Professional Documents
Culture Documents
Figure (a) Diagram showing the separation of a mixture of components A and B by column
elution chromatography. (b) The output of the signal detector at the various stages of elution
shown in (a).
• In modern chromatography, continuous addition is accomplished by means of a pump in
LC and SFC or by the application of pressure in GC.
Working:
• With the first introduction of fresh mobile phase, the eluent— the portion of the sample
contained in the mobile phase—moves down the column, where it further distributes
between the mobile phase and the stationary phase (time t1).
• Interaction of the solute with the fresh mobile phase and the stationary phase takes place
as the mobile phase washes the sample through the column.
• As fresh mobile phase flows through the column, it carries solute molecules down the
column in a continuous series of interactions with the two phases
• Because solute movement can occur only in the mobile phase, the average rate at which a
solute zone migrates down the column depends on the fraction of time it spends in the
mobile phase.
• This fraction is small for solutes strongly retained by the stationary phase (component B)
and large when the solute resides mostly in the mobile phase (component A).
• Ideally, the resulting differences in rates cause the components in a mixture to separate
into bands, or zones, along the length of the column (t2).
• Isolation of the separated species is then accomplished by passing a sufficient quantity of
mobile phase through the column to cause the individual zones to pass out the end, where
they can be detected or collected (times t3 and t4).
Analyte Dilution:
• Figure above illustrates an important general characteristic of the chromatographic
process—namely, dilution of analytes always accompanies chromatographic separation.
• Thus, the size of the original zone containing the analytes in the figure is noticeably
smaller than either of the two zones that reach the detector, meaning that significant
dilution of the analytes and band spreading have occurred while they were being
separated.
• As a result, the detectors employed for separated analytes must often be more sensitive
than would be required if the separation process were unnecessary.
Chromatograms
• If a detector that responds to solute concentration is placed at the end of the column and
its signal is plotted as function of time (or of volume of the added mobile phase), a series
of peaks is obtained
• Such a plot, called a chromatogram, is useful for both qualitative and quantitative
analysis.
• The positions of peaks on the time axis can be used to identify the components of the
sample; the areas under the peaks provide a quantitative measure of the amount of each
component.
4.2 Band broadening and optimization of column performance
• A chromatographic separation is optimized by varying experimental conditions until the
components of a mixture are separated cleanly in a minimum amount of time.
• Optimization experiments are aimed at either
(1) reducing zone broadening (the component moves through the column as a zone which
is detected by the detector and translated into a peak). or
(2) altering relative migration rates of the components.
• Zone broadening is increased by those kinetic variables that increase the plate height of a
column.
• Migration rates, on the other hand, are varied by changing those variables that affect
retention and selectivity factors.
where
u is the linear-flow velocity (cm/s),
r is the radius of the tube (cm), and
DM is the diffusion coefficient of the solute in the mobile phase (cm2/s).
Extracolumn broadening can become quite serious when small-bore columns are used.
Here, the radius of the extracolumn components should be reduced to 0.010 inch or less,
and the length of extracolumn tubing made as small as feasible to minimize this source of
broadening.
4.3 Liquid chromatography - Instrumentation
• Pumping pressures of several hundred atmospheres are required to achieve reasonable
flow rates with packings of 3–10 mm, which are common in modern LC.
• Because of these high pressures, the equipment for HPLC tends to be more elaborate and
expensive than equipment for other types of chromatography.
• Figure below is a diagram showing the important components of a typical LC instrument.
Mobile-Phase Reservoirs and Solvent Treatment Systems :
• A modern LC apparatus is equipped with one or more glass reservoirs, each of which
contains 500 mL or more of a solvent.
• Provisions are often included to remove dissolved gases and dust from the liquids.
FIGURE Block diagram showing components of a typical apparatus for HPLC.
• Since, dissolved gases can lead to irreproducible flow rates and band spreading; in
addition, both bubbles and dust interfere with the performance of most detectors.
• To avoid this , Sparger is used.
• In Sparging method , the dissolved gases are swept out of the solution by fine bubbles
of an inert gas that is not soluble in the mobile phase.
• Another convenient way of treating solvents before introduction into the reservoir is to
filter them through a millipore filter under vacuum.
• This treatment removes gases as well as suspended matter.
• An elution with a single solvent or solvent mixture of constant composition is termed an
isocratic elution.
• In gradient elution, two (and sometimes more) solvent systems that differ significantly in
polarity are used and varied in composition during the separation.
• The ratio of the two solvents is varied in a preprogrammed way, sometimes continuously
and sometimes in a series of steps.
