Professional Documents
Culture Documents
CASE™ V6.7
Exercise Testing System
2060290-235 Revision 3
NOTE: The information in this manual applies only to CASE system version 6.7 with product code STE. Due to continuing
product innovation, specifications in this manual are subject to change without notice.
CASE, MAC, MARS, and MUSE are trademarks owned by GE Medical Systems Information Technologies, Inc. a
General Electric Company going to market as GE Healthcare.
2 Equipment Overview
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Revision 3 CASE 1
2060290-235
Motherboard Power and I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
3 Installation
Installation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Installing the Keyboard and the Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Connecting the System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Mains Connection, Power Up, Functional Test . . . . . . . . . . . . . . . . . . . . . . . 3-8
4 InSite
Installing InSite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Pre-Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
How to Activate the InSite Taskbar Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
2 CASE Revision 3
2060290-235
InSite Configuration Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
5 Maintenance
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Required Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Cleaning/Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Interior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
6 Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Revision 3 CASE 3
2060290-235
Windows 10 IoT Enterprise Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Repairing the Custom Windows 10 IoT Enterprise Image Installation . . . . . . . . . 6-14
Setup Windows 10 IoT Enterprise for CASE after Repair . . . . . . . . . . . . . . . . . . 6-16
Activation of Windows 10 IOT Enterprise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Installation of the CASE V6.73 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Logon to the Repaired CASE System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
Installing InSite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
7 Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
System Backup and Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Saving to SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Saving System Configuration to the SD Card . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Saving Custom Settings to an SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Restoring from SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Restoring System Configuration from SD Card . . . . . . . . . . . . . . . . . . . . . . . 7-7
Restoring Custom Setups from SD Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Recovering from a CASE BTRV Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Remote Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Controlling Electrostatic Discharge Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Replacing the Display Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Replacing the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
Removing the Flat Panel Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
Mounting a Flat Panel Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Servicing the Keypad Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
Removing the Worksurface Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Thermal Writer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Servicing the Power Distribution Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
Replacing the Solid State Drive (SSD) FRU . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-25
Servicing the I/O Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
4 CASE Revision 3
2060290-235
8 System Rebuild
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
CASE V6.7 and Windows 10 Repair Instructions Product Code STE . . . . . . . . . . 8-3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Process Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Replacing the Solid State Drive (SSD) FRU containing the CASE Custom Windows
Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
First Setup of Windows 10 IoT Enterprise for CASE . . . . . . . . . . . . . . . . . . . . . . . 8-6
Installation of the CASE V6.73 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Replacing the Hard Disk Drive (HDD) FRU containing the CASE Custom Windows
Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
First Setup of Windows 10 IoT Enterprise for CASE . . . . . . . . . . . . . . . . . . . . . . . 8-8
Installation of the CASE V6.73 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Revision 3 CASE 5
2060290-235
Power Cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
CASE Interface Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
FRU Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-10
A Electromagnetic Compatibility
CASE System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Guidance and Manufacturer's Declaration – Electromagnetic Emissions . . . . . . .A-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity . . . . . . . .A-4
Recommended Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-6
Compliant Cables and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-7
B CASE Files
CASE Files–Directory Structure and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
File name nomenclature of the examination procedure . . . . . . . . . . . . . . . . . . . . .B-4
Files in the Application Data Folder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-4
C Miscellaneous
DICOM Interface Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Terminology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
Special DICOM setup items without UI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-5
Protocol Code Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-6
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-6
Method of Protocol Code mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-7
Syntax of the Protocol Code mapping per scheme: . . . . . . . . . . . . . . . . . . . .C-7
Adding a new Protocol Scheme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
Default protocol scheme and codes used in DICOM.INI . . . . . . . . . . . . . . . .C-8
Description of dialogs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
DICOM tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-12
Network configuration - DSS/Order Filler, PPS Manager . . . . . . . . . . . . . .C-16
Network Configuration - Image Manager, Image Archive . . . . . . . . . . . . . .C-17
Worklist customization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-18
6 CASE Revision 3
2060290-235
Data String Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-20
Shared File Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-22
Data File Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-22
General Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-24
Reloading CASE Application for Connection to Network Folder . . . . . . . . . . . . .C-24
Reconfigure the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-26
User Configuration (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-26
Install Option Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-26
Revision 3 CASE 7
2060290-235
8 CASE Revision 3
2060290-235
1 Introduction
Manual Information
Revision History
Each page of the document has the document part number and revision letter at the
bottom of the page. The revision letter identifies the document's update level.
3 2020-01-31 Updated the “Service Manual Language Information” section for Greek language.
Added “Setup Windows 10 and CASE Application” section in chapter 3.
Updated “Setup Windows 10 IoT Enterprise for CASE after Repair” section in chapter 6.
警告 本维修手册仅提供英文版本。
ZH-CN 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式
的伤害。
警告 本维修手册仅提供英文版本。
(ZH-TW) 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
不重視本警告可能導致維修人員、操作人員或病患因電擊、機械因素或 其他
因素而受到傷害。
UPOZORENJE Ove upute za servisiranje dostupne su samo na engleskom jeziku.
(HR)
Ukoliko korisnički servis zahtijeva neki drugi jezik, korisnikova je
odgovornost osigurati odgovarajući prijevod.
Nemojte pokušavati servisirati opremu ukoliko niste konzultirali i razumjeli ove
upute.
警告 このサービスマニュアルは英語版しかありません。
(JA)
サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業
はその業者の責任で行うものとさせていただきます。
このサービスマニュアルを熟読し、十分に理解をした上で装置のサービス
を行ってください。
この警告に従わない場合、サービスを担当される方、操作員あるいは患者
が、感電や機械的又はその他の危険により負傷する可能性があります。
Manual Purpose
This manual supplies technical information for service representative and technical
personnel so they can maintain the equipment to the assembly level. Use it as a
guide for maintenance and electrical repairs considered field repairable. Where
necessary the manual identifies additional sources of relevant information and or
technical assistance.
See the operator manual for the instructions necessary to operate the equipment
safely in accordance with its function and intended use.
Intended Audience
This manual is intended for the person who uses, maintains, or troubleshoots this
equipment.
Illustrations
All illustrations in this manual are provided as examples only. They may not
necessarily reflect your equipment setup or data displayed.
In this manual, all names appearing in examples and illustrations are fictitious. The
use of any real person's name is purely coincidental.
Safety Information
Responsibility of the Manufacturer
GE Medical Systems Information Technologies, Inc. is responsible for the effects of
safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs are
carried out by persons authorized by GE Healthcare.
The electrical installation of the relevant room complies with the requirements
of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
General
This device is not intended for home use.
Parts and accessories used must meet the requirements of the applicable IEC 60601
series safety standards, and/or the system configuration must meet the requirements
of the IEC 60601-1-1 medical electrical systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all functions.
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety of the resulting
system. Consideration relating to the choice shall include:
If the installation of the equipment, in the USA, will use 240V rather than 120V, the
source must be a center-tapped, 240V, single-phase circuit.
WARNING
SHOCK HAZARD – Disconnect power before replacing fuses.
WARNING
FIRE HAZARD – For continued protection against risk of fire
replace fuse only with fuse of same type and rating.
Term Definition
DANGER Indicates an imminent hazard which, if not avoided, will result in death or
serious injury.
WARNING Indicates a potential hazard or unsafe practice which, if not avoided, could
result in death or serious injury.
CAUTION Indicates a potential hazard or unsafe practice which, if not avoided, could
result in minor personal injury or product/property damage.
NOTE Provides application tips or other useful information to assure that you get
the most from your equipment.
Equipment Symbols
On/Standby
Signal input
Signal output
Fuse
12V 12 V DC
The number found under this symbol is the date of manufacture in the YYYY-
MM format.
Manufacturer's identification
Order number
Revision
Serial number
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed as unsorted municipal waste and must be collected
separately. Please contact an authorized representative of the manufacturer
for information concerning the decommissioning of your equipment.
This symbol indicates that the device was tested and certified according to
the Russian safety standards specified by GOST.
Mass
Service Information
Service Requirements
Follow the service requirements listed below.
Refer equipment servicing to GE Healthcare authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that
warranty.
It is the user's responsibility to report the need for service to GE Healthcare or
to one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using
this equipment to implement a satisfactory maintenance schedule may cause
undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the
equipment will always be functional when required.
Equipment Identification
The Serial Number and Unit ID number tags are on the back of the device in the
location shown below.
Serial
Number
location
A B C D E F
A Product Code
B Year Manufactured (00-99)
00 = 2000
01 = 2001
02 = 2002
(and so on)
C Fiscal Week Manufactured
D Production Sequence Number
E Manufacturing Site
F Miscellaneous Characteristic
General Description
This section is an overview of a CASE system. See the “CASE V6.7 Operator's
Manual” for details on operating procedures.
Front View
1 Monitor
2 Function keypad
5 Paper tray
Rear View
DP video
1 Monitor AC power and worksurface light cables are internally
routed in display arm
3 Access panel
Assembly Descriptions
Familiarity with the components of the CASE assembly will aid in troubleshooting
the equipment. The CASE consists of the following components.
Monitor
CASE product code STE uses a 22 inch widescreen monitor rated IEC-950 universal
AC input. It has an IEC320 power cord and a DP (Display port) connector on a cable
for connection to the motherboard DP (DisplayPort). It displays optimal waveform
quality at the system setting of 1680 × 1050 resolution.
Keypad
The keypad is designed as a bus powered, low speed USB peripheral using a
Cypress MCU. There is protection from ESD on the PCB as well.
Thermal Writer
The thermal writer assembly contains a thermal print head with cue hole sensor, a
motor/encoder assembly to drive the platen roller, and a GE Healthcare designed
PCB to receive scan line data and motor commands from the PC over USB.
The PCB is based on a Cypress high speed USB MCU with an 8051 core. A type 'B'
USB connector provides the link to the PC.
The functioning of this USB writer PCB depends on communication with the host
PC motherboard USB controller which is under the control of a software driver. A
special driver is always needed to communicate with this writer PCB (this driver is
built into the application software). Check operation in the CASE system, since full
functionality depends on cooperation with the motherboard USB ports and
associated software drivers.
Installation
1. Open the paper compartment lid and fix the small A4 paper adapter by using
the two M3x8 screws.
2. Screw the large A4 paper adapter to the left side of the paper compartment by
using the two M3x5 screws.
Power System
AC power comes into the unit through an IEC-320 inlet module that has an
equipotential ground stud mounted near it.
routes AC power to and DC power from the 24Vdc medical grade power supply
routes AC power to the isolation transformer to provide isolated AC power to
the IEC-950 flat-screen monitor
provides 24Vdc to the thermal writer
provides 12Vdc, -12Vdc, 5V standby, 5Vdc, and 3.3Vdc to the motherboard
and peripherals
Provides EMI filtering and over current protection
NOTE
No switching of AC power is needed with AC inputs ranging from 100-240
VAC since all supplies, regulators and isolation transformer contained in the
unit are universal input type.
Standard PC Components
PC Subsystem
The PC subsystem provides processing, video, and network capability using
industry standard PC components.
Acquisition/Interface Card
This PCB plugs into an expansion slot on the motherboard and provides an interface
to the PC that allows CAM-14 acquisition and analog/TTL I/O.
Keyboard
Medical grade Windows keyboard with USB interface.
You must use the USB-to-PS/2 adapter shipped with your system to connect the
keyboard to the PC I/O panel PS/2 keyboard connector.
NOTE
Do NOT connect the keyboard to a USB port.
Mouse
Two-button USB mouse with scroll wheel.
