Professional Documents
Culture Documents
CO2 Bypass
and Condenser
APL valve Option
0-70 cm H2O
Vent Engine
Atmosphere
NO
Free
Absorber
Drain
Breathing
Check Valve NC Canister
Bag Drain
NC Button
NO
Mechanical Popoff
Overpressure Valve Valve Bag/Vent
(110 cm H2O)
(10 cm H2O) Atmosphere
Drive Gas Check Condenser
Valve (3.5 cm Exhalation Valve
H2O bias) (2.0 cm H2O bias)
0-10 LPM Drive Gas
0-10 LPM Patient and Fresh Gas
Inspiratory 0-20 LPM Total Typical Flow
Flow Control To Scavenging
Valve
O2 Flush 4 Sample Gas Return
(opt)
Aux O2
0-10 LPM
(opt) 35 psi
2nd O2
Reg 200 mL Reservoir Negative Pressure
0-120 LPM Flow Vent Drive 25 psi Expiratory Gas Monitor
@15 LPM relief valve Flow Sensor
Gauge P Gas
Inlet Control Bleed to Ambient
System Switch Valve 1.0 LPM @ 3.0 cm H2O
O2 P-LINE if continuous (rate dependent)
Drain
NC
NO
Patient
O2 Cyl (opt)
P Gauge
* 5.4 psi
Drive Balance
Gas 30 psi Regulator Air (opt) O2
NO
Select Flush
Venturi
Switch O2 Inspiratory
Gauge P
NC Sensor Flow Sensor
O2 Cyl (opt)
Paw Trans
Drive Pressure
Limit Switch
Relief Gauge
Selectatec Manifold 5.5 psi Pressure
Relief Valve
* = Max 3 cylinders
= Test Port Plug Note: For System Software prior to 4.8,
= ABS Bulkhead Ports the DPL switch is plumbed to the
Manifold Pressure Transducer,
2
refer to Section 2.6.5.
1
VAP VAP
4 3
*
Scavenging
From System *
Scavenging
From System
Scavenging
From System*
Room
High or Low Air
Filter Flow Orifice
High or Low
Flow Connector to
Needle Valve Disposal System
30 mm Male Assembly with
Connector DISS EVAC
Connector
AB.74.031
Key to Symbols
= Plugged port (1/8 inch) for sample gas return.
= Plugged port (30 mm) for auxiliary breathing system scavenging.
= Open port (30 mm) for auxiliary breathing system scavenging.
* = Zero to 10 l/min drive gas; zero to 10 l/min patient and fresh gas; zero to 20 l/min total typical flow.
Note: Active AGSS systems with a 12.7 mm connector do not include the Flow Orifice and the Flow Indicator.
Control Module
Ventilator Display Upper Chassis
EL Display AC Power
Membrane Switches Task Light Outlets
Connector Board
Optical Encoder
Speaker
O2
Supply
Switch
RS232 Download
Vent
Engine
Board
Inlet Flow
Valve Control
Valve
with
O2 Circuit Breaker
Flush and
Breathing System Line Filter
Tabletop Switch Lower Chassis
Power
Cord
Control/Display Module
Task Light
RS-232
O2 N2O Air
Bag/Vent
VEB
white
black Integrated
CPU
ABS ON white
AB.74.125
black
orange
Canister
white
CO2
Bypass
Power
Supply
ESIB GND
ACGO
Line
Filter 12V
ACGO Battery
Selector
Key to Symbols Switch AC
Fuses Inlet
VEB = Vent Engine Board
ESIB = Enhanced Sensor Interface Board
ACGO = Auxiliary Common Gas Outlet
Auxiliary O2 Flowmeter
From
AB.74.083
Scavenging Downtube O2 Supply
Vaporizer Manifold
Sample
Return
* From
Vaporizer Manifold
Flowmeter
ACGO
Vent Engine
O2 N2O Air
See Note
White
Black
Blue
Yellow
White
Black
Aux O2
Enhanced
Sensor N2O
Interface
Board
AB.74.072
O2
See Vacuum
Note
ACGO
OUTLET
BREAKERS
COUNTRY SPECIFIC
OUTLETS
RV3
D6521ZOV141RA20 WHITE
140V
Common
Mode 1
G2
U-5B
250V
2 15%
GND (GRN/YEL) CHASSIS GROUND
Power Supply
EQUIPOTENTIAL STUD L
G
CHASSIS GROUND
* CONNECTOR
LINE FILTER
5A
FUSE
L L
5A
AB.74.084
N FUSE
N G
OUTLET
BREAKERS
COUNTRY SPECIFIC
OUTLETS
Power Supply
EQUIPOTENTIAL STUD
L
N
CