You are on page 1of 7

en

Anti-CCP
09P27
Anti-CCP Reagent Kit ABBL502R02
B9P270
Read Highlighted Changes: Revised March 2018.

09P2720

Instructions must be carefully followed. Reliability of assay results ll


BIOLOGICAL PRINCIPLES OF THE PROCEDURE
cannot be guaranteed if there are any deviations from these This assay is a two-step immunoassay with an automated sample
instructions. pretreatment for the semi-quantitative determination of the IgG
ll
NAME class of autoantibodies specific to cyclic citrullinated peptide (CCP)
Alinity i Anti-CCP Reagent Kit in human serum or plasma using chemiluminescent microparticle
immunoassay (CMIA) technology.
ll
INTENDED USE Sample and wash buffer are combined. An aliquot of the prediluted
The Alinity i Anti-CCP assay is a chemiluminescent microparticle sample, CCP coated paramagnetic microparticles, and sample
immunoassay (CMIA) used for the semi-quantitative determination of diluent are combined and incubated. The anti-CCP antibodies
the IgG class of autoantibodies specific to cyclic citrullinated peptide present in the sample bind to the CCP coated microparticles. The
(CCP) in human serum or plasma on the Alinity i analyzer. mixture is washed. Anti-human IgG acridinium-labeled conjugate is
The Alinity i Anti-CCP assay is to be used as an aid in the diagnosis added to create a reaction mixture and incubated. Following a wash
of Rheumatoid Arthritis (RA) and should be used in conjunction cycle, Pre-Trigger and Trigger Solutions are added.
with other clinical information. Autoantibody levels represent one The resulting chemiluminescent reaction is measured as relative
parameter in a multicriterion diagnostic process, encompassing both light units (RLUs). There is a direct relationship between the amount
clinical and laboratory-based assessments. of anti-CCP antibody in the sample and the RLUs detected by the
ll
SUMMARY AND EXPLANATION OF THE TEST system optics.
Rheumatoid Arthritis (RA) is a common, systemic autoimmune For additional information on system and assay technology, refer to
disease affecting 0.5-1% of the population. It is characterized by the Alinity ci-series Operations Manual, Section 3.
chronic inflammation of the synovium, which commonly leads to ll
REAGENTS
progressive joint destruction and in most cases, to disability and
reduction of quality of life.1 Evidence gained over the last few years
Kit Contents
suggests that aggressive therapy given early in the disease has the Alinity i Anti-CCP Reagent Kit 09P27
greatest therapeutic potential.2, 3 Volumes (mL) listed in the table below indicate the volume per
The serum of RA patients contains a variety of antibodies cartridge.
directed against self-antigens. The most widely known of these 09P2720
autoantibodies is the rheumatoid factor (RF) antibody directed Tests per cartridge 100
against the constant domain of IgG molecules. The presence Number of cartridges per kit 2
of RF is one of the American College of Rheumatology's (ACR)
Tests per kit 200
criteria for the classification of RA.4 Although the RF test has good
6.6 mL
sensitivity for RA, it is not very specific for the disease as it can
also be detected in the serum of patients with other rheumatic 6.1 mL
or inflammatory diseases and even in a substantial percentage 10.4 mL
of the healthy (elderly) population.5 For several years it has been
recognized that antibodies to anti-perinuclear factor (APF) and  CCP coated microparticles in phosphate buffer with
anti-keratin (AKA) are highly specific for RA. It was subsequently protein (bovine) stabilizer and surfactant. Minimum concentration:
reported that both of these antibodies reacted with native filaggrin 0.05% solids. Preservative: sodium azide.
and are now referred to as anti-filaggrin antibodies (AFA).6-8 More  Mouse anti-human IgG: acridinium-labeled conjugate in
recently it has been shown that all of these antibodies are directed MES buffer with protein (bovine) stabilizer and surfactant. Minimum
to citrulline-containing epitopes.9 Citrulline is a non-standard amino concentration: 10 ng/mL. Preservatives: Nipasept and Sarafloxacin.
acid, as it is not incorporated into proteins during protein synthesis.
It can, however, be generated via post-translational modification of  Phosphate buffer with protein (bovine) stabilizer and
arginine residues by the enzyme peptidyl arginine deiminase (PAD).10 surfactant. Preservative: sodium azide.
In 1998, Schellekens and colleagues reported that linear peptides
containing citrulline (CP) were very specific for RA antibodies (96%) Warnings and Precautions
in an ELISA based assay.11 Subsequent work demonstrated that •
cyclic variants of these peptides, termed cyclic citrullinated peptides • For In Vitro Diagnostic Use
(CCP), were equally specific for RA, but with a higher sensitivity •
than linear peptides.12 To improve the sensitivity of the CCP test Safety Precautions
further, several dedicated libraries of citrulline-containing peptides
CAUTION: This product requires the handling of human specimens.
were screened with RA sera and a new set of peptides (CCP2)
It is recommended that all human-sourced materials be considered
were discovered which gave superior performance compared to the
potentially infectious and handled in accordance with the OSHA
CCP1 test.13 Over the last few years, many independent studies
Standard on Bloodborne Pathogens. Biosafety Level 2 or other
have confirmed the diagnostic performance of the CCP2 test.14, 15 In
appropriate biosafety practices should be used for materials that
2007, the European League against Rheumatism (EULAR) published
contain or are suspected of containing infectious agents.17-20
guidelines for the diagnosis of early RA, and the measurement of
antibodies to anti-CCP was included as a serology marker.16