You must use the USB-to-PS/2 adapter shipped with your system to connect the
mouse to the PC I/O panel PS/2 mouse connector.
NOTE
Do NOT connect the mouse to a USB port.
USB Ports
There are two USB ports located next to the system power button on the right side of
the CASE system. The USB ports connect internally to the motherboard.
SD Card Reader
The external SD Cards are read using an adapter in the USB ports on the right side
of CASE system.
Connections
Shown below is the CASE rear connector panel. See Chapter 3 of this document and
the CASE Operator's Manual, “B Miscellaneous, Connecting Peripheral Devices,”
for information on connecting peripheral devices to the CASE.
Connectors Usage
COM 1 Treadmill
Mouse NOTE:
Mouse use PS2/Adapters, DO NOT USE the USB ports
LAN
COM 2 Tango BP
Connectors Usage
Analog I/O Common output for Echo machines and Tango. Splitter cable
needed if connecting both
COM 3
COM 4
COM 5
Com 6
Audio.Speaker
4 x USB
Keyboard NOTE:
Keyboard use PS2/Adapters, DO NOT USE the USB
ports.
NOTE
A reference diagram of the CASE’s back panel, complete with connector labels,
is located inside the access door to the back of the unit. Refer to this diagram for
information on where to make system connections.
USB
Table 1. USB Jack or Plug
Pin Name
1 Vcc
2 - Data
3 + Data
4 Ground
8 CH4OUT
9 AN1IN
SHLD CHASSIS
Pin Name
2 GND
3 AMTX+
4 AMTX-
5 AMRX+
6 AMRX-
7, 8, 9, 10 N/C
Pin Name
1 Main Link_Lane 0+
2 GND
4 Main Link_Lane 1+
5 GND
6 Main Link_Lane 1-
7 Main Link_Lane 2+
8 GND
9 Main Link_Lane 2-
10 Main Link_Lane 3+
11 GND
12 Main Link_Lane 3-
13 CONFIG 1
14 CONFIG 2
15 AUX CH+
16 GND
17 AUX CH-
19 Return
20 DP_PWR
DVI
C3
24 17
C5 C4
9
C1 8 1
C2
Table 5. DVI
Pin Name
1 T.M.D.S. DATA 2-
2 T.M.D.S. DATA 2+
4 T.M.D.S. DATA 4-
5 T.M.D.S. DATA 4+
6 DDC CLOCK
7 DDC DATA
8 ANALOG V-SYNC
9 T.M.D.S. DATA 1-
10 T.M.D.S. DATA 1+
12 T.M.D.S. DATA 3-
13 T.M.D.S. DATA 3+
14 +5V POWER
15 GND
17 T.M.D.S. DATA 0-
18 T.M.D.S. DATA 0+
20 T.M.D.S. DATA 5-
21 T.M.D.S. DATA 5+
23 T.M.D.S. CLOCK+
24 T.M.D.S. CLOCK-
C1 ANALOG RED
C2 ANALOG GREEN
C3 ANALOG BLUE
C4 ANALOG H-SYNC
C5 ANALOG GROUND
Pin Name
1 +3.3V
2 +3.3V
3 GROUND
4 +5V
5 GROUND
6 +5V
7 GROUND
10 +12V
11 +12V
12 +3.3V
13 +3.3V
14 -12 V
15 GROUND
17 GROUND
18 GROUND
19 GROUND
20 -5V
21 +5V
22 +5V
23 +5V
24 GROUND
Pin Name
1 GROUND
2 GROUND
3 +12V
4 +12V
Pin Name
1 DATA
2 N/C
3 GROUND
4 +5V (fused)
5 CLOCK
6 N/C
Table 9. COM 1 - 6
Pin Name
1 DCD #
2 SERIAL IN
3 SERIAL OUT
4 DTR #
5 GROUND
6 DSR #
7 RTS #
8 CTS #
9 RI #
Installation Overview
SYSTEM WARNINGS
EXCESSIVE LEAKAGE CURRENT – A secondary display
or printer that is a non-medical grade device and is used
within the patient environment, must always be powered from
an additional transformer providing at least basic isolation
(isolating or separating transformer). Using without an
isolating transformer could result in unacceptable enclosure
leakage currents.
InSite is the service option that allows diagnostics for the CASE system. Refer
to Chapter 4, “Installing InSite” for details on installing and configuring InSite.
8. Configure the CASE system to use the DICOM interface (CASE V6.61 and
higher). Refer to “DICOM Interface Configuration” on page C-3 for details.
9. Configure the CASE system for an EMR application, if applicable.
By interfacing with an EMR application, the CASE system can be controlled by
and transfer data with the EMR application. Refer to “Configuring a CASE
System with the EMR Application” on page C-19 for details.
After the installation is complete, the CASE system is ready for the user.
Assembly
For this new CASE product code STE, first time assembly has been simplified as
follows:
"The display arm and display have been fully assembled and mounted at the
factory.
"The DP video cable contained in display arm has been connected from the
display to the motherboard.
"The AC power cable contained in display arm has been connected to the
display.
3b
3a
28A
161A
NOTE
The provided cable clamp is for the following cables: ACQ, Treadmill and
Network cables.
Cable
clamp
Reference
diagram
2-15A
A reference diagram of the CASE's back panel, complete with connector labels,
is located inside the access door to the back of the unit.
NOTE
It is recommended you label the mouse and keyboard connectors that
connect to the USB-to-PS/2 adapters.
NOTE
Coil the excess cable for storage as shown above. DO NOT squeeze the
cables when closing the access panel.
Mains
connection
(power line
Mains power switch connection)
Potential
equalization
system
connection
Back Panel
Power
button
Side Panel
1. Connect the CASE to the power line and fasten with the clamp provided.
NOTE
DO NOT energize the system at this time.
6. Login to system.
If logging in as Admin, double click on the CASE icon when it displays.
Select the Admin icon and enter admin1,3,5,7 for the password.
If logging in as GEService, double click on the CASE icon when it
displays. Select the GEService icon and enter service for the password.
If logged in as User double click on the CASE icon when it displays. The
initial CASE screen will open.
7. Modify the date, time and time zone (in the lower right side of the Windows
display), if applicable.
NOTE
Do not turn off the mains power switch until you have:
The power button on the side panel does not disconnect the system from
AC power.
7. Click Change.
9. Click OK.
A message “You must restart your computer to apply the changes” appears.
10. Click OK.
12. On the home screen, right-click This PC and then click Properties.
13. Verify that the entered serial number appears on the system properties screen.
4. From the drop-down menu, select the preferred regional date and time format.
9. Click Apply.
Region Settings dialog box appears.
10. From the drop-down menu, select the preferred region settings.
5. Click Programming.
6. Click Close after the Programming Status field displays PROGRAMMING OK.
7. Launch the service tool (ServiceTool.exe) form the CASE CAM-USB ACQ
Board Manufacturing Tools folder in the C drive.
14. Shut down the system. Wait 5 seconds before restart. The default login screen
opens.
NOTE
In the proceeding window, there may be a warning symbol next to the GE
Pegasus Thermal Writer USB Driver. This warning is normal and it is OK
to proceed; it will go away after the CASE 6.7 Application Software is
installed and the system reboots as covered later in this manual. Close the
Device Manager window.
You are now ready to install the CASE application by using the CASE
Application DVD Advantech V6.73- P/N 2109926-006
d. Click Next.
e. Call on the suggested phone number presented on the screen and follow the
You are now ready to install the CASE application by using the CASE Application
DVD Advantech V6.73- P/N 2109926-006.
.
2. At the Choose Program Files Destination prompt, click Next to accept the
D:\CASE (default) as the Destination Folder.
b.Select the Drive and Folder for the mapped database directory, check the
Reconnect at login box.
c.If needed, click the Connect using a different user name link and enter a
user name and password that will connect to the network folder at
login. (The default user is Administrator.)
d.If needed, click the Sign up for online storage or connect to a network
server link to add a network place through the Network Place Wizard.
e.Click Finish and Next.
4. At the Setup Type window, select US Factory Setup or European Factory Setup
as appropriate for the installation, and click Next.
5. Click on Install.
NOTE
External SD card reader which initially shipped with the CASE unit is required
for this section.
You can print the option codes by selecting the Print icon or you have to manually
write down all of your option codes.
1. From within the CASE application, select the System Configuration icon.
The System Configuration window opens.
2. Plug the external SD card reader into one of the USB ports on the right side of
the CASE unit, then insert a SD card into the SD card slot of the external SD
card reader.
3. On the System Configuration tab, select the Save icon.
A dialog window will open.
4. Select YES.
The message System Configuration successfully stored displays.
NOTE
Make sure to delete the three dots before entering the name.
CAUTIONS
Before installation, close ALL programs (CASE) on ALL
stations. The CASE server itself must not be used as a CASE
station. If an application is open on any system, it may cause
failure of installation.
The following table outlines the installation procedures for both network
configurations. They are presented side-by-side to call out differences and
similarities.
Create a Domain User or Domain User Create a user or create users and a user
group to be used by the CASE systems group on the server.
Create a shared folder on the server with full Create a shared folder on the server with
control permissions for the Domain User or full control permissions for the user or
Domain User group. user group.
Provide the server name (or IP address) and path to the created shared folder for
mapping a drive letter to the folder or UNC path.
Provide static IPs for each system (if Remote View function will be used).
NOTE
If existing CASE systems can communicate with the MUSE server, the MUSE
server access permissions will need to be reconfigured if a new CASE user is
created.
Domain Configuration
To install CASE on a domain, you first configure the server and then configure each
CASE system to be added to the network. A Domain Administrator may have to
perform both steps.
Set up Shared Folder – The procedure for setting up a shared folder differs
depending on whether you are installing a new CASE server or updating an existing
CASE server.
Verify the share permissions on the CASE Database shared directory currently in
use: The designated Network User or Network User Group must have full control
permissions.
a. First, you need to configure your networking settings, so that CASE is able
to properly communicate with your Domain Controller. To make sure that
CASE can reach your Domain Controller, try to ping it (Right-click Start >
b. Enter ping [your Domain Controller] and verify that the reply is working. If
the Domain Controller is unreachable, check your TCP/ IP settings, for
example, the DNS server configuration (Right-click Start > click Network
Connections > right-click the Connection you want to configure > select
Properties > select TCP/IP > click Properties).
Directory group).
Once User Name and Password are entered, click OK.
Note that, as shown in the following picture, your Domain Name (for example,
CARDIONET) is displayed under the User Name and Password.
If the local Computer Name is shown and not your Domain Name, then enter
the User Name in this format: [Domain Name]\[Network User].
12. Optional: To verify that CASE joined the Domain, open the basic system
properties page (right-click Computer > click Properties) and check your full
Computer Name as well as the name of the Domain.
13. Repeat the section CASE Configuration in a Domain for all systems before
proceeding.
Workgroup Configuration
To install CASE in a Workgroup, you first configure the server and then configure
each CASE system to be added to the network. An administrator may have to
perform both steps.
Verify the share permissions on the CASE Database shared directory currently in
use: The designated Network User or Network User Group must have full control
permissions.
Under Computer name, domain and workgroup settings, click Change Settings.
To configure the user accounts and network setting, use the following procedure.
This section must be completed if at least one CASE system added to the CASE
network has locally stored examinations. If all systems (CASE) and the shared
CASE server folder are new, this section can be skipped.
