You are on page 1of 8

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/23191593

Determination of serum creatinine by Jaffe method and how to calibrate to


eliminate matrix interference problems

Article  in  Clinical Chemistry and Laboratory Medicine · February 2008


DOI: 10.1515/CCLM.2008.224 · Source: PubMed

CITATIONS READS

65 21,254

3 authors, including:

P. Sedlák

5 PUBLICATIONS   74 CITATIONS   

SEE PROFILE

All content following this page was uploaded by P. Sedlák on 02 February 2019.

The user has requested enhancement of the downloaded file.


Article in press - uncorrected proof
Clin Chem Lab Med 2008;46(8):1127–1133  2008 by Walter de Gruyter • Berlin • New York. DOI 10.1515/CCLM.2008.224 2007/88

Determination of serum creatinine by Jaffe method and how


to calibrate to eliminate matrix interference problems

Vratislav Chromý1,*, Kateřina Rozkošná2 and Keywords: artificial serum matrix for calibration of
Pavel Sedlák3 Jaffe method; compensation of serum interferents in
1 Jaffe method; Jaffe serum creatinine.
Institute of Chemistry, Faculty of Science, Masaryk
University, Brno, Czech Republic
2
Diagnostics Research Department, Pliva-Lachema
Introduction
Diagnostika, Brno, Czech Republic
3
Department of Clinical Biochemistry, Municipal
Most chemical methods for measuring creatinine are
Hospital, Čáslav, Czech Republic
still based on the reaction with alkaline picrate (Jaffe
1886). Creatinine reacts with picrate ion in an alkaline
medium to yield an orange-red adduct (1). The reac-
Abstract tion is not specific and many compounds are known
to produce Jaffe-like chromogens, including protein,
Background: The calibration of Jaffe method for glucose, ascorbic acid, ketone bodies, pyruvate, gua-
serum creatinine using one serum-based standard nidine, cephalosporins (2) and bilirubin, the last one
complemented with artificial matrix compensating being the negative interferent (3). The degree of inter-
standard’s Jaffe-interfering substances allows two- ference caused by these serum-present compounds
point calibration with results well comparable with (i.e., by serum matrix) is also dependent on the spe-
enzymatic methods. cific reaction conditions chosen. Manual methods
Method: Spectrophotometry was used. have traditionally been equilibrium methods with a
Results: Jaffe procedures compensating serum/plas- reaction time of approximately 15 min at room
ma intereferents by subtracting a constant amount of temperature.
creatinine poorly overcompensate creatinine in chil- Kinetic assays were developed for analyses with
dren. Two-point calibration with a pair of primary improved specificity, because some Jaffe-like adducts
serum standards certified by the reference measure- with picrate are formed rapidly during the first 20 s
ment procedure (isotope-dilution, mass spectrometry) after mixing reagent and sample (such as acetoace-
or with a pair of secondary standards linked to pri- tate) and some react slowly within 80–100 s after mix-
mary materials could provide results well agreeable ing (such as protein) and kinetic measurement
with enzymatic determination. Such a calibration between 20–60 or 20–80 s was attributed predomi-
comprises an absorbance offset corresponding to oth- nantly to the creatinine-picrate reaction (4).
er-than-creatinine Jaffe-interfering chromogens pres- The reaction specificity is also affected by picrate/
ent in standards in the calibration line, while a alkali concentration. The Jaffe reaction is pseudo-first
two-point calibration combining one standard with order with respect to picrate concentration up to
physiological saline/water always grossly distorts cal- 30 mmol/L and hydroxide concentration more than
ibration line. We calculated/prepared artificial serum 0.5 mol/L (4). Ferrous/ferric ions are known to catalyse
matrices capable of compensating Jaffe-interfering oxidation reactions. The system with ferricyanide was
chromogens in serum standards. The combination of used in Jaffe reaction to catalyse the oxidation of
even one standard with its artificial matrix also ena- interfering bilirubin to biliverdin (3). Similarly, cupric/
bles two-point calibration with practically the same cuprous ions and Cu(II)-tartrate were also used as a
results as with a pair of primary standards. catalyser in Jaffe reaction instead of ferricyanide (5).
Conclusions: A two-point calibration of Jaffe method Several enzymatic methods have been developed
for serum creatinine combining only one serum stan- for measuring creatinine, but they are also not inter-
dard with creatinine solution matching standard’s ference-free (4): bilirubin interference is usually
allow matrix-present interferents present results well solved by adding ferricyanide into a reagent, potential
comparable with enzymatic determination, providing interference caused by ascorbic acid can be overcome
the standard is attested/linked to the reference meas- by inclusion of ascorbate oxidase and the influence of
urement procedure. intermediate creatine and urea is usually solved in the
Clin Chem Lab Med 2008;46:1127–33. first enzyme sequence by a preincubation with crea-
tininase. The reference measurement procedure
(RMP) employing isotope-dilution with mass spec-
*Corresponding author: Prof. Dr. Vratislav Chromý, trometry (ID-MS) (6) is used for creatinine assay in
Institute of Chemistry, Faculty of Science, Masaryk
Standard Reference Material (SRM) calibrators.
University, Komenského 2, Brno, Czech Republic
Phone: q420-549494555, E-mail: vrch@chemi.muni.cz Nevertheless, Jaffe kinetic reaction with so-called
Received February 18, 2008; accepted April 10, 2008 window measurement (selected reaction time meas-
Article in press - uncorrected proof
1128 Chromý et al.: Calibration of serum creatinine determination by Jaffe method

