PXR Type: B
[CALIBRATION MANUAL]
CONTENTS
CONTENTS
WIRE CONNECTION
Pre-Installation
REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5
TABLE OF CONTENTS
1. INTRODUCTION 1
2. PRE-INSTALLATION
2.1 Physical Specifications
2.2 Mounting and Service Area
3. REQUIREMENTS
3.1 Environmental Requirements
3.1.1 Altitude, Temperature and Humidity
1. INTRODUCTION
Pre-Installation document provides the information and data needed to plan and
quality the customer site prior to equipment delivery and installation.
This document considers only the generator and its associated components.
Product information, environmental and electrical requirements are specified.
For system-related requirements, such as room layout, and system interconnections,
refer to documentation provided with other subsystems.
2. PRE-INSTALLATION DATA
400. 0 mm
200. 0 mm
36. 0 mm
42. 0 mm
3. REQUIREMENTS
• Devices such as Safety Switch, Warning Light, and Door Interlock Switch should
be supplied and installed by the customer.
Installation
REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5
TABLE OF CONTENTS
1. INTRODUCTION
1. Open the shipping box, then remove the Operation Panel, Interconnecting Cables
and other furnished parts.
2. Remove the HFG from the shipping pallet and place it near to its site in the room.
At least two people are required for this operation.
3. When the HFG is unpacked, check the part numbers and serial numbers of each
component with its identification labels, and inspect all pieces for visible damage.
If any damaged parts are found, repair or order replacements to prevent
unnecessary delay in installation.
4. Remove the HFG covers for cable connections.
3. CABLE CONNECTION
3.1 INTRODUCTION
This section provides the information necessary to connect the HFG with the system
and options.
This HFG is permanently connected to the power line, and powered on unless the
safety switch installed in the room electrical cabinet is off.
Internal parts of the generator are permanently powered on through power line
although the OP is off. Be sure that the safety switch is off before handling any
internal part of the equipment.
1. Verify that the power supply line is “OFF” in the room electrical cabinet.
2. The power supply line should be according to the HFG model defined in the
“Pre-Installation” document.
[Fig. 1]
[Fig. 2]
• Make sure that stator wires are properly connected. Before making any exposure,
check that the anode rotates correctly.
• If the X-ray tube is provided with the Temperature switch signal, the cable should
be routed to the terminal block TB 502 of Interface Board in the HFG and
connected to the following terminals.
Stator Wires INTERFACE BOARD [TB 502]
TEMPERATURE RELAY THERM. PR.
TEMPERATURE RELAY THERM PR
• If an X-ray tube is not provided with thermostat signal, jump both connections in
the terminal block TB 502.
• The HV Transformer contains isolation oil. Check that there is not oil leakage.
• If there is, check that the oil level is approximately 12 ㎜ of the top surface of the HV
transformer.
• Check that all electrical connections are firm and secure, and all cables are
correctly routed.
• Re-install the HFG covers and connect its internal ground wires.
• Pass the power line, high voltage and interconnections cable through the
respective cable outlets.
5. SYSTEM INTERCONNECTIONS
• The HV Transformer contains isolation oil. Check that there is not oil leakage.
• If there is, check that the oil level is approximately 12 ㎜ of the top surface of the HV
transformer.
T2 T1 OIL INJECTION
CATHOD ANODE
GND
TERMINAL BLOCK
[Fig. 3]
* TERMINAL BLOCK
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)
TB502
CN503
CN502
TB501
TB503
TB504
[Fig. 4]
[Fig. 5]
Ion Chamber
AEC BOARD
AEC2
9pin D-sub
(BUCKY2)
CN602 Cable
CN603
Ion Chamber
CN601 CN604
9pin D-sub
Cable
AEC1
(BUCKY1)
Remark
Remark
shield cable
1
1
1 1
2 2
3 3
4 4
5
5
6 6
7 7
8 8
9 9
9 9
외함 GND
15m
Connection Table
NO Wire Color Start Point End Point Length Remark Function
1 BLACK Ion -CN1 1 AEC CN60 5 15m N/C
2 UL2464 BROWN Ion -CN1 2 AEC CN60 4 15m FIELD 2
3 AWG24 RED Ion -CN1 3 AEC CN60 3 15m FIELD 1
4 9P(SHIELD CABLE) ORANGE Ion -CN1 4 AEC CN60 2 15m RESET
5 YELLOW Ion -CN1 5 AEC CN60 1 15m OUTPUT
6 GREEN Ion -CN1 6 AEC CN60 9 15m FIELD 3
BLUE Ion -CN1 7 AEC CN60 8 15m -15V
8 VIOLET Ion -CN1 8 AEC CN60 7 15m +15V
9 WHITE/SHIELD Ion -CN1 9 AEC CN60 6 15m GND
[Fig. 6]
Board Calibration
REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5
TABLE OF CONTENTS
1. INTRODUCTION
2. BOARD CALIBRATION
VR3 Small Scale (10mA ~ 80mA) mA FEEDBACK to mAs adjustment with VF converter
Measuring Point:TP5
Adjust the VR3 until the TP5 set to 25kHz± 0.5kHz.