• Modern HPLC instruments are often equipped with proportioning valves that introduce
liquids from two or more reservoirs at ratios that can be varied continuously.
Pumping Systems :
• The requirements for liquid chromatographic pumps include
1) the generation of pressures of up to 6000 psi (lb/in.) or 414 bar
2) pulse-free output,
3) flow rates ranging from 0.1 to 10 mL/min,
4) flow reproducibilities of 0.5% relative or better, and
5) resistance to corrosion by a variety of solvents.
• The high pressures generated by liquid chromatographic pumps are not an explosion
hazard because liquids are not very compressible.
• Thus, rupture of a component results only in solvent leakage.
• However, such leakage may constitute a fire or environmental hazard with some solvents.
• Two major types of pumps are used in LC:
1. the screw-driven syringe type and
2. the reciprocating pump.
• Reciprocating pumps are used in almost all modern commercial chromatographs.
Reciprocating Pumps :
• Reciprocating pumps usually consist of a small chamber in which the solvent is pumped
by the back and forth motion of a motor driven piston
• Typical silica surfaces contain about 8 μmol/m2 of OH groups and have surface areas of
100–300 m2/g.
Normal- and Reversed-Phase Packings:
• Two types of partition chromatography are distinguishable based on
the relative polarities of the mobile and stationary phases.
Normal-phase chromatography
Reversed-phase chromatography
In normal-phase chromatography , highly polar stationary phases is triethylene glycol or
water; a relatively mobile phase. nonpolar solvent is hexane or i-propyl.
In reversed-phase chromatography, the stationary phase is nonpolar, often a
hydrocarbon, and the mobile phase is a relatively polar solvent (such as water, methanol)
In normal-phase chromatography, the least polar component is eluted first; increasing the
polarity of the mobile phase then decreases the elution time.
With reversed-phase chromatography, however, the most polar component elutes first,
and increasing the mobile-phase polarity increases the elution time.
Bonded-phase packings are classified as reversed phase when the bonded coating is
nonpolar in character and as normal phase when the coating contains polar functional
groups.
The major advantage of reversed-phase separations is that water can be used as the
mobile phase. Water is an inexpensive, nontoxic, UV-transparent solvent compatible with
biological solutes. Also, mass transfer is rapid with nonpolar stationary phases
Method Development in Partition Chromatography:
• Method development tends to be more complex in LC because in a liquid mobile phase
the sample components interact with both the stationary phase and the mobile phase .
• The success of a partition chromatographic separation is often critically dependent on
whether the mobile phase is, say, acetonitrile, hexane, or dioxane.
Column Selection in Partition Chromatographic Separations:
• Successful chromatography with interactive mobile phases requires a proper balance of
intermolecular forces among the three active participants in the separation process—the
solute, the mobile phase, and the stationary phase.
• These intermolecular forces are described qualitatively in terms of the relative polarity of
each of the three reactants.
• The polarities of the various analyte functional groups increase in the following order:
• Hydrocarbons < ethers < esters < ketones < aldehydes < amides < amines < alcohols.
• Water is more polar than compounds containing any of the preceding functional groups.
• Often, in choosing a column for a partition chromatographic separation, the polarity of
the stationary phase is matched roughly with that of the analytes; a mobile phase of
considerably different polarity is then used for elution.
• This procedure is generally more successful than one in which the polarities of the solute
and mobile phase are matched but different from that of the stationary phase.
• Here, the stationary phase often cannot compete successfully for the sample components;
retention times then become too short for practical application.
• At the other extreme, of course, is the situation where the polarity of the solute and
stationary phases are too much alike and totally different from that of the mobile phase.
• Here, retention times become inordinately long.
• In summary, then, polarities for solute, mobile phase, and stationary phase must be
carefully blended if good partition chromatographic separations are to be realized in a
reasonable time.
Effect of Solvent Strength on Retention Factors :
• Solvents that interact strongly with solutes are often termed “strong” solvents.
• Strong solvents are often, but not always, polar solvents.
• Solvent strength depends on the nature of the analyte and stationary phase. Several
indexes have been developed for quantitatively describing the polarity of solvents.
• The most useful of these for partition chromatography is the polarity
index P’, which was developed by Snyder.
• The polarity index is a numerical measure of the relative polarity of various solvents.
Applications of Partition Chromatography :
• The table below lists the few typical examples of the multitude of uses of partition
chromatography in various fields
4.5 Adsorption chromatography
• Adsorption, or liquid-solid, chromatography is the classic form of LC first introduced by
Tswett at the beginning of the last century.
• Because of the strong overlap between normal- phase partition chromatography and
adsorption chromatography, many of the principles and techniques used for the former
apply to adsorption chromatography.
