Professional Documents
Culture Documents
1.1. GENERAL :
10.3.1. Each in house calibrated instrument that has Internal Calibration Certificate
(Exhibit 10-02) is maintained by QC Inspector in the calibration file.
10.3.2. All equipment is attached by a Calibration Label (Exhibit 10-03) which records
the serial No., Validity of calibration date.
10.3.3. QC Manager shall also verify records of calibration of subcontractors (e.g.
heat treatment recorder, thermocouples, universal and impact test
machines). Record shall be maintained on file.
SECTION : 10
QUALITY CONTROL MANUAL Page : 2 of 2
Rev. : 0
EDITION NO.1
Date : 17/09/2023
CALIBRATION OF MEASUREMENT AND TEST EQUIPMENT
10.4.1. When instruments are found out of calibration or damaged, the QC Inspector
tags the item with a “NOT CALIBRATED” label, removes it from the work
area and arranges for calibration or replacement.
10.4.2. If a Code item is checked using discrepant instruments, a Nonconformance
Report shall be issued and dealt with as described in Sec. 6 of this Manual.
Traceability of the instruments used is established from the inspection
reports / test reports.
10.4.3. The identification of measuring equipment in fulfillment of Code requirements
shall be entered in relevant reports for traceability.
10.4.4. Care shall be taken during critical measurements such as use of multiple
pressure gauges during pressure test and use of instruments with higher
accuracy for critical dimensions to avoid possibility of error in
measurements.