urement within 20–80 s) still remains the leading sera with results well comparable with enzymatic
method for both serum and urine creatinine (4), even methods.
though it is known to overestimate ‘‘true plasma/
serum creatinine by approximately 20% at physiolog-
ical values’’, which is now undoubtedly known to be Materials and methods
caused by nonspecificity of Jaffe reaction and thus
still persisting problems which interferents are always Equipment
present in a serum matrix (4, 7, 8). The main serum
A Shimadzu UV-160 spectrophotometer (Shimadzu, Tokyo,
interferents in Jaffe reaction are proteins (9), which
Japan) with tungsten lamp, 10"0.01 mm light-path quartz
content differ significantly wcf. (10)x in adults and in cells and 6-position cell-rack changer heated to 258C or
children. Unfortunately, reaction time recommended 37"0.18C by a Peltier device (Shimadzu), with wavelength
for Jaffe procedure (4) has its lag phase and reaction 492"0.4 nm was employed. A Shimadzu spectrophotometer
time too short to be used even for small series of cre- was used only for manual Jaffe procedure with one-part re-
atinine samples made on analysers which are usually agent to obtain the data given in Figure 1 and Table 1.
preset to measure in intervals within 60–180 s. A Hitachi 911 Automatic Analyser (Hitachi Ltd., Tokyo,
Some producers of in vitro diagnostics (analytical Japan) with concave diffraction grating was employed. All
kits, standards, analysers, etc.) solved interference analyses were made at 37"0.28C at wavelengths 505"2 nm
and 570 nm (auxiliary) in 6-mm reaction cells with automatic
problems in their Jaffe-based procedures in analyser
recalculation of absorbance to 10 mm. A Hitachi analyser
software by subtracting a constant 26 mmol/L of cre-
was used for all automated Jaffe and enzymatic measure-
atinine as ‘‘an average value (11) of Jaffe-positive ments involving two-part reagents. Data are given in Figure
compounds in a serum matrix from resulting creati- 2 and Tables 2 and 3.
nine concentration estimated by their compensated
Jaffe method’’. According to Junge et al. (12), ‘‘an off-
set of 21 mmol/L of creatinine corresponds to the dif-
ferences between a modified Jaffe and enzymatic
methods’’. Wuyts et al. (9) compared Jaffe compen-
sated/uncompensated procedures for serum creati-
nine with enzymatic methods in more detail and
found a big difference in children caused mainly by
interfering proteins.
While the calibration of urine creatinine with cre-
atinine solutions seems to be without difficulties (no
proteins or traces of proteins), the improvement of
serum/plasma creatinine measurement is still the
focus of analytical and kidney function programmes
(9, 12). According to Myers et al. (13), ‘‘standardiza-
tion of calibration does not correct for analytical inter-
Figure 1 Calibration rate-curves with serum standards and
ferences (nonspecificity bias) and nonspecificity
creatinine solutions.
problems associated with some of the routine meth- One-reagent manual Jaffe procedure (MJ), time measure-
ods must be addressed’’. ment between 60 and 168 s and temperature of 378C. Full
To be explicitly defined, every calibration line must line with black squares: two-point calibration with SRM 967
have at least two calibration points. Two-point cali- LI–LII, line prolongation to x-axis intersects y/x-axes showing
bration of Jaffe reaction with two serum standards overall matrix effect as mAU/creatinine. Dashed line with
with the true creatinine value (standards attested by white circles: two-point calibration with LyoC-ASM; line pro-
longation to x-axis intersects y/x-axes showing ASM as
RMP or linked to SRM) automatically compensates
mAU/creatinine. Full line with black triangles: three-point
Jaffe interfering substances implementing them as an calibration with solutions of creatinine 50, 100 and 300
absorbance offset/nonspecificity bias into the calibra- mmol/L.
tion equation/line. Unfortunately, a lot of producers of
diagnostics supply their customers with only one
Table 1 An overall effect of SRM 967 LIqLII matrices (as
serum-based multi-purpose standard/calibrator and a creatinine, mmol/L).
user is forced to combine such a standard with water/
physiological saline to achieve two-point calibration Reaction time, s Temperature, 8C
which badly distorts Jaffe calibration line eliminating
25 37
its unavoidable absorbance offset.
We tested several approaches in serum creatinine 20–80 31 34
calibration (14) using a modified kinetic Jaffe method 30–180 33 37
60–168 32 38
with Cu(II)-tartrate as a catalyser for bilirubin oxida-
60–150a – 36a
tion (5). We believe we found a simple two-point cal- Average 32 36
ibrating procedure combining one serum standard Range 31–33 34–38
with its artificial serum matrix (ASM) compensating Manual JP (Jaffe reagent and procedure by Pliva-Lachema
standard’s matrix-present Jaffe interferents. Such cal- Diagnostika)-one-reagent procedure. aAutomated JP-two-re-
ibration could be used for both adult and children’s agent procedure.
Article in press - uncorrected proof
Chromý et al.: Calibration of serum creatinine determination by Jaffe method 1129

