Professional Documents
Culture Documents
905–910, 2004
Copyright © 2004 Elsevier Inc.
Printed in the USA. All rights reserved
0360-3016/04/$–see front matter
doi:10.1016/j.ijrobp.2004.02.030
PHYSICS CONTRIBUTION
DAVID S. FOLLOWILL, PH.D., DENISE S. DAVIS, M.S., AND GEOFFREY S. IBBOTT, PH.D.
Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, TX
Purpose: To determine the relationships between electron beam depth dose characteristics, depth of maximum
dose (dmax), depth of 80% dose (d80), and depth of 50% dose (d50), and the nominal energy designation of electron
beams from multiple linear accelerators for the purpose of electron beam treatment planning and quality
assurance.
Methods: The Radiological Physics Center Staff, during its on-site dosimetry review visits to institutions
participating in clinical trials, measured depth dose characteristics for more than 2000 electron beams. Mea-
surements were performed on Varian, Siemens, and Elekta/Philips accelerators generating beams with nominal
energy values ranging from 4 –22 MeV. The depth dose data were determined at the nominal source-to-skin
distance with the reference cone size in accordance with recommendation of the American Association of
Physicists in Medicine Task Group 25 report.
Results: The important depth dose characteristics dmax, d80, and d50 varied in a predictable fashion when plotted
against the true beam quality indicator, R50. However, d80 and d50 values overlapped considerably when plotted
against the manufacturers’ nominal electron energy values. For a specific nominal electron energy value, the
values of dmax, d80, and d50 varied by as little as 3 mm for low energy levels to nearly as much as 20 mm for high
energies.
Conclusions: The manufacturer’s nominal electron energy value does not adequately describe the depth dose
characteristics of an electron beam for treatment planning purposes. Clinicians and physicists should determine
and use only the specific depth dose data for their clinical beams and not the manufacturer’s nominal value.
© 2004 Elsevier Inc.
Reprint requests to: David S. Followill, Ph.D., The University of Health and Human Services.
Texas M. D. Anderson Cancer Center, Department of Radiation Acknowledgments—The authors thank the many physicists of the
Physics, Unit 547, 1515 Holcombe Boulevard, Houston, TX Radiological Physics Center for assisting in gathering the data as
77030.Tel:(713)745-8989;Fax:(713)794-1364;E-mail:dfollowi@ part of the Center’s quality audit program. The authors further
mdanderson.org acknowledge Morris Tatcher, Alexander Nevelsky, and Raquel
Presented in part at the 37th Annual Meeting of the American Bar-Deroma for useful discussions regarding these data.
Association of Physicists, 1995, in Boston, MA. Received Aug 22, 2003, and in revised form Feb 13, 2004.
This work was supported by Public Health Service Grant no. CA Accepted for publication Feb 16, 2004.
10953, awarded by the National Cancer Institute, Department of
905
906 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004
Table 1. Average depth dose characteristics for electron beams from Varian accelerators
and among models of accelerator from the same manufac- less than 10 cm ⫻ 10 cm. The depth dose curves, however,
turer, depending on the design of the treatment head. The do not differ significantly for field sizes 10 cm ⫻ 10 cm or
manufacturers’ displayed nominal electron energy values larger.
are generally defined in terms of the most probable (modal) We used measured data obtained in more than 20 years of
energy of the discrete electron energy spectrum before it on-site dosimetry review visits to establish the relationship
leaves the vacuum window of the accelerating structure between the manufacturer’s nominal electron energy and
within the head of the accelerator (4). Over the years, the actual beam energy and dose distribution. We report mea-
manufacturers of accelerators have used different methods sured depth dose data for electron beams from three man-
to describe electron beam energies (5). Because of variabil- ufacturers: Elekta/Philips, Siemens, and Varian. The data
ity in the manufacturers’ nominal electron energy specifi- presented in this report provide clinical physicists a useful
cations, the nominal electron energy value is not meaningful quality assurance tool to provide redundant checks of their
for physics or dosimetry purposes. The AAPM calibration electron beam dosimetry parameters. In addition, clinicians
protocol (6), an AAPM electron dosimetry report (7), and an will benefit from the data by better understanding the spe-
independent publication (8) all state that electron depth dose cific dosimetry characteristics associated with the manufac-
values best define the actual electron energy of the spectrum turers’ nominal electron energy values. Electron beam ther-
reaching the patient. In particular, the d50 along the central apy planning should be based on the actual beam energy and
axis adequately characterizes the electron beam quality for dose distribution, not on the manufacturer’s nominal value.
calibrations and dosimetry calculations (6, 7, 9).
