Professional Documents
Culture Documents
https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
1 Introduction
A cardiac defibrillator is a medical electrical device used to defibrillate the heart by
applying electrical pulses to the patient's skin (external electrodes) or exposed heart (internal
electrodes) through electrodes. The width of the instantaneous high-energy pulse is (4-10)
ms, and the output energy is (40-400) J. The output energy of the cardiac defibrillator is very
important. It may not reach the effect if too small, or the patient may be injured if too large.
In JJF1149-2014 Calibration Specification for Cardiac Defibrillators [1], defibrillator
analyzers are used to measure the output energy of cardiac defibrillators. Therefore, the
energy accuracy of the defibrillator analyzer is an important parameter.
By querying related literature, and combining years of experience, we summarized two
methods, comparison method [2,3] and power integration method [4-8].
a) comparison method
This method requires a defibrillator and a standard defibrillator analyzer as a standard
device. Firstly, use the standard defibrillator analyzer to calibrate the energy value of the
defibrillator, denoted as E0. Then use the calibrated defibrillator analyzer to measure the
energy value of the defibrillator, denoted as E1. Set ∆𝐸𝐸𝐸𝐸 the error of energy value, then
*
Corresponding author: zhanzq@simt.com.cn
© The Authors, published by EDP Sciences. This is an open access article distributed under the terms of the Creative Commons
Attribution License 4.0 (http://creativecommons.org/licenses/by/4.0/).
MATEC Web of Conferences 336, 02010 (2021) https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
The schematic diagram of the equipment is shown in Fig. 1. The voltage divider module
uses a voltage divider, and divider ratio r is a fixed value, such as 1000:1. The sampling
module uses an oscilloscope, whose voltage is donated as 𝑉𝑉𝑉𝑉𝑜𝑜𝑜𝑜 (𝑛𝑛𝑛𝑛). The defibrillator analyzer
has a fixed resistance, usually 50Ω, which can be calibrated by a digital multimeter. Then
𝑇𝑇𝑇𝑇
𝐸𝐸𝐸𝐸 = ∑𝑛𝑛𝑛𝑛1(𝑉𝑉𝑉𝑉𝑜𝑜𝑜𝑜 (𝑛𝑛𝑛𝑛) × 𝑟𝑟𝑟𝑟)2 × 𝑅𝑅𝑅𝑅 (4)
Defibrillator Defibrillator
Analyzer
The discrete Fourier transform (DFT) of a finite-length discrete signal x(n) is defined as
2𝜋𝜋𝜋𝜋
−𝑗𝑗𝑗𝑗
𝑋𝑋𝑋𝑋(𝑘𝑘𝑘𝑘) = ∑𝑁𝑁𝑁𝑁−1 𝑘𝑘𝑘𝑘𝑛𝑛𝑛𝑛
𝑛𝑛𝑛𝑛=0 𝑥𝑥𝑥𝑥(𝑛𝑛𝑛𝑛)𝑊𝑊𝑊𝑊𝑁𝑁𝑁𝑁 , 𝑘𝑘𝑘𝑘 = 0,1, … , 𝑁𝑁𝑁𝑁 − 1, 𝑊𝑊𝑊𝑊𝑁𝑁𝑁𝑁 = 𝑒𝑒𝑒𝑒 𝑁𝑁𝑁𝑁 (6)
Fig. 2.b shows the curve of the pulse signal in the frequency domain by MATLAB. The
energy of the signal is mainly concentrated in the low frequency part, that is, the part from
DC to 1kHz. Therefore, the defibrillation analyzer calibration system needs to meet
requirements of the AC voltage peak value greater than 3.16kV, DC to 1kHz with high
accuracy.
2
MATEC Web of Conferences 336, 02010 (2021) https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
2.a 2.b
Fig. 2.a Waveform generated by the defibrillator 2.b curve in frequency domain.
