3A Analytical Chemistry
Analytical Balance- highly sensitive lab instruments designed to accurately measure mass. Their
readability has a range between 0.1mg - 0.01mg. Analytical balances have a draft shield or weighing
chamber to prevent the very small samples from being affected by air currents.
2. Take a picture of the instrument and label all the parts. (10 pts.)
3. What are the necessary steps and reminders in using the instrument. Name all. (20 pts.)
Check the balance is level using the level indicator (see Figure 1). If the
bubble is not in the center adjust the level, normally by twisting the feet,
• Check that the balance is on and that the door is closed. Press the “Tare”
button and wait 5-10 secs for a ‘*’ or similar symbol to appear in the upper
left/right hand corner of the display, and the mass to read 0.0000 g.
• Open the door and place a weigh boat, weight paper, or other container on
• Close the door and wait for the digital readout to stabilize (‘*’).
• Remove the container from the balance and add the substance to be
weighed. Avoid adding substances on the balance pan as this can result in
• Return container to balance and wait 5-10 secs (may take up to a minute)
Trouble shooting
checked at regular intervals. There are many factors that govern whether an
changes. Also, hot or cold objects can create convection currents in the air
temperature.
weight. This may particularly be an issue for objects that have recently been
during measurement.
• Air flows: Air movement in the laboratory across the pan will cause
variations in the measurement. A draft shield reduces this but it will take
time for the air within the draft shield to stabilize once the door is closed.
Changes in air temperature within the draft shield will also cause air
hands, etc. Reducing air flow incident on the balance in the laboratory is key
to reducing this issue and ensuring all items entering the draft shield are
• Static electricity: This can be one of the biggest causes of frustration when
using a balance. If the mass you are measuring wonders up or down and
will not stabilize then there is a good chance that you have a static issue.
balance. To prevent this you can use an anti-static device which will “fire”
positive and negative ions at the weight boat, powder, etc. to neutralize the
static charge. The good anti-static system are incredibly effective and can
save hours of pain and frustration. Anti-static plastic weigh boats or metal