DOCUMENT: RD045300761

REVISION NO.: 2.0

PXR Type: B

Date: May,10. 2005

POSKOM RESEARCH INSTITUTE

[CALIBRATION MANUAL]

POSKOM HIGH FREQUENCY GENERATOR

PXR SERIES (FOR 1-TUBE)

POSKOM CO., LTD


#405, Unitech ville, 1141-2, BAEKSUK-DONG, ILSAN-GU, KOYANG, REPUBLIC OF KOREA
TEL: 82-31-906-9007,FAX: 82-31-908-4208, E-mail: poskom@poskom.co.kr, URL:www.poskom.com
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

CONTENTS

INPUT POWER SECTION

1. Input Power Selection ------------------------------- 4

BOARD kV & mA CALIBRATION

1. Board Calibration ------------------------------- 6


1.1 HT Control Calibration ------------------------------- 7
1.2 HT Control SILK SIRCUIT ------------------------------- 7
1.3 FILAMENT Calibration ------------------------------- 8
1.4 Aging ------------------------------- 8
1.5 kV Calibration ------------------------------- 9
1.6 mA Calibration ------------------------------- 10
1.7 Bucky Type Selection ------------------------------- 11
12
14
14

DOCUMENT NO.: PXR-DF-03001


-1-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

CONTENTS

OPERATION PANEL CALIBRATION

1. Operation Panel Description ---------------------------- 17


. 1.1 DIP SW Function ---------------------------- 20
2. Radiography Selection in Normal Mode ---------------------------- 20
. 2.1 Radiography Selection in Normal Mode ---------------------------- 21
. 2.2 Setting APR Radiography ---------------------------- 23
. 2.3 Setting AEC Radiography ---------------------------- 24
. 2.4 Radiography ---------------------------- 25
3. Calibration Mode ---------------------------- 26
3.1 Display of Panel in Calibration Mode ---------------------------- 33
3.2 KVREF/MAREF Setting ---------------------------- 34
3.3 KVREF/MAREF Calibration Details ---------------------------- 36
3.4 Guide to Each Calibration Details ---------------------------- 44
4.5 Initialization of Calibration Data ---------------------------- 49
4.6 Initialization Details of Calibration Data ---------------------------- 50
4 Counter Mode ---------------------------- 51
Panel Fuction On AEC Calibration Mode ---------------------------- 55
How to Calibrate AEC Calibration Mode ---------------------------- 59
. Trouble ShootingGuide ---------------------------- 62
. Error Code Guide ---------------------------- 66
Check Code Guide ---------------------------- 69
9. HT-Controller A/D Reference Table ---------------------------- 70

WIRE CONNECTION

1. HV Tank Connection ---------------------------- 76


2. I/F Board Connection ---------------------------- 77
3. AEC Board Connection ---------------------------- 79

DOCUMENT NO.: PXR-DF-03001


-2-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Pre-Installation

DOCUMENT NO.: PXR-DF-03001


-3-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5

DOCUMENT NO.: PXR-DF-03001


-4-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TABLE OF CONTENTS

1. INTRODUCTION 1
2. PRE-INSTALLATION
2.1 Physical Specifications
2.2 Mounting and Service Area
3. REQUIREMENTS
3.1 Environmental Requirements
3.1.1 Altitude, Temperature and Humidity

3.2 Electrical Requirements


3.2.1 Power Line Requirements
3.2.2 Recommended Wire Size
3.2.3 Interconnection and Grounding Requirements
3.3 Safety Devices

DOCUMENT NO.: PXR-DF-03001


-5-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

1. INTRODUCTION

Pre-Installation document provides the information and data needed to plan and
quality the customer site prior to equipment delivery and installation.
This document considers only the generator and its associated components.
Product information, environmental and electrical requirements are specified.
For system-related requirements, such as room layout, and system interconnections,
refer to documentation provided with other subsystems.

DOCUMENT NO.: PXR-DF-03001


-6-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. PRE-INSTALLATION DATA

This section provides product information and illustrations showing physical


dimensions, weight, mounting holes and normal access areas for cabling and service.

2.1 PHYSICAL SPECIFICATIONS

Fig. 2-1 STRUCTURE AND FIGURES (20kW ~ 50kW)

DOCUMENT NO.: PXR-DF-03001


-7-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

400. 0 mm

200. 0 mm
36. 0 mm

42. 0 mm

Fig.2-2 CONFIGULATION OF OPERATION PANEL

2-2. MOUNTING AND SERVICE AREA

The normal method of mounting each generator component is the following :


• HFG : Floor freestanding or anchor to floor
• Console : Desk freestanding or anchor to an optional pedestal.

The minimum recommended free area for service


SURFACE
Left side Right side Front Rear Top Bottom
HFG 0.5m 0.5m 1m - Free -
Console 0.1m 0.1m Free 0.1m Free -

DOCUMENT NO.: PXR-DF-03001


-8-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3. REQUIREMENTS

3.1 ENVIRONMENTAL REQUIREMENTS

3.1.1 ALTITUDE, TEMPERATURE AND HUMIDITY


• Altitude : Under 2400m(8000ft) sea level
• Temperature : 10℃(50℉) ~ 38℃(100℉)
• Humidity : Average relative humidity not exceed 95%

3.2 ELECTRICAL REQUIREMENTS

3.2.1 POWER LINE REQUIREMENTS


• Operation
PXR-201B PXR-321B PXR-401B PXR-501B
Max. kW 20 kW 32 kW 40 kW 50 kW
mA Range 10 ~ 250 mA 10 ~ 400 mA 10 ~ 500 mA 10 ~ 630 mA
kV Range 40 ~ 125 kV 40 ~ 125 kV 40 ~ 125 kV 40 ~ 150 kV
Power Line 220/240 VAC 220/240 VAC 220/240 VAC 220/240 VAC
1Φ, 50/60Hz 1Φ, 50/60Hz 1Φ, 50/60Hz 1Φ, 50/60Hz

• RMS line current


Line Voltage PXR-201B PXR-321B PXR-401B PXR-501B
220 VAC 100 A 177 A 208 A 210 A
240 VAC 83 A 155 A 184 A 186 A
Stand-by 0.6 A 0.6 A 0.6 A 0.6 A

• Max. Power line impedance


Line Voltage PXR-201B PXR-321B PXR-401B PXR-501B
220 VAC 0.94 0.78 0.64 0.49
240 VAC 0.98 0.85 0.71 0.56

DOCUMENT NO.: PXR-DF-03001


-9-
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3.2.2 RECOMMENDED WIRE SIZE


• Minimum wire size from distribution transformer to room electrical cabinet
Line Wire Size
Model
Voltage 15m 30m
220VAC 30 ㎟ AWG 1/0 65 ㎟ AWG 4/0
PXR-201B
240VAC 30 ㎟ AWG 1/0 65 ㎟ AWG 4/0
220VAC 50 ㎟ AWG 1/0 105 ㎟ AWG 4/0
PXR-321B
240VAC 50 ㎟ AWG 1/0 105 ㎟ AWG 4/0
220VAC 70 ㎟ AWG 2/0 120 ㎟ AWG 5/0
PXR-401B
240VAC 70 ㎟ AWG 2/0 120 ㎟ AWG 5/0
220VAC 80 ㎟ AWG 3/0 120 ㎟ AWG 5/0
PXR-501B
240VAC 80 ㎟ AWG 3/0 120 ㎟ AWG 5/0

3.3 SAFETY DEVICES

• Devices such as Safety Switch, Warning Light, and Door Interlock Switch should
be supplied and installed by the customer.

3.3.1 SAFETY SWITCH


• The main Safety Switch should be installed in the Room Electrical Cabinet, and
provided with light indicators for “Power On / Off”.

3.3.2 DOOR INTERLOCK SWITCH


• The door interlock switch indicates to the operator when doorways to the X-ray
room are open.
• It inhibits or not the X-ray generation, according to the local standards and customer
preference.
• This switch should be installed in the entrance doors and its connecting cable should
be routed to the HFG.

3.3.3 WARNING LIGHT


• The warning light is a signal lamp installed outside of the X-ray room that
indicates an X-ray exposure is On.
• The warning light connection cable should be routed to the HFG.

DOCUMENT NO.: PXR-DF-03001


- 10 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Installation

DOCUMENT NO.: PXR-DF-03001


- 11 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5

DOCUMENT NO.: PXR-DF-03001


- 12 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TABLE OF CONTENTS

DOCUMENT NO.: PXR-DF-03001


- 13 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

1. INTRODUCTION

The installation process depends on generator and system configuration.

DOCUMENT NO.: PXR-DF-03001


- 14 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. UNPACKING AND INSTALLATION

The generator is shipped in one box to facilitate transport and installation.


Upon receipt of the X-ray unit and associated equipment, inspect all shipping
contains for signs of damage. If damage is found, notify the carrier or his agent
immediately.

1. Open the shipping box, then remove the Operation Panel, Interconnecting Cables
and other furnished parts.
2. Remove the HFG from the shipping pallet and place it near to its site in the room.
At least two people are required for this operation.
3. When the HFG is unpacked, check the part numbers and serial numbers of each
component with its identification labels, and inspect all pieces for visible damage.
If any damaged parts are found, repair or order replacements to prevent
unnecessary delay in installation.
4. Remove the HFG covers for cable connections.

DOCUMENT NO.: PXR-DF-03001


- 15 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3. CABLE CONNECTION

3.1 INTRODUCTION

This section provides the information necessary to connect the HFG with the system
and options.

3.2 POWER LINE CONNECTION

This HFG is permanently connected to the power line, and powered on unless the
safety switch installed in the room electrical cabinet is off.

Internal parts of the generator are permanently powered on through power line
although the OP is off. Be sure that the safety switch is off before handling any
internal part of the equipment.

1. Verify that the power supply line is “OFF” in the room electrical cabinet.
2. The power supply line should be according to the HFG model defined in the
“Pre-Installation” document.

DOCUMENT NO.: PXR-DF-03001


- 16 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[Fig. 1]

[Fig. 2]

Input Connection Remark


AC 200V ①(AC 0V) + ②(AC 200V)
AC 220V ①(AC 0V) + ③(AC 220V)
AC 240V ①(AC 0V) + ④(AC 240V)

DOCUMENT NO.: PXR-DF-03001


- 17 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3.3 X-RAY TUBE CONNECTION

3.3.1 HIGH VOLTAGE CABLES


• Anode and Cathode cables are furnished according to the room layout.
1. Assemble the mounting accessories of each termination plug following the cable
manufacturer’s instruction.
2. Prepare the high voltage terminals that will be installed in the X-ray Tube
receptacles. Apply silicone paste over the entire surface of the plug including the
pins.
3. Carefully connect the Anode and Cathode cables from the HV Transformer(High
Voltage Tank) into the respective X-ray Tube receptacles. Insure that all
connections are made correctly, maintaining correct Anode and Cathode
orientation. Tighten the cable nuts securely.
4. Carefully connect the Anode and Cathode cables from X-ray tube into the
respective HV Transformer receptacles. Insure that all connections are made
correctly, maintaining correct Anode and Cathode orientation.

3.3.2 STATOR(ROTOR) CABLE


• X-ray tubes equipped with the stator cable installed.
• Route the stator cable together with the HV cable to the HFG.
• Connect the stator cable terminals to the indicated terminal block TB 502 on
Interface Board.
Stator Wires INTERFACE BOARD [TB 502]
MAIN ROTOR 1
AUX ROTOR 2
COM ROTOR 3

• Make sure that stator wires are properly connected. Before making any exposure,
check that the anode rotates correctly.
• If the X-ray tube is provided with the Temperature switch signal, the cable should
be routed to the terminal block TB 502 of Interface Board in the HFG and
connected to the following terminals.
Stator Wires INTERFACE BOARD [TB 502]
TEMPERATURE RELAY THERM. PR.
TEMPERATURE RELAY THERM PR

DOCUMENT NO.: PXR-DF-03001


- 18 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

• If an X-ray tube is not provided with thermostat signal, jump both connections in
the terminal block TB 502.

