Personalization of the generator V5.

3
CHAPTER 2
TECHNICAL MANUAL

Ref : 32-40-013-C
2
Generator GTI HF

Contents
Programming the configuration parameters ....................... 3
1 Entry into the configuration mode ...................................................... 3
2 Contents of eeprom 1: configuration of the system .............................. 3
3 Parameter description ..................................................................... 12
4 End of the configuration mode ......................................................... 56

Configuration list .............................................................. 56

Tube characteristics definition ......................................... 61


1 Passage into the tube characteristics definition modeE...................... 61
2 Contents of eeprom 2 and 3 characteristic definition of tubes 1 & 2 .... 62
3 Record of the tube load characteristics ............................................ 64
4 Measurement of the tube emission characteristics............................. 66
5 Compensation of the heating current ................................................ 67
6 Tube caracteristic list ...................................................................... 68

Ref : 32-40-013-C
3
Chapter 2 / Technical manuel

Programming the configuration parameters


1 Entry into the configuration mode
This programm has been made to allows the adding of a display for heat integrator (op-
tionnel).
It allows to work in automatic exposure mode with falling load.
On start you will have in the display unit used for kV the state of the heat integrator:

11 Less than 3 hours has been passed.The integrator calculate the time.
44 Error of checksum. The integrator is set at 0.
55 More than one day without power supply. The integrator is set at 0.
66 More than 3 hours without power supply. The integrator is set at 0.

Turn the commutator S2 on the control card so as to illuminate the LED «CONFIGURA-
TION.
At the end of modifying the configuration parameters, do not forget to reset the commutator
on «NORMAL» position, so as to be able to use the generator.
The generator starts in the mode which allows you to define the installation.
The display unit normally used to adjust the graphy kV becomes the EEPROM configura-
tion parameter number display. The display unit normally used to adjust the graphy current
becomes the parameter value display. The adjustment keys corresponding to these display
units allow a selection of the parameter to be modified and the adjustment of the value of
the selected parameter. The validation / invalidation key of the automatic graphy mode
permits the modified parameter to be stored in the memory. The corresponding blinking
LED reminds the user that he must store on memory the value that has been modified.
Forgetting to press this key will not affect the contents of the configuration EEPROM.
The configuration EEPROM contains 256 bytes of which only 196 are accessable to the
user.
Note: It is the operator’s responsibility to correctly program the configuration memory. A
faulty functioning could be the result of a faulty programming of the configuration memory.

Ref : 32-40-013-C
4
Generator GTI HF

2 Contents of eeprom 1: configuration of the system


The following list is a condensed form of the contents of EEPROM which contain the
configuration parameters of the system. The pages which follow the list describe in detail
each parameter.

N° min. max. name

0 1 80 max. power (kW)


1 0 2 tube linked with post 1(0=post inhibited)
2 0 2 tube linked with post 2(0=post inhibited)
3 0 2 tube linked with post 3(0=post inhibited)
4 0 2 tube linked with post 4(0=post inhibited)
5 0 2 tube linked with post 5(0=post inhibited)
6 0 2 tube linked with post 6(0=post inhibited)

7 5 13 mA graphy maximum in tomography


5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12: 800mA 13: 1000mA

8 0 1 mA internal meter display if 1


10 0 1 pulsed fluoroscopy mode when starting if 1
11 1 6 post number when starting
12 0 255 Stand-by time in mn(auto-dodo)
14 0 255 SPECIAL:Confuguration by default of the eeprom

18 0 3 auto. exposure n° used for post 1;(0=no auto.exposure)


19 0 3 auto. exposure n° used for post 2;(0=no auto. exposure)
20 0 3 auto. exposure n° used for post 3;(0= no auto.exposure)
21 0 3 auto. exposure n° used for post 4;(0=no auto. exposure)
22 0 3 auto. exposure n° used for post 5;(0=no auto. exposure)
23 0 3 auto. exposure n° used for post 6;(0=no auto. exposure)

24 0 23 auto. exposure mode by default from post 1


25 0 23 auto. exposure mode by default from post 2
26 0 23 auto. exposure mode by default from post 3
27 0 23 auto. exposure mode by default from post 4
28 0 23 auto. exposure mode by default from post 5

Ref : 32-40-013-C
5
Chapter 2 / Technical manuel

N° min. max. name


29 0 23 auto. exposure mode by default from post 6
mode LMR HML
0 — L
1 —R L
2 -M- L
3 -MR L
4 L— L
5 L-R L
6 LM- L
7 LMR L
8 — M
9 —R M
10 -M- M
11 -MR M
12 L— M
13 L-R M
14 LM- M
15 LMR M
17 — H
18 —R H
19 -M- H
20 -MR H
21 L— H
22 L-R H
23 LM- H
24 LMR H
30 0 1 fluoroscopy authorization on post 1 if at 0
31 0 1 fluoroscopy authorization on post 2 if at 0
32 0 1 fluoroscopy authorization on post 3 if at 0
33 0 1 fluoroscopy authorization on post 4 if at 0
34 0 1 fluoroscopy authorization on post 5 if at 0
35 0 1 fluoroscopy authorization on post 6 if at 0
36 0 1 0=normal
1=tube 1 equiped with a high speed starter
and taken into account the authorization entry of the anode
37 0 1 0=normal
1=tube 2 equiped with a high speed starter
and taken into account the authorization entry of the anode
38 0 1 rotation of the tube 1 anode in fluoroscopy if 1
39 0 1 rotation of the tube 2 anode in fluoroscopy if 1
40 1 255 small focus standby heating current
1=2000+14*parameter value
41 1 255 large focus standby heating current
1=2000+14*parameter value
42 120 150 kV maximum for tube 1
43 120 150 kV maximum for tube 2

Ref : 32-40-013-C
6
Generator GTI HF

N° min. max. name


44 0 1 1=tomography mode for tube 1
45 0 1 1=tomography mode for tube 2
46 0 1 0=normal function
1=second time external authorization for post 1
47 0 1 0=normal function
1=second time external authorization for post 2
48 0 1 0=normal function
1=second time external authorization for post 3
49 0 1 0=normal function
1=second time external authorization for post 4
50 0 1 0=normal function
1=second time external authorization for post 5
51 0 1 0=normal function
1=second time external authorization for post 6
52 0 255 tempo before passage into live memory after beginning
(1=10ms)
53 0 255 tempo of fluoroscopy prolongation for memorization
(1=10ms)
54 0 255 start of pulsed fluoroscopy=stablization time of the
automatic kV (1=100ms)
55 0 255 pulsed fluoroscopy standby time (1=100ms)
56 0 255 live memory time (1=10ms)
57 0 13 triggering of large focus tube 1
0=10mA; 1=20mA; 3=40mA;4=50mA;5=100mA;
6=150mA;7=200mA; 8=300mA;9=400mA;
10=500mA;11=600mA;12=800mA;13=1000mA;
14=pas de gros foyer.
58 0 13 triggering of large focus tube 2
0=10mA; 1=20mA; 3=40mA ;4=50mA;5=100mA;
6=150mA;7=200mA; 8=300mA;9=400mA;
10=500mA;11=600mA;12=800mA;13=1000mA
14=pas de gros foyer.
59 0 255 inhibition for some keys for post 1
60 0 255 inhibition for some keys for post 2
61 0 255 inhibition for some keys for post 3
62 0 255 inhibition for some keys for post 4
63 0 255 inhibition for some keys for post 5
64 0 255 inhibition for some keys for post 6
bit 7: inhibition for automatic fluoroscopy
bit 6: inhibition for pulsed fluoroscopy
bit 5: inhibition for slow screen
bit 4 :inhibition for normal screen
bi t 3 :inhibition for fast screen
bit 2 ;inhibition for left field
bit 1 :inhibition for central field
bit 0 :inhibition for right field

Ref : 32-40-013-C
7
Chapter 2 / Technical manuel

N° min. max. name


65 0 255 compensation for exposure time in ms
The contents of this memory is substracted from the exposure time
asked by the user
66 0 1 Heat integrator stop the exposition
67 0 100 Percentage of the tube power used with post 1
68 0 100 Percentage of the tube power used with post 2
69 0 100 Percentage of the tube power used with post 3
70 0 100 Percentage of the tube power used with post 4
71 0 100 Percentage of the tube power used with post 5‘
72 0 100 Percentage of the tube power used with post 6
73 0 63 Post condition relay 1
74 0 63 Post condition relay 2
75 0 63 Post condition relay 3
76 0 63 Post condition relay 4
77 0 63 Post condition relay 5
78 0 63 Post condition relay 6
79 0 63 Post condition relay 7
80 0 63 Post condition relay 8
81 0 63 Post condition relay 9
82 0 63 Post condition relay 10
7 6 5 4 3 2 1 0
- PL6 &Profil PL6 PL5 PL4 PL3 PL2 PL1
83 0 64 State condition relay 1
84 0 64 State condition relay 2
85 0 64 State condition relay 3
86 0 64 State condition relay 4
87 0 64 State condition relay 5
88 0 64 State condition relay 6
89 0 64 State condition relay 7
90 0 64 State condition relay 8
91 0 64 State condition relay 9
92 0 64 State condition relay 10
7 6 5 4 3 2 1 0
mask inh 1st time bucky graphy scopy ready
93 0 2 FPS by default of cinema post
0=12 images per second
1=25 images per second
2=50 images per second
94 0 4 Serial time by default of cinema post
0=2 seconds
1=5 seconds
2=10 seconds
3=15 seconds
4=20 seconds
95 3 20 Exposure time in cinema in ms
96 0 1 Operating mode post 5
0=post 5 is normal
1=post 5 is a cinema post

Ref : 32-40-013-C
8
Generator GTI HF

N° min. max. name

99 0 255 Cooling factor for anode tube 1


100 0 255 Cooling factor for anode tube 2
101 0 255 Threshold for triggering heating alarm anode tube 1(lsb)joules
102 0 255 Threshold for triggering heating alarm anode tube 1()joules
103 0 255 Threshold for triggering heating alarm anode tube 1(msb)joules
104 0 255 Threshold for triggering heating alarm anode tube 2(lsb)joules
105 0 255 Threshold for triggering heating alarm anode tube 2()joules
106 0 255 Threshold for triggering heating alarm anode tube 2(msb)joules
107 0 255 Cooling factor for housing tube 1
108 0 255 Cooling factor for housing tube 2
109 0 255 Threshold for triggering heating alarm housing tube 1(lsb)joules
110 0 255 Threshold for triggering heating alarm housing tube 1()joules
111 0 255 Threshold for triggering heating alarm housing tube 1(msb)joules
112 0 255 Threshold for triggering heating alarm housing tube 2(lsb)joules
113 0 255 Threshold for triggering heating alarm housing tube 2()joules
114 0 255 Threshold for triggering heating alarm housing tube 2(msb)joules
115 0 255 Servocontrol factor for photomultiplier
116 0 255 Power threshold for interface relays
117 0 8 Blackening for post 1
118 0 8 Blackening for post 2
119 0 8 Blackening for post 3
120 0 8 Blackening for post 4
121 0 8 Blackening for post 5

Ref : 32-40-013-C
9
Chapter 2 / Technical manuel

N° min. max. name


122 0 8 Blackening for post 6
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4
123 0 1 1=anatomical keyboard
0=no anatomical keyboard
124 0 10 Offset of recopy scopy>graphy
125 40 150 kV graphy default value post 1
126 40 150 kV graphy default value post 2
127 40 150 kV graphy default value post 3
128 40 150 kV graphy default value post 4
129 40 150 kV graphy default value post 5
130 40 150 kV graphy default value post 6
131 0 11 mA graphy default value post 1
132 0 11 mA graphy default value post 2
133 0 11 mA graphy default value post 3
134 0 11 mA graphy default value post 4
135 0 11 mA graphy default value post 5
136 0 11 mA graphy default value post 6
0: 10mA
1: 20mA
2: 30mA
3: 40mA
4: 50mA
5: 100mA
6: 150mA
7: 200mA
8: 300mA
9: 400mA
10: 500mA
11: 600mA
12:800mA
13:1000mA
137 0 1 automatic graphy mode on post 1 if 1
138 0 1 automatic graphy mode on post 2 if 1
139 0 1 automatic graphy mode on post 3 if 1
140 0 1 automatic graphy mode on post 4 if 1
141 0 1 automatic graphy mode on post 5 if 1
142 0 1 automatic graphy mode on post 6 if 1

