00809-0100-3166
Rev. AD
March 2020
Rosemount™ 1066
Smart-Enabled, 2-Wire Transmitter
Essential Instructions
Read this page before proceeding
Emerson designs, manufactures, and tests its Rosemount products to meet many national and
international standards. Because these instruments are sophisticated technical products, you must
properly install, use, and maintain them to ensure they continue to operate within their normal
specifications. The following instructions must be adhered to and integrated into your safety
program when installing, using, and maintaining Rosemount products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of life; personal injury;
property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Reference
Manual is not the correct manual, telephone 1-800-854-8257 and the requested manual will
be provided. Save this Reference Manual for future reference.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Reference
Manual and per applicable local and national codes. Connect all products to the proper
electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product’s
performance and place the safe operation of your process at risk. Look alike substitutions may
result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal
injury.
NOTICE
If a 475 universal haRT® Communicator is used with these transmitters, the software within the 475 may
require modification. If a software modification is required, please contact your local Emerson Service Group
or National Response Center at 1-800-654-7768.
Essential Instructions I
3. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm.
Care must be taken to avoid electrostatic charge build-up. The Model 1066 Transmitter must
not be rubbed or cleaned with solvents or a dry cloth.
4. Special Condition of Use of 1066 C FF/FII5 and 1066T FF/FII5. For use with simple apparatus
model series 140, 141, 142, 150, 400, 401, 402, 402VP, 403, 403VP, 404, and 410VP contact-
ing conductivity sensors and model series 222, 225, 226, 228 toroidal sensors.
WaRNING
physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of
end users’ equipment. This could be intentional or unintentional and needs to be protected
against.
Physical security is an important part of any security program and fundamental to protecting
your system. Restrict physical access by unauthorized personnel to protect end users’ assets.
This is true for all systems used within the facility.
II
Reference Manual Table of Contents
00809-0100-3166 March 2020
Contents
Section 1: Quick Start Guide
1.1 Quick start guide ..........................................................................................................1
Section 3: Installation
3.1 Unpacking and Inspection..........................................................................................13
3.2 Installation – General Information .............................................................................13
3.3 Preparing Conduit Openings......................................................................................13
Section 4: Wiring
4.1 General ...................................................................................................................... 17
4.1.1 General Information......................................................................................17
4.1.2 Digital Communication.................................................................................17
4.2 Power Supply/Current Loop – 1066 HT......................................................................17
IV Table of Contents
Reference Manual Table of Contents
00809-0100-3166 March 2020
8.6.1.3 Filter................................................................................................54
8.6.1.4 Free Chlorine pH Correction ...........................................................54
8.6.1.5 Manual pH Correction ....................................................................54
8.6.1.6 Resolution ......................................................................................54
8.6.2 Total Chlorine Measurement Programming .................................................55
8.6.2.1 Description .....................................................................................55
8.6.2.2 Measure..........................................................................................55
8.6.2.3 Units ...............................................................................................55
8.6.2.4 Filter................................................................................................55
8.6.2.5 Resolution ......................................................................................55
8.6.3 Monochloramine Measurement Programming............................................56
8.6.3.1 Measure: Monochloramine ............................................................56
8.6.3.2 Units ...............................................................................................56
8.6.3.3 Filter................................................................................................57
8.6.3.4 Resolution ......................................................................................57
8.7 Oxygen Measurement Programming ........................................................................57
8.7.1 Oxygen Measurement Application.................................................58
8.7.2 Units ...............................................................................................58
8.7.3 Partial Press ....................................................................................58
8.7.4 Salinity ............................................................................................58
8.7.5 Filter................................................................................................58
8.7.6 Pressure Units.................................................................................58
8.8 Ozone Measurement Programming ..........................................................................59
8.8.1 Units ...............................................................................................59
8.8.2 Filter................................................................................................59
8.8.3 Resolution ......................................................................................59
Section 9: Calibration
9.1 Introduction ..............................................................................................................67
9.2 Calibration..................................................................................................................67
9.2.1 Auto Calibration .........................................................................................................68
9.2.2 Manual Calibration – pH................................................................................68
9.2.3 Entering a Known Slope Value – pH ..............................................................68
9.2.4 Standardization – pH ....................................................................................69
9.2.5 SMART sensor auto calibration upload – pH .................................................69
9.3 ORP and Redox Calibration ........................................................................................70
9.4 Contacting Conductivity Calibration..........................................................................71
9.4.1 Entering the Cell Constant ............................................................................72
9.4.2 Zeroing the Instrument.................................................................................72
9.4.3 Calibrating the Sensor in a Conductivity Standard (in process cal)................72
9.4.4 Calibrating the Sensor To A Laboratory Instrument (meter cal)....................73
9.4.5 Cal Factor ......................................................................................................73
VI Table of Contents
Reference Manual Table of Contents
00809-0100-3166 March 2020
VIII
Reference Manual Section 1: Quick Start Guide
00809-0100-3166 March 2020
Specifications 3
Section 2: Description and Specifications Reference Manual
March 2020 00809-0100-3166
Non-Incendive:
Class I, Division 2 Groups A-D
Class I, Div. 2, Groups A-D* Dust Ignition proof Class II & III, Div 1, Groups EFG
Dust Ignition Proof Class II & III, Div 1, Groups EFG Class II & III, Division 1, Groups E-G
Class II & III, Div. 1, Groups E-G Tamb = -20°C to 65°C, IP66 enclosure
For Non-Incendive Field Wiring Installation, see drawing 1400670
Type 4/4X Enclosure
T4 Tamb = -20 °C to 65 °C
For Non-Incendive Field Wiring Installation, see drawing 1400669
*Additionally approved as a system with models 140,141,142, 150, 400, 400VP, 401, 402, 402VP, 403,403VP, 404 & 410VP contacting conductivity
sensors and models 222, 225, 226 & 228 inductive conductivity sensors.
4 Specifications
Reference Manual Section 2: Description and Specifications
00809-0100-3166 March 2020
1500
1364
1250 ohm s
1000
Load,ohm s
750
w ith H ART 545
com m unication ohm s
500
250
w ithoutH ART
com m unication
0
12 18 24 30 36 42
Pow ersupply voltage,Vdc
H ART option
analog Outputs: Two-wire loop powered (Output 1 only). Two 4-20 mA electrically isolated cur-
rent outputs (Output 2 must be externally powered). Superimposed HART digital signal on
Output 1. Fully scalable over the operating range of the sensor.
Weight/Shipping Weight: 2 lbs/3 lbs (1 kg/1.5 kg)
Specifications 5
Section 2: Description and Specifications Reference Manual
March 2020 00809-0100-3166
FIGuRE 2-2. General purpose and high performance ph sensors 3900, 396pvp
and 3300hT
6 Specifications
Reference Manual Section 2: Description and Specifications
00809-0100-3166 March 2020
conductivity sensors
Temperature specifications:
Temperature Accuracy,
± 0.1 °C
Pt-1000, 0-50 °C
Temperature Accuracy,
± 0.5 °C
Pt-1000, Temp. > 50 °C
0.01
0.01µS/cm to 200µS/cm 200µS/cm to 2000µS/cm
1 µS/cm to 20mS/cm
1.0
20mS/cm to 200mS/cm
4-electrode 2µS/cm to
1400mS/cm
Specifications 7
Section 2: Description and Specifications Reference Manual
March 2020 00809-0100-3166
Temperature specifications:
Temperature Accuracy,
± 0.5 °C
Pt-100, -25 to 50 °C
Temperature Accuracy,
± 1 °C
Pt-100, 50 to 210 °C
226
50µS/cm to 500mS/cm 500mS/cm to 2000mS/cm
225 & 228
50µS/cm to 1500mS/cm 1500mS/cm to 2000mS/cm
242
100µS/cm to 2000mS/cm
222
(1in & 2in) 500µS/cm to 2000mS/cm
8 Specifications
Reference Manual Section 2: Description and Specifications
00809-0100-3166 March 2020
2.6.4 Monochloramine
The 1066 is compatible with the 499A CL-03 Monochloramine sensor. The 1066 fully compensates
readings for changes in membrane permeability caused by temperature changes. Because
monochloramine measurement is not affected by pH of the process, no pH sensor or correction is
required. For more information concerning the use and operation of the amperometric chlorine
sensors, refer to the product data sheets.
Specifications 9
Section 2: Description and Specifications Reference Manual
March 2020 00809-0100-3166
Communication
HT HART® Digital Communication Superimposed on 4-20mA Output
FF FOUNDATION™ fieldbus Digital Output
FI FOUNDATION™ fieldbus Digital Output with FISCO
agency approval
60 None Required
FM Approved, Intrinsically Safe (appropriate sensor & safety barrier
67
required), and Non-Incendive
CSA Approved , Intrinsically Safe (appropriate sensor & safety barrier
69
required), and Non-Incendive
73 ATEX/IECEx Approved, Intrinsically Safe (safety barrier required)
Specifications 11
Section 2: Description and Specifications Reference Manual
March 2020 00809-0100-3166
12 Specifications
Reference Manual Section 3: Installation
00809-0100-3166 March 2020
Section 3: Installation
3.1 unpacking and inspection
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions.
Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the
packing list are present. If items are missing, notify Rosemount immediately.
NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit
hub to the conduit before attaching the fitting to the transmitter.
Electrical installation must be in accordance with the National Electrical Code (aNSI/NFpa-70) and/or any
other applicable national or local codes.
Installation 13
Section 3: Installation Reference Manual
March 2020 00809-0100-3166
14 Installation
Reference Manual Section 3: Installation
00809-0100-3166 March 2020
FIGuRE 3-2. pipe and wall mounting dimensions (Mounting bracket pN: 23820-00)
Installation 15
Section 3: Installation Reference Manual
March 2020 00809-0100-3166
16 Installation
Reference Manual Section 4: Wiring
00809-0100-3166 March 2020
Section 4: Wiring
4.1 General
4.1.1 General Information
The 1066 is easy to wire. All wiring connections are located on the main circuit board. The front
panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
For digital communications, the load must be at least 250 ohms. To supply the 12.7 Vdc lift off
voltage at the transmitter, the power supply voltage must be at least 17.5 Vdc.
1000
Load,ohm s
750
w ith HART 545
com m unication ohm s
500
250
w ithoutHART
com m unication
0
12 18 24 30 36 42
Pow ersupply voltage,Vdc
HART option
Wiring 17
Section 4: Wiring Reference Manual
March 2020 00809-0100-3166
18 Wiring
Reference Manual Section 4: Wiring
00809-0100-3166 March 2020
INNER ENCLOSURE
4-20m A /-24VDC RETURN
4-20m A /+24VDC
_
+
TB7 RTD IN
SOL
HINGED PANEL TB6 +V SHLD
1
2 -V pH
TB4 TB1
Wiring 19
Section 4: Wiring Reference Manual
March 2020 00809-0100-3166
FOuNDaTION Fieldbus
Figure 4-4 shows a 1066PFF being used to measure and control pH and chlorine levels in drinking
water. The figure also shows three ways in which Fieldbus communication can be used to read
process variables and configure the transmitter.
20 Wiring
Reference Manual Section 4: Wiring
00809-0100-3166 March 2020
FIGuRE 4-6. ph/ORp sensor wiring to the 1066 printed circuit board
DW G NO.
40106612
Wiring 21
Section 4: Wiring Reference Manual
March 2020 00809-0100-3166
FIGuRE 4-7. Contacting and Toroidal Conductivity sensor wiring to the 1066 circuit board
DW G NO.
40106615
22 Wiring
Reference Manual Section 4: Wiring
00809-0100-3166 March 2020
FIGuRE 4-8. Chlorine, oxygen, ozone sensor wiring to 1066 printed circuit board (1066-Cl, 1066-DO,
1066-Oz)
TB4 TB1
1) TB5/ANODE ANO DE
& CATHODE CATHODE
2) TB3/RTD RETURN
SENSE
RTD IN
Wiring 23
Section 4: Wiring Reference Manual
March 2020 00809-0100-3166
INSTALL PLUG
S IN LL
A OTHER
OPENINGS AS NEEDED
INNER ENCLOSURE
3
(OUTPUT1) SENSOR WIRING
OUTPUT 2 LOOP PWR TB5 TB3 TB2
+24V +24V ANODE RTN REF
TH UM
RTD IN GND
TB7 SOL
TB6 +V SHLD
1
2 -V pH
TB4 TB1
W HITE
WIRE NUT
1066 HART CIRCUIT BOARD
(pH/CL/DO/OZ)
ASSY24406-xx
3 SPLICE CO
NNECTOR - O
PRVIDEDBY END US
ER.
WIRELESS
THUM
ADAPTOR
DW G NO.
40106614
24 Wiring
Reference Manual Section 4: Wiring
00809-0100-3166 March 2020
INNER ENCLOSURE
_
+
TB7 RTD IN
SOL
HINGED PANEL TB6 +V SHLD
1
2 -V pH
TB4 TB1
Wiring 25
Section 4: Wiring Reference Manual
March 2020 00809-0100-3166
26 Wiring
5.1
00809-0100-3166
Reference Manual
APPROVED M ODELS
1066-AA-BB-CC XM TR
18
17
16
15
TABLE III
1066...HT...LOOP POW ER
SIGNAL TERM INALS TB6 -1,-2 & -3 30 200 0.9 0 0
all Intrinsically Safe Installations
1066...HT...ANALOG OUTPUT 2
SIGNAL TERM INALS TB7 -1 & -2 30 200 0.9 0 0
DW G NO REV
E
March 2020
Section 4: Wiring
27
28
October 2017
NICLASS I,DIV 2
GRPS A-D
CLASS II,DIV 2
GRPS E-G
THUM
G ND
OR TB7 FOR ALL OTHER OPTIONS.)
G ND
SHLD SN S C A THO D E G ND
LOAD
THUM
G ND
OR TB7
BOTH M AY BE SHLD SN S C A THO D E
G ND
G ND FIELD COM M UNICATOR REM OTE TRANSM ITTER
INSTALLED REF RTN ANO DE INTERFACE FOR USE IN CLASS IAREA
TB3 TB2 TB5 LOOP PW R OUTPUT 2
SENSOR W IRING (OUTPUT1) (SEE NOTE 3 AND TABLE III) LOAD
1066-CL/DO/OZ...ONLY
LOAD
W HITE
SM ART THUM YELLOW
W IRELESS RED
ADAPTER GREEN N/C
BLACK N/C SPLICE CONNECTOR DW G NO REV
D E
00809-0100-3166
Reference Manual
DRV_B SHLD
RTDIN
THUM
RSHLD
15 16 17
PH SENSOR
17 18
CSA APPROVED DEVICE OR TB1 TB4
THUM
G ND OR TB7
G ND
SHLD SN S C A THO D E G ND
EITHER OR BOTH REF RTN ANO DE
15 16 17
DSHLD
DRV_A
15 16 17
DRV_B SHLD
ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066....HT...
RTDIN
THUM
RSHLD
OR
M ODELS 140,141,142,150,400,400VP,401,402,402VP,403,403VP,404 &
410VP CONTACTING CONDUCTIVITY SENSOR
OR 17 18
ANY SIM PLE APPARATUS W ITH LESS THAN 200 FEET OF ATTACHED CABLE
DW G NO REV
D E
March 2020
Section 5: Intrinsically Safe Installation
29
30
March 2020
A G EN C Y C O N TRO LLED D O C UM EN T
A N Y C HA N G E W ILL REQ UIRE
C ERTIFIC A TIO N A G EN C Y
SUBM ITTA L/A PPRO V A L
FIGuRE 5-6. FM installation
19. A 1066 W ITH THE -FIO PTIO N M A Y BE IN STA LLED PER THE FISC O IN STRUC TIO N S O R PER THE IN TRIN SIC A LLY SA FE IN STRUC TIO N S O N THIS D RA W IN G .
TABLE III
SIGNAL
14. M ETA L C O N D UIT IS N O T REQ UIRED FO R IN TRIN SIC A LLY SA FE IN STA LLA TIO N S.HO W EV ER,IF C O N D UIT IS USED ,BO N D IN G BETW EEN C O N D UIT IS N O T
A UTO M A TIC A N D M UST BE PRO V ID ED A S PA RT O F THE IN STA LLA TIO N .
13. N O REV ISIO N TO D RA W IN G W ITHO UT PRIO R FM A PPRO V A L.A G EN C Y C O N TRO LLED D O C UM EN T. SIGNAL
12. THE A SSO C IA TED A PPA RA TUS M UST BE FM A PPRO V ED A N D M UST BE RESISTIV ELY LIM ITED HA V IN G LIN EA R O UTPUTS.
11. C O N TRO L EQ UIPM EN T C O N N EC TED TO A SSO C IA TED A PPA RA TUS M UST N O T USE O R G EN ERA TE M O RE THA N 250 V rm s O R V dc.
M ODEL NO.
9. THE IN TRIN SIC A LLY SA FE EN TITY C O N C EPT A LLO W S IN TERC O N N EC TIO N O F IN TRIN SIC A LLY SA FE D EV IC ES W ITH A SSO C IA TED A PPA RA TUS W HEN
THE FO LLO W IN G IS TRUE:
FIELD D EV IC E IN PUT A SSO C IA TED A PPA RA TUS O UTPUT
20. THE FISC O C O N C EPT A LLO W S IN TERC O N N EC TIO N O F IN TRIN SIC A LLY SA FE A PPA RA TUS TO A SSO C IA TED A PPA RA TUS N O T SPEC IA LLY EXA M IN ED
8. RESISTA N C E BETW EEN IN TRIN SIC A LLY SA FE G RO UN D A N D EA RTH G RO UN D M UST BE LESS THA N 1.0 O hm . D ELIV ERED BY THE A SSO C IA TED A PPA RA TUS,C O N SID ERIN G FA ULTS A N D A PPLIC A BLE FA C TO RS.IN A D D ITIO N ,THE M A XIM UM UN PRO TEC TED
C A PA C ITA N C E (C i) A N D THE IN D UC TA N C E (Li) O F EA C H A PPA RA TUS (O THER THA N THE TERM IN A TIO N ) C O N N EC TED TO THE FIELD BUS M UST BE
7. D UST-TIG HT C O N D UIT SEA L M UST BE USED W HEN IN STA LLED IN C LA SS IIA N D C LA SS IIIEN V IRO N M EN TS. LESS THA N O R EQ UA L TO 5nF and 10 H RESPEC TIV ELY.
