WI - 12 - Final Unit Test - Infusion Pump
WI - 12 - Final Unit Test - Infusion Pump
: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
The following document shall be used by testing engineer at production to conduct the final
unit testing. This document contains the detailed procedure of each test to be conducted.
DAY 1
1.0 Service settings:
Menu key long press when set key is disabled which will enter into service settings.
Compare the setting values available to default values.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
1.2 Menu setting:
Enable the set key then press the menu key and enter into Menu setting.
Compare the setting values available to default values.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
1.5 Factory Settings:
Set disabled condition, long press Clear key + dec. key to enter into factory settings.
Manual calibration:
1. Fix the IV set properly
2. Take 10ml syringe and set the piston at 10ml mark and place output of IV to syringe.
3. Run 100mL/Hr. Once 10mL reached stop the machine. Check total drop counts. Drops/ml
Ratio= Total drop counts divided by 10
**Go to service settings and Edit Drop/ml ratio for Adult & Pedia IV
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
For checking the event logs, enable the set key and press menu key and enter into the menu
settings to view the events by generating the above-mentioned events.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
2. Lo-FL Alarm - Run the machine in High Flow rate and Lock the IV regulator. Machine should
give
Lo-FL Alarm.
3. Hi-FL Alarm - Run the machine in Low Flow rate and cut the Drop Sensor sensing area as fast as
you can repeatedly. Machine should give Hi-FL Alarm.
4. tr-1 Alarm - Set some Target volume 1.0 ml and run the machine in high flow rate. After
reaching target volume machine should give tr-1 Alarm.
5. tr-2 Alarm - Set some Target time1 minute and run the machine in low flow rate. After
reaching target time machine should give tr-2 Alarm.
6. AL3 Alarm - AL3 ALARM should not come through out the testing days.
7. Pressure drop alarm-Block the IV tube then remove the blocking scissors machine should give
pressure drop alarm.
8. KVO - SET any KVO flow rate in MENU settings. Now set the target volume or time and run
the machine. After target reached machine should run in KVO Flow rate.
9. KVO OFF - SET KVO OFF in MENU settings. Now set the target volume or time and run the
machine. After target reached machine should give only TARGET alarm and KVO should not
run.
10. AIR alarm - Turn air alarm ON in MENU Settings. Generate air bubble at chamber side and
run the machine. After few movements air alarm should come.
11. Battery OFF - Verify this OFF alarm when Battery backup is testing. After Battery dry then
this alarm should come.
12.Door open alarm-Verify the door open alarm or close the door on the time of run
condition.
13.Occlusion Alarm-Verify the blockage in the IV tube or check the mode of levels in the
menu setting.
Note: Mute key should be used to mute the Alarm.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
2.0: Accuracy test (AC mode)- Adult tube:
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode and F/R mode to set dr/min.
2. Set the flowrate and enable set key and press target key to set target time.
Set F/R- 100 dr/min. Target Time- 5 mins.
3. Collect the volume either in syringe or Collecting beaker.
4. Note the accuracy using the reference table mentioned the value in Accuracy Table.
5. Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6. Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
2.1: Accuracy test (AC mode)- Adult tube:
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode and F/R mode to set ml/hr.
2. Set the flowrate and enable set key and press target key to set target time.
Set F/R- 500 ml/hr. Target Volume- 450 ml.
3. Collect the volume either in syringe or Collecting beaker.
4. Note the accuracy using the reference table mentioned the value in Accuracy Table.
5. Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6. Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
2.2: Accuracy test (AC mode)- Adult tube-Night loop test:
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode and F/R mode to set ml/hr.
2. Set the flowrate for 300 ml/hr.
3. Put the tube into the loop and run the machine for ≤ 7hrs
4. If the machine run for ≤ 7hrs. Then it is pass
DAY 3
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
3.0: Accuracy test (Battery mode)- Romsons tube:
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode and F/R mode to set ml/hr.
2. Remove the Power cord for Battery mode and set the flowrate to 25 ml/hr.
3. Collect the volume either in syringe or Collecting beaker until the machine OFF.
4. Backup hrs: ≥ 4hrs.
5. Note down the time when machine is offed due to battery drained.
6. Calculate the start time and end time. If the time is reached more than 3 hours enter
the testis passed else enter the test is failed.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
2. Set the flowrate and enable set key and press target key to set target time.
Set F/R- 10 dr/min. Target Time- 15 mins.