Ref : 32-40-013-C
10
Generator GTI HF

N° min. max. name


143 0 1 automatic fluoroscopy mode on post 1 if 1
144 0 1 automatic fluoroscopy mode on post 2 if 1
145 0 1 automatic fluoroscopy mode on post 3 if 1
146 0 1 automatic fluoroscopy mode on post 4 if 1
147 0 1 automatic fluoroscopy mode on post 5 if 1
148 0 1 automatic fluoroscopy mode on post 6 if 1
149 0 255 LSB graphy number
150 0 255 SB graphy number
151 0 255 MSB graphy number
total number of exposures =(149)+(150)*256 +(151)*65536
Presing the memo key reset at 0 the total number of exposures.
152 0 255 starting time tube 1(1/10ème de seconde)
153 0 255 starting time tube 2(1/10ème de seconde)
154 0 255 inhibition time fluoroscopy after a radiography tube 1
(1/10ème de seconde)
155 0 255 inhibition time fluoroscopy after a radiography tube 2
(1/10ème de seconde)
156 0 1 reading of kV graphy reference on An5 poste 1
157 0 1 reading of kV graphy reference on An5 poste 2
158 0 1 reading of kV graphy reference on An5 poste 3
159 0 1 reading of kV graphy reference on An5 poste 4
160 0 1 reading of kV graphy reference on An5 poste 5
161 0 1 reading of kV graphy reference on An5 poste 6
162 2 36 graphy time default value post 1
163 2 36 graphy time default value post 2
164 2 36 graphy time default value post 3
165 2 36 graphy time default value post 4
166 2 36 graphy time default value post 5
167 2 36 graphy time default value post 6
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000
168 40 150 kV graphy by default pos 1 in Inferior Members
169 40 150 kV graphy by default pos 2 in Inferior Members
170 40 150 kV graphy by default pos 3 in Inferior Members
171 40 150 kV graphy by default pos 4 in Inferior Members
172 40 150 kV graphy by default pos 5 in Inferior Members
173 0 1 Operating mode post 6
0=post 6 normal
1=post 6 Inferior Members

Ref : 32-40-013-C
11
Chapter 2 / Technical manuel

N° min. max. name


174 0 2 0=Pused fluoroscopy with video memory
1=Pulsed fluoroscopy in radiographic mode on tube 1
2=Pulsed fluoroscopy in radiographic mode on tube 2
175 0 7 Heating current at 50kV
176 0 7 Heating current at 150kV
0=10 mA
1=20 mA
2=30 mA
3=40 mA
4=50 mA
5=100 mA
6=150 mA
7=200 mA
WARNING:
1.Be sure that HV current doesn’t reach the tube’s caracteristics
because there is no control by the microprocessor.
2. Choose a value of current higher at 120kV than the one at 50kV.
3. Do not exceed 200mA.

177 120 150 Maximum kV in pulsed fluoroscopy


178 1 100 Percentage of tube used.
179 0 4 Number of minutes of time counting
180 0 59 Number of seconds of time counting
181 0 255 Optimization of Video Ref in automatic
182 0 255 Servocontrol of Video Ref in automatic
183 0 1 Constant heating current on post 1
184 0 1 Constant heating current on post 2
185 0 1 Constant heating current on post 3
186 0 1 Constant heating current on post 4
187 0 1 Constant heating current on post 5
188 0 1 Constant heating current on post 6
0=Normal
1=No Constant heating current
189 0 1 Use of a dual speed starter on tube 1
0=Normal: 3000rpm
1=DSS on: 9000rpm
190 0 1 Use of a dual speed starter on tube 2
0=Normal: 3000rpm
1=DSS on: 9000rpm
191 0 1 Falling load on Post 1
192 0 1 Falling load on Post 2
193 0 1 Falling load on Post 3
194 0 1 Falling load on Post 4
195 0 1 Falling load on Post 5
196 0 1 Falling load on Post 6

Ref : 32-40-013-C
12
Generator GTI HF

3 Parameter description
PARAMETER 0: MAXIMUM POSSIBLE POWER OF THE POWER SUPPLY
Minimum value: 1
Maximum value: 80
Parameter unit: kW
Parameter signification: Maximum possible power of the power supply.
Explication: This parameter is used by the calculator on one part for calculating the inter-
dictions to determine if the power requested by the operator for an exposure exceeds the
power supply’s capacity; in order to avoid its deterioration. On the other part this parameter
displays the highest values possible of the high-voltage current in the case where an
80kW power supply is used.
You can enter any value between 1kW and 80kW. But the typical values are: 40 for a
50kW generator not equiped with a primary transformer, 50 for a power supply equiped
with a primary transformer, and 80 for a power supply of 80kW.
Note: The calculator uses a binary calculus arithmetic exempt from decimal numbers; it
is possible you will notice a small error of calculus due to a rounding-off in the intermediary
calculus of interdiction. In any case, the error by the calculator will not affect the function
security of the generator.

PARAMETER 1: TUBE CORRESPONDING WITH POST 1


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 1.
Explication: This parameter fixes the number of the tube corresponding to post 1.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

PARAMETER 2: TUBE CORRESPONDING WITH POST 2


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 2.
Explication: This parameter fixes the number of the tube corresponding to post 2.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

Ref : 32-40-013-C
13
Chapter 2 / Technical manuel

PARAMETER 3: TUBE CORRESPONDING WITH POST 3


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 3.
Explication: This parameter fixes the number of the tube corresponding to post 3.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

PARAMETER 4: TUBE CORRESPONDING WITH POST 4


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 4.
Explication: This parameter fixes the number of the tube corresponding to post 4.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

PARAMETER 5: TUBE CORRESPONDING WITH POST 5


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 5.
Explication: This parameter fixes the number of the tube corresponding to post 5.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

PARAMETER 6: TUBE CORRESPONDING WITH POST 6


Minimum value: 0
Maximum value: 2
Parameter signification: Tube corresponding with Post 6.
Explication: This parameter fixes the number of the tube corresponding to post 6.
If you enter the value 0, the post will be inhibited and it will be impossible for the user to
work in this technic.
If you enter the value 1, the post will be accessable and the corresponding tube will be
tube number 1.
If you enter the value 2, the post will be accessable and the corresponding tube will be
tube number 2.

Ref : 32-40-013-C
14
Generator GTI HF

PARAMETER 7: MAXIMUM MA GRAPHY IN TOMOGRAPHY


Minimum value: 5
Maximum value: 13
Parameter unit: (pointer of the table)
Parameter signification: Maximum mA graphy in tomography.
Explication: This current fixes the maximum current possible that the operator can request
in the tomography technic.
In tomography, the high-voltage current is generally limited so that sometimes the operator
is left with the possibility of working only on the small focus of the tube.
The value that you enter for this parameter gives the maximum value possible:
Value entered: 5Maximum graphy current: 100mA
Value entered: 6Maximum graphy current: 150mA
Value entered: 7Maximum graphy current: 200mA
Value entered: 8Maximum graphy current: 300mA
Value entered: 9Maximum graphy current: 400mA
Value entered: 10 Maximum graphy current: 500mA
Value entered: 11 Maximum graphy current: 600mA
Value entered: 12 Maximum graphy current: 800mA
Value entered: 13 Maximum graphy current: 1000mA

PARAMETER 8: YES/NO DISPLAY OF THE INTERNAL MAS-METER


Minimum value: 0
Maximum value: 1
Parameter signification: Yes/No display of the internal mAs-meter.
Explication: This parameter allows for the internal mAs-meter display during a graphy.
If you enter 0 as the value of this parameter, there will not be an internal mAs-meter
display except in the case of working with the automatic exposure controller.
The mAs display unit serves both to display the mAs requested by the operator during the
manual adjust-ment of exposure constants, and the display of the mAs during an exposure.
If you enter 1 as the value of this parameter, there will always be a display of the internal
mAs-meter during an exposure.
Note: The internal mAs-meter display starts the X-ray emissions and becomes normal
as soon as the graphy command is released. Also, the internal mAs-meter’s precision
can be faulty in the case of a very short exposure or if the high-voltage current is weak.

PARAMETER 9: AUTOMATIC SCOPIE WHEN STARTING


Minimum value: 0
Maximum value: 1
Parameter signification: Automatic scopy when starting the apparatus.
Explication: If you enter 0 there will be no automatic scopy by default when the generator
is being started.
If you enter 1 there will be a validation of the automatic scopy mode by default when the
generator is being started.

Ref : 32-40-013-C
15
Chapter 2 / Technical manuel

PARAMETER 10: YES/NO VALIDATION OF THE PULSED MODE WHEN STARTING


Minimum value: 0
Maximum value: 1
Parameter signification: Yes/No validation of the pulsed mode when starting the apparatus.
Explication: This parameter allows you to define the state of the validation key for the
pulsed scopy.If you enter the value 1, post 1 will be activated at the startup.
PARAMETER 11: SELECTION OF THE ACTIVE POST WHEN STARTING
Minimum value: 1
Maximum value: 6
Parameter signification: Selection of the active post when starting the apparatus.
Explication: You enter as the value for this configuration parameter the number of the
poste that you want when the apparatus is started.
If you enter the value 1, post 1 will be activated at the startup.
If you enter the value 2, post 2 will be activated at the startup.
If you enter the value 3, post 3 will be activated at the startup.
If you enter the value 4, post 4 will be activated at the startup.
If you enter the value 5, post 5 will be activated at the startup.
If you enter the value 6, post 6 will be activated at the startup.

PARAMETER 12: YES/NO VALIDATION OF THE AUTOMATIC STANDBY


Minimum value: 0
Maximum value:255
Parameter signification: validation of the automatic standby.
Explication: The generator is equiped with an automatic economizing system which place
it in a state of standby after( parameter 12 )xminutes of non-use. To leave the standby
mode, the operator presses either, any one of the control panel keys, the graphy control,
or the scopy pedal.
The automatic standby mode is validated if you enter a valuedifferent of 0 in this parameter.
If you enter the value 0 for this parameter, the generator will never go into the standby
mode.

PARAMETER 14: SPECIAL


Minimum value: 0
Maximum value: 255
Parameter signification: If you program 255 the eeprom will be filled by values by default

Ref : 32-40-013-C
16
Generator GTI HF

PARAMETER 18: AUTOMATIC EXPOSURE NUMBER USED FOR POST 1


Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 1.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 1.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.

PARAMETER 19: AUTOMATIC EXPOSURE NUMBER USED FOR POST 2


Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 2.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 2.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.

PARAMETER 20: AUTOMATIC EXPOSURE NUMBER USED FOR POST 3


Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 3.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 3.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.

Ref : 32-40-013-C
17
Chapter 2 / Technical manuel

PARAMETER 21: AUTOMATIC EXPOSURE NUMBER USED FOR POST 4


Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 4.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 4.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.