6. C O N TA C TIN G C O N D UC TIV ITY SEN SO RS,A M PERO M ETRIC A N D pH SEN SO RS W ITHO UT PREA M PS SHA LL M EET THE REQ UIREM EN TS O F SIM PLE A PPA RA TUS IN EA C H SEG M EN T O N LY O N E A C TIV E D EV IC E,N O RM A LLY THE A SSO C IA TED A PPA RA TUS,IS A LLO W ED TO PRO V ID E THE N EC ESSA RY EN ERG Y
A S D EFIN ED IN A N SI/ISA RP12.6 A N D THE N EC ,A N SI/N FPA 70.THEY C A N N O T G EN ERA TE N O R STO RE M O RE THA N 1.5 V ,100 m A ,25 m W O R A PA SSIV E FO R THE FIELD BUS SYSTEM .THE V O LTA G E Uo (O R V oc O R V t) O F THE A SSO C IA TED A PPA RA TUS IS LIM ITED TO A RA N G E O F 14 V TO 24 V dc.A LL
C O M PO N EN T THA T D O ES N O T D ISSIPA TE M O RE THA N 1.3W . O THER EQ UIPM EN T C O N N EC TED TO THE BUS C A BLE HA S TO BE PA SSIV E,M EA N IN G THA T THEY A RE N O T A LLO W ED TO PRO V ID E EN ERG Y TO THE
SYSTEM ,EXC EPT A LEA KA G E C URREN T O F 50 FO R EA C H C O N N EC TED D EV IC E.SEPA RA TELY PO W ERED EQ UIPM EN T N EED S G A LV A N IC
ISO LA TIO N TO A SSURE THA T THE IN TRIN SIC A LLY SA FE FIELD BUS C IRC UIT REM A IN S PA SSIV E.
O R G REA TER THA N THE V O LTA G E (V oc O R V t) A N D C URREN T (Isc O R It) W HIC H C A N BE D EV ELO PED BY THE A SSO C IA TED A PPA RA TUS (SA FETY BA RRIER).
IN A D D ITIO N ,THE M A XIM UM UN PRO TEC TED C A PA C ITA N C E (C i) A N D IN D UC TA N C E (Li) O F THE IN TRIN SIC A LLY SA FE A PPA RA TUS,IN C LUD IN G
IN TERC O N N EC TIN G W IRIN G ,M UST BE EQ UA L O R LESS THA N THE C A PA C ITA N C E (C a) A N D IN D UC TA N C E (La) W HIC H C A N BE SA FELY C O N N EC TED TO
THE A PPA RA TUS.(REF.TA BLES I,IIA N D III). A T EA C H EN D O F THE TRUN K C A BLE A N A PPRO V ED IN FA LLIBLE LIN E TERM IN A TIO N W ITH THE FO LLO W IN G PA RA M ETERS IS SUITA BLE:
F
2. THE M O D EL 1066-C /T HA S SYSTEM A PPRO V A L FO R USE W ITH M O D ELS 222,225,226 & 228 TO RO ID A L SEN SO RS O R 140,141,142,150,400,401,402,
O N E O F THE A LLO W ED TERM IN A TIO N S M IG HT A LREA D Y BE IN TEG RA TED IN THE A SSO C IA TED A PPA RA TUS.THE N UM BER O F PA SSIV E A PPA RA TUS
1066 M O D ELS W ITH P/L/D /Z O PTIO N S HA V E O UTPUT EN TITY PA RA M ETERS W HIC H A LLO W THE USE O F V A RIO US SEN SO RS W HIC H M A Y C O N N EC TED TO THE BUS SEG M EN T IS N O T LIM ITED
D UE TO I.S.REA SO N S.IF THE A BO V E RULES A RE
TO THE SEN SO R TERM IN A LS D O N O T EXC EED THE V A LUES SPEC IFIED IN TA BLE IW HERE: RESPEC TED ,UP TO A TO TA L LEN G TH O F 1000 m
(SUM O F TRUN K A N D A LL SPUR C A BLES) O F C A BLE
IS PERM ITTED .THE IN D UC TA N C E A N D THE
1. A N Y SIN G LE SHUN T ZEN ER D IO D E SA FETY BA RRIER A PPRO V ED BY FM HA V IN G THE FO LLO W IN G O UTPUT PA RA M ETERS:SUPPLY/SIG N A L TERM IN A LS TB6-1 C A PA C ITA N C E O F THE C A BLE W ILL N O T IM PA IR THE
A N D 2 FO R FIELD BUS O PTIO N O R TB6-1,2 A N D 3 FO R HA RT O PTIO N .A LSO TB7-1 A N D 2 IF A N A LO G O UTPUT 2 IS USED . IN TRIN SIC SA FETY O F THE IN STA LLA TIO N .
SIZE D W G N O REV
C LA SS II,D IV 2
G RPS E-G
A N A LO G O UTPUT 2 O N LY A V A ILA BLE SA FETY BA RRIER UN SPEC IFIED PO W ER SUPPLY
PH SEN SO R O N 1066...HT... ( SEE N O TES 2,8,9 & 10 30 V D C M A X FO R IN TRIN SIC SA FETY
TB1 TB4
THUM
GND
GND
SHLD SNS CATHODE GND
+24V
REF RTN ANODE +24V RO SEM O UN T M O D EL 375 O R 475
EITHER O R TB2 TB3 TB5 LOOP PW R OUTPUT 2
pH -V
SHLD +V TB6
SOL RTD IN OR RO SEM O UN T M O D EL 375 O R 475
TB7
THUM
GND
SHLD +V TB6
SOL RTD IN
TB7 LO A D
THUM
GND
GND
SHLD SNS CATHODE GND
+24V
REF RTN ANODE +24V
W HITE
M O D EL 775 YELLOW
SM A RT THUM RED
W IRELESS GREEN SIZE D W G N O REV
A D A PTER
N O TES:UN LESS O THERW ISE SPEC IFIED SPLICE CONNECTOR D 1400670 E
SC A LE:1:1 W EIG HT: SHEET 2 O F 3
March 2020
Section 5: Intrinsically Safe Installation
31
32
March 2020
IS C LA SS I,G RPS A -D
C LA SS II,G RPS E-G
C LA SS III
N IC LA SS I,D IV 2
G RPS A -D
C LA SS II,D IV 2
G RPS E-G
FIGuRE 5-8. FM installation
A N A LO G O UTPUT 2 O N LY
A V A ILA BLE O N 1066...HT... SA FETY BA RRIER UN SPEC IFIED PO W ER SUPPLY
( SEE N O TES 2,8,9 & 10 30 V D C M A X FO R IN TRIN SIC SA FETY
DSHLD FO R A LL O THER O PTIO N S.) (24 V D C TYPIC A L)
SHLD
RSHLD RTDIN
THUM
SENSE GND RO SEM O UN T M O D EL 375 O R 475
GND
RTN
15 17
16 17
PH SEN SO R
FM A PPRO V ED D EV IC E O R TB1 TB4
15 17
pH -V
SIM PLE A PPA RA TUS SHLD +V TB6
SOL RTD IN
A N A LO G O UTPUT 2 O N LY A V A ILA BLE O N 1066...HT...
OR TB7
THUM
GND
EITHER O R SHLD SNS CATHODE
GND
GND
+24V
BO TH M A Y BE REF RTN ANODE +24V
DSHLD
15 17
SHLD
RSHLD RTDIN A N A LO G O UTPUT 2 O N LY A V A ILA BLE O N 1066...HT...
THUM
SENSE GND
GND
RTN
The MENu key is used to access menus for programming and calibrating the instrument. Four top-
level menu items appear when pressing the MENU key:
• Calibrate: calibrate the attached sensor and analog output(s).
• Hold: Suspend current output(s).
• Program: Program outputs, measurement, temperature, security and reset.
• Display: Program display format, language, warnings, and contrast
Pressing MENu from the main (live values) screen always causes the main menu screen to appear.
Pressing MENU followed by EXIT causes the main screen to appear.
Pressing the DIaG key displays active Faults and Warnings, and provides detailed instrument infor-
mation and sensor diagnostics including: Faults, Warnings, Sensor information, Out 1 and Out 2
live current values, model configuration string e.g. 1066-P-HT-60 and Instrument Software ver-
sion. Pressing DIAG provides useful diagnostics and information (as applicable): Measurement,
Sensor Type, Raw signal value, Cell constant, Zero Offset, Temperature, Temperature Offset,
selected measurement range, Cable Resistance, Temperature Sensor Resistance, software ver-
sion.