DOCUMENT NO.: PXR-DF-03001


- 19 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

4. FINAL INSTALLATION AND CHECKS

4.1 HV TRANSFORMER(HIGH VOLTAGE TANK)

• The HV Transformer contains isolation oil. Check that there is not oil leakage.
• If there is, check that the oil level is approximately 12 ㎜ of the top surface of the HV
transformer.

4.2 CABLE FASTENING AND COVERS

• Check that all electrical connections are firm and secure, and all cables are
correctly routed.
• Re-install the HFG covers and connect its internal ground wires.
• Pass the power line, high voltage and interconnections cable through the
respective cable outlets.

DOCUMENT NO.: PXR-DF-03001


- 20 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5. SYSTEM INTERCONNECTIONS

5.1 SYSTEM INTERCONNECTIONS

• The HV Transformer contains isolation oil. Check that there is not oil leakage.
• If there is, check that the oil level is approximately 12 ㎜ of the top surface of the HV
transformer.

5.2 HV TRANSFORMER(HV TANK) INTERCONNECTIONS

T2 T1 OIL INJECTION

CATHOD ANODE

GND

TERMINAL BLOCK

[Fig. 3]

DOCUMENT NO.: PXR-DF-03001


- 21 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* TERMINAL BLOCK
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)

DOCUMENT NO.: PXR-DF-03001


- 22 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5.3 INTERFACE BOARD INTERCONNECTIONS

TB502

CN503

CN502

TB501

TB503
TB504

CN501 CN505 CN504

[Fig. 4]

DOCUMENT NO.: PXR-DF-03001


- 23 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5.3.1 Interface Board Connection Map


TB 504 (BR-944C) TB 503 (BR-944C)
PIN I/O Signal PIN I/O Signal
1 O BUCKY1 SPLY 1 I DOOR
2 O BUCKY1 DR 2 I GND
3 O BUCKY1 EXP1 3 I OTHER DEVICE 1
4 O BUCKY1 EXP2 4 I GND
5 O SPARE OUT1-1 5 I OTHER DEVICE 2
6 I SPARE OUT1-2 6 I GND
7 I SPARE OUT2-1 7 I SPARE IN 1
8 O SPARE OUT2-2 8 I GND
9 O BUCKY2 SPLY 9 I SPARE IN 2
10 O BUCKY2 DR 10 I GND
11 O BUCKY2 EXP1 11 O LIGHT IN USE-1
12 O BUCKY2 EXP2 12 O LIGHT IN USE-2
13 I BUCYK2 OK 13 O WARNING 1-1
14 I BUCYK2 OK (GND) 14 O WARNING 1-2
15 O BUCYK1 OK 15 O WARNING 1-3
16 O BUCYK1 OK (GND) 16 O WARNING 2-1
17 O +24V 17 O WARNING 2-2
18 O 24V GND 18 O WARNING 2-3

TB 502 (BR-944C) TB 501 (BR-944C)


PIN I/O Signal PIN I/O Signal
1 O ROTOR 1 1 25 VAC
2 O ROTOR 2 2 16 VAC
3 O ROTOR 3 3 0 VAC
4 O GND 4 220 VAC
5 I TUBE THERMAL 5 110 VAC
6 I TUBE THERMAL 6 0 VAC
7 O 25 VAC(10A)
8 O 16 VAC(10A)
9 O 0 VAC
10 O 220 VAC
11 O 110 VAC
12 O 0 VAC

DOCUMENT NO.: PXR-DF-03001


- 24 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[Fig. 5]

DOCUMENT NO.: PXR-DF-03001


- 25 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5.4 AEC BOARD INTERCONNECTIONS

Ion Chamber
AEC BOARD

AEC2
9pin D-sub
(BUCKY2)
CN602 Cable
CN603

Ion Chamber
CN601 CN604
9pin D-sub
Cable
AEC1
(BUCKY1)

Remark

Remark
shield cable

1
1
1 1
2 2
3 3
4 4
5
5
6 6
7 7
8 8
9 9

9 9
외함 GND
15m

[Ion Chamber Side] [AEC Board Side]

Connection Table
NO Wire Color Start Point End Point Length Remark Function
1 BLACK Ion -CN1 1 AEC CN60 5 15m N/C
2 UL2464 BROWN Ion -CN1 2 AEC CN60 4 15m FIELD 2
3 AWG24 RED Ion -CN1 3 AEC CN60 3 15m FIELD 1
4 9P(SHIELD CABLE) ORANGE Ion -CN1 4 AEC CN60 2 15m RESET
5 YELLOW Ion -CN1 5 AEC CN60 1 15m OUTPUT
6 GREEN Ion -CN1 6 AEC CN60 9 15m FIELD 3
BLUE Ion -CN1 7 AEC CN60 8 15m -15V
8 VIOLET Ion -CN1 8 AEC CN60 7 15m +15V
9 WHITE/SHIELD Ion -CN1 9 AEC CN60 6 15m GND

[Fig. 6]

DOCUMENT NO.: PXR-DF-03001


- 26 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Board Calibration

DOCUMENT NO.: PXR-DF-03001


- 27 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5

DOCUMENT NO.: PXR-DF-03001


- 28 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TABLE OF CONTENTS

DOCUMENT NO.: PXR-DF-03001


- 29 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

1. INTRODUCTION

DOCUMENT NO.: PXR-DF-03001


- 30 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. BOARD CALIBRATION

2.1 HT CONTROL BOARD CALIBRATION

VR1 Oscillator frequency calibration for filament drive


• Measuring Point:U18-3 PIN
• Adjust Frequency:36kHz ± 0.5kHz(55.6±1.6μsec).
• Caution:Measure as 2-PULSE since it is alternating current.

VR2 Filament LAMP DEAD TIME calibration


• Caution: Do not calibrate but just turn to the left (counter clock-wise).

VR3 Small Scale (10mA ~ 80mA) mA FEEDBACK to mAs adjustment with VF converter
• Measuring Point:TP5
• Adjust the VR3 until the TP5 set to 25kHz± 0.5kHz.
• Simple Adjustment: Connect mAs tester to the connector 8,9(mA TEST)of HV tank
and shoot x-ray and adjust mAs.
• Conditions: 40kV 50mA / 10mAs
• Measuring: 10 ± 0.2mAs

VR4 Oscillator Frequency Adjustment for kV DRIVE


• Measuring Point:TP7
• Adjust Frequency:40kHz ± 0.5kHz (50±1.2μsec)
• Caution:Measure as 2-PULSE since it is alternating current.

VR6 Floroscophy Scale Calibration


• Measuring Point: U19-9 Pin
• Adjust Frequency:10.35kΩ
• Caution:Calibrate only when Floroscophy Calibration.

DOCUMENT NO.: PXR-DF-03001


- 31 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.2 HT CONTROL SILK CIRCUIT

DIP SW1

TP15
GND

VR4
DIP SW2

VR6

TP8
kV FB

VR2

TP4
mA FB VR3 VR1

[Fig. 1]

DOCUMENT NO.: PXR-DF-03001


- 32 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.3 FILAMENT DRIVE CALIBRATION

VR1 FILAMENT TRANS DIRVE Power Voltage Calibration


• Measuring Point : TP8, TP9
• Adjusting Point:DC130V±1V

DOCUMENT NO.: PXR-DF-03001


- 33 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.4 FILAMENT DRIVE SILK CURCUIT

TP8

VR1

TP9

[Fig. 2]

DOCUMENT NO.: PXR-DF-03001


- 34 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.5 AGING

How to Calibrate : Execute on Operation Panel


Measuring Point : (Actual Current : less than 40mA )

CONDITIONS SHOOTING TIME SHOOTING INTERVAL


40kV/50mA 0.01sec Two Times 20~30 Sec.
50kV/50mA 0.01sec Two Times 20~30 Sec.
60kV/50mA 0.01sec Two Times 20~30 Sec.
80kV/50mA 0.01sec Two Times 20~30 Sec.
40kV/50mA 0.1sec Two Times 50~60 Sec.
50kV/50mA 0.1sec Two Times 50~60 Sec.
60kV/50mA 0.1sec Two Times 50~60 Sec.
80kV/50mA 0.1sec Two Times 50~60 Sec.
90kV/50mA 0.1sec Two Times 20~30 Sec.
100kV/50mA 0.01sec Two Times 20~30 Sec.
110kV/50mA 0.01sec Two Times 20~30 Sec.
120kV/50mA 0.01sec Two Times 20~30 Sec.

DOCUMENT NO.: PXR-DF-03001


- 35 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.6 kV CALIBRATION

Calibration Method: Act in the panel calibration Mode(From C1-00 to 20)


• Setting range:
mA Sec kV Remark
10mA 0.05sec 40kV, 80kV, 125kV
100mA 0.05sec 40kV, 80kV, 125kV
200mA 0.05sec 40kV, 80kV, 125kV
320mA 0.05sec 40kV, 80kV, 125kV

• Measuring Point:HT-CONT TP8:kV FEEDBACK


• Converted Value:1V=33.3 kV
• Adjusting Value:±10% for the setting kV
• Caution: Execute the final test with correct mA value. kV could be changed by the mA.

DOCUMENT NO.: PXR-DF-03001


- 36 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.7 mA CALIBRATION

Calibration Method: Execute in panel calibration mode.


(from C1-21 39 : mA, from C2-21 39 : Pre_mA)
• Setting Range:
mA Sec kV Remark
10mA 0.1sec 40kV, 80kV, 125kV
100mA 0.1sec 40kV, 80kV, 125kV
200mA 0.1sec 40kV, 80kV, 125kV
320mA 0.1sec 40kV, 80kV, 125kV

• Measuring point:HT-CONT TP4


• mA FEEDBACK conversion
(10~99mA : 1V = 10mA, 100~999mA : 1V = 100mA)
• Adjusting value :±10% for the setting kV
• Caution : Execute the mA variation from C1-21 to 39.
In case of overshoot and undershoot、act from C2-21to 39 for the stable mA levels.

DOCUMENT NO.: PXR-DF-03001


- 37 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.8 mA OSCILLOSCOPE WAVE FORM

1) NORMAL CONDITIONS

[Fig. 3]

2) UNDERSHOOT WAVE FORM


Operate in panel calibration mode (From C2-21 to 39). Decrease DATA.

[Fig. 4]

DOCUMENT NO.: PXR-DF-03001


- 38 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3) OVERSHOOT WAVE FORM


Act in Panel calibration mode (From C2-21 to 39). Decrease DATA.

[Fig. 4]

DOCUMENT NO.: PXR-DF-03001


- 39 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2.9 BUCKY TYPE SELECTION

Select the bucky type by choosing no. 4 of DIP SW 1 on HT-Controller.

DIP SW 1-4 Bucky Type Remark


On Motor-Type Bucky
Off Solenoid-Type Bucky

2.10 BUCKY O.K SIGNAL

Select the bucky O. K. signal by choosing no.1 of DIP SW 2 on HT-Controller.

