13.1 All examination, measurement and test equipment used to fulfill applicable
Code requirements shall be calibrated in accordance with ‘ASY Examination,
Measurement and Test Equipment Calibration Procedure QA 18.
13.6 Status of Calibrated items shall be identified through the use of a "Calibration
Decal" (Exhibit QA 24) attached to the item or to its container if the item is too
small for the decal to be attached. The Calibration Decal shall list the date
calibrated, the date calibration is due, the agency/person that performed the
calibration and the equipment identification number.
13.8 Pressure gauges used for hydrostatic, pneumatic or proof testing are
calibrated against dead weight tester or calibrated master gauge.
13.9 Impact test temperature measuring instruments shall be calibrated every six
months. Impact testing machines shall be calibrated every 12 months.
13.11 Baking / Holding ovens shall be calibrated for proper working condition of
temperature indicators.
13.12 Hand quivers shall be verified for its function and temperature (65°C
minimum).
13.13 Welding machines shall be verified for current and voltage display accuracy
using calibrated gauges.