1
Indications of Reagent Deterioration
The following warnings and precautions apply to:
and Deterioration of the reagents may be indicated when a calibration
error occurs or a control value is out of the specified range.
Contains sodium azide.
Associated test results are invalid, and samples must be retested.
EUH032 Contact with acids liberates very toxic gas.
Assay recalibration may be necessary.
P501 Dispose of contents / container in
For troubleshooting information, refer to the Alinity ci-series
accordance with local regulations.
Operations Manual, Section 10.
Safety Data Sheets are available at www.abbottdiagnostics.com or
contact your local representative.
ll
INSTRUMENT PROCEDURE
The Alinity i Anti-CCP assay file must be installed on the Alinity i
For a detailed discussion of safety precautions during system analyzer prior to performing the assay.
operation, refer to the Alinity ci-series Operations Manual, Section 8.
For detailed information on assay file installation and viewing and
Reagent Handling editing assay parameters, refer to the Alinity ci-series Operations
• Upon receipt, gently invert the unopened reagent kit by rotating Manual, Section 2.
it over and back for a full 180 degrees, 5 times with green label For information on printing assay parameters, refer to the Alinity ci-
stripe facing up and then 5 times with green label stripe facing series Operations Manual, Section 5.
down. This ensures that liquid covers all sides of the bottles For a detailed description of system procedures, refer to the Alinity
within the cartridges. During reagent shipment, microparticles ci-series Operations Manual.
can settle on the reagent septum.
–– Place a check in the square on the reagent kit to indicate to ll
SPECIMEN COLLECTION AND PREPARATION
others that the inversions have been completed. FOR ANALYSIS
• After mixing, place reagent cartridges in an upright position for Specimen Types
1 hour before use to allow bubbles that may have formed to The specimen types listed below were verified for use with this
dissipate. assay.
• If a reagent cartridge is dropped, place in an upright position Other specimen types and collection tube types have not been
for 1 hour before use to allow bubbles that may have formed to verified with this assay.
dissipate. Specimen Types Collection Tubes
• Reagents are susceptible to the formation of foam and bubbles. Serum Serum
Bubbles may interfere with the detection of the reagent level in Serum separator
the cartridge and cause insufficient reagent aspiration that may
Plasma Lithium heparin plasma separator
adversely affect results.
Potassium EDTA
For a detailed discussion of reagent handling precautions during
system operation, refer to the Alinity ci-series Operations Manual, • Performance has not been established for the use of cadaveric
Section 7. specimens or the use of body fluids other than human serum or
Reagent Storage plasma.
• Liquid anticoagulants may have a dilution effect resulting in lower
Storage Maximum Additional Storage
Temperature Storage Time Instructions concentration values for individual specimens.
Unopened 2 to 8°C Until Store in upright position. • Plasma specimens from different anticoagulant tube types
expiration If cartridge does not should not be used interchangeably for monitoring anti-CCP.
date remain upright, gently • The instrument does not provide the capability to verify specimen
invert the cartridge 10 types. It is the responsibility of the operator to verify that the
times and place in an correct specimen types are used in the assay.
upright position for 1 hour Specimen Conditions
before use. • Do not use:
Onboard System 30 days –– heat-inactivated specimens
Temperature –– pooled specimens
Opened 2 to 8°C Until Store in upright position. –– grossly hemolyzed specimens
expiration If cartridge does not –– specimens with obvious microbial contamination
date remain upright during
• For accurate results, serum and plasma specimens should be
storage, discard the
free of fibrin, red blood cells, and other particulate matter. Serum
cartridge.
specimens from patients receiving anticoagulant or thrombolytic
Do not reuse original therapy may contain fibrin due to incomplete clot formation.
reagent caps or
• To prevent cross contamination, use of disposable pipettes or
replacement caps due to
pipette tips is recommended.
the risk of contamination
and the potential to Preparation for Analysis
compromise reagent • Follow the tube manufacturer’s processing instructions for
performance. collection tubes. Gravity separation is not sufficient for specimen
preparation.
Reagents may be stored on or off the system. If removed from the
• Specimens should be free of bubbles. Remove bubbles with an
system, store reagents with new replacement caps in an upright
applicator stick before analysis. Use a new applicator stick for
position at 2 to 8°C. For reagents stored off the system, it is
each specimen to prevent cross-contamination.
recommended that they be stored in their original trays or boxes to
To ensure consistency in results, recentrifuge specimens prior to
ensure they remain upright.
testing if
For information on unloading reagents, refer to the Alinity ci-series
• they contain fibrin, red blood cells, or other particulate matter
Operations Manual, Section 5.
• they require repeat testing.