DIP SW 2-1 Use or Non-use of signal Remark


On Use Bucky O. K Signal
Off Not use Bucky O. K Signal

DOCUMENT NO.: PXR-DF-03001


- 40 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Operation Panel Calibration

DOCUMENT NO.: PXR-DF-03001


- 41 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5

DOCUMENT NO.: PXR-DF-03001


- 42 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TABLE OF CONTENTS

DOCUMENT NO.: PXR-DF-03001


- 43 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

OPERATION PANEL DESCRIPTION

22
10
12
16 5
9 3 4 11 1 2 6

23 24

13 14 15 20 17 21 7 8
18 19

1 Power ON switch 13 Setting switch for APR mode


2 Power OFF switch 14 Selection switch of Region and Display
3 KV Display/Setting switch 15 Setting switch of Direction and Display
4 mAs Display/Setting switch 16 Setting switch of a patient size and Display
5 ‘mAs/sec Display/Setting switch 17 Setting switch of AEC mode and Display
6 ‘mAs/sec selection switch 18 BACKUP Release switch
7 Preparation switch for Exposure 19 Setting switch of DENSITY and Display
8 X-ray Exposure switch 20 BUCKY1 & 2 selection switch
9 “OVER” LED (Red) 21 AEC Field Selection and Display
10 “ERROR” LED (Red) 22 Exposure hand switch
11 Ready LED (Green) 23 Small Focus Display
12 X-ray Exposure (Yellow) 24 Large Focus Display

DOCUMENT NO.: PXR-DF-03001


- 44 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[1]Power on Switch
Switch on the power supply

[2]Power off Switch


Switch off the power supply

[3]kV Display / Setting Switch


Set the kV for the exposures. The calibrated kV will be displayed on the menu.

[4] mA Display / Setting Switch


Set the mA for the exposures. The calibrated mA will be displayed on the menu.

[5] mA/sec Display / Setting Switch


Set the mA/sec for the exposures. The calibrated mA/sec will be shown on the menu.

[6]mAs/sec selection Switch


Select mAs/sec exposure mode.
• mAs/sec selection「mAs」Mode
•「sec」Mode for the exposure time
Indicate the selected exposure mode on the menu.

[7]Exposure Ready Switch (READY Switch)


When this switch is pushed, the exposure ready action [READY] is executed.

[8]X-ray Exposure Switch(X-RAY Switch)


Press Exposure Ready Switch and when the [11] Exposure ready LED is lighted on, press
the x-ray exposure switch at the same time.

[9]OVER Indication(Red)
The LED is switched on when operation exceeds radiography conditions. The exposure
is prohibited while [the over LED] is lighted on. In case the LED is turned on due to
changed exposure conditions in the APR mode, the memory of condition should not be
saved. Under the AEC mode selection, if radiography time exceeds backup time, it
blinks. Backup release switch cancels over display at the time of AEC mode.

[10]ERROR Indicator(Red)
The LED is turned on when an error occurs.

[11]Ready Condition Indicator(Green)


The LED is turned on when X-ray exposure changes into a possible state.

[12]X-ray Exposure Indicator (X-Ray Signal : Yellow)


The light is switched on during the exposure.

DOCUMENT NO.: PXR-DF-03001


- 45 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[13]APR Mode setting switch


This switch sets APR mode as On or Off.

[14]Selection Switch of Parts and Display


It sets the body parts to be radiographed. LED displays the selected location.

[15]Setting switch of Direction and Display.


The direction of radiography is set up. LED displays the selected direction.

[16]Setting switch of a Patient Size and Display and Radiography memory switch
This switch is used for selecting Exposure size. And selected size is displayed. And in
that case of APR Exposure condition revision, store it.
* It is prohibited under DENSITY≠0, and OVER Display on to store APR Exposure
condition.

[17]AEC Mode Set-up switch and display


This switch sets AEC mode as On or Off.

[18]BACKUP Release Switch


This switch resets BACKUP and CHECK CODE. Some parts of check code are excluded.

[19] Setting switch of Density and Display (APR,AEC Selection)


The density level sets exposure density, which is displayed.
It could affect Step setting of mAs/sec by APR Selection.

[20] BUCKY1 Selection Switch and Display


It sets BUCKY1(Table Type)
[21] BUCKY2 Selection Switch and Display
It sets BUCKY2(Stand Type)
[22] Hand Switch for exposure
It is for X-Ray Exposure Operation. 1st step Switch is for X-ray Exposure ready and 2nd
step Switch is for operation of X-ray Exposure.
[23]Small-focus Display
The light is switched on when the small focus of the tube is selected.
[24]Large- focus Display
The light is switched on when the large focus of the tube is selected.

DOCUMENT NO.: PXR-DF-03001


- 46 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. DIP SW FUNCTION

Adjust DIP Switch SW32 (8P) located on the back front of control panel and operate the each
calibration.
DIP SW (SW 32: 8P) Function Display
DIPSW-1 ON (DIPSW-2,3,4 OFF):Calibration Mode
DIPSW-2 ON (DIPSW-1,3,4 OFF):Not in use
DIPSW-3 ON (DIPSW-1,2,4 OFF):Not in use
DIPSW-4 ON (DIPSW-1,2,3 OFF):Not in use
DIPSW-5 ON :AEC selection admission
DIPSW-6 ON :Not in use
DIPSW-7 ON :Not in use
DIPSW-8 ON :Select Exposure End Buzzer
DIPSW-1,2,3,4 ON:Counter Mode

DIP SW-1 : ON (DIP SW-2,3,4 OFF) Calibration Mode


Execute the initializing of KVREF, mAREF, setting of each calibration and data.

DIP SW-5:ON Admission of AEC selection


Enable to Select SW-ON:AEC Mode.

DIP SW-8:ON Select Exposure End Buzzer


SW-OFF: Select Exposure On Buzzer.

DIP SW-1,2,3,4 ON:Counter Mode


Display Frequency of exposure, exposure time and frequency of Error & Check Code
occurrence.

DOCUMENT NO.: PXR-DF-03001


- 47 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3. RADIOGRAPHY SELECTION IN NORMAL MODE

This equipment is available with following exposure mode.

1) MANUAL EXPOSURE MODE


Execute x-ray exposure by manual settings

2) APR EXPOSURE MODE


Execute X-ray Exposure after setting up radiographic conditions with selection of target
regions, directions and patient size. You can save each exposure conditions and levels.

3.AEC Exposure Mode


Execute the X-ray exposure with Ion Chamber control.
You can calibrate the range(-3 ~±0~ +3) with density setting switch.

DOCUMENT NO.: PXR-DF-03001


- 48 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3-1.SETTING MANUAL RADIOGRAPHY

1) Select the manual radiography mode


Clear up the conditions by pushing [13] APR setting switch and [17] AEC setting switch when
the APR mode and AEC mode are selected.

2) Select “BUCKY1" or "BUCKY 2"


Push the BUCKY1 [20] (table BUCKY) or BUCKY2 [20] (Stand BUCKY) switch and select the
mode if a radiography using BUCKY should be used.

3) A setup of tube voltage "kV"


Set up the tube voltage with “kV” [3] setting switch.
Setting range: 40kV ~ 150kV (1kV step)

4) A set up of tube current “mA”


Set up the tube current with “mA” [4] setting switch.
Setting range : 10、12.5、16、20、25、32、40、50、63、80、100、
125、160、 mA ....(Small Focus)
200、250、320、400、500、630 mA....(Large Focus)

*
Model mA Range Remark
PXR-201B 10~250 mA
PXR-321B 10~400 mA
PXR-401B 10~500 mA
PXR-501B 10~630 mA

5) Setup of Exposure mode (mAs /sec radiography)


6) Execute Exposure Mode with mAs/sec selection switch
mAs setting of Exposure accumulation time.
Select the exposure accumulation time with mAs/sec setting switch
- Setting range:0.1mAs ~ 800mAs
- Step: 0.1、0.2、0.3、0.4、0.5、0.6、0.8、1.0、1.3、1.6、2.0、2.5、3.2、4.0、
5.0、6.3、8.0、10.0、12.5、16、20、25、32、40、50、63、80、100、125、
160、200、250、320、400、500、630、800 mAs

DOCUMENT NO.: PXR-DF-03001


- 49 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Model mAs Range Remark


PXR-201B 0.1~500mAs
PXR-321B 0.1~630mAs
PXR-401B 0.1~630mAs
PXR-501B 0.1~800mAs

Exposure time『sec』setting
Select the exposure time with mAs/sec [5] setting switch.
- Setting range:0.001sec ~ 10sec
- Step: 0.001、0.002、0.003、0.004、0.005、0.006、0.008、0.010、0.013、
0.016、0.020、0.025、0.032、0.040、0.050、0.064、0.080、0.100、
0.125、0.160、0.200、0.250、0.320、0.400、0.500、0.640、0.800、
1.00、1.25、1.60、2.00、2.50、3.20、4 .00、5.00、6.3、8.0 、10sec

* It could display “OVER” when the exposure condition exceeds the item’s permitted
capacity. During that period, Exposure is prohibited. Please reset the exposure
condition.

Model sec Range Remark


PXR-201B 0.001~10 sec
PXR-321B 0.001~10 sec
PXR-401B 0.001~10 sec
PXR-501B 0.001~10 sec

DOCUMENT NO.: PXR-DF-03001


- 50 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3-2.SETTING APR RADIOGRAPHY

1) Select the APR mode


Set up the APR mode with APR [13] setting switch.
When it is selected, the LED of the switch upper-left part LED lights on.
2) Select Target region
Set up with Region [14] setting switch.
LED display on a panel moves up and down by pushing up and down switch.
Choose the target region to be taken radiography. The number of parts that can be
chosen out of 9.
3) Selection of the Directions
Set up with [15] Radiography Direction [15] setting switch.
The radiography direction is displayed on the right-hand side of the selected location.
Push the direction setting switches 1 2 3 4, and choose it out of 4.
4) Select Patient Size
Patient size is chosen with Size [16] setting switch.
The patient size that can be chosen is out of LL L M S SS BABY.
5) Selection of DENSITY
The set-up value of mAs/sec step is changed by pushing [19] “DENSITY” switch.
.DENSITY:range –1 to -3: Step down
.DENSITY:±0 : No change
.DENSITY:range+1 to +3: Step up
6) Change of Radiography Conditions
The set-up radiography conditions can be changed by the same method as a setup of manual
radiography conditions. [20][21]BUCKY1•2are decided beforehand therefore you cannot
change them. If “OVER” lights on during exposure condition revision, Esposure is prohibited.
7) Memory of the Changed Radiography Conditions
The radiography conditions could be changed by above 6) Size clauses. The memory method
pushes Patient Size setting switch(LL, L, M, S, SS, BABY) 6 chosen now for about 1 second.
If a buzzer sounds, it is an end of memorizing conditions. Patient Size switch chosen is
displayed by lighting of attached LED. If the patient size switch not chosen is pushed, it will be
set as other radiography conditions and memory will not be carried out. It does not memorize
in the following condition.
(1) OVER display lights up.
(2) DENSITY setting ≠0

DOCUMENT NO.: PXR-DF-03001


- 51 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3-3.SETTING AEC RADIOGRAPHY

1) AEC Mode Enable


Select the AEC mode with [17] AEC setting switch.

2) Selection of Field

Select the target region with setting switch. You can choose the field by pushing switches.
The selected field of Ion Chamber will be displayed on the LED panel.
You cannot make radiography without selecting the field.

3) Time should be higher than 100msec under AEC mode.

4) You can select AEC 1 or AEC 2 by setting [20]. [21] bucky 1, 2.


It can not be the ready condition if Bucky and field are not selected.
The actual mAs value will be displayed on the [5] mAs / sec LED after AEC exposure ended.

ERROR AND ERROR CHECK


-Error 29: occur when there is no feedback from Ion-chamber
CAUTION -Check 10: Occur when there is no enough x-rays for for x-ray
setting level. Make the radiography again after pushing
back-up switch and rearrange conditions in this case.