PARAMETER 22: AUTOMATIC EXPOSURE NUMBER USED FOR POST 5


Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 5.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 5.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.
PARAMETER 23: AUTOMATIC EXPOSURE NUMBER USED FOR POST 6

Minimum value: 0
Maximum value: 3
Parameter signification: Automatic exposure number used for post 6.
Explication: When using the automatic exposure controller, it is necessary to to define for
each post, the number of the chamber corresponding to that post.
When you use a remote control automatic exposure controller, the transfer of the chamber
number is done transparantly.
It is possible to chose one chamber out of three.
If you enter 0 as the value of this parameter, you inhibit the function of the automatic
exposure controller for post 6.
If you enter 1 as the value for this parameter, you associate chamber 1 to post 1.
If you enter 2 as the value for this parameter, you associate chamber 2 to post 1.
If you enter 3 as the value for this parameter, you associate chamber 3 to post 1.

Ref : 32-40-013-C
18
Generator GTI HF

PARAMETER 24: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 1
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 1.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure controller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 1 each time that the user requests this post.

PARAMETER 25: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 2
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 2.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure controller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 2 each time that the user requests this post.

PARAMETER 26: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 3
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 3.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure controller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 3 each time that the user requests this post.

PARAMETER 27: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 4
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 4.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure controller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 4 each time that the user requests this post.

Ref : 32-40-013-C
19
Chapter 2 / Technical manuel

PARAMETER 28: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 5
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 5.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure controller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 5 each time that the user requests this post.

PARAMETER 29: PREFERED FUNCTION MODE FOR THE AUTOMATIC EXPOSURE


CONTROLLER FOR POST 6
Minimum value: 0
Maximum value: 23
Parameter signification: Prefered function mode for the automatic exposure controller for
post 6.
Explication: It is possible to define a prefered function mode to work with the automatic
exposure con-troller for each operation technic.
In programming a value in this parameter, you define the active exposure field and the
screen re-enforcer used by default for post 6 each time that the user requests this post.
If you enter 0 for the value of this parameter, you do not select an exposure field, but you
do select the slow screen re-enforcer.
If you enter 1 for the value of this parameter, you select the right exposure field and the
slow screen re-enforcer.
If you enter 2 for the value of this parameter, you select the middle exposure field and the
slow screen re-enforcer.
If you enter 3 for the value of this parameter, you select the middle and right exposure
fields and the slow screen re-enforcer.
If you enter 4 for the value of this parameter, you select the left exposure field and the slow
screen re-enforcer.
If you enter 5 for the value of this parameter, you select the left and right exposure fields
and the slow screen re-enforcer.
If you enter 6 for the value of this parameter, you select the middle and left exposure fields
and the slow screen re-enforcer.
If you enter 7 for the value of this parameter, you select the left, middle and right exposure
fields and the slow screen re-enforcer.
If you enter 8 for the value of this parameter, you do not select an exposure field, but you
select the normal screen re-enforcer.
If you enter 9 for the value of this parameter, you select the right exposure fields and the
normal screen re-enforcer.
If you enter 10 for the value of this parameter, you select the middle exposure field and the
normal screen re-enforcer.
If you enter 11 for the value of this parameter, you select the middle and right exposure

Ref : 32-40-013-C
20
Generator GTI HF

fields and the normal screen re-enforcer.


If you enter 12 for the value of this parameter, you select the left exposure field and the
normal screen re-enforcer.
If you enter 13 for the value of this parameter, you select the left and right exposure fields
and the normal screen re-enforcer.
If you enter 14 for the value of this parameter, you select the middle and left exposure
fields and the normal screen re-enforcer.
If you enter 15 for the value of this parameter, you select the left, middle, and right exposure
fields and the normal screen re-enforcer.
If you enter 16 for the value of this parameter, you do not select an exposure field, but you
do select the rapid screen re-enforcer.
If you enter 17 for the value of this parameter, you select the right exposure field and the
rapid screen re-enforcer.
If you enter 18 for the value of this parameter, you select the midlle exposure field and the
normal screen re-enforcer.
If you enter 19 for the value of this parameter, you select the middle and right exposure
fields and the normal screen re-enforcer.
If you enter 20 for the value of this parameter, you select the left exposure field and the
rapid screen re-enforcer.
If you enter 21 for the value of this parameter, you select the left and right exposure fields
and the rapid screen re-enforcer.
If you enter 22 for the value of this parameter, you select the middle and left exposure
fields and the rapid screen re-enforcer.
If you enter 23 for the value of this parameter, you select the left, middle, and right exposure
fields and the rapid screen re-enforcer.

PARAMETER 30: SCOPY AUTHORIZATION / INHIBITION ON POST 1


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 1.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

PARAMETER 31: SCOPY AUTHORIZATION / INHIBITION ON POST 2


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 2.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

Ref : 32-40-013-C
21
Chapter 2 / Technical manuel

PARAMETER 32: SCOPY AUTHORIZATION / INHIBITION ON POST 3


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 3.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

PARAMETER 33: SCOPY AUTHORIZATION / INHIBITION ON POST 4


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 4.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

PARAMETER 34: SCOPY AUTHORIZATION / INHIBITION ON POST 5


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 5.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

PARAMETER 35: SCOPY AUTHORIZATION / INHIBITION ON POST 6


Minimum value: 0
Maximum value: 1
Parameter signification:Authorization / inhibition of fluoroscopy on post 6.
Explication: Certain technics are not equiped with an image intensifier. It is possible to
inhibit the scopy on these posts.
If you enter 0 as the value of this parameter, you authorize the scopy for this technic.
If you enter 1 as the value of the parameter, you inhibit the scopy for this technic.

PARAMETER 36: CONSIDERATION OF A HIGH SPEED STARTER FOR TUBE 1


Minimum value: 0
Maximum value: 1
Parameter signification: Taking into consideraiton the presence of a high speed starter
for tube 1.
Explication: If you desire to use an external rotor power supply instead of the original
power supply of the generator, you must consequently program this parameter. Indicating
the presence of an extermal starter will have the following conseqences:

Ref : 32-40-013-C
22
Generator GTI HF

•The mangament program will inhibit the internal rotor power supply commands.
•You must command your rotor power supply with the help of the relay contacts of
the interface board.
•You must signal the interface board that the nominal rotation speed is reached by
validating the anode authorization input signal.
To command your rotor power supply, you must use the post contacts associated with
the tube and the relay contacts of the first graphy time.
If you enter 0 for the value of this parameter, the generator’s internal rotor power supply is
used.
If you enter 1 for the value of this parameter, you indicate that tube 1 is equiped with an
external rotor power supply, and you must validate the anode authorization input of the
interface board.

PARAMETER 37: CONSIDERATION OF A HIGH SPEED STARTER FOR TUBE 2


Minimum value: 0
Maximum value: 1
Parameter signification: Taking into consideraiton the presence of a high speed starter
for tube 2.
Explication: If you desire to use an external rotor power supply instead of the original
power supply of the generator, you must consequently program this parameter. Indicating
the presence of an extermal starter will have the following conseqences:
•The mangament program will inhibit the internal rotor power supply commands.
•You must command your rotor power supply with the help of the relay contacts of
the interface board.
•You must signal the interface board that the nominal rotation speed is reached by
validating the anode authorization input signal.
To command your rotor power supply, you must use the post contacts associated with
the tube and the relay contacts of the first graphy time.
If you enter 0 for the value of this parameter, the generator’s internal rotor power supply is
used.
If you enter 1 for the value of this parameter, you indicate that tube 2 is equiped with an
external rotor power supply, and you must validate the anode authorization input of the
interface board.

PARAMETER 38: ANODE ROTATION FOR TUBE 1 IN FLUOROSCOPY(NOT USE IF


HIGH SPEED STARTER)
Minimum value: 0
Maximum value: 1
Parameter signification: The rotation of the anode during fluoroscopy for tube 1.
Explication: If you desire to rotate the anode in fluoroscopy, enter the value 1 as the value
for this parameter.
If you desire a fluroscopy without rotation of the anode, enter the value 0 for this parameter.

Ref : 32-40-013-C
23
Chapter 2 / Technical manuel

PARAMETER 39: ANODE ROTATION FOR TUBE 2 IN FLUOROSCOPY


Minimum value: 0
Maximum value: 1
Parameter signification: The rotation of the anode during fluoroscopy for tube 2.
Explication: If you desire to rotate the anode in fluoroscopy, enter the value 2 as the value
for this parameter.

PARAMETER 40: SMALL FOCUS STANDBY HEATING CURRENT


Minimum value: 1
Maximum value: 255
Parameter signification: The standby heating current for the small focus.
Explication: This parameter fixes the standby heating current for the small focus. The
value of the current is obained by the following formula:
Small focus current (mAs) = Parameter 40 * 14 + 2000
Example: A value of 1 gives a current of 2014 mA; a value of 30 gives a current of 2420
mA.
PARAMETER 41: LARGE FOCUS STANDBY HEATING CURRENT
Minimum value: 1
Maximum value: 255
Parameter signification: The standby heating current for the large focus.
Explication: This parameter fixes the standby heating current for the large focus. The
value of the current is obained by the following formula:
Large focus current (mAs) = Parameter 41 * 14 + 2000
Example: A value of 1 gives a current of 2014 mA; a value of 30 gives a current of 2420
mA.
PARAMETER 42: kV MAXIMUM FOR TUBE 1
Minimum value: 120
Maximum value: 150
Parameter unit: kV
Parameter signification: Maximum kV for tube 1.
Explication: This parameter fixes the maximum value that can be requested by the user
during a graphy with tube 1. Normally, the generator can deliver a maximum high-voltage
of 150 kV. However, the tube housing ensembles have a lower high-voltage limit. It is
necessary to fix the maximum value permited to avoid high-voltage arcing due to this
limitation.
PARAMETER 43: kV MAXIMUM FOR TUBE 2
Minimum value: 120
Maximum value: 150
Parameter unit: kV
Parameter signification: Maximum kV for tube 2.
Explication: This parameter fixes the maximum value that can be requested by the user
during a graphy with tube 2. Normally, the generator can deliver a maximum high-voltage
of 150 kV. However, the tube-housing ensembles have a lower high-voltage limit. It is
necessary to fix the maximum value permited to avoid high-voltage arcing due to this
limitation.

Ref : 32-40-013-C
24
Generator GTI HF

PARAMETER 44: VALIDATION / INVALIDATION OF TOMOGRAPHY FOR TUBE 1


Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the tomography mode for tube 1.
Explication: The tomography mode is characterized by the possiblitiy of requesting
exposures with a high voltage current as low as 10 mA. Validating the tomography mode
allows for a current as low as 10 mA to be obtained in graphy.
Enter 0 as the value for this parameter to inhibit the tomography mode and prohibit the
access of graphy currents lower than 50 mA.
Enter 1 as the value for this parameter to validate the tomography mode and allow the
access of graphy currents as low as 10 mA.

PARAMETER 45: VALIDATION / INVALIDATION OF TOMOGRAPHY FOR TUBE 2


Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the tomography mode for tube 2.
Explication: The tomography mode is characterized by the possiblitiy of requesting
exposures with a high voltage current as low as 10 mA. Validating the tomography mode
allows for a current as low as 10 mA to be obtained in graphy.
Enter 0 as the value for this parameter to inhibit the tomography mode and prohibit the
access of graphy currents lower than 50 mA.
Enter 1 as the value for this parameter to validate the tomography mode and allow the
access of graphy currents as low as 10 mA.

PARAMETER 46: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 1
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization
for post 1.
Explication: During the installation of the generator, you may find it necessary to validate
the graphy second time authorization when an external apparatus is ready. In this case
the interface board contains an input which must be validated when outside equipment
authorizes the graphy second time.
By validating this parameter you create a condition for the request of the graphy second
time which will result in the following: the operator requests the graphy second time and
the input of external validation of second time is validated. If during an exposure, the
external validation input of second time disappears, the exposure is interrupted in the
same way as if the operator had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the
input state of second time external validation.