• In fact, in many normal-phase separations, adsorption-displacement processes govern
retention.
• Finely divided silica and alumina are the only stationary phases that find use for
adsorption chromatography.
• Silica is preferred for most applications because of its higher sample capacity.
• The adsorption characteristics of the two substances parallel one another. For both,
retention times become longer as the polarity of the analyte increases.
• For both, retention times become longer as the polarity of the analyte increases.
• The polarity index P’, which gives idea to the strengths of solvents for adsorption
chromatography.
• Eluent strength Ɛ0 is a much better index, which is the adsorption energy per unit area of
solvent.
• The eluent strength depends on the adsorbent, with Ɛ0 values for silica being about 0.8 .
• Because of the versatility and ready availability of bonded stationary phases, use of
traditional adsorption chromatography with solid stationary phases has decreased in
recent years in favor of normal-phase chromatography.
4.6 Ion exchange chromatography
• Ion exchange chromatography definition (or ion chromatography) is a process that
allows the separation of ions and polar molecules based on their affinity to the ion
exchanger.
• It can be used for almost any kind of charged molecule including large proteins, small
nucleotides, and amino acids.
• The principle of separation is thus by reversible exchange of ions between the target ions
present in the sample solution to the ions present on ion exchangers.
• In this process two types of exchangers i.e., cationic and anionic exchangers can be used.
• Cationic exchangers possess negatively charged group, and these will attract positively
charged cations. These exchangers are also called “Acidic ion exchange” materials,
because their negative charges result from the ionization of acidic group.
• Anionic exchangers have positively charged groups that will attract negatively charged
anions. These are also called “Basic ion exchange” materials.
• Ion exchange chromatography is most often performed in the form of column
chromatography.
Principle of ion exchange chromatography:
• This form of chromatography relies on the attraction between oppositely charged
stationary phase, known as an ion exchanger, and analyte.
• The ion exchangers basically contain charged groups covalently linked to the surface of
an insoluble matrix.
• The charged groups of the matrix can be positively or negatively charged.
• When suspended in an aqueous solution, the charged groups of the matrix will be
surrounded by ions of the opposite charge.
• In this “ion cloud”, ions can be reversibly exchanged without changing the nature and the
properties of the matrix.
• Ion-exclusion chromatography finds numerous applications for identification and
determination of acidic species in milk, coffee, wine, and many other products of
commerce.
• Salts of weak acids can be analyzed because they are converted to the corresponding acid
by the hydrogen ions in the exchanger. Weak bases and their salts can also be determined
by ion-exclusion chromatography.
Working:-
• Consider a column having E- Y+ cation exchanger in which E - is negative charged
exchanger and Y+ is the mobile counter ion.
• Let X+ be the cation in the sample having charge greater than Y+ .
The X+ ion can exchange sites with the counter ion Y+ with satisfying the following relationship;
E - Y+ X+ → E - X+ Y+ + The remaining neutral and negatively charged particles
Desired bounded cation (X+ ) can now be eluted by either of the two ways;
1. By adding a component M+ having magnitude of charge more than that of X+ so that M+
will replace X+ and X+ will be eluting out.
2. By changing pH of the solvent (mobile phase so that X+ have no charge and is then
unbounded from the matrix and can be eluted out.
COMPONENTS OF A SEC
1. Stationary Phase
2. The Mobile Phase
3. The Columns
4. The Pump
5. Detectors
STATIONARY PHASE:
• Stationary Phase Semi-permeable, porous beads with welldefined range of pore sizes .
• Beads are crosslinked polymers
• Degree of crosslinking is controlled carefully to yield different pore sizes.
• Smaller pore sizes are used for rapid desalting of proteins or for protein purification.
• Intermediate pore sizes are used to separate relatively small proteins.
• Very large pore sizes are used for purification of biological complexes.
• Stationary phase used for gel exclusion chromatography include dextran (Sephadex™),
polyacrylamide and dextranpolyacrylamide (Sephacryl™).
• Each is available with a variety of different ranges of pore size in the beads, permitting
separation of macromolecules of different size.
PUMP
A highly constant flow rate has to be maintained during the entire chromatogram. This is
very important in SEC.
A change of the flow rate of only 0.1% can cause an error in molar mass of up to 10%.
Most pumps can only reproduce the flow rate to 0.2–0.3%.
In-line filters in the solvent reservoir may prevent particles from coming into the pump
heads, which might damage the check valves or the pump seals.
