SERVICE MANUAL
707
UNIT

MK serial changer rel. 1. 3. 7

4512 202 00781

DMC Hamburg

Printed in Hamburg, Federal Republic of Germany


ã 2002 Philips Medical Systems
ALL RIGHTS RESERVED

4512 984 25831 1


25831_titel
Mk serial changer rel. 1. 3. 7.

SERVICE MANUAL - UNIT

MK serial changer rel. 1. 3. 7 Author: A. Ehlers

Type No: 4512 202 00781

In case there are any questions concerning this manual,


please send this LOPAD via fax to 49/(0)40/5078 2481

File: MK serial changer rel. 137_25831

List of pages and drawings (LOPAD) Manual Order No: 4512 984 25831
released: 12/20001

0.5 223 mm (Rosa Karton)

1
2

3... 26 (02.0)
_________________________________________
P--List 4512 202 00781
_________________________________________ Z1--5.1 (b/01.0) Duo DIAGNOST
Z1--6.1 (b/01.0) Duo DIAGNOST
Z1--2 (a/01.0) Serial changer DD
Z1--7.1 (b/01.0) Duo DIAGNOST
Z1--0.2 (a/01.0) Serial changer DD
_________________________________________ Z1--8.1 (b/01.0) Duo DIAGNOST
Z1--9.2 (b/01.0) Duo DIAGNOST
Z0--1 (b/01.0) Duo DIAGNOST _________________________________________
Z1--0.1 (c/01.0) Duo DIAGNOST
Z0--2 (b/01.0) Duo DIAGNOST
Z1--0.2 (d/01.0) Duo DIAGNOST
Z2--0.2 (b/01.0) Duo DIAGNOST
Z1--0.8 (01.0) Duo DIAGNOST
Z2--0.4 (c/01.0) Duo DIAGNOST
Z1--1.1 (d/01.0) Duo DIAGNOST
Z2--1 (c/01.0) Duo DIAGNOST
Z1--1.2 (c/01.0) Duo DIAGNOST
Z2--2.1 (b/01.0) Duo DIAGNOST
Z1--1.3 (c/01.0) Duo DIAGNOST
Z2--2.4 (01.0) Duo DIAGNOST
Z1--1.4 (c/01.0) Duo DIAGNOST
_________________________________________
Z1--1.5 (c/01.0) Duo DIAGNOST
Z1--2.1 (c/01.0) Duo DIAGNOST Z1--0.1 (c/01.0) Central Part X
Z1--3.1 (b/01.0) Duo DIAGNOST Z2--0.2 (b/01.0) Central Part X
_________________________________________
Z1--4.1 (b/01.0) Duo DIAGNOST

ã 2002 Philips Medical Systems


2 ALL RIGHTS RESERVED 4512 984 25831
25831_lopad
MK serial changer rel. 1. 3. 7

TEXT

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1. Preparatory work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Removal of the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3. Installation of the DC power supply unit SC2 for the serial changer . . . 5
4. Removal of the serial changer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
5. Exchange of the main board D80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
6. Exchange of the Opto2 board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7. Exchange of the tensioning mechanism of the grid frame . . . . . . . . . . . . . 14
8. Exchange of the flat cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
9. Check and adjustment of the play of the two guideways of the I/O drive 19
10. Check and adjustment of the angle of the two cassette holders . . . . . . . . 20
11. Check and adjustment of the collimators close to the film . . . . . . . . . . . . . 22
12. Finishing work and installation of the serial changer . . . . . . . . . . . . . . . . . 23

Note
After the installation of this FCO the serial changer has the new code number 4512 202 0078.

Introduction:
The major modification on the serial changer concerns the voltage supply.
It is changed over from AC 28V to DC 36V. After the serial changer is modified, it must be operated only with the
new DC power supply SC2 (installed in the geometry control unit and part of this FCO).

MK serial changer (02.0) 3


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

1. Preparatory work
S Switch the system ON.
S Select exposure technique Bucky.
S Tilt the tabletop into the 0° position.
S Move the Bucky unit in longitudinal direction into the
center of the tabletop.
S Move the tabletop in transverse direction into the
middle position.
S Switch the system OFF.

2. Removal of the tabletop

Note
A second person is needed to remove the tabletop!

2.1. Uncoupling the tabletop from the drive

S Remove the two square cover panels of the


sectional rail of the table at the column side at the
head end and the foot end of the tabletop
(the illustration shows the cover plate at the head
end of the tabletop).

S Remove the two screws M6 of the drive at the head


end of the table.

Note
Before removing the last two screws secure the tabletop
with the aid of a second person so that it cannot roll off
the table base!
S Remove the two screws M6 of the drive at the foot end
of the table.

2.2. Remove the tabletop by pushing it laterally


from the table base with the aid of a second
person.

2.3. Position the tabletop at a safe location and


make sure that it cannot tip over.

4 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

3. Installation of the DC power supply unit SC2 for the serial changer

Note
After the DC power supply unit is installed, the system must be switched on only when plug VAX4 at the serial
changer is pulled off. Only after the D80 board is exchanged in the serial changer, the serial changer can be
operated again on the DC power supply unit SC2.

3.1. Remove the cover of the housing of the


SC unit at the base of the unit.

3.2. Pull off plugs SC X31 and SC X35.

3.3. Remove the two upper outer screws and the


lower outer screw M5 from the PCB housing.

3.4. Install the SC2 unit and fix it with the three
screws M5.

3.5. Attach the new target designation labels on


the two cables:
Formerly SC X31 becomes SC2 X1 and
formerly SC X35 becomes SC2 X3.

3.6. Connect the earth conductor of the SC 2 unit


to earth point SCX50.

3.7. Place the two plugs with the new target


designations SCX1 and SCX3.

3.8. Connect SC2 X2 with SC X31 and SC2 X4


with SCX35.

3.9. Install the cover of the housing of the SC unit


at the base of the unit.

MK serial changer (02.0) 5


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

4. Removal of the serial changer

4.1. Remove the cover plate (about 2.2m long) at


the column side.

4.2. Remove the cover plate (about 2.2m long) at


the operating side.

4.3. Remove the two anti-scatter collimators


screwed on at the two sides of the Bucky
frame.

4.4. Remove the mushroom-colored narrow cover


screwed on at the operating side of the serial
changer at the Bucky frame.

4.5. Unscrew the front panel (with green LED and


I/0 button) and pull off the plug from the small
D83--1 board (if present).

6 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

4.6. Mark the lateral installation location of the Marking of


serial changer in the Bucky frame with a thin “lateral position”
pencil.

Note
In case it is not possible to detect the lateral installation
location of the serial changer in the Bucky frame when
the serial changer is re-installed, this location must be
determined again under radiation by means of a
centering cassette!

4.7. Cancel all electrical connections of the serial


changer:
-- VA X4 power supply
-- VA X2 CAN connection
-- VA X1 radiation release
-- Earth 2 times

4.8. Remove the four screws fixing the serial


changer in the Bucky frame
(two screws each on each side).

4.9. Remove the mechanical coupling between the


motorized collimators and the drive.

MK serial changer (02.0) 7


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

4.10. Open the two collimators manually fully up to


the stop.

4.11. Remove the left-hand lateral retaining angle


(two nuts M3) of the anti-scatter grid.

4.12. Loosen the two screws (two nuts M3) of the


right-hand lateral retaining angle. 1. Loosen 2 nuts M3

3. Remove retaining angle

2. Unscrew 2 nuts M3

4.13. Lift the serial changer by about 5cm and push


out the anti-scatter grid to the left through a
slit in the serial changer, carefully lower the
serial changer back into position.

8 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

4.14. Remove the protective plate attached at the


operating side inside the serial changer for
the Amplimat measuring chamber
(6x countersunk screw M3).

4.15. Pull out the Amplimat measuring chamber to


the operating side (by about 15cm) until the
Amplimat connector becomes visible.

4.16. Unscrew the Amplimat connector and


separate it from the measuring chamber.

4.17. Pull the Amplimat measuring chamber plug


fully to the rear out of the serial changer and
the Bucky frame and let it hang down.

4.18. Pull out the Amplimat measuring chamber fully


out towards the operating side and position it
at a safe location.

Note
Remove the four plastic guide nipples inserted laterally
in the Amplimat measuring chamber and discard them
-- they are replaced by new ones when the measuring
chamber is re--installed!

MK serial changer (02.0) 9


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

5. Exchange of the main board D80

5.1. Remove the mushroom-colored cover of the


main board D80
(3 x Phillips screw M4).

Remove 3x Phillips screws M4

5.2. Prepare the antistatic--kit for use.

5.3. Pull off the following plugs:

-- X11, X12, X13, X17 and X19


-- Flat cable X5

5.4. Remove all fixing screws of the D80 board


(4x countersunk screws M3 at the side and Remove 4x countersunk screws M3
3x Phillips screws M4 at the soldering side)
>Ill.: Remove_D80_old<.

Remove 3x Phillips screws M4

5.5. Pull off connector X16 (cable going to the


collimator drive motor SM).

5.6. Remove the termination resistor for the CAN


from X3 and install it on the new D80 board.

Note
Applicable only if the new D80 board is not equipped with
a termination resistor at X3.

10 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

5.7. Place connector X16 including the extension


cable supplied (about 15 cm long).

Note
The extension cable must under all circumstances be
installed when the socket X16 is located on the soldering
side of the D80 board.
If the socket X16 is located on the component side, the
extension cable must not be installed.

5.8. Fasten the D80 board on the serial changer


(4x countersunk screws M3 at the side and
3x Phillips screws M4 on the soldering side).

5.9. Place the following plugs:

-- X11, X12, X13, X17 and X19


-- Flat cable X5
(placing is not necessary if the flat cable still
needs to be exchanged, see chapter “Exchange
of the flat cable“).

MK serial changer (02.0) 11


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

6. Exchange of the Opto2 board

Note
From serial number 336 of the serial changer onwards
an exchange of the Opto2 board is not necessary since
the most recent version should already be installed. The
most recent version has a metal cover that reaches over
the complete length of the board -- this can easily be
recognized when the serial changer is removed.

Opto 2 without metal cover:


Exchange is necessary!

Opto 2 with metal cover:


Exchange is not necessary!

12 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

6.1. Pull off plug X19 going to the Opto2 board on


main board D80.

6.2. Remove all tie-wraps of this lead.

6.3. Remove the two countersunk screws M3.

2x countersunk screws M3

6.4. Remove the Opto 2 board including cable.

6.5. Insert the new Opto 2 board and screw it on.

6.6. Route the connection cable going to the main


board D80.

6.7. Connect plug X19 with the main board D80.

6.8. Fix the cable with the tie-wraps supplied.

6.9. Install the mushroom-colored cover of the


main board D80
(3 x Phillips screws M4).

MK serial changer (02.0) 13


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

7. Exchange of the tensioning


mechanism of the grid frame

Note
From serial number 370 of the serial changer onwards
a modification of the tensioning mechanism is not
necessary, since the most recent version should already
be installed. The most recent version has an additional
recess on the tension lever and a magnet holder which
is narrower than that of the older version (about 15mm). New

Old

7.1. Remove the two countersunk screws M3 on


the base plate of the tensioning mechanism.

Remove 2x countersunk screws M3

7.2. Remove the gear motor from the base plate.

7.3. Remove the tension lever from the axle.

7.4. Unscrew the magnet holder from the grid


frame (2 x Phillips screws M4).

Note
When loosening the screws a relatively high force must
be exerted on each of the two Phillips screws. This force
must under all circumstances be compensated by, for
instance, a screwdriver placed under the magnet holder
for otherwise the oscillating springs of the grid frame
would be damaged.

14 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

7.5. Install the new magnet holder on the grid


frame.

7.6. During oscillation the magnet holder must


pass the Hall sensor centrically on the
D82 board; if necessary, bend the magnet
holder respectively.
A
7.7. Distance A between the magnet holder and
the coil form of the electromagnet on the
D81 board must be between 2.0 and 3.0mm;
if necessary, loosen the fixing screws of the
electromagnet, displace the magnet and
screw it on again. B

7.8. Distance B between the magnet holder and


the Hall sensor on the D81 board must be
between 0.5mm and 1.0mm when the grid is
tightened;
if necessary, readjust with the fixing screws of
the board (which also serve as adjusting
screws).

Note
In case the adjustment via the fixing screws of the
D82 and D81boards is not sufficient, further
adjustment can be performed by loosening the fixing
screws(4 x 2 x M3 5.5mm across flats) of the springs
on the Bucky frame. Be aware that this adjustment is
relatively complicated since it changes also the
distance of the grid frame with respect to the
Amplimat measuring chamber as well as the
oscillation direction.

7.9. Fit the new tension lever on the axle.


If necessary, exchange the present axle with
the new axle supplied.

7.10. Install the gear motor on the base plate.

Note
Usually the gear motor is in position „Grid tightened“, i.e.
when installing the gear motor make sure that also the
Bucky grid is in position “Grid tightened”.

MK serial changer (02.0) 15


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

8. Exchange of the flat cable

Note
From serial number 289 of the serial changer onwards
an exchange of the flat cable is not necessary since the
most recent version should already be installed. The
most recent version has a flat cable with the kinking
pointing at the D80 board.

8.1. Unscrew the holder of the toothed belt from


the I/O drive (2 countersunk screws M3).

8.2. Remove the toothed belt, the I/O carriage can


be easily moved back and forth hand.