Simple Adjustment: Connect mAs tester to the connector 8,9(mA TEST)of HV tank
and shoot x-ray and adjust mAs.
Conditions: 40kV 50mA / 10mAs
Measuring: 10 ± 0.2mAs
DIP SW1
TP15
GND
VR4
DIP SW2
VR6
TP8
kV FB
VR2
TP4
mA FB VR3 VR1
[Fig. 1]
TP8
VR1
TP9
[Fig. 2]
2.5 AGING
2.6 kV CALIBRATION
2.7 mA CALIBRATION
1) NORMAL CONDITIONS
[Fig. 3]
[Fig. 4]
[Fig. 4]
REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5
TABLE OF CONTENTS
22
10
12
16 5
9 3 4 11 1 2 6
23 24
13 14 15 20 17 21 7 8
18 19
[1]Power on Switch
Switch on the power supply
[9]OVER Indication(Red)
The LED is switched on when operation exceeds radiography conditions. The exposure
is prohibited while [the over LED] is lighted on. In case the LED is turned on due to
changed exposure conditions in the APR mode, the memory of condition should not be
saved. Under the AEC mode selection, if radiography time exceeds backup time, it
blinks. Backup release switch cancels over display at the time of AEC mode.
[10]ERROR Indicator(Red)
The LED is turned on when an error occurs.
[16]Setting switch of a Patient Size and Display and Radiography memory switch
This switch is used for selecting Exposure size. And selected size is displayed. And in
that case of APR Exposure condition revision, store it.
* It is prohibited under DENSITY≠0, and OVER Display on to store APR Exposure
condition.
2. DIP SW FUNCTION
Adjust DIP Switch SW32 (8P) located on the back front of control panel and operate the each
calibration.
DIP SW (SW 32: 8P) Function Display
DIPSW-1 ON (DIPSW-2,3,4 OFF):Calibration Mode
DIPSW-2 ON (DIPSW-1,3,4 OFF):Not in use
DIPSW-3 ON (DIPSW-1,2,4 OFF):Not in use
DIPSW-4 ON (DIPSW-1,2,3 OFF):Not in use
DIPSW-5 ON :AEC selection admission
DIPSW-6 ON :Not in use
DIPSW-7 ON :Not in use
DIPSW-8 ON :Select Exposure End Buzzer
DIPSW-1,2,3,4 ON:Counter Mode
*
Model mA Range Remark
PXR-201B 10~250 mA
PXR-321B 10~400 mA
PXR-401B 10~500 mA
PXR-501B 10~630 mA
Exposure time『sec』setting
Select the exposure time with mAs/sec [5] setting switch.
- Setting range:0.001sec ~ 10sec
- Step: 0.001、0.002、0.003、0.004、0.005、0.006、0.008、0.010、0.013、
0.016、0.020、0.025、0.032、0.040、0.050、0.064、0.080、0.100、
0.125、0.160、0.200、0.250、0.320、0.400、0.500、0.640、0.800、
1.00、1.25、1.60、2.00、2.50、3.20、4 .00、5.00、6.3、8.0 、10sec
* It could display “OVER” when the exposure condition exceeds the item’s permitted
capacity. During that period, Exposure is prohibited. Please reset the exposure
condition.
2) Selection of Field
Select the target region with setting switch. You can choose the field by pushing switches.
The selected field of Ion Chamber will be displayed on the LED panel.
You cannot make radiography without selecting the field.
4.RADIOGRAPHY
The selection of exposure on and off buzzer is operated by the DIP SW-8.
(SW-OFF/ select exposure on buzzer, SW-ON/ select exposure(during 1sec) off buzzer)
5. DISPLAY OF PANEL
mAs
・ Data Format
sec
Push the [2]power off switch and turn the power off.
Press 1st switch of DIPSW(SW32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as a calibration mode.
C-1:KVREF&MAREF SETTING
Select the numbers in the calibration mode with [4] mA setting switch.
The selected number in calibration mode is appeared on mA display on the panel.
The calibration data is shown on the [5] mAs/sec display after the calibration mode and
numbers are selected.
-20∼ 00 ∼+20
* Execute the x-ray exposure with following methods after the data is selected in
KVREF/MAREF setting mode.
[18] Execute the conversion of monitor(the monitor after calibration)with backup release
switch. During the monitor recording, it is possible to make x-ray exposures. The setting
of mA during calibration is fixed for the time. APR•AEC Mode is not in option this time.