Materials

Creatinine was from Sigma-Aldrich (Munich, Germany;


C4255, anhydrous G99%, Mrs113.12). For human sera,
patient sera and serum pools were used in this study. A
more detailed description of used sera is given in the legend
to Figure 2 and in the footnotes to Tables 2 and 3. Sera from
a group of 30 children used for comparing enzymatic and
Jaffe method (Table 3) was prepared from samples collected
in a children’s hospital before analysis of serum creatinine
(i.e., this group also contained Jaffe-uncompensated sam-
ples). Sera from a group of 18 children (10 samples, aged
1–5 years and 8 samples, aged 6–10 years) used for analy-
ses, presented in Figure 2, contained only samples analysed
in advance by compensated Jaffe procedure and contained
only samples with creatinine )18 mmol/L, all lower samples
being cut out by an analyser software program set to avoid
negative results.

Figure 2 Creatinine in children’s sera, comparison of Jaffe


and enzymatic methods. Reagents
Two-reagent automated Jaffe procedures, JP (uncompen-
Reagent grade chemicals were used throughout this study.
sated) and JR (enzymatic creatinine reagent and procedure
by Roche Diagnostics compensated) vs. EP. Time measure- Physiological saline (PS) was prepared by dissolving 9 g of
ment of 60–168 and 60–150 s, respectively, temperature of NaCl in distilled water in a 1000-mL volumetric flask. Creat-
378C. The same group of sera (ns18) from a children’s hos- inine stock solution of 1000 mmol/L was prepared by dis-
pital analyser, selected by their compensated JR procedure solving 113.12 mg of creatinine in a 1000-mL volumetric flask
(10 samples, aged 1–5 years plus 8 samples, aged in PS and was further diluted with PS to obtain desired
6–10 years). Enzymatic/uncompensated JPs were calibrated concentration.
with a pair of LyoC-PS and LyoC-ASM, respectively. Circles:
JP vs. EP. Squares: compensated JR vs. EP. Creatinine Calibrators We used the National Institute of Standards
average/mean values found (mmol/L): EP 52.9/31–85; and Technology SRM 967 as primary creatinine standards
EP 51.4/27–83; JR 34.5/18–63. throughout this study (15): frozen human serum level I (LI)
was certified for creatinine 66.5"1.9 and level II (LII) for
Table 2 Preliminary experiments. The effect of calibration 346.2"7.3 mmol/L.
on serum creatinine assay by Jaffe method (mmol/L).
Lyonorm calibrator (LyoC) and its artificial serum matrix
Calibration mode/calibrator Creatinine found (ASM) We recalibrated/reset the concentration of creatinine
concentration, mmol/L in Pliva-Lachema universal human multi-purpose serum-
Low pool High pool
based standard/calibrator (Pliva-Lachema Diagnostika, Brno,
Two-point/PS-LI (66.5) 74 208 Czech Republic) against SRM using automated Jaffe proce-
Two-point/PS-LII (346.2) 123 330 dure (JP) to 331"12 mmol/L. We determined that the total
Two-point/LI-LII (66.5 and 346.2) 86* 323* amount of Jaffe interferents present in LyoC serum matrix
Three-point/creatinine solutionsa 125 362 corresponds to 38"2 mmol/L of creatinine and we prepared
Two-point/ASM-LyoC (325.0)b 85* 328* LyoC-matching ASM as a solution containing 38 mmol/L of
Automated JP-two-reagent procedure. Both pools contained creatinine. A pair of LyoC-ASM and/or LyoC-PS was used for
an equal mixture of 20 male/female adult sera. *Probably the two-point calibration for both Jaffe and enzymatic methods,
true serum creatinine free of interferents. aCreatinine solu- respectively. As an example, we also prepared and used
tions of 50, 100 and 300 mmol/L. bASM contained 38.0 ASM based on glucose and bovine serum albumin solutions,
mmol/L of creatinine. respectively (see Results).