The ENTER key. Pressing ENTER stores numbers and settings and moves the display to the next
screen.
The ExIT key. Pressing EXIT returns to the previous screen without storing changes.
Display and Operation 33
Section 6: Display and Operation Reference Manual
March 2020 00809-0100-3166
Selection Keys:
Surrounding the ENTER key, four Selection keys – up, down, right and left, move the cursor to all
areas of the screen while using the menus.
Selection keys are used to:
1. Select items on the menu screens
2. Scroll up and down the menu lists
3. Enter or edit numeric values
4. Move the cursor to the right or left
5. Select measurement units during operations
Secondary values:
Up to four secondary values are shown in display quadrants at the bottom half of the screen. All four
secondary value positions can be programmed by the user to any displayable parameter available.
If the transmitter detects a problem with itself or the sensor the word Fault or Warning will appear
at the bottom of the display. A fault requires immediate attention. A warning indicates a problem-
atic condition or an impending failure. For troubleshooting assistance, press Diag.
The main display screen can be programmed to show primary process variables, secondary
process variables and diagnostics.
1. Press MENU
2. Scroll down to Display. Press ENTER.
3. Main Format will be highlighted. Press ENTER.
4. The sensor 1 process value will be highlighted in reverse video. Press the selection keys to
navigate down to the screen sections that you wish to program. Press ENTER.
5. Choose the desired display parameter or diagnostic for each of the four display sections in
the lower screen.
6. Continue to navigate and program all desired screen sections. Press MENu and ExIT. The
screen will return to the main display.
The default display shows the live process measurement in the upper display area and tempera-
ture in the center display area. The user can elect to disable the display of temperature in the cen-
ter display area using the Main Format function. See Figure 4-1 to guide you through program-
ming the main display to select process parameters and diagnostics of your choice.
Transmitter Programming 37
Section 7: Transmitter programming Reference Manual
March 2020 00809-0100-3166
7.2.2 Procedure
Follow the Reset Analyzer procedure (Section 7.8) to reconfigure the transmitter to display new
measurements or measurement units. To change the specific measurement or measurement
units for each measurement type, refer to the Program menu for the appropriate measurement
(Section 6.0).
Temperature
require temperature compensation. The 1066 performs tempera-
Units: °C
ture compensation automatically by applying internal temperature
Temp Comp: Auto
correction algorithms. Temperature correction can also be turned
off. If temperature correction is off, the 1066 uses the temperature Manual: +25.0 °C
entered by the user in all temperature correction calculations.
7.4.2 Definitions
1. CURRENT OUTPUTS. The transmitter provides a continuous output current (4-20 mA)
directly proportional to the process variable or temperature. The low and high current out-
puts can be set to any value.
2. ASSIGNING OUTPUTS. Assign a measurement or temperature to Output 1 or Output 2.
3. DAMPEN. Output dampening smooths out noisy readings. It also increases the response
time of the output. Output dampening does not affect the response time of the display.
4. MODE. The current output can be made directly proportional to the displayed value (linear
mode) or directly proportional to the common logarithm of the displayed value (log mode).
38 Transmitter Programming
Reference Manual Section 7: Transmitter programming
00809-0100-3166 March 2020
FIGuRE 7-1. Configuring and 1.234 S/cm 25.0”C 1.234 S/cm 25.0”C
Scale
Linear
Log
O2 SN 4mA: 0 °C
Enter a value for 4 mA and 20 mA for each output. Follow the
O2 SN 20mA: 100 °C
menu screens in Figure 7-1 to assign values to the outputs.
7.5.2 Procedure
1. Press MENU. The main menu screen appears. Choose program.
2. Scroll down to Security. Select Security.
3. The security entry screen appears. Enter a three digit security code for each of the desired
security levels. The security code takes effect two minutes after the last key stroke. Record the
security code(s) for future access and communication to operators or technicians as needed.
4. The display returns to the security menu screen. Press EXIT to return to the previous screen.
To return to the main display, press MENU followed by EXIT.
Figure 7-2 displays the security code screens.
Transmitter Programming 39
Section 7: Transmitter programming Reference Manual
March 2020 00809-0100-3166
MAIN MENU
Program
Program
Outputs
Measurement
Temperature
1.234 µS/cm 25.0 ºC
Security
Security Calibration/Hold: 000
All: 000
Diagnostic Setup
Reset Analyzer
7.6.2 Procedure
1. If a security code has been programmed, selecting the Calibrate, Hold, Program or Display top
menu items causes the security access screen to appear
2. Enter the three-digit security code for the appropriate security level. 1.234 µS/cm 25.0 ºC
Security Code
000
3. If the entry is correct, the appropriate menu screen appears. If the
entry is incorrect, the Invalid Code screen appears. The Enter
Security Code screen reappears after 2 seconds.
40 Transmitter Programming
Reference Manual Section 7: Transmitter programming
00809-0100-3166 March 2020
MAIN MENU
1.234 µS/cm 25.0 ºC
Hold outputs?
Hold
No
Hold
Hold: No
Yes
7.8.2 Procedure
To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the
Reset Analyzer flow diagram.
Transmitter Programming 41
Section 7: Transmitter programming Reference Manual
March 2020 00809-0100-3166
42 Transmitter Programming
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
Programming Measurements 43
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for pH programming and the 1066 live screen
prompts for each function to complete configuration and programming.
8.2.2 Measurement
1.234 µS/cm 25.0 ºC
SN Measurement
The display screen for selecting the measurement is shown. The
default value is displayed in bold type. Refer to the pH/ORP pH
ORP
Redox
Programming flow diagram to complete this function.
8.2.3 Preamp
1.234 µS/cm 25.0 ºC
The display screen for identifying the Preamp location is shown. The
default value is displayed in bold type. Refer to the pH/ORP SN Preamp
Programming flow diagram to complete this function. Analyzer
Sensor/JBox
Note: Sensor/JBox must be selected to support SMART pH sensors
from Rosemount.
44 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
High pH
algorithm is shown. The default value is displayed in bold type. Refer
Custom
to the pH/ORP Programming flow diagram to complete this function.
- 0.032pH/ºC
coefficient is shown. The default value is displayed in bold type. Refer
to the pH/ORP Programming flow diagram to complete this function.
8.2.7 Filter
1.234 µS/cm 25.0 ºC
SN Input filter
The display screen for entering the input filter value in seconds is
04 sec
shown. The default value is displayed in bold type. Refer to the
pH/ORP Programming flow diagram to complete this function.
Programming Measurements 45
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for ORp programming at the end of Sec. 6 and the
1066 live screen prompts for each function to complete configuration and programming.
8.3.2 Preamp
1.234 µS/cm 25.0 ºC
SN Preamp
Analyzer
The display screen for identifying the Preamp location is shown. The
Sensor/JBox
default value is displayed in bold type. Refer to the pH/ORP
Programming flow diagram to complete this function.
8.3.3 Filter
1.234 µS/cm 25.0 ºC
SN Input filter
The display screen for entering the input filter value in seconds is
shown. The default value is displayed in bold type. Refer to the 04 sec
pH/ORP Programming flow diagram to complete this function.
46 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
Filter: 2 sec
program any displayed function, scroll to the desired item and press
Custom Setup
ENTER.
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for contacting conductivity programming at the end
of Section 8 and the 1066 live screen prompts for each function to complete configuration and
programming.
Programming Measurements 47
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
50 µS
shown. The default value is displayed in bold type. Note: Ranges are
500 µS
shown as conductance, not conductivity. Refer to the contacting
conductivity Programming flow diagram to complete this function. 2000 µS
20 mS
200 mS
600 mS
48 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
02 sec
The default value is displayed in bold type. Refer to the contacting
conductivity Programming flow diagram to complete this function.
recommended.
entered data by displaying this screen:
Programming Measurements 49
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for toroidal conductivity programming at the end
of Section 8 and the 1066 live screen prompts for each function to complete configuration and
programming.
Other
50 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
NaOH (0-12%)
HCl (0-15%)
Low H2SO4
High H2SO4
NaCl (0-20%)
Custom Curve
8.5.4 Range
The display screen for Selecting Auto-ranging or a specific range is
1.234 µS/cm 25.0 ºC
SN Range
shown. The default value is displayed in bold type. Note: Ranges are
Auto
shown as conductance, not conductivity. Refer to the toroidal
conductivity Programming flow diagram to complete this function. 50-600 µS
0.5-100 mS
90-1500 mS
NOTE: when manually changing ranges, a zero calibration and
recalibration in a solution of known conductivity must be performed
with the toroidal sensor wired to the instrument. Refer to Section 9.5.2
zeroing the Instrument and Section 9.5.3 Calibrating the Sensor in a
Conductivity Standard.
Raw
Refer to the toroidal conductivity Programming flow diagram to
complete this function.
Programming Measurements 51
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
8.5.7 Slope
1.234 µS/cm 25.0 ºC
SN Slope
The display screen for Entering the conductivity/temp Slope is shown.
