You are on page 1of 1

QUICK

START ™

GUIDE 01 COLLECT
FECAL MATTER
02 PREPARE
FECAL SAMPLE
03 ADD SAMPLE
TO SLIDE

CHECK LIST
Ensure you have the following
materials ready to run a test:
Stir Stick Glass Slide ParaPro Sample
Tube + Scoop
• Unscrew the tube cap from • Shake the tubes and then • Place a glass slide in the
the collection tube place them in a centrifuge slide template with the
Transfer Loop Coverslip • Add 1 scoop, about 1 gram, • Spin the sample for 2 name or barcode towards
ParaPro Collection of fecal matter to the minutes at 500 RCF the top
Tube + Cap
collection tube • Unscrew the collection • Pull material from the top
• Use a stir stick to remove tube with the scoop and of the sample tube using
Slide Template Centrifuge Scanner the matter from the scoop discard both the transfer loop
and mix it • Place the sample tube in • Place 2-4 loops of fecal
• Screw the sample tube onto a tray matter onto the glass slide
the collection tube around the edge of the
• Continue twisting until the target circle
tubes are firmly screwed
together

04 PLACE 05 INITIATE 06 LOAD AND 07 REVIEW


COVERSLIP TEST SCAN SLIDE SCAN

Coverslip Top

• Orient the coverslip with • Select Scan for the • Unlock the slide holder and • Open the sample results,
the IMAGYST text readable correct sample; Sample place the glass slide (with review any captured
in the lower left-hand corner information will come coverslip) on the slide tray images, and select
of the coverslip pre-populated from any • Place any label or writing examples for inclusion
• Place the coverslip on FUSE-connected software towards the right-side of in the report
the glass slide over the system the slide tray and lock the • Add any comments and
targeted area of the slide • If no integration is available, slide holder select Finalize
template select the Add Sample icon • Select Scan Slide • Select the View Report
(+), enter in the sample icon to view a shareable
information, select Create, PDF or review results in
and then select Scan FUSE-connected software
• Select a scanner, and make system
sure the tray is open (select
Open Tray if necessary)

For questions concerning IMAGYST please contact VETSCAN Technical Support at 800-822-2947 or dxsupport@zoetis.com.
All trademarks are the property of Zoetis Services LLC or a related company or licensor unless otherwise noted.
© 2020 Zoetis Services LLC. All rights reserved. VTS-00248