NOTE
This procedure copies the examination files and the custom setups from each
CASE system. However, it copies the system configuration, the user list, and
any custom configured reports from only one CASE system, which should be
designated as the “special station.” If the CASE network existed before this
installation began, it is recommended you use one of the CASE systems
NOTE
External SD card reader which initially shipped with the CASE unit is required
for this section.
NOTE
Refer to “Saving Custom Settings to an SD Card” on page 7-5.
Existing Environment
Adding a new CASE system to an existing network environment, do the following:
Load the system configuration that was copied from the special station as part
of the data backup process.
For more information, refer to “Restoring System Configuration from SD Card”
on page 7-7.
Verify that all settings are correct for each tab in the System Configuration
window.
Verify the connection from the CASE system to the MUSE server was not
broken, if the CASE system should communicate with MUSE.
For more information, refer to “Data Interface with MUSE” on page 3-47.
Existing Environment
Adding a New CASE system to an existing networked environment.
Load the custom settings that were copied from the special station as part of the
data backup process.
The custom settings can be saved to the CASE server from one CASE system
and then copied from the server to each CASE system. For more information,
refer to “Restoring System Configuration from SD Card” on page 7-7.
Verify that the settings are correct for each tab in the Setup window.
Installing InSite
NOTE
On CASE systems, InSite is pre-installed, but not configured. In InSite, the
CASE Program directory and the CASE Database directory point to D:\CASE
by default.
If you encounter any problems with InSite (for example, after reinstalling the
CASE application with the Network CASE Database Directory), uninstall
InSite (see “Uninstalling InSite” on page 4-14). After uninstalling, reinstall
InSite as described in section “Installation Procedure” on page 4-5).
Introduction
After the installation of InSite two services are available for use by GE Service:
This document outlines the steps involved in installing InSite for CASE Windows
10 IoT Enterprise.
Pre-Installation Checklist
NOTE
During installation, InSite will use the following information as default values
(CASE Program Directory, CASE Database Directory, Proxy Server Address
and Proxy Server Port of Internet Explorer). To successfully complete the
installation of the GE CASE InSite Agent, you should have this information
available. To complete the installation of the GE CASE InSite Service Agent,
you will need the following information.
1. Login as Administrator.
2. Start CASE application.
3. Select System Configuration.
4. On the General tab, fill out the Institution Name and Address fields (Name,
Street, City).
5. Exit CASE application.
6. If a proxy server will be used, please contact your site system administrator for
installation details. The site system administrator should determine whether the
proxy server requires authentication (method, user name, password).
Guidance for proxy server settings to be used by site system administrator:
a. Start Internet Explorer. Select Tools > Internet Options > Connection tab
>L AN Settings button.
b. Select the Use a proxy server for your LAN… checkbox and enter the proxy
server address and port provided by the site system administrator.
NOTE
In order to operate InSite, administrator rights are necessary.
Installation Procedure
1. Log on to the system.
a. Select Start > Run > Browse.
b. Locate the file <drive>:\CASE V6.7\InsiteExC\setup.exe.
4. Check whether the InSite desktop icon (A) has been installed.
A
B
2. Click the Yes button to acknowledge the upcoming User Account Control
window.
3. Check whether the InSite taskbar icon (C) has been installed.
3. Check the task bar: the InSite task bar icon (D) must be visible.
1. Login as Administrator.
2. Right-click the InSite icon and select Run As Administrator.
3. Left-click Options … (B).
4. Select the Automatically start InSite (i.e. this) application check box.
5. Select the Automatically start InSite and Remote Desktop Services check box.
NOTE
The Manual Mode is the default mode of the InSite Services.
Check the InSite taskbar icon to find out whether the Services have been
started.
1. Login as Administrator.
2. Right-click the InSite icon and select Run As Administrator.
3. Left-click Options … (B).
4. Clear the Automatically start InSite (i.e. this) application check box.
5. Clear the Automatically start InSite and Remote Desktop Services check box.
NOTE
With this configuration, the InSite and VNC Server services will not
automatically start up after a reboot.
To reactivate the services, select the Automatically start InSite and Remote
Desktop Services check box and restart your PC.
Check the InSite taskbar icon to find out whether the Services have been
started.
Uninstalling InSite
1. Close all windows before uninstalling InSite.
a. Select Start > Run > Browse.
b. Locate the file <drive>:\CASE V6.7\InsiteExC\setup.exe.
c. Double-click the file name.
d. Click the OK button.
e. At the UAC prompt, click Yes.
3. Click Finish.
4. Check the task bar:
the InSite icon is no longer visible.
Introduction
A regular equipment maintenance program helps prevent unnecessary equipment
and power failures and reduces possible health hazards. This chapter contains
instructions for the following recommended maintenance:
WARNING
Failure– on the part of all responsible individuals, hospitals or
institutions employing the use of this device– to implement the
recommended maintenance schedule may cause equipment failure
and possible health hazards. The manufacturer does not, in any
manner, assume responsibility for performing the recommended
maintenance schedule unless an Equipment Maintenance
Agreement exists. The sole responsibility rests with the
individuals, hospitals, or institutions utilizing the device.
Cleaning/Disinfection
Warning
Shock Hazard — Disconnect the power cord from the wall outlet
before cleaning or disinfecting the system.
1. Wipe exterior of the device with a soft lint-free cloth, using the following
solutions as recommended in the APIC Guidelines for Selection and Use of
Disinfectants (1996):
0.55% Sodium hypochlorite (10:1 “household” bleach with water)
Any sodium hypochlorite wipe product that meets the above guidelinesof
can be used
NOTE
Wring excess disinfectant from wipe before using.
NOTE
Any contact of disinfectant solutions with metal parts may cause corrosion.
Cautions
Never immerse the device in liquid.
Do not pour or spray any liquid directly on the device or permit fluid to seep
into connections or openings.
Never use conductive solutions, solutions that contain chlorides, wax or wax
compounds to clean the device.
Never use solutions or products that contain the following:
Any type of Ammonium Chloride such as, but not limited to:
Dimethyl Benzyl Ammonium Chloride
Quaternary Ammonium Chlorise solutions
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol-based cleaning agents
Sodium salts
Never autoclave or steam clean the device.
Do not coil leadwires/cables tightly around device.
Interior Cleaning
Disassembly
To clean the interior, log off the CASE application, shut down the system, turn the
power switch to OFF, and unplug the unit from the power outlet. Remove the screws
from the CPU access door, then remove the door.
General
Check for dust buildup on the surfaces of the interior circuit boards, components,
and assemblies. Use commercially available compressed air to blow away
accumulated dust. Follow the manufacturer's directions for use.
Thermal Printhead
Clean the thermal printhead every three months, or more often with heavy use. A
build-up of thermal paper coating on the printhead can cause light or uneven
printing.
Visual Inspection
Perform a visual inspection of all equipment and peripheral devices daily. Turn off
the unit and remove power before making an inspection or cleaning the unit.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Verify that all cords and connectors are securely seated.
Inspect keys and controls for proper operation.
Toggle keys should not stick in one position.
Knobs should rotate fully in both directions.
What follows is a general description of the suggested tests; refer to the governing
standard for a more detailed description. Conduct the tests under normal ambient
conditions of temperature, humidity, pressure, and line voltage. The leakage
currents correspond to 110% of the rated voltage for the test unit. Most Safety
Testers take this into account; otherwise, the measured values must be calculated.
If the device has a power cord, the test measures the impedance between the cord's
protective earth pin and any accessible metal part which is protectively earthed. If
the device does not have a power cord, the test measures the impedance between the
AC inlet's protective earth terminal and any accessible metal part which is
protectively earthed. Conduct the test both with and without a power cord.
To conduct this test, the test circuit must conform to the following values:
50 Hz/60 Hz 25 A 6V
The test fails if the measured value is greater than indicated in the following table:
The following diagram shows the Measuring Circuit [M] required for leakage
current. The reading in mV corresponds to µA (leakage current). The Safety Testers
generally work with this Measuring Circuit [M] and the displayed values are already
converted to leakage current.
This test measures leakage current of the device's internal power supply. The device
has to be turned on and off, and connected to your safety-testing equipment.
Measurements should be taken under the following conditions (refer to the safety
tester diagram):
The test has failed if the measured values are greater than:
300 µA (UL)
Use the following settings on your Safety Tester during normal conditions:
Switch Setting
GND closed
S1 (neutral) closed
Use the following settings on your Safety Tester during single fault conditions:
Switch Setting
The test fails if the measured values are greater than indicated in the following table:
NOTE
For your protection, the following values of resistor R may be used:
During N.C., measurements should be taken under the following conditions (refer to
the safety tester diagram:
During S.F.C., measurements should be taken under the following conditions (refer
to the safety tester diagram):
The test has failed if the measured values are greater than:
N.C. S.F.C.
100 µA (AC, IEC & UL) 500 µA (AC, IEC & UL)
WARNING
PERSONAL SAFETY–MAINS VOLTAGE is applied to floating
input (patient connection) and patient cables/test plug conductors
and safety tester PATIN JACK.
This test performs leakage current tests during single fault conditions (S.F.C.) with
line voltage applied to the floating (patient connection) inputs.
In all cases, the leakage current is measured from the floating (patient) inputs jack of
the device to ground. To setup this leakage current test:
During S.F.C, measurements should be taken under the following conditions (refer
to safety tester diagram):
The test has failed if the measured values are greater than:
N.C. S.F.C.
After completing this test, disconnect the safety tester from line voltage before
disconnecting the patient cable/test plug and/or the device.
Overview
This chapter discusses the tools and techniques for troubleshooting problems with
the CASE system. The topics include:
All event logs, system settings and CASE configuration files will be automatically
saved to a single compressed ZIP file.
e.g., D:\CASE\LOG
Compatibility
The CreateSystemInfoZipFile.bat file is compatible with following systems:
D:\CASE\Support\
Usage
1. Log in as local administrator.
2. Double click on the file CreateSystemInfoZipFile.bat located in one of above
directories.
This will activate the process of creating the support information zip file.
Location of the generated zip file:
e.g., D:\CASE\LOG
<serialnumber>_<date>_<time>.zip
Troubleshooting Tools
CASE provides three groups of tools to help you diagnose issues with the system:
Service Screen
Diagnostic Tests
Windows 10 IoT Enterprise Repair
Service Screen
The Service Screen is used to help diagnose issues with the CASE system. It
provides access to the system log, key settings, software tests, and service functions.
It can be accessed in the General Settings via the “For Service” key. When prompted
for a password, enter helserv or case8k.
Element Description
Log Level Level on which the logging mechanism filters incoming messages
Error: Only messages classified as ERROR are written to the log
file
Warning: Error and warning messages are written to the log file
Information: All kind of messages are written to the log file
Enable key log If enabled, key events are written to the log file
Enable Log Messages If enabled, log messages are written to the log file
Number of log files Defines the number of log files kept for previous events (log
history)
Possible Values: 1 – 15 log files
Log files size [byte] Defines the maximum size of a single log file in byte
Possible Values: 10 – 1024 Byte
Browse Opens the currently selected log file (currently selected in the log
file drop down list) within the Microsoft Notepad.
Delete Deletes the currently selected log file (currently selected in the log
file drop down list)
Export Provides means to export one or more log files to a local directory
or SD card
CORINA time constant (HP This field does not pertain to CASE.
filter)
CAM-14 High Pass Filter User selectable High Pass filter for the CARDIO application and
Analog Out. The High pass filter allows the following cutoff
frequencies: 1: 0.01 Hz, 2: 0.05 Hz The default is 0.05 Hz. The
default should not be changed, if not requested by the customer.
