Topcon SL-D7 Slitlamp - Service Manual
Topcon SL-D7 Slitlamp - Service Manual
2003
REVISED ............
SLIT LAMP
S L-D 7
TOPCON CORPORATION
R-SLD7-0-0308-11
CONTENTS
Page
1. INTRODUCTION ....................................................................................................................................1
1-1 How to Use This Repair Manual ...........................................................................................................1
1-2 Precautions for Repairs ..........................................................................................................................2
1-3 Nomenclature .........................................................................................................................................3
1-4 Optical Arrangement .............................................................................................................................7
1-5 Electric Circuit .......................................................................................................................................9
R-SLD7-0-0308-11
1. INTRODUCTION
1-1 How to Use This Repair Manual
(1) Follow the procedure described in the flowchart below.
Trouble
NO
Need to service?
YES
Repair
Check
NO Is operation
normal?
YES
Cleaning
Packing
Completed
1
R-SLD7-0-0308-11
1-2 Precautions for Repairs
(1) Disassemble, assemble, repair or adjust the optical system in the clean area where no dust or foreign mat-
ter will affect the instrument.
(2) For assembly and inspection, refer to the assembly and instruction manual of SL-D7.
(3) When repairing this instrument, use proper materials and tools correctly according to this manual.
(4) As a rule, a metallic part should be replaced as a single part and an optical part as a unit in this manual.
(5) Refer to the service parts list for the disassembly or assembly of the component not mentioned in this
manual.
(6) Don’t use other lubricating oils or adhesives except those specified.
(7) Order repair parts according to the service parts list.
2
R-SLD7-0-0308-11
1-3 Nomenclature
For Japan
3
R-SLD7-0-0308-11
1 Control lever
2 Chin rest base plate
3 Cross slide fixing knob
4 Rail cover
5 Cross slide
6 Table
7 Accessory drawer
8 Brightness control dial
9 Power switch
10 Pilot lamp
11 Up-and-down balance spring
12 Microscope arm fixing knob
13 Angle scale
14 Click stop roller
15 Illumination arm fixing knob
16 Arm cover
17 Chin rest up-and-down handle
18 Eyepiece
19 12.5× eyepiece (with click)
20 Magnification change handle
21 Eyepiece tightening knob
22 Barrier filter unit
23 Objective lens
24 Slit width adjusting handle
25 Inclination stop lever
26 Centering knob
27 Mirror
28 R/L lock switch
29 Aperture/slit change handle
30 Filter selection lever
31 Aperture/slit display window
32 Lamp house cover
33 Chin rest
34 Forehead rest
35 Diffusion lens
36 Connector for light adjustment (5P)
37 Connector for image taking-in operation (8P)
38 Color temperature conversion filter
4
R-SLD7-0-0308-11
For TMS, TE, TS and General
5
R-SLD7-0-0308-11
1 Control lever
2 Chin rest base plate
3 Cross slide fixing knob
4 Rail cover
5 Cross slide
6 Table
7 Color temperature conversion filter
8 Brightness control dial
9 Power switch
10 Pilot lamp
11 Up-and-down balance spring
12 Microscope arm fixing knob
13 Angle scale
14 Click stop roller
15 Illumination arm fixing knob
16 Arm cover
17 Chin rest up-and-down handle
18 Eyepiece
19 12.5× eyepiece (with click)
20 Magnification change handle
21 Eyepiece tightening knob
22 Barrier filter unit (depending on destination, TE)
23 Objective lens
24 Slit width adjusting handle
25 Inclination stop lever
26 Centering knob
27 Mirror
28 Fixation target
29 Aperture/slit change handle
30 Filter selection lever
31 Aperture/slit display window
32 Lamp house cover
33 Chin rest
34 Forehead rest
35 Diffusion lens
36 Connector for optical adjustment (5P)
37 Connector for image taking-in operation (8P)
6
R-SLD7-0-0308-11
1-4 Optical Arrangement
Filament position
Slit position
7
R-SLD7-0-0308-11
Names of optical units Functions
1 Condenser lens Adjusts the condensing position for the main bulb filament.
2 Filter Selects the blue, red-free, ND, heat absorbing, UV cut, IR cut or
exciter filter according to the clinical purpose.
