Professional Documents
Culture Documents
Service Procedure
Date 28/03/14 SP14_02
Replacement of CB_Slave Board by 4G_SAPS Board
Page 1 / 3
Introduction:
This procedure is aimed to explain how to replace the electronic board CB Slave by the electronic board 4G SAPS.
NOTE: Both power input and serial interface cables of the machine do not change.
Needed items: Before with CB Slave. Later with 4G SAPS.
4G SAPS
Procedure:
1. Remove from the Saps following parts: The CB Slave Board; The Stirring Motor cable; Board plastic supports >>
2. Assemble on the Saps: The 4G Saps Board; The Stirring Motor cable 210mm. NOTE: Take care of inserting the board protection cover.
3. Connect the stepper motor and the stirring motor.
Stirring motor(s).
Parallel output for
5lt canister.
Stepper motor
4. Address the 4GSAPS board in the same way as it was the CB Slave board.
CB Slave hardware addressing chart: 4GSAPS hardware addressing chart:
Example:
JM1 = ON
JM2-JM3-JM4-JM6 = OFF
Example:
S1 = 0
S2 = 1
1 207014 KIT RETROFIT 4GSAPS (The kit is recommended. Here below is all the material related to the kit)
1 4000189 CB4GSAPS BL B.O
1 4000057 CABLE l=210 mm. STIRRER MOTOR
2 4000713 SCREW CRPH M3 8 DIN 7500C INOX
NOTE: The number of the hardware circuit address may be different from the circuit number on the label. (Configuration transcoding)