Table 3 Comparison of Jaffe and enzymatic methods (as creatinine, mmol/L).

Procedure Adults Children

1: JP-ER 2: JP-EP 3: JR-EP 4: JP-EP 5: JP-EP


Average, Y 77 219 209 40 35
Range 45–150 103–659 100–629 8–47 6–50
Average, X 81 217 205 35 38
Range 47–144 97–649 104–659 7–44 8–49
Automated two-point calibration using two-reagent procedures combining Jaffe and enzymatic methods. For adults, proce-
dure 1: sera from a group of 100 male/female (aged 20–70 years) with normal/elevated creatinine levels. Both calibrations
with SRM LI–LII (rs0.993; ys0.996x–3.6). Procedure 2: sera from a group of 60 male/female (aged 17–64 years) with signif-
icantly elevated creatinine levels. EP calibration with LyoC-PS, JP with LyoC-ASM (rs0.991, ys1.007x–4.7). Procedure 3: the
same sera as in 2. EP calibration with SRM LI–LII, JR compensated data by subtracting 26 mmol/L of creatinine were taken
from the hospital laboratory (rs0.992; ys0.979xq0.5). For children, sera from a group of 30 children (20 samples, aged 1–5
years and 10 samples, aged 6–10 years). Procedure 4: both calibrations with SRM LI–LII (rs0.960, ys0.989x–4.5). Procedure
5: EP calibration with LyoC-PS (rs0.967; ys0.988xq2.4), JP calibration with LyoC-ASM.
Article in press - uncorrected proof
1130 Chromý et al.: Calibration of serum creatinine determination by Jaffe method