3 Projection lens Projects and focuses the slit image.
4 Diffusion lens Widens the illumination field when photographing the anterior
eye section.
5 Reflection mirror Reflects the slit image to project it into the anterior eye section.
6 Microscope objective lens
7 Magnification drum lens
8 Eyepiece objective lens Used to focus the observed image in the eyepiece and to align the
optical axis.
9 Image erecting prism Changes the inverted image in the microscope to the erect one
and adjusts the rotation centering axis to prevent the image from
moving due to the pupillary distance shift.
10 Eyepiece Magnifies the observed image in the microscope and adjusts the
diopter to the patient's eye.
11 Color temperature conversion filter
12 Barrier filter By combining with the exciter filter installed to the illumination
system, observes and photographs the anterior eye section in flu-
orescent light.
8
R-SLD7-0-0308-11
SL-D7 For Japan
Chin rest unit
* RED is not connected among 7 core wires.
ORG : Orange
CABLE (Bind YEL, GRY and ORG and connect them.)
PUR : Purple
PIN : Pink CABLE (Bind PUR, PIN and LBL and connect them.)
BRN : Brown
BLU : Blue
1-5 Electric Circuit
J・S SW
RED AWG#26
LDE1 Halogen lamp CN2 1
ORG AWG20 CN2 2 SW4
P53 1 1 J1 P2 1 YEL AWG#26
Pilot LED PIN AWG20 CN2 3
P53 2 2 J1 P2 2 4
PUR AWG22 CN2
9
1 P52 3 J1 P2 3 5
R/L detection SW
PIN AWG22 RED AWG22 CN2
2 P52 P54 1 4 J1 P2 4 6 SW2
YEL AWG22 CN2
P54 2 5 J1 P2 5 7
PIN AWG22 CN2 1 COM
P54 3 6 J1 P2 6 8
GRY AWG#26
BRN AWG18 Stabilization power supply YEL/GRN CN2 2
3 P1
Volume AWG18 1 3
2 P1 ORG
BLU AWG18 SW3 4
1 P1 AWG#26
External fixation target E 5
2 3
6
RED AWG22 ORG
P51 1 1 J2 7
P51 2 2 J2 AWG#26 8
SW1 PIN AWG22 3 J2 GRY AWG#26 9
GRY Connector for image
P51 taking-in operation 10
BRN BLU AWG18
To IMAGEnet 11
2 P55
1 P55
AWG18 AWG18 12
E
SRRN134300
F1 F2 R/L LOCK SW
RED AWG22
1 J3
YEL AWG22
BRN BLU 2 J3
DC-1 Power supply 3 J3
AWG18 AWG18 GRY
AWG18 4 J3
(Output 12V/1A)
YEL/GRN
L G N AWG18
E
E
CORD1
R-SLD7-0-0308-11
SL-D7
For Export
J・S SW
RED AWG#26
10
RED CN2 5 1 COM
PIN AWG22 2 P52
AWG22 4 4
P54 1 J1 P2 CN2 6 SW2
YEL AWG22
P54 2 5 J1 P2 5 7 2 3
PIN AWG22 CN2 NO
P54 3 6 J1 P2 6 ORG AWG#26 NC
BRN AWG18 Stabilization power supply YEL/GRN CN2 8
3 P1
Volume AWG18
2 P1 Connector for image
BLU AWG18 1 P1 taking-in operation
External fixation target E
To IMAGEnet
RED AWG22
P51 1 1 J2 P3 1
SW1 P51 2 2 J2 P3 2
PIN AWG22 3 J2 P3 3
P51 GRY GRY
BLU AWG18 D2 Earphone code
BRN AWG28
2 P55
1 P55
AWG18 AWG18 YEL/GRN
E AWG18 R1 Earphone code LED2
GRY External fixation target LED
AWG28
F1 F2
RED AWG22
1 J3
YEL AWG22
BRN BLU 2 J3
DC-1 Power supply 3 J3
AWG18 AWG18 GRY
AWG18 4 J3
(Output 12V/1A)
YEL/GRN
L G N AWG18
E
E
CORD1
R-SLD7-0-0308-11
2. REPAIR GUIDE
2-1 Accuracy Check
How to adjust
No. Item How to check and allowance Tools and No.