Ref : 32-40-013-C
25
Chapter 2 / Technical manuel

PARAMETER 46: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 1
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization for
post 1.
Explication: During the installation of the generator, you may find it necessary to validate the
graphy second time authorization when an external apparatus is ready. In this case the
interface board contains an input which must be validated when outside equipment authorizes
the graphy second time.
By validating this parameter you create a condition for the request of the graphy second time
which will result in the following: the operator requests the graphy second time and the input
of external validation of second time is validated. If during an exposure, the external validation
input of second time disappears, the exposure is interrupted in the same way as if the operator
had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the input
state of second time external validation.

PARAMETER 47: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 2
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization
for post 2.
Explication: During the installation of the generator, you may find it necessary to validate the
graphy second time authorization when an external apparatus is ready. In this case the
interface board contains an input which must be validated when outside equipment authorizes
the graphy second time.
By validating this parameter you create a condition for the request of the graphy second time
which will result in the following: the operator requests the graphy second time and the input
of external validation of second time is validated. If during an exposure, the external validation
input of second time disappears, the exposure is interrupted in the same way as if the operator
had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the input
state of second time external validation.

Ref : 32-40-013-C
26
Generator GTI HF

PARAMETER 48: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 3
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization for
post 3.
Explication: During the installation of the generator, you may find it necessary to validate the
graphy second time authorization when an external apparatus is ready. In this case the
interface board contains an input which must be validated when outside equipment authorizes
the graphy second time.
By validating this parameter you create a condition for the request of the graphy second time
which will result in the following: the operator requests the graphy second time and the input
of external validation of second time is validated. If during an exposure, the external valida-
tion input of second time disappears, the exposure is interrupted in the same way as if the
operator had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the input
state of second time external validation.

PARAMETER 49: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 4
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization for
post 4.
Explication: During the installation of the generator, you may find it necessary to validate the
graphy second time authorization when an external apparatus is ready. In this case the
interface board contains an input which must be validated when outside equipment authorizes
the graphy second time.
By validating this parameter you create a condition for the request of the graphy second time
which will result in the following: the operator requests the graphy second time and the input
of external validation of second time is validated. If during an exposure, the external valida-
tion input of second time disappears, the exposure is interrupted in the same way as if the
operator had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the input
state of second time external validation.

Ref : 32-40-013-C
27
Chapter 2 / Technical manuel

PARAMETER 50: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 5
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization
for post 5.
Explication: During the installation of the generator, you may find it necessary to validate
the graphy second time authorization when an external apparatus is ready. In this case
the interface board contains an input which must be validated when outside equipment
authorizes the graphy second time.
By validating this parameter you create a condition for the request of the graphy second
time which will result in the following: the operator requests the graphy second time and
the input of external validation of second time is validated. If during an exposure, the
external validation input of second time disappears, the exposure is interrupted in the
same way as if the operator had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the
input state of second time external validation.

PARAMETER 51: VALIDATION / INVALIDATION OF THE SECOND TIME EXTERNAL


AUTHORIZATION FOR POST 6
Minimum value: 0
Maximum value: 1
Parameter signification: Validation / invalidation of the second time external authorization
for post 6.
Explication: During the installation of the generator, you may find it necessary to validate
the graphy second time authorization when an external apparatus is ready. In this case
the interface board contains an input which must be validated when outside equipment
authorizes the graphy second time.
By validating this parameter you create a condition for the request of the graphy second
time which will result in the following: the operator requests the graphy second time and
the input of external validation of second time is validated. If during an exposure, the
external validation input of second time disappears, the exposure is interrupted in the
same way as if the operator had released the second time button.
Enter 0 as the value for this parameter in order to have the graphy second time take place
without validation by an external signal.
Enter 1 as the value for this parameter in order that the generator takes into account the
input state of second time external validation.

Ref : 32-40-013-C
28
Generator GTI HF

PARAMETER 52: WAITING TIME BETWEEN THE MOMENT OF THE FLUOROSCOPY


REQUEST AND THE PASSAGE INTO LIVE MEMORY
Minimum value: 0
Maximum value: 255
Parameter unit: 1/100th of a second.
Parameter signification: Waiting time between the moment of the fluoroscopy request
and the passage into live memory.
Explication: So as to avoid that the moniter linked to the video memory does not stay
black immediately after the request for scopy, this parameter fixes the duration of the time
delay at the end of which the memory displays the live image.

PARAMETER 53: FLUOROSCOPY PROLONGATION TIME FOR THE MEMORIZATION


Minimum value: 0
Maximum value: 255
Parameter unit: 1 / 100th of a second.
Parameter signification: Fluoroscopy prolongation time for the memorization.
Explication: A digital video memory stores the last image at the moment the scopy
command is released. This parameter fixes the duration of the time delay during which
the scopy is prolonged so that the mem-orization can be completed.

PARAMETER 54: STABILIZATION TIME OF THE AUTOMATIC kV = STARTING TIME


OF PULSED FLUOROSCOPY
Minimum value: 0
Maximum value: 255
Parameter unit: 1/100th of a second.
Parameter signification: Stabilization time of the automatic kV = starting time of pulsed
fluoroscopy.
Explication: The pulsed fluoroscopy does not begin immediately after a command for
scopy. Program this parameter so as to allow time for the automatic kV system to find its’
equilibrium.

PARAMETER 55: PULSED FLUOROSCOPY STANDBY TIME


Minimum value: 0
Maximum value: 255
Parameter unit: 1/10th of a second.
Parameter signification: Pulsed fluoroscopy standby time.
Explication: This parameter fixes the duration of the cutoff or X-rays in pulsed fluoroscopy.

PARAMETER 56: LIVE MEMORY TIME


Minimum value: 0
Maximum value: 255
Parameter unit: 1/100th of a second.
Parameter signification: Live memory time.
Explication: In pulsed fluorscopy, this parameter fixes the time where the memory is live.

Ref : 32-40-013-C
29
Chapter 2 / Technical manuel

PARAMETER 57: TRIGGERING OF LARGE FOCUS TUBE 1


Minimum value: 0
Maximum value: 14
Parameter signification:triggering of large focus tube 1
If you enter 0 large focus will be always activated
If you enter 1 large focus will be activated at 10 mA
If you enter 2 large focus will be activated at 20 mA
If you enter 3 large focus will be activated at 30 mA
If you enter 4 large focus will be activated at 40 mA
If you enter 5 large focus will be activated at 50 mA
If you enter 6 large focus will be activated at 100 mA
If you enter 7 large focus will be activated at 200 mA
If you enter 8 large focus will be activated at 300 mA
If you enter 9 large focus will be activated at 400 mA
If you enter 10 large focus will be activated at 500 mA
If you enter 11 large focus will be activated at 600 mA
If you enter 12 large focus will be activated at 800 mA
If you enter 13 large focus will be activated at 1000 mA
If you enter 14 large focus will never be activated

PARAMETER 58: TRIGGERING OF LARGE FOCUS TUBE 2


Minimum value: 0
Maximum value: 14
Parameter signification:triggering of large focus tube 1
If you enter 0 large focus will be always activated
If you enter 1 large focus will be activated at 10 mA
If you enter 2 large focus will be activated at 20 mA
If you enter 3 large focus will be activated at 30 mA
If you enter 4 large focus will be activated at 40 mA
If you enter 5 large focus will be activated at 50 mA
If you enter 6 large focus will be activated at 100 mA
If you enter 7 large focus will be activated at 200 mA
If you enter 8 large focus will be activated at 300 mA
If you enter 9 large focus will be activated at 400 mA
If you enter 10 large focus will be activated at 500 mA
If you enter 11 large focus will be activated at 600 mA
If you enter 12 large focus will be activated at 800 mA
If you enter 13 large focus will be activated at 1000 mA
If you enter 14 large focus will never be activated

Ref : 32-40-013-C
30
Generator GTI HF

PARAMETER 59: inhibition for some keys for post 1


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field
Example:to inhibit automatic fluoroscopy and pulsed fluoroscopy keys,you just have to
make the following code:
bit 7/ bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
0 / 0 / 1 / 1 / 1 / 1 /1 / 1
This gives 1+2+4+8+16+32+0x64+0x128=63

PARAMETER 60: inhibition for some keys for post 2


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field

PARAMETER 61: inhibition for some keys for post 3


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field

Ref : 32-40-013-C
31
Chapter 2 / Technical manuel

PARAMETER 62: inhibition for some keys for post 4


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field

PARAMETER 63: inhibition for some keys for post 5


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field

PARAMETER 64: inhibition for some keys for post 6


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to inhibits some keys not used in the
concerned post
Explication:To inhibit the key, you just have to set the corresponding bit to 0
bit 7:inhibit automatic fluoroscopy
bit 6:inhibit pulsed fluoroscopy
bit 5:inhibit low screen
bit 4:inhibit normal screen
bit 3:inhibit fast screen
bit 2:inhibit left field
bit 1:inhibit middle field
bit 0:inhibit right field

Ref : 32-40-013-C
32
Generator GTI HF

PARAMETER 65: compensation for exposure time in ms


Minimum value: 0
Maximum value: 255
Parameter signification:compensation for exposure time in ms.
Explication:The contents of this memory is substracted from the exposure time asked by
the user,to compensate the difference between command for XRay and real exposure

PARAMETER 66: Heat integrator prevent exposures


Minimum value: 0
Maximum value: 1
Parameter signification:Heat integrator prevent exposures
Explication:This parameter allows to stop XRay if the heat integrator has reach the maxi-
mum.
If you enter 0 Heat integrator stop exposures
If you enter 1Heat integrator allows exposures but the buzzer is still activated

PARAMETER 67: Percentage of the tube power used with post 1


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 1
Explication:This parameter allows to reduce the tube power in function with the selected
post

PARAMETER 68:Percentage of the tube power used with post 2


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 2
Explication:This parameter allows to reduce the tube power in function with the selected
post

PARAMETER 69: Percentage of the tube power used with post 3


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 3
Explication:This parameter allows to reduce the tube power in function with the selected
post

PARAMETER 70: Percentage of the tube power used with post 4


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 4
Explication:This parameter allows to reduce the tube power in function with the selected
post

Ref : 32-40-013-C
33
Chapter 2 / Technical manuel

PARAMETER 71: Percentage of the tube power used with post 5


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 5
Explication:This parameter allows to reduce the tube power in function with the selected
post

PARAMETER 72: Percentage of the tube power used with post 6


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used with post 6
Explication:This parameter allows to reduce the tube power in function with the selected
post

PARAMETER 73: Post condition relay 1


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 1
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.The condition PL6&Pofil is used only in
Inferiors members.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1
Example:if we needs to activate relay 1 with post 1 & 2 we just have to make the following
code
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
0 0 0 0 0 0 1 1
0x128+0x64+0x32+0x16+0x8+0x4+1x2+1x1=3
Warning:state of relay depends of parameters post condition and state condition

PARAMETER 74: Post condition relay 2


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 2
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 75: Post condition relay 3


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 3
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

Ref : 32-40-013-C
34
Generator GTI HF

PARAMETER 76: Post condition relay 4


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 4
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 77: Post condition relay 5


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 5
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.The condition post6 & profil is used for
inferior members.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 78: Post condition relay 6


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 6
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 79: Post condition relay 7


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 7
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 80: Post condition relay 8


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 8
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

Ref : 32-40-013-C
35
Chapter 2 / Technical manuel

PARAMETER 81: Post condition relay 9


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 9
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 82: Post condition relay 10


Minimum value: 0
Maximum value: 63
Parameter signification:Post condition relay 10
Explication:This parameter allows the relay to be active with the differents posts used.the
choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
- PL6&PROFIL PL6 PL5 PL4 PL3 PL2 PL1

PARAMETER 83: State condition relay 1


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 1
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.The condition mask is
used only in Inferiors members.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready
Example:if we needs to activate relay 1 in fluoroscopy we just have to make the following
code
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
0 0 0 0 0 0 1 0
0x128+0x64+0x32+0x16+0x8+0x4+1x2+1x0=2