Research Full Disclosure ECG \ONL500 stores Full Disclosure ECG with 500 samples per
(in directory \ONL500, second. \ONL2000 stores Full Disclosure ECG with 2000 samples
\ONL2000) per second.
Element Description
Test dictionary… Compares the Text Strings of the selected language with the actual
selected language of the system configuration.
Set CAM-USB/CORINA Serial Set the CAM-USB or the CORINA Serial number of a CAM-USB or
Number… CORINA from Service with Serial Number 0.
Delete Network Locks… All other stations of the network using this server must be stopped
before this Network Locks should be deleted.
Save Log Files Save the Log Files what have been activated from Engineering and
send this Log Files please to Engineering.
Store data cyclically during This setting enables continuous storage of test data during an
Exercise Test Exercise Test, thus limiting the loss of data if any issue occurs with
the system (system/application unresponsive) and the ECG data
cannot be saved in the test end phase.
When the system is restarted after a crash or unresponsiveness,
the temporarily saved Exercise Test will automatically be entered
into the database.
Max. number of retained The maximum number of retained backups can be defined in the
backups range from 1 to 5.
Transfer to MUSE: Use When tests are transferred to MUSE, the location number of the
Location Number from System current system configuration is used rather than the number that
Configuration was stored at the time the test was acquired.
Transfer to MUSE: Use Cart/ When tests are transferred to MUSE, the cart/device number of the
Device Number from System current system configuration is used rather than the number that
Configuration was stored at the time the test was acquired.
Diagnostic Tests
The diagnostic tests check operation of the acquisition module and board, writer,
and keypad. A burn-in test in the menu is for factory production only.
Acquisition Test
Click ACQ TEST at the bottom of the Service Screen. The acquisition test screen
opens.
Status—the fields in this window show the status of the CAM-USB acquisition
driver.
Lead Status—this window displays the lead impedance measurement. The normal
range of impedance is 0-999. If the box next to a lead is checked, an error code used
for reading samples is displayed in the Impedance column. For example if the LA
Lead is disconnected, the checkbox will be marked.
Analog I/O—the Channel fields for Output/Input are for test purpose only, it does
not affect the application.
TTL Pulse—when this button is pressed, a TTL Pulse occurs on the TTL output pin
of the analog output connector.
Dec. Gain/Inc.Gain—these buttons are used for the sensitivity of the gain for
displaying the ECG curve.
Keypad Test
1. Click KEYPAD TEST at the bottom of the Service Screen.
2. Press each key on the keypad, and watch the interactive screen and verify that
each corresponding key on the screen functions when the keypad key is pressed.
NOTE
The button with the red X is for the desk light and is crossed out on the
interactive screen. This is normal. Press this button on the keypad and
verify that the light under the monitor goes on and off.
Interactive Screen
The keycodes appearing in the field under the interactive screen are use by
software engineering to troubleshoot the buttons should the keypad test fail.
Thermal Writer
Click WRTR TEST at the bottom of the Service Screen.
If you run the thermal writer tests as user Admin or GEService, the following
message window shows up:
If you only want to run a speed test or a roller tests, then click on the OK button. The
message window will disappear and the writer tests can be run without any issue.
Print Test
1. To run a Print Test, select Writer Speed from the pull-down list. There are 4
selectable writer speeds you can choose before running the Speed Test.
2. Next select Speed Test to print out a test graph. The speed selected determines
the print density (light/dark) and how fast the print runs.
Roller Test
The Roller Test option allows you to test the printer. When selected, a full page test
prints allowing you to view the density (light/dark) of the roller.
The screen provides information about the writer settings. If the printhead or the
PCB of the writer is replaced, the resistance value must be adjusted.
If the default resistance is not adequate, the value can be changed to another stage.
If you need to adjust the printhead resistance value, run the CASE application as
Administrator, not as User, Admin or Service.
1. Select System Configuration and Quit Program to close the CASE application.
2. Right-click the CASE icon.
3. Select Run as administrator.
This will restart the CASE application and the writerutil tool will now give you
the possibility to adjust the printhead resistance values.
4. Enter the value and then select Recalc and Save.
5. Finally, select System Configuration and Quit Program to close the CASE
application.
0=r 5=m
1=z 6=f
2=x 7=n
3=v 8=a
4=c 9=s
Example: the six-digit number for January 15, 2002 is 021501. Using the chart
above, the password would convert into “rxzmrz”.
There can also be an issue restoring the Windows installation due to a lack of
disk space on drive C:.
NOTE
Be aware that every additional application and all related data will get lost when
using the repair installation method. It is mandatory to save all documents and
files before you do the following installations.
Only the old Windows files will be stored in the folder WINDOWS.OLD and
will still be accessible. This folder will increase the used disk space on drive C.
Delete this WINDOWS.OLD folder if there is a need to do a Repair Installation
a second time.
a. Restart the CASE system and press F8 key continuously during startup.
The Advanced Boot Options dialog box is opens.
b. Press the Escape key.
The Windows Boot Manager dialog box is opens.
NOTE
Do not install Windows 10 IoT Enterprise from an original Microsoft Windows
10 IoT Enterprise DVD.
3. Leave the product key field blank and select Do this Later.
5. Click on Admin and then login by typing admin and press Enter.
6. In the Enter Computer Name field, type the CASE serial number and click OK.
NOTE
Make sure that you enter the correct computer name that is same as the serial
number labeled on the back panel of the equipment.
7. Click on Admin and type in the admin password and then, press enter. You must
change your password when prompted.
13. In Select Target System, select CASE and verify that a generated serial number
displays under status.
14. Click Programming to verify that the Generated Serial Number is programmed.
(This will not be the same as the 13-digit serial number).
15. Click Close after the Programming Status field displays Programming OK.
Configuration of the system is finished and the system displays the startup
screen.
16. Shut down the system. Wait 5 seconds before restart. The default login screen
opens.
17. After login, check the USB devices in Device Manager.
a. Right-click on Start button, select Device Manager.
b. Open Universal Serial Bus controllers.
c. Check for GE Pegasus Thermal Writer.
NOTE
In the proceeding window, there may be a warning symbol next to the GE
Pegasus Thermal Writer USB Driver. This warning is normal and it is OK
to proceed; it will go away after the CASE 6.7 Application Software is
installed and the system reboots as covered later in this manual. Close the
Device Manager window.
2. Double-click the CASE shortcut icon on the desktop (to launch the CASE
system) and verify the initial screen lists
GE CASE 6.73.
3. Close the CASE application to return to the Windows desktop.
Installing InSite
For installing instructions, refer to “Installing InSite” on page 4-3.
NOTE
After the successful Windows 10 repair, the first start of the InSite installer will
uninstall the previous InSite installation. Therefore, to properly install the InSite
software, the installer has to be run a second time.
NOTE
Take the time to make all the recommended visual checks (refer to the table
below) before starting any detailed troubleshooting procedures
Power-up Self-test
On power-up, the system automatically runs an internal self-test. If all circuits test
good, the start up screen displays.
General Troubleshooting
When activating the REORG procedure you can reorganize defective BTRIEVE
files from CASE. This repair procedure may become necessary when the data
becomes damaged for some reason or another (e.g. after a power failure). To carry
out the repair procedure successfully, CASE must first of all be terminated. There
must be sufficient memory space on your SSD.
As a general rule, the following applies: memory space required = twice as much
free memory space as the sum of all CASE files with the suffix ”BTR”.
The original CASE files are still available in the form of “OLD” files after
reorganization and should only be deleted when the reorganization has been
completed successfully.
Problem: System displays a message indicating the test is open at another system. Users
confirm the test is not open at any other CASE or client PC.
Solution: Run Delete Network Locks in the For Service Only screen.
Problem: More than one user is trying to use the Remote View tool to view the same test on
separate client PCs.
Solution: None! A stress test currently being performed can be monitored from one station
only at any one time.
Problem: On some occasions, the program does not calculate the target load.
Solution: To calculate the target load, the program needs to know the patient's gender, date of
birth, height, and weight.
In the following situations, the program is not able to calculate the target load:
Symptom: Application seems to lock up at Test End in Exercise Test. If the database is
damaged and the user initiates to store the test, there is a BTRV error message. But
this error message may display behind the acquisition window. The problem is that
the user will not see the message box and for him it seems that the system locked up.
Solution: Press one of the keys ALT, ESC, RETURN. This will bring the error message into
the foreground.
Symptom: Application seems to lock up while printing a Post Test Report on the thermal
printer or a taskbar popup displays “This document failed to print”.
Problem: The administrator password cannot be entered because the “Default user account
settings” in “Regional and Language Options” are enabled for non Latin languages
(e.g. Russian, Japanese, Chinese, Korean, etc.).
Solution: During start-up of CASE, in the password dialog, change the keyboard language
by simultaneously pressing <ALT> and <SHIFT> on the LEFT side of the
keyboard.
Then you can enter the administrator password “admin1,3,5,7”. You must change
your password, when prompted.
Problem: When you double click on the "CardioWeb" application on your desktop , an error
message "Invalid call blocked an attempt to exploit C:\...\W3WP.exe called from
module HASP_WINDOWS_4234327.DLL" is displayed.
Solution: This issue might be beacuse of the McAfee antivirus installed on your system.
Update the access protection rules of the antivirus on your system to allow the IIS
worker process executable W3WP.EXE to be run.
Unlike all other electrodes, the RL-electrode is not an input electrode, so no signal is
acquired and the electrode is always indicated as OK.
Solution: There are two main causes which may lead to the loss/corruption of data on SSD:
A read error, or
The destructive influence of some viruses.
Restore the system partition by using the last BACKUP file. If this does not help,
restore the complete system by using the last BACKUP file. This will cause the loss
of tests taken since the backup was done!
Problem: Small differences in the measurement values after transferred from MAC 1200 to
CASE.
Solution: The protocol used to transfer the data doesn't use the same amplitude unit for the
measurement values from MAC 1200. These values have to be converted when the
data is received in CASE/CardioSoft/CS and the results may not be exactly the same
values.
Solution: This Error Message is generated if the print job cannot be started.
Problem: The MUSE Web Server works, but an error message displays when the CASE
patient list is invoked.
Solution: A MUSE Web user account must be created under Windows and assigned to the
MUSE group. The user account must also be created in MUSE.
Solution: During exercise tests, the heart rate is calculated for 30-second tabular data. The
heart rate is averaged over 30 seconds at 5-second intervals and entered in the table.
The max. heart rate is derived from the tabular data. If the heart rate is sent to the
metabolic cart at 5-second intervals, the value displayed there may be higher than in
CASE.
Parameter Designation
Parameter Designation
Answer: The CASE application has the follow restrictions for data storage:
TWA Median 450 TWA memory full, no more TWA data can be
stored!
When the maximum number is reached, as identified in the table above, no more
objects can be stored. The data coming in after that limit is reached will be lost and
the message at the screen warns the user of this fact. However, in the CASE
CardioSoft/CS V6.7 application these limits are so high, that even a long Exercise
Test with many events will not reach them all. An example of reaching one of the
limits would be if the user did something extremely abnormal by clicking the
“PRINT 12 LEAD” button more than 99 times during a single Exercise Test, then
the memory for ECG strips would be full.
When the user has enabled to store the Full Disclosure ECG, even though the object
data cannot be saved beyond the limits, the user has the possibility to review or print
the complete Exercise Test and can analyze the ECG waveforms beyond that limit.
Problem: No picture
Solutions: The signal cable should be completely connected to the monitor computer.