and repair
1 Same focal point Focus at the maximum magnification and Focuse setting of Diopter telescope
due to magnifica- then decrease the magnification in turn, 25×, microscope 1
tion 16×, 10× and the minimum. Measure the ................ P. 12
diopter difference at each magnification.
2 Image shifts Place the standard scale on the object. Set the Accuracy setting Eyepiece with
when adjusting pupillary distance at the minimum and then at of pupillary dis- scale 2
the pupillary dis- the maximum and measure the maximum shift tance prism box Standard scale 3
tance. width of the optical axis. [within 0.05mm] ................ P. 15
3 The slit width Check if the frictional force of handle is too Adjustment of
narrows of itself. weak to obtain the desired position when slit width adjust-
changing the slit width. ing force
................ P. 19
4 When releasing Check if the slit illumination inclines too Adjustment of
the inclination quickly when removing the stopper. inclination unit
unit, it inclines rotary force
too quickly. ................ P. 16
5 Difference Measure the difference of the center in the slit Center setting of
between the open/close direction. aperture and slit
aperture and slit Aperture φ0.2: Within 0.03 image
image Aperture φ1, φ2: Within 0.1 ................ P. 16
6 The slit image Set the aperture to φ0.2. Rotate the slit verti- Slit image shift Eyepiece with
shifts due to the cally and then horizontally and measure the setting when scale 2
slit rotation. maximum shift distance of the slit image on rotating the slit Standard scale 3
the focusing rod. ................ P.17
7 The slit image is Even after setting the same focal point in the Focus setting of
out of focus. microscope, the slit image is out of focus. slit image
................ P.17
8 The slit image Rotate the illumination arm in 30° right and Slit image shift Eyepiece with
shifts when rotat- left each and measure the shifting distance of setting when scale 2
ing the illumina- the slit image. [Within 0.1mm] rotating the slit Standard scale 3
tion unit illumination unit
horizontally horizontally
(in 30° right and ................ P.18
left each).
9 The slit image Project the slit image on the focusing rod hor- Slit image shift Eyepiece with
shifts due to izontally and incline it. Measure the vertical setting when scale 2
inclination. shifting distance of the slit image. inclined Standard scale 3
[Within 0.05mm] ................ P.18
10 Trouble of click Check the click power by placing the small Adjustment of
power of small and large arms at stops on the same line. click power of
and large arms small and large
arms
................ P.19
11 Inclination of slit Project the slit image on the focusing rod and Adjustment of Eyepiece with
image measure its inclination with the eyepiece. inclination of slit scale 2
[Within 0.24mm at the slit length 14mm] image Focusing rod 4
................ P.19
* The width of a line provided on the eyepiece with scale is 0.02mm. When checking the accuracy, use this
line as a guideline.
11
R-SLD7-0-0308-11
2-2 Adjustment
* Before carrying out adjustment, if the screw lock has been applied to a screw, dissolve it using Isoamyl acetate.
Illustration Procedure
6S4×12S SUS
(3 pcs.)
Displayed Diopter
magnification × difference (dpt)
25 0.5
16 0.6
10 0.15
(Minimum) 0.35
12
R-SLD7-0-0308-11
Illustration Procedure
2. Focusing
1 When the microscope unit does not well focus and
the slit image is not clear
(with magnification drum unit at 16×):
1 Remove the eyepiece unit.
Place the diopter telescope and sight a target as
Diopter telescope
far as possible to check for diopter zero setting.
