DIRECTION 2340897-100, REVISION 08 HISPEED QX/I SERVICE MANUAL - GENERAL


2.) Normal rail voltage should only be used to test for dielectric breakdowns. Turn on one side (cath-
ode/anode) at a time, since the bang-bang circuit was not designed for accurate KV loop control.
3.) High potting the tube is very dangerous if not done correctly, even with the HVDC bus set to
test mode.

Figure 2-112 HVDC Test Setup

3.6.8 kV Fiber-Optic Test

2 - SW Tools
SERVICE DESKTOP -> DIAGNOSTICS -> KV LOOP -> HV FIBER OPTICS

Figure 2-113 kV Fiber-Optic Screen

Diagnostic Description
This diagnostic sequentially enables the fiber optic drivers to the KV inverters.

When (to use)


1.) KV values drop as the mA increases.
2.) Suspect bad fiber-optic connection.
3.) Verification that all fiber optic connections are not miss wired.

What (to look for)


Verify LEDs located on the KV inverters are enabled in the correct sequence.

Chapter 2 - Service Desktop, Tools, and Diagnostics Page 145