PARAMETER 84: State condition relay 2


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 2
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

Ref : 32-40-013-C
36
Generator GTI HF

PARAMETER 85: State condition relay 3


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 3
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 86: State condition relay 4


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 4
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 87: State condition relay 5


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 5
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 88: State condition relay 6


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 6
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 89: State condition relay 7


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 7
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

Ref : 32-40-013-C
37
Chapter 2 / Technical manuel

PARAMETER 90: State condition relay 8


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 8
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 91: State condition relay 9


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 9
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 92: State condition relay 10


Minimum value: 0
Maximum value: 64
Parameter signification:State condition relay 10
Explication:This parameter allows the relay to be active with the differents states of the
generator.The choice is made with a 1 with the corresponding bit.
bit 7 / bit 6 / bit 5 / bit 4 / bit 3 / bit 2 / bit 1 / bit 0
mask/ inh/ - /prep/bucky/graphy/scopy/ready

PARAMETER 93: FPS by default of cinema post


Minimum value: 0
Maximum value: 2
Parameter signification:FPS by default of cinema post
Explication:This parameter displays on start of post 5 the FPS.
0=12 images per second
1=25 images per second
2=50 images per second

PARAMETER 94: serial time by default of cinema post


Minimum value: 0
Maximum value: 4
Parameter signification:serial time by default of cinema post
Explication:This parameter displays on start of post 5 the serial time.
0=2 seconds
1=5 seconds
2=10 seconds
3=15 seconds
4=20 seconds

Ref : 32-40-013-C
38
Generator GTI HF

PARAMETER 95: Exposure time in cinema in ms


Minimum value: 3
Maximum value: 20
Parameter signification:Exposure time in cinema in ms
Explication:This parameter defines the exposure time in ms during cinema serial.
PARAMETER 96: Operating mode post 5
Minimum value: 0
Maximum value: 1
Parameter signification:Operating mode post 5
Explication:This parameter defines the post 5 (cinema or not)
0=post 5 is normal
1=post 5 is a cinema post

Ref : 32-40-013-C
39
Chapter 2 / Technical manuel

PARAMETER 99: Cooling factor for anode tube 1


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter is calculate from the cooling curve of the tube

PARAMETER 100: Cooling factor for anode tube 2


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter is calculate from the cooling curve of the tube

PARAMETER 101: Threshold for triggering heating alarm anode tube 1(lsb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 1(lsb)joules

PARAMETER 102: Threshold for triggering heating alarm anode tube 1()joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 1()joules

PARAMETER 103: Threshold for triggering heating alarm anode tube 1(msb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 1(msb)joules
Explication:We use the maximum anode heating storage in JOULES;we divise by 2 to the
power to 16 ,it gives the msb with the whole part;we divise the rest by 2 to the power to 8
,it gives the()with the whole part;the rest gives the lsb

PARAMETER 104: Threshold for triggering heating alarm anode tube 2(lsb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 2(lsb)joules

PARAMETER 105: Threshold for triggering heating alarm anode tube 2()joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 2()joules

PARAMETER 106: Threshold for triggering heating alarm anode tube 2(msb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm anode tube 2(msb)joules
Explication:We use the maximum anode heating storage in JOULES;we divise by 2 to the
power to 16 ,it gives the msb with the whole part;we divise the rest by 2 to the power to 8
,it gives the()with the whole part;the rest gives the lsb

Ref : 32-40-013-C
40
Generator GTI HF

PARAMETER 107: Cooling factor for housing tube 1


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter is calculate from the cooling curve of the tube

PARAMETER 108:Cooling factor for housing tube 2


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter is calculate from the cooling curve of the tube

PARAMETER 109: Threshold for triggering heating alarm housing tube 1(lsb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 1(lsb)joules

PARAMETER 110: Threshold for triggering heating alarm housing tube 1()joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 1()joules

PARAMETER 111: Threshold for triggering heating alarm housing tube 1(msb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 1(msb)joules
Explication:We use the maximum housing heating storage in JOULES;we divise by 2 to
the power to 16 ,it gives the msb with the whole part;we divise the rest by 2 to the power
to 8 ,it gives the()with the whole part;the rest gives the lsb

PARAMETER 112: Threshold for triggering heating alarm housing tube 2(lsb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 2(lsb)joules

PARAMETER 113: Threshold for triggering heating alarm housing tube 2()joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 2()joules

PARAMETER 114: Threshold for triggering heating alarm housing tube 2(msb)joules
Minimum value: 0
Maximum value: 255
Parameter signification:Threshold for triggering heating alarm housing tube 2(msb)joules
Explication:We use the maximum housing heating storage in JOULES;we divise by 2 to
the power to 16 ,it gives the msb with the whole part;we divise the rest by 2 to the power
to 8 ,it gives the()with the whole part;the rest gives the lsb

Ref : 32-40-013-C
41
Chapter 2 / Technical manuel

PARAMETER 115: Servocontrol factor for photomultiplier


Minimum value: 0
Maximum value: 255
Parameter signification:Servocontrol factor for photomultiplier
Explication:This parameter allows to adjust reaction of kV graphy in automatic cinema

PARAMETER 116: Power threshold for interface relays


Minimum value: 0
Maximum value: 255
Parameter signification:Power threshold for interface relays
Explication:This parameter define in kW the threshold where the power signal will be
activate(see parameters 82>92)

PARAMETER 117: Blackening for post 1


Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 1
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4
PARAMETER 118: Blackening for post 2
Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 2
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3

Ref : 32-40-013-C
42
Generator GTI HF

8 +4
PARAMETER 119: Blackening for post 3
Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 3
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4

PARAMETER 120: Blackening for post 4


Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 4
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4

PARAMETER 121: Blackening for post 5


Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 5
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4

Ref : 32-40-013-C
43
Chapter 2 / Technical manuel

PARAMETER 122: Blackening for post 6


Minimum value: 0
Maximum value: 8
Parameter signification:Blackening for post 6
Explication:This parameter defines on start the select value of blackening
Code
0 -4
1 -3
2 -2
3 -1
4 0
5 +1
6 +2
7 +3
8 +4

PARAMETER 123: Presence of an anatomical keyboard


Minimum value: 0
Maximum value: 1
Parameter signification:Presence of an anatomical keyboard
Explication: 1=anatomical keyboard
0=no anatomical keyboard

PARAMETER 124: Offset of recopy


Minimum value: 0
Maximum value: 10
Parameter signification:possibility to substract an offset value till 10kV on the copy of
scopy kV on the graphy.kV
Explication:Copy graphy kV=scopy kV-parameter 124

PARAMETER 125: kV graphy default value post 1


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 1

PARAMETER 126: kV graphy default value post 2


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 2

PARAMETER 127: kV graphy default value post 3


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 3

Ref : 32-40-013-C
44
Generator GTI HF

PARAMETER 128: kV graphy default value post 4


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 4

PARAMETER 129: kV graphy default value post 5


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 5

PARAMETER 130: kV graphy default value post 6


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of Kv for post 6

PARAMETER 131: mA graphy default value post 1


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 1
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

PARAMETER 132: mA graphy default value post 2


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 2
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

PARAMETER 133: mA graphy default value post 3


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 3
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

PARAMETER 134: mA graphy default value post 4


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 4
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

Ref : 32-40-013-C
45
Chapter 2 / Technical manuel

PARAMETER 135: mA graphy default value post 5


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 5
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

PARAMETER 136: mA graphy default value post 6


Minimum value: 0
Maximum value: 13
Parameter signification:This parameter allows to fix the starting value of mA for post 6
0: 10mA 1: 20mA 2: 30mA 3: 40mA 4: 50mA
5: 100mA 6: 150mA 7: 200mA 8: 300mA 9: 400mA
10: 500mA 11: 600mA 12:800mA 13:1000mA

PARAMETER 137: automatic graphy mode on post 1 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 1 if 1

PARAMETER 138: automatic graphy mode on post 2 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 2 if 1

PARAMETER 139: automatic graphy mode on post 3 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 3 if 1

PARAMETER 140: automatic graphy mode on post 4 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 4 if 1

PARAMETER 141: automatic graphy mode on post 5 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 5 if 1

Ref : 32-40-013-C
46
Generator GTI HF

PARAMETER 142: automatic graphy mode on post 6 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic graphy mode
on post 6 if 1

PARAMETER 143: automatic fluoroscopy mode on post 1 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 144: automatic fluoroscopy mode on post 2 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 145: automatic fluoroscopy mode on post 3 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 146: automatic fluoroscopy mode on post 4 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 147: automatic fluoroscopy mode on post 5 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 148: automatic fluoroscopy mode on post 6 if 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to select automatic fluoroscopy mode if 1

PARAMETER 149: total graphy number


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to the LSB of radiographic exposures.
To reset this counter memeorize with the automatic graphy when you are at parameter
149.

PARAMETER 150: total graphy number


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to the SB of radiographic exposures.

Ref : 32-40-013-C
47
Chapter 2 / Technical manuel

PARAMETER 151: total graphy number


Minimum value: 0
Maximum value: 255
Parameter signification:This parameter allows to the MSB of radiographic exposures.
Explication:To know the total graphy number you have to do:
(parameter 149 )X1+(parameter 150)X256 +(parameter 151)X256X256

PARAMETER 152: Rotor starting time tube 1


Minimum value: 0
Maximum value: 255
Parameter unit:1/10th of a second
Parameter signification:This parameter allows to fix the duration of the boost of the stator
of tube 1.NB In fluoroscopy the duration is fixed at 0.15s. t

PARAMETER 153: Rotor starting time tube 2


Minimum value: 0
Maximum value: 255
Parameter unit:1/10th of a second
Parameter signification:This parameter allows to fix the duration of the boost of the stator
of tube 2.NB In fluoroscopy the duration is fixed at 0.15s. t

PARAMETER 154: Inhibition time of fluoroscopy after a radiography tube 1


Minimum value: 0
Maximum value: 255
Parameter unit:1/10th of a second
Parameter signification:This parameter allows to fix the inhibition time of fluoroscopy after
a radiography tube 1

PARAMETER 155: Inhibition time of fluoroscopy after a radiography tube 2


Minimum value: 0
Maximum value: 255
Parameter unit:1/10th of a second
Parameter signification:This parameter allows to fix the inhibition time of fluoroscopy after
a radiography tube 2

PARAMETER 156: reading of kV graphy reference on An5 poste 1


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

PARAMETER 157: reading of kV graphy reference on An5 poste 2


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

Ref : 32-40-013-C
48
Generator GTI HF

PARAMETER 158: reading of kV graphy reference on An5 poste 3


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

PARAMETER 159: reading of kV graphy reference on An5 poste 4


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

PARAMETER 160: reading of kV graphy reference on An5 poste 5


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

PARAMETER 161: reading of kV graphy reference on An5 poste 6


Minimum value: 0
Maximum value: 1
Parameter signification:if you put 1 in this parameter the display of kV will depend of the
signal on An5 (if the key automatic radiography is on.)