2
NOTE: If fibrin, red blood cells, or other particulate matter are If testing will be delayed more than 22 hours for specimens stored
observed, mix by low speed vortex or by inverting 10 times prior to at room temperature or more than 7 days for specimens stored at
recentrifugation. 2 to 8°C, remove serum or plasma from the clot, red blood cells, or
Prepare frozen specimens as follows: separator gel and store at -20°C or colder.
• Frozen specimens must be completely thawed before mixing. Avoid more than three freeze/thaw cycles.
• Mix thawed specimens thoroughly by low speed vortex or by Specimen Shipping
inverting 10 times. Package and label specimens in compliance with applicable state,
• Visually inspect the specimens. If layering or stratification is federal, and international regulations covering the transport of clinical
observed, mix until specimens are visibly homogeneous. specimens and infectious substances.
• If specimens are not mixed thoroughly, inconsistent results may
be obtained.
ll
PROCEDURE
• Recentrifuge specimens. Materials Provided
Recentrifugation of Specimens 09P27 Alinity i Anti-CCP Reagent Kit
• Transfer specimens to a centrifuge tube and centrifuge at Materials Required but not Provided
100 000 g-minutes. • Alinity i Anti-CCP assay file
• Examples of acceptable time and force ranges that meet this • 09P2701 Alinity i Anti-CCP Calibrators
criterion are listed in the table below. • 09P2710 Alinity i Anti-CCP Controls or other control material
Centrifugation time using alternate RCF values can be calculated • Alinity Trigger Solution
using the following formula: • Alinity Pre-Trigger Solution
100 000 g-minutes • Alinity i-series Concentrated Wash Buffer
Minimum Centrifugation time (minutes) =
RCF For information on materials required for operation of the instrument,
Recentrifugation Time refer to the Alinity ci-series Operations Manual, Section 1.
(Minutes) RCF (x g)* g-Minutes For information on materials required for maintenance procedures,
10 10 000 100 000 refer to the Alinity ci-series Operations Manual, Section 9.
20 5000 100 000
Assay Procedure
40 2500 100 000
For a detailed description of how to run an assay, refer to the Alinity
* To ensure consistency in results, specimens must be centrifuged ci-series Operations Manual, Section 5.
using an appropriate tube at a minimum of 2500 RCF to obtain a • If using primary or aliquot tubes, refer to the Alinity ci-series
minimum of 100 000 g-minutes. Operations Manual, Section 4 to ensure sufficient specimen is
RCF = 1.12 × rmax (rpm/1000)2 present.
RCF - The relative centrifugal force generated during • To minimize the effects of evaporation, verify adequate sample
centrifugation. cup volume is present prior to running the test.
rpm - The revolutions per minute of the rotor on which • Maximum number of replicates sampled from the same sample
the specimens are being spun (usually the digital cup: 10
readout on the centrifuge will indicate the rpm). –– Priority:
Centrifugation The time should be measured from the time the ◦◦ Sample volume for first test: 60 µL
Time - rotor reaches the required RCF or rpm to the time it ◦◦ Sample volume for each additional test from same sample
begins decelerating. cup: 10 µL
rmax - Radius of the rotor in millimeters. NOTE: If custom –– ≤ 3 hours on the reagent and sample manager:
tube adapters (i.e., adapters not defined by the
◦◦ Sample volume for first test: 150 µL
centrifuge manufacturer) are used, then the radius
(rmax) should be manually measured in millimeters ◦◦ Sample volume for each additional test from same sample
and the RCF calculated. cup: 10 µL
g-minutes - The unit of measure for the product of RCF (× g) –– > 3 hours on the reagent and sample manager:
and centrifugation time (minutes). ◦◦ Replace with a fresh aliquot of sample.
• Refer to the Alinity i Anti-CCP calibrator package insert and/
• Transfer clarified specimen to a sample cup or secondary tube or Alinity i Anti-CCP control package insert for preparation and
for testing. For centrifuged specimens with a lipid layer, transfer usage.
only the clarified specimen and not the lipemic material.
• For general operating procedures, refer to the Alinity ci-series
Specimen Storage Operations Manual, Section 5.
Maximum • For optimal performance, it is important to perform routine
Specimen Storage maintenance as described in the Alinity ci-series Operations
Type Temperature Time Special Instructions
Manual, Section 9. Perform maintenance more frequently when
Serum/ Room 22 hours Specimens may be required by laboratory procedures.
Plasma temperature stored on or off the
(study clot, red blood cells, or Sample Dilution Procedures
performed at separator gel. Samples with anti-CCP value exceeding 195.6 U/mL are flagged
30°C) with the code "> 195.6 U/mL" and may be diluted using either the
Automated Dilution Protocol or the Manual Dilution Procedure.
2 to 8°C 7 days Specimens may be
Automated Dilution Protocol
stored on or off the
clot, red blood cells, or The system performs a 1:6 dilution of the sample and automatically
separator gel. calculates the concentration by multiplying the result by the dilution
factor.

