電子承認完了図書
No. 2C201-166EN
INSTALLATION MANUAL
FOR
64-SLICE SYSTEM KIT CGS-38A
FOR
WHOLE-BODY X-RAY CT SCANNER
TSX-101A
(2C201-166EN)
No. 2C201-166EN
Trademarks
Aquilion is a trademark of Toshiba Medical Systems Corporation.
This manual may include trademarks of other companies.
The marks and are not shown in this manual.
IMPORTANT!
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
3. The contents of this manual are correct to the best of our knowledge.
Please inform us of any ambiguous or erroneous descriptions, missing
information, etc.
2
No. 2C201-166EN
REVISION RECORD
3
No. 2C201-166EN
CONTENTS
Page
1. SCOPE----------------------------------------------------------------------------------------------------------------------- 5
2. COMPOSITION ------------------------------------------------------------------------------------------------------------ 5
4
No. 2C201-166EN
1. SCOPE
This manual describes the procedures for installing the optional 64-slice system kit for whole-body
X-ray CT scanner Aquilion 32 (TSX-101A).
2. COMPOSITION
(2) Label--------------------------------------------------------------- 1
5
No. 2C201-166EN
3h
An option license can be handled by saving it to a 3.5-inch floppy disk or as a key code. Usually, a
3.5-inch floppy disk is used for shipment.
Use the window shown in figure 1 below which is displayed by selecting [[Utility]] → [[Engineering]],
entering the password, and then selecting [[Option License]].
Figure 1
(1) Procedure for checking the license on the disk (3.5-inch floppy disk)
Insert the floppy disk in the drive and select [[Media check]].
The window shown in figure 2 is displayed, showing the option name in the License Name
column.
At this time, if the license is not available, "----" is displayed in the UNIQUE ID field.
The license name is "SCAN64ROW".
Figure 2
6
No. 2C201-166EN
The window shown in figure 3 is displayed, indicating that the license is available in the
system.
Use the method described in (1) to confirm that the UNIQUE ID is on the disk.
This work must be performed for both the S-CON and D-CON.
Figure 3
In the case where the license for the D-console is incorrectly applied to the S-console, etc.,
the license can be returned to the disk. Place the corresponding disk in the system in which
the license was incorrectly applied and then select [[Unload license]]. After operation is
completed, use the method described in (1) to check the disk and confirm that "----" is
displayed in the UNIQUE ID field.
7
No. 2C201-166EN
When [[List]] is selected in the window shown in figure 1, the window shown in figure 4 is
displayed. In this window, licenses that have already been made available and licenses that
have not been made available can be checked.
Although the license status can also be checked by selecting [[Utility]] → [[Option]] →
[[License List]], this method displays only licenses that have already been made available.
Figure 4
(6) Perform RDD setup at the S-CON. Select "Restore configuration file from RDD" and "Write
FC, PRE and REC params to RDD".
8
No. 2C201-166EN
(7) After initialization is completed, reboot the system and acquire calibration data for 64-slice
acquisition using the calibration data acquisition function.
After the setting procedures up to those described in subsection 3.4 are completed, perform
operational checks.
(1) Select an eXam Plan. Confirm that 64-slice acquisition (0.5 × 64) can be selected for a
helical scan.
Attach the model label of this option kit at a location near the system model label on the gantry.
9 E