Application note 57
Most Megger test equipment is supplied complete with a certificates have a box within which the instrument
full 12-month calibration certificate included in the box. commissioning date may be entered. This date should be
evidenced by a copy of the official sales invoice for that
After carrying out suitable risk assessments you may instrument.
determine that the 12-month period should run from the
date of purchase, rather than the date the instrument was Re-calibration will then be required within 12 months of
originally tested. To facilitate this, Megger calibration this commissioning date.
There are test boxes available such as the MTB7671/2 that Megger oil tester calibration can also be confirmed using the
can check the calibration of low voltage installation test digital voltage checkers, VCM80D for voltages up to 80 kV
equipment by testing them against known values. These are and VCM100D for voltages up to 100 kV.
not a calibration method but by checking the tester at more
frequent intervals than the annual calibration, a trend can be Calibration shunts are available for checking calibration on
spotted to alert the user to changes in the instrument that low resistance ohmmeters. These four terminal calibrating
require investigation. resistance standards feature a conservative current rating to
ensure against overheating. Each calibration shunt provides
Some scheme providers request evidence of test two current terminals with knurled Allen-head cap screws
equipment’s accuracy and consistency so by recording and two potential terminals with slotted screw heads.
results when using a check box is the easiest way of
evidencing this. It is important to have the calibration checked on the
calibration checking equipment to ensure continuing
It is not only low voltage installation testers that can be accuracy and traceability to national standards. This is
checked, Megger manufacture a CB101 calibration check usually performed every 12 months.
box for use with up to 5kV insulation testers to check
accuracy. Four resistors are connected in the unit to enable
a wide range of values to be checked.
Calibration certificates
Megger as a manufacturing company provides three types of laboratory on recalibration of a tester or on calibrating an
calibration certificates as can be seen below. The first instrument that is supplied without a certificate and the
certificate is an example of the certificate that is supplied third is an example of a UKAS (United Kingdom
with many of the Megger products, the second is an Accreditation Services) certificate, which meets the
example of a certificate issued by Megger’s calibration requirements of ISO17025.
Understanding calibration certificates
Application note 57
Calibration
equipment
information
Date of purchase
should be written
here and evidenced
by a copy of the
invoice
Amount of
variance This column
acceptable says if the
from instrument is
applied within the
value tolerances
Range
under test This column
states the
accuracy of
the equipment
used for
calibration
Understanding calibration certificates
Application note 57
Page 1 of 2
Date of
calibration
Details of
calibrated
instrument
Details of
calibration
equipment
used
Understanding calibration certificates
Application note 57
Page 2 of 2
Range
under test
Page 1 of3
UKAS
certificate
number
Date of UKAS
calibration accreditation
number
Identifies
Calibrated calibration
instrument equipment
details used
Calibration
environment
Understanding calibration certificates
Application note 57
Page 2 of3
Range
under test
Page 3 of3