PARAMETER 162: graphy time default value post 1


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 1
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

PARAMETER 163: graphy time default value post 2


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 2
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

Ref : 32-40-013-C
49
Chapter 2 / Technical manuel

PARAMETER 164: graphy time default value post 3


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 3
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

PARAMETER 165: graphy time default value post 4


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 4
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

PARAMETER 166: graphy time default value post 5


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 5
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

PARAMETER 167: graphy time default value post 6


Minimum value: 2
Maximum value: 36
Parameter signification:This parameter allows to fix the starting value of time for post 6
2:0.003 3:0.004 4:0.005 5:0.006 6:0.007 7:0.008
8:0.009 9: 0.010 10: 0.020 11: 0.030 12: 0.040 13: 0.050
14: 0.060 15: 0.070 16: 0.080 17: 0.090 18: 0.100 19: 0.200
20: 0.300 21: 0.400 22: 0.500 23: 0.600 24: 0.700 25: 0.800
26: 0.900 27: 1.000 28: 1.200 29: 1.500 30: 2.000 31: 2.500
32: 3.000 33: 3.500 34: 4.000 35: 4.500 36: 5.000

Ref : 32-40-013-C
50
Generator GTI HF

PARAMETER 168: kV graphy default value inferior members position 1


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of kV for position 1

PARAMETER 169: kV graphy default value inferior members position 2


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of kV for position 2

PARAMETER 170: kV graphy default value inferior members position 3


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of kV for position 3

PARAMETER 171: kV graphy default value inferior members position 4


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of kV for position 4

PARAMETER 172: kV graphy default value inferior members position 5


Minimum value: 40
Maximum value: 150
Parameter signification:This parameter allows to fix the starting value of kV for position 5

PARAMETER 173: Operating mode post 6


Minimum value: 0
Maximum value: 1
Parameter signification:Operating mode post 6
Explication:This parameter defines the post 5 (Inferior members or not)
0=post 5 is normal
1=post 5 is a inferior members post

PARAMETER 174: Pulsed fluoroscopy mode


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows toselect the using mode of the pulsed
fluoroscopy.
0=Pused fluoroscopy with video memory
1=Pulsed fluoroscopy in radiographic mode on tube 1
2=Pulsed fluoroscopy in radiographic mode on tube 2

PARAMETER 175: Heating current at 50kV


Minimum value: 0
Maximum value: 7
Parameter signification:This parameter allows to choose the fixed heating current at 50 KV

Ref : 32-40-013-C
51
Chapter 2 / Technical manuel

PARAMETER 176: Heating current at 150kV


Minimum value: 0
Maximum value: 7
Parameter signification:This parameter allows to choose the fixed heating current at 150
KV
0=10 mA
1=20 mA
2=30 mA
3=40 mA
4=50 mA
5=100 mA
6=150 mA
7=200 mA
WARNING:
1.Be sure that HV current doesn’t reach the tube’s caracteristics because
there is no control by the microprocessor.
2. Choose a value of current higher at 120kV than the one at 50kV.
3. Do not exceed 200mA.

PARAMETER 177: Maximum kV in pulsed fluoroscopy


Minimum value: 120
Maximum value: 150
Parameter signification:This parameter gives the maximum kV in pulsed fluoroscopy in
radiographic mode

PARAMETER 178: Percentage of the tube power in pulsed fluoroscopy


Minimum value: 0
Maximum value: 100
Parameter signification:Percentage of the tube power used in pulsed fluoroscopy in
radiography mode

PARAMETER 179: Display of number of minutes in fluoroscopy mode


Minimum value: 0
Maximum value: 4
Parameter signification:You can display the maximum minutes of pulsed fluoroscopy in
radiographic mode on the display generally used for counting the fluoroscopy time.

PARAMETER 180: Display of number of seconds in fluoroscopy mode


Minimum value: 0
Maximum value: 59
Parameter signification:You can display the maximum seconds of pulsed fluoroscopy in
radiographic mode on the display generally used for counting the fluoroscopy time.

Ref : 32-40-013-C
52
Generator GTI HF

PARAMETER 181: Optimization of Video Ref in automatic


Minimum value: 0
Maximum value: 255
Parameter signification: Adjustment of the step of the video referencein radiographic mode
in radiographic mode on the input An6 of the IC U31.
Explication: This parameter allows a good adjustmentthe value of kV according to the
ffedback on the ref kV.You just have to measure the feedback and enter the value equal to
(V/5)X 256.
Example: V measure =2,5 V parameter 181 = 2,5/5 X 256 = 128

PARAMETER 182: Servocontrol of Video Ref in automatic


Minimum value: 0
Maximum value: 255
Parameter signification: Precise adjustment of the servocontrol of the video referencein
radiographic mode in radiographic mode on the input An6 of the IC U31.

PARAMETER 183: Fixed current at on post 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

PARAMETER 184: Fixed current at on post 2


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

PARAMETER 185: Fixed current at on post 3


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

Ref : 32-40-013-C
53
Chapter 2 / Technical manuel

PARAMETER 186: Fixed current at on post 4


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

PARAMETER 187: Fixed current at on post 5


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

PARAMETER 188: Fixed current at on post 6


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a fixed heating current independant
of the displayed kV. This will evoid a forbidden given by a fast moving of kV which will not
let the time to the filament to cool.
if you enter 0=normal heating
if you enter1=heating with fixed current

PARAMETER 189: Use of a dual speed starter on tube 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to work with a dual spped starter Varian
integrated into the generator.
If you enter 0=no dual speed starter or use in 3000 rpm
If you use a dual sped starter in low speed you have to put parameters 152 or 153 at 0
If you enter 1=Dual speed starter Varian and use in 9000 rpm

PARAMETER 190: Use of a dual speed starter on tube 2


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to work with a dual speed starter Varian
integrated into the generator.
If you enter 0=no dual speed starter or use in 3000 rpm
If you use a dual sped starter in low speed you have to put parameters 152 or 153 at 0
If you enter 1=Dual speed starter Varian and use in 9000 rpm

Ref : 32-40-013-C
54
Generator GTI HF

PARAMETER 191: Falling load on post 1


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current
if you enter1=falling load

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

PARAMETER 192: Falling load on post 2


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current
if you enter1=falling load

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

PARAMETER 193: Falling load on post 3


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current
if you enter1=falling load

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

PARAMETER 194: Falling load on post 4


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current
if you enter1=falling load

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

Ref : 32-40-013-C
55
Chapter 2 / Technical manuel

PARAMETER 195: Falling load on post 5


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current
if you enter1=falling load

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

PARAMETER 196: Falling load on post 6


Minimum value: 0
Maximum value: 1
Parameter signification:This parameter allows to choose a falling heating current
independant of the displayed mA. The selection of small or large focus will be made by the
mA key.
if you enter 0=normal current

Note : The falling load only works if the triggering of the large focus (parameters 57
or 58 depending on the tube) is at 150 mA = value 7.

4 End of the configuration mode


When you have finished the modification of the configuration parameters, turn the
commutator S2 on the control card so as to illuminate the LED «NORMAL».

Ref : 32-40-013-C
56
Generator GTI HF

Configuration list
You are advised to record the values programmed.

Parameter Function Value


0 maximum power (kW)
1 tube linked with post 1
2 tube linked with post 2
3 tube linked with post 3
4 tube linked with post 4
5 tube linked with post 5
6 tube linked with post 6
7 mA graphy maximum in tomography
8 mA internal meter display
10 pulsed fluoroscopy mode when starting
11 post number when starting
12 0=inhibition of the automatic standby
14 special
18 auto. exposure n° used for post 1
19 auto. exposure n° used for post 2
20 auto. exposure n° used for post 3
21 auto. exposure n° used for post 4
22 auto. exposure n° used for post 5
23 auto. exposure n° used for post 6
24 auto. exposure mode by default from post 1
25 auto. exposure mode by default from post 2
26 auto. exposure mode by default from post 3
27 auto. exposure mode by default from post 4
28 auto. exposure mode by default from post 5
29 auto. exposure mode by default from post 6
30 fluoroscopy authorization on post 1
31 fluoroscopy authorization on post 2
32 fluoroscopy authorization on post 3
33 fluoroscopy authorization on post 4
34 fluoroscopy authorization on post 5
35 fluoroscopy authorization on post 6
36 tube 1 equiped with a high speed starter
37 tube 1 equiped with a high speed starter
38 rotation of the tube 1 anode in fluoroscopy
39 rotation of the tube 2 anode in fluoroscopy
40 small focus standby heating current
41 large focus standby heating current
42 kV maximum for tube1
43 kV maximum for tube 2
44 1=tomography mode for tube 1
45 1=tomography mode for tube 2

Ref : 32-40-013-C
57
Chapter 2 / Technical manuel

Parameter Function Value


46 second time external authorization for post 1
47 second time external authorization for post 2
48 second time external authorization for post 3
49 second time external authorization for post 4
50 second time external authorization for post 5
51 second time external authorization for post 6
52 tempo before passage into live memory after beginning
53 tempo of fluoroscopy prolongation for memorization
54 start of pulsed fluoroscopy=automatic kV stablization time
55 pulsed fluoroscopy standby time
56 live memory time
57 triggering of large focus tube 1
58 triggering of large focus tube 2
59 inhibition for some keys for post 1
60 inhibition for some keys for post 2
61 inhibition for some keys for post 3
62 inhibition for some keys for post 4
63 inhibition for some keys for post 5
64 inhibition for some keys for post 6
65 compensation for exposure time in ms
66 Heat integrator stop XRay
67 Percentage of the tube power used with post 1
68 Percentage of the tube power used with post 2
69 Percentage of the tube power used with post 3
70 Percentage of the tube power used with post 4
71 Percentage of the tube power used with post 5‘
72 Percentage of the tube power used with post 6
73 Post condition relay 1
74 Post condition relay 2
75 Post condition relay 3
76 Post condition relay 4
77 Post condition relay 5
78 Post condition relay 6
79 Post condition relay 7
80 Post condition relay 8
81 Post condition relay 9
82 Post condition relay 10
83 State condition relay 1
84 State condition relay 2
85 State condition relay 3
86 State condition relay 4
87 State condition relay 5
88 State condition relay 6
89 State condition relay 7

Ref : 32-40-013-C
58
Generator GTI HF

Parameter Function Value


90 State condition relay 8
91 State condition relay 9
92 State condition relay 10
93 FPS by default of cinema post
94 Serial time by default of cinema post
95 Exposure time in cinema in ms
96 Operating mode post 5
99 factor for anode tube 1
100 Cooling factor for anode tube 2
101 Threshold for triggering heating alarm anode tube 1(lsb)
102 Threshold for triggering heating alarm anode tube 1()
103 Threshold for triggering heating alarm anode tube 1(msb)
104 Threshold for triggering heating alarm anode tube 2(lsb)j
105 for triggering heating alarm anode tube 2()j
106 Threshold for triggering heating alarm anode tube 2(msb)j
107 Cooling factor for housing tube 1
108 Cooling factor for housing tube 2
109 Threshold for triggering heating alarm housing tube 1(lsb)j
110 Threshold for triggering heating alarm housing tube 1()j
111 Threshold for triggering heating alarm housing tube 1(msb)
112 Threshold for triggering heating alarm housing tube 2(lsb)j
113 Threshold for triggering heating alarm housing tube 2()j
114 Threshold for triggering heating alarm housing tube 2(msb)
115 Servocontrol factor for photomultiplier
116 Power threshold for interface relays
117 Blackening for post 1
118 Blackening for post 2
119 Blackening for post 3
120 Blackening for post 4
121 Blackening for post 5
122 Blackening for post 6
123 anatomical keyboard
124 Offset of recopy scopy>graphy
125 kV graphy default value post 1
126 kV graphy default value post 2
127 kV graphy default value post 3
128 kV graphy default value post 4
129 kV graphy default value post 5
130 kV graphy default value post 6