3
Manual Dilution Procedure • If quality control results do not meet the acceptance criteria
Suggested dilution: 1:10 defined by your laboratory, sample results may be suspect.
Add 50 μL of the sample to 450 μL of Alinity i Anti-CCP Negative Follow the established quality control procedures for your
Control. laboratory. Recalibration may be necessary. For troubleshooting
Avoid contamination of Negative Control when transferring volume for information, refer to the Alinity ci-series Operations Manual,
manual dilution. Section 10.
Refer to the Alinity i Anti-CCP control package insert for preparation • Review quality control results and acceptance criteria following a
and storage. change of reagent or calibrator lot.
The operator must enter the dilution factor in the Specimen or Quality Control Guidance
Control tab of the Create Order screen. The system will use this Refer to “Basic QC Practices” by James O Westgard, Ph.D. for
dilution factor to automatically calculate the concentration of the guidance on laboratory quality control practices.22
sample and report the result. Verification of Assay Claims
If the operator does not enter the dilution factor, the result must be For protocols to verify package insert claims, refer to Verification of
manually multiplied by the appropriate dilution factor before reporting Assay Claims in the Alinity ci-series Operations Manual.
the result. If a diluted sample result is less than the lower value of
the measuring interval of 0.5 U/mL, do not report the result. Rerun
ll
RESULTS
using an appropriate dilution. Calculation
For detailed information on ordering dilutions, refer to the Alinity ci- The Alinity i Anti-CCP assay utilizes a 4 Parameter Logistic Curve fit
series Operations Manual, Section 5. data reduction method (4PLC, Y-weighted) to generate a calibration
and results.
Calibration
For instructions on performing a calibration, refer to the Alinity ci- Flags
series Operations Manual, Section 5. Some results may contain information in the Flags field. For a
Each assay control must be tested to evaluate the assay calibration. description of the flags that may appear in this field, refer to the
Once a calibration is accepted and stored, all subsequent samples Alinity ci-series Operations Manual, Section 5.
may be tested without further calibration unless: Measuring Interval
• A reagent kit with a new lot number is used. Measuring interval is defined as the range of values in U/mL which
• Daily quality control results are outside of statistically-based meets the limits of acceptable performance for linearity, imprecision,
quality control limits used to monitor and control system and bias.
performance, as described in the Quality Control Procedures The measuring interval of the Alinity i Anti-CCP assay is 0.5 to
section of this package insert. 195.6 U/mL.
–– If statistically-based quality control limits are not available,
then the calibration should not exceed a 30-day limit for
ll
LIMITATIONS OF THE PROCEDURE
• For diagnostic purposes, the Alinity i Anti-CCP results should
recalibration frequency.
be used in conjunction with other clinical data; e.g., symptoms,
This assay may require recalibration after maintenance to critical medical history, etc.
parts or subsystems or after service procedures have been
• If the Alinity i Anti-CCP results are inconsistent with clinical
performed.
evidence, additional testing is suggested to confirm the result.
Quality Control Procedures • The value of anti-CCP in juvenile arthritis has not been
The recommended control requirement for the Alinity i Anti-CCP determined.
assay is that a single sample of each control level be tested once • Some specimens may not dilute linearly because of
every 24 hours each day of use. the heterogeneity of the autoantibodies with respect to
Additional controls may be tested in accordance with local, state, physiochemical properties.
and/or federal regulations or accreditation requirements and your • Alinity i Anti-CCP results should not be used interchangeably
laboratory’s quality control policy. with other manufacturers' methods for anti-CCP
To establish statistically-based control limits, each laboratory should determinations.
establish its own concentration target and ranges for new control lots • Plasma specimens from different anticoagulant tube types
at each clinically relevant control level. This can be accomplished should not be used interchangeably for monitoring anti-CCP.
by assaying a minimum of 20 replicates over several (3-5) days and • Specimens from patients who have received preparations of
using the reported results to establish the expected average (target) mouse monoclonal antibodies for diagnosis or therapy may
and variability about this average (range) for the laboratory. Sources contain human anti-mouse antibodies (HAMA).23, 24 Specimens
of variation that should be included in this study in order to be containing HAMA may produce anomalous values when tested
representative of future system performance include: with assay kits such as Alinity i Anti-CCP that employ mouse
• Multiple stored calibrations monoclonal antibodies.23
• Multiple reagent lots • Heterophilic antibodies in human specimens can react
• Multiple calibrator lots with reagent immunoglobulins, interfering with in vitro
• Multiple processing modules (if applicable) immunoassays.25 Patients routinely exposed to animals or to
• Data points collected at different times of the day animal serum products can be prone to this interference and
Refer to published guidelines for information or general control anomalous values may be observed. Additional information may
recommendation, for example Clinical and Laboratory Standards be required for diagnosis.
Institute (CLSI) Document C24-A3 or other published guidelines, for • Refer to the SPECIMEN COLLECTION AND PREPARATION FOR
general quality control recommendations.21 ANALYSIS section of this package insert for specimen limitations.
• If more frequent control monitoring is required, follow the
established quality control procedures for your laboratory.