Jaffe reagent, Pliva (JP) A two-part reagent analytical kit 1 part of R2, see Jaffe reagent JP. In a cell heated to the
was used, with Creatinine Liquid 500, Cat. No. 10010227 selected temperature, mix 1.00 mL of MJ reagent with
(Pliva-Lachema Diagnostika) (Reagents; R1: NaOH 0.24 0.050 mL of a sample and register immediately the rate curve
mol/L; R2: picric acid 26 mmol/L). JP was used both in man- against picrate reagent with PS at 492 nm. Calculate the dif-
ual and automated procedures. In this kit, Cu(II)-tartrate is ference DAs(A2 –A1) for reaction time between 20–80 and
used to catalyse bilirubin oxidation (5). Performance data 60–168 s, respectively.
(serum): limit of detection 2.33 mmol/L, low limit of quanti-
fication 7.09 mmol/L, working range 7.09–3000 mmol/L, bias Automated Jaffe procedures Automated Jaffe procedures
–2.54% and –1.01% (for 80 and 334 mmol/L of creatinine, were used only in two-reagent mode with both Jaffe/enzy-
respectively), precision (intra-/inter-assay): 1.8%/2.16%, cor- matic methods. Analyses, which were carried out in dupli-
relation with enzymatic creatinine Pliva (ns96, rs0.999, cate, were used for all data presented in Tables 2 and 3 and
ys1.008x–6.4 mmol/L). in Figure 2. The analyser was programmed accordingly for
two-reagent procedure JP and JR, respectively, with a wave-
Jaffe reagent, Roche (JR) A two-part reagent analytical kit length of 505 and 570 nm, respectively.
was used, with CREA, Creatinine Jaffe method, rate-blanked
and compensated, Cat. No. 11875418 (Roche Diagnostics, Automated Jaffe procedure, Pliva (JP) Mix 0.015 mL of a
Mannheim, Germany) (Reagents; R1: NaOH 0.8 mol/L; R2: sample with 0.240 mL of R1 and incubate for 5 min, add
picric acid 25 mmol/L; see automated Jaffe procedure 0.060 mL of R2 and after 60 s read the absorbance A1 and
Roche). Performance data (serum): working range after 168 s read the absorbance A2, reaction time measure-
18–2210 mmol/L, low limit of quantification 18 mmol/L, ment between 60 and 168 s.
precision (inter-/intra-assay): 0.7%/2.3%, correlation with
enzymatic creatinine Roche (ns50, rs0.998, Automated Jaffe procedure, Roche (JR) Mix 0.006 mL of
ys1.014x–0.07 mmol/L). sample with 0.250 mL of R1 and incubate for 5 min, add
0.125 mL of R2 and after 60 s read the absorbance A1 and
Enzymatic creatinine, Pliva (EP) A two-part reagent analyt- after 150 s read the absorbance A2, reaction time measure-
ical kit was used, with Creatinine Enzymatic Liquid 204, Cat. ment between 60 and 150 s. Rate-blanked and compensated
No. 10010231 (Pliva-Lachema Diagnostika) (Reagents; R1: JR and Roche calibrator was used only in a hospital labo-
Good’s buffer pH 7.5, 25 mmol/L, creatinase G12 kU/L, sar- ratory which supplied us with patient sera together with their
cosine oxidase G8 kU/L, ascorbate oxidase G2 kU/L, catalase creatinine values.
G200 kU/L, indicator N-ethyl-N-sulphopropyl-m-toluidine
(ESPMT) 0.47 mmol/L, detergent -1%, gentamycin -0.1%; Automated enzymatic procedure, Pliva (EP) Mix 0.018 mL
R2: Good’s buffer pH 7.5, 100 mmol/L, creatininase G300 of sample with 0.81 mL of R1, incubate for 5 min and read
kU/L, peroxidase G20 kU/L, 4-aminoantipyrine 2.95 mmol/L, the first absorbance (A1), add 0.270 mL of R2, incubate for
detergent -0.5%, sodium azide G0.1%). Performance data 10 min and read the second absorbance (A2) and calculate
(serum): limit of detection 1.50 mmol/L, low limit of quanti- the difference. Wavelength: 546 nm.
fication 4.60 mmol/L, working range 4.6–5700 mmol/L, bias
–3.60% and –2.20% (for 90 and 360 mmol/L of creatinine, Automated enzymatic procedure, Roche (ER) Mix 0.006 mL
respectively), precision (intra-/inter-assay): 1.75%/3.14%, cor- of a sample with 0.65 mL of R1, incubate for 5 min and read
relation with other enzymatic creatinine (ns98, rs0.999, the first absorbance (A1), add 0.20 mL of R2, incubate for
ys1.051x–14.9 mmol/L). 5 min and read the second absorbance (A2) and calculate the
difference. Wavelength: 546 nm.
Enzymatic creatinine, Roche (ER) A two-part reagent analyt-
ical kit was used, with Creatinine plus, Cat. No. 11775642 (Roche
Diagnostics) (Reagents; R1: N-tris(hydroxymethyl)methyl-3-
Results
aminopropanesulphonic acid (TAPS) buffer pH 8.1,
30 mmol/L, creatinase G333 mcat/L, sarcosine oxidase
G133 mcat/L, ascorbate oxidase G33 mcat/L, indicator 3-
Preliminary experiments
hydroxy-2,4,6-triiodobenzoic acid (HTIB) 5.9 mmol/L, deter-
Calibration lines and equations in Figure 1 were
gent and stabiliser; R2: TAPS buffer pH 8.0, 50 mmol/L,
obtained by JP-one-reagent manual procedure and
creatininase G500 mcat/L, peroxidase G16.7 mcat/L, 4-ami-
noantipyrine 2.0 mmol/L, potassium hexacyanoferrate were used to calculate the amount of serum/matrix
163 mmol/L, detergent and stabiliser). Performance data interfering substances (Table 1) present in used stan-
(serum): working range 2.7–2652 mmol/L, low limit of quan- dards. Figure 1 shows the overall effect of interfering
tification 2.7 mmol/L, precision (inter-/intra-assay): 0.8%/ substances in a pair of SRM both as absorbance/
2.1%, correlation with HPLC method (ns64, rs0.999, creatinine nonspecificity bias. Compared with a curve
ys0.989xq0.036 mg/dL). constructed with creatinine solutions, SRM curve is
shifted to a higher absorbance which is caused by
Procedures other-than-creatinine positive chromogens present in
SRM matrices. All calibration lines combining a
All measurements carried out manually were performed in
serum standard and/or standards with water or PS as
triplicate with one-part Jaffe reagent (MJ) at 258C and 378C,
respectively and were used only for the calculation/deter-
zero will distort calibration curves, a distortion being
mination of Jaffe-interfering substances in standards and in dependent on the standard and the calibration mode
two types of control sera (see Figure 1 and Table 2). chosen. Figure 1 also shows that a pair of LyoC-ASM
enables determination of serum creatinine with accu-
Manual Jaffe procedure, Pliva (MJ) The manual Jaffe pro- racy well comparable with SRM, bringing the calibra-
cedure, Pliva (MJ), was used only in one-reagent mode. MJ- tion line tightly to that constructed with a pair of
reagent was prepared from JP by mixing 4 parts of R1 with primary standards.
Article in press - uncorrected proof
Chromý et al.: Calibration of serum creatinine determination by Jaffe method 1131