8.3. Remove the flat cable.

16 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

8.4. Drill one hole 5mm diameter on the rear side


of the serial changer, 253mm from the left and
45mm from the bottom.
Drill two holes 4mm diameter on the rear side
of the serial changer, 140mm and 160mm
from the left and 66mm from the bottom.

140
132

2 holes 4 mm

2 holes 5 mm
for 3rd tie--wrap
1 hole 5 mm

66
60

45
30

140
160
165
253

Note
It may happen that on some serial changers the two
holes for the third tie-wrap are missing.
In this case these holes must additionally be drilled with
a diameter of 5mm. The position of these holes is
165mm from the left and 30mm respectively 60mm from
the bottom.

MK serial changer (02.0) 17


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

8.5. Install the additional holder for the new flat


cable in the new 5mm hole.
Install the holder for the Amplimat plug in the
new 4mm holes.

8.6. Install the flat cable.

8.7. Fasten the cable with three tie-wraps on the


rear side.

8.8. Fix the cable with the additionally attached


holder on the rear side of the serial changer
and with the already present support to the
I/O drive.

8.9. Make sure that the flat cable does not touch
any parts inside the serial changer such as
Amplimat measuring chamber or toothed belt
of the I/O drive in any of the operating
positions (--15°, 0°, +90°) while the I/O drive is
moved back and forth.

Note
The rubber buffer attached to the carriage of the
I/O drive must not hit the flat cable. If necessary,
displace the rubber buffer!

8.10. Put on back the toothed belt and screw it on


(not necessary if work proceeds with “Check
and adjustment of the play of the two
guideways of the I/O drive”).

18 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

9. Check and adjustment of the play


of the two guideways of the I/O drive

9.1. Unscrew the holder of the toothed belt from


the I/O drive (2 countersunk screws M3).

9.2. Remove the toothed belt, now the I/O carriage


can easily be moved back and forth by hand.

9.3. Slowly move the cassette carriage of the


I/O drive back and forth over the complete
range of movement.

9.4. While doing so check whether there is any


play at the right-hand or left-hand guide rail or
whether the ease of movement is impaired.

9.5. If necessary, put the serial changer on its left


side, slightly loosen the four sockethead cap
screws of the left-hand guide rail and eliminate
the play or, respectively, the impairment of the
ease of movement by displacing the rail within
the fixing holes.

9.6. Proceed accordingly for the right-hand guide


rail.

9.7. Put on back the toothed belt and screw on the


holder of the toothed belt.

MK serial changer (02.0) 19


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

10. Check and adjustment of the angle of the two cassette holders

Note
To guarantee that also heavy (large) cassettes can
reliably be held and correctly be positioned, the two
holders must have certain inward angle.
This angle makes sure that the holders get a firm grip of
the cassette .

10.1. Insert the serial changer in the Bucky frame


and establish the following electrical
connections:

-- Earth
-- VA X4 power supply
-- VA X2 CAN connection
-- VA X1 radiation release

10.2. Press the movement stop button on the table.

10.3. Switch the system ON.

10.4. Insert a cassette 35cm or 14“ in lengthwise


direction of the table in the serial changer.

10.5. Select at the system control desk „full-size


exposure“, now the cassette is centered in the
serial changer and fixed by the two cassette
holders.

10.6. Switch the system OFF.

10.7. Pull the cassette out of the serial changer to


the front, only little force is required.

20 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

10.8. Insert a cassette that is smaller than 35cm or


14“ (e.g. 25cm/10“, 28cm/11“ or 30cm/12“) in
lengthwise direction of of the table in the serial
changer.

10.9. Push the cassette in until it fully touches the


rear end of the cassette carriage with its rear
surface.

10.10. Use a feeler gauge to measure the gap 2.2 mm ± 0.1 mm


between the rear end of the right-hand
cassette holder and the cassette inserted,
nominal value = 2.2mm +/-- 0.1mm.

Note
It is also possible to use a 2.0mm Allen key instead of
a feeler gauge to measure the gap because its width
across flats is 2.2mm.

10.11. In case the deviation is too high, bend the


right-hand cassette holder carefully by hand
until the nominal value is obtained.

10.12. Repeat steps 10.9 to 10.11 for the left-hand


cassette holder accordingly.

10.13. Remove the cassette from the serial changer.

MK serial changer (02.0) 21


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

11. Check and adjustment of the collimators close to the film

11.1. Pull off the plug for the voltage supply of the
serial changer from VAX4.

11.2. Press the movement stop switch.

11.3. Switch the system ON.

11.4. Reconnect supply cable VAX4.

Note
The serial changer power--ON selftest starts
automatically which tests all axes over the complete
range of movement.

11.5. Pull off plug VAX4 as soon as the two


collimators near the film have reached their
maximum opening position.

11.6. Switch the system OFF.

11.7. Disconnect the mechanical coupling between


drive belt and the frontal collimator near the
film (2 Phillips screws M4).

11.8. Displace the collimator until its edge is flush


with the edge of the housing underneath.

11.9. In case the edge of the collimator is not


parallel to the edge of the housing, loosen the
two sockethead cap screws M4 fixing the
collimator plate on the ROLLON guide and
adjust it such that it is parallel to the edge of
the housing underneath.

11.10. Tighten the two sockethead cap screws M4


and the two Phillips screws M4.

11.11. In case the edge of the rear collimator is not


parallel to the edge of the housing underneath,
loosen the two sockethead cap screws M4
fixing the collimator plate on the ROLLON
guide and adjust it such that it is parallel to the
edge of the housing underneath.

Note
Opposed to the frontal collimator, the position of the rear
collimator cannot be changed!

11.12. Tighten the two sockethead cap screws M4.

22 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

12. Finishing work and installation of


the serial changer

12.1. Push the Amplimat measuring chamber


including the four new guide nipples fully into
the serial changer, slightly bend aside the
holder of the Amplimat plug so that the
Amplimat plug can be placed.

12.2. Screw on the Amplimat plug and bend back


the holder into position until it holds the
Amplimat plug.
Holder for Amplimat plug
12.3. Check visually by looking into the serial bent aside
changer through the flap on the operating side
whether Amplimat measuring chamber, plug
and cables are installed and routed in such a
way that they do not touch the cassette
carriage located underneath and the grid
frame installed above.

12.4. Check whether the Amplimat measuring


chamber has a bulge at the top or the bottom.

Note
If, for instance, too much lateral force is exerted upon
the Amplimat measuring chamber it may happen that it
has a bulge which either collides with the anti-scatter
grid or the cassette carriage. In this case enlarge the
Holder for Amplimat plug
slits in the four lateral guide nipples (plastic) with a sharp
introduced
knife.

12.5. Install the protective plate for the Amplimat


measuring chamber on the operating side
inside the serial changer.

12.6. Install the front panel (with green LED and


I/0 button).

12.7. Lift the serial changer by about 5 cm and push


the anti-scatter grid from the left through the
slit into the serial changer. Carefully lower the
serial changer back into position.

12.8. Install the left-hand lateral retaining angle (2


nuts M3) of the anti-scatter grid.

12.9. Remove the mechanical coupling between the


motorized collimators and the drive.

MK serial changer (02.0) 23


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

12.10. Open the two collimators manually fully to the


mechanical stop.

12.11. Position the anti-scatter grid so that the grid


bar with the tube symbol crosses the center of
the centering hole in the grid frame.
Centering hole
12.12. Tighten the four nuts M3 of the two lateral
retaining angles.

12.13. Clean the ROLLON guide of the collimator


pair near the film and slightly grease it with the
lubricant supplied. Grid bar

12.14. Install the mechanical coupling between the


motorized collimators and the drive.

12.15. Screw on the mushroom-colored narrow cover


on the operating side of the serial changer at
the Bucky frame.

12.16. Screw on the serial changer in the Bucky


frame at the position marked when the serial
changer was removed.

24 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


MK serial changer rel. 1. 3. 7

12.17. Install the two anti-scatter collimators at the


two sides of the Bucky frame in such a way
that they do not touch the collimators near the
film.

12.18. Switch the system ON.

12.19. Check whether the serial changer functions


properly as follows:
-- Load cassettes of different sizes.
-- Check whether all cassette sizes are recognized
correctly on the display of the tube hand grip.
-- Select different cassette subdivisions
(2, 3, 4 and 6 subdivisions).
-- Release different exposures and exposure
series (full-size and subdivision) with the
different cassette sizes in the 0° and 88° tilting
position.
-- Check all developed exposures (full-size and
subdivision).
-- Check whether the ease of movement of the
Bucky grid is unimpaired.
Note
When subdivision exposures are concerned, the grid is
not tightened again after the first exposure but kept
oscillating electrically so that a check of the ease of
movement is relatively easy at that moment.

-- Check whether the ease of movement of the


collimators near the film is unimpaired,
especially with respect to the anti-scatter
collimators.

12.20. Check visually whether the ease of movement


of the cassette holders is unimpaired during
subdivision exposures:
-- Insert a 24x30 cm (or similar) cassette in
transverse direction with respect to the table.
-- Select subdivision 6 at the control desk of the
system.
-- Check whether the distance between left-hand
cassette holder and Amplimat measuring
chamber is at least 1mm (use a pocket light).
-- Release three exposures.
-- Check whether the distance between right-hand
cassette holder and Amplimat measuring
chamber still is at least 1mm.

12.21. Switch the system OFF and stick the label with the new code
number (new code number: 4512 131 00781) on the present old
label located at the right-hand side near the D80 board.

MK serial changer (02.0) 25


MK_serial_changer_25831 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
MK serial changer

12.22. Push the tabletop with a second person onto


the table frame.

Note
before the first two screws M6 are tightened secure the
tabletop with a second person so that it does not roll off
the table base!

12.23. Screw in all four screws M6 to couple the


tabletop to the drive.

12.24. Install the cover panel (about 2.2m long) at the


operating side.

12.25. Install the cover panel (about 2.2m long) at the


column side.

12.26. Install the two square cover panels of the


sectional rails of the table at the column side
at the head and the foot end of the table.

12.27. Switch the system ON.

12.28. Check whether the transverse drive of the


table works properly.

12.29. Switch the system OFF.

26 (02.0) MK serial changer


E 2002 Philips Medical Systems MK_serial_changer_25831

ALL RIGHTS RESERVED


Philips Medical Systems

PARTS LIST Service

Product group: 707 Publication No.: 4512-984-25891

Description : Serial changer duo DIAGNOST Ref.No.: 4512-202-00781

List of pages and drawings:

4512-202-00781 (02.0) P-0.1

4512-202-00780 (02.0) P-1 4512-202-00780 (02.0) PZ-1


(02.0) PZ-2

4512-202-00781 (02.0) - P-0.1


© 2000 Philips Medical Systems
ALL RIGHTS RESERVED
Technical No.
Serial changer duo DIAGNOST
4512-202-.....

Page Item Description REC CDS Technical data Codenummer Qty. from up to

1.00.00 Serial changer duo DIAGNOST CS 4512-202-00781 1 00781

PZ-2 1.01.00 Main PCB X CS D80 4512-131-50311 1 00781

1.02.00 Power supply unit CS SC2, for item 1.00.00 4512-131-50181 1 00781

PZ-1 1.03.00 PCB Opto1, with cable and plug X CS R-0214 D65, with cable and plug 4512-131-32131 1 00781

PZ-1 1.04.00 PCB Motor connection CS R-0215 D66 4512-131-32141 1 00781

PZ-2 1.05.00 PCB control panel X CS D83, with cable X11 4512-131-28491 1 00781

PZ-2 1.06.00 PCB opto 2 X CS D71, D72, with cables X19, X17 4512-131-28502 1 00781

PZ-2 1.07.00 PCB grid oscilation X CS D81, with electromagnet 4512-131-28511 1 00781

PZ-2 1.08.00 PCB hall sensor X CS D82, with cable X15 4512-131-28521 1 00781

PZ-2 1.09.00 Cable DS X13 - X14 4512-131-28531 1 00781

PZ-1 1.10.00 Flat cable DS X5 - X6 4512-131-28542 1 00781

1.11.00 Fixation knobs DS for amplimat chamber 4512-131-32531 1 00781

PZ-1 1.12.00 Drive belt fixation DS 820-206-4 for cass. clamping device 4512-131-32231 2 00781

PZ-1 1.13.00 Drive belt for RM/LM X DS R-0223 755 lg. for RM/LM 4512-131-32241 1 00781

PZ-1 1.14.00 Drive belt X DS for IOM 4512-131-28551 1 00781

PZ-2 1.15.00 Drive belt X DS for SM 4512-131-28561 1 00781

PZ-1 1.16.00 Motor with gear and encoder X DS 820-340-3 RM, with gear, encoder 4512-131-32261 1 00781
and cable
PZ-1 1.17.00 Motor with gear and encoder, left X DS 820-350-3 LM, with gear, encoder and 4512-131-32271 1 00781
cable
PZ-1 1.18.00 Motor with gear and encoder, in-out X DS 820-330-3 IOM, with gear, encoder 4512-131-32281 1 00781
and cable
PZ-2 1.19.00 Motor X DS SM, with gear, encoder and cable 4512-131-28571 1 00781

PZ-2 1.20.00 Grid motor, assy. X CS 820-300-2 GM, with switches RS, GS 4512-131-32521 1 00781
and cable X12
PZ-1 1.21.00 Grid frame complete CS 4512-131-28582 1 00781

PZ-1 1.22.00 Set of springs CS for grid frame 4512-131-28591 1 00781

PZ-1 1.23.00 Front cover DS with lid 4512-131-28601 1 00781

1.24.00 Set of springs CS for front lid 4512-131-28611 1 00781

4512-202-00780 (02.0) - P-1


© 2000 Philips Medical Systems
ALL RIGHTS RESERVED
1.23.00 1.18.00 1.13.00 1.17.00 1.14.00 1.03.00
A4 02--01--21 Gri.