The “0” on the diopter adjusting ring should be
at the zero marker with 0.5 diopter allowance.
If not, correct it.
Eyepiece unit 2 Put into or put out the image lens frame (44670
1501) so that the finest focus is obtained at the
“0” diopter on the scale.
After adjustment, apply the screw lock.
3 Mount the eyepiece unit on the magnification
unit.
Turn the magnification unit clamp knob clock-
wise to couple the eyepiece unit with the magni-
2 Image lens frame fication unit, and then set the magnification
(44670 1501) drum at 25.6×. Focus each unit by moving the
magnification lens frame (44670 1016) forward
and backward.
4 Apply the screw lock after focusing.
Clamp knob
3 Magnification
lens frame Objective lens
CQ2×5S
Each 3 pcs.
left and right
Magnification unit
13
R-SLD7-0-0308-11
Illustration Procedure
6 CQ2×5S
Each 3 pcs. left and right
14
R-SLD7-0-0308-11
Illustration Procedure
Standard scale
5 6SV3×6S (3 pcs. each)
4
15
R-SLD7-0-0308-11
Illustration Procedure
Stop lever
[1.000~1.400g]
φ0.2
16
R-SLD7-0-0308-11
Illustration Procedure
Aperture φ0.2
Eyepiece
with scale
Illumination
objective lens
17
R-SLD7-0-0308-11
Illustration Procedure
Illumination arm
(Right eye)
[Movement: Within 0.05] U3×3S (2 pcs.)
18
R-SLD7-0-0308-11
Illustration Procedure
U4×4S
19
R-SLD7-0-0308-11
2-3 Repairing Order
Precautions for repairs
20
R-SLD7-0-0308-11
2-3-1 Disassembly and assembly of illumination unit
Illustration Procedure
Aperture/filter unit
3 4
6S3×6SUS
(4 pcs.)
1
CR2×6SUS
(2 pcs.)
3
U2.6×2.5S
(2 pcs.)
Cover
(44770 2601)
Aperture change (For TAC: 44675 2524)
handle
21
R-SLD7-0-0308-11
Important points for assembly
Illustration Points
Pin hole
3 Rod pin
Slit width
control link
2
Slit width control
link plate
CQ3×8S
Main blade
Wedge-shaped slit
22
R-SLD7-0-0308-11
Illustration Points
5 In case of the broken slit blade:
Marking line (a) Make a marking line before disassembly to
Marking line
set the slit blade into the original position.
(b) Remove the slit blade set screws.
(c) First, whet the main blade. (Refer to the
illustration on the previous page.)
* Place the slit blade on the slit holder tool perpen-
Slit blade set screw
Main blade dicularly and evenly and whet it uniformly and
horizontally. After whetting, remove burrs. When
removing burrs, be careful not to damage the slit
edge and scoop only the burred part lightly by
hand. After whetting, wipe off the machine oil
with ether and apply it thinly to the whetstone
Slit blade Slit holder tool again.
White whetstone
U2.6×2.5S
(2 pcs.)
23
R-SLD7-0-0308-11
2-3-2 Disassembly and assembly of power supply unit
Illustration Procedure
Cord retainer
Clamp
Screw D
Screw E
Plug P1
24
R-SLD7-0-0308-11
2-3-3 Trouble shooting (Main body)
Reference
No. Trouble Cause Remedy Tool
page
1 The slit does not Dust is stuck to the Wipe off dust with • Eyepiece with P.22, 23
close. slit. ether many times. scale
The slit is wedge- Adjust the slit blade. • Eyepiece with P.22, 23
shaped. scale
2 There is a differ- The accuracy is not Loosen the slit box • Standard scale P.16
ence between the good in the incorpo- tightening screws (4 • Eyepiece with
aperture of φ0.2 and ration of the slit unit pcs.) to position scale
the slit center. and illumination them accurately.
unit.