2.00%/ºC
The default value is displayed in bold type. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
using the highest range (90 mS to 1500 mS) in very low conductivity
processes below 100 µS (conductance) might generate a high noise
value relative to the actual process value. In these cases, it is
recommended to increase the input filter setting above the default
value of 2 sec. to suppress the effect of noise.
When the custom curve data entry is complete, press ENTER. The 1.234 µS/cm 25.0 ºC
SN Calculate Curve
Custom curve
display will confirm the determination of a custom curve fit to the
fit completed.
entered data by displaying this screen:
In Process Cal
recommended.
If the custom curve fit is not completed or is unsuccessful, the display 1.234 µS/cm 25.0 ºC
SN Calculate Curve
Failure
will read as follows and the screen will return to the beginning custom
curve screen.
52 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for chlorine programming at the end of Sec. 8 and
the 1066 live screen prompts for each function to complete configuration and programming.
Programming Measurements 53
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
8.6.1.2 Units
1.234µS/cm 25.0ºC
SN Units
The display screen for selecting units as ppm or mg/L is shown. The
default value is displayed in bold type. Refer to the Chlorine ppm
Programming flow diagram to complete this function. mg/L
05 sec
The default value is displayed in bold type. Refer to the Chlorine
Programming flow diagram to complete this function.
Live/Continuous
Manual pH correction is shown. The default value is displayed in bold
Manual
type. Refer to the Chlorine Programming flow diagram to complete this
function.
54 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
8.6.2.1 Description
This Chlorine sub-section describes how to configure the 1066 transmitter for Total Chlorine
measurement using amperometric chlorine sensors. The following programming and
configuration functions are covered:
1. Measure: Free Chlorine Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display resolution 0.01 or 0.001
4. Filter: 5sec Override the default input filter, enter 0-999 seconds
The adjacent screen format will appear (factory defaults are shown). To program any displayed
function, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and
the 1066 live screen prompts for each function to complete configuration and programming.
8.6.2.3 Units
1.234 µS/cm 25.0 ºC
SN Units
The display screen for selecting units as ppm or mg/L is shown. The
default value is displayed in bold type. Refer to the Chlorine ppm
Programming flow diagram to complete this function mg/L
8.6.2.4 Filter
1.234 µS/cm 25.0 ºC
SN Input filter
The display screen for entering the input filter value in seconds is
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function. 05 sec
Programming Measurements 55
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
0.01
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function.
The following screen format will appear (factory defaults are shown). To program any displayed
function, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each
configuration function. Use the flow diagram for chlorine programming at the end of Sec. 8 and
the 1066 live screen prompts for each function to complete configuration and programming.
56 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
05 sec
shown. The default value is displayed in bold type. Refer to the
Chlorine Programming flow diagram to complete this function.
1.234µS/cm 25.0ºC
To configure Oxygen:
SN Configure
Type: Water/Waste
1. Press MENU
Units: ppm
Partial Press: mmHg
2. Scroll down to Program. Press ENTER.
3. Scroll down to Measurement. Press ENTER.
Salinity: 00.0‰
Filter: 5sec
Pressure Units: bar
Use Press: At Air Cal
The adjacent screen format will appear (factory defaults are shown). To
Custom Setup
program any displayed function, scroll to the desired item and press
ENTER.
The following sub-sections show the initial display screen that appears for each configuration
function. Use the flow diagram for oxygen programming at the end of Sec. 6 and the 1066 live
screen prompts for each function to complete configuration and programming.
Programming Measurements 57
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
Brewing
Programming flow diagram to complete this function.
Oxygen In Gas
% Saturation
Partial Pressure
% Oxygen In Gas
ppm Oxygen In Gas
8.7.4 Salinity
The display screen for Entering the Salinity (as parts per thousand) of 1.234 µS/cm 25.0ºC
SN Salinity
00.0 ‰
the process liquid to be measured is shown. The default value is dis-
played in bold type. Refer to the Oxygen Programming flow diagram
to complete this function.
Enter Salinity as ‰
05 sec
shown. The default value is displayed in bold type. Refer to the Oxygen
Programming flow diagram to complete this function.
atm
sure is shown. This selection is needed for the display of atmospheric
kPa
pressure. The default value is displayed in bold type. The user must
mbar
enter a known value for local atmospheric pressure. Refer to the
Oxygen Programming flow diagram to complete this function. bar
58 Programming Measurements
Reference Manual Section 8: programming Measurements
00809-0100-3166 March 2020
To configure Ozone:
1.234 µS/cm 25.0ºC
SN Configure
Units: ppm
1. Press MENU
2. Scroll down to Program. Press ENTER. Filter: 5 sec
Resolution: 0.001
3. Scroll down to Measurement. Press ENTER.
The adjacent screen format will appear (factory defaults are shown). To program any displayed
function, scroll to the desired item and press ENTER.
The following sub-sections show the initial display screen that appears for each configuration
function. Use the flow diagram for ozone programming at the end of Sec. 6 and the 1066 live
screen prompts for each function to complete configuration and programming.
Note: Ozone measurement boards are detected automatically by the analyzer. No measurement
selection is necessary.
05 sec
The display screen for entering the input filter value in seconds is shown.
The default value is displayed in bold type. Refer to the Ozone
Programming flow diagram to complete this function.
Programming Measurements 59
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
60 Programming Measurements
1.234 S/cm 25.0…C 1.234 S/cm 25.0…C
TD S
Salinity 1.234 S/cm 25.0…C
Reference Manual
SN R ange
1.234 S/cm 25.0…C
Auto
Programming Measurements
SN C ellC onstant
0-50 S
50-500 S 1.00000 /cm
500-2000 S
2000 S -20 m S 1.234 S/cm 25.0…C
20 m S -200 m S
200 m S -600 m S SN R TD O ffset
0.00¡C
1.234 S/cm 25.0…C
1.234 S/cm 25.0…C
SN C onfigure
M Program M easure: C ond 1.234 S/cm 25.0…C
A P
O utputs 1.234 S/cm 25.0…C Type: 2-Electrode
I r SN R TD Slope
M easurem ent R ange: Auto
N o C onfigure?
Figure 8-2. Configure Contacting Measurements
2.00% /…C
g Tem perature M easurem ent C ellK: 1.00000/cm
M r contacting R TD O ffset: 0.00…C 1.234 S/cm 25.0…C
E a Security
R TD Slope: 0 SN Tem p C om p
N m R esetAnalyzer Tem p C om p: Slope
Slope
U Frequency Slope: 2.00% /¡C
1.234 S/cm 25.0…C N eutralSalt
R efTem p: 25.0¡C
C ation
Filter: 2 sec SN Slope
R aw
C ustom Setup 2.00 % /…C
1.234 S/cm 25.0…C
1.234 S/cm 25.0…C SN R efTem p
1.234 S/cm 25.0…C
SN C ustom C urve
(25.0…C norm al)
SN C ustom C onfig
C onfigure U nits: ppm +25.0…C
EnterD ata Points # ofPoints: 2
C alculate C urve R efTem p: 25.0 …C
Slope: 2.00 % /¡C 1.234 S/cm 25.0…C
61
62
1.234 S/cm 25.0…C
SN Measurement
March 2020
Conductivity
Resistivity
TDS
Salinity
1.234 S/cm 25.0…C
NaOH (0-12%) 1.234 S/cm 25.0…C
SN Model HCl (0-15%)
Low H2SO4 SN Cell Constant
228 High H2SO4
225 3.00000 /cm
NaCl (0-20%)
226
247 1.234 S/cm 25.0…C
SN Range
1.234 S/cm 25.0…C
Auto
50-600 S SN RTD Offset
1.234 S/cm 25.0…C 1.2
34 S/cm 25.0…C
0.5-100 mS
Section 8: programming Measurements
0.00¡C
M SN Configure 90-1500 mS
Program
A P Model: 228 200 S
1.234 S/cm 25.0…C
Figure 8-3. Configure Toroidal Measurements
Programming Measurements
1.234 S/cm 25.0”C
00809-0100-3166
SN Units
Reference Manual
ppm
mg/L
Programming Measurements
ppb
g/L
1.234 S/cm 25.0”C % Saturation
Partial Pressure
SN Type
% Oxygen In Gas
Water/Waste ppm Oxygen In Gas
Trace Oxygen
BioRx-Rosemount
BioRx-Other 1.234 S/cm 25.0”C
Brewing SN Partial Press
Oxygen In Gas
mm Hg
in Hg
Figure 8-4. Configure Oxygen Measurements
Presssure Units
1.234 S/cm 25.0”C mm Hg
in Hg
SN Resolution
atm
0.001 kPa
0.01 mbar
bar
SN Use Pressure?