1.22.00 1.21.00 1.16.00 1.04.00 1.12.00 1.13.00 1.10.00


451220200780PZ--1

4512 202 00780 (02.0) PZ--1


E 1995 Philips Medizin Systeme
ALL RIGHTS RESERVED
1.09.00 1.08.00 1.01.00 1.15.00 1.19.00

1.06.00

~
A4 02--01--21 Gri.
451220200780PZ--2

1.20.00 1.07.00 1.05.00

4512 202 00780 (02.0) PZ--2


E 1995 Philips Medizin Systeme
ALL RIGHTS RESERVED
EROS- easy real time operating system
+5V +5V
+30V Local LED usage
X4 TP5 +30V +5V 2 24

45V 1 F2 5 4 VCC 3 21 R64 2


8 R62 2 8 LED Usage
Legend: +30V in 21/20 4/5 out 19 7
U5
DC Z1 Z2 19/18 6/7 20 R63 4 IC22 R61 4 7
X5 LED-No Description
CLL : Control LED on front 3A 3
C9 U1A U1B
17/16 8 6 IC23 6
TP6 1000µ 3 5
IOM : Cassette tray in/out motor see Z1-1.1
1 3 TP2 TP1 IC21 17
3
5 N14 14
LED 0 EROS is running
2 4 5/6 F1
IOS : In/out button duo DIAGNOST 2 7/8 IC13 16
Gnd LED 1 CAN bus is disconnected
IOE : Encoder on IOM motor TP1 N13 12 LM
LM : Left side clamping device motor U2A U2B ~28V AC 5/7
R45 LED 2 Startup error active
6 right
LE : Encoder on LM motor GND 2
7/8 5/6 23 1/2 11/12 13/14/15
Gnd
1 4 14/13 11/12 LED 3 Runtime error active
3
GM : Grid motor U4 TP3 GND 5V Switch mode GND GND GND
GS : Grid motor stop switch 7 left
power supply +30V LED 4 Opto1 (Cas Present) (only valid if EROS running)
RM : Right side clamping device motor X2 +5V CAN
2 24 (ON if Cassette present)
RE : Encoder on RM motor 7 R9
OC5 3 21 R57 2
8 R59 2 8
LED 5 Opto2 (Front Optos) (only valid if EROS running)
CAN H 7 1 2 20 7
RS : Grid ready/start switch 20 R58 4 R60 4 7
(ON if opto2 is active)
SE : Encoder on SM motor IC19 6
IC18 6
2 3 5
SM : Shutter device motor CAN L 6 IC20 17
3
5 N12 10 LED 6 Release Grid (only valid if EROS running)
OPTO 1 : Detector for inserted cassette (D65 PCB) input 6 CAN TxD (ON while RELEASE GRID is low)
9 IC2 D9 91 16
Gnd
OPTO 2 : Detector for the cassette rear end CAN V+ 3 7
TP2 N11 8 RM LED 7 HALL Sensor (only valid if EROS running)
5/7
(D71 and D72 PCB's) 11/12 13/14/15
R47 (ON while grid magnet is right opposite)
3 Gnd
D65 : OPTO 1 printed circuit board CAN GND 8 4 3
D66 : "Switch board" for cassette tray connectors GND GND
D71 : OPTO 2 printed circuit board, transmitter R66 +30V LED control function is polled by EROS ms timer and updates LED status all 0,5s:
D72 : OPTO 2 printed circuit board, receiver
System CAN 2 24
J1 LED 0...3 functionality
D80 : Main controlPCB with CPU see Z1-10.1 8 R51 2 8
3 21 R49 2
D81 : PCB with grid oscillation electromagnet duo DIAGNOST X3 21 7
7
Time Description LED0 LED1 LED2 LED3
OC6 20 R50 4 IC16 6 R52 4
D82 : PCB with grid oscillation hall-sensor 7
CAN H 8 1 8 3 5
IC17 6 LEDs at startup
D83 : Front panel 3 N10 6
IC15 17 5 t0 program starts running after power on ON ON ON ON
2 16
CAN L Gnd t1 after RAM check has finished ON ON ON OFF
R94 CAN RxD 4 IOM
3 90 5/7 TP3 N9 t2 after cstart has finished (environment ON ON OFF OFF
Termination
resistor 130R
9
CAN V+
5 IC1 7 11/12 13/14/15
R46
Gnd initialisation
6 t3 after ROM check has finished ON OFF OFF OFF
3 6
CAN GND 7 GND GND t4 after EROS is initialisized and OFF OFF OFF OFF
+30V
2 24
application is started
OPTO2 3 21 R114 2
8 R116 2 8
LEDs in runtime
22 7 1a indication EROS running BLINK X X X
X19 +5V 7
D71 1 R55
20 R115 4
3
IC11 6
5
R117 4
IC14 6
X16 1b indication EROS not running ON X X X
3 6 2a indication CAN disconnected BLINK BLINK X X
IC10 17 5 N7
D2 D1 2b indication CAN connected BLINK ON X X
2 16
Gnd 3a indication startup error active BLINK X BLINK
Gnd IC6 TP4 N8 8 SM X
5/7 3b indication startup error not active BLINK X ON X
R48
11/12 13/14/15
X17 CPU Gnd 4a indication runtime error active BLINK X X BLINK
T1 T2 80C167 4b indication runtime error not active BLINK X X ON
D72 1
GND GND +5V 28V AC 5 EROS not running OFF OFF OFF OFF
Gnd CR 28V AC X12
2 N1 R56/R65 8 bl
28
LD9 SSR1
Q7 R14 6
R12 rd Grid
GM
Gnd
12 gn
5 motor
X12 Gnd
C14
4 7 blk
Gnd
R34
X13 X14
Open = stop GS 2 R18
6 +30V
2 2
D81-1 TR1
8 Q4

D1 Grid
Q3 4 4
frame
Open = ready RS 3 R32
electro magnet
7
Gnd

X11
D83-1 X5 X6 X7
1
8 R37
30 12 4 D66-1 6
1
in/out N13
IOS
4 R54
+5V 3
5 D35
3
R25 N3 5 9 4
4
58
LD1 2
LM Left
5 R76 Q8 3
2 R29
Grid Gnd 14 2 8
HALL
frame 10 2
N14

X15 X14 X13 +5V Gnd X9


3
1 D81-1 7 7 4 12 6
1
N9
1 2 8 8 R19
3
5 D35
3 1
29 R27 N2 3 11 4
4
IC1 57
3 5 5 IOM IN/OUT
2 2 2
3
Gnd HALL
Magnet 6 10 8
Gnd 2
N10
X10 X6 X5
D65-3 OPTO1 6 5 9 N5 R23
+5V
N6
X8
27 8 8 6
1
N11
D66-1 Gnd Gnd 13 1 3
5 D35
X1
T2 T1 6
D10 R10 LD10 OC1 R24 R26 N4 7 7 4
4
59
2 +24V gn
1 3 5 RM Right
D66-1 X6:1 11 3 2
Start 3
D2 D1 8 R1 grid Gnd
Gnd
HALL
C19 10 6 8
D66-1 X6:3 4 2
2 4 N12
0V
Gnd
+5V
X16
Schr.

6
+5V 1
R1
LD7 61 3 D35
5
A1/A3 01-11-28

R2
LD6 62
R53 4
4
60
R3 2
SM Shutters
LD5 63 3
HALL
+30V 8
LD4 R5
64
LD8
Gnd
2
VA
X1
LD3 R6
65
D4
K1 3 Unit manual
LD2 R7
R11
Q1
Exposure 2
SYNC
Serial changer
VA D80
66
3
+30V
R8 Gnd
LD1 67

Serial changer DD (a/01.0) Main control board LD0 R15


LD11
Z1-2
6
Serial changer
68
D5
K2
c Philips Medical Systems 8x1K5 Q2 not
ALL RIGHTS RESERVED R13 Cassette ready 7 used
VSS 4
Gnd
A3/A4 01-11-27 Schr.

Serial changer dD
C 1994 Philips Medical Systems
TP1

picky pack pcb


TP6
TP5

TP3

D80-1

TP6
TP5
TP3
TP2
TP1
TP2

+5V

GND

(a/01.0)
+30V
28AC 2
28AC 1
OPTO 1 LD8 1
LD9 Exposure
1 IC K1
8 X12 IC 13
33 LD10 X5
1 1 TP1 1
9 9 9
D80

X16 14
1 8 TP6
X13 IC6 1 8
IC4 PROM TP5 rear side
OPTO 2 X4

rear side
1 8 X11 X2 X3 TP3 X1
2 8 Start grid IC 10 IC 15 IC 20 IC 21
System CAN J2
1 1
X18 TP2
X17 X19 LD 7 6 543210
28V AC
50/60Hz 3A

Z1-0.2
Main control board
VA-D80
duoDIAGNOST Section Z

Schematic Diagrams

Survey of units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.1


Connection plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.2
ERA 102 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.3
SC 210 Relais interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.4
SC 100/200 Stand Geometry Control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.5
LA Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.6
UE 100 Tube handle grip (THG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.7
Power supply SC 2 (optional for serial changer) . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.8
CG Geometry operation control (GOC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--0.9
ERA power supply geometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--1.1
Power supply table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--1.2
Power supply column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--1.3
Power supply bucky . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--1.4
Power supply LA/TSO/GOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--1.5
Safety circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--2.1
Door contact X-ray indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--2.2
Tilting drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--3.1
Bucky frame longitudinal and I.I. long drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--4.1
Bucky frame park drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--5.1
Table lateral drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--6.1
Fulcrum frontal drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--7.1
Compressor frontal drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--8.1
Bucky control / sensing / option bucky wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--9.1
Serial changer (Option bucky wall) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--9.2
CAN bus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--10.1
Signal bus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--11.1
Foot switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--12.1
Foot switch II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--12.2
LA Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--13.1
Tube hand grip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--14.1
User interface TSO/GOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z1--15.1

duoDIAGNOST (b/02.0) Z0--1


DuoD_z01_b020 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
BA = Image intensifier
BD = Photosensor
BGC= II Tube 23/31/38 cm
CB = System controller Cockpit
UB UD CG = Geometry operation control
CGII= Geometry operation control II optional
Generator rack E Option rack R FS = Foot switch
Remote service FSII = Foot switch II optional

LA = Collimator
PH = Bucky measuring chamber
ERBA
SC = Geometry control unit (GCU)
UE1 SC1 = Tilting drive
ERA SC1.1= Rear, front cover E-Box
EZ SC2 = Power supply, optional for serial changer
SD = Connection field tower
SD1= Fulcrum frontal drive
SD2= Compressor frontal drive
UA XA LA UE PH SH1 SD3= Safety end switches BFP /Ang.
EN SD2 SD VA SV SH SH4 SD4 = Compressor frontal force switches
RC SD4 SF1
SF SE = Emergency stop
SD1 SF3 SF = Bucky frame park drive
EQ SH3 SF1 = Bucky frame long./ II long. drive
SH = Coding field table frame
SD3
SH1 = Safety end switch footend/ headend
Bucky frame long./ bucky frame park/
SH1 II long. drive
G SH2= Table side operation (TSO)
SF1 SH3= Table lateral drive
SH4 = Potential equalization
UC SQ = Collision switch II
SV = Scattered radiation protection park switch
UA = Tube arm coupling
UB = Position and safety switch column
UC = Brake and collision switch tube longitudinal
UD = Brake tube vertical
UE = Tube handle grip (THG)
SC2 SC SC1.1 SC1 BA BGC SQ BD WK SH2 SE UE1 = Tube holder
VA = Bucky
CG II (GOCII)
optional WK = TV-camera
XA = X-ray tube

ERA = Power supply for geometry and II/TV


ERBA = II/TV control unit / CPX-Rack

Trolley RC = Digital X-ray imaging (DXI) optional


VA
GOC

Monitor

CG CB
PH PC-
Keyboard
Schr.

(DIM)
FSFL II
optional
01-12-03

FSFL
Survey of units
A2/A3

Duo Diagnost (c/01.0) Z1-0.1


c Philips Medical Systems
ALL RIGHTS RESERVED
SFBFP Coding field
SH Coding field UB Top of the column
SC
Brakes longit.

M1
X1 Z1-5.1 SCK6 S1 UC Collisionsswitch
in the foot of
the column
D9 X12 CAN Z1-10.1 X2 D9
Z1-2.1 Z1-2.1
S2
Y1 X2 Z1-5.1 X26 M3 D9 X20 Z1-14.1 X1 D9 Z1-14.1
D15 X15 Signal Bus and supply Z1-11.1 X4 D15
X1 S1
Z1-5.1 X6 D9
Y2
G1 Z1-14.1
Y1
Z1-5.1 X3 D9
R1

S1 Z1-5.1 X8 D9

UE THG Hand grip UA Tube support


Solder
X17 X1 connector
SF1 M12 X24 I.I. long brake Z1-4.1 X38 M15 D15 Z1-14.1 D15 X1
M3 X3 Z1-14.1
BFL/II long D9 X1 CAN Z1-10.1 D9 X11 &circuit Y1
M1
X1 Z1-4.1 SCK3
PCB
Y1
X1.1 X2 CGII S1
X3 D9 X2 Z1-14.1
optional
cable
Y2 Z1-4.1 X27 M6 D9 S2
CAN X10
Y3

Z1-4.1 X4 D9
G1 D9 X9 Z1-2.1 X15 D9
D9 X23 Z1-14.1 X1
M3
Y1/Y2
R2
Z1-4.1 X2 D9 D9 X1 I.I. long pos. Z1-4.1 X13 D9
Z1-10.1 UD Brakes vertik.
R1 D9 X10 Z1-9.1 X6 D9
LA2 Collimator
CAN X11 D9

SE Emergency stop D9 X1 CAN

S1
Z1-2.1 X7 D9 SD Coding field
EZX43.1 D9
X2 M4
D9 X11 CAN Z1-10.1 X11 D9 EZX42 D9
Z1-10.1
D9 X12 Z1-9.1 X5 D9
SH3 GGW
Table later. Corrugated hose of the stand PH X1
Amplimat X32 X14 CAN X12
X1 Z1-6.1 X25 M3 Mod 2x10
Z1-1.1
M1 Sp. S. 3+

R1 Z1-6.1 X5 D9 VA Release Z1-9.1 X13 D9 RC DXI Option


X1
Built in resistor of the
motor X2
CAN Z1-10.1 X10 D9 SC2
X11 M4 CG
D9 X1 Z1-7.1 X11 D9 Power supply mains GOC X3 CAN
CBX110
Z1-10.1
X31 X10
24V X1 M9
M6 X2 X1 400V Z1-1.2 X1
SH1 Safety switch
BFL-BFP SV
M6 Z1-1.1 M4
X5
Sign bus
CBX61
Z1-2.1 X29 M3 X8 D15 Z1-2.1 M9 Z1-11.1
S1
X1
Z1-2.1
D15 X9 Z1-8.1 X12 D15 M9 X35 X4 X3 Z1-2.1 X8 M9
head
S4 S3
S2 Z1-2.1 X28 M3 Coll. switch BFP
M9 X34 Z1-2.1 X7 M9
foot Scat. rad.
prot. park Z1-8.1 F201
switch S1 M6 X21 Z1-7.1 X36 M6 ENX2100
safety
X3 ERA
sw. BFP Power supply for
SH2 X1 D15 Sign. bus and supply
Z1-11.1 X14 D15 S2
Supply ser. ch.
GGW
mains 230V
CPX
geometry and II/TV

Corrugated hose of the stand


TSO (option) X2 CAN X1
X2 D9 CAN Z1-10.1 X13 D9 CB X11
X30
X3 D9 CAN Z1-10.1 X10 D9 SD1 FF CPX Controller
Z1-7.1 X23 M3
101/102
D9 X3 EZX61 D9 OPTIMUS
M1
CAN-Termination
Z1-7.1 X3 D9 ERBA X24
(Bucky tray-SEL)
R1 CGII X5 optional
Monitor X13A X20
CPX II/TV CB X60
RCX1 X14
Controller
(DXI option) X21
EZX23
II/TV SQ II Collision
switch S1 D25 X3 Sign bus Z1-11.1 X23 D25 X22
II-Control Central part X Z1-1.1 RCX6
BGC X1 ERBAX6 S1
Z1-2.1 X7 D9

Photosensor Central part X Z1-1.1


BD X1 ERBAX15
D9 X14 Z1-2.1
S1 SC1
WK X9
X1/X2
TV Camera Central part X Z1-1.1
ERBAX11 SD2 CF D9 Z1-3.1
R1
X1 Z1-8.1 X22 M3
M1 D9 X10 Z1-3.1 X1
Y1
GGW Z1-8.1 X2 D9
R1
Corrugated hose of the stand
M12 X32

M9 X37
SD3 X19 Part of
XA Angulate safety switch S1 /S1A/S1B X6
D9
GCU Z1-3.1
M1 Tilting
Schr.

Thermo X2/X1 D9
X1 EZX3 S1A S1
SCK2 M 3~
S1B
Z1-2.1 D15 X16
K1, L1, F1,
X2/X1 BFL position switch S2 K2, L2, F2
S2
X100 Stator
EX1100 D9 X18 K3, L3, F3
A1/A3 01-12-03

etc.

D9 X8
X31 EZX21
+HT EG HT+
Mod 2x10 D15 X21
SD4 Compr. Connection plan
force D9 Z1-2.2
switch Door contact
-HT EG HT- S1 D9 X22 Z1-2.2
S2 X1 Z1-8.1 X5 D9 X-Ray ON
Corrrugated hose S3 D9 X7
for column
(63 x 69)
S4 Wall Bucky

Duo Diagnost (d/01.0) Z1-0.2


c
Philips Medical Systems
ALL RIGHTS RESERVED
ERA Cabinet
Power supply
see Z1-1.1
1 2 3 4 5 6 X1 X8 1 2 4 5 7 8 9

1 2 3 4 5 6 X1 X3 1 2 4 5 7 8 9

F1 F2
2,5AT 3,15AT

X1:3 X6:2
D3
230V 45V
D7
X1:1 X6:1 gr

3 1

IFRP150

SC2
Power supply
1 2 3 4 5 6 X2 X5 X4 1 2 3 4 5 6 7 8 9

1 2 3 4 5 6 X31 X35 1 2 3 4 5 6 7 8 9
X50
SC Backpanel

VA
see Z1-9.2
A4 01-11-30 Schr.

SC2
Power supply

Duo DIAGNOST (01.0) Z1-0.8


C Philips Medical Systems
ERAF1 10A ERAL1
X203 X107 X108 X1 X1 X2 SCX31
1 2 1 1 L1 L1 1 1 1 SC2 1 L1 1
L1 R R/

3 4 2 2 N N 2
Filter 2 2 2 N 2
L2 S S/

5 6 3 3 L2 L2 3 3 3 3 L2 3
L3 T T/
ERAX201 Drives in stand
4 4 N N 4
Filter 4 4 4 N 4 see Z1-1.2
N N/
N
5
5 5 L3 L3 5 5 5 5 L3
PE
6 6 6 PE 6

X4 Overvoltage X2 RCX4
1 1 L3 1
check
2
Monitor 2 N 2
DXI
ERAX202 X98 see Central Part X Z1-1.2
3 3 3
PE
X5 X3 ERBAX30
1 1 L3 1

2 Option remote service 2 N 2 CPX Controller


X97 3 3
see Central Part X Z1-1.1
3
PE
X29 X11 LA2X2
X6
1
U9 =12-16V= 3 U9 3

2 Option 2. monitor 1 U6 1
H9 2 2 Collimator
3
see Z1-16
X26 4 4
R8/R13
U6 =25V=

H6 X7 SCX34
7 0V CHK+ 7

X21 8 0V CHK- 8
R2/R3
X104 U1 =25V= 1 U1 1
ERAT1 vi 1
X24 H1 2 U4 2
20V R6/R11
U4 =25V= 3 U5 3
gy 2
1 4 GND U1 4 Stand geometry
Converter 3 H4 see Z1-1.2
R65 "Collimator X25 5 GND U4 5
see Z1-1.3
yw 3 12V - 16V U9 light" R7/R12 see Z1-1.4
2 U5 =25V= 6 GND U5 6

9 9
20V or 4
H5
F1 X102 blk
X22 X8 X3 X4 SCX35
R4/R9
U2 =25V= 1 1 SC2 1 U2 1

230V ERA102 H2 2 2 2 U3 2
R10
X103 blk Power supply X23 3 3 3 3
R5/R10
U3 =25V= U2 4 4 4 GND 4
Stand geometry
5
X105 5
U3 5 5 5 GND see Z1-1.2 /1.4
4 bl 1 H3 6
6 6 6 ERAX10.2
3 28V 7
STOP MV1 7 7 7 1
K1 br 2 8 option
STOP MV2 8 8 8 2
Converter
X27 9 9 9 9 3 GOCII
25V R22/R23
rd 3 U7 =25V= 4
5
H7 X10 see Z1-0.8 CGX1
28V wt 4 X91 X92 1 U7 1 ERAX10.1
GND 2 GND U7 2 1
Schr.

STOP MV1 3 2
K1 2 3 GOC
4 STOP MV2 4 see Z1-1.5 3 GOC
1 X81
01-12-03

R24/R39 5 5 4
U8-1 5

X82 H8
R37/R38 ERA
A2/A3

U8-2
Power supply
geometry
Duo Diagnost (d/01.0) Z1-1.1
c Philips Medical Systems
X4
1 L3 1

2 N 2
I I/TV
3 3 Monitor

X5 TDX100
1 L3 3
2 N 1 Remote service optional
3 7

X6
1 L3 1

2 N 2
2. Monitor optional
3 3 Main safety
relay k1
X1 X1 SC2 X2 SCX31 SCK1
SCX32:1
1 1 1 L1 1
1 3
L1
X5:L1
L1 T1
SCX32:7
3 L2 3
3 3
SC U1
SCL1 Motor controller
5 5 5 L3 5
see Z1-3.1
SCX32:4
2 2 2 N1 2
2 4
N1
X5:N
SCX32:10
4 4 4 N2 4
1 3
L1
X5:L1
L2 T2
6 PE 6
6 6
SC U3
SCL2 Motor controller
X2 RCX4
1 L3 1 see Z1-4.1
2 N 2
I I/TV 2 4
N1
X5:N
3 3 Central Part X
see Z1-1.2 (optional)
X3 ERBAX30
1 L3 1

2 N 2
I I/TV
3 3 Central Part X
see Z1-1.1

ERA102 X11 LA2X2


1 U6 1
see Z1-1.1 3 U9 3

2 2 Collimator
see Z1-1.5 Deadman
4 4
relay 3
SCK6
X7 SCX34 SCX37:8
7
SFY1 see Z1-5.1
7 0VCHK+ 13 14

8 0VCHK- 8 Dead man


relay 1
1 U1 1
SCX37:6
SCX37:5
Dead man SCK2 SCX10:1
2 relay 2
2 U4 SC1Y1 see Z1-3.1
13 14 43 44
3 U5 3 SCK3 SCX37:7
SCX38:7
SF1Y1
4 GND U1 4 13 14
see Z1-4.1
5 GND U4 5 SF2Y1
SCX38:1
6 GND U5 6 SF2Y2
9 9

24V U1
X1:Z18
X8 X3 X4 SCX35
SC2 1 SCX37:3 SCX37:4
1 1 1 U2 U2 1.2
X1:D6/B6
2 2 2 U3 2 SCX37:1 L3 T3 43 44
U3 1.2
SCX37:2
X1:D10/B10
SC210
3 3 3
43 44 53 54
U3 1.1
SCX37:9 Relais
3
4
X1:B2/B4 interface
4 4 4 GND U2 GND U2
X1:D8/B8 see Z1-2.1
5 5 5 GND U3 5 GND U3
X1:D12/B12
6 6 6 6 GND U1
X1:D4/ 4
7 7 7 7

8 8 8 8
0V CHK-
9 X1:B21
9 9 9
24V U1
X2:B2/D2
24V U1
X2:B30/D30
see Z1-0.8 24V U3
X2:Z2
GND U1
X2:B4/D4
SC100/200
Stand geo
GND U3
X2: 4 controller

SCX4:5
CGII X1 SHX15:5
X10 CGX1 SHX14:5
24V U1
1 U7 1 X1:5 power +5V
switching
2 2
regulator
3
3 GOC/GOCII SCX4:12
4 SHX15:12 SH2
Schr.

4 see Z1-1.5 SHX14:12


5 5 GND U1
X1:12 Table side
operation
see Z1-11.1
A2/A4 01-12-03

Power supply
Table base

Duo DIAGNOST (c/01.0) Z1-1.2


c
Philips Medical Systems
ALL RIGHTS RESERVED
X4
1 L3 1

2 N 2
I I/TV
3 3 Monitor

X5 TDX100
1 L3 3
2 N 1 Remote service optional
3 7

X6
1 L3 1

2 N 2
2. Monitor optional
3 3

X1 X1 SC2 X2 SCX31
1 1 1 L1 1

3 3 3 L2 3

5 5 5 L3 5 Drives in
stand
2 2 N1 2 see Z1-1.2
2

4 4 4 N2 4

6 6 6 PE 6

X2 RCX4
1 L3 1
2 N 2 I I/TV
Central Part X
3 3 see Z1-1.2 (optional)

X3 ERBAX30
1 L3 1

2 N 2
I I/TV SCX23:1
UD
Central Part X 24V U5
3 3 see Z1-1.1 Tube vertical brake
see Z1-14.1

ERA102 X11 LA2X2


1 U6 1
see Z1-1.1 3 U9 3

2 2 Collimator
see Z1-1.5 SCX20:5
4 4
UBX1:5
24V U5
Tube longitudinal brake
X7 SCX34
7 0VCHK+ 7
SCX20:2
UC see Z1-14.1

8 UBX1:2
8 0VCHK- 24V U1
Collision switch column
1 U1 1

2 U4 2

3 U5 3

4 GND U1 4

5 GND U4 5
SCX20:1
UBX1:1
UB
24V U1
Position switch vertical
6 GND U5 6 see Z1-14.1

9 9

X8 X3 SC2 X4 SCX35
1 1 1 U2 1
SCX17:1
2 2 2 U3 2 24V U1
X1:1
3 3 3 3 SCX17:5/6
4 4 4 GND U2 4 Stand 24V U5
X1:5/6/7 UE 100
5
geometry SCX17:2
Tube hand grip
5 5 5 GND U3 see Z1-1.2
GND U1
X1:2
6 6 see Z1-14.1
6 6
7 X3:1 X2:1
7 7 7
8 8 24V U5 24V U1
8 8
9 9 9 9

see Z1-0.8
UA
Couple mechanism
see Z1-14.1

CGII X1
X10 CGX1
1 U7 1
2 2
3
3 GOC/GOCII
Schr.

