVEL 1 LEVEL 2 LEVEL 3 Config Info


Application V/R3.1x or V/ Application V/R4.0x
R2.00 * (OS4.5.1)_Be1 (OS1.8.12)_Twin Linux
Secondary IDE 0 Disabled
Secondary IDE 1 Disabled
FDD, COM, LPT Port Disabled
PCI PIRQ[A-D]# Disabled
PS2 Keyboard Power ON Disabled
PS2 Mouse Power ON Disabled
PWRON After PWR-Fail On
PnP/PCI Configurations Reset Configuration Data Enabled
Resources Controlled By Auto(ESCD)
PCI/VGA Palette Snoop Disabled
INT Pin 1 Assignment Auto
INT Pin 2 Assignment Auto
INT Pin 3 Assignment Auto
INT Pin 4 Assignment Auto
INT Pin 5 Assignment Auto
INT Pin 6 Assignment Auto
INT Pin 7 Assignment Auto
INT Pin 8 Assignment Auto
PC Health Status CPU Fan Speed Control Smart Fan
CPU Temp Tag 64
Frequency/Voltage Control Auto Detect PCI CLK Enabled
Spread Spectrum Disabled
CPU $ AGP/PCI Clock Set Auto Detect
CPU Voltage Set Auto
DDR Voltage Set Auto
AGP Voltage Set Auto

4. Re-set the original password (the password is "linux").


Note
BIOS Setting Password MUST be set. Only trained service personnel can enter BIOS Setting.
5. Then the system continues to boot up and begins detecting SCSI Devices. Make sure DVD-RAM is shown in
the list of SCSI Device.

3-21
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

3-9 INSTALL SOFTWARE AND OPTION

Detail installation procedure refer to Function Check/Adjustment -> System -> Software Loading.

Product CD-ROM Functional Check/Adjustment Manual


CT/e, CT ProSpeed AI/FI Series 2323696 2244775
CT/e Dual, ProSpeed AII/FII Series 2337938 2326039

After OS started, click Reconfig -> Hardware Setting, Select DVD-RAM as State Storage Type. (Click DVD-RAM
twice to make sure it is selected.)

e
pl
am
Ex

3-10 CONFIGURE SYSTEM & PROTOCOLS


Refer to the configuration and protocol information written down before to reconfigure the system.

3-11 FULL CALIBRATION SYSTEM


Detail calibrate procedure refer to Function Check/Adjustment -> System -> System Calibration.

3-12 FINALIZATION
Perform a test scan and verify that the image is normal.

3-22
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
REV 4 5161957-100

SECTION 4 - FINAL CHECK

4-1 ATTACHING THE RATING PLATE


Attach the provided rating plate on the rear of OC, beside the OC rating plate.
Note
If other options have been installed in the system, several rating plates should have been attached.
In this case, attach this rating plate above the others.

4-2 SENDING THE PRODUCT LOCATOR CARD


1. Fill in the blanks on the Product Locator Card provided in this kit.
2. Send the Product Locator Card to the appropriate address provided below.
Address for GEMS-AM sites:
Send to the address printed on the Product Locator Card.
Address for GEMS-E sites:
Send to the address printed on the Product Locator Card.
Address for GEMS-A sites:
Send to the local PLS or Service Administrator.

4-1
CT/e, ProSpeed AI/FI Series
CT/e Dual, ProSpeed AII/FII Series
GE HEALTHCARE O2 PC TO LINUX PC865 UPGRADE INSTRUCTION
blank

4-2
www.gehealthcare.com

Imagination at work