At Air Cal
m A Input
March 2020
Section 8: programming Measurements
63
64
March 2020
SN Measurement
Free Chlorine 1.234 S/cm 25.0”C
Total Chlorine SN U nits
Monochloramine ppm
m g/L
Section 8: programming Measurements
Program
Filter: 5sec
Security Chlorine
M AIN M EN U
Dual Cal: Disable
Diagnostic Setup Free Cl Correct: Live
Reset Analyzer Manual pH: 7.00 pH
Frequency Resolution: 0.001
1.234 S/cm 25.0”C
SN Manual pH
07.00 pH
1.234 S/cm 25.0”C
SN Resolution
0.001
0.01
00809-0100-3166
Reference Manual
Programming Measurements
00809-0100-3166
Reference Manual
Programming Measurements
1.234 S/cm 25.0”C
1.234 S/cm 25.0”C SN U nits
ppm
FIGuRE 8-5. Configuring Ozone Measurements
Program
m g/L
Outputs 1.2
34 S/cm 25.0”C ppb
1.234 S/cm 25.0”C g/L
Measurement
Measurement
Temperature SN Configure
Configure?
Measurement Units: ppm 1.234 S/cm 25.0”C
Program
Security Ozone
SN Input filter
M AIN M EN U
Filter: 5sec
Reset Analyzer 05 sec
Resolution: 0.001
Frequency
SN Resolution
0.001
0.01
March 2020
Section 8: programming Measurements
65
Section 8: programming Measurements Reference Manual
March 2020 00809-0100-3166
66 Programming Measurements
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Section 9: Calibration
9.1 Introduction
Calibration is the process of adjusting or standardizing the transmitter to a lab test or a calibrated
laboratory instrument, or standardizing to some known reference (such as a commercial buffer).
The auto-recognition feature of the transmitter will enable the appropriate calibration screens to
allow calibration for any configuration of the transmitter. Completion of Quick Start upon first
power up enables live measurements but does not ensure accurate readings in the lab or in
process. Calibration should be performed with each attached sensor to ensure accurate,
repeatable readings.
This section covers the following calibration functions:
1. Auto buffer cal for pH (pH Cal - Section 9.2)
2. Manual buffer cal for pH (pH Cal - Section 9.2)
3. Set calibration stabilization criteria for pH (pH Cal - Section 9.2)
4. Standardization calibration (1-point) for pH, ORP and Redox (pH Cal - Section 9.2 and 9.3)
5. SMART sensor auto calibration upload
9.2 Calibration
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto
calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces
errors. The transmitter recognizes the buffers and uses temperature-corrected pH values in the
calibration. Once the 1066 successfully completes the calibration, it calculates and displays the
calibration slope and offset. The slope is reported as the slope at 25 °C.
To calibrate the pH loop with a connected pH sensor, access the Calibration screen by pressing
ENTER from the main screen and press ENTER.
To calibrate pH:
1. Press the MENu button
2. Select Calibrate. Press ENTER.
3. Select ph. Press ENTER.
1.234 µS/cm 25.0 ºC
SN Calibrate?
pH
The following sub-sections show the initial display screen that appears
Temperature
for each calibration routine. Use the flow diagram for ph calibration at
the end of Section 7 and the live screen prompts to complete
Calibration 67
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
Note that pH auto calibration criteria can be changed. The following criteria can be adjusted:
• Stabilization time (default 10 sec.)
• Stabilization pH value (default 0.02 pH)
1.234 µS/cm 25.0 ºC
SN Setup
• Type of Buffer used for AUTO CALIBRATION (default is Standard,
Offset: 60 mV
return to the pH Buffer Cal Menu.
Calculated: 61.22mV
Max: 60.00mV
Press EXIT
68 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Buffer 2
being used; otherwise, use auto calibration. Auto calibration avoids
common pitfalls and reduces errors. The adjacent appears after
selecting Manual pH calibration.
9.2.4 Standardization – pH
1.234 µS/cm 25.0 ºC
The pH measured by the 1066 transmitter can be hanged to match
the reading from a second or referee instrument. The process of SN Enter Value
making the two readings agree is called standardization. During 07.00pH
standardization, the difference between the two pH values is
converted to the equivalent voltage. The voltage, called the reference
offset, is added to all subsequent measured cell voltages before they are converted to pH. If a
standardized sensor is placed in a buffer solution, the measured pH will differ from the buffer pH
by an amount equivalent to the standardization offset.
Calibration 69
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
To calibrate the ORP loop with a connected ORP sensor, access the Calibration screen by pressing
ENTER from the main screen and press ENTER.
1. Standardization ORP 1 point buffer calibration with manual buffer value entry.
To calibrate ORP:
1.234 µS/cm 25.0 ºC
1. Press the MENU button. SN Calibrate?
2. Select Calibrate. Press ENTER. ORP
3. Select ORP and Redox. Press ENTER. Temperature
The following sub-sections show the initial display screen that appears for each calibration rou-
tine. Use the flow diagram for ORp calibration at the end of Section 8 and the live screen prompts
to complete calibration.
If the ORP and Redox Cal is successful, the screen will return to the Cal sub-menu.
70 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Calibration 71
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
Meter Cal
routine to complete calibration.
The adjacent screen appears after selecting Conductivity calibration: Cell K: 1.00000/cm
In Water
signal that are present even when there is no conductivity to be
measured. This procedure is affected by the length of extension cable and
should always be repeated if any changes in extension cable or sensor
have been made. Electrically connect the conductivity probe as it will 1.234 µS/cm 25.0 ºC
SN Zero Cal
Sensor Zero Done
actually be used and place the measuring portion of the probe in air. Be
sure the probe is dry.
The adjacent screen will appear after selecting Zero Cal from the
1.234 µS/cm 25.0 ºC
Conductivity Calibration screen:
SN Zero Cal
Sensor Zero Fail
The adjacent screen will appear if zero Cal is successful. The screen will
Offset too high
return to the conductivity Cal Menu.
Press EXIT
The adjacent screen may appear if zero Cal is unsuccessful.
72 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Updated cell
will return to the conductivity Cal Menu.
constant:
1.00135/cm
Press EXIT
If the meter cal is successful the screen will return to the conductivity Cal
1.234 µS/cm 25.0 ºC
Menu.
SN Meter Cal
Calibration
The adjacent screen will appear if Meter Cal is unsuccessful. The screen Error
Press EXIT
will return to the conductivity Cal Menu.
Calibration 73
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
In Process Cal
Cell K: 2.7000/cm
9.5.1 Entering the Cell Constant
New toroidal sensors always need to be calibrated. The cell constant provided on the sensor label
is a nominal value and does not need to be entered.
This procedure sets up the transmitter for the probe type connected to the transmitter. Each type
of probe has a specific cell constant
1.234 µS/cm 25.0 ºC
Cell Constant
The display screen for entering a cell constant for the sensor is shown.
2.7000/cm
The default value is displayed in bold type.
74 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Press EXIT
Calibration
return to the conductivity Cal Menu.
Error
Press EXIT
Calibration 75
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
The following sub-sections show the initial display screen that appears for each calibration rou-
tine. Use the flow diagram for Chlorine calibration at the end of Section 7 and the live screen
prompts to complete calibration.
1.234 µS/cm 25.0 ºC
The adjacent screen appears after selecting Free Chlorine calibration: SN Calibration
Zero Cal
In Process Cal
76 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Press EXIT
If the In Process Cal is successful, the screen will return to the Cal sub-menu.
1.234 µS/cm 25.0 ºC
SN InProcess Cal
The adjacent screen may appear if In Zero Cal is unsuccessful. The screen
Calibration
will return to the Amperometric Cal Menu.
Error
Press EXIT
Calibration 77
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
Calibration is to establish the slope of the calibration curve. Because stable total chlorine
standards do not exist, the sensor must be calibrated against a test run on a grab sample of the
process liquid. Several manufacturers offer portable test kits for this purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER from the main
screen, and press ENTER.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution with zero total chlorine
2. Grab Cal Standardizing to a sample of known total chlorine concentration
To calibrate total chlorine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
Temperature
Temperature, scroll to the desired item and press ENTER
The following sub-sections provide you with the initial display screen that appears for each
calibration routine. Use the flow diagram for Chlorine calibration at the
1.234 µS/cm 25.0 ºC
SN Calibration
end of Sec. 9 and the live screen prompts to complete calibration.
This adjacent screen appears after selecting Total Chlorine calibration: Zero Cal
In Process Cal
1.234µS/cm 25.0ºC
SN Zero Cal
Sensor zero done
The adjacent screen will appear if In Zero Cal is successful. The screen will
return to the Amperometric Cal Menu.
Press EXIT
Calibration error
screen will return to the Amperometric Cal Menu.
Press EXIT
78 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
The following sub-sections provide you with the initial display screen that appears for each
calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 9 and the live
screen prompts to complete calibration.
In Process Cal
Calibration 79
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
Press EXIT
Press EXIT
80 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
saturated water are identical. The equivalence comes about because the sensor really measures
the chemical potential of oxygen. Chemical potential is the force that causes oxygen molecules to
diffuse from the sample into the sensor where they can be measured. It is also the force that
causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is
saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two
phases (air and water) is the same. Oxygen sensors generate a current directly proportional to the
rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor.