In our preliminary experiments (Table 2), we tested Creatinine determined by Jaffe and enzymatic
various standards/calibration modes on two serum methods
pools with low/high level of creatinine using JP-two-
reagent automated procedure. Jaffe method was cal- The two-point calibration of Jaffe method with a pair
ibrated either with a pair of SRM (15) or with a of SRM and/or with LyoC-ASM we used in preliminary
secondary standard LyoC paired with its ASM or with experiments compensates for serum matrix effects
solutions of pure creatinine. Calibration with pure cre- (Table 2). We also verified it in more detail by two
atinine and also with a serum standard combined enzymatic procedures, EP/ER, on two groups of adult
with PS/water used to establish zero will always dis- sera, the first one with normal/elevated creatinine, the
tort calibration line and overestimate true serum cre- second one with greatly elevated creatinine level
atinine, the overestimation being major in serum (Table 3, procedures 1 and 2, see footnotes). It is evi-
samples with low creatinine. dent that the enzymatic procedures can be two-point
calibrated both with a pair of SRM and with one sec-
ondary standard (such as LyoC) complemented with
Jaffe-like interferents in standards
PS. A Jaffe method can be calibrated either with a pair
of serum standards or with one standard comple-
A calibration line in Figure 1, constructed with a pair
mented with its ASM. Averages/ranges of creatinine
of SRM (LI and LII) enables estimation of an overall
concentration and also regression lines/relationships
effect of matrices-present interferents as its intersec-
found did not show notable differences. We used the
tion with y/x-axes or calculation of either the overall
same group of samples as in procedure 2 also for
(Table 1) or even individual matrix effects combining
comparing enzymatic method with compensated JR
appropriate equations for SRM with those for creati-
procedure using compensation by subtracting the
nine solutions. We calculated the following individual
constant amount of 26 mmol/L of creatinine from all
matrix effects (as creatinine, mmol/L): SRM-LI ;13,
samples (Table 3, procedure 3). We were surprised by
SRM-LII ;23, LyoC ;38, and two control sera, low
the very good results we found with compensated
;9 and high ;40. As expected, the higher matrix
Jaffe method (11) in adults.
effect was always found in multi-purpose standards/
Special attention was given to the analysis of sera
controls spiked generally to higher levels. An overall
from children (Table 3, procedures 4 and 5). Results
SRM matrix could also be found by an analyser using
achieved by uncompensated Jaffe and enzymatic cre-
two-reagent procedures: a Hitachi analyser was
atinine correlate closely and regression data are quite
two-point calibrated with a pair of SRM, then several
acceptable. It is evident a two-point calibration, either
creatinine solutions with known concentration
with a pair of SRM and/or with a pair of LyoC-ASM,
(10–60 mmol/L) were analysed as ‘‘unknown sam-
respectively, used for uncompensated Jaffe assay in
ples’’. The zero intersection between two straight adults that functions well even in children. In our sec-
lines which the analyser printed as found/known cre- ond experiment with samples from children, we tried
atinine in ‘‘unknown samples’’ corresponded to the to compare compensated (JR) and uncompensated
interferents present in SRM matrices. A matrix cor- (JP) procedures with enzymatic creatinine (Figure 2).
responding to an absorbance matching ;36 mmol/L For such comparison, we used sera in which creati-
of creatinine could also be simulated by solutions of nine was analysed in advance by compensated Jaffe
glucose (;280 mmol/L) or by bovine serum albumin procedure in a hospital laboratory, which means this
(;109 g/L). However, ASM with creatinine is the best group contained only sera with creatinine )18
because of the same reaction rate. Proteins (9) are the mmol/L. A correlation graph comparing enzymatic
major interferents in Jaffe method. It is easy to cal- and compensated Jaffe assays indicates overcompen-
culate that an average offset of 21 mmol/L of Jaffe- sation in creatinine of approximately ;16 mmol/L in
serum-creatinine found by Junge et al. (12) children. The correlation between EP/JP procedures is
corresponds to the interference caused by ;64 g/L of quite good and a two-point calibration used allows
protein (as bovine albumin). much better results in children, compared with com-
pensated Jaffe method.
Reaction time measurement