3 Difference between The reflection mir- Position the reflec- • Standard scale P.17, 18
the slit center and ror in the illumina- tion mirror in the • Eyepiece with
microscope center. tion unit is not illumination unit scale
positioned correctly. accurately.
25
R-SLD7-0-0308-11
2-3-4 Trouble shooting (Power supply unit)
Does the
pilot lamp 2 light NO
when the power switch The pilot lamp does not light. Inspection 1
1 is ON?
Does the NO
halogen lamp light? The halogen lamp does not light. Inspection 2
YES
Does the
light intensity of
the halogen lamp change NO The light intensity of the halogen
continuously by using the lamp does not change continuously. Inspection 3
light adjusting
volume 30 ?
YES
Is the
brightness of the halogen NO
The halogen lamp is dark. Inspection 4
lamp normal?
YES
YES
NO
Is R/L detected? R/L is not detected. Inspection 6
YES
Normal
26
R-SLD7-0-0308-11
Inspection 1 Inspection Procedure Inspection Details
NO
Is the
power switch turned NO The power switch is faulty.
ON/OFF? Replace 1 .
YES
Is the
output voltage of the NO The SW power supply is faulty.
SW power supply normal Replace the power supply unit.
(5.1V/15V)?
YES
NO
Is the
connector for light
adjustment (5P) or the NO
Connect it correctly.
illumination connector
set correctly?
27
R-SLD7-0-0308-11
Inspection 2 Inspection Procedure Inspection Details
NO
Wiring check (2)
Does the NO
light adjusting volume The light adjusting volume is
change? faulty. Replace 30 .
YES
Wiring check (3)
28
R-SLD7-0-0308-11
Inspection 4 Inspection Procedure Inspection Details
Is the
inside of the YES
halogen lamp soiled with The halogen lamp is faulty.
white or black Replace 28 .
materials?
NO
Is the
output voltage of the NO The SW power supply is
SW power supply normal faulty. Replace 3 .
(5.1V/15V)?
YES
Is the
resistance of the light NO The light adjusting volume is
adjusting volume normal faulty. Replace 30 .
(0~30kΩ)?
YES
Wiring check (4)
Is the YES
external fixation LED The external fixation LED is
broken? faulty. Replace 26 .
NO
NO
Wiring check (5)
29
R-SLD7-0-0308-11
Inspection 6 Inspection Procedure Inspection Details
Is the R/L NO
lock switch set to “A” Set it to “A” (auto).
(auto)?
YES
Does the
microswitch operate NO Adjust the height of the
(sound “click”)? microswitch.
YES
Wiring check (6)
30
R-SLD7-0-0308-11
2-3-5 Wiring check
This section describes the special checks that are required when the problem is diagnosed as “Wiring check”
in the frowcharts of “2-3-4 Trouble shooting”. The number in parentheses for the wiring check in “2-3-4 Trou-
ble shooting” corresponds to “No.” in the following table.
Use the tester to check between “From” and “To” in “Area for checking”.
If the wire passes through a part on the way, such a part is shown in “Via”.
Check it at the same time. Refer to “Electric parts arrangement”.
31
R-SLD7-0-0308-11
2-3-6 Electric parts arrangement
J1
J3
J2
32
R-SLD7-0-0308-11
(2) Halogen lamp cord
33
R-SLD7-0-0308-11
(3) Halogen lamp and socket
34
R-SLD7-0-0308-11
(4) External fixation target
CN2 CN1
35
R-SLD7-0-0308-11
2-3-7 List of electric part numbers
36
R-SLD7-0-0308-11
2-4 List of Repair Tools
2-4-1 List of special repair tools
Tool Function/
Name Illustration Application Remarks
No. accuracy
1 Diopter Observation • Setting the same focus by
telescope magnification magnification
(3.4×)
37
R-SLD7-0-0308-11
TOPCON MEDICAL SYSTEMS, INC