The diffusion rate depends on the difference in chemical potential between oxygen in the sensor
and oxygen in the sample. An electrochemical reaction, which destroys any oxygen molecules
entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the
sensor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines
the diffusion rate and the sensor current. When the sensor is calibrated, the chemical potential
of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air
or air-saturated water is immaterial. The chemical potential of oxygen is the same in either
phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it
is convenient to use water-saturated air for calibration. Automatic air calibration is standard. The
user simply exposes the sensor to water-saturated air. The transmitter monitors the sensor
current. When the current is stable, the transmitter stores the current and measures the
temperature using a temperature element inside the oxygen sensor. The user must enter the
barometric pressure. From the temperature the transmitter calculates the saturation vapor
pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure
from the barometric pressure. Using the fact that dry air always contains 20.95% oxygen, the
transmitter calculates the partial pressure of oxygen. Once the transmitter knows the partial
pressure of oxygen, it uses the Bunsen coefficient to calculate the equilibrium solubility of
atmospheric oxygen in water at the prevailing temperature. At 25 °C and 760 mm Hg, the
equilibrium solubility is 8.24 ppm. Often it is too difficult or messy to remove the sensor from the
process liquid for calibration. In this case, the sensor can be calibrated against a measurement
made with a portable laboratory instrument. The laboratory instrument typically uses a
membrane-covered amperometric sensor that has been calibrated against water-saturated air.
To calibrate the oxygen sensor, access the Calibration screen by pressing ENTER from the main
screen, select and press ENTER.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in a medium with zero oxygen
2. Air Cal Calibrating the sensor in a water-saturated air sample
3. In Process Cal Standardizing to a sample of known oxygen concentration
4. Sen@ 25 °C:2500 nA/ppm Entering a known slope value for sensor response
To calibrate oxygen:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Oxygen. Press ENTER.
Temperature
scroll to the desired item and press ENTER
Calibration 81
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
The following sub-sections provide you with the initial display screen that appears for each
calibration routine. Use the flow diagram for Oxygen calibration at the end of Section 9 and the
live screen prompts for each routine to complete calibration.
1.234 nA
9.7.1 Zeroing the Sensor SN Zero Cal
Zeroing
Wait
The adjacent screen will appear during Zero Cal.
1.234 nA
SN Zero Cal
The adjacent screen will appear if In Zero Cal is successful. The
Sensor zero done
screen will return to the Amperometric Cal Menu.
1.234 nA
The adjacent screen may appear if In Zero Cal is unsuccessful. The SN Zero Cal
screen will return to the Amperometric Cal Menu. Sensor zero failed
Press EXIT
Done
will return to the Amperometric Cal Menu.
Press EXIT
82 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
Calibration 83
Section 9: Calibration Reference Manual
March 2020 00809-0100-3166
The following sub-sections provide you with the initial display screen that appears for each calibra-
tion routine. Use the flow diagram for Ozone calibration at the end of Section 9 and the live screen
prompts to complete calibration.
1.234 nA
SN Zero Cal
The following screen will appear if In Zero Cal is successful. The screen
1.234 nA
SN Zero Cal
The following screen may appear if In Zero Cal is unsuccessful. The screen
Press EXIT
Press EXIT
84 Calibration
Reference Manual Section 9: Calibration
00809-0100-3166 March 2020
To calibrate temperature, access the Calibration screen by pressing ENTER from the main screen,
select Temperature and press ENTER.
To calibrate temperature:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Temperature. Press ENTER. 1.234 µS/cm 25.0 ºC
SN Calibrate
+0 25.0 °C
The adjacent screen will appear.
The following sub-section provides you with the initial display screen that appears for temperature
calibration. Use the flow diagram for Temp calibration at the end of Sec. 7 to complete calibration.
9.9.1 Calibration
1.234 µS/cm 25. 0 ºC
The adjacent screen will appear during Temperature Cal. SN Calibrate
Cal in progress.
Please wait.
If the sensor Temperature offset is greater than 5 °C from the default
value, the following screen will appear:
Calibration 85
86
M C
March 2020
E a O zone HCl
O xygen Low H 2SO 4 1.234 S/cm 25.0…C
FIGuRE 9-1. Calibrate ph
1.234 S/cm 25.0…C 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C
1.234 S/cm 25.0…C
SN pH C al SN pH Auto C al SN Setup SN pH Auto C al
SN pH BufferC al
BufferC al StartAutoC al Stable Tim e: 10 sec Place Sensorin
Auto Setup Stable D elta: 0.02 pH Buffer1
Standardize
M anual Buffer: Standard Press EN TER
Slope: 59.16m V/pH
O ffset: 600 m V
1.234 S/cm 25.0…C 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C
1.234 S/cm 25.0…C
SN pH M anualC al M anualBufferM SN M anualC al SN pH M anualC al
SN pH Auto Buffer2
Buffer1 07.00pH C alin progress. Slope: 59.16m V/pH
W ait
Buffer2 Please w ait. O ffset: 60 m V
Calibration
Calibration
1.234 S/cm 25.0”C
1.234 S/cm 25.0”C 1.234 S/cm 25.0”
1.234 S/cm 25.0”C
M
00809-0100-3166
C SN Calibrate?
SN C alibration SN Zero C al SN Zero C al
A a
Reference Manual
SN InProcess C al
Figure 9-2. Calibrate Contacting and Toroidal Conductivity
W aitforstable
reading.
87
88
March 2020
Section 9: Calibration
SN Enter Value
10.00 ppm
Figure 9-3. Calibrate Free Chlorine, Total Chlorine, and Monochloramine
SN InProcess Cal
Cal in progress.
Please wait.
00809-0100-3166
Reference Manual
Calibration
1.234 S/cm 25.0”C
1.234 S/cm 25.0”C
Calibration
1.234 S/cm 25.0”C
M C SN Calibrate?
SN AirC al 1.234 S/cm 25.0”C
A a SN C alibration
TDS StartC alibration SN AirPressure
I l 1.234 S/cm 25.0”C Air Cal
00809-0100-3166
CustomConc n
Conductivity Temperature 1.234 S/cm 25.0”C
SN AirC al
D one
SN Stable D elta
0.05 ppm
89
90
March 2020
Section 9: Calibration
M C SN Calibrate?
A a
TDS
I l 1.234 S/cm 25.0”C
Free Chlorine Salinity 1.234 S/cm 25.0”C 1.234 S/cm 25.0”C
N i 1.234 S/cm 25.0”C
Total Chlorine NaOH
b SN Zero C al SN Zero C al
Calibrate? Chloramine HCl SN C alibration
M r Ozone Low H2SO4 Zeroing Sensorzero done
Zero C
al
Oxygen W ait
E a High H2SO4 In Process
C al
N t pH NaCl
U e ORP Resistivity
Redox Custom Conc n
1.234 S/cm 25.0”C 1.234 S/cm 25.0”C 1.234 S/cm 25.0”C
Conductivity Temperature
SN InProcess C al SN InProcess C al SN InProcess C al
W aitforstable PressEN TER if Take sam ple;
reading reading is stable. PressEN TER .
SN Enter Valu
e
10.00 ppm
SN InProcess Cal
Cal in progress.
Please wait.
00809-0100-3166
Reference Manual
Calibration
Calibration
00809-0100-3166
Reference Manual
M C SN C alibrate?
A a Free C hlorine TD S
I l 1.234 S/cm 25.0…C Salinity
N i TotalC hlorine N aO H
b C alibrate? C hloram ine HCl
M r O zone Low H 2SO 4 1.234 S/cm 25.0…C
1.234 S/cm 25.0…C
E a O xygen H igh H 2SO 4 SN Standardize
pH SN EnterValue
N t N aC l
O RP R esistivity C alin progress.
U e +0600 m V
R edox Please w ait.
C ustom C onc n
91
92
March 2020
Section 9: Calibration
A a SN C alibrate?
I l 1.234 S/cm 25.0…C
N i Free C hlorine TD S
b C alibrate? TotalC hlorine Salinity
M r C hloram ine N aO H
E a O zone HCl
N t O xygen Low H 2SO 4
pH H igh H 2SO 4
U e
O RP N aC l
R edox R esistivity
C onductivity C ustom C onc n 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C 1.234 S/cm 25.0…C
SN C alibrate SN C alibrate
Tem
Tem per
perat
atur
uree SN Tem p O ffset> 5¡C
+025.0¡C C alin progress. C ontinue?
Please w ait. No
Yes
00809-0100-3166
Reference Manual
Calibration
Reference Manual Section 10: haRT Communication
00809-0100-3166 March 2020
10.1 Introduction
The 1066 transmitter can communicate with a HART host using HART Revision 5 or HART Revision
7. The revision of HART used by the 1066 can be selected using the keypad/display or a HART
master such as AMS or the 475 Handheld Communicator. The default version of HART is Revision
5. Since some HART hosts cannot accommodate HART 7, the choice of HART Revision should be
based on the capabilities of the host, and should be chosen as a first step in configuration. If HART
Revision 7 can be used with the host, it does offer a number of advantages over Revision 5,
including long tag name, time stamped data, and measurement status, and enhanced burst mode.
When haRT 5 is chosen, the Device Revision of the 1066 is Device Revision 1; when haRT 7 is chosen
the Device Revision is Revision 2 (or greater). The Device Revision of the DD (Device Description)
and install files for AMS and DeltaV used should be the same as the Device Revision of the 1066.