DOCUMENT NO.: PXR-DF-03001


- 52 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

4.RADIOGRAPHY

1) X-ray Radiography using Radiography switch of Control Panel


(1) Push [7] X-ray exposure preparation switch. [11] X-ray exposure ready LED is lighted
up after completing preparation.
(2) After READY LED lighting up, if [8] exposure switch and [7] exposure ready switch are
pressed together, X-ray shall exposed.
(3) During exposure, X-rays Exposure [12] displays (X-RAY) lights up, and the buzzer
turns on.
(4) A check code is displayed as to the appliances. And when a finger is lifted from a switch
on the way and exposure is interrupted.

2) X-ray radiography with hand-switch


(1) Press the 1st step (preparation switch) of Hand Switch. [11] Ready LED is lighted up
after finishing preparation for exposure.
(2) After ready LED is lighting up, press the 2nd step of Hand Switch in order to expose
x-ray.
(3) [12] x-ray Exposure LED (X-RAY) lights up during the exposure, and the buzzer is
turned on.
(4) If a finger was lifted from a switch on the way, the check code will be displayed and
exposure is interrupted.

The selection of exposure on and off buzzer is operated by the DIP SW-8.
(SW-OFF/ select exposure on buzzer, SW-ON/ select exposure(during 1sec) off buzzer)

DOCUMENT NO.: PXR-DF-03001


- 53 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5.CALIBRATION MODE (DIPSW1: ON)

* CALIBRATION MODE OUTLINE

[C-1]KVREF&MAREF SETTING (1/2)

Code Description Remark


C-1 __0 [10mA] KVREF
C-1 __1 [12.5mA] KVREF
C-1 __2 [16mA] KVREF
C-1 __3 [20mA] KVREF
C-1 __4 [25mA] KVREF
C-1 __5 [32mA] KVREF
C-1 __6 [40mA] KVREF
C-1 __7 [50mA] KVREF
C-1 __8 [63mA] KVREF
C-1 __9 [80mA] KVREF
C-1 _10 [100mA] KVREF
C-1 _11 [125mA] KVREF
C-1 _12 [160mA] KVREF
C-1 _13 [200mA] KVREF
C-1 _14 [250mA] KVREF
C-1 _15 [320mA] KVREF
C-1 _16 [400mA] KVREF
C-1 _17 [500mA] KVREF
C-1 _18 [630mA] KVREF
C-1 _19 [800mA] KVREF
C-1 _20 [999mA] KVREF

DOCUMENT NO.: PXR-DF-03001


- 54 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-1]KVREF&MAREF SETTING (2/2)

Code Description Remark


C-1 _21 [10mA] MAREF
C-1 _22 [12.5mA] MAREF
C-1 _23 [16mA] MAREF
C-1 _24 [20mA] MAREF
C-1 _25 [25mA] MAREF
C-1 _26 [32mA] MAREF
C-1 _27 [40mA] MAREF
C-1 _28 [50mA] MAREF
C-1 _29 [63mA] MAREF
C-1 _30 [80mA] MAREF
C-1 _31 [100mA] MAREF
C-1 _32 [125mA] MAREF
C-1 _33 [160mA] MAREF
C-1 _34 [200mA] MAREF
C-1 _35 [250mA] MAREF
C-1 _36 [320mA] MAREF
C-1 _37 [400mA] MAREF
C-1 _38 [500mA] MAREF
C-1 _39 [630mA] MAREF
C-1 _40 -
C-1 _41 -
C-1 _42 -
C-1 _43 -
C-1 _44 Small Focus MAREF All Shift
C-1 _45 Large Focus MAREF All Shift
C-1 _46 -

DOCUMENT NO.: PXR-DF-03001


- 55 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-2] SETTING EACH CALIBRATION LEVELS (1/3)

Code Description Remark


C-2 __0 [10mA] mAs Pulse Count Calibration
C-2 __1 [12.5mA]mAs Pulse Count Calibration
C-2 __2 [16mA] mAs Pulse Count Calibration
C-2 __3 [20mA] mAs Pulse Count Calibration
C-2 __4 [25mA] mAs Pulse Count Calibration
C-2 __5 [32mA] mAs Pulse Count Calibration
C-2 __6 [40mA] mAs Pulse Count Calibration
C-2 __7 [50mA] mAs Pulse Count Calibration
C-2 __8 [63mA] mAs Pulse Count Calibration
C-2 __9 [80mA] mAs Pulse Count Calibration
C-2 _10 [100mA]mAs Pulse Count Calibration
C-2 _11 [125mA]mAs Pulse Count Calibration
C-2 _12 [160mA]mAs Pulse Count Calibration
C-2 _13 [200mA]mAs Pulse Count Calibration
C-2 _14 [250mA]mAs Pulse Count Calibration
C-2 _15 [320mA]mAs Pulse Count Calibration
C-2 _16 [400mA]mAs Pulse Count Calibration
C-2 _17 [500mA]mAs Pulse Count Calibration
C-2 _18 [630mA]mAs Pulse Count Calibration
C-2 _19 [800mA]mAs Pulse Count Calibration

DOCUMENT NO.: PXR-DF-03001


- 56 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-2] SETTING EACH CALIBRATION LEVELS (2/3)

Code Description Remark


C-2 _21 [10mA] PREP_MAREF Calibration Data
C-2 _22 [12.5mA] PREP_MAREF Calibration Data
C-2 _23 [16mA] PREP_MAREF Calibration Data
C-2 _24 [20mA] PREP_MAREF Calibration Data
C-2 _25 [25mA] PREP_MAREF Calibration Data
C-2 _26 [32mA] PREP_MAREF Calibration Data
C-2 _27 [40mA] PREP_MAREF Calibration Data
C-2 _28 [50mA] PREP_MAREF Calibration Data
C-2 _29 [63mA] PREP_MAREF Calibration Data
C-2 _30 [80mA] PREP_MAREF Calibration Data
C-2 _31 [100mA] PREP_MAREF Calibration Data
C-2 _32 [125mA] PREP_MAREF Calibration Data
C-2 _33 [160mA] PREP_MAREF Calibration Data
C-2 _34 [200mA] PREP_MAREF Calibration Data
C-2 _35 [250mA] PREP_MAREF Calibration Data
C-2 _36 [320mA] PREP_MAREF Calibration Data
C-2 _37 [400mA] PREP_MAREF Calibration Data
C-2 _38 [500mA] PREP_MAREF Calibration Data
C-2 _39 [630mA] PREP_MAREF Calibration Data
C-2 _40 [800mA] PREP_MAREF Calibration Data
C-2 _41 [999mA] PREP_MAREF Calibration Data

DOCUMENT NO.: PXR-DF-03001


- 57 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-2] SETTING EACH CALIBRATION LEVELS (3/3)

Code Description Remark


C-2 _42 Not in Use
C-2 _43 kV Feedback Monitor setting Data
C-2 _44 BUCKY1(Table)Selection Data
C-2 _45 BUCKY2(Stand)Selection Data
C-2 _46 BUCKY1(Table) Exposure DELAY-TIME control Data
C-2 _47 BUCKY2(Stand) Exposure DELAY-TIME control Data
C-2 _48 Not in Use
C-2 _49 [AEC Mode] AEC1 Type Selection
C-2 _50 [AEC Mode] AEC2 Type Selection
C-2 _51 [AEC Mode] Exposure part1: AEC data table selection
C-2 _52 [AEC Mode] Exposure part2: AEC data table selection
C-2 _53 [AEC Mode] Exposure part3: AEC data table selection
C-2 _54 [AEC Mode] Exposure part4: AEC data table selection
C-2 _55 [AEC Mode] Exposure part5: AEC data table selection
C-2 _56 [AEC Mode] Exposure part6: AEC data table selection
C-2 _57 [AEC Mode] Exposure part7: AEC data table selection
C-2 _58 [AEC Mode] Exposure part8: AEC data table selection
C-2 _59 [AEC Mode] Exposure part9: AEC data table selection
C-2 _60 [AEC Mode] DENSITY=-3 Shift Data
C-2 _61 [AEC Mode] DENSITY=-2 Shift Data
C-2 _62 [AEC Mode] DENSITY=-1 Shift Data
C-2 _63 [AEC Mode] DENSITY=+1 Shift Data
C-2 _64 [AEC Mode] DENSITY=+2 Shift Data
C-2 _65 [AEC Mode] DENSITY=+3 Shift Data
C-2 _66 AEC Data Table 1:DENSITY Shift Data
C-2 _67 AEC Data Table 1:DENSITY Shift Data
C-2 _68 AEC Data Table 1:DENSITY Shift Data
C-2 _69 AEC Data Table 1:DENSITY Shift Data
C-2 _70 [AEC Mode] AEC1 Reference Level Data
C-2 _71 [AEC Mode] AEC2 Reference Level Data
C-2 _72 HFG kW Selection Data
C-2 _73 -

DOCUMENT NO.: PXR-DF-03001


- 58 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Code Description Remark


C-2 _74 -
C-2 _75 -
C-2 _76 -
C-2 _77 -
C-2 _78 -
C-2 _79 HFG kW Selection Data
C-2 _80 Tube Selection Data
C-2 _81 kV Range Selection Data
C-2 _82 mA Range Selection Data
C-2 _83 mAs Range Selection Data

[C-3] INITIALIZING SETTING DATA


C-3 __0: Initializing Every Data

DOCUMENT NO.: PXR-DF-03001


- 59 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5. DISPLAY OF PANEL

5-1. DISPLAY OF PANEL IN CALIBRATION MODE

[1] Power on Switch

[2] Power off Switch

[3] kV display/setting switch


Calibration Mode
kV

・ Calibration Mode Display [C-1 ⇔ C-2 ⇔ C-3]


・ Mode Selection Switch

[4]mA display/setting switch


Calibration Mode
mA

・ Calibration numbers display


・ Select the numbers

[5] mAs/sec display/setting switch


mA s Calibration Mode C-1] [C-2]
sec
・ Calibration Data Display
・ Data Change Switch

[6] mAs/sec display/setting switch


Calibration Mode [C-3]

mAs
・ Data Format
sec

[18] BACKUP Cancellation Switch


Calibration Mode [C-1]
•Monitor Calibration ⇔ Monitor Conversion Switch

DOCUMENT NO.: PXR-DF-03001


- 60 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5-2. KVREF/MAREF SETTING

• Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.

• Push the power on [1] switch and turn the power on.
Control panel is displayed as a calibration mode.

• Select the calibration mode with[3] kV setting switch.


The selected calibration mode is appeared on kV display on the panel.

C-1:KVREF&MAREF SETTING

• Select the numbers in the calibration mode with [4] mA setting switch.
The selected number in calibration mode is appeared on mA display on the panel.

*THE RANGE FOR SELECTION

__0~_18, _21~_39, _44~45

The calibration data is shown on the [5] mAs/sec display after the calibration mode and
numbers are selected.

* CALIBRATION DATA SETTING RANGE

-20∼ 00 ∼+20

DOCUMENT NO.: PXR-DF-03001


- 61 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* Execute the x-ray exposure with following methods after the data is selected in
KVREF/MAREF setting mode.

[18] Execute the conversion of monitor(the monitor after calibration)with backup release
switch. During the monitor recording, it is possible to make x-ray exposures. The setting
of mA during calibration is fixed for the time. APR•AEC Mode is not in option this time.