3.Collect the volume either in syringe or Collecting beaker.
4.Note the accuracy using the reference table mentioned the value in Accuracy Table.
5.Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6.Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
DAY 4
4.0: Accuracy test (AC mode)- Micro tube (Romson):
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
micro mode and F/R mode to set ml/hr.
2. Set the flowrate and enable set key and press target key to set target time.
Set F/R- 100 ml/hr. Target time- 30 min.
3. Collect the volume either in syringe or Collecting beaker.
4. Note the accuracy using the reference table mentioned the value in Accuracy Table.
5. Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6. Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
micro mode and F/R mode to set ml/hr.
2. Set the flowrate for 200 ml/hr.
3. Put the tube into the loop and run the machine for ≤ 7hrs
4. If the machine run for ≤ 6hrs. Then it is pass
DAY 5
5.0: Bolus test- Purge bolus (Manual Bolus)- AC mode:
1. Enable the set key and press menu key to enter menu settings. Go to bolus mode and
set purge bolus.
2. Press set key and press bolus key to set bolus Flowrate for 300 ml/hr.
3. Collect the volume for 5ml with the syringe.
4. Observe the value in syringe and enter the value in table. If the syringe is collected in
5 ml enter the test is passed, else enter the test is fail.
5. Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6. Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
1. Enable the set key then press the menu key and enter into Menu setting and change
the mode Auto bolus.
2. Remove the power cord for battery mode. Press set key and press bolus key to set
bolus Flowrate for 60 ml/hr.
3. Set the target for 1 hr.
4. Check the drop count on machine that matches with actual drop of 1200 drops.
5. Backup hrs: >3hrs
6. Note down the time when machine is offed due to battery drained.
7. Calculate the start time and end time. If the time is reached more than 3 hours enter
the testis passed else enter the test is failed.
DAY 6
6.0 Occlusion Blocking test- Adult & micro:
Procedure for setting the PSI& mmHg
Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
First go to the service setting number 10 and change the pressure units by pressing
target key (PSI or mmHg) then turn ON the pressure unit display by using the
increment key.
Go to the menu setting number 6 and set the occlusion levels (low, medium and high)
one by one by using increment decrement key
Run the flow rate according to test chart, block the edge of the IV tube with the help
of blocking scissor.
Finally check the occlusion alarm testing for different levels of 300 mmHg,
600mmHg, 900mmHg/ 6PSI, 12PSI, 18PSI.
The tolerance level should be ±3PSI or ±150 mmHg.
Perform occlusion for the flowrates 50ml/hr and 50 dr/min.
6.1 Accuracy night test- Romsons Adult:
1. Enable the set key and press menu key to enter menu settings. Go to Tube type to set
adult mode and F/R mode to set ml/hr.
2. Set the flowrate to 10 ml/hr and enable set key and press target key to set target time
for 16hrs
3. Collect the volume in beaker for 160 ml.
4. Note the accuracy using the reference table mentioned the value in Accuracy Table.
5. Calculation of Actual ml: Actual Accuracy % = {Collected O/p mL}/ {Actual mL
should deliver} x 100.
6. Calculation of Actual ml: Machine Accuracy % = {Elapsed mL shown in Machine}/
{Collected mL} x 100.
DAY 7
7.1 Packing Day test:
A. Used Hours Resetting:
1. In machine stop state, long press the unit down key and Target key to see
machine Used Hours (in hours).
2. To reset the Used Hours, long press Clear key in the used hours settings then the
Prepared by K. Sai Kumar Approved by U. Suganth
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
D. Verification by QC:
To verify the below functions by QC
1. Keep KVO Flow rate as 0.1ml/hr.
2. Machine Run in AC and DC condition.
3. Plug in the power Chord Orange colour indication mains’ icon should glow in LCD
display.
4. Plug out the power Chord battery icon should glow in DC indicator symbol.
5. tAr.1 ALARM-Target Volume reached Alarm.
Signature Signature
Doc. No.: AME/WI/12
AKAS MEDICAL EQUIPMENT
Rev. No.: 00
WORK INSTRUCTION Issue No.: A
FINAL UNIT TEST – INFUSION PUMP Date: 29.12.2022
Signature Signature