4
ll
EXPECTED VALUES Analytical Specificity
This study was performed on the ARCHITECT i System. This study was performed on the ARCHITECT i System.
Representative performance data are provided in this section. To assess the potential cross-reactivity of the CCP antigen used
Results obtained in individual laboratories may vary. in the ARCHITECT Anti-CCP assay with other autoantibodies, the
It is recommended that each laboratory determine its own reference assay was evaluated with 20 samples positive for various other
range based upon its particular locale and population characteristics. autoantibodies and negative for CCP antibodies. The following
In a representative study, serum specimens from 199 asymptomatic, autoantibodies (1-4 samples of each) were tested in the assay: SSA,
apparently healthy males (n = 126) and females (n = 73), with SSB, Sm, RNP, ds-DNA, Jo-1, Scl-70, Ribo-P, TPO, ANA, and AMA.
an age range of 19 to 67 years, were tested with the ARCHITECT The study showed no significant cross-reactivity of the CCP antigen
Anti-CCP assay. No differences attributable to gender or age were with any of these other autoantibodies.
observed. Specimen values ranged from < 0.5 U/mL to 2.5 U/mL. Interference
A cut-off of 5.0 U/mL was chosen, whereby a result of ≥ 5.0 U/mL This study was performed on the ARCHITECT i System.
is considered positive and a result of < 5.0 U/mL is considered Potentially Interfering Endogenous Substances
negative. A study was performed based on guidance from CLSI EP07-A2.29
ll
SPECIFIC PERFORMANCE CHARACTERISTICS Serum samples with anti-CCP levels across the assay range of
Representative performance data are provided in this section. 0.5 U/mL to 200.0 U/mL were supplemented with the potentially
Results obtained in individual laboratories may vary. interfering compounds listed in the table below. The maximum
The Alinity i analyzer and the ARCHITECT i System utilize the same deviation of anti-CCP concentration observed in serum samples
reagents and sample/reagent ratios. during these studies ranged from:
Unless otherwise specified, all studies were performed on the Alinity • -7.6% to 0.8% for anti-CCP concentrations > 10.0 U/mL
i analyzer. • -1.0% to 7.5% for anti-CCP concentrations ≥ 5.0 U/mL to
≤ 10.0 U/mL
Precision
• -0.3 U/mL to 0.2 U/mL for anti-CCP concentrations < 5.0 U/mL
Within-Laboratory Precision
Potentially Interfering Substance Interferent Level
A study was performed based on guidance from CLSI EP05-A2.26
Bilirubin 20 mg/dL
Testing was conducted using 1 lot of the Alinity i Anti-CCP Reagent
Hemoglobin 800 mg/dL
Kit, 1 lot of the Alinity i Anti-CCP Calibrators, and 1 lot of the Alinity
i Anti-CCP Controls and 1 instrument. One control and 6 human Total Protein 12 g/dL
serum panels were assayed in a minimum of 2 replicates at 2 Triglycerides 3000 mg/dL
separate times per day on 20 different days. Rheumatoid Factor 200 IU/mL
Within-Run Within-Laboratory Tube Type Matrix Comparison
(Repeatability) (Total)a
Sample n Mean (U/mL) SD %CV SD %CV
This study was performed on the ARCHITECT i System.
Positive Control 80 21.5 0.41 1.9 0.45 2.1 The specimen collection tubes listed below were verified for use with
Panel 1 80 10.1 0.16 1.5 0.18 1.8 the ARCHITECT Anti-CCP assay.
Panel 2 80 30.2 0.40 1.3 0.70 2.3 • serum, serum separator, lithium heparin plasma separator, and
Panel 3 80 66.7 1.18 1.8 1.56 2.3
potassium EDTA.
Panel 4 80 138.6 3.90 2.8 5.67 4.1 When compared to the control tube type (serum), the tube types
Panel 5 80 3.0 0.06 1.9 0.06 2.1
evaluated for samples with anti-CCP values < 5.0 U/mL showed less
than a 0.5 U/mL difference on average, and the tube types evaluated
Panel 6 80 183.4 8.83 4.8 9.84 5.4
for samples with anti-CCP values ranging from 5.3 to 178.8 U/mL
a Includes within-run, between-run, and between-day variability. showed less than a 10% difference on average. The distribution of
Lower Limits of Measurement the differences or percent differences per tube type is listed in the
following table.
A study was performed based on guidance from CLSI EP17-A2.27
Testing was conducted using 3 lots of the Alinity i Anti-CCP Reagent Distribution of Absolute Distribution of Absolute Percent
Differences < 0.5 U/mL Differences for Samples with Anti-CCP
Kit on each of 2 instruments over a minimum of 3 days. The Values 5.3 to 178.8 U/mL
for Samples with Anti-
maximum observed Limit of Blank (LoB), Limit of Detection (LoD), Tube Type CCP Values < 5.0 U/mL < 10% ≥ 10% to ≤ 20% > 20%
and Limit of Quantitation (LoQ) values are summarized below.
Serum Separator 100% 76% 16% 8%
U/mL (19/19) (19/25) (4/25) (2/25)
LoBa 0.1 Potassium EDTA 100% 72% 24% 4%
LoDb 0.3 (19/19) (18/25) (6/25) (1/25)
LoQc 0.4 Lithium Heparin 100% 80% 16% 4%
a Plasma (19/19) (20/25) (4/25) (1/25)
The LoB represents the 95th percentile from n ≥ 60 replicates of Separator
zero-analyte samples.
b The LoD represents the lowest concentration at which the analyte
Method Comparison
can be detected with 95% probability based on n ≥ 60 replicates of A study was performed based on guidance from CLSI EP09-A330
low-analyte level samples. using the Passing-Bablok regression method.
c The LoQ was determined from n ≥ 60 replicates of low-analyte
Correlation Concentration
level samples and is defined as the lowest concentration at which a Units n Coefficient Intercept Slope Range
maximum allowable precision of 20 %CV was met. Alinity i Serum U/mL 120 0.99 0.08 0.94 0.5-183.1
Linearity Anti-CCP vs
ARCHITECT
A study was performed based on guidance from CLSI EP06-A.28 Anti-CCP
This assay is linear across the measuring interval of 0.5 to
195.6 U/mL.