Calibtation Mode Calibtation Mode mA in the exposure


(KVREF setting) (MAREF setting) Monitor (Fixed)
C-1__0 C-1_21 10 mA
C-1__1 C-1_22 12.5 mA
C-1__2 C-1_23 16 mA
C-1__3 C-1_24 20 mA
C-1__4 C-1_25 25 mA
C-1__5 C-1_26 32 mA
C-1__6 C-1_27 40 mA
C-1__7 C-1_28 50 mA
C-1__8 C-1_29 63 mA
C-1__9 C-1_30 80 mA
C-1_10 C-1_31 100 mA
C-1_11 C-1_32 125 mA
C-1_12 C-1_33 160 mA
C-1_13 C-1_34 200 mA
C-1_14 C-1_35 250 mA
C-1_15 C-1_36 320 mA
C-1_16 C-1_37 400 mA
C-1_17 C-1_38 500 mA
C-1_18 C-1_39 630mA
C-1_19 C-1_40 800mA
C-1_20 C-1_41 999mA
As the mA setting is fixed, make exposure with selecting other conditions such as
kV•mAs/sec and check the movement of calibration data. When the exposure condition
exceeds the capacity, the ‘over’ LED is lighted on. While the light is turned on, exposure of
x-ray is forbidden. In case of rearrangement after x-ray exposure, operate the data setting
again with back up release switch and return the exposure monitor to calibration monitor.
• How to end the calibration mode
Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Calibration Mode.

DOCUMENT NO.: PXR-DF-03001


- 62 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

5-3. KVREF/MAREF CALIBRATION DETAILS

C-1__0 : [10mA : Small Focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 10.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(10 mA fixed).
•

C-1__1 : [12.5mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 12.5.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
• It is possible to change on the screen with [18] backup switch(12.5 mA fixed).

C-1__2 : [16mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 16.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(16 mA fixed).
•

C-1__3 : [20mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 20.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(20 mA fixed).
•

C-1__4 : [25mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(25 mA fixed).

C-1__5 : [32mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(32 mA fixed).

DOCUMENT NO.: PXR-DF-03001


- 63 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1__6 : [40mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(40 mA fixed).

C-1__7 : [50mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
• It is possible to change on the screen with [18] backup switch(50 mA fixed).

C-1__8 : [63mA : small focus] KVREF Shift Data

Shift the KVREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
• It is possible to change on the screen with [18] backup switch(63 mA fixed).

C-1__9 : [80mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 80.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(80 mA fixed).
•

C-1_10 : [100mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 100.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(100 mA fixed).
•

C-1_11 : [125mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 125.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(125 mA fixed).
•

DOCUMENT NO.: PXR-DF-03001


- 64 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1_12 : [160mA : small focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(160 mA fixed).
•

C-1_13 : [200mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 200.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(200 mA fixed).
•

C-1_14 : [250mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 250.
Data range is from –20 to +20 and starting value is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(250 mA fixed).
•

C-1_15 : [320mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 320.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(320 mA fixed).
•

C-1_16 : [400mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 400.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(400 mA fixed).
•

C-1_17 : [500mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 500.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(500 mA fixed).
•

DOCUMENT NO.: PXR-DF-03001


- 65 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1_18 : [630mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 630.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(630 mA fixed).
•

C-1_19 : [800mA : large focus] KVREF Shift Data


Shift the KVREF value from the total base when the mA is 800.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W It is possible to change on the screen with [18] backup switch(800 mA fixed).
•

C-1_20 : [999mA : large focus] KVREF Shift Data

C-1_21 : [10mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 10.
Data range is from –20 to +20 and starting point is 00.(Defalult)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(10 mA fixed).
•

C-1_22 : [12.5mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 12.5.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(12.5 mA fixed).
•

C-1_23 : [16mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 16.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(16 mA fixed).

DOCUMENT NO.: PXR-DF-03001


- 66 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1_24 : [20mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 20.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(20 mA fixed).
•

C-1_25 : [25mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(25 mA fixed).
•

C-1_26 : [32mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(32 mA fixed).
•

C-1_27 : [40mA : small focus] MAREF Shift Data

Shift the MAREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W
• It is possible to change on the screen with [18] backup switch(40 mA fixed).

C-1_28 : [50mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(50 mA fixed).
•

C-1_29 : [63mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(63 mA fixed).
•

DOCUMENT NO.: PXR-DF-03001


- 67 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1_30 : [80mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 80.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(80 mA fixed).
•

C-1_31 : [100mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 100.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(100 mA fixed).

C-1_32 : [125mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 125.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(125 mA fixed).
•

C-1_33 : [160mA : small focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 160.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(160 mA fixed).
•

C-1_34 : [200mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 200.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(200 mA fixed).
•

C-1_35 : [250mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 250.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(250 mA fixed).
•

DOCUMENT NO.: PXR-DF-03001


- 68 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1_36 : [320mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 320.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(320 mA fixed).
•

C-1_37 : [400mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 400.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(400 mA fixed).
•

C-1_38 : [500mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 500.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(500 mA fixed).
•

C-1_39 : [630mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 630.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(630 mA fixed).
•

C-1_40 : [800mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 800.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(800 mA fixed).
•

C-1_41 : [999mA : large focus] MAREF Shift Data


Shift the MAREF value from the total base when the mA is 999.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W It is possible to change on the screen with [18] backup switch(999 mA fixed).
•

DOCUMENT NO.: PXR-DF-03001


- 69 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-1 _44: Small focus MAREF ALL SHIFT Data


Shift the MAREF value from the total base .
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(Small focus : 40mA fixed)

C-1 _45 : Large Focus MAREF All Shift Data


Shift the MAREF value from the total base .
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
• It is possible to change on the screen with [18] backup switch(Large Focus:20mA fixed)

DOCUMENT NO.: PXR-DF-03001


- 70 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-2] How to calibrate each condition

• Power off by pushing switch[2]


Turn on only the first switch of DIP SW(SW32: 8P)

•Power on by pushing switch[1]


Operation Panel displays Calibration Mode

•Select setting mode by kV setting switch[3]


On kV display you can see the Selected setting mode
C-2 : setting of each calibration

•Select calibration no. by handling kV setting switch[4]


On kV display you can see the Selected setting no.
*setting no. range
__0~_83

• After selecting setting mode, no., you can see setting data on mAs/sec display[5]
You can change calibrate data with mAs/sec setting switch
Calibration data range is different from each other.

• How to end Calibration Mode


Using Power OFF switch, you can turn the power off.
It can go back to DIP SW pre calibration Mode.

DOCUMENT NO.: PXR-DF-03001


- 71 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-2]. GUIDE TO INDIVIDUAL CALIBRATION DETAILS

C-2 __0 : [10mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 10.
Data range is from 000 to 004 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __1 : [12.5mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 12.5.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __2 : [16mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 16.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __3 : [20mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 20.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __4 : [25mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 25.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __5 : [32mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 32.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __6 : [40mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 40.

Data range is from 000 to 040 and starting point is 001.


• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual –mAs
DOCUMENT NO.: PXR-DF-03001
- 72 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 __7 : [50mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 50.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __8 : [63mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 63.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __9 : [80mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 80.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __10 : [100mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 100.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __11 : [125mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 125.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C -2 __12 : [160mA : small focus] mAs Pulse Count Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 160.
Data range is from 000 to 040 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __13 : [200mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 200.
Data range is from 000 to 009 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __14 : [250mA :large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 250.

DOCUMENT NO.: PXR-DF-03001


- 73 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Data range is from 000 to 009 and starting point is 001.


• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __15 : [320mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 320.
Data range is from 000 to 014 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __16 : [400mA : large focus] PREP/MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 400.
Data range is from 000 to 019 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __17 : [500mA :large focus] PREP/MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 500.
Data range is from 000 to 024 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __18 : [630mA :large focus] PREP/MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 630.
Data range is from 000 to 024 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

C-2 __19 : [800mA :large focus] PREP/MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 630.
Data range is from 000 to 024 and starting point is 001.
• For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.

DOCUMENT NO.: PXR-DF-03001


- 74 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 __21: [10mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __22: [12.5mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __23: [16mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __24: [20mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __25: [25mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __26: [32mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __27: [40mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __28: [50mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

DOCUMENT NO.: PXR-DF-03001


- 75 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 __29: [63mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __30: [80mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __31: [100mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __32: [125mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __33: [160mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 001.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __34: [200mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 002.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __35: [250mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 002.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __36: [320mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 002.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

DOCUMENT NO.: PXR-DF-03001


- 76 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 __37: [400mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 002.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __38: [500mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 002.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __39: [630mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 003.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __40: [800mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 003.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

C-2 __41: [999mA: Small Focus] PREP_MAREF Calibration Data


Set the PREP/MAREF Calibration data in case the mA is 10mA.
Data range is from 000 to 010 and starting point is 003.
For Data= 001: mA PRE_MAREF-1 is output when exposure ready.

DOCUMENT NO.: PXR-DF-03001


- 77 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2_43 : kV Feedback Monitoring Data


It sets the levels, which monitors kV Feedback in case of exposure.
Data range is from 001 to 070 and starting point is 070. [Monitoring 70%]

C-2_44 : Bucky 1(Table) Selection Data

Data range is from 000 to 002 and starting point is 000.


000=Prohibit Bucky 1(Table) Selection
001=Permit Bucky 1(Table) Selection-control delay time
002= Permit Bucky 1(Table) Selection- control input sign for bucky movement check

C-2_45 : Bucky 2(Stand) Selection Data


Data range is from 000 to 002 and starting point is 000.
000=Prohibit Bucky 2(stand) Selection
001=Permit Bucky 2(stand) Selection-control delay time
002= Permit Bucky 2(stand) Selection- control input sign for bucky movement check

C-2_46 : Bucky 1(Table) Exposure delay time control data

Use in case C-2 __44 Bucky 1(Table) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
• Operation in Bucky 1(Table) selection
Magnet for Bucky 1(Table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.

C-2_47 : Bucky 2(stand) Exposure delay time control data

Use in case C-2 __45 Bucky 2(stand) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
• Operation in Bucky 1(Table) selection
Magnet for Bucky 2(table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.

C-2 _48 : Not IN USE

DOCUMENT NO.: PXR-DF-03001


- 78 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2_49 : [AEC Mode] AEC 1 Type Selection Data


AEC 1 Type Selection Data, Data range is from 000 to 002 and starting point is 000.
000= AEC 1 –OFF : [AEC1 : Prohibited for selection]
001= AEC 2 –ON : [AEC1 : Photo control selection]
002= AEC 2 –ON : [AEC1 : Iron chamber control selection]

C-2_50 : [AEC Mode] AEC 2 Type Selection Data


AEC 2 Type Selection Data, Data range is from 000 to 002 and starting point is 000.
000= AEC 2 –OFF : [AEC2 : Prohibited for selection]
001= AEC 2 –ON : [AEC2 : Photo control selection]
002= AEC 2 –ON : [AEC2 : Iron chamber control selection]

C-2_51: [AEC Mode] Exposure part1 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part1=AEC-OFF)

C-2_52: [AEC Mode] Exposure part2 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part2=AEC-OFF)

C-2_53: [AEC Mode] Exposure part3 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part3=AEC-OFF)

C-2_54: [AEC Mode] Exposure part4 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part4=AEC-OFF)

C-2_55: [AEC Mode] Exposure part5 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part5=AEC-OFF)

C-2_56: [AEC Mode] Exposure part6 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part6=AEC-OFF)

C-2_57: [AEC Mode] Exposure part7 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part7=AEC-OFF)

C-2_58: [AEC Mode] Exposure part8 AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part8=AEC-OFF)

DOCUMENT NO.: PXR-DF-03001


- 79 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2_59: [AEC Mode] Exposure part9AEC Data Table Selection


Select the Table among 4 types of setting AEC Data Tables
Data range is from 000 to 004, starting point is 000(exposure part9=AEC-OFF)

C-2_60 : [AEC Mode] Density =-3, Shift Data


Set the level of density as –3.
Data range is from 050 to 150 and starting point is 070.
In case data is 070, VREF setting value is 70% of the original value.

C-2_61 : [AEC Mode] Density =-2, Shift Data


Set the level of density as –2.
Data range is from 050 to 150 and starting point is 080.
In case data is 080, VREF setting value is 80% of the original data.

C-2_62 : [AEC Mode] Density =-1, Shift Data


Set the level of density as –1.
Data range is from 050 to 150 and starting point is 090.
In case data is 090, VREF setting value is 90% of the original data.

C-2_63 : [AEC Mode] Density =+1, Shift Data


Set the level of density as =+1.
Data range is from 050 to 150 and starting point is 110.
In case data is 110, VREF setting value is 110% of the original data.

C-2_64 : [AEC Mode] Density =+2, Shift Data


Set the level of density as =+2.
Data range is from 050 to 150 and starting point is 120.
In case data is 120, VREF setting value is 120% of the original data.

C-2_65 : [AEC Mode] Density =+3, Shift Data


Set the level of density as =+3.
Data range is from 050 to 150 and starting point is 130.
In case data is 130, VREF setting value is 130% of the original data.

C-2 _66: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added

DOCUMENT NO.: PXR-DF-03001


- 80 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 _67: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added

C-2 _68: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added

C-2 _69: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added

C-2_70 : [AEC Mode] AEC1 Reference Level Data


Set the level AEC1 Reference Level.
Data range is from 000 to 255(hex) and starting point is 128(2.5V).
In case the data is 026, AEC Reference Level is 0.5V.
In case the data is 051, AEC Reference Level is 1.0V.
In case the data is 077, AEC Reference Level is 1.5V.
In case the data is 102, AEC Reference Level is 2.0V.
In case the data is 153, AEC Reference Level is 3.0V.
In case the data is 179, AEC Reference Level is 3.5V.
In case the data is 204, AEC Reference Level is 4.0V.
In case the data is 230, AEC Reference Level is 4.5V.
In case the data is 255, AEC Reference Level is 5.0V.

C-2_71 : [AEC Mode] AEC2 Reference Level Data


Set the level AEC2 Reference Level.
Data range is from 000 to 255(hex) and starting point is 128(2.5V).
In case the data is 026, AEC Reference Level is 0.5V.
In case the data is 051, AEC Reference Level is 1.0V.
In case the data is 077, AEC Reference Level is 1.5V.
In case the data is 102, AEC Reference Level is 2.0V.
In case the data is 153, AEC Reference Level is 3.0V.
In case the data is 179, AEC Reference Level is 3.5V.
In case the data is 204, AEC Reference
DOCUMENTLevelNO.:
is 4.0V.
PXR-DF-03001
In case the data is 230, AEC Reference Level is 4.5V.
- 81 -
In case the data is 255, AEC Reference Level is 5.0V.
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

C-2 _79 : HFG kW Select Data


Set the kW data of HFG
Data range is 000~004, starting point is 002(32kW)
Data=000 : HFG kWÆ16kW
Data=001 : HFG kWÆ20kW
Data=002 : HFG kWÆ32kW
Data=003 : HFG kWÆ40kW
Data=004 : HFG kWÆ50kW

C-2_80 : Tube Select Data


Set Tube of HFG
Data range is 000~002, starting point is 000(Toshiba E7239X)
Data=000 : Tube=Toshiba E7239X
Data=001 : Tube=Toshiba E7252X
Data=002 : Tube=Varian RAD-14

C-2 _ 81 : kV range Select Data


Set kV Range Data
Data range is 000~001, starting point is 000(125kV)
Data=000 : kV RangeÆ40~125kV.
Data=001 : kV RangeÆ40~150kV.

C-2 _82 : mA Range Select Data


Set mA range Data
Data range is 000~004, starting point is 002(400mA)
Data=000 : mA rangeÆ200mA
Data=001 : mA rangeÆ250mA
Data=002 : mA rangeÆ400mA
Data=003 : mA range->500mA
Data=004 : mA range->630mA

C-2 _83 : mAs Range select Data


Set mAs Range Data
Data range is 000~002, starting point is 001(500mAs)
Data=000 : mAs rangeÆ320mAs
Data=001 : mAs rangeÆ500mAs
Data=002 : mAs rangeÆ630mAs

DOCUMENT NO.: PXR-DF-03001


- 82 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-3] INITIALIZATION OF CALIBRATION DATA

• Push the [2]power off switch and turn the power off.
Press 1st switch of DIP SW(SW32 : 8P)and turn it on.

• Push the [1] power on switch and turn the power on.
Control panel is shown as a calibration mode.

• Select the calibration mode with[3] kV setting switch.


The selected calibration mode appears on the kV display on the panel.

C-3:Initiation of Calibration Data

• Select setting no. of setting mode with kV setting switch.


You can see the selected no. on kV display.
Setting no. selection range
--0

• Execute the initialization of calibration data after selecting calibration mode and numbers.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.

• How to end the calibration mode


Turn it off with power off switch and the position will return to the DIP SW mode before the
start of calibration Mode.

DOCUMENT NO.: PXR-DF-03001


- 83 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

[C-3] DETAILS OF CALIBRATION DATA INITIALIZATION

C-3__0 : Initialization of the Whole Data


•(C-1__0~C-1_41)Initialization of KVREF & MAREF calibration data
•(C-2__0~C-2_69)Initialization of each data setting
•Initialization of AEC data
•Initialization of APR radiography data

Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.

DOCUMENT NO.: PXR-DF-03001


- 84 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

. AEC CALIBRATION MODE

Panel fuction at AEC Calibration Mode selection

1 Power ON switch 15 Setting switch of Direction and Display


2 Power OFF switch 16 Setting switch of a patient size and Display
3 kV setting switch display 17 Setting switch of AEC mode and Display
4 mA setting switch display
5 mAs/sec setting switch display
14 Selection switch of Region and Display

DOCUMENT NO.: PXR-DF-03001


- 85 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

APR CALIBRATION MODE SETTING

(1)Notice about APR Calibration Mode

• The operation of x-ray Exposure is prohibited under the APR Calibration Mode.
Only the changes of APR radiography conditions are possible.

• AEC Calibration Mode can be set under Calibration Mode(DIP SW-1).

• Only Ion-Chamber can be controlled.


Pls refer to the AEC Connection to know how to connect.

(2)Selection of APR Calibration Mode

• Push the [2]power off switch and turn the power off.
Press only 1st switch of DIPSW(SW 32 : 8P)and turn it on.

• Push the power on [1] switch and turn the power on.
Control panel is displayed as Calibration Mode.

DOCUMENT NO.: PXR-DF-03001


- 86 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

(3) How to calibrate AEC DATA

• AEC Reference Data Setting


1) C-2_70 : Calibrate the Level AEC1 Reference Level Data and set AEC Level by referring
HT-Controller A/D Reference Table.
(initial Default is supposed to be 128(2.5V))
2) Calibrating Density by setting AEC Level Data can cause AEC level % calibrated from Level
Data Density.
Ex)
AEC Level Data : 128 (2.5V) AEC Level Data : 153 (3V)
Density % DEC. Level(V) Density % DEC. Level(V)
-3 70 1.75V Î -3 70 2.1V
-2 80 2V LEVEL -2 80 2.4V
-1 90 2.25V Data -1 90 2.7V
0 100 128 2.5V Revisio 0 100 153 3V
+1 110 2.75V n +1 110 3.3V
+2 120 3V +2 120 3.6V

+3 130 3.25V +3 130 3.9V

• AEC Density Data setting


1) C-2_60 ~ 65 : Calibrating AEC Level Shift Data enables to calibrate AEC Level % shift
When calibrating Density.
2) Data of AEC Level Shift is Density = 0 so it could shift AEC Reference Level Data(%).
Ex)
AEC Level Data : 128(2.5V) AEC Level Data : 128(2.5V)
Density Data % Level Density Data % Level
-3 70 1.75V Î -3 25 0.62V
-2 80 2V Shift -2 50 1.25V
-1 90 2.25V Data -1 75 1.87V
0 128 100 2.5V Revision 0 128 100 2.5V
+1 110 2.75V +1 125 3.12V
+2 120 3V +2 150 3.75V
+3 130 3.25V +3 175 4.37V

DOCUMENT NO.: PXR-DF-03001


- 87 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

• Notice for AEC Level Shift setting


You cannot set data of Density –2 larger than that of Density –1.
You should set Data gradually increase or decrease.
ex)

Default Abnormal setting Normal setting


Density Data Density Data Density Data

-3 70 -3 70 -3 85
-2 80 -2 90 -2 90
-1 90 -1 80 -1 95
0 0 0
+1 110 +1 130 +1 105
+2 120 +2 120 +2 110

+3 130 +3 110 +3 115

(4)How to end AEC Calibration Mode

• Power off by using Power-OFF switch [2].


Return DIP SW before Calibration Mode started.

DOCUMENT NO.: PXR-DF-03001


- 88 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Counter Mode (DIP SW 1,2,3,4 : ON)

*Panel Function when you choose Counter Mode.

5
3 4 1 2

17

1 Power ON switch 5 mAs/sec setting switch Display


2 Power OFF switch 17 Setting switch of AEC mode and Display
3 Counter Display/
Error Code & Check Code
Display selection switch
4 kV setting switch Display

DOCUMENT NO.: PXR-DF-03001


- 89 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

How to handle Counter Mode

(1) Counter Display


Display following counter values.
• Accumulated Frequency of X-ray exposures
• Accumulated time of X-ray exposures
• Accumulated Frequency of Error code
• Accumulated Frequency of Check code

(2)Counter Mode Selection


• Push the [2]power off switch and turn the power off.
Press 1•2•3•4 switch of DIPSW(SW32 : 8P)and turn it on.
• Push the power on [1] switch and turn the power on.
Control panel appears as a Counter Mode.

(3)Selection of Counter Mode


• Push the[15]APR direction setting switch(1•2•3•4)and select counter mode menu.

A. Switch: Accumulated Frequency of X-ray exposures


Each sign of[3][4][5]are as following under the exposure frequency display
000 kV 000 mA 00 sec
[3]kV Display:Freqency
[4]mA Display:Frequency
[5]sec Display:Blank
Range:000000∼999999
(In case exceeding the maximum value : 000000)

DOCUMENT NO.: PXR-DF-03001


- 90 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

B. Switch: Accumulated Period of X-ray exposures


Each sign of accumulated period of X-ray exposure are as following
0 kV 00 mA 00 sec
[3]kV Display:Time [Hour]
[4]mA Display:Time [Minute]
[5]sec Display: Time [seconds]
Range:0 Hr. 00 Min. 00 Sec. ≈999 Hr. 59 Min. 59 Sec.
(In case exceeding the maximum value :0 Hr. 00 Min. 00 Sec. )

C. Switch: Accumulated Frequency of Error code


Each sign of accumulated frequency of Error code are as following
Err kV 01 mA 0 sec
[3]kV Display, [4]mA Display :Error code set up
[5]sec Display:Frequency of error
Selection of error code is set by kV setting switch. Range:Err01 ∼ Err30
Frequency :0 ∼9999, In case exceeding the maximum value :0

D. Switch: Accumulated Frequency of Check code


Each sign of accumulated frequency of check code are as following
Ch kV 01 mA 0 sec
[3]kV Display, [4]mA Display :Error code set up
[5]sec Display:Frequency of error
Selection of error code is set by kV setting switch. Range:Ch-01 ∼ Ch-o9
Frequency :0 ∼9999, In case exceeding the maximum value :0

DOCUMENT NO.: PXR-DF-03001


- 91 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

(4) How to end the Counter calibration mode


• Turn it off with power off switch and the position will return to the DIP SW mode before the
start of Counter Mode.

DOCUMENT NO.: PXR-DF-03001


- 92 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

ERROR CODE & CHECK CODE

When problems occurred during the use of equipment, they should be marked on the [3]kV,
[4]mA setting section. As well, the [10] error LED will be lighted on with the occurrence of
error.

• Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted
on again.

• Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code. While the check code is displayed, radiography operation is forbidden for safety. Check
code is cleared when the backup release switch is switched on. Please refer to the check code
list for the case the clearance method is different.

DOCUMENT NO.: PXR-DF-03001


- 93 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

ERROR CODE LIST

Error Code Description


Err 01 CHARGE MONITOR ERROR
• When the power voltage drops below AC160V and DC150V.

Err 02 TIME OVER ERROR AT STANDBY(Serial Communication Fail Error)


• Standby position is not cleared after the 10 sec. at the operation panel
power on( No reply from HT-CONT)

Err 03 READY OUT SIGNAL ERROR AT STANDBY


• Defects of Operation panel or HT control board

Err 04 X-RAY OUT SIGNAL ERROR AT STANDBY


• Defects of Operation panel or HT control board

Err 05 PANEL READY SWITCH INPUT ERROR AT STANDBY


• When the operation panel ready switch input is ‘on’ position.

Err 06 PANEL X-RAY SWITCH INPUT ERROR AT STANDBY


• When the operation panel ready switch input is ‘on’ position.

Err 07 HAND S/W READY INPUT ERROR AT STANDBY


• When the hand s/w ready input is ‘on’ position.

Err 08 HAND S/W X-RAY INPUT ERROR AT STANDBY


• When the hand s/w x-ray input is ‘on’ position.

Err 09 ROTOR CURRENT ERROR AT NON-EXPOSURE POSITION


• TP18 (RTR_I_FB) of HT Control Board is higher than 1V at non-exposure
action

Err 10 FILLAMENT CURRENT ERROR AT NON-EXPOSURE POSITION


• TP2 (FIL_I_FB) of HT Control Board is higher than 1.5V at non-exposure
action

Err 11 kV FEEDBACK CURRENT ERROR AT NON-EXPOSURE POSITION


• TP8 (kV_FB) of HT Control Board is higher than 1V TP8 (33.3kV) at
non-exposure action

Err 12 mA FEEDBACK CURRENT ERROR AT NON-EXPOSURE POSITION


• TP4 (mA_FB) 10-80mA setting of HT Control Board is higher than o.5V
(5mA), In case of 100-500mA setting of HT Control Board is higher than
o.1V (11mA) TP8 (33.3kV) at non-exposure action

Err 13 ROTOR CURRENT ERROR AT EXPOSURE POSITION


• TP18 (RTR_I_FB) of HT Control Board is lower than 3.0V at exposure
ready action after the 0.8 sec. With ready switch input.

Err 14 FILLAMENT CURRENT ERROR AT EXPOSURE POSITION


• TP2 (FIL_I_FB) of HT Control Board is lower than 1.5V at exposure ready
action after the 1.5 sec. With ready switch input.

DOCUMENT NO.: PXR-DF-03001


- 94 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Err 15 kV FEEDBACK CURRENT ERROR AT EXPOSURE POSITION


• After Exposure is ready, X-ray switch input and turn on HV-ENABLE
signal and within 3msec HT8 (kV_FB)setting is under %70
During exposure TP8(kV_FB) setting is under 70%
*kV Feedback monitoring data(70%) can be set on calibration
mode[C-2_43]

Err 16 mA FEEDBACK CURRENT ERROR AT EXPOSURE POSITION


• Exposure Ready Condition
TP4 (mA_FB) 10-80mA setting of HT Control Board is lower than o.5V
(5mA), In case of 100-630mA setting of HT Control Board is lower than
o.1V (11mA) after the 3 msec. with kV Feedback signal.
• During Exposure
TP4 (mA_FB) 10-80mA setting of HT Control Board is lower than o.5V
(5mA), In case of 100-630mA setting of HT Control Board is lower than
o.1V (11mA) .
Err 17 Not Defined

Err 18 X-RAY TIME OVER ERROR AT SEC MODE EXPOSURE


• X-ray end signal is not reached to the panel from HT exceeding the
setting time at the sec mode exposure

Err 19 X-RAY TIME OVER ERROR AT mAs MODE EXPOSURE


• X-ray end signal is not reached to the panel from HT exceeding the
setting time calculated by the mA×(1/2) formula at the mAs mode
exposure.

Err 20 X-RAY TIME OVER ERROR AT AEC MODE EXPOSURE


• X-ray end signal is not reached to the panel from HT exceeding the
setting time calculated by backup-timer at the AEC mode exposure.

Err 21 HV INTERLOCK ERROR


• Unlink the HV trans connection

Err 22 IGBT FAULT ERROR


• HV Tans is under overloading condition about mA (350A).

Err 23 kV FEEDBACK OVER CURRENT ERROR AT EXPOSURE POSITION


• TP8 (kV_FB) of HT Control Board is higher than 20kV during the
exposure.

Err 24 mA FEEDBACK OVER CURRENT ERROR AT EXPOSURE POSITION


• TP8 (mA_FB) of HT Control Board is higher than [mA setting value plus
50%] during the exposure.

Err 25 NO ZERO CROSS ERROR


• Impossible to detect the frequency signal.

Err 26 ZERO CROSS OVER ERROR


• Frequency signal detection over.

Err 27 BUCKY 1 MOTION ERROR


• No signal input exceeding 3 sec. during the bucky 1 motion.
• Execute error check for the calibration mode [C-2_44]=002 setting.

DOCUMENT NO.: PXR-DF-03001


- 95 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

Err 28 BUCKY 2 MOTION ERROR


• No signal input exceeding 3 sec. during the bucky 2 motions.
• Execute error check for the calibration mode [C-2_45]=002 setting.

PHOTO –TIMER SIGNAL INCREASE ERROR


Err 29 • PHOTO-TIMER SIGNAL does not increase on AEC Exposure operation.
Does not increase when 100msec have past after monitoring Exposure

Err 30 PHOTO TIMER SIGNAL RESET ERROR


• Photo timer input signal does not reset at the AEC exposure ready
position (ready switch input).
• The input value of HT control board is higher than 1.0V.

Err 31 I/F_BOARD-IN ERROR (HT-CONT)

Err 32 AEC_BOARD-IN ERROR (HT-CONT)

Err 33 SMALL-FOCUS SELECT ERROR (HT-CONT)

Err 34 LARGE-FOCUS SELECT ERROR (HT-CONT)

DOCUMENT NO.: PXR-DF-03001


- 96 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

CHECK CODE LIST

CODE Description
Ch_01 TUBE THERMOSTAT OPERATION
• The temperature of the slim and long shaped light bulb is higher than 80C.
• Thermo switch is in open position.
• Release the check code at Thermo switch off condition.
Ch_02 When another switch is pushed during exposure
• Release the check code with back up reset switch.
Ch_03 When PREP. Switch on OP PANEL is detached during exposure
• Release the check code with back up reset switch.
Ch_04 When X-RAY switch on OP PANEL is detached during exposure
• Release the check code with back up reset switch.
Ch_05 When a hand switch (PREP) is detached during exposure
• Release the check code with back up reset switch.
Ch_06 When a hand switch (X-RAY) is detached during exposure
• Release the check code with back up reset switch.
Ch_07 Shortage of power-supply capacity for exposure
• The power voltage drop is below AC160V and DC 150V.
• Release the check code with input of power.
Ch_08 Simultaneous exposure prohibition
• Other x-ray system is executed simultaneously when the exposure ready switch
was pushed.
• Release the check code by off the power of other x-ray system.
Ch_09 Warning to X-rays room door opening
• The door of x-ray room was opened when the exposure ready switch was pushed.
• Stop the exposure of x-ray in this case and the check code is released when the
door is closed.
Ch_10 AEC
• X-ray exposure does not satisfy AEC Reference Level(Voltage).
Expose again after pushing back up switch and resetting the condition .

DOCUMENT NO.: PXR-DF-03001


- 97 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

APR INITIAL TRANSMISSION DATA

[C-3]It implys the APR Initial Data transmitted by the initialization of calibration data.

*If possible, initial data is recommended to be 1 type, and in case you need several kinds of
initial data, different APR conditions about ROM should be prepared and correspond.

• APR Radiography Conditions


This data is transmitted in case of calibration data initialization[C-3__0].
Please refer to the APR radiography conditions Table for the contents of Data.

• Composition of APR Radiography Condition Data (9×4×6=216 sets)

LEVEL-1 LEVEL-2 LEVEL-3


Selections of Target Selections of Directions Selections of
region(9sets) (4sets) Patient size
(6sets)
Head 1 LL
Cervical Vertebral 2 L
Chest 3 M
Thoracic Vertebrae 4 S
Rib SS
Abdomen BABY
Vertebra
Upper Limbs
Lower Limbs

DOCUMENT NO.: PXR-DF-03001


- 98 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TROUBLE SHOOTING GUIDE

1. System & PCB Check Point (1/2)


Error No. Parts Check Point
Error 01 IGBT PC1(PC817)
LED LD1, LD2
CN404 Connection
U2(MIC4584)
POWER Q2, Q3(MTD3055EL)
CN208 Connection
MAIN R101(60W20Ω)
BR1(VUO150-16NO7)
SW1(Toggle Switch Position)
Error 02 POWER +12V
CN210 Connection
HT-Cont. CN103 Connection
CN104 Connection
MAIN J1 Connection
Error 03 OP Ready S/W
Error 04 OP X-Ray S/W
Error 05 MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 06 MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 07 OP Hand S/W Connection
MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection
Error 08 OP Hand S/W Connection
MAIN J1 Connection
HT-Cont. U14(TLP521-4)
U10(74HC14)
CN103 Connection

DOCUMENT NO.: PXR-DF-03001


- 99 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

1. System & PCB Check Point (2/2)


Error No. Parts Check Point
Error 09 - -
Error 10 - -
Error 11 - -
Error 12 - -
Error 13 HT-Cont. U29(VM48)
I/F Rotor Signal Connection
POWER F2(5A) Fuse
Error 14 FILAMENT RY1(CH2-MD12)
R11(5W2Ω)
Q2, Q5(IRFP40B)
Error 15 - -
Error 16 - -
Error 17 Not In Use
Error 18 HT-Cont. HT-Cont. PCB
Error 19 HT-Cont. U22(VFC32)
Error 20 HT-Cont. U22(VFC32)
Error 21 HT-Cont. CN107 Connection
Error 22 - -
Error 23 - -
Error 24 - -
Error 25 HT-Cont. HT-Cont. PCB
Error 26 HT-Cont. HT-Cont. PCB
Error 27 I/F CN501 Connection
HT-Cont. CN101 Connection
Error 28 I/F CN501 Connection
HT-Cont. CN101 Connection
Error 29 AEC/ABC CN603, 604 Connection
Error 30 HT-Cont. U12(74HC14)
Error 31 HT-Cont. U12(74HC14)
Error 32 HT-Cont. U12(74HC14)
Error 33 FILAMENT RY1(CH2-MD12)
Error 34 FILAMENT RY1(CH2-MD12)

DOCUMENT NO.: PXR-DF-03001


- 100 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. Check Point
Check Code No. Parts Check Point
Check 01 I/F Connection
Tube Tube Temperature
Check 02 OP Panel S/W Input
Check 03 OP Panel S/W Input
Check 04 OP Panel S/W Input
Check 05 Hand S/W Hand S/W Input
Check 06 Hand S/W Hand S/W Input
Check 07 MAIN AC Input
High-Voltage Capacitor Voltage Level
(DC 300~310V)
Check 08 I/F I/F Connection
Check 09 I/F I/F Connection
Check 10 Ion Chamber Ion Chamber Input Level