Shift the KVREF value from the total base when the mA is 25.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(25 mA fixed).
Shift the KVREF value from the total base when the mA is 32.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(32 mA fixed).
Shift the KVREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(40 mA fixed).
Shift the KVREF value from the total base when the mA is 50.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-:
W KVREF goes down
It is possible to change on the screen with [18] backup switch(50 mA fixed).
Shift the KVREF value from the total base when the mA is 63.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: KVREF goes up
Data=-: KVREF goes down
W
It is possible to change on the screen with [18] backup switch(63 mA fixed).
Shift the MAREF value from the total base when the mA is 40.
Data range is from –20 to +20 and starting point is 00.(Default)
Data=+: MAREF goes up
Data=-: MAREF goes down
W
It is possible to change on the screen with [18] backup switch(40 mA fixed).
After selecting setting mode, no., you can see setting data on mAs/sec display[5]
You can change calibrate data with mAs/sec setting switch
Calibration data range is different from each other.
C-2 __0 : [10mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 10.
Data range is from 000 to 004 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __1 : [12.5mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 12.5.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __2 : [16mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 16.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __3 : [20mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 20.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __4 : [25mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 25.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __5 : [32mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 32.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __6 : [40mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 40.
C-2 __7 : [50mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 50.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __8 : [63mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 63.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __9 : [80mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 80.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __10 : [100mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 100.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __11 : [125mA : small focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 125.
Data range is from 000 to 040 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __13 : [200mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 200.
Data range is from 000 to 009 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
C-2 __14 : [250mA :large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 250.
C-2 __15 : [320mA : large focus] mAs Pulse Count Calibration Data
Set the PREP/MAREF Calibration data in case the mA is 320.
Data range is from 000 to 014 and starting point is 001.
For mAs exposure, set the Pulse count to -1
For AEC exposure, set the Pulse count to +1 and act actual -mAs.
Use in case C-2 __44 Bucky 1(Table) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
Operation in Bucky 1(Table) selection
Magnet for Bucky 1(Table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
Use in case C-2 __45 Bucky 2(stand) selection Data delay time control.
Data range is from 000 to 200 and starting point is 020
When data is 020, delay time is 200msec.
Operation in Bucky 1(Table) selection
Magnet for Bucky 2(table) is called as ON under ready-up condition.
Turn the magnet off with x-ray signal input and the action after delay time is called as
HV-Enable on.
C-2 _66: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
C-2 _67: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
C-2 _68: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
C-2 _69: [AEC Mode] AEC Data Table1 Density Shift Data
Set the whole density level of AEC Data Table1
Data range is from 050 to 150 and starting point is 10.
Data= 050 : VREF setting value is 50%
Data= 150: VREF setting value is 150%
Data= 100: VREF setting value is same as before
* Density(-3~+3) added
Push the [2]power off switch and turn the power off.
Press 1st switch of DIP SW(SW32 : 8P)and turn it on.
Push the [1] power on switch and turn the power on.
Control panel is shown as a calibration mode.
Execute the initialization of calibration data after selecting calibration mode and numbers.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.
Press the [6]mAs/sec switch for a second. When the check buzzer sounds, the data
initialization is completed.
The operation of x-ray Exposure is prohibited under the APR Calibration Mode.
Only the changes of APR radiography conditions are possible.
Push the [2]power off switch and turn the power off.
Press only 1st switch of DIPSW(SW 32 : 8P)and turn it on.
Push the power on [1] switch and turn the power on.
Control panel is displayed as Calibration Mode.
-3 70 -3 70 -3 85
-2 80 -2 90 -2 90
-1 90 -1 80 -1 95
0 0 0
+1 110 +1 130 +1 105
+2 120 +2 120 +2 110
5
3 4 1 2
17
When problems occurred during the use of equipment, they should be marked on the [3]kV,
[4]mA setting section. As well, the [10] error LED will be lighted on with the occurrence of
error.
Error Code
The error code is displayed when problems, such as abnormalities, occur to equipment.
While the error code is displayed, radiography operation is forbidden for safety.
A power supply is re-switched off and on, then an error code is cleared. If the problem of
equipment is not solved at the time of a power-supply re-switched on, the display is lighted
on again.
Check code
Check codes are warning displays when an error is on operation of equipment unlike an error
code. While the check code is displayed, radiography operation is forbidden for safety. Check
code is cleared when the backup release switch is switched on. Please refer to the check code
list for the case the clearance method is different.
CODE Description
Ch_01 TUBE THERMOSTAT OPERATION
The temperature of the slim and long shaped light bulb is higher than 80C.
Thermo switch is in open position.