A single HART 5 (Device Revision 1) or HART 7 (Device Revision 2 or greater) DD (Device
Description) is used for all model codes of the 1066, which include the pH/ORP, conductivity,
oxygen, chlorine, and ozone transmitters. All 1066 transmitters have the same HART device
identification, as outlined below:
HART Communications 93
Section 10: haRT Communication Reference Manual
March 2020 00809-0100-3166
• univ Cmd Rev – toggles between HART version 5 and HART version 7. If the HART host being
used can accommodate HART 7, HART 7 should be chosen due to its larger feature set. If the
host can only use HART 5, then HART 5 must be chosen.
Note: If the 1066 is connected to a HART host and the HART version is changed, the host will
likely detect the transmitter as a new transmitter with a different device revision number.
• polling address – Choose “0” unless Multidrop is being used. If Multidrop is being used, each
transmitter should have its own polling address of from 1 to 15.
• loop current mode – Set Output 1 current to a minimum value for multidrop applications
(HART 7 only).
• Find Device Cmd – Setting Find Device to “On”, enables the 1066 to be identified by the host.
The transmitter returns identity information including device type, revision level, and device
ID (HART 7 only).
94 HART Communications
Reference Manual Section 10: haRT Communication
00809-0100-3166 March 2020
• Burst Message 0, 1, 2 – Toggles burst messages 0, 1, and/or 2 on or off (HART 7 only). See the
end of section 10.2 for the HART burst commands available.
If HART 5 is chosen (Univ Cmd Rev = 5), the following controls are available:
• univ Cmd Rev – toggles between HART version 5 and HART version 7. If the HART host being
used can accommodate HART 7, HART 7 should be chosen due to its larger feature set. If the
host can only use HART 5, then HART 5 must be chosen.
Note: If the 1066 is connected to a HART host and the HART version is changed, the host will
likely detect the transmitter as a new transmitter with a different device revision number.
• polling address – Choose “0” unless Multidrop is being used. If Multidrop is being used, each
transmitter should have its own polling address of from 1 to 15.
• Burst Mode – Toggles the single HART 5 burst message on or off. See below for the HART
burst commands available.
HART Communications 95
Section 10: haRT Communication Reference Manual
March 2020 00809-0100-3166
96 HART Communications
Reference Manual Section 10: haRT Communication
00809-0100-3166 March 2020
HART Communications 97
Section 10: haRT Communication Reference Manual
March 2020 00809-0100-3166
98 HART Communications
Reference Manual Section 10: haRT Communication
00809-0100-3166 March 2020
HART Communications 99
Section 10: haRT Communication Reference Manual
March 2020 00809-0100-3166
1066 ph/ORp
Type Enumeration / Read/Write Default
Range
zero_offset FLOAT R/W 0.0 mV
The zero offset of a pH sensor resulting from a calibration.
ph_slope FLOAT 40 -- 62 mV/pH R / W 59.16 mV/pH
The slope of a pH sensor resulting from a buffer calibration.
zero_offset_limit FLOAT R/W 60
The maximum value of the zero offset allowed for a successful
calibration.
min_ph_slope FLOAT 40 -- 62 mV/pH R / W 40 mV/pH
The minimum value of the slope allowed for a successful calibration.
max_ph_slope FLOAT 40 -- 62 mV/pH R / W 62 mV/pH
The maximum value of the slope allowed for a successful calibration.
ph_stabilization_time FLOAT 0 -- 99 seconds R/W 10 seconds
The period of time used to determine stability of a pH sensor during
automatic calibration.
ph_stabilization_value FLOAT 0.01 -- 1.0 pH R/W 0.02 pH
The pH change used to determine stability of a pH sensor during
automatic calibration.
ph_buffer_standard ENUM Standard / Nist R/W Standard / Nist
The types of pH buffers available for automatic calibration. DIN 19267
Ingold
Merck
Fisher
preamp_location ENUM analyzer R/W
The location of the preamplifier in a pH measurement. sensor / J-box
reference_impedance_level ENUM Low R/W Low
Configures the transmitter to use reference electrodes with High
impedances less than 500 kohm (low) or greater than 500 kohm
(high).
ph_solution_temperature_correction ENUM Off Off
Configures the transmitter to correct the pH measurement for ultra pure water
changes in the pH of the solution with temperature. high pH
ammonia
custom
ph_solution_temperature_coefficient FLOAT -9.999 -- 9.999 R/W 0.000 pH / °C
The coefficient used to correct solution pH when using custom pH / °C
solution pH temperature correction.
1066C
Type Enumeration / Read/W Default
Range rite
1066T
Type Enumeration / Read/Write Default
Range
toroidal_sensor_model ENUM 228 R/W 228
Selects the model of toroidal sensor being used for enhanced 225
accuracy. 226
247
Other
cond_range ENUM Automatic R/W Automatic
Selects a particular range for the conductivity measurement or auto- 50-600 µS
matic ranging for toroidal sensors. 0.5-100 mS
90-1500 mS
temp_comp_type ENUM Linear slope R/W Linear slope
The type of temperature compensation used by the transmitter. Raw / None
temp_comp_slope FLOAT R/W 2.0% / °C
Provides a slope value used in linear temperature compensation.
reference_temperature FLOAT R/W 25 °C
The temperature that temperature compensation corrects to. It is
usually 25 °C.
1066Oz/Cl/DO
Type Enumeration / Range Read/Write Default
polar_voltage FLOAT R/W 0
The polarization voltage used by the transmitter; automatical-
ly set by sensor selection.
temp_coeff FLOAT R/W 0
The temperature coefficient used to compensate temperature;
automatically set by sensor selection.
amp_sensor_sensitivity FLOAT 0.1 -- 1,000,000.0 R/W 2,500 nA/ppm
The response of the sensor to changes in concentration, deter- nA/ppm
mined by calibration.
amp_sensor_zero_current FLOAT -999.9 -- 999.9 nA R/W 0.0 nA
The current output of the sensor when the concentration is 0.
It is determined by a zero calibration.
1066Oz
Type Enumeration / Range Read/Write Default
chlorine_type ENUM 0.001 R/W 0.001
0.01
The resolution of the displayed ozone measurement.
1066Cl
Type Enumeration / Range Read/Write Default
chlorine_type ENUM Free Chlorine R/W Free Chlorine
Total Chlorine
The type of chlorine being measured. Chloramine
1066DO
Type Enumeration / Range Read/Write Default
2.2.8.2. Output 2
2.2.8.2.1. Secondary Variable
2.2.8.2.2. SV URV
2.2.8.2.3. SV LRV
2.2.8.2.4. Scale
2.2.8.2.5. Dampening
2.2.8.2.6. Fault Mode
2.2.8.2.7. Fault Value
2.2.9. Communications
2.2.9.1. Burst Message 1
2.2.9.1.1. Burst Message
2.2.9.1.2. Message Content
2.2.9.1.3. Update Rate
2.2.9.1.3.1. Trigger Mode
2.2.9.1.3.2. Trigger Level
2.2.9.1.3.3. Trigger Level Unit
2.2.9.1.3.4. Classification
2.2.9.1.3.5. Update Rate
2.2.9.1.3.6. Default Update Rate
2.2.9.2. Burst Message 2
2.2.9.2.1. Burst Message
2.2.9.2.2. Message Content
2.2.9.2.3. Update Rate
2.2.9.2.3.1. Trigger Mode
2.2.9.2.3.2. Trigger Level
2.2.9.2.3.3. Trigger Level Unit
2.2.9.2.3.4. Classification
2.2.9.2.3.5. Update Rate
2.2.9.2.3.6. Default Update Rate
2.2.9.3. Burst Message 3
2.2.9.3.1. Burst Message
2.2.9.3.2. Message Content
2.2.9.3.3. Update Rate
2.2.9.3.3.1. Trigger Mode
2.2.9.3.3.2. Trigger Level
2.2.9.3.3.3. Trigger Level Unit
2.2.9.3.3.4. Classification
2.2.9.3.3.5. Update Rate
2.2.9.3.3.6. Default Update Rate
2.2.9.4. Event Notification
2.2.9.4.1. Event Message
2.2.9.4.2. Unack Update Rate
2.2.9.4.3. Default Update Rate
2.2.9.4.4. Debounce Interval
2.2.9.4.5. Pending Events
2.2.9.4.6. Acknowledge Event (If there are pending events)
2.2.9.5. 12-Point Sample
2.2.9.5.1. Data Sampling
2.2.9.5.2. Device Variable
2.2.9.5.3. Sample Interval
2.2.9.5.4. 12-Point Sample (Data Sampling is not Off)
2.2.9.6. Variable Mapping
2.2.9.6.1. PV is
2.2.9.6.2. SV is
2.2.9.6.3. TV is
2.2.9.6.4. QV is
2.2.9.7. Multidrop
2.2.9.7.1. Polling Address
2.2.9.7.2. Loop Current Mode
NOTICE
Consult the factory for additional information regarding service or repair.