5
Negative and Positive Percent Agreement ARCHITECT Anti-CCP
Cut-off Specimen Category Total n Positive n % Specificity
Using a cut-off of 5.0 U/mL for the ARCHITECT Anti-CCP assay, the Non-RA Specimens 499 9 98.2
concordance was calculated to be 99.3%. in Total
Percent Agreement Non-RA Healthy 200 1 99.5
Asymptomatic
A study was performed based on guidance from CLSI EP12-A2.31 Non-RA Disease 299 8 97.3
ARCHITECT Anti-CCP Specimensa
Positive Negative aThe non-RA diseases were Ankylosing Spondylitis, Autoimmune
Alinity i Anti-CCP Positive 119 0
Thyroiditis/Hashimoto's Disease, Crohn's Disease, Dermatomyositis,
Negative 0 108
Epstein-Barr Virus, Lyme Disease, Osteoarthritis, Polymyalgia
Positive % agreement = 100.00% (119/119); 95% confidence interval Rheumatica, Polymyositis, Psoriatic Arthritis, Reactive Arthritis/
(96.95 - 100.00%) Reiter's Syndrome, Scleroderma, Sjögren's Syndrome, Systemic
Negative % agreement = 100.00% (108/108); 95% confidence interval Lupus Erythematosus, and Ulcerative Colitis.
(96.64 - 100.00%) ll
BIBLIOGRAPHY
Receiver Operator Characteristic (ROC) Analysis 1. Feldmann M, Brennan FM, Maini RN. Rheumatoid arthritis. Cell
A ROC analysis was carried out using ARCHITECT Anti-CCP data. 1996;85:307-310
The area under the curve (AUC) for the ARCHITECT Anti-CCP assay 2. Landewé RB. The benefits of early treatment in rheumatoid arthritis:
was 0.873 (95% confidence interval: 0.849-0.897). The ROC analysis confounding by indication, and the issue of timing. Arthritis Rheum
2003;48(1):1-5.
curve is shown below. 3. Lard LR, Visser H, Speyer I, et al. Early versus delayed treatment
in patients with recent-onset rheumatoid arthritis: comparison of
two cohorts who received different treatment strategies. Am J Med
2001;111:446-451.
4. Arnett FC, Edworthy SM, Bloch DA, et al. The American Rheumatism
Association 1987 revised criteria for the classification of rheumatoid
arthritis. Arthritis Rheum 1988;31(3):315-324.
5. van Venrooij WJ, Hazes JM, Visser H. Anticitrullinated protein/
peptide antibody and its role in the diagnosis and prognosis of early
rheumatoid arthritis. Neth J Med 2002;60(10):383-388.
6. Nienhuis RL, Mandema E, Smids C. A new serum factor in patients
with rheumatoid arthritis. The antiperinuclear factor. Ann Rheum Dis
1964;23:302-305.
7. Young BJ, Mallya RK, Leslie RD, et al. Anti-keratin antibodies in
rheumatoid arthritis. Br Med J 1979;2:97-99.
8. Hoet RM, Boerbooms AM, Arends M, et al. Antiperinuclear factor, a
marker autoantibody for rheumatoid arthritis: colocalisation of the
perinuclear factor and profilaggrin. Ann Rheum Dis 1991;50:611-618.
9. Sebbag M, Simon M, Vincent C, et al. The antiperinuclear factor
and the so-called antikeratin antibodies are the same rheumatoid
arthritis-specific autoantibodies. J Clin Invest 1995;95:2672-2679.
10. Vossenaar ER, Zendman AJ, van Venrooij WJ, et al. PAD, a growing
family of citrullinating enzymes: genes, features and involvement in
x 1 - Specificity (false positives) No discrimination disease. BioEssays 2003;25:1106-1118.
11. Schellekens GA, de Jong BA, van den Hoogen FH, et al. Citrulline
y Sensitivity (true positives) ARCHITECT is an essential constituent of antigenic determinants recognized
by rheumatoid arthritis-specific autoantibodies. J Clin Invest
High Dose Hook 1998;101(1):273-281.
This study was performed on the ARCHITECT i System. 12. Schellekens GA, Visser H, de Jong BA, et al. The diagnostic
High dose hook is a phenomenon whereby very high level specimens properties of rheumatoid arthritis antibodies recognizing a cyclic
may read within the dynamic range of the assay. No high dose citrullinated peptide. Arthritis Rheum 2000;43(1):155-163.
13. Vossenaar ER, van Venrooij WJ. Anti-CCP antibodies, a highly
hook effect was observed when a sample containing approximately
specific marker for (early) rheumatoid arthritis. Clin Applied Immunol
2000 U/mL of anti-CCP antibody was assayed. Rev 2004;4:239-262.
Clinical Sensitivity and Specificity 14. Pruijn GJ, Vossenaar ER, Drijfhout JW, et al. Anti-CCP antibody
This study was performed on the ARCHITECT i System. detection facilitates early diagnosis and prognosis of rheumatoid
arthritis. Current Rheumatology Reviews 2005;1(1):1-7.
The clinical sensitivity was determined for 496 confirmed RA 15. Nishimura K, Sugiyama D, Kogata Y, et al. Meta-analysis: diagnostic
individuals, and clinical specificity was determined for 499 non- accuracy of anti-cyclic citrullinated peptide antibody and rheumatoid
RA specimens (299 from patients with other rheumatic and factor for rheumatoid arthritis. Ann Intern Med 2007;146(11):797-808.
non-rheumatic disorders and 200 from asymptomatic apparently 16. Combe B, Landewe R, Lukas C, et al. EULAR recommendations
healthy individuals). Using a cut-off of 5.0 U/mL, the sensitivity was for the management of early arthritis: report of a task force of the
calculated to be 70.6% with a specificity of 98.2%. The results are European Standing Committee for International Clinical Studies
Including Therapeutics (ESCISIT). Ann Rheum Dis 2007;66:34-45.
summarized in the following tables.
17. US Department of Labor, Occupational Safety and Health
ARCHITECT Anti-CCP Administration, 29 CFR Part 1910.1030, Bloodborne pathogens.
Specimen Category Total n Positive n % Sensitivity 18. US Department of Health and Human Services. Biosafety in
Confirmed RAa 496 350 70.6 Microbiological and Biomedical Laboratories. 5th ed. Washington, DC:
US Government Printing Office; December 2009.
a RA patients were classified according to the ACR Criteria.4 19. World Health Organization. Laboratory Biosafety Manual. 3rd ed.
Geneva: World Health Organization; 2004.