In contrast, clinicians do not have a standard method of
METHODS AND MATERIAL
specifying therapeutic electron beam energy. Traditionally,
clinicians are instructed that the prescription depth (in cm) Beam characteristics of accelerators installed at institu-
is equal to either one fourth or one third of the manufactur- tions participating in cooperative clinical trials were mea-
er’s nominal energy (MeV), corresponding to the approxi- sured by the staff of the RPC. The RPC is responsible for
mate depths of the 90% or 80% isodose lines, respectively. providing an independent quality assurance audit program
Clinicians are also taught that the approximate range of the to institutions entering patients into clinical trials. The qual-
electrons (cm) is equal to one half of the nominal electron ity assurance audit program includes on-site dosimetry re-
energy value (MeV), and this value is often used as a guide view visits to many of the participating institutions by RPC
for protecting sensitive underlying, normal tissues, such as physicists. During these visits, the RPC staff perform a
lung (10). These rules of thumb can be misleading, because series of dosimetry measurements to ensure that the dosim-
the manufacturer’s nominal electron energy does not fully etry data used to calculate beam-on time for patients are
describe the dosimetry characteristics of the specific elec- accurate. As a part of the measurements, each electron beam
tron beam. The depths of 80% and 90% dose depend very output is calibrated, and the depth dose curve is verified at
strongly on the electron field size, in particular for field sizes dmax, d80, and d50.
Comparison of electron beam characteristics ● D. FOLLOWILL et al. 907
Table 2. Average depth dose characteristics for electron beams from Siemens accelerators
Measurements were made using electron beams from were taken with a cone size of 15 cm ⫻ 15 cm. The SSD
Varian accelerators (Clinac models 12, 18, 20, 1800, 2100 was 100 cm for nearly all (99%) of the measurements. An
series, 2300 series, and 2500), Siemens accelerators (mod- SSD of 95 cm was used for a few of the Elekta/Philips
els 12, 20, 67, 74, 77, MD, MD2, KD, MXE, and Primus), machines. All electron beam measurements incorporated
and Elekta/Philips accelerators (models SL75, SL18, SL20, a shift of 0.5rcav to the effective point of measurement.
SL25, and Precise). The data consist of 2127 electron beams Values of dmax were measured in increments of 1 mm or
ranging in energy from 4 –22 MeV. 2 mm for electron beam energies less than 10 MeV. For
The RPC measurements were made in a water phantom electron energies greater than or equal to 10 MeV, the
that provided adequate lateral scatter and backscatter institution’s value of dmax was used because a flat plateau
using a Farmer-type 0.6 cm3 ion chamber- either NE region around dmax makes the precise location less crit-
model 2571 (NE Technology, Ltd., Berkshire, England) ical for these higher energies. Exact values of d80 and d50
or PTW model N23333 (PTW, Freiburg, Germany), and were interpolated from shallower (⬎80%) and deeper
read with a Keithley model 602 electrometer (Keithley (⬍50%) measurements.
Instruments, Inc., Cleveland, OH). All of the depth dose The average d80 values should aid clinicians in determin-
values were measured at nominal SSD and used a refer- ing the effective therapeutic depth for different electron
ence cone size in accordance with the AAPM TG-25 energies while the average d50 values are important to
report (7). The reference cone size was 10 cm ⫻ 10 cm physicists in determining the beam quality of each electron
for 93% of the measurements while the remaining data beam for calibration purposes.
908 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004
Table 3. Average depth dose characteristics for electron beams from Elekta/Philips accelerators
Fig. 1. Electron beam depth dose characteristics of dmax, d80, and d50 depths plotted against electron beam quality, R50
for Varian, Siemens, and Elekta/Philips accelerators. The function fit and its R2 value are listed.
value is used frequently to determine the prescription depth, cists during their on-site dosimetry review visits as a redun-
meaning that the choice of energy is based on agreement dant check of the electron depth dose ion chamber measure-
between the greatest depth of the target volume and the depth ments. For clinical purposes, specification of prescription
of the 80% isodose line. Our data, however, show that this depth should be based on actual measured data, not on rules
agreement does not hold. Indeed, we show that the use of of thumb or other estimators. Reliance on nominal electron
estimators or rules of thumb may lead to significant errors in energies and rules of thumb can be very misleading, as
dose prescription. In the example above, the rule of thumb shown in Fig. 2.