4 4 see Z1-1.5
5 5
A2/A4 01-12-03

Power supply
Column

Duo DIAGNOST (c/01.0) Z1-1.3


c
Philips Medical Systems
ALL RIGHTS RESERVED
X4
1 L3 1

2 N 2
I I/TV
3 3 Monitor

X5 TDX100
1 L3 3
2 N 1 Remote service optional
3 7

X6
1 L3 1

2 N 2
2.Monitor optional
3 3

X1 X1 SC2 X2 SCX31
1 1 1 L1 1

3 3 3 L2 3

5 5 5 L3 5 Drives
in
2 2 2 N1 2 stand
ss Z1-1.2
4 4 4 N2 4

6 6 6 PE 6

X2 RCX4 24V U1
X2:1 X8
1 L3 1
2 N 2
I I/TV
SVA2
3 3 Central Part X Cassette sensing
see Z1-1.2 (optional) see Z1-9.1

X3 ERBAX30 SCX5:3
1 L3 1 SDX12:3
SDX13:3
2 N 2
I I/TV 24V U1
3 3 Central Part X
see Z1-1.1 SCX5:7

ERA102 X11 LA2X2 SDX12:7


1 U6 1 SDX13:7 24V U1
see Z1-1.1 3 U9 3

2 2 Collimator SCX5:6
see Z1-1.5 SDX12:6
4 4 SDX13:6
24V U1
X1:6
SVA1
X7 SCX34
7
Bucky control
7 0VCHK+
see Z1-9.1
8 0VCHK- 8
1 SCX5:1
1 U1 SDX12:1
2 SDX13:1
2 U4 24V U5
X1:1
3 U5 3

GND U1 4 SCX5:5
4
SDX12:5
5 SDX13:5
5 GND U4 GND U5
X1:5
6 GND U5 6

9 9
SCX7:7
VPX1:7
24V U1
X8 X3 X4 SCX35 X1:7
SC2 1
1 1 1 U2 Stand
2 2 U3 2 geometry SCX7:3
2 VPX1:3
see Z1-1.2 24V U1
X1:3
3 3 3 3
4 4 4 GND U2 4
SCX7:6
5 GND U3 5
5
6
5
6 6
VPX1:6
24V U1
X1:6
VA1
6
7 7 7 Bucky wall
7
see Z1-9.1
8 8 8 8
9 SCX7:1
9 9 9 VPX1:1
24V U5
X1:1

see Z1-0.8 SCX7:5


VPX1:5
GND U5
X1:5

24V U8
X1:9
CGII X1
X10 CGX1 SVA (optional)
1 U7 1 GND U8
2
X1:10 Serial changer
2
3
3 GOC/GOCII
Schr.

4 4 see Z1-1.5
5 5
A2/A4 01-12-03

Power supply
Bucky

Duo DIAGNOST (c/01.0) Z1-1.4


c
Philips Medical Systems
ALL RIGHTS RESERVED
X4
1 L3 1
2 N 2
I I/TV
3 3 Monitor

X5 TDX100
1 L3 3
2 N 1 Remote service optional
3 7

X6
1 L3 1
2 N 2
2. Monitor optional
3 3

X1 X1 SC2 X2 SCX31
1 1 1 L1 1

3 3 3 L2 3

5 5 5 L3 5
Drives in
stand
2 2 2 N1 2
see Z1-1.2

4 4 4 N2 4

6 6 6 PE 6

X2 RCX4
1 L3 1

2 N 2
I I/TV
3 3 Central Part X
see Z1-1.2 (optional)
X3 ERBAX30
1 L3 1

2 N 2
I I/TV
Central Part X
3 3
see Z1-1.1 LA2 Collimator
+5V
X11 see Z1-13.1 +24V
VCC
1 U6 X2:1
VDD
3 U9 L1 C20 A1 C1 +UREF
ERA102 2 X2:2
VSS
see Z1-1.1 4 LAH1
X2:3 K1 X4:1
5/8 4/7
X7 SCX34
X2:4 X4:2
7 0VCHK+ 7

8 0VCHK- 8

1 U1 1

2 U4 2 Stand
3 geometry
3 U5 see Z1-1.2
4 GND U1 4

5 GND U4 5

6 GND U5 6

9 9

X8 X3 SC2 X4 SCX35 option


1 1 1 U2 1
X11 CGII Geometry operation control
2 2 2 U3 2
1 see Z1-15.1
3 2 option
3 3 3 ERAX10.2
4 4 GND U2 4 X1 LED IC13 LED
4 Stand
5 1 1 L1 101 24V 102
5 5 5 GND U3 geometry 5V +5V
6 6 6 6 see Z1-1.2
2 2
7 7 7 7
8
3 3 KL1
8 8 8 Emergency
9 9 4 4 KL2 Stop
9 9
5

see Z1-0.8

X11 CG Geometry operation control


1 see Z1-15.1
2
ERAX10.1
X10 X1 LED IC13 LED
1 U7 1 1 L1 101 24V 102
+5V
5V
GND U7 2
2 2

3 STOP MV1 3 3 KL1


Emergency
Schr.

4 STOP MV2 4 4 KL2 Stop

5 5
A2/A4 01-12-03

Power supply
Collimator LA
Table side operation (TSO)
Geometry operation control (GOC)
Duo DIAGNOST (c/01.0) Z1-1.5
c
Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 +5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal Flaching: not NV RAM Stand geo controller
error Initialisation used write
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus

Data bus
P8 Bus P2 Bus P7 BUS P7 BUS P8 Bus P2 Bus

D26-A D27-A D38-B D26-A D30-A D16-A D30-A D15-A D16-A D35-A D27-A +24V U1 D26-A D26-A V28-A U3 D28-A D35-A D27-A
+24V U1 A/D
+24V U1 +24V U1 +24V U1 +24V U1 VCC
K4-B
K5-A K4-A
V131-B V103-C V102-B K5-B V130 V101
EN MV
V130
switched
D33-B V54-A
off collision
X1:B21 X1:A18 X2:D10 X1:B20 X2:D24 X1:B18 X2:B20 X2:B30 X2:D30 X2:Z30 X2:D32 X1:B15 X2:Z6 X1:C18 X2:Z28 X1:B8 X1:A4 X1:C8 X1:B22 X2:D6 X1:C20 X1:A19 X2:D12 X1:C13 X2:Z26 X1:C22

SD3S1 function SCX16:2 SCX16:1


32 SCX3:25 SCX3:14

U1 Check
SCV1 RBAX5
Safety circuit line SCK1 SCX15:4 SCX15:6 SCX16:6 SCX16:5 SCV4/V5 PO+V
SCX4:1 SCX4:8

CONNECT ENX ENMV STATUS


Fulcrum height in mm main
31 safety SCW2 SCX22:2 SCX21:2
+111...+235 SCX16:3 SCX3:18
Over voltage check

-165...+110 relay K1 SCX3:15

Door contact
SCX4:10

CONNECT ENX ENMV

X RAY ON
2 1 SCV6/V7 SCX16:14 SCX16:13 SCX16:11 SCX16:10 1 1
1 1 SCX14:6 S...
SD3S1 S...

I I Collision
2 4 2 4
safety 12 GOC
TSO SCW1 SCR2 SCR1
switch SC1S1

EN X STATUS
2 10K 2 40K
enable dead man K2
dead man status K2

11
tilting PO 0V PO 0V

enable dead man K8

enable dead man K3


SCX14:1

dead man status K8

dead man status K3


SD3X1:4 SD3X1:5
movement stopped status

lamp movement stopped

Delay K2

Delay K3
SD3X2:4 SD3X2:5 SCX20:3

EN MV
PO +V

PO 0V
UBX1:3 SCX12:13 SCX12:2

EN X
safety
2 2 SDX9:13 SDX9:2
switch 2
A SD3S1 B tube
UBS2 SDX8:13 SDX7:6
1 1 SVX1:2
1 vertikal UBX1:4
SCX20:4 extern
X1:D2 X1:Z2 X1:B32 X1:B30 X1:D32 X1:D30 X1:Z16 X1:Z30 X1:Z32 12
SCX12:3 32 X-RAY ON
SDX9:3 SCK2 SCV2 32
SCK6 SCV6 SQS1
SDX6:8 11 I I Collision
SD3X1:8 Tilting 31 BFP R1 R3
2
SD3X1:3 31 SVS1 SDX7:1
1 Scat. rad.
32 SCV3 4 1 SDX8:6
SD3S1 2 22 SCK3 D18
K7-B K7-C 11 9 prok. park.
SVX1:7
4 22 D1 D2
SCK6 SCK2 BFL table
R2 R4
SD3X1:4 SD3X1:5 21 21 31 8 6 D19
lat. 13 2
SCX34:8 SD3X2:4 SD3X2:5 9 11 K8-A 6 8 SVS4 extern
X7:8 2 2 1 BFP collision Door
Delay circuit
SD3S1 A B SCX16:8 SCX16:4 SCX16:9 K8-C K8-B SVX1:1 contact
SCX34:3 1 "switch off
24V U5 safety
1
delay"
13 4 SDX8:1 SVX1:6 SCX22:1
X7:3 SD3X2:9 SHX9:4 SDX9:1
see Z1-1.1 switch u. SHX9:6
2
angulate
SD3X1:9 24V U1 SCK2/3 SVS3
SDX6:9 1 BFP collision
C1 C2 SCX21:1
SDX8:5 EN MV Generation for tomography
SCX34:6 2200u 2200u
GND U5 safety 2 SVX1:4 SVX1:5

ERA X7:6 1: Connect ENX ENMV: Geoready+Tomo selected SDX8:11


switch SVS2 +Preperation pressed SDX9:11
BFP 1
SCX12:11
Cabinet SVX1:3
SDX8:4 SC210
2: ENX : Enable X-ray generation
Fluoro or Exposure SCX20:7
Power supply SDX9:4 UBX1:7
SCX12:4 Relay interface 3: EN MV : Enable movement
SCX15:5 X1:D4 X1:Z4 X1:Z18
safety SHX9:5
ERAX10.1:3 end SHX29:1
CGX1:3 switch 1
X10:3 head
SH1S1 2
2
BFL/BFP UC S1
1 CG ES1 SHX29:2 1 Column collisi.
SHX28:2
2 GOC safety
2
end
switch
SH1S2
UBX1:2

GND U1
1
ERAX10.1:4 foot SHX28;1 SCX20:2
CGX1:4 BFL/BFP SHX7:5 SHX7:4 SHX7:6

2 4 X1
Schr.

ERAX10.2:4
option SES1
CGIIX1:4 SEH1
GND U1

see Z1-1.1 2 1 Emergency 3

24V U1
CGII ES1 Stop X2
ERAX10.2:3
01-12-04

CGIIX1:3 1 GOC II
X10:4 SHX7:2 SHX7:1 SHX7:3
SHX9:2
SCX16:8 SHX9:1
SCX15:2
SCX15:1
24V U1

24V U1

24V U1

24V U1

24V U1
X8:7 SCX35:7
SC2X4:7 SCX35:8
SCX34:1/7 SC2X3:7 SC2X4:8
X8:8 24V U1 SC2X3:8
A1/A3

X7:1/7
GND U1 SCX34:4
X7:4 Safety circuit
BFP: Bucky frame park

BFL: Bucky frame longitudinal

Duo DIAGNOST (c/01.0) Z1-2.1


c Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
+5V VCC
SC100/200
1/24 14/11
VCC
A9
C70
DC/DC
C69
H1 H2 H3 H4
Stand geo controller
X7
12/13 10/15 D1-F rd yw gr gr
Fatal Flaching: not NV RAM
SCX34 error Initialisation used write
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H X1:B2 B3 S1-A
CAN H NV RAM
1
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus

Data bus

P2 BUS
D37 U1 P5 BUS
D35-A

U3
VCC VCC VCC A/D D25-A

V28 24V U1 24V U1


B13-B B14-A B13-A B18-A
+15V
POFC G
24V 24V
GND V75 D34-A
10V -10V
S D

X1:B28 X1:A28 X1:C28 X1:C27 X1:A27 X1:A29 X2:D18 X1:A14 X1:A10 X1:B14

SCX8:3 SCX8:4 SCX8:6 SCX8:7 SCX8:2 SCX8:8 SCX10:6 SCX9:6 SCX9:2 SCX9:1
SCK6
see Z1-2.1

TILTERRSTAT
Mnemonic:
ERACabinet 2 1

ENTILTDW

see Z1-5.1
POFCGND
SCK2

TILTSEPO
ENTILTUP

NGUREF
TILTPOS
POUREF
RESTILT
SCK1 Dead man

TILTBR
Power supply Main safety 4 3 see Z1-2.1 ENTILTDW Enable tilting down
see Z1-2.1 ENTILTUP Enable tilting up
6 5 NGUREF Negative reference voltage
POFCGND Power FC ground
SC2X1:1 X1:3 X1:4 X1:6 X1:7 X1:2 X1:8 SC1X1:2
SC2X2:1 POUREF Positive reference voltage
L1 SCX31:1 SCX32:1 L1-1.2 U 1 U1 W2
SC1M1 RESTILT Restart tilting
L1-1.1 X5:L1 X5:U 4.6A/6,3A
X1:1 1 3 TILTBR Tilting brake
L1 T1 2 1 SC1Y1
3~ 2 V1 TILTERRSTAT Tilting error status
Load
Line

SC2X1:2 V 2 V1 U2
N1
SC2X2:2
SCX31:2 SCX32:4
N1 N1-1.1
SC U1 X5:V
4 3
M1 1N5060 TILTPOS
TILTSEPO
Tilting position
Tilting setpoint
X1:2 2 SCL1 4 X5:N
Motor controller
X5:W
W 3 W1 V2
1
2
Filter 6 5 3 1
SC1X1:1
X5:PE SCX50 SC1R1
PE
SCX52 X5:- X5:PTC X5:+
Tilting motor brake Tilting position potentiometer
Power supply SCX51
see Z1-1.1
SCX10:1
2- 1+ SCX37:6
SC U2
Brake chopper circuit

+24V U4 SCX34:2 SCX37:5


X7:2
Schr.

13 14 43 44
GND U4 SCX34:5
X7:5
A1/A3 02-02-05

Tilting drive

Duo DIAGNOST (b/02.0) Z1-3.1


c Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal Flaching: not NV RAM Stand geo controller
error Initialisation used write
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus
Data bus
P2 Bus P2 Bus P2 Bus
+5V
P2 Bus W5 P2 BUS
D37-A PO EC GND

VCC VCC VCC U3 D28-A 24V U1 U3 D25-A 24V U1 D30-A U3 D25-A D30-A D30-A
A2-B A/D A/D A/D
+24V U1 +24V U1 +24VU1
W4 U2
24V 24V
B15-A B16-B B16-A D15-A D24-B D34-B V132-D D33-A V102-A V54-D
D15-A A2-A +10V -10V
GND U1

X1:A16 X1:B30 X1:C30 X1:A32 X1:C31 X1:C32 X1:A31 X1:C15 X2:D4/B4 X1:A13 X1:A14 X1:C10 X1:B14 X2:Z18 X1:A11 X2:B18 X2:Z22

SCX12:5
SDX9:5
SDX6:2 SCX18:3 SCX18:4 SCX18:6 SCX18:7 SCX18:2 SCX18:8 SCR8 SCX13:9 SCX13:8 SCX13:6 SCX13:7 SCX13:1 SCX38:11 SCX13:4 SCX38:10 SCX38:12
SD3X1:2 1K
3
SD3S2

I.I. long clutch


I.I. long brake
24V U1

I.I. long pos.


Position
GND U1 SCX34:4

BFL. POS.

NE UREF

BFL. brake
PO UREF

PO UREF
BFL. Velo

NE UREF
1 switch 2

GND U1
X7:4
SD3X1:1 SCW4
SDX6:1
1
Enable BFL down

SDX9:1
24V U1 SCX34:1 SCX12:1
Opto GND BFL
Enable BFL up

+24V U1 GND U1

BFL error stat


Set point BFL
X7:1 SHX1:9 SHX1:8 SHX1:6 SHX1:7 SHX1:1 SHX24:11 SHX1:4 SHX24:10 SHX24:12
SHX2:7 SHX2:2
Reset BFL

SHX4:6 SHX4:1 SHX2:6 SHX2:5 SHX2:1 SHX27:4 SHX2:4 SHX27:5 SHX27:6


SCK3
Dead man
SCK1 see Z1-2.1
Main safety
see Z1-2.1
1K
see Z1-1.1 2 2

X1:3 X1:4 X1:6 X1:7 X1:2 X1:8


+ G - 1 3 3 1

SC2X1:3 SCX31:3
U SF1X1:1 U1 W2 SF1M1 SF1 R1 SF1 R2
L2 SC2X2:3 SCX32:7 L2-1.1 X5:U SF1 G1 Position poti Position poti
X1:3 1 3 X5:L1 2 1
L2 T2 BFL. I.I. long
SF1X1:2 3~
V V1 U2
Load
Line

N2
SC2X1:4
SC2X2:4
SCX31:4
SCX32:10
SC U3 X5:V
4 3
M1
N2
X1:4 2 4 X5:N Motor controller W SF1X1:3 W1 V2
X5:W
6 5
SF1X1:PE
SCX51
SCL2 X5:PE
PE
Filter X5:- X5:PTC X5:+ SCX50

ERA SCX52
SCX50
BFL and I.I. long
Cabinet 2- 1+

Power supply SC U4
Brake chopper circuit Brake BFL

SCX37:7 Brake Clutch


SCX34:2 SCX38:7 I.I. long I.I. long
24V U4 SCX37:5 SHX24:7 SHX27:1 SF1 Y1
X7:2
13 14 13 14
see Z1-1.1
GND U4 SCX34:5
X7:5 SHX27:2 SHV2 SF1 Y2
Schr.

SCX38:1
SHX24:1
SHX27:3 SHV3 SF1 Y3
A1/A3 01-12-04

Function brakes and clutch SF: SHS4

Brake Brake Clutch Movement


SF1Y1 SF1Y2 SF1Y3 function
ON ON ON Scanning I.I. Long/tomo/ang.
OFF OFF tomo/ang.
ON
OFF OFF OFF Stand by Drive
ON 24V active brake free clutch active (transmitting torsion)
Duo DIAGNOST (b/01.0) Z1-4.1
c Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal
error
Flaching:
Initialisation
not
used
NV RAM
write
Stand geo controller
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND
D6-A D31-A
P8 Bus P7 Bus

Data bus

D37 U1 P5 BUS P2 Bus P2 Bus

P2 Bus
VCC VCC VCC

D36-A U3 D28-A 24V U1 U3 D25-A 24V U1


D15-A A/D A/D
B18-A +24V U1
+15V B13-B B14-A B13-A
POFC
GND 24V 24V
V53-B D24-D D34-C
10V 10V

X1:B16 X1:B28 X1:A28 X1:C28 X1:C27 X1:A27 X1:A29 X2:B16 X2:D4/B4 X1:C12 X1:A14 X1:B10 X1:B14

SCX15:9
SHX9:9
SCX8:3 SCX8:4 SCX8:6 SCX8:7 SCX8:2 SCX8:8
SHX8:6 SCX38:9 SCX13:9 SCX13:5 SCX13:6 SCX13:3 SCX13:1

SFS1 4
Position
BFP
GND U1 SCX34:4
switch 1 SCK2
X7:4 SHX8:1 Dead man
SHX9:1 see Z1-2.1

BFP. brake

BFP. POS.
BFP. Velo

PO UREF

NE UREF
SCX15:1
24V U1

GND U1
SCX34:1
X7:1 +24V U1

2 1

see Z1-3.1
SHX24:9
4 3

TILTERRSTAT
SHX26:2

TILTSEPOI
ENTILTDW

POFCGND
ENTILTUP

RESTILT

6 5
SCK1
Main safety
see Z1-1.1
see Z1-2.1 SCK6
see Z1-2.1 SFX2:2

X1:3 X1:4 X1:6 X1:7 X1:2 X1:8


SC2X1:1 SCX31:1
L1 SC2X2:1 SCX32:1 L1-1.2 U SFX1:1 U1 W2 SFM1
L1-1.1 X5:L1 X5:U BFP
X1:1 1 3
L1 T1 2 1 SF Y1
SFX1:2 3~ 2
Load
Line

V V1 U2 Brake BFP
N1
SC2X1:2
SC2X2:2
SCX31:2
SCX32:4 N1 N1-1.1
SC U1 X5:V
4 3
M1
1
SHV1
SHX1:9 SHX1:5 SHX1:6 SHX1:3 SHX1:1
X1:2 2 4 X5:N
Motor controller W SFX1:3 SHX6:6 SHX6:1 SHX3:6 SHX3:2 SHX3:1
SCL1 X5:W
6 5
W1 V2
1)
Filter SFX1:PE
X5:PE SCX50
SFX2:1
2
SCX52 PE
ERA
X5:- X5:PTC X5:+
SCX51 - G + 1 3

SF R1
Cabinet SF G1 Position poti
2- 1+ BFP.
Power supply
SC U2
Brake chopper circuit
Schr.

1) Short circuit comnector SHX6 4512 131 11761 built in


SCX34:2 SCX37:8 SHX24:4 from table s/n 318 onwards and is needed after
A1/A3 01-12-04

24V U4 SCX37:5 SCX38:4 SHX26:1 exchange of the BFP drive unit 4512 131 26052!
X7:2 Tacho generatot never had a function!
13 14 13 14
see Z1-1.1
GND U4 SCX34:5
X7:5

BFP. Drive

Duo DIAGNOST (b/01.0) Z1-5.1


c Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 +5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal
error
Flaching:
Initialisation
not
used
NV RAM
write
Stand geo controller
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus

Data bus
P2 Bus
P7 Bus

D36-A D36-A U3 D25-A 24V U1


24V U1 A/D
+24V U1 +24V U1 +24V U1

24V 24V
V101 V131-A V102-C D34-C
10V 10V

X2:D8 X2:B12 X2:Z12 X1:A14 X1:B11 X1:B14

PWM Table lat

Direction Table lat

Enable Table lat


SCK1 SCK3
Main safety Deadman
+24V U1 SCX13:6 SCX13:2 SCX13:1
see Z1-2.1 see Z1-2.1

X1:D26 X1:Z18 X1:D24 X1:D22


SC2X3:1 SCX35:1
+24V U2 SC2X4:1 SCX37:3 SCX37:4
X1:D6/B6 K2 K1
BAX8:1

Table lateral pos.


L3 T3 43 44

Power supply K1

PO UREF

NE UREF
see Z1-1.1
G Q1
SC2X3:4
GND U2 SC2X4:4 SCX35:4
X1:D8/B8 S
BAX8:4 D

SC 210
ERA Relay interface
Cabinet
Power supply
K2 K2
SHX1:6 SHX1:2 SHX1:1
SHX5:6 SHX5:2 SHX5:1
X1: 6/ 8 X1: 10/ 12

SCX38:3 SCX38:6
SHX24:3 SHX24:6
2
SHX25:1 SHX25:2
SH3X1:1 SH3X1:2 3 1
SH3 M1 SH3 R1
2.5A 24V
blk rd
Position poti
M
A B

Motor table lateral


Schr.
A1/A3 01-12-04

Table lateral

Duo DIAGNOST (b/01.0) Z1-6.1


c
Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 +5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal
error
Flaching:
Initialisation
not
used
NV RAM
write
Stand geo controller
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus
Data bus
P2 Bus
P7 Bus

D36-A D36-A 24V U1 U3 D25-A 24V U1


+24V U1 +24V U1 +24V U1 A/D

24V 24V
V131-D V102-D V103-A D24-A
10V 10V

X2:B10 X2:D14 X2:B14 X1:A14 X1:C11 X1:B14

Direction Fulcrum
PWM Fulcrum

Enable Fulcrum
SCK1
Main safety
see Z1-2.1 +24V U1 SCX11:6 SCX11:2 SCX11:1

X1:B26 X1:Z18 X1:B24 X1:B22


SC2X3:2 SCX35:2
+24V U3 SC2X4:2 SCX37:1 SCX37:9
X1:B2/B4 K4 K3
BAX8:2
43 44

Fulcrum pos.
Power supply
K3

PO UREF

NE UREF
see Z1-1.1
G Q2
SC2X3:5
GND U3 SC2X4:5 SCX35:5
X1:D12/B12 S
BAX8:5 D

SC 210
ERA Relay interface
Cabinet
Power supply K4 K4
SDX1:6 SDX1:2 SDX1:1
SDX3:6 SDX3:2 SDX3:1
X1:B14/D14 X1:B16/D16
SCX36:1 SCX36:4
SDX21:1 SDX21:4
SDX23:1 SDX23:2
2
3 1
SD1 M1
4.7A 24V SD1 R1
br yw Position poti
M
A B

Fulcrum frontal drive


Schr.
A1/A3 01-12-04

Fulcrum frontal drive

Duo DIAGNOST (b/01.0) Z1-7.1


c
Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 +5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal
error
Flaching:
Initialisation
not
used
NV RAM
write
Stand geo controller
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
see Z1-1.1
A10 3/22
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus

Data bus
P7 Bus P2 Bus

D36-A D36-A U3 D25-A D27-A D16-A


+24V U1 +24V U1 +24V U1 24V U1 A/D 24V U1

24V 24V
V131-C V103-D V53-A D34-D
10V 10V

X2:Z10 X2:Z14 X2:D16 X1:A14 X1:A12 X1:B14 X1:C23 X1:A24 X1:B24 X1:C24

Direction Compr.H

Enable Compr.
PWM Compr.
+24V U1

SCK1 SCK3 SCX34:1


Main safety Dead man +24V U1 SCX11:6 SCX11:3 SCX11:1 SCX12:6 SCX12:7 SCX12:8 SCX12:9 SCX12:1
see Z1-2.1 see Z1-2.1

X1:Z26 X1:Z18 X1:Z24 X1:Z22


SC2X3:2 SCX35:2

Comps. emer gency 160N


+24V U3 SC2X4:2 SCX37:1 SCX37:2
X1:D10/B10 K6 K5
BAX8:2
43 44 53 54

Comps. force 70NH

Comps. force 100N

Comps. force 140N


Compr. pos.
Power supply K5

PO UREF

NE UREF
see Z1-1.1
G Q3
SC2X3:5
GND U3 SC2X4:5 SCX35:5
X1:D12/B12 S
BAX8:5 D

SC 210
ERA Relay interface
Cabinet
Power supply
K6 K6
SDX1:6 SDX1:3 SDX1:1 SDX9:6 SDX9:7 SDX9:8 SDX9:9 SDX9:1
SDX2:6 SDX2:2 SDX2:1 SDX5:6 SDX5:7 SDX5:8 SDX5:9 SDX5:1
SD4X1:2 SD4X1:3 SD4X1:4 SD4X1:5 SD4X1:1
X1:B18/D18 X1:B20/D20
SCX36:2 SCX36:5
SDX21:2 SDX21:5 2 2 2 2
SDX22:1 SDX22:2
2 1
SD4S11 SD4S21 SD4S31 SD4S4
SD2X1:1 SD2X1:2
3 1
SD2 M1
rd 2.5A 24V blk
SD2 R1
Position poti
M Compr. Force switch

Compr. frontal drive


Schr.
A1/A3 01-12-04

Compr. frontal drive

Duo DIAGNOST (b/01.0) Z1-8.1


c
Philips Medical Systems
ALL RIGHTS RESERVED
VCC VCC VCC VCC VCC
X8 X9
1/24 14/11 5V
VCC
VCC
C70 A9 C69
DC/DC
X7
D1-F
H1 H2 H3 H4 SC100/200
12/13 10/15 rd yw gr gr

SCX34
Fatal
error
Flaching:
Initialisation
not
used
NV RAM
write
Stand geo controller
done protection
1
+24V U1 X2:B2/D2 1/24 11
+24V U1 +15V D1-E off:
CAN connected
C15 C72
10/15 A1-C
Power supply
A10 3/22
see Z1-1.1
DC/DC D1-D

4
GND U1 X2:B4/D4 12/13 2/13
GND U1 -15V
VCC D14 P5 Bus D1-A
SCX1 VCC Controller
7
CAN H
CAN H X1:B2 B3 S1-A
1 NV RAM
VCC
D9-A D19-A 2 write
2 protection
CAN L X1:A1 3
CAN L B5
CAN Bus S2-A
see Z1-10.1 P2 Bus 1
R12 2
9
CAN V+ X1:C2
CAN V+ A5 +5V CAN 3
POH POL
CPU reset
3
CAN GND X1:B1
CAN GND D6-A D31-A
P8 Bus P7 Bus

Data bus
P2 Bus P2 Bus
P7 Bus

D15-A D35-A D26-A


+24V U1 +24V U1

V132-B D15-A V130

X1:A17 X2:D28 X1:A21 X1:C16 X2:B28

present WB

Sync WB
Sync TB

Start grid

Termination

Cassette

Start grid
24V U1 24V U1
System CAN see Z1-10.1 resistor 130R
24V U1 24V U1
SD X10
SCX5:2 SCX5:4 SCX5:6 6 3 9 2 7 5 5 7 2 9 3 6 SCX7:9 SCX7:2 SCX7:4 SCX7:6
SDX12:2 SDX12:4 SDX12:6
R5 VPX1:9 VPX1:2 VPX1:4 VPX1:6
SDX13:2 SDX13:4 SDX13:6
1K R8 R9
SCX5:3 X1:2 X1:4 X1:6 X2:6 X2:3 X2:9 X2:2 X2:7 X2:5 X3:5 X3:7 X3:2 X2:9 X3:3 X3:6 1K 1K X1:9 X1:2 X1:4 X1:6
SDX12:3
SDX13:3
3 SCX7:7
VPX1:7
6 3
RY1 D4 D3
+15V
VA1 +24V

X1:3 4 5 shield
CAN H
X1:7 7
8 5
4
SYNC Bucky wall 2
RY1
K1 +36V RY2 P1 delay old D1
exposure CAN L free P2 1 SYNC
delay new 6-60ms
SCX7:3 exposure
LD8 CAN V+ VPX1:3 2-10ms +15V
X1:3 IC1B T2
CAN GND R2 R4
K1 timer
exposure CAN (GND) ~20s R8 VAS1
SCX34:3 W1 1 2 3 X24 2 4
CAN GND J1 100R SCX7:1
D2 IC1A
VPX1:1 1
X1:1 C7 timer X25
IC6 24V U5 +24V
LD10 CPU IC1 IC2 +24V
X26
VAY1
SCX34:6 DC/DC +15V +15V
SCX7:5 R11 1
OC1 IC2 D5
VPX1:5
X1:5 timer X27 2
GND U5
R14 ~ 2s
R12 START & T3
OC1
R13

C8
SCX35:3
SCX12:10/12
SDX9:10/12
SDX8:10/12 TP5
L2 X1:3 2,5AT 3,15AT
X4:3
SVX1:9
X4:1
F2 5 4
X1:3 U5 +30V
C9 Z1 U1A
3 1000µ Z2 U1B
1 3 TP2
+ 4m7 SCX35:6 2 4 5/6 F1
7/8
230V 45V gr
SCX12:14/15 TP1
see Z1-1.1 SDX9:14/15
TP6 1 +5V U2A U2B ~28V AC
SDX8:14/15 6 right 7/8
VCC 5/6
SVX1:10 2 4
N X1:4 X4:6 X4:2 1 3
X1:4 U4
left TP3
VA
7
GND D-80
ERA
Cabinet SC2 Main control board
Serial changer
Schr.

Power supply Power supply


A1/A3 01-11-29

Serial changer
Option bucky wall
Duo DIAGNOST (b/01.0) Z1-9.2
c Philips Medical Systems
ALL RIGHTS RESERVED
duoDIAGNOST Section Z

SERVICE MANUAL -- UNIT

Wiring Diagrams

E-Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--0.1
E-Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--0.2
R-Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--0.3
Earthing diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--0.4
ERA Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--1
SC Backpanel Geometry control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--2.1
SC Wiring Diagram Geometry control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--2.2
SC/SC1 Tilting drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--2.3
Power supply SC 2 (optional for serial changer) . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--2.4
SD Backpanel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--3.1
SD1 Fulcrum frontal drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--3.2
SD2 Compressor frontal drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--3.3
SD3 Safety end switches BFP / Angolate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--3.4
SD4 Compressor frontal force switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--3.5
SE Emergency stop
SF Bucky frame park drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--4
SF1 Bucky frame longitudinal I.I. longitudinal drive . . . . . . . . . . . . . . . . . . . . . . . . . Z2--5
SH Backpanel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--6.1
SH1 Safety end switch footend Headend bucky frame longitudinal
Bucky frame park I.I. long drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--6.2
SH2 Table side operation (TSO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--6.3
SH3 Table lateral drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--6.4
SQ Collision switch I.I. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--7
SV Scattered radiation protection park switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--8
UA Tube arm coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--9.1
UB Position and safety switch column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--9.2
UC Brake and collision switch tube longitudinal . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--9.3
UD Tube vertical brake . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--9.4
UE / UE 100 Tube handle grip (THG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--10
CG/CG II Geometry operation control (GOC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Z2--11

duoDIAGNOST (b/01.0) Z0--2


DuoD_z02_b020 E 2002 Philips Medical Systems
ALL RIGHTS RESERVED
CBX60
EZX23
RCX6 (DXI) option
SCX3
LAX1 rear side CGIIX5 option
CGX1 CBX110
SC2X3
SCX34

X1 X30 ERAX3
ERA Power supply or
EZX42
X20 21 22 23 24 X2 X33
X7 X8
X13B X13A X14 X4 X6
X104 X9 X11 X34
X10 X11 BGCX1
T1 X15
X105 WKX1
X3 BDX1
option 2nd GOC ERBA
RAX10.1 CPXII/TV Controller
CGX1 EZX61
RAX10.2 RCX1
ENX2100 X103 (DXI option)
CGX1.2
L1' L2' L3' N
Monitor
X102 X108
L1 L2 L3 loard
ERA102
X107
X203

L1
PE N F1 L1 Backpanel EZ Basis rack
X1
T1 T2 T3
line X4 X5 X6 X2 X3
L1 L2 L3 N

230V monitor
230V remote service
230V 2nd monitor
RCX4 (DXI) option
ERBAX30
SC2X1
EY Rotor control

EQ Converter

G H.V. generator
Schr.
01-12-04
A3/A4

E-Cabinet
Duo Diagnost (b/01.0) Z2-0.2
c Philips Medical Systems
ALL RIGHTS RESERVED
COVER HEAD
X11
UB1
Option rack R Generator rack E INSULATED COVERS :
X11 X11 WCL
X12 X11 FOLLOWING METALIC COVERS, WITH ELECTRICAL EQUIPMENT UNDERNEATH,
cabinet back
X30:3 ARE TO PROVIDE WITH SELFLAMINATING INSULATING FOIL or TAPE:

TD
REMOTE SERVICE ERA UC UB
SELECTOR (Option) COLUMN HEAD No. COVER FOR
X100:7
Power supply Geo PCB
COLUMN- COL.FOOT
UD 1 POTI Table Lateral SH3 R1 SPECIFICATION FOR INSUL. TAPE :
ERBA X11 X11 BRAKES TUBE 2 POTI BFP SF R1 UL, DIELECTR.STRENGTH MIN. 500 V,
CPX Backpanel VERT. 3 POTIs BFL/IIl. SF1 R1/R2 MECH. RESISTANT,WELL LAMINATING.
WCL 4 Backpanel SH SH
UA UE1 5
6
Saf.SWITCH BF.head SH1 S1
Saf.SWITCH BF.foot SH1 S2
FOR EXAMPLE: Polyestertape
Beiersdorf tesafilm 4108 or 4113 or
TUBE ARM TUBE HOLDER
7 Coll.SWITCH II SQ S1 3M scotch-No. 5 or 10
X4:3 X2:3 X100 X11 8 Couple Stat.SWITCHES column UA S1/S2
EZ * X98
X5:3 X3:3 in THG
9 Decouple Bloc.MAGNET column UA X1/Y1
X202

LA UE
Monitor COLLIMATOR TUBE HANDLE GRIP
earthpot. multiplier
fixed cab. housing
cabinet front WTB
eart bolt
cab. housing
ERY
cabinet front
GOC
SC GEOMETRIE
X4:3
Mon.
CONTROLLER UNIT
X40
SD
RC DXI (Option/if no SIM) EN * X50 FULCRUM
Gener.(main PE)
X45
BACKPLANE
BA
SD II-TUBE (BV) CONNECTORHOUSING CONNECTORHOUSING
X52 B F L / II DRIVE SF1 X1 B F P DRIVE SF X1

SC1.1 BACKPLANE ON GCU X51 BA1


REAR
COVER
for E-BOX
X12 WCL
FRONT
COVER
for E-BOX
M1
EQ * X40 X1
X11 X13 X12
BACKPLANE SH
SH
E-BOX MAINFRAME TABLE BASE
X11 SH4 X2

WTB
SF1 POTENTIAL
BALANCE
X10 B F L / II
DRIVE
SC1 X50
X51

SC2 GEOM.FOOT
CENTRAL EARTH
X61
INSULATED X62
Power supply
X5

X11 SH1 X12 SF


Chassis f.switches
M1 B F P DRIVE
G* M1
SAF.SWITCHES BF.. HEAD/ FOOT

X11
X63
X11
SH1 X13

SC1 Long cover for


backw.table base
Chassis f.M1
Amplimat chamber
TILTING-DRIVE
VA
BUCKYMODUL
VA X5
X11

Chassis f.M1

EARTHING LEAD THROUGH "CORRUGATED HOSE" X64

WTB= CORRUGATED HOSE TABLE BASE and SH3


WCL= CORRUGATED HOSE COLUMN TABLE LAT.
Schr.

VAC DRIVE
X1
* Part of generator
A1/A3 01-12-03

X12
SH2 TABLESIDE
OPERATION
Housing TSO mounted with earthing washers
Earthing diagram
Duo Diagnost (c/01.0) Z2-0.4
c Philips Medical Systems
ALL RIGHTS RESERVED
ERA Power supply ERA102 Backpanel

X104
T1:vi 1
T1:gy 2 X7 X8 X9 X10 X11
1 1 1
T1:yw 3 1 2 3 1 2 3
2 2 2
T1:or 4 4 5 6 4 5 6
3 3 3
T1 X103 7 8 9 7 8 9
blk 4 4 4
blk X102
X105:4 5
wt
rd X105:3
X105:2 SCX34 SC2X3 LA2X2
br
bl X105:1 CGX1
or X104:4
ERYX4 yw X104:3
gy X104:2
vi X104:1

X97

X108:2
X108:4
X108:1

X108:3

X108:5
ENX2100 X105
T1:bl 1
N L1 L2 L3 L1' L2' L3' N'
T1:br 2
T1:rd 3
T1:wt 4
loard

L1 L2 L3 T1:blk
X103
X108
PE N T1:blk
X102 L1:L1' 1
on L1:N' 2
L1:L2' 3
off
X202
X201

F201 L1 L1:N' 4
L1:L3' 5
6
PE N
X107 X203 X1
L1:L1 1 F1:T1 1 1
T1 T2 T3
L1:N 2 2 2
L1:L2 3 F1:T2 3 X4 X5 X6 X2 X3 3
Schr.