6
20. Clinical and Laboratory Standards Institute (CLSI). Protection
of Laboratory Workers From Occupationally Acquired Infections;
ll
Key to Symbols
Approved Guideline—Fourth Edition. CLSI Document M29-A4. Wayne, ISO 15223 Symbols
PA: CLSI; 2014. Consult instructions for use
21. Clinical and Laboratory Standards Institute (CLSI). Statistical Quality
Control for Quantitative Measurement Procedures: Principles and Manufacturer
Definitions; Approved Guideline—Third Edition. CLSI Document
C24-A3. Wayne, PA: CLSI; 2006.
22. Westgard JO. Basic QC Practices. 3rd ed. Madison, WI: Westgard Sufficient for
Quality Corporation; 2010.
23. Primus FJ, Kelley EA, Hansen HJ, et al. “Sandwich”-type Temperature limitation
immunoassay of carcinoembryonic antigen in patients receiving
murine monoclonal antibodies for diagnosis and therapy. Clin Chem
1988;34(2):261-264. Use by/Expiration date
24. Schroff RW, Foon KA, Beatty SM, et al. Human anti-murine
immunoglobulin responses in patients receiving monoclonal antibody
therapy. Cancer Res 1985;45(2):879-885. In Vitro Diagnostic Medical
25. Boscato LM, Stuart MC. Heterophilic antibodies: a problem for all Device
immunoassays. Clin Chem 1988;34(1):27-33.
Lot Number
26. Clinical and Laboratory Standards Institute (CLSI). Evaluation of
Precision Performance of Quantitative Measurement Methods; List Number
Approved Guideline—Second Edition. CLSI Document EP05-A2.
Wayne, PA: CLSI; 2004. Serial number
27. Clinical and Laboratory Standards Institute (CLSI). Evaluation of
Other Symbols
Detection Capability for Clinical Measurements; Approved Guideline–
Second Edition. CLSI Document EP17-A2. Wayne, PA: CLSI; 2012. Conjugate
28. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Contains Sodium Azide. Contact
the Linearity of Quantitative Measurement Procedures: A Statistical with acids liberates very toxic
Approach; Approved Guideline. CLSI Document EP06-A. Wayne, PA:
CLSI; 2003. gas.
29. Clinical and Laboratory Standards Institute (CLSI). Interference Distributed in the USA by
Testing in Clinical Chemistry; Approved Guideline—Second Edition. Information needed for United
CLSI Document EP7-A2. Wayne, PA: CLSI; 2005.
States of America only
30. Clinical and Laboratory Standards Institute (CLSI). Measurement
Procedure Comparison and Bias Estimation Using Patient Samples; Inversions Performed
Approved Guideline—Third Edition. CLSI Document EP09-A3. Wayne, Microparticles
PA: CLSI; 2013.
31. Clinical and Laboratory Standards Institute (CLSI). User Protocol for Produced for Abbott by
Evaluation of Qualitative Test Performance; Approved Guideline—
Product of Great Britain
Second Edition. CLSI Document EP12-A2. Wayne, PA: CLSI; 2008.
Note for number formatting: For use by or on the order of a
• A space is used as thousands separator (example: 10 000 physician only (applicable to USA
specimens). classification only).
• A period is used to separate the integer part from the fractional Sample Diluent
part of a number written in decimal form (example: 3.12%).
Alinity and ARCHITECT are trademarks of Abbott Laboratories in
various jurisdictions. All other trademarks are property of their
respective owners.

Abbott GmbH & Co. KG


Max-Planck-Ring 2
65205 Wiesbaden
Germany
+49-6122-580

Axis-Shield Diagnostics Limited Abbott Laboratories


Luna Place, The Technology Park, Abbott Park, IL 60064 USA
Dundee DD2 1XA
United Kingdom
Customer Service: Contact your local representative
or find country-specific contact information on
www.abbottdiagnostics.com
Revised March 2018.
©2017, 2018 Abbott Laboratories

You might also like