would indicate the prescription depth for a 12 MeV beam Physicists should find the data in Fig. 1 and Tables 1–3
should be 12/3, or 4 cm. Measurements showed that the actual useful as a quality assurance tool for the verification of
depth of the 80% isodose line varied from 0.5 cm shallower to measured electron depth dose data. Once electron depth
0.8 cm deeper than the estimate. dose curves are measured, institutional data can be com-
In addition, the data in Tables 1 and 2 show that the depth pared to the data presented here as a redundant check to
of the 50% isodose line for a 12 MeV beam was as little as 4.2 identify potential errors in the dosimetry data.
cm. This suggests that the use of a rule of thumb may result in
the selection of an electron beam having a dose at the prescrip-
tion depth of only slightly more than 50%. In fact, because of CONCLUSION
the wide range of d80 and d50 values for the 12 MeV beams, 1
linac’s d80 is another’s d50. This observation is true for the The manufacturer’s nominal energy is not an appropriate
Varian, Siemens, and Elekta/Philips accelerators. electron beam indicator for clinical radiotherapy. Beams of the
However, the data in Fig. 1 demonstrate a reproducible same nominal electron energy from different accelerators may
and linear relationship between R50, the TG-51 electron exhibit significant differences in electron beam penetration,
beam quality specification, and the d80 and d50 values. The even for the reference 10 cm ⫻ 10 cm field size.
linear relationship between R50 and d50 is used by the Radiation oncologists and physicists should be familiar with
TG-51 calibration protocol for the calibration of electron the specific depth dose characteristics for each electron beam
beams. This linear relationship is also used by RPC physi- used in their clinic. Clinical decisions as to which electron
910 I. J. Radiation Oncology ● Biology ● Physics Volume 59, Number 3, 2004
Fig. 2. The range of electron beam depth dose characteristics of dmax, d80, and d50 depths plotted against manufacturer’s
nominal electron energy value for Varian, Siemens, and Elekta/Philips accelerators.
beam is best used for an electron treatment should be made the data presented in this work are for a single field size and do
based on the actual measured depth dose data rather than rules not contain the effects of smaller field sizes, beam obliquity, or
of thumb. Clinicians are cautioned when using these results, as increased treatment distances.
REFERENCES
1. Kirby TH, Gastorf RJ, Hanson WF, et al. Electron beam 8. Brahme A, Svensson H. Specification of electron beam quality
central axis depth dose measurements. Med Phys 1985;12: from central-axis depth absorbed-dose distribution. Med Phys
357–361. 1976;3:95–102.
2. American Association of Physicists in Medicine, RTC Task 9. International Atomic Energy Agency. Absorbed dose deter-
Group 21. A protocol for the determination of absorbed dose mination in external beam radiotherapy. An international code
from high-energy photon and electron beams. Med Phys 1983; of practice for dosimetry based on standards of absorbed dose
10:741–771. to water. Technical report series no. 398. Vienna: International
3. Davis DS, Followill DS, Kennedy P, Hanson WF. Electron Atomic Energy Agency, 2000.
percent depth dose and cone ratio data from various machines. 10. Vaeth JM, Meurk ML. Electron beam therapy in clinical
Med Phys 1995;22:1007.
practice. In: Leibel SA, Phillips TL, editors. Textbook of
4. Klevenhagen SC. Physics and dosimetry of therapy electron
radiation oncology, 1st ed. Philadelphia: WB Saunders, 1998:
beams. Madison, WI: Medical Physics Publishing, 1993.
5. International Electrotechnical Commission (IEC). Interna- p. 216.
tional standard—medical electrical equipment. Geneva: IEC 11. Kennedy PM, Hanson WF. A review of high-energy photon
60601-2-1; 1998. beam characteristics measured by the Radiological Physics
6. Almond PR, Biggs PJ, Coursey BM, et al. AAPM’s TG-51 Center. Med Phys 1992;19:838.
protocol for clinical reference dosimetry of high-energy pho- 12. International Commission on Radiation Units and Measure-
ton and electron beams. Med Phys 1999;26:1847–1870. ments. Radiation dosimetry: Electron beams with energies
7. Khan FM, Doppke KP, Hogstrom KR, et al. Clinical electron- between 1 and 50 MeV. ICRU report 35. Bethesda, MD:
beam dosimetry: Report of AAPM Radiation Therapy Com- International Commission on Radiation Units and Measure-
mittee Task Group 25. Med Phys 1991;18:73–109. ment, 1984.