Release the check code at Thermo switch off condition.
Ch_02 When another switch is pushed during exposure
Release the check code with back up reset switch.
Ch_03 When PREP. Switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_04 When X-RAY switch on OP PANEL is detached during exposure
Release the check code with back up reset switch.
Ch_05 When a hand switch (PREP) is detached during exposure
Release the check code with back up reset switch.
Ch_06 When a hand switch (X-RAY) is detached during exposure
Release the check code with back up reset switch.
Ch_07 Shortage of power-supply capacity for exposure
The power voltage drop is below AC160V and DC 150V.
Release the check code with input of power.
Ch_08 Simultaneous exposure prohibition
Other x-ray system is executed simultaneously when the exposure ready switch
was pushed.
Release the check code by off the power of other x-ray system.
Ch_09 Warning to X-rays room door opening
The door of x-ray room was opened when the exposure ready switch was pushed.
Stop the exposure of x-ray in this case and the check code is released when the
door is closed.
Ch_10 AEC
X-ray exposure does not satisfy AEC Reference Level(Voltage).
Expose again after pushing back up switch and resetting the condition .
[C-3]It implys the APR Initial Data transmitted by the initialization of calibration data.
*If possible, initial data is recommended to be 1 type, and in case you need several kinds of
initial data, different APR conditions about ROM should be prepared and correspond.
2. Check Point
Check Code No. Parts Check Point
Check 01 I/F Connection
Tube Tube Temperature
Check 02 OP Panel S/W Input
Check 03 OP Panel S/W Input
Check 04 OP Panel S/W Input
Check 05 Hand S/W Hand S/W Input
Check 06 Hand S/W Hand S/W Input
Check 07 MAIN AC Input
High-Voltage Capacitor Voltage Level
(DC 300~310V)
Check 08 I/F I/F Connection
Check 09 I/F I/F Connection
Check 10 Ion Chamber Ion Chamber Input Level
WIRE CONNECTION
REVISION HISTORY
Revision No. Date Details Remarks
0
1
2
3
4
5
TABLE OF CONTENTS
1. HV TANK CONNECTION
T2 T1 OIL INJECTION
CATHOD ANODE
GND
CORE
* CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+) mA & mAs Meter
2 mA TP- mA Test Point (-) Connection
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)
TB502
CN503
CN502
TB501
TB503
TB504
Bucky Motion
(RY3)
BUCKY1 (RY1)
BUCKY2 (RY2)
0.1sec
간찰기 (RY7)
LIGHT IN USE
(RY4)
TB 504 (BR-944C)
PIN I/O Signal
1 O BUCKY1 SPLY
2 O BUCKY1 DR BUCKY1 POWER
3 O BUCKY1 EXP1 (AC220 or AC110)
4 O BUCKY1 EXP2
5 O SPARE OUT1-1
6 I SPARE OUT1-2
7 I SPARE OUT2-1
8 O SPARE OUT2-2
9 O BUCKY2 SPLY
10 O BUCKY2 DR BUCKY1 POWER
11 O BUCKY2 EXP1 (AC220 or AC110)
12 O BUCKY2 EXP2
13 I BUCYK2 OK
14 I BUCYK2 OK (GND)
15 O BUCYK1 OK
16 O BUCYK1 OK (GND)
17 O +24V
18 O 24V GND
DOCUMENT NO.: PXR-DF-03001
- 113 -
PXR SERIES (FOR 1 TUBE) CALIBRATION MANUAL POSKOM
Ion Chamber
AEC BOARD
AEC2
9pin D-sub (BUCKY2)
CN602 Cable
CN603
Ion Chamber
CN601 CN604
9pin D-sub
Cable
AEC1
(BUCKY1)
Remark
shield cable
1
1
1 1
2 2
Ion - CN1
3 3
AEC CN60
4 4
5
5
6 6
7 7
8 8
9 9
9 9
외함 GND
15m
Connection Table
NO Wire Color Start Point End Point Length Remark Function
1 BLACK Ion - CN1 1 AEC CN60 5 15m N/C
2 UL2464 BROWN Ion - CN1 2 AEC CN60 4 15m FIELD 2
3 AWG24 RED Ion - CN1 3 AEC CN60 3 15m FIELD 1
4 9P (SHIELD CABLE) ORANGE Ion - CN1 4 AEC CN60 2 15m RESET
5 YELLOW Ion - CN1 5 AEC CN60 1 15m OUTPUT
6 GREEN Ion - CN1 6 AEC CN60 9 15m FIELD 3
7 BLUE Ion - CN1 7 AEC CN60 8 15m -15V
8 VIOLET Ion - CN1 8 AEC CN60 7 15m +15V
9 WHITE/SHIELD Ion - CN1 9 AEC CN60 6 15m GND