Check for a Vacation Switch on the rear or side of the monitor. If it is present,
confirm it is in the ON position.
Front Power Switch on the monitor and the CASE power switch should be in
the ON position.
Check the monitor and your computer with respect to recommended settings.
Check the signal cable connector for bent or pushed-in pins.
Solutions: Unplug the power cord of the monitor from the AC outlet to turn off and reset
the monitor.
Check for a Vacation Switch on the rear or side of the monitor. If it is present,
confirm it is in the ON position.
Problem: The 22 inch monitor’s picture becomes too dark in dark environment.
Solution: Set the Auto EcoView mode to Off by entering the appropriate menu using the
leftmost EcoView button (under the leaf symbol) located on the front of the 22 inch
monitor.
Solution: Image persistence is when a residual or “ghost” image of a previous image remains
visible on the screen. Unlike CRT monitors, LCD monitors, image persistence is not
permanent, but constant images being displayed for a long period of time should be
avoided.
To alleviate image persistence, turn off the monitor for as long as the previous image
was displayed. For example, if an image was on the monitor for one hour and a
residual image remains, the monitor should be turned off for one hour to erase the
image.
We recommend that when your CASE is sitting idle for long periods of time, turn
OFF the system or the monitor.
Problem: LED on monitor is not lit (no green or amber color can be seen)
Solution: Power Switch should be in the ON position and power cord should be connected.
Solutions: Use the OSM Image Adjust controls to increase or decrease the Coarse total.
Check to make sure that a supported mode has been selected on the system
being used. (Please consult system manual to change graphics mode.)
Problem: No Video
Solutions: If no video is present on the screen, turn the Power button OFF and ON again.
Make certain the computer is not in a power-saving mode (touch the keyboard
or mouse).
Printer Debugging
Select Start > Settings > Printers and select your standard printer.
Solution: Select menu “File” the entry “Properties”. Select “General” Tab and press button
“Printing Preferences” and than press button “Advanced”. In the displayed list is in
the topic “Document Options” the Entry “Print Optimizations”. This “Print
Optimizations” must be disabled.
Solution: The laser printer may not have enough memory. A minimum of 8 MB is required.
Open Printer properties menu and change the printer scheduling to “Print
direct.”
Do the following entries to the LOC_WIN.INI, section.[CARDIO]
GRA_DontRemoveGrid=1
Solution: You can reduce the Transmission Retry Timeout to decrease the delay of the
displayed Paper out message, but it could cause a timeout message during the
printing of pages with noisy ECG curves.
Reduce the Transmission Retry Timeout in Windows Printer Settings (tab Port,
button Configure Port...).
Overview
This chapter describes the general procedures to perform the most common service
tasks. These include:
Software-Related Service
“System Backup and Restore”
“Recovering from a CASE BTRV Error”
“Remote Service”
Hardware-Related Service
“Controlling Electrostatic Discharge Damage”
“Replacing the Display Arm”
“Replacing the Monitor”
“Servicing the Keypad Assembly”
“Removing the Worksurface Assembly”
“Servicing the Power Distribution Assembly”
“Replacing the Solid State Drive (SSD) FRU”
“Servicing the I/O Assembly”
“Configuring the Acquisition Interface Board”.
NOTE
External SD card reader which initially shipped with the CASE unit is required
for this section.
Saving to SD Card
These procedures are used to back up the following information to an SD card:
System configurations
System settings
If possible, perform these procedures before replacing the Solid State Drive (SSD).
You can also use these procedures to save these settings so they can be copied from
one unit to another.
You can print the option codes by selecting the Print icon or you have to manually
write down all of your option codes.
1. From within the CASE application, select the System Configuration icon.
The System Configuration window opens.
2. Plug the external SD card reader into one of the USB ports on the right side of
the CASE unit, then insert a SD card into the SD card slot of the external SD
card reader.
3. On the System Configuration tab, select the Save icon.
A dialog window will open.
4. Select YES.
The message System Configuration successfully stored displays.
NOTE
Make sure to delete the three dots before entering the name.
System configurations
System settings
If available, perform these procedures after replacing the Solid State Drive (SSD).
You can also use these procedures to restore settings from one unit to another.
2. Plug the external SD card reader into one of the USB ports on the right side of
the CASE unit, then insert a SD card into the SD card slot of the external SD
card reader.
In the System Configuration window select the LOAD icon.
3. Select YES.
The message System configuration successfully loaded (software options not
loaded, please use option disk to load) displays.
4. Select OK.
5. Select the System Configuration icon.
The System Configuration window opens.
8. Select YES.
9. Enter all of the option codes that you manually wrote down prior to replacing
the Solid State Drive (SSD).
10. Select the Check button.
Verify that all option codes are dimmed. If not, this option code was not
correctly entered. Please correct the option code.
CAUTION
Option codes are serial number specific and should not be
duplicated.
11. After entering the option codes in the System Configuration window, select OK.
7. Select YES.
The message Overwrite test protocols and phases? displays.
8. Select YES.
9. Select the Custom Setup icon.
The Load Settings window opens.
13. If you need to restore additional custom setups, repeat step 5 through step 12 for
each setting BEFORE proceeding to the following step.
14. If you loaded more than one custom setup, select the custom setup under
Settings that you would like to use as your default setup.
15. Select Load.
The message Load setting and overwrite current settings? displays.
The Btrieve recovery process that follows is used to reorganize the corrupt database
in order to possibly regain access to tests already stored and to allow new tests to be
stored until the system can be repaired.
For further information on recovering from a CASE BTRV error, contact technical
support.
Procedure
Follow these steps:
1. Log on as Administrator.
a. Logon as:
User name: Admin
Password: admin1,3,5,7
Click OK.
Remote Service
Remote service is available using InSite. Refer to Chapter 4, “Installing InSite” for
more information.
CAUTION
The CASE contains components that are susceptible to
electrostatic discharge damage. Observe all static precautions
while performing service. Failure to observe these precautions
may result in failure of components.
The following guidelines help make a service CASE system more resistant to the
ESD damage:
Discharge any static charge you may have built up before handling
semiconductors or assemblies containing semiconductors.
A grounded, antistatic wristband (3M part number 2046 or equivalent) or heel
strap should be worn at all times while handling or repairing assemblies
containing semiconductors.
Use properly grounded soldering and test equipment.
Use a static-free work surface (3M part number 8210 or equivalent) while
handling or working on assemblies containing semiconductors.
Do NOT remove semiconductors or assemblies containing semiconductors
from antistatic containers (Velo-stat bags) until absolutely necessary.
Make sure power to an assembly is turned off before removing or inserting a
semiconductor.
Do NOT slide semiconductors or electrical/electronic assemblies across any
surface.
Do NOT touch semiconductor leads unless absolutely necessary.
Semiconductors and electrical/electronic assemblies should be stored only in
antistatic bags or boxes.
These guidelines may not guaranty a 100% static-free CASE system, but greatly
reduce the potential for failure of any electrical/electronic assemblies.
2-10A
Item Description
A Display Arm
D Display Adapter
8. Pull up the Display Arm and Adapter Assembly to access the black circular
bushing and then push the bushing out of the hole of the Adapter Mounting
Plate.
9. Pull the three cables out of the slot in the Adapter Mounting Plate so that the
Arm and Adapter Assembly can be removed and place on a flat surface.
10. Remove the Arm and Adapter Set Screw so the Display Adapter can be
separated from the Display Arm. Move the Display Arm further away from the
Display Adapter while carefully making sure the cables are pulled through the
center of the Display Adapter.
CAUTION
APPROPRIATE CABLE USAGE - The CASE system is
configured to operate ONLY with the Digital Panel (DP) video
cable that is within the CASE Display Arm Assembly.
DO NOT use any DVI or VGA cables that may have been
delivered in the monitor's accessory box to configure the CASE
system. Discard any cables you may find in the monitor's
accessory box.
1. Remove the keypad overlay to gain access to the keypad assembly locking tabs.
2. Working from right to left, release the locking tabs. Extract the keypad
assembly.
To gain access the writer and power supply assemblies, the CASE worksurface must
first be removed. Follow these steps to remove the CASE worksurface:
1. Disconnect the keyboard and mouse connectors from the I/O drawer and pull
the cable through the cable guide and remove.
CAUTION
Lift the front of the CASE system only far enough to reach
underneath and disconnect the keypad cables. Lifting the CASE
system any higher could damage the connectors on the keypad
board.
4. While holding the front of the worksurface up, reach underneath it and carefully
disconnect the two keypad cables located near the left of the keypad assembly.
Log off the CASE application, shut down the system, turn the power switch to OFF,
and unplug the unit from the power outlet.
3. Disconnect the cables that are connected to the power supply assembly. Make
note of the cable locations before disconnecting them.
4. Loosen the two lock nuts holding the power supply in place.
6. If you need to exchange the Power Distribution Unit (PDU) circuit board,
unscrew the eight screws shown in the picture below and exchange the board.
7. If you need to exchange the 24 volt power supply, unscrew the power supply
screws as shown below and exchange the power supply.
4. Remove five mounting screws from the I/O drawer rear panel. Note where any
non-standard cables are currently connected for reassembly later on.
6. Carefully slide out the I/O drawer to gain access to the I/O boards and mother
board assembly.
CAUTION
When moving the I/O drawer, keep it to the left or press in on
circuit board to prevent the side of the chassis from damaging it.
8. Remove the connectors at the back of the Solid State Drive assembly.
9. Remove the assembly.
10. Remove the two screws holding the Solid State Drive to the bracket.
13. After installing the replacement SSD FRU, see Chapter 8, “System Rebuild”
and execute the instructions.
1. Use a screw driver to loosen and open the I/O access drawer located on the back
of the unit.
Cable clamps – used for external
devices, e.g., the CAM-14.
2. Remove five mounting screws from the I/O drawer rear panel. Note where any
non-standard cables are currently connected for reassembly later on.
4. Carefully slide out the I/O drawer to gain access to the I/O boards and mother
board assembly.
CAUTION
When moving the I/O drawer, keep it to the left or press in on
circuit board to prevent the side of the chassis from damaging it.
5. Hang the drive bay to position I/F cables appropriately when reinserting the I/O
drawer assembly back into the CASE chassis.
6. Reverse the procedure to reassemble.
7. Reconnect the CASE unit to AC power.
The following is a picture of the interior of the I/O assembly drawer with mother
board fully mounted.
NOTE
For the purpose of this picture, cables are unmounted and moved out of the way.
1. Make sure that you are at the initial CASE application screen.
2. Log on as a service user with the following procedure:
a. Right click on the Windows Start button.
b. Click the arrow next to Shut down or sign off then select Shut Down.
c. Once the system shuts down, click the Power button on the side panel to
restart.
d. Double click the CASE icon when it displays.
e. Select GEService and enter service for the password.
3. On the Windows desktop, click the Windows Start button then select Control
Panel >> System Security >> System
4.
5. s.
6. From the window that opens, verify that the Computer Name is the unit serial
number (This Computer).
NOTE
You can rename the computer by selecting Change settings (to the far right)
and then Change. Enter the new computer name in the Computer Name
field.
19. Check the Target System is selected as CASE and verify a generated serial
number is displayed.
20. Click the Programming button to ensure the Generated Serial Number is
programmed. (This will not be the same as 13-digit SN.)
21. Click Close after the Programming Status displays PROGRAMMING OK.
22. Close the CASE CAM-USB ACQ Board Service Tool window, after the
Acquisition System Serial Number changes to the CASE label Serial Number.