Ref : 32-40-013-C
59
Chapter 2 / Technical manuel

Parameter Function Value


131 mA graphy default value post 1
132 mA graphy default value post 2
133 mA graphy default value post 3
134 mA graphy default value post 4
135 mA graphy default value post 5
136 mA graphy default value post 6
137 automatic graphy mode on post 1 if 1
138 automatic graphy mode on post 2 if 1
139 automatic graphy mode on post 3 if 1
140 automatic graphy mode on post 4 if 1
141 automatic graphy mode on post 5 if 1
142 automatic graphy mode on post 6 if 1
143 automatic fluoroscopy mode on post 1 if 1
144 automatic fluoroscopy mode on post 2 if 1
145 automatic fluoroscopy mode on post 3 if 1
146 automatic fluoroscopy mode on post 4 if 1
147 automatic fluoroscopy mode on post 5 if 1
148 automatic fluoroscopy mode on post 6 if 1
149 LSB graphy number
150 SB graphy number
151 MSB graphy number
152 starting time tube 1(1/10ème de seconde)
153 starting time tube 2(1/10ème de seconde)
154 inhibition time fluoroscopy after a radiography tube 1
155 inhibition time fluoroscopy after a radiography tube 2
156 reading of kV graphy reference on An5 poste 1
157 reading of kV graphy reference on An5 poste 2
158 reading of kV graphy reference on An5 poste 3
159 reading of kV graphy reference on An5 poste 4
160 reading of kV graphy reference on An5 poste 5
161 reading of kV graphy reference on An5 poste 6
162 graphy time default value post 1
163 graphy time default value post 2
164 graphy time default value post 3
165 graphy time default value post 4
166 graphy time default value post 5
167 graphy time default value post 6
168 kV graphy by default post 1
169 kV graphy by default post 1
170 kV graphy by default post 1

Ref : 32-40-013-C
60
Generator GTI HF

Parameter Function Value


171 kV graphy by default post 1
172 kV graphy by default post 1
173 Operating mode post 6
174 Pused fluoroscopy modewith video memory
175 Heating current at 50kV
176 Heating current at 150kV
177 Maximum kV in pulsed fluoroscopy
178 Percentage of tube used.
179 Number of minutes of time counting
180 Number of seconds of time counting
181 Optimization of Video Ref in automatic
182 Servocontrol of Video Ref in automatic
183 Constant heating current on post 1
184 Constant heating current on post 2
185 Constant heating current on post 3
186 Constant heating current on post 4
187 Constant heating current on post 5
188 Constant heating current on post 6
189 Use of a dual speed starter on tube 1
190 Use of a dual speed starter on tube 2
191 Falling load on Post 1
192 Falling load on Post 2
193 Falling load on Post 3
194 Falling load on Post 4
195 Falling load on Post 5
196 Falling load on Post 6

Ref : 32-40-013-C
61
Chapter 2 / Technical manuel

Tube characteristics definition


1 Passage into the tube characteristics definition modeE
If you desire to define or modify the characteristics of tube 1, turn the commutator S2 on
the control card so as to illuminate the LED «DEFINITION TUBE 1 «.
If you desire to define or modify the characteristics of tube 2, turn the commutator S2 on
the control card so as to illuminate the LED «DEFINITION TUBE 2 «.
The generator then starts in the chosen tube characteristics definition mode.
At the end of a modification or definition phase of the tube characterisitics, do not forget to
return the commutator S2 to position «NORMAL» so as to allow for the use of the generator.
The display unit normally used to adjust the graphy kV becomes the EEPROM parameter
number display. The display unit normally used for the adjustment of the graphy current
becomes the display of the data value of the display address.
The adjustment keys corresponding to these display units allow for the selection of the
address to be modified and the adjustment of the selected parameter’s value.
The automatic graphy validation / invalidation key allows for the modified contents to be
put on memory. The blinking LED corresponding to this key reminds the user that he
must put into memory the value that he has just modified. Forgetting to press this key will
not make a change on the contents of EEPROM.
The EEPROM configuration contains 256 bytes accessable to the user.
Note: It is the operator’s responsiblility to correctly fill these EEPROMs. A faulty function
can be the result of a faulty programming of the memory.

Ref : 32-40-013-C
62
Generator GTI HF

2 Contents of eeprom 2 and 3 characteristic definition of tubes 1 & 2


The following list gives a résumé of the contents of the configuration EEPROMs. For a more
complete description see the text which follows it.
Note: There are two EEPROMs; one contains the charactristics of tube 1, the other the
characterisitics of tube 2.

Address Contents
0 MAXIMUM power permitted on the small focus for t=0.02s.
1 MAXIMUM power permitted on the small focus for t=0.05s.
2 MAXIMUM power permitted on the small focus for t=0.1s.
3 MAXIMUM power permitted on the small focus for t=0.2s.
4 MAXIMUM power permitted on the small focus for t=0.5s.
5 MAXIMUM power permitted on the small focus for t=1s.
6 MAXIMUM power permitted on the small focus for t=2s.
7 MAXIMUM power permitted on the small focus for t=5s.

8 MAXIMUM power permitted on the large focus for t=0.02s.


9 MAXIMUM power permitted on the large focus for t=0.05s.
10 MAXIMUM power permitted on the large focus for t=0.1s.
11 MAXIMUM power permitted on the largefocus for t=0.2s.
12 MAXIMUM power permitted on the large focus for t=0.5s.
13 MAXIMUM power permitted on the large focus for t=1s.
14 MAXIMUM power permitted on the large focus for t=2s.
15 MAXIMUM power permitted on the large focus for t=5s.

16 MAXIMUM power permitted at 40 kV on the small focus1


17 MAXIMUM power permitted at 50 kV on the small focus
18 MAXIMUM power permitted at 60 kV on the small focus
19 MAXIMUM power permitted at 70 kV on the small focus
20 MAXIMUM power permitted at 80 kV on the small focus
21 MAXIMUM power permitted at 900 kV on the small focus
22 MAXIMUM power permitted at 100 kV on the small focus
23 MAXIMUM power permitted at 110 kV on the small focus
24 MAXIMUM power permitted at 125 kV on the small focus

25 MAXIMUM power permitted at 40 kV on the large focus


26 MAXIMUM power permitted at 50 kV on the large focus
27 MAXIMUM power permitted at 60 kV on thelarge focus
28 MAXIMUM power permitted at 70 kV on the large focus
29 MAXIMUM power permitted at 80 kV on the large focus
30 MAXIMUM power permitted at 90 kV on the large focus
31 MAXIMUM power permitted at 100 kV on the large focus
32 MAXIMUM power permitted at 110 kV on the large focus
33 MAXIMUM power permitted at 125 kV on the large focus

Ref : 32-40-013-C
63
Chapter 2 / Technical manuel

Address Contents
34 heating current to obtain 10 mA at 40 kV
35 heating current to obtain 20 mA at 40 kV
36 heating current to obtain 30 mA at 40 kV
37 heating current to obtain 40 mA at 40 kV
38 heating current to obtain 50 mA at 40 kV
39 heating current to obtain 100 mA at 40 kV
40 heating current to obtain 150 mA at 40 kV
41 heating current to obtain 200mA at 40 kV
42 heating current to obtain 300mA at 40 kV
43 heating current to obtain 400mA at 40 kV
44 heating current to obtain 500mA at 40 kV
45 heating current to obtain 600mA at 40 kV
46 heating current to obtain 800mA at 40 kV
47 heating current to obtain 1000mA at 40 kV

48 heating current to obtain 10 mA at 80 kV


49 heating current to obtain 20 mA at 80 kV
50 heating current to obtain 30 mA at 80 kV
51 heating current to obtain 40 mA at 80 kV
52 heating current to obtain 50 mA at 80 kV
53 heating current to obtain 100 mA at80 kV
54 heating current to obtain 150 mA at 80 kV
55 heating current to obtain 200mA at 80 kV
56 heating current to obtain 300mA at 80 kV
57 heating current to obtain 400mA at 80 kV
58 heating current to obtain 500mA at 80 kV
59 heating current to obtain 600mA at 80 kV
60 heating current to obtain 800mA at 80 kV
61 heating current to obtain 1000mA at 80 kV

62 heating current to obtain 10 mA at 120 kV


63 heating current to obtain 20 mA at 120 kV
64 heating current to obtain 30 mA at 120 kV
65 heating current to obtain 40 mA at 120 kV
66 heating current to obtain 50 mA at 120 kV
67 heating current to obtain 100 mA at 120 kV
68 heating current to obtain 150 mA at 120 kV
69 heating current to obtain 200mA at 120 kV
70 heating current to obtain 300mA at 120 kV
71 heating current to obtain 400mA at 120 kV
72 heating current to obtain 500mA at 120 kV
73 heating current to obtain 600mA at 120 kV
74 heating current to obtain 800mA at 120 kV
75 heating current to obtain 1000mA at 120 kV

Ref : 32-40-013-C
64
Generator GTI HF

Address Contents
76 heating compensation at 10mA
77 heating compensation at 20mA
78 heating compensation at 30mA
79 heating compensation at 40mA
80 heating compensation at 50mA
81 heating compensation at 100mA
82 heating compensation at 150mA
83 heating compensation at 200mA
84 heating compensation at 300mA
85 heating compensation at 400mA
86 heating compensation at 500mA
87 heating compensation at 600mA
88 heating compensation at 800mA
89 heating compensation at 1000mA

*The power is determined by steps of 500 watts.

3 Record of the tube load characteristics


From the pages on the X-ray tube characteristics you should determine the numerical
values of the maximum power permited for each one of the two focuses in function with
the specific exposure times and the high-voltage values.

The power values are given in kW by units of 0.5 kW.


Determine the maximum power permited on the small focus for exposures during:
0.020 second
0.050 second
0.100 second
0.200 second
0.500 second
1.000 second
2.000 seconds
5.000 seconds
Enter these eight values in cases 0 to 7 of the tube’s EEPROM memory.

Ref : 32-40-013-C
65
Chapter 2 / Technical manuel

Determine the maximum power permited on the large focus for exposures during:
0.020 second
0.050 second
0.100 second
0.200 second
0.500 second
1.000 second
2.000 seconds
5.000 seconds
Enter these eight values in the cases 8 to 15 of the tube’s EEPROM memory.

Determine the maximum power permited on the small focus for the following high-voltage
values:
50 kV
60 kV
70 kV
80 kV
90 kV
100 kV
110 kV
125 kV
150 kV
Enter these nine values in cases 16 to 24 of the tube’s EEPROM memory.

Determine the maximum power permited on the large focus for the following high-voltage
values:
50 kV
60 kV
70 kV
80 kV
90 kV
100 kV
110 kV
125 kV
150 kV
Enter these nine values in cases 25 to 33 of the tube’s EEPROM memory.

Ref : 32-40-013-C
66
Generator GTI HF

4 Measurement of the tube emission characteristics


The management program of the generator allows for a listing of the tube’s emission
characteristics.
Beginning with address 34 the display unit normally used for the graphy mA, displays the
value of the heating current in mA.

The listing of these emission characteristics consists of finding the heating current value
supplied to each tube focus to obtain a high-voltage current at a certain high-voltage va-
lue.

The automatic listing program for the emission characteristics effectuates short exposures
of about 20 ms to 30 ms measuring the high-voltage current and then automatically sto-
res the results in the EEPROM memory.

The automatic system begins by displaying certain specific addresses from EEPROM:
•Address 34 permits a measure of the group of emission characteristics at 40 Kv
•Address 48 permits a measure of the group of emission characteristics at 80 Kv.
•Address 62 permits a measure of the group of emission characteristics at 120 Kv.

When you display an address permiting a listing of the emission characteristics, the display
unit normally used for the scopy kV, displays the kV value where it affects the measurement.

Note that if you interrupt the measuring phases during a group of tests, you must start the
phase again from the beginning. It is impossible to begin a group of tests in the middle.
When the program obtains an emission characteristic that it is seeking, it stores it on
memory. The operator is informed by a buzzer signal, and the program passes to the
following address. If the address that follows is the beginning of another phase of tests,
the system stops so you can mark a pause.

To start a group of measurements you must press the graphy second time. The graphy
first time is not taken into account, as it is automatically created by the measurement
managing program.

During an exposure, the program displays the value of the high-voltage current
measurement. The heat is raised each time up to approximately a maximum value of 5.5
ampéres. If the listing becomes impossible due to a limitation of the load characteristics
or if the maximum heating current is attained without having obtained the necessary high-
voltage current, the program sounds a buzzer and stores a special value in EEPROM
indicating the impossibility of the measurement.