When using a traditional oscilloscope for measurement, we must save data into a csv file,
then copy the file to a computer for data processing. Thus, the results cannot be displayed in
real time.
We write LabVIEW programs for NI oscilloscope, and the program can display result in
real time.
The front panel of the energy test program is as follows. Resistance is used to set the
3
MATEC Web of Conferences 336, 02010 (2021) https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
resistance value, that is, the internal resistance value of the defibrillation analyzer measured
by the 3458A multimeter. The Energy indicator displays the calculated energy value.
4
MATEC Web of Conferences 336, 02010 (2021) https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
∑𝑛𝑛𝑛𝑛
𝑖𝑖𝑖𝑖=1(𝑥𝑥𝑥𝑥𝑖𝑖𝑖𝑖 −𝑥𝑥𝑥𝑥̅ )
2
𝑠𝑠𝑠𝑠(𝑥𝑥𝑥𝑥) = � (7)
𝑛𝑛𝑛𝑛−1
The uncertainty components 𝑢𝑢𝑢𝑢1rel , 𝑢𝑢𝑢𝑢2rel , 𝑢𝑢𝑢𝑢3rel , 𝑢𝑢𝑢𝑢4rel are independent of each other.
According to the formula
𝑇𝑇𝑇𝑇
𝐸𝐸𝐸𝐸 = ∑𝑛𝑛𝑛𝑛1(𝑉𝑉𝑉𝑉𝑜𝑜𝑜𝑜 (𝑛𝑛𝑛𝑛) × 𝑟𝑟𝑟𝑟)2 × 𝑅𝑅𝑅𝑅 (8)
5
MATEC Web of Conferences 336, 02010 (2021) https://doi.org/10.1051/matecconf/202133602010
CSCNS2020
In summary, there are two main sources of uncertainty. One is 𝑢𝑢𝑢𝑢𝐸𝐸𝐸𝐸rel, and one is 𝑢𝑢𝑢𝑢6rel .
These two components are independent of each other.
Calculate the relative combined standard uncertainty at different energy level respectively.
Table 5. Relative combined standard uncertainty
Energy level 30J 50J 100J 200J 300J 360J
𝑢𝑢𝑢𝑢𝑐𝑐𝑐𝑐rel 0.808% 0.797% 0.792% 0.790% 0.790% 0.790%
4 Results
We firstly analyzed the defibrillation signal generated by a defibrillator, then designed the
measurement equipment, including voltage divider module, sampling module, and data
processing module. With LabView program, the equipment can measure the energy value in
real time. At last, the uncertainty evaluation results are given.
Acknowledgments
This work is supported by Shanghai Municipal Bureau of Quality and Technical Supervision
2018-24, and Shanghai Institute of Measurement and Technology F00RY1611.
References
1. JJF1149-2014 Calibration Specification for Cardiac Defibrillators, GAQSIQ, (2014).
2. QED-6 user’s guide, BIO-TEK, (2002).
3. Cui Li. Analysis and evaluation of the uncertainty of the defibrillation analyzer
verification results. Chinese Medical Equipment, 36-38, 11, 12(2014).
4. Zhang Huanghui. Development of calibration system for defibrillator analyzers[J].
Electrical Instrument & Measurement. 56, 11(2019).
5. Zhang Yadong, Li Yanfeng, et al. Research on defibrillation pulse energy measurement
system based on STM32. Medical and Medical Equipment, 8-10, 36, 1(2015).
6. Yao Guohong, Zhang Yadong, Li Yanfeng, et al. Development of defibrillation energy
standard source. Medical and Medical Equipment, 10-13, 36, 9(2015).
7. Su Yanping, Yao Nianling, Cui Li. Analysis and research on quality verification methods
of cardiac defibrillators. China Medical Equipment, 22-24, 11,3 (2014).
8. Wei Chuan, Wang Lin, Yan Ling. Research on Traceability Method of Cardiac
Defibrillator Testing and Calibration Equipment. China Test, 34-36, 37, 5(2011).