We were not able to confirm any positive effect of a Discussion


reaction time on the compensation of serum interfe-
rents in Jaffe kinetic methods (Table 1). All matrix- SRMs used in this study were based on two frozen
present interferents, even when measured at different SRMs which are intended (15) ‘‘primarily for use in
times, were situated in a very narrow range. The evaluating the accuracy of procedures for the deter-
greater matrix effect found at 378C is caused by tem- mination of creatinine in human serum and for use in
perature-increased reaction velocity of some interfe- validating secondary reference materials’’. We used
rents. Moreover, DA measured within the generally SRMs as primary standards for all procedures com-
recommended 20–80 s (4) is low. In this respect, it is pared, for creatinine validation in the secondary stan-
more convenient to read the absorbance within dard LyoC and also for the calculation/determination
;60–180 s, where DA is approximately 50% higher of the Jaffe-matrix-effect in some other serum mate-
and which timing is also convenient for most of the rials. Creatinine in SRM, being certified by RMP (6), is
automatic analysers. supposed to be free of interferences caused by Jaffe-
Article in press - uncorrected proof
1132 Chromý et al.: Calibration of serum creatinine determination by Jaffe method

like serum constituents, while the absorbance found diseases, it would be more suitable to calibrate
by Jaffe procedure with SRM and/or with serum serum/plasma creatinine with standards in which cre-
samples should also include the effect of other-than- atinine levels range between 50 and 150 mmol/L. It
creatinine Jaffe-like chromogens as an absorbance would also be more convenient for the standardised
offset. calibration of sera for both children and adults.
Our study proved that two-point calibration of a
Jaffe method using a pair of serum standards with
true creatinine value and/or a secondary standard
Acknowledgements
paired with its artificial matrix yield results well com-
parable with enzymatic methods. Such two-point cal-
ibration respects natural and unavoidable absorbance We would like to express our gratitude to Drs. Ludmila
Malášková and Miroslava Beňovská, Department of Clinical
offset representing nonspecificity Jaffe bias and
Biochemistry, Faculty Hospital, Brno-Bohunice, Czech
incorporates it into the calibration equation/line. Any Republic, for patient sera analysed by their Jaffe-based, rate-
calibration combining a serum standard with PS/ blanked and compensated Roche analytical method. We
water closes calibration line incorrectly to zero result- would also like to thank Dr. Jan Havliš, Department of Func-
ing in falsely increased serum creatinine. tional Genomics and Proteomics, Institute of Experimental
In agreement with Wuyts et al. (9), we confirmed Biology, Faculty of Science, Masaryk University, Brno, for
that the greatest errors/nonspecificity bias in Jaffe advice and help during manuscript preparation. V.C. would
methods is caused by proteins. Reference interval (10) like to express thanks to PLIVA-Lachema Diagnostika, Barr
for total serum protein varies greatly in adults and its Group, Brno, Czech Republic, for financial support of this
work.
variation in newborns is even greater ("15% and
"35% from the mean value, respectively). Moreover,
children’s sera and particularly sera of toddlers can
contain only a fraction of creatinine compared with References
adults wcf. (10)x. In this respect, to find a universal
compensation constant for compensated Jaffe meth- 1. Vasiliades J. Reaction of alkaline sodium picrate with
od valid for both adults/children is impossible. While creatinine: kinetics and mechanism of formation of the
non creatinine picric acid complexes. Clin Chem 1976;22:
compensated Jaffe method (11) applied to adults
1664–71.
yielded serum creatinine well comparable with enzy- 2. Spencer K. Analytical reviews in clinical biochemistry:
matic procedure, its application to children poorly the estimation of creatinine. Ann Clin Biochem 1986;23:
overcompensates creatinine in children and a cut-out 1–25.
filter implemented to compensated Jaffe procedures 3. Knapp ML, Mayne PD. Development of an automated
even disallows the determination of a real low creat- kinetic Jaffe method designed to minimize bilirubin