X203:4

line L1:N 4 X201:N 4 1 1 1 1 1 4


X203:1

X203:3

X203:5
X98

L1:L3 5 F1:T3 5 2 2 2 2 2 5
L1 L2 L3 N
01-12-04

6 6 3 3 3 3 3 6
X98
X107:1

X107:3

X107:5

X107:2
X107:4

Monitor TDX100 2. Monitor RCX4 ERBAX30 SC2X1


X202:PE Remote optional
A3/A3

Option
service
optional
ERA Power supply
Duo DIAGNOST (c/01.0) Z2-1
c Philips Medical Systems
ALL RIGHTS RESERVED
2 SCR1:2
100X1:A1

4 100X1:B1

X1:1
X1:2
X1:3
X1:4
X1:5

1 X12:2
X2:2
X2:1

X2:3

X2:4
X2:5
optional

100X1:B4
GOCII

1
2
3

2
3
4
5
UE X1 CAN THG SH W1
SC X11 CAN GOCII optional X2 CAN TSO
X12

6
7
8
9

6
7
8
9
X1.1 W2
X31

100X1:C2
W3

100X1:B2
400V in

24V U1 1

2
X1:6
X1:7

X1:9
X1:8
X2:6

X2:7

X2:9
X2:8

X16:3
100X1 1 2 3 4 5 6 SC2X2

100X1:A7

6 100X1:C7

10 100X1:C5
4 100X1:B7
5 100X1:B6

9 100X1:B5
ABC

5 24V U1
PO 0V

13 PO +V

X3:25

8 X3:14
X1:3
X1:2 1 X1:7

N2

L3
N1
2

L1

L2
X1:9

1
11

2
12

7
6
4
W3

1
Signal bus 3
ERBA SH

2
3

7
8
X3 X4 X3:18

11
12
10

14
15
Signal bus

13
& supply TSO X3:23

9
X23 4
X15 X3:16

14
210X1:B32 15
100X1:A5 16
17
100X1:A4 18
19

21
20

100X1:A6 22
100X1:C4 23
24
25
5 X3:9
X3:5

210X1:D12
X3:10 1 4 7 10
100X1:C8

GND U1
100X1:B8
X3:22 6
SCK1

X3:18
X4:10
X3:4 7 X3:6
X4:8

X4:1

R2
X3:3 8 X3:14
X13:3
100X1:C22

100X1:A14
100X1:B12
9 X3:25 2 5 8 11
R3 X9:2 X32

GND U1
X13:7

4 100X2:D28
10

2 100X1:A17

100X1:A23
R1

24V U1
X13:4

GND U5
24V U1 R1 X13:2

24V U5
24V U1
11
X6:5 X11:2
12

GND U1
X11:3 X13:5 3 6 9 12
SH

1
2
3
4
5
X13:8 13 R2
R2 1 R5 X6 sensing X6:6
X10 X6:4

5
14
SD X9:1

100X1:B13 6
100X1:B23 7
8
9
X5 exposure 15
24V U1

X12 X18:8

X6:8
R4 X15:9

110X1:D/B/Z30

110X1:D/B/Z28
6
7
8
9
16 X16:14
X12:5

R5
R4 17 X7:2

X18:7
X5:2
18 X16:11
R6 X16:2
19 X16:6
X12:13

4 100X1:A28
2 100X1:A27
3 100X1:B28
100X1:A21 X19:7
2 100X1:C16

4 100X2:B28
GND U5

X19:6
24V U1

20
24V U5
24V U1

X19:8 210X1:Z32
21 X15:4
X7:9
X34:8 22 X17:4
X17:3 23 X21:2 X33 1 2 3 4 not connected

5
1

X7 Wall bucky control SC U1 X8 Tilting and BFP 210X1:Z16 24 X6:7


X5:9 X12:6

6
7
8
9
25
6
7
8
9

X12:7 X12:9

100X1:C28
100X2:D18 100X1:C27

100X1:A29
26
R7 X12:8 X19:5
X20:6 27 X8:7
100X1:B14
100X1:A10
PO UREF

X8:2 28 X8:6
X8:4 29 X8:3
X8:8 30 X18:4

X37:6
X18:3
31

GND U3

GND U5

GND U4
X18:2

X12:12
SC1

24V U3

24V U4
245 U5
X12:14

24V U2
1

X18:7
2
3
4
5

X18:6 32
R1 X9 Poti tilting
7
8
9
6

1
2
3
4
5
SC1
X1 X10 Brake tilting

6
7
8
9
NG UREF
PO UREF
2 100X1:C11
3 100X1:A12

3 2 1 3 2 1

100X1:A24
100X2:D12

100X1:C23
100X1:A16

100X1:B24
24V U1

X20:4
X15:5
W1
X35 SC2 X34
SD ERAX7
1

4
5

6 5 4 24V in & 6 5 4
X1 X11 Poti FF & CF 24V in,
X4 Overvolt.
1
2
3

6
7
8
5
SD
4
Switches
6
7
8
9

11 X12 check
12
10

14
15
13
X4 & supply serial changer
9
100X1:C12
100X1:B11

100X1:A11

100X1:A19
3 100X1:B10

100X1:C24
1 NG UREF

9 8 7 9 8 7
X20:7

X35:6
X35:3
X20:3

GND U2
210X1

100X1:B21
24V U1
200X2

X15:2

GND U1
24V U1
2

4
5

SH Poti BFP &Table lateral


Z B D
X13 & BFL/ I.I./TV generator BFP Z B D
2
3
4
5

SC1 Safety switch


X1 X14 X37:9
1
6
100X1:C10 7
100X1:A13 8
9

& BFL/ I.I./TV tilting


S1 X16:9 2 2 2 V6
6
7
8
9

24V U3 2 2 2 24V U1
PO UREF

GND U1

210X1:B14
V5

210X1:D10

210X1:D18

210X1:D14
210X1:B10

210X1:B18
210X1:D6
210X1:B6
4 4 4 GND U1

24V U3
24V U4
100X2:D10
100X1:A18

100X2:B20
100X1:B18

24V U2
GND U3 4 4 4 GND U1

X10:1
24V U1
W2:2

V6

X38:3 6 6 6 X37:4
X16:3 6 6 6 210X1:Z30
100X1:B16

8 8 8 GND U2
1
100X1:B20
24V U1

2
3

6
7
8
4
5
100X2:D24

SCK1/K2/3/K6 8 8 8 210X1:D26
X16 3 2 1 3 2 1
11
12
10

14
15
13
X35:7

X12:4

X38.6 10 10 10 X37:2 X36


210X1:Z26 10 10 10
100X1:A31 100X1:C18

100X1:B15

X16:1
SDX21
100X2:Z28
210X1:Z2

100X2:Z6

Motors
210X1:B26
FF & CF
1
2
3
4
5

SH X15 Emergency OFF 210X1:D22 12 12 12 210X1:D24 12 12 12 GND U3


X37 6 5 4 6 5 4
X9 X12:2 SCK1/K2/K3/K6
6
7
8
9

210X1:Z24 14 14 14 210X1:B24 14 14 14 X36:1


4 100X1:C30
3 100X1:B30

210X1:B22

210X1:D20

210X1:D16
210X1:B20

210X1:B16
4 100X1:C21
3 100X1:A22

8 100X2:B24
2 GND U1

X20:8 16 16 16 210X1:Z22 100X1:B22 16 16 16 X36:4


1 24V U1

24V U5

9 8 7
X38:9
X38:11 18 18 18 X18:6 24V U1 18 18 18 X36:2
1
2

210X1:B2
210X1:B4
SCU3 X18 X38:10

X38:7
X38:4
UE BFL
7
5
6

X17 Supply & contr. X19:2 20 20 20 X23:6 20 20 20 X36:5 X38 Motor table lateral, Brakes BFP
6
7
8
9
11
12
10

14
15
13

X1 & BFL/ I.I./TV & Clutch BFL/ I.I./TV


9

THG X16:5
100X1:C31
100X1:A32

X38:12 22 22 22 100X2:D16 22 22 22 102X2:Z12


100X1:C15

X19:3 210X1:Z6 24V U4


W4

3 2 1
5 100X1:B25

X19:4 102X2:B14
3 100X2:D22
4 100X2:B22
2 100X2:Z20

210X1:Z8
24V U1

24 24 24 X15:6 100X2:Z14 24 24 24 102X2:B12


X17:8 102X2:D14

210X1:D2
210X1:Z2
1 24V U1

5 24V U5

X22:2 26 26 26 100X2:Z10 26 26 26 102X2:D8


X14:1
X12:3

X16:4
210X1:Z10 X37:8
1

102X2:B10 6 5 4
not connected 210X1:Z12
X19 X16:10 28 28 28 X5:4 28 28 28
UB
2
3
4
6
7
8
9

Brakes & switches X7:4


X1 X20 collision long column 30 30 30 100X1:C20 30 30 30 PO 0V V5 V6
100X1:A20
100X1:B19

GND U1
100X1:C19

6
7
8
9

V4 PO+V 100X2:B16 X37:7


9 8 7
100X1:A25

100X2:Z16

24V U1 32 32 32 100X1:C20 32 32 32 X4:10


X12:11

X3:15

100X2:Z22 100X2:B18
12 11 10
2 100X1:C22

Geometry Relais
2 100X2Z26
24V U1

control unit interface 100X2:Z18


1 24V U1

Door V4 V7 15 14 13 SHX24
1

X-Ray
3
4
5

3
4
5

contact X21 indicat. X22


Schr.

6
7
8
9

7
8
9

100X2:B30
6
24V U5
A1/A3 01-12-04

X45
1

1
2
3
4
5

2
3
4
5

UDX1 X23 Brakes tube vertical X24 not connected

SC Backpanel
6
7
8
9

6
7
8
9

SC Backpanel
100X2:D20

Geometry control unit

Duo DIAGNOST (b/01.0) Z2-2.1


c Philips Medical Systems
ALL RIGHTS RESERVED
ERAX8 ERAX1
3,15AT 45V 0

1 4 7
X3 2 5 8
3 6 9

X1 1 2 3 4 5 6
1 4 7 10
X4 2 5 8 11
3 6 9 12
X2 1 2 3 4 5 6

SCX35 SCX31
0 230V 2,5AT
A4 01-11-29 Schr.

SC2
Power supply

Duo DIAGNOST (01.0) Z2-2.4


C Philips Medical Systems
BA = Image intensifier
BD = Photosensor
BGC= II Tube 23/31/38 cm
CB = System controller Cockpit
UB UD CG = Geometry operation control
CGII= Geometry operation control II optional
Generator rack E Option rack R FS = Foot switch
Remote service FSII = Foot switch II optional

LA = Collimator
PH = Bucky measuring chamber
ERBA
SC = Geometry control unit (GCU)
UE1 SC1 = Tilting drive
ERA SC1.1 = Rear, front cover E-Box
EZ SC2 = Power supply, optional for serial changer
SD = Connection field tower
SD1= Fulcrum frontal drive
SD2= Compressor frontal drive
UA XA LA UE PH SH1 SD3= Safety end switches BFP /Ang.
EN SD2 SD VA SV SH SH4 SD4 = Compressor frontal force switches
RC SD4 SF1
SF SE = Emergency stop
SD1 SF3 SF = Bucky frame park drive
EQ SH3 SF1 = Bucky frame long./ II long. drive
SH = Coding field table frame
SD3
SH1 = Safety end switch footend/ headend
Bucky frame long./ bucky frame park/
SH1 II long. drive
G SH2= Table side operation (TSO)
SF1 SH3= Table lateral drive
SH4 = Potential equalization
UC SQ = Collision switch II
SV = Scattered radiation protection park switch
UA = Tube arm coupling
UB = Position and safety switch column
UC = Brake and collision switch tube longitudinal
UD = Brake tube vertical
UE = Tube handle grip (THG)
SC2 SC SC1.1 SC1 BA BGC SQ BD WK SH2 SE UE1 = Tube holder
VA = Bucky
CG II (GOCII)
optional WK = TV-camera
XA = X-ray tube

ERA = Power supply for geometry and II/TV


ERBA = II/TV control unit / CPX-Rack
RC = Digital X-ray imaging (DXI) optional
Trolley
VA
GOC

Monitor

CG CB
PH PC-
Keyboard
Schr.

(DIM)
FSFL II
optional
01-12-04

FSFL
Survey of units
A2/A3

Central Part X (c/01.0) Z1-0.1


c Philips Medical Systems
ALL RIGHTS RESERVED
CBX60
EZX23
RCX6 (DXI) option
SCX3
LAX1 rear side CGIIX5 option
CGX1 CBX110
SC2X3
SCX34

X1 X30 ERAX3
ERA Power supply or
EZX42
X20 21 22 23 24 X2 X33
X7 X8
X13B X13A X14 X4 X6 X34
X104 X9 X11
X10 X11 BGCX1
T1 X15 WKX1
X105
X3 BDX1
option 2nd GOC ERBA
RAX10.1 CPXII/TV Controller
CGX1 EZX61
RAX10.2 Monitor
ENX2100 X103 CGX1.2
L1' L2' L3' N
X102 X108
L1 L2 L3 loard
ERA102
X107
X203

L1
PE N F1 L1 Backpanel EZ Basis rack
X1
T1 T2 T3
line X4 X5 X6 X2 X3
L1 L2 L3 N

230V monitor
230V remote service
230V 2nd monitor
RCX4 (DXI) option
ERBAX30
SC2X1
EY Rotor control

EQ Converter

G H.V. generator
Schr.
01-12-04
A3/A4

E-Cabinet
Central Part X (b/01.0) Z2-0.2
c Philips Medical Systems