23. Shutdown the computer as normal.
Verification
After configuring the acquisition interface board, verify its success by running a
burn-in test with the following instructions:
Overview
In the event of a catastrophic system failure, it will be necessary to replace the Solid
State Drive (SSD) with the SSD FRU (PN 2109926-002) which contains the CASE
Windows 10 custom image pre-installed on the SSD and contains a DVD for the
CASE ADVANTECH v6.73 application.
For installation and setup instructions for the new SSD FRU, please refer to the
section below.
CAUTION
LOSS OF CONFIGURATIONS—This procedure will scrub all custom
configurations and modifications to standard configurations and settings. If
lost, these custom and modified configurations and settings CANNOT be
retrieved electronically.
Verify that the following items are in the FRU kit you received. If any item is
missing, contact technical support.
If you are upgrading the CASE system which has SNP product code, verify that the
following items are in the FRU kit you received.
Tools Required
External DVD drive (option available through sales)
Process Overview
The repair process consists of the following steps:
Installing InSite
Refer to “Installing InSite” on page 4-3.
4. Remove the four lock nuts from the drive assembly front panel.
6. Hang the drive assembly by the hooks on its side panels to access connectors
and individual drive mounting screws. The drive is now accessible for
replacement.
7. Disconnect the SATA power and the SATA data cables from the hard drive.
CAUTION
ELECTRIC SHOCK - Improper placement between two drive
might cause electric shock.
Place the ESD bag of the HD between the two drives to avoid
shorting any open contacts.
8. Plug the SATA power and SATA data cable connectors into the new HDD.
14. Click and drag the highlighted contents of the D: drive to the F: drive of the
other File Explorer window.
15. Verify that all the files and folders from the D: drive copied to the F: drive.
16. Close both windows and shut down the CASE system.
17. Remove power from the CASE system and unplug the power cord.
18. Disconnect the cables from the new hard drive.
19. Remove the two screws on each side.
Installing InSite
Refer to “Installing InSite” on page 4-3.
Procedure
Use the scripts in this chapter when debriefing a Service Dispatch.
The checkout procedures on the following pages apply to CASE Systems with
product code STE. Checkout procedures are separated into two categories.
i. Clean if necessary.
3. Power on
b. Log on as administrator
(Name: Admin PW: admin1,3,5,7).
4. Check software
i. Press Pre-test.
Pass/Fail
Expected Result: Test starts.
5. Check hardware
a. Screen
Start CASE application if not already started. Pass/Fail
Expected Result: Screen shows CASE application.
b. Desk light
Switch desk light on and off on the keypad. Pass/Fail
Expected Result: Light turns on and off.
c. Speaker
Boot CASE device. Pass/Fail
Expected Result: Beeps during boot up.
c. Double-click Services.
a. Open Explorer.
Leakage Current
Step Condition UUT - ON Result
Limits in µA
C Touch Current in µA
Leakage Current
Step Condition UUT - ON Result
Limits in µA
NC = Normal Condition
SFC = Single Fault Condition
NA = Not Applicable
UUT = Unit Under Test
NOTE
Verify part numbers before ordering service parts (field replaceable units). See
the tech memo series for this product for changes or additions to this list.
Cables, Interface
Select External FRUs
Power Cords
Software
FRU Diagrams
Cables, Interface
The following table identifies the FRUs available for connecting the CASE system
to the exercise products.
External FRUs
The following table identifies the external FRUs for CASE system.
CASE System
Display
Power Cords
The following table identifies the power cords available for the CASE system.
FRU Diagrams
The diagrams on the following pages identify the various assemblies, components,
and FRUs of CASE systems. Key FRUs are identified by numbered callouts. The
table following each diagram identifies the Item Number and description of each
callout in the drawing.
* Note: After replacing this writer, one ferrite from the old writer, if not
already present, must be moved to the new replacement writer and
positioned in the same location.
Drawing 2:
CASE System Back
86 2104239-001 FRU KIT LABELS GE LOGO DUAL USB AND REAR I/O
Includes item 86: 2055267-010 LABEL ADH IO PANEL
CASE W. ADVANTECH BD
And additionally:
2007786-022 NAMEPLATE 40MM GE LOGO GRANITE
GRAY
2103011-001 3M USB LABEL FOR CHASSIS LEG
* Note: These three parts are already included when ordering 2091039-001
motherboard assembly. Parts are listed here if it is desired to order
separately without motherboard.
NOTE
Display Port (DP) video cable, display power cord and work surface light with
cable are part of display arm.
Loc
Num Item Number Item Description
87 2104239-001 FRU KIT LABELS GE LOGO DUAL USB AND REAR I/OFRU KIT LABELS GE LOGO DUAL USB AND
REAR I/O
Includes item 87: 2007786-022 NAMEPLATE 40MM GE LOGO GRANITE GRAY
And additionally:
2055267-010 LABEL ADH IO PANEL CASE W. ADVANTECH BD
2103011-001 3M USB LABEL FOR CHASSIS LEG
88 Various see ASSY STRESS KEYPAD 22 KEY CASE SERIES ROHS COMPLIANT
earlier table
titled
“Worksurface
Pad,
Keypads, and
Keypad
Labels”
119 2098963-002 FRU CABLE USB 10 PIN 2.54MM PITCH TO 2X USB TYPE A
50CM
153 2095970-002 FRU HARN CASE USB 2X INTERNAL WITH TWO FERRITES
** Note: After replacing this harness one ferrite from the old harness, if not already
present, must be moved to the new replacement harness and positioned in the same
location.
*** Note: Option only on CASE with KISS which is Product Number 2012492-032.
2 2109926-002 FRU SSD CASE WINDOWS 10 IoT 64bit IMG AND DVD CASE
ADVANTECH V6.7
39 2098467-002 FRU HARNESS SOFTSTART WITH POWER SYMBOL
SWITCH
82 2095972-002 FRU MECH HARN PC SERIAL FEMALE WITH TWO
FERRITES
153 2095970-002 FRU HARN CASE USB 2X INTERNAL WITH TWO FERRITES
Documentation FRUs
The following table identifies the documentation available for CASE v6.7 and their
part numbers.
If the cables do not have the ferrites installed as pictured in this section, please
continue.
A ferrite unlocking key should be ordered, if not already part of the FE tool kit, GE
P/N 2055233-001 is recommended.
To open and remove the ferrite use unlocking key, GE P/N 2055233-001.
To open and remove the ferrite use unlocking key, GE P/N 2055233-001.
CASE System
Changes or modification to this system not expressly approved by GE Healthcare
could cause EMC issues with this or other equipment. This system is designed and
tested to comply with applicable regulation regarding EMC and needs to be installed
and put into service according to the EMC information stated as follows.
WARNINGS
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
RF emissions
Class A
CISPR11 The equipment is suitable for use in all
establishments other than domestic
Harmonic Emissions
Class A establishments and those directly connected to
EN 61000-3-2
the public low-voltage power supply network
Voltage fluctuations/ that supplies buildings used for domestic
Flicker emissions Complies purposes.
EN 61000-3-3
Electromagnetic Environment -
Immunity Test EN 60601 Test Level Compliance Level
Guidance
Electrical Fast ± 2 kV for power supply ± 2 kV for power supply Mains power should be that of a typical
Transient/ lines lines commercial or hospital environment.
burst (EFT) ±1 kV for input/output ±1 kV for input/output
IEC 61000-4-4 lines lines
Surge +/- 1 kV line(s) to line(s) +/- 1 kV line(s) to line(s) Mains power should be that of a typical
IEC 61000-4-5 +/- 2 kV line(s) to earth +/- 2 kV line(s) to earth commercial or hospital environment.
Voltage dips, short < 5% Ut (>95% dip in Ut) < 5% Ut (>95% dip in Ut) Mains power should be that of a typical
interruptions, and for 0.5 cycles for 0.5 cycles commercial or hospital environment.
voltage variations on < 40% Ut (>60% dip in < 40% Ut (>60% dip in Ut is the AC mains voltage prior to
power supply input Ut) for 5 cycles Ut) for 5 cycles application of the test level.
lines < 70% Ut (>30% dip in < 70% Ut (>30% dip in If the user requires continued system
IEC 61000-4-11 Ut) for 25 cycles Ut) for 25 cycles operation during power mains
< 5% Ut (>95% dip in Ut) < 5% Ut (>95% dip in Ut) interruptions, it is recommended that the
for 5 s for 5 s system is powered from an applicably rated
uninterruptible power supply or a battery.
Electromagnetic Environment -
Immunity Test EN 60601 Test Level Compliance Level
Guidance
NOTE:
Do not use portable or mobile RF communications equipment closer to any part of the system, including the
cables, than the recommended separation distance calculated for the equation applicable to the frequency of
the transmitter.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radio, AM and FM radio broadcast, and TV broadcast cannot, theoretically, be predicted with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, consider conducting an
electromagnetic site survey. If the measured field strength in the location the system is used exceeds the
applicable RF compliance level listed in this table, observe the system to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the system.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by the reflection
from structures, objects, and people.
The system is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the system can help
prevent electromagnetic interference by maintaining the following recommended
minimum distance between portable and mobile RF communications equipment
(transmitters) and the system, according to the maximum output power of the
communications equipment.
1. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.
2. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
For transmitters rated at a maximum output power not listed above, estimate the recommended separation distance d in
meters (m) using the equitation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
WARNING
The use of accessories, transducers and cables other than those
specified may result in increased emissions or decreased
immunity performance of the equipment or system.
The table below lists cables, transducers, and other applicable accessories with
which GE Healthcare claims EMC compliance.
NOTE
Any supplied accessories that do not affect EMC compliance are not included.
Filename Description
erg_prot_v5an.dat Data file that stores the settings for the stress test
driver (e.g., ergometer, treadmill).
NOTE
Up to 30 different setups can be stored (see Custom Setup in System
Configuration), making them available to all users of a network. The file
names are SETUP0.INI to SETUP29.INI. Stress test drivers are stored in
files PA5V0.DAT to PA5V29.DAT.
Terminology
IHE: Integrating the Healthcare Enterprise
DICOM: Digital Imaging and Communication in Medicine
SCP: DICOM Service Class Provider
SCU: DICOM Service Class User
PPS: Performed Procedure Step
DSS: Department System Scheduler
MWL: Modality Work List
AE title: Application entity title
General
NOTE
The DICOM Conformance Statement (P/N 2040396-087) for this product
provides all necessary information to assess the interoperability with other
systems.
Setup
To complete a standard IHE configuration the following steps are needed:
NOTE
All data entered in this section is automatically stored in the local file
<Application Data Folder>\CA_SYS.INI (section [DICOM]).
Enter the AE title for this CASE system in the SCU AE title field.
NOTE
All data entered in this section is automatically stored in the server-based file
DICOM.INI, located where the database is installed.
Enter the network configuration for DSS/Order Filler and PPS Manager (may
be identical to the Order Filler) using the button "Network Configuration" in the
section "Worklist".
AE title, host name and port number of the remote host are required to complete
this configuration.
Click both "Verify remote host" buttons to check the configuration (verification
service, using ECHO Messages).
Enter the network configuration for the Image Archive and Image Manager
(may be identical to the Image Archive) using the button "Network
Configuration" in the section "Image Archiving".
AE title, host name and port number of the remote host are required to complete
this configuration.
Click both "Verify remote host" buttons to check the configuration (verification
service, using ECHO Messages).
The SCU port is used to receive storage commitment responses. Click the
button "Verify SCU port" to check the configuration (verification service, using
ECHO messages).