interference in plasma creatinine assays. Clin Chim Acta
inine level in children within 1–5 years of age. In fact,
1987;168:239–46.
the overcompensation we found in creatinine levels 4. Lamb E, Newman DJ, Price CP. Kidney function tests. In:
in children must be even greater, because our tested Burtis CA, Ashwood ER, Bruns DE, editors. Tietz text-
group contained only selected sera from children with book of clinical chemistry, 4th ed. St. Louis, Missouri:
creatinine levels )18 mmol/L. Without such an ana- WB Saunders Company, 2006:797–801.
lyser made correction, a lot of creatinine levels in chil- 5. Fischer J, Chromý V. Reagent for photometric kinetic
dren found by compensated Jaffe procedure should determination of creatinine suppressing the interference
of bilirubin. Czechoslovak patent, Authors’ Certificate
have been negative. Both types of two-point calibra- 1982:228836.
tion we used in this study for uncompensated Jaffe 6. Welch MJ, Cohen A, Hertz HS, Ng KJ, Schafer R, Van
method function well for both adults and children. Der Lijn P, et al. Determination of serum creatinine by
Unfortunately, the scatter of serum creatinine levels isotope dilution mass spectrometry as a candidate defin-
in children will always be large because of a generally itive method. Anal Chem 1986;58:1681–5.
very low DA measured with children’s samples. 7. Carobene A, Ferrero C, Ceriotti F, Modenese A, Besozzi
M, De Giorgi E, et al. Creatinine measurement proficien-
The daily calibration with a pair of SRMs is expen-
cy testing: assignment of matrix-adjusted ID GC-MS
sive and is not common. We used SRMs in our study target values. Clin Chem 1997;43:1342–7.
only to compare various creatinine procedures, to 8. Lawson N, Lang T, Broughton A, Prinsloo P, Turner C,
determine/calculate an absorbance offset/nonspecifi- Marenah C. Creatinine assays: time for action? Ann Clin
city bias caused by Jaffe-interfering substances in Biochem 2002;39:599–602.
serum standards and to prepare matched artificial 9. Wuyts B, Bernard D, Van Den Noortgate N, Van Den
Walle J, Van Vlem B, De Smet R, et al. Reevaluation of
serum matrices.
formulas for predicting creatinine clearance in adults
Thus far, many companies producing diagnostics and children, using compensated creatinine methods.
offer only one type of multi-purpose serum-based Clin Chem 2003;49:1011–4.
standard/calibrator insufficient for two-point calibra- 10. Roberts WL, McMillin QA, Burtis CA, Bruns DE. Refer-
tion of Jaffe method, and we hope our study would ence information for the clinical laboratory. In: Burtis CA,
help to solve calibration problems still connected with Ashwood ER, Bruns DE, editors. Tietz textbook of clinical
chemistry, 4th ed. St. Louis, Missouri: WB Saunders
Jaffe procedure. Unfortunately, the concentration of
Company, 2006:2264, 2292–3.
creatinine in all multi-purpose calibrators is usually 11. CREA. Creatinine Jaffe method, rate-blanked and com-
set too high. For reliable serum creatinine glomerular pensated. Prague, Czech Republic: Roche Diagnostics,
filtration rate and for the treatment of chronic kidney Cat. No. 11875418.
Article in press - uncorrected proof
Chromý et al.: Calibration of serum creatinine determination by Jaffe method 1133

12. Junge W, Wilke B, Halabi A, Klein G. Determination of 14. Chromý V, Rozkošná K, Sedlák P. Serum Jaffe creatinine,
reference intervals for serum creatinine, creatinine how to calibrate and eliminate serum matrix problems
excretion and creatinine clearance with an enzymatic wabstractx. VIII National Congress of Czech Society of
and a modified Jaffe method. Clin Chim Acta 2004;344: Clinical Biochemistry (with international participation),
137–48. České Budějovice, 23–25 September, 2007. Klin Biochem
13. Myers GL, Miller WG, Coresh J, Fleming J, Greenberg Metab 2007;15(Suppl):S7-4, P-6.
N, Greene T, et al. Recommendations for improving 15. National Institute of Standards and Technology. Stan-
serum creatinine measurement: a report from the labo- dard reference material 967. Gaithersburg, MD: National
ratory working group of the National kidney disease edu- Institute of Standards and Technology. Certificate issue
cation program. Clin Chem 2006;52:5–18. date: 24 January, 2007.

View publication stats

You might also like