DOCUMENT NO.: PXR-DF-03001


- 101 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

HT-CONTROLLER A/S REFERENCE

* HT-Controller CPU Analog-to-Digital Input Value (1/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
1 0.02 0.039 31 0.608 1.216
2 0.039 0.078 32 0.627 1.255
3 0.059 0.118 33 0.647 1.294
4 0.078 0.157 34 0.667 1.333
5 0.098 0.196 35 0.686 1.373
6 0.118 0.235 36 0.706 1.412
7 0.137 0.275 37 0.725 1.451
8 0.157 0.314 38 0.745 1.49
9 0.176 0.353 39 0.765 1.529
10 0.196 0.392 40 0.784 1.569
11 0.216 0.431 41 0.804 1.608
12 0.235 0.471 42 0.824 1.647
13 0.255 0.51 43 0.843 1.686
14 0.275 0.549 44 0.863 1.725
15 0.294 0.588 45 0.882 1.765
16 0.314 0.627 46 0.902 1.804
17 0.333 0.667 47 0.922 1.843
18 0.353 0.706 48 0.941 1.882
19 0.373 0.745 49 0.961 1.922
20 0.392 0.784 50 0.98 1.961
21 0.412 0.824 51 1 2
22 0.431 0.863 52 1.02 2.039
23 0.451 0.902 53 1.039 2.078
24 0.471 0.941 54 1.059 2.118
25 0.49 0.98 55 1.078 2.157
26 0.51 1.02 56 1.098 2.196
27 0.529 1.059 57 1.118 2.235
28 0.549 1.098 58 1.137 2.275
29 0.569 1.137 59 1.157 2.314
30 0.588 1.176 60 1.176 2.353

DOCUMENT NO.: PXR-DF-03001


- 102 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* HT-Controller CPU Analog-to-Digital Input Value (2/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
61 1.196 2.392 91 1.784 3.569
62 1.216 2.431 92 1.804 3.608
63 1.235 2.471 93 1.824 3.647
64 1.255 2.51 94 1.843 3.686
65 1.275 2.549 95 1.863 3.725
66 1.294 2.588 96 1.882 3.765
67 1.314 2.627 97 1.902 3.804
68 1.333 2.667 98 1.922 3.843
69 1.353 2.706 99 1.941 3.882
70 1.373 2.745 100 1.961 3.922
71 1.392 2.784 101 1.98 3.961
72 1.412 2.824 102 2 4
73 1.431 2.863 103 2.02 4.039
74 1.451 2.902 104 2.039 0.078
75 1.471 2.941 105 2.059 4.118
76 1.49 2.98 106 2.078 4.157
77 1.51 3.02 107 2.098 4.196
78 1.529 3.059 108 2.118 4.235
79 1.549 3.098 109 2.137 4.275
80 1.569 3.137 110 2.157 4.314
81 1.588 3.176 111 2.176 4.353
82 1.608 3.216 112 2.196 4.392
83 1.627 3.255 113 2.216 4.431
84 1.647 3.294 114 2.235 4.471
85 1.667 3.333 115 2.255 4.51
86 1.686 3.373 116 2.275 4.549
87 1.706 3.412 117 2.294 4.588
88 1.725 3.451 118 2.314 4.627
89 1.745 3.49 119 2.333 4.667
90 1.765 3.529 120 2.353 4.706

DOCUMENT NO.: PXR-DF-03001


- 103 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* HT-Controller CPU Analog-to-Digital Input Value (3/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
121 2.373 4.745 151 2.961 5.922
122 2.392 4.784 152 2.98 5.961
123 2.412 4.824 153 3 6
124 2.431 4.863 154 3.02 6.039
125 2.451 4.902 155 3.039 6.078
126 2.471 4.941 156 3.059 6.118
127 2.49 4.98 157 3.078 6.157
128 2.51 5.02 158 3.098 6.196
129 2.529 5.059 159 3.118 6.235
130 2.549 5.098 160 3.137 6.275
131 2.569 5.137 161 3.157 6.314
132 2.588 5.176 162 3.176 6.353
133 2.608 5.216 163 3.196 6.392
134 2.627 5.255 164 3.216 6.431
135 2.647 5.294 165 3.235 6.471
136 2.667 5.333 166 3.255 6.51
137 2.686 5.373 167 3.275 6.549
138 2.706 5.412 168 3.294 6.588
139 2.725 5.451 169 3.314 6.627
140 2.745 5.49 170 3.333 6.667
141 2.765 5.529 171 3.353 6.706
142 2.784 5.569 172 3.373 6.745
143 2.804 5.608 173 3.392 6.784
144 2.824 5.647 174 3.412 6.824
145 2.843 5.686 175 3.431 6.863
146 2.863 5.725 176 3.451 6.902
147 2.882 5.765 177 3.471 6.941
148 2.902 5.804 178 3.49 6.98
149 2.922 5.843 179 3.51 7.02
150 2.941 5.882 180 3.529 7.059

DOCUMENT NO.: PXR-DF-03001


- 104 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* HT-Controller CPU Analog-to-Digital Input Value (4/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
181 3.549 7.098 211 4.137 8.275
182 3.569 7.137 212 4.157 8.314
183 3.588 7.176 213 4.176 8.353
184 3.608 7.216 214 4.196 8.392
185 3.627 7.255 215 4.216 8.431
186 3.647 7.294 216 4.235 8.471
187 3.667 7.333 217 4.255 8.51
188 3.686 7.373 218 4.275 8.549
189 3.706 7.412 219 4.294 8.588
190 3.725 7.451 220 4.314 8.627
191 3.745 7.49 221 4.333 8.667
192 3.765 7.529 222 4.353 8.706
193 3.784 7.569 223 4.373 8.745
194 3.804 7.608 224 4.392 8.784
195 3.824 7.647 225 4.412 8.824
196 3.843 7.686 226 4.431 8.863
197 3.863 7.725 227 4.451 8.902
198 3.882 7.765 228 4.471 8.941
199 3.902 7.804 229 4.49 8.98
200 3.922 7.843 230 4.51 9.02
201 3.941 7.882 231 4.529 9.059
202 3.961 7.922 232 4.549 9.098
203 3.98 7.961 233 4.569 9.137
204 4 8 234 4.588 9.176
205 4.02 8.039 235 4.608 9.216
206 4.039 8.078 236 4.627 9.255
207 4.059 8.118 237 4.647 9.294
208 4.078 8.157 238 4.667 9.333
209 4.098 8.196 239 4.686 9.373
210 4.118 8.235 240 4.706 9.412

DOCUMENT NO.: PXR-DF-03001


- 105 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

* HT-Controller CPU Analog-to-Digital Input Value (5/5)


DEC A/D Input (A/D) X 2 DEC A/D Input (A/D) X 2
241 4.725 9.451 249 4.882 9.765
241 4.745 9.49 250 4.902 9.804
243 4.765 9.529 251 4.922 9.843
244 4.784 9.569 252 4.941 9.882
245 4.804 9.608 253 4.961 9.922
246 4.824 9.647 254 4.98 9.961
247 4.843 9.686 255 5 10
248 4.863 9.725

DOCUMENT NO.: PXR-DF-03001


- 106 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

WIRE CONNECTION

DOCUMENT NO.: PXR-DF-03001


- 107 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5

DOCUMENT NO.: PXR-DF-03001


- 108 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TABLE OF CONTENTS

DOCUMENT NO.: PXR-DF-03001


- 109 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

1. HV TANK CONNECTION

T2 T1 OIL INJECTION

CATHOD ANODE

GND

CORE

* CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)

DOCUMENT NO.: PXR-DF-03001


- 110 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

2. I/F BODARD CONNECTION

TB502

CN503

CN502

TB501

TB503
TB504

CN501 CN505 CN504

DOCUMENT NO.: PXR-DF-03001


- 111 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

TB 504 (BR-944C) TB 503 (BR-944C)


PIN I/O Signal PIN I/O Signal
1 O BUCKY1 SPLY 1 I DOOR
2 O BUCKY1 DR 2 I GND
3 O BUCKY1 EXP1 3 I OTHER DEVICE 1
4 O BUCKY1 EXP2 4 I GND
5 O SPARE OUT1-1 5 I OTHER DEVICE 2
6 I SPARE OUT1-2 6 I GND
7 I SPARE OUT2-1 7 I SPARE IN 1
8 O SPARE OUT2-2 8 I GND
9 O BUCKY2 SPLY 9 I SPARE IN 2
10 O BUCKY2 DR 10 I GND
11 O BUCKY2 EXP1 11 O LIGHT IN USE-1
12 O BUCKY2 EXP2 12 O LIGHT IN USE-2
13 I BUCYK2 OK 13 O WARNING 1-1
14 I BUCYK2 OK (GND) 14 O WARNING 1-2
15 O BUCYK1 OK 15 O WARNING 1-3
16 O BUCYK1 OK (GND) 16 O WARNING 2-1
17 O +24V 17 O WARNING 2-2
18 O 24V GND 18 O WARNING 2-3

TB 502 (BR-944C) TB 501 (BR-944C)


PIN I/O Signal PIN I/O Signal
1 O ROTOR 1 1 25 VAC
2 O ROTOR 2 2 16 VAC
3 O ROTOR 3 3 0 VAC
4 O GND 4 220 VAC
5 I TUBE THERMAL 5 110 VAC
6 I TUBE THERMAL 6 0 VAC
7 O 25 VAC(10A)
8 O 16 VAC(10A)
9 O 0 VAC
10 O 220 VAC
11 O 110 VAC
12 O 0 VAC

DOCUMENT NO.: PXR-DF-03001


- 112 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

READY READY UP(ON) EXP. EXP. END

Bucky Motion
(RY3)

BUCKY1 (RY1)

BUCKY2 (RY2)

0.1sec

간찰기 (RY7)

POWER ON POWER OFF

LIGHT IN USE
(RY4)

TB 504 (BR-944C)
PIN I/O Signal
1 O BUCKY1 SPLY
2 O BUCKY1 DR BUCKY1 POWER
3 O BUCKY1 EXP1 (AC220 or AC110)
4 O BUCKY1 EXP2
5 O SPARE OUT1-1
6 I SPARE OUT1-2
7 I SPARE OUT2-1
8 O SPARE OUT2-2
9 O BUCKY2 SPLY
10 O BUCKY2 DR BUCKY1 POWER
11 O BUCKY2 EXP1 (AC220 or AC110)
12 O BUCKY2 EXP2
13 I BUCYK2 OK
14 I BUCYK2 OK (GND)
15 O BUCYK1 OK
16 O BUCYK1 OK (GND)
17 O +24V
18 O 24V GND
DOCUMENT NO.: PXR-DF-03001
- 113 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM

3. AEC BODARD CONNECTION

Ion Chamber

AEC BOARD
AEC2
9pin D-sub (BUCKY2)
CN602 Cable
CN603

Ion Chamber

CN601 CN604
9pin D-sub
Cable
AEC1
(BUCKY1)

AEC Output Signal : Positive


Remark

Remark
shield cable

1
1
1 1
2 2
Ion - CN1

3 3

AEC CN60
4 4
5
5
6 6
7 7
8 8
9 9

9 9
외함 GND
15m

[Ion Chamber Side] [AEC Board]

Connection Table
NO Wire Color Start Point End Point Length Remark Function
1 BLACK Ion - CN1 1 AEC CN60 5 15m N/C
2 UL2464 BROWN Ion - CN1 2 AEC CN60 4 15m FIELD 2
3 AWG24 RED Ion - CN1 3 AEC CN60 3 15m FIELD 1
4 9P (SHIELD CABLE) ORANGE Ion - CN1 4 AEC CN60 2 15m RESET
5 YELLOW Ion - CN1 5 AEC CN60 1 15m OUTPUT
6 GREEN Ion - CN1 6 AEC CN60 9 15m FIELD 3
7 BLUE Ion - CN1 7 AEC CN60 8 15m -15V
8 VIOLET Ion - CN1 8 AEC CN60 7 15m +15V
9 WHITE/SHIELD Ion - CN1 9 AEC CN60 6 15m GND

DOCUMENT NO.: PXR-DF-03001


- 114 -