Configure the extended character set for the DICOM Interface depending on the
language used: add the item CHARACTER_SET to “Special DICOM setup
items without UI”).
The protocol code mapping is responsible for starting the correct procedures out of
the DICOM work list.
Therefore it is crucial that the received and the used protocol scheme and codes
match.
To verify this, create at least one order per desired procedure type in the DSS/ Order
Filler system.
Select each of the orders and verify that the "Start procedure" button is always
active.
If this button is disabled for one of the orders, the corresponding protocol code map-
ping has to be modified.
Either the DSS/ Order Filler adapts its used protocol scheme/codes to this pre-
defined scheme/codes or the protocol code mapping has to be modified within this
system (See “Protocol Code Mapping” on page C-6).
0: ISO_IR 6 (ASCII)
1: ISO_IR 100 (ISO 8859-1: Latin 1: Western European languages)
2: ISO_IR 101 (ISO 8859-2: Latin 2: Central/Eastern European languages)
3: ISO_IR 144 (ISO 8859-5: Cyrillic: Russian)
4: ISO_IR 148 (ISO 8859-9: Latin 5: Turkish)
Use the selected ECG Report formats of the Exercise test setup for ECG
Waveform Images (only valid if the "General ECG Waveform SOP Class" is
disabled and 12-leads are selected in the Exercise test setup):
IA_USE_SELECTED_ECG_REPORT_FORMATS=1 (Default: 0, use always
1x10s format)
ECHO Retry Number for sending ECHO_RQ during “Verify remote host”
action.
ECHO_RETRY=5 (Default: 3; Range 1 to 20000)
There is a predefined DICOM protocol scheme ("SRT") for Exercise test that CASE
uses as default data.
However, the DSS/Order filler may use other schemes and protocol codes.
In this case you can add or adjust schemes and protocol codes.
CASE has also implemented it's own scheme per procedure type that could be used
by the DSS/Order filler.
The DICOM interface provides the Protocol Code Mapping in both directions:
That sequence is used to perform the Protocol Code mapping for scheduled tests
and must contain the Protocol Code in its Code Value (DICOM tag 0008,0100)
and the Scheme in its Scheme Designator (DICOM tag 0008,0102).
Using the DICOM worklist, the received (scheduled) DICOM Protocol Scheme/
Code is mapped to the
CASE Procedure Type (e.g. Exercise test, Resting ECG...) and to the
CASE Protocol Code I and II for detailed information
(e.g. for Exercise test: Use of treadmill or ergometer. Use of protocol, e.g.
BRUCE, WHO...).
If CASE cannot find a corresponding procedure type, the "Start examination" button
in the DICOM worklist is disabled.
The DICOM Performed Protocol Code may be different than the DICOM Scheduled
Protocol Code (but identical scheme), because the user could have changed it.
Example:
The user starts a scheduled order, that initiates an Exercise test with a Treadmill and
the BRUCE protocol.
Before going to the pretest phase, the user switches to the NAUGHTON protocol.
So the DICOM Performed Protocol Code reflects what was executed in reality.
Changing or appending additional schemes for the mapping can only be done by
manually changing the data in this file using an editor (e.g. Notepad).
Changes always affect all CASE systems and clients in the CASE network.
The DICOM Protocol Scheme/ Code applies to the DICOM tag 0040, 0008 (Sched-
uled Protocol Code Sequence) and 0040, 0260 (Performed Protocol Code
Sequence).
1. Determine the procedure type for which you want to add a new scheme and use
the corresponding test type identifier, e.g.
Procedure type is Exercise test --- corresponding procedure type identifier is
"STRESS".
2. Append the new scheme for unscheduled tests in section
[PROTOCOL_SCHEMES], using the item with the corresponding procedure
type identifier, e.g.
New scheme is "ABCD":
STRESS_PROT_SCHEME_UNSCHED=SRT,99IHE,CSO_STRESS,ABCD
3. Use the new scheme as default protocol scheme for the determined procedure
type in section [PROTOCOL_SCHEMES], if the performed protocol code
could not be mapped to a DICOM protocol code, e.g.
STRESS_PROT_SCHEME_USED_IF_UNKNOWN=ABCD
4. Determine DICOM default protocol codes for this scheme, if the performed
protocol code could not be mapped to a DICOM protocol code, e.g.
or
[ABCD]
ABCD_UNKNOWN_0_PROT_CODE=ABCD_ERG_UNKNOWN
ABCD_UNKNOWN_1_PROT_CODE=ABCD_TRM_UNKNOWN
ABCD_NO=2
ABCD_NO_0=STRESS,ABCD_WHO,0,WHO
ABCD_NO_1=STRESS,ABCD_BRUCE,1,BRUCE
SPI_PROT_SCHEME_UNSCHED=CSO_SPI
; Default performed protocol scheme used, if the performed protocol code could not
be
; mapped to a DICOM protocol code
STRESS_PROT_SCHEME_USED_IF_UNKNOWN=CSO_STRESS
ESPIRO_PROT_SCHEME_USED_IF_UNKNOWN=CSO_ESPIRO
ECG_PROT_SCHEME_USED_IF_UNKNOWN=CSO_ECG
ABP_PROT_SCHEME_USED_IF_UNKNOWN=CSO_ABP
SPI_PROT_SCHEME_USED_IF_UNKNOWN=CSO_SPI
[SRT]
; DICOM protocol scheme SRT for procedure type Exercise test
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
SRT_NO=9
SRT_NO_0=STRESS,P2-7131C,1,BALKE
SRT_NO_1=STRESS,P2-7131A,1,BRUCE
SRT_NO_2=STRESS,P2-7131D,1,ELLESTAD
SRT_NO_3=STRESS,P2-7131B,1,MODBRUCE
SRT_NO_4=STRESS,P2-713A1,1,MODNAUGHTON
SRT_NO_5=STRESS,P2-713A0,1,NAUGHTON
SRT_NO_6=STRESS,P2-7131F,1,PEPPER
SRT_NO_7=STRESS,P2-7131E,1,RAMP
SRT_NO_8=STRESS,P2-31102,0,WHO
[99IHE]
; DICOM protocol scheme 99IHE for procedure type Exercise test
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
99IHE_NO=4
99IHE_NO_0=STRESS,PHARMSTRESS,1,MODBRUCE
99IHE_NO_1=STRESS,PERSANTINE,1,PERSANTINE
99IHE_NO_2=STRESS,ADENOSINE,1,ADENOSINE
99IHE_NO_3=STRESS,DOBUTAMINE,1,DOBUTAMINE
[CSO_STRESS]
; User defined protocol scheme for procedure type Exercise test
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
CSO_STRESS_NO=21
CSO_STRESS_NO_0=STRESS,CSO_WHO,0,WHO
CSO_STRESS_NO_1=STRESS,CSO_WHO50,0,WHO50
CSO_STRESS_NO_2=STRESS,CSO_WHO75,0,WHO75
CSO_STRESS_NO_3=STRESS,CSO_HOLLMANN,0,HOLLMANN
CSO_STRESS_NO_4=STRESS,CSO_BAL,0,BAL
CSO_STRESS_NO_5=STRESS,CSO_STD.FRANCE,0,STD.FRANCE
CSO_STRESS_NO_6=STRESS,CSO_MODWHO,0,MODWHO
CSO_STRESS_NO_7=STRESS,CSO_CONCONI,0,CONCONI
CSO_STRESS_NO_8=STRESS,CSO_BRUCE,1,BRUCE
CSO_STRESS_NO_9=STRESS,CSO_MODBRUCE,1,MODBRUCE
CSO_STRESS_NO_10=STRESS,CSO_NAUGHTON,1,NAUGHTON
CSO_STRESS_NO_11=STRESS,CSO_ELLESTAD,1,ELLESTAD
CSO_STRESS_NO_12=STRESS,CSO_MODBALKE,1,MODBALKE
CSO_STRESS_NO_13=STRESS,CSO_USAFSAM,1,USAFSAM
CSO_STRESS_NO_14=STRESS,CSO_SLOWUSAFSAM,1,SLOWUSAFSAM
CSO_STRESS_NO_15=STRESS,CSO_CORNELL,1,CORNELL
CSO_STRESS_NO_16=STRESS,CSO_BALKE,1,BALKE
CSO_STRESS_NO_17=STRESS,CSO_MODBALKEWARE,1,MODBALKE-
WARE
CSO_STRESS_NO_18=STRESS,CSO_ADENOSINE,1,ADENOSINE
CSO_STRESS_NO_19=STRESS,CSO_DOBUTAMINE,1,DOBUTAMINE
CSO_STRESS_NO_20=STRESS,CSO_PERSANTINE,1,PERSANTINE
[CSO_ESPIRO]
; User defined protocol scheme for procedure type Ergospirometry test
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
CSO_ESPIRO_NO=21
CSO_ESPIRO_NO_0=ESPIRO,CSO_E_WHO,0,WHO
CSO_ESPIRO_NO_1=ESPIRO,CSO_E_WHO50,0,WHO50
CSO_ESPIRO_NO_2=ESPIRO,CSO_E_WHO75,0,WHO75
CSO_ESPIRO_NO_3=ESPIRO,CSO_E_HOLLMANN,0,HOLLMANN
CSO_ESPIRO_NO_4=ESPIRO,CSO_E_BAL,0,BAL
CSO_ESPIRO_NO_5=ESPIRO,CSO_E_STD.FRANCE,0,STD.FRANCE
CSO_ESPIRO_NO_6=ESPIRO,CSO_E_MODWHO,0,MODWHO
CSO_ESPIRO_NO_7=ESPIRO,CSO_E_CONCONI,0,CONCONI
CSO_ESPIRO_NO_8=ESPIRO,CSO_E_BRUCE,1,BRUCE
CSO_ESPIRO_NO_9=ESPIRO,CSO_E_MODBRUCE,1,MODBRUCE
CSO_ESPIRO_NO_10=ESPIRO,CSO_E_NAUGHTON,1,NAUGHTON
CSO_ESPIRO_NO_11=ESPIRO,CSO_E_ELLESTAD,1,ELLESTAD
CSO_ESPIRO_NO_12=ESPIRO,CSO_E_MODBALKE,1,MODBALKE
CSO_ESPIRO_NO_13=ESPIRO,CSO_E_USAFSAM,1,USAFSAM
CSO_ESPIRO_NO_14=ESPIRO,CSO_E_SLOWUSAFSAM,1,SLOWUSAFSAM
CSO_ESPIRO_NO_15=ESPIRO,CSO_E_CORNELL,1,CORNELL
CSO_ESPIRO_NO_16=ESPIRO,CSO_E_BALKE,1,BALKE
CSO_ESPIRO_NO_17=ESPIRO,CSO_E_MODBALKEWARE,1,MODBALKE-
WARE
CSO_ESPIRO_NO_18=ESPIRO,CSO_E_ADENOSINE,1,ADENOSINE
CSO_ESPIRO_NO_19=ESPIRO,CSO_E_DOBUTAMINE,1,DOBUTAMINE
CSO_ESPIRO_NO_20=ESPIRO,CSO_E_PERSANTINE,1,PERSANTINE
[CSO_ECG]
; User defined protocol scheme for procedure type Resting ECG
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
CSO_ECG_NO=1
CSO_ECG_NO_0=ECG,CSO_ECG,0,Resting ECG
[CSO_ABP]
; User defined protocol scheme procedure procedure type Ambulatory Blood Pres-
sure
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
CSO_ABP_NO=1
CSO_ABP_NO_0=ABP,CSO_ABP,0,ABP-Test
[CSO_SPI]
; User defined protocol scheme for procedure type Spirometry
; list of defined protocol codes; first item must specify the number of protocol codes
in this scheme.