Types-)Syringe pumps, Reciprocating pumps
DETECTOR
Concentration sensitive detectors
Bulk Property Detectors- Refractive Index (RI) Detector
Solute Property Detectors- Ultraviolet (UV) Absorption Detector
Evaporative Detectors- Evaporative Light Scattering Detector (ELSD)
Molar mass sensitive detectors
Light Scattering Detectors
Low Angle Light Scattering (LALS) Detectors
Multiangle Light Scattering (MALS) detectors
Viscosity Detectors- Differential Vscometers
Other :- Flame Ionization Detector (FID), A Mass
Spectrometer or A Fourier Transform Infrared (FTIR) Spectrometer
ADVANTAGES
Short analysis time.
Well defined separation.
Narrow bands and good sensitivity.
There is no sample loss.
Small amount of mobile phase required.
The flow rate can be set.
DISADVANTAGES
Limited number of peaks that can be resolved within the short time scale of the GPC run.
Filtrations must be performed before using the instrument to prevent dust and other
particulates from ruining the columns and interfering with the detectors.
The molecular masses of most of the chains will be too close for the GPC separation to
show anything more than broad peaks
APPLICATION
Proteins fractionation
Purification
Molecular weight determination.
Separation of sugar, proteins, peptides, rubbers and others on the basis of their size.
This technique can be determine the quaternary structure of purified proteins.
SEC is a widely used technique for the purification and analysis of synthetic and
biological polymers, such as protein, polysaccharides and nucleic acid.
Various species of RNA and viruses have been purified using agarose gels.
For Desalting
For copolymerisation studies
where
F is the average volumetric flow rate within the column;
V and t are retention volumes and times, respectively; and
The subscripts R and M refer to species that are retained and not retained on the
column.
The flow rate within the column is not directly measurable. Instead, the rate of gas flow
as it exits the column is determined experimentally with a flow meter
For popular soap-bubble-type flow meters, where the gas is saturated with water, the
average flow rate F is related to the measured flow rate Fm by
Where
Tc is the column temperature in kelvins,
T is the temperature at the flow meter, and
P is the gas pressure at the end of the column.
Usually P and T are the ambient pressure and temperature.
The term involving the vapor pressure of water, PH2O, is a correction for the pressure used when
the gas is saturated with water.
• Both VR and VM depend on the average pressure within the column—a quantity that lies
intermediate between the inlet pressure Pi and the outlet pressure P (atmospheric pressure).
• The pressure drop correction factor j, also known as the compressibility factor, accounts for
the pressure within the column being a nonlinear function of the Pi/P ratio.
• Corrected retention volumes and , which correspond to volumes at the average
column pressure, are obtained from the relationships
-------------(1)
where j can be calculated from the relationship
----------------------------------------------------(2)
The specific retention volume Vg is then defined as
-------------------------(3)
where mS is the mass of the stationary phase, a quantity determined at the time of column
preparation.
Relationship between Vg and K:
• The specific retention volume Vg can be related to the distribution constant Kc by
------------------------------------------------------------(4)
Combining this with Equation (2) we, get
------------------------------------------------------------------(5)
--------------------------------------------------------------------(6)
The density of the liquid on the stationary phase ρS is given by
------------------------------------------------------------------------(7)
------(8)
Applications of GC :
• GC has two roles
• First, GC is a tool for performing separations.
• In this role, GC methods are unsurpassed when applied to complex organic, metal-
organic, and biochemical systems made up of volatile species or species that can be
derivatized to yield volatile substances.
• The second role that GC plays is in the completion of an analysis.
• In this role, retention times or volumes are used for qualitative identification, and peak
heights or peak areas provide quantitative information.
• For qualitative purposes, GC is much more limited than most of the spectroscopic
methods.
• Thus, an important trend in the field has been in the direction of combining the
remarkable separation capabilities of GC with the superior identification properties of
such instruments as mass, infrared, and nuclear magnetic resonance spectrometers
Capillary electrophoresis
Capillary electrophoresis (CE) yields high-speed, high-resolution separations on
exceptionally small sample volumes (0.1–10 nL in contrast to slab electrophoresis, which
requires samples in the μL range).
Additionally, the separated species are eluted from one end of the capillary, so
quantitative detectors, similar to those found in high-performance liquid chromatography
(HPLC)
Migration Rates in CE
As Equation -1 shows, the migration rate of an ion v depends on the electric field strength. The
electric field in turn is proportional to the magnitude of the applied voltage V and inversely
proportional to the length L over which it is applied.
Thus
-----------------------------------------(2)
This relationship indicates that high applied voltages are desirable to achieve rapid ion
migration and fast separations.
It is desirable to have rapid separations, but it is even more important to achieve high-
resolution separations.