Quick Reference Guide

Weekly Calibration of the


i-CAT FLX V-Series
Weekly i-CAT Calibrations Recommended
It is recommended that a Panel Calibration and Shutter Calibration be performed once a week to
ensure optimal image quality. Calibration instructions may be found in the i-CAT FLX User Manual. (It is
recommended that a GeoCal be performed once a year to ensure optimal image quality or if the image
quality is degraded. Instructions for a Geometric Calibration may be found in the i-CAT FLX Technical Guide.
To request a manual or if assistance is needed, please call i-CAT Technical Support at (800) 205-3570.)
Begin by opening Smart Scan Studio on the Scanner Controller (the computer that the touch panel is
connected to) and pressing to access the Utilities menu.

Performing the Panel Calibration


The Panel Calibration is performed in both
landscape and portrait positions for 4 x 4 and 2
x 2 resolutions. Several tests run as part of the
panel calibration. A pie chart displays status.
Proceed as follows:
1. Select PanelCal.
2. E
 nsure the field of view on the
scanner is clear. (Remove the
headrest and open the gate. The
scanner’s gantry will not rotate so
you do not need to worry about it
colliding with the chin cup or chin
rest.)
3. Press . The scanner initializes.
4. W
 hen prompted, press the Scan
button on the operator control box. An audible alarm sounds and the X-ray OK light is illuminated
during radiation exposure.
5. You
 will be prompted to press the Scan button for each test.
NOTE: The panel will rotate to the portrait position at the start of the portrait tests. (V17 systems
only).
6. When the Panel Calibration is complete, Calibration Complete is displayed.
7. Press to display Complete screen and select ShutterCal.
Panel Calibration Failure
If the calibration fails, the following message is displayed: Panel Calibration Processing Failure.
Press to exit. Check that the field of view is clear of all objects and that there are no obstacles
to the rotation of the gantry. Re-run PanelCal. If failure persists, contact i-CAT Technical Support.

Product indications for use at http://www.i-cat.com/customer-care/IFU/

IC-0113 www.i-CAT.com | P 800.205.3570 AWARD-WINNING CONE BEAM 3D DENTAL IMAGING


Quick Reference Guide

Performing the Shutter Calibration


The Shutter Calibration runs several tests in both
landscape and portrait (the panel will rotate)
positions. As with the PanelCal, a pie chart
displays status. Proceed as follows:
1. Select ShutterCal.
2. E
 nsure the field of view on the
scanner is clear.
3. Press . The scanner initializes.
4. W
 hen prompted, press the Scan
button on the operator control box.
An audible alarm is sounded and the
X-ray OK light is illuminated during
radiation exposure.
5. You
 will be prompted to press the
Scan button for each test.
NOTE: The panels will automatically rotate to the portrait position at the start of the portrait tests.
(V17 systems only.)
6. When the Shutter Calibration is complete, Calibration Complete is displayed. Thumbnail images of
each ShutterCal test are available to select and view.
7. Press to display Complete screen and another option.

Shutter Calibration Failure


If the calibration fails, the following message is displayed: “Shutter Calibration Processing Failure”.
Press to exit. Check that the field of view is clear of all objects and that there are no obstacles to
the rotation of the gantry. Re-run ShutterCal. If failure persists, contact i-CAT Technical Support.