CSO_SPI_NO=1
CSO_SPI_NO_0=SPI,CSO_SPI,0,Spirometry-Test
;The user can add other protocol schemes or change existing protocol schemes as
needed, but the required syntax must be followed.
Description of dialogs
DICOM tab
Figure 1
e. Default procedure type for starting the test if the scheduled protocol code
did not match:
Exercise test
Resting ECG
Spirometry
ABP
Ergospirometry
Not used
3. Service support
a. Check the DICOM-Interface according to the DICOM configuration setup:
The following checks are implemented:
DICOM library (Merge3-COM) initialization
"Open Association" and "Close Association" action to the DSS/Order
filler is performed, if the work list is enabled; the Service List
[DMWL_Service_List] in file MERGECOM.APP is used for
association negotiation.
"Open Association" and "Close Association" action to the PPS
Manager is performed, if the work list is enabled; the Service List
[MPPS_Service_List] in file MERGECOM.APP is used for association
negotiation.
"Open Association" and "Close Association" action to the Image
Archive is performed, if one of the TCP/IP functions for image
archiving is enabled and the ECG Waveform object is enabled and the
"General ECG Waveform Object SOP Class" is disabled; the Service
List [IA_Service_List_12LECG] in file MERGECOM.APP is used for
association negotiation.
"Open Association" and "Close Association" action to the Image
Archive is performed, if one of the TCP/IP functions for image
archiving is enabled and the ECG Waveform object is enabled and the
"General ECG Waveform Object SOP Class" is enabled; the Service
List [IA_Service_List_GenECG] in file MERGECOM.APP is used for
association negotiation.
"Open Association" and "Close Association" action to the Image
Archive is performed, if one of the TCP/IP functions for image
archiving is enabled and the PDF Report object is enabled; the Service
List [IA_Service_List_EncPDF] in file MERGECOM.APP is used for
association negotiation.
"Open Association" and "Close Association" action to the Image
Manager is performed, if one of the TCP/IP functions for image
archiving is enabled and the "Storage Commitment Request" is
enabled; the Service List [IM_Service_List] in file MERGECOM.APP
is used for association negotiation.
The default event logging level is level 0, which is used for errors.
Adds more informations (that are not necessarily errors) to the log files
Lists the DICOM messages of one transaction as files in the DICOM
LOG folder
To change the event logging level, use the Notepad to edit the file
<Application Data Folder>\CA_SYS.INI, section [DICOM], entry
DICOM_LOG_LEVEL=1.
e. Use the "General ECG Waveform SOP Class" for archiving ECG
Waveforms (instead of "12-Lead ECG Waveform SOP Class"):
There may be two scenarios to enable this item:
The Image Archive cannot process/view the DICOM 12-lead ECG
Waveform object.
The Exercise test is performed with 3, 6 or 15 leads.
Perform the automatic archiving via TCP/IP (default; uses the Image
Archive SCP configuration and the Image Manager SCP configuration,
see 4g)
Perform the automatic archiving via File Export. The destination folder
of 4b is used. The file name is used as configured in the Export setup
dialog (see User Manual, System Configuration)
Disable the automatic archiving
Perform the manual archiving via TCP/IP (uses the Image Archive SCP
configuration and the Image Manager SCP configuration, see 4g)
Perform the manual archiving via File Export (uses the destination
folder of 4b and the filename as configured in the Export setup dialog,
see User Manual, System Configuration)
Disable the manual archiving (default)
Figure 2
The following data are stored server-based (unless local storage was explicitly set in
the configuration) accessible for all CASE systems and Clients (file DICOM.INI in
Database folder):
c
f
d
g
j
Figure 3
The following data are stored server-based (unless local storage was explicitly set in
the configuration) accessible for all CASE systems and Clients (file DICOM.INI in
Database folder):
Worklist customization
Figure 4
The worklist customization dialog allows to customize the following fields in the
DICOM worklist:
a. Check box: allows you to enable or disable the corresponding field in the
DICOM worklist.
b. Drop-down list: available DICOM data object to select
c. Text box for entry of the DICOM data object prompt
d. Use default text button to restore the default text prompt for the selected
DICOM data object
3. Worklist Default button to reset all fields of the DICOM worklist to their default
values.
Metabolic Interface
The system can communicate with the data acquisition device over the CASE
system's serial port or it can record test data to a shared file on the network.
Communication Commands
To enable the serial port communication, select the “Metabolic Cart” checkbox in
System Configuration -> Devices and select the appropriate COM port.
The stress test module can process the following commands from the serial port:
'E' echo on
'T' start treadmill
'O' stop treadmill
'H' stage hold
'R' request measurements
'W' start pretest
'S' start exercise
'M' start recovery
parity no
data bits 8
stop bits 1
baud rate 1200
handshake no
You can modify the settings for the serial port by modifying the following values in
the [CAS_RECORD] section of the file
<Application Data Folder>\CARDIO.INI.
Field Values
SpiroParity 0 = no parity
1 = even parity
2 = odd parity
SpiroHandShake 0 = no handshake
1 = XonXoff
2 = hardware handshake
3 = hardware handshake and interpretation of CR and NULL
For example, to configure the serial port for even parity, 7 data bits, 2 stop bits, 9600
baud, and XonXoff, cardio.ini would look like this:
[CAS_RECORD]
SpiroParity = 1
SpiroDataBits = 0
SpiroStopBits = 1
SpiroBaudrate = 5
SpiroHandShake = 1
[AAAA_BBBB_C_DDDD_E_FFFF_GGGG_HHHH_IIII_JJJJ_KKKK_LLLL_M-
MMM_]NNPP
where a space is shown as an '_'.
KKKK - ST Level for most significant ECG lead in [0.01 mV] = [0.1 mm] (-999 if
unavailable)
LLLL - ST Slope for most significant ECG lead in [0.1 mV/s] (-999 if unavailable)
The most significant ECG lead is the lead with the largest ST-depression.
The leads aVR, aVL and V1 are excluded.
If no ST segment depression is found, lead V5 is defined as the most
significant lead.
MMMM - ST Integral for most significant ECG lead in [µVs] (-999 if unavailable)
The checksum is the sum of the numeric values of all characters from "[" to "]"
inclusive. (Example: Numeric value of character 'A' is 65).
Field Values
For example, if you want the shared file ergospir.dat to be written to a directory
called spirapp\data located on the c:\ drive, the entry in LOC_WIN.INI would look
like this:
[CAS_RECORD]
SpiroOptionWriteToFile = 1
SpiroDirectory = C:\SPIROAPP\DATA
KKKK - ST Level for most significant ECG lead in [0.01 mV] = [0.1 mm] (-999 if
unavailable)
LLLL - ST Slope for most significant ECG lead in [0.1 mV/s] (-999 if unavailable)
The most significant ECG lead is the lead with the largest ST-depression.
The leads aVR, aVL and V1 are excluded.
If no ST segment depression is found, lead V5 is defined as the most
significant lead.
MMMM - ST Integral for most significant ECG lead in [µVs] (always -999 since
not used)
The checksum is the sum of the numeric values of all characters from "[" to "]"
inclusive. (Example: Numeric value of character 'A' is 65).
Example –
Meaning
General Settings
For serial output and for shared file output:
The availability of the blood pressure values can be controlled by the following
entries in the file <Application Data Folder>\LOC_WIN.INI:
ExerciseBpGreyoutTime:?
Range: 5..360, default = 60, unit = seconds
The BP values will be greyed out after this period of time
SpiroNegateGreyedBp=?
Range: 0..2, default 1
BP values older than defined with entry
'ExerciseBpGreyoutTime' will be converted as
follows:
0: old BP value will be positive
1: old BP value will be invalid (-999)
2. old BP value will be negative
CAUTIONS
Reload the application only after you configure the CASE
network (“Configure the Network” on page 3-32) and save
the existing data (“Copying Setup and Configuration for
Existing CASE System” on page 3-45)
Please choose to use either a UNC path or a mapped drive to access the CASE
server.
NOTE
If a network error message is shown, please be sure that the server firewall does
not block the incoming request.
NOTE
Make sure the Reconnect at login box is checked.
Also, Connect as should be a Network User.
If the Network User is a domain user, then enter [Domain Name]\[Network
User].
If a network error message is shown, please be sure that the server firewall does
not block the incoming request.
CAUTION
Do not start the application on any CASE system until all CASE
systems have been reconfigured. Attempting to start the CASE
application before all CASE systems have been configured may
corrupt the installation on all CASE systems.
The described procedure uses the SEND and RECEIVE functionality of the CASE
application to merge data from a CASE system to a network-based database.
If there are more than 1000 examinations to transfer per CASE system, it is
recommended to use Date filters in the test list or to merge by using the Database
Repair Tool V6.7 (to be ordered by trained field service engineers; P/N 2063485-
001 SERVICE KIT DATABASE REPAIR TOOL V6.7).
NOTE
The following description assumes that you have the rights to create a shared
temporary folder on a network server. If you are not permitted to do this, create
a temporary shared folder on the local CASE system and use this folder for
sending/receiving the examinations to the network server.
NOTE
If the amount of local disk space is low, the Send/Receive process may be
done in several steps.
Instead of using the mapped drive letters, UNC paths (paths according to
the Universal Naming Convention) can be used as well.
Condition
The CASE application version on the network server must be equal to or higher than
the CASE system used.
Network Server
On the server, create a temporary folder (e.g., F:\CASE_EXA)
(F: = drive letter where the temporary folder is stored)
Share this folder
CASE System
Map a local drive letter to the temporary network folder (e.g., X:\CASE_EXA)
(X: = network drive letter where the temporary folder is stored)
Open the Test List within the CASE application
Send all tests to the temporary folder (e.g., X:\CASE_EXA)
Manual Mode
Network Server
From within the CASE application, open the window Receive Data from ECG
device via System Configuration >> General tab.
Select the mode Data from storage medium.
Select the folder where the tests from the CASE system are stored
F:\CASE_EXA
(F: = drive letter where the temporary folder is stored).
Transfer the tests.
Automatic Mode
Network Server
Verify that the database folder is setup correctly to the one you want to merge
the data to.
Close the CASE application.
Select the folder where the tests from the CASE system are stored
F:\CASE_EXA
(F: = drive letter where the temporary folder is stored)
Copy all .exa files to D:\CASE\RESTORE.
Then start CASE application. All files will be moved automatically to the
selected database.
NOTE
Make sure the Reconnect at login box is checked.
Also, Connect as should be a Network User.
If the Network User is a domain user, then enter [Domain Name]\[Network
User].
If a network error message is shown, please be sure that the server firewall does
not block the incoming request.
NOTE
If Expert Mode was enabled at the beginning of the installation procedure, the
Application Data directory can be selected additionally. This directory contains
CASE-relates files, such as configuration and log files. By default, the
Application Data directory is set to D:\CASE.
CAUTION
Do not start the application on any workstation until all
workstations have been reconfigured. Attempting to start the
CASE application before all workstations have been configured
may corrupt the installation on all workstations.
User List
Creating a user list is done by clicking on the User List button on the General tab in
System Configuration. Refer to the operating system manual if necessary.
Antivirus Software
Antivirus software can be installed as long as the CASE meets the minimum
specifications. D:\CASE folder should be excluded. Scans and updates should not be
scheduled during patient testing hours.
www.gehealthcare.com