Weebz Mandiri

Heated Sample Line Indo Warehouse Blok C1


Jl. Kaliabang Tengah No. 22
Calculator Bekasi Utara 17125 – Indonesia
telp/fax. +62.21.8838.0008
CS2 H2S SO2 Analyzer - SPV - Lenzing

Input Data
Maintenance Temperature 55 degC Min. Ambient Temp 25 degC
Max. Product Temp. 65 degC Max. Ambient Temp 35 degC
Max. Exposure Temp. 40 degC Safety Value 20%

Process Tube Spec


Process Tube 1 Process Tube 3
Outer Diameter 1/4 inch Min. Ambient Temp - inch
Tube Material PFA Max. Ambient Temp -
Wall Thickness TBA inch Safety Value - inch
Type Heated Type None

Process Tube 2 Process Tube 4


Outer Diameter 1/4 inch Min. Ambient Temp - inch
Tube Material PFA Max. Ambient Temp -
Wall Thickness TBA inch Safety Value - inch
Type Unheated Type None

Calculation of Sample Condensate Accumulation Formula used:


Cooled Air flow: 240 Nl/h 𝑤_𝑡𝑡𝑙=𝐶𝑜𝑜𝑙𝑒𝑑 𝑎𝑖𝑟 𝑓𝑙𝑜𝑤 ×(𝑀𝑜𝑖𝑠𝑡𝑢𝑟𝑒 𝐴𝑐𝑐𝑢𝑚.)/((100 𝑁𝑙 𝑐𝑜𝑜𝑙𝑒𝑑 𝑎𝑖𝑟)/ℎ𝑟)
Sample Dew Point: 30 °C
Moisture Content: 4.25120084358225 %Vol
Moisture Accumulation: 2.13095132018349 ml/h per 100 Nl cooled air/h 𝑤_𝑡𝑡𝑙=𝑇𝑜𝑡𝑎𝑙 𝑐𝑜𝑛𝑑𝑒𝑛𝑠𝑎𝑡𝑒 𝑎𝑐𝑐𝑢𝑚𝑢𝑙𝑎𝑡𝑖𝑜𝑛
Condensate Accumulation: 5.11428316844039 ml/h
Convertion
Vol/Length Unit Constant to m3/m Length Unit Convertion Constant to m
m3/m 1 m 1
cc/ft 3.28E-06 ft 0.304785126485827
inch 0.0254
mm 1.00E-03
Flow Rate
Pressure Unit Convertion Constant to psia Time Unit Convertion Constant to Unit
psia 1 s 1 m3/s
bar 0.0689464975179261 minute 60 m3/h
atm 14.6959 cc/min
l/h
l/min
Convertio
Convertion Convertion n
Constant first Constant first Constant
unit Volume Unit unit Temperaturfirst unit
1 m3 1 degC 1
0.000277777778 cc 1.00E-06 degF x
1.66666667E-08 l 0.001
2.78E-07 ft3 0.0283168
1.67E-05