Ref : 32-40-013-C
67
Chapter 2 / Technical manuel

A measurement impossibility is shown by the display unit as «- - - -» on the heating


current display. During a listing be sure to survey the temperature of the different elements
such as the high-voltage power supply and the tube housing. Also, be sure to observe a
pause of 10 minutes between each group of measurements.

If a problem arises during a measurement phase, it can sometimes be resolved simply by


switching the generator off and on.

After the completion of the measurement phase, it is advisable to verify the emission
characteristics measured by going back to the normal function of the generator and trying
several radiography exposures in several groups of constants which cover the range of
the generator and tube’s possibilities; for example:

40 kV 10 mA 0.050 second
40 kV 20 mA 0.050 second
40 kV 50 mA 0.050 second
40 kV 100 mA 0.050 second
40 kV 150 mA 0.050 second
80 kV 50 mA 0.050 second
80 kV 100 mA 0.050 second
80 kV 150 mA 0.050 second
120 kV 50 mA 0.050 second
120 kV 100 mA 0.050 second
120kV 150 mA 0.050 second
40 kV 600 mA 0.050 second
etc...

If you notice that certain values do not give the expected results, it would be necessary
either to re-do the corresponding group of tests, or manually correct the heating current
value. While certainly not indispensible, an oscilloscope can be connected on the test
point of the tank’s high voltage current measurement for verification.

5 Compensation of the heating current


Finally, you must adjust the compensation parameters of the heating current (parameters
76 to 89) so as to obtain a stable high-voltage current during an exposure.

Ref : 32-40-013-C
68
Generator GTI HF

6 Tube caracteristic list


It is advisable to record the contents of EEPROMs 2 and 3:
DATE:
TUBE N°:
TYPE OF TUBE:

Address Contents Value

0 MAXIMUM power permitted on the small focus for t=0.02s.


1 MAXIMUM power permitted on the small focus for t=0.05s.
2 MAXIMUM power permitted on the small focus for t=0.1s.
3 MAXIMUM power permitted on the small focus for t=0.2s.
4 MAXIMUM power permitted on the small focus for t=0.5s.
5 MAXIMUM power permitted on the small focus for t=1s.
6 MAXIMUM power permitted on the small focus for t=2s.
7 MAXIMUM power permitted on the small focus for t=5s.

8 MAXIMUM power permitted on the large focus for t=0.02s.


9 MAXIMUM power permitted on the large focus for t=0.05s.
10 MAXIMUM power permitted on the large focus for t=0.1s.
11 MAXIMUM power permitted on the largefocus for t=0.2s.
12 MAXIMUM power permitted on the large focus for t=0.5s.
13 MAXIMUM power permitted on the large focus for t=1s.
14 MAXIMUM power permitted on the large focus for t=2s.
15 MAXIMUM power permitted on the large focus for t=5s.

16 MAXIMUM power permitted at 40 kVon the small focus


17 MAXIMUM power permitted at 50 kVon the small focus
18 MAXIMUM power permitted at 60 kVon the small focus
19 MAXIMUM power permitted at 70 kVon the small focus
20 MAXIMUM power permitted at 80 kVon the small focus
21 MAXIMUM power permitted at 90 kVon the small focus
22 MAXIMUM power permitted at 100 kVon the small focus
23 MAXIMUM power permitted at 110 kVon the small focus
24 MAXIMUM power permitted at 125 kVon the small focus

25 MAXIMUM power permitted at 40 kVon the large focus


26 MAXIMUM power permitted at 50 kVon the large focus
27 MAXIMUM power permitted at 60 kVon thelarge focus
28 MAXIMUM power permitted at 70 kVon the large focus
29 MAXIMUM power permitted at 80 kVon the large focus
30 MAXIMUM power permitted at 90 kVon the large focus
31 MAXIMUM power permitted at 100 kVon the large focus
32 MAXIMUM power permitted at 110 kVon the large focus
33 MAXIMUM power permitted at 125 kVon the large focus

Ref : 32-40-013-C
69
Chapter 2 / Technical manuel

Address Contents Value

34 heating current to obtain 10 mA at 40 kV


35 heating current to obtain 20 mA at 40 kV
36 heating current to obtain 30 mA at 40 kV
37 heating current to obtain 40 mA at 40 kV
38 heating current to obtain 50 mA at 40 kV
39 heating current to obtain 100 mA at 40 kV
40 heating current to obtain 150 mA at 40 kV
41 heating current to obtain 200mA at 40 kV
42 heating current to obtain 300mA at 40 kV
43 heating current to obtain 400mA at 40 kV
44 heating current to obtain 500mA at 40 kV
45 heating current to obtain 600mA at 40 kV
46 heating current to obtain 800mA at 40 kV
47 heating current to obtain 1000mA at 40 kV

48 heating current to obtain 10 mA at 80 kV


49 heating current to obtain 20 mA at 80 kV
50 heating current to obtain 30 mA at 80 kV
51 heating current to obtain 40 mA at 80 kV
52 heating current to obtain 50 mA at 80 kV
53 heating current to obtain 100 mA at80 kV
54 heating current to obtain 150 mA at 80 kV
55 heating current to obtain 200mA at 80 kV
56 heating current to obtain 300mA at 80 kV
57 heating current to obtain 400mA at 80 kV
58 heating current to obtain 500mA at 80 kV
59 heating current to obtain 600mA at 80 kV
60 heating current to obtain 800mA at 80 kV
61 heating current to obtain 1000mA at 80 kV

Ref : 32-40-013-C
70
Generator GTI HF

Address Contents Value

62 heating current to obtain 10 mA at 120 kV


63 heating current to obtain 20 mA at 120 kV
64 heating current to obtain 30 mA at 120 kV
65 heating current to obtain 40 mA at 120 kV
66 heating current to obtain 50 mA at 120 kV
67 heating current to obtain 100 mA at 120 kV
68 heating current to obtain 150 mA at 120 kV
69 heating current to obtain 200mA at 120 kV
70 heating current to obtain 300mA at 120 kV
71 heating current to obtain 400mA at 120 kV
72 heating current to obtain 500mA at 120 kV
73 heating current to obtain 600mA at 120 kV
74 heating current to obtain 800mA at 120 kV
75 heating current to obtain 1000mA at 120 kV

76 heating compensation at 10mA


77 heating compensation at 20mA
78 heating compensation at 30mA
79 heating compensation at 40mA
80 heating compensation at 50mA
81 heating compensation at 100mA
82 heating compensation at 150mA
83 heating compensation at 200mA
84 heating compensation at 300mA
85 heating compensation at 400mA
86 heating compensation at 500mA
87 heating compensation at 600mA
88 heating compensation at 800mA
89 heating compensation at 1000mA

Ref : 32-40-013-C
71
Chapter 2 / Technical manuel

DATE:
TUBE N°:
TYPE OF TUBE:

Address Contents Value

0 MAXIMUM power permitted on the small focus for t=0.02s.


1 MAXIMUM power permitted on the small focus for t=0.05s.
2 MAXIMUM power permitted on the small focus for t=0.1s.
3 MAXIMUM power permitted on the small focus for t=0.2s.
4 MAXIMUM power permitted on the small focus for t=0.5s.
5 MAXIMUM power permitted on the small focus for t=1s.
6 MAXIMUM power permitted on the small focus for t=2s.
7 MAXIMUM power permitted on the small focus for t=5s.

8 MAXIMUM power permitted on the large focus for t=0.02s.


9 MAXIMUM power permitted on the large focus for t=0.05s.
10 MAXIMUM power permitted on the large focus for t=0.1s.
11 MAXIMUM power permitted on the largefocus for t=0.2s.
12 MAXIMUM power permitted on the large focus for t=0.5s.
13 MAXIMUM power permitted on the large focus for t=1s.
14 MAXIMUM power permitted on the large focus for t=2s.
15 MAXIMUM power permitted on the large focus for t=5s.

16 MAXIMUM power permitted at 40 kVon the small focus


17 MAXIMUM power permitted at 50 kVon the small focus
18 MAXIMUM power permitted at 60 kVon the small focus
19 MAXIMUM power permitted at 70 kVon the small focus
20 MAXIMUM power permitted at 80 kVon the small focus
21 MAXIMUM power permitted at 90 kVon the small focus
22 MAXIMUM power permitted at 100 kVon the small focus
23 MAXIMUM power permitted at 110 kVon the small focus
24 MAXIMUM power permitted at 125 kVon the small focus

25 MAXIMUM power permitted at 40 kVon the large focus


26 MAXIMUM power permitted at 50 kVon the large focus
27 MAXIMUM power permitted at 60 kVon thelarge focus
28 MAXIMUM power permitted at 70 kVon the large focus
29 MAXIMUM power permitted at 80 kVon the large focus
30 MAXIMUM power permitted at 90 kVon the large focus
31 MAXIMUM power permitted at 100 kVon the large focus
32 MAXIMUM power permitted at 110 kVon the large focus
33 MAXIMUM power permitted at 125 kVon the large focus

Ref : 32-40-013-C
72
Generator GTI HF

Address Contents Value

34 heating current to obtain 10 mA at 40 kV


35 heating current to obtain 20 mA at 40 kV
36 heating current to obtain 30 mA at 40 kV
37 heating current to obtain 40 mA at 40 kV
38 heating current to obtain 50 mA at 40 kV
39 heating current to obtain 100 mA at 40 kV
40 heating current to obtain 150 mA at 40 kV
41 heating current to obtain 200mA at 40 kV
42 heating current to obtain 300mA at 40 kV
43 heating current to obtain 400mA at 40 kV
44 heating current to obtain 500mA at 40 kV
45 heating current to obtain 600mA at 40 kV
46 heating current to obtain 800mA at 40 kV
47 heating current to obtain 1000mA at 40 kV

48 heating current to obtain 10 mA at 80 kV


49 heating current to obtain 20 mA at 80 kV
50 heating current to obtain 30 mA at 80 kV
51 heating current to obtain 40 mA at 80 kV
52 heating current to obtain 50 mA at 80 kV
53 heating current to obtain 100 mA at80 kV
54 heating current to obtain 150 mA at 80 kV
55 heating current to obtain 200mA at 80 kV
56 heating current to obtain 300mA at 80 kV
57 heating current to obtain 400mA at 80 kV
58 heating current to obtain 500mA at 80 kV
59 heating current to obtain 600mA at 80 kV
60 heating current to obtain 800mA at 80 kV
61 heating current to obtain 1000mA at 80 kV

Ref : 32-40-013-C
73
Chapter 2 / Technical manuel

Address Contents Value

62 heating current to obtain 10 mA at 120 kV


63 heating current to obtain 20 mA at 120 kV
64 heating current to obtain 30 mA at 120 kV
65 heating current to obtain 40 mA at 120 kV
66 heating current to obtain 50 mA at 120 kV
67 heating current to obtain 100 mA at 120 kV
68 heating current to obtain 150 mA at 120 kV
69 heating current to obtain 200mA at 120 kV
70 heating current to obtain 300mA at 120 kV
71 heating current to obtain 400mA at 120 kV
72 heating current to obtain 500mA at 120 kV
73 heating current to obtain 600mA at 120 kV
74 heating current to obtain 800mA at 120 kV
75 heating current to obtain 1000mA at 120 kV

76 heating compensation at 10mA


77 heating compensation at 20mA
78 heating compensation at 30mA
79 heating compensation at 40mA
80 heating compensation at 50mA
81 heating compensation at 100mA
82 heating compensation at 150mA
83 heating compensation at 200mA
84 heating compensation at 300mA
85 heating compensation at 400mA
86 heating compensation at 500mA
87 heating compensation at 600mA
88 heating compensation at 800mA
89 heating compensation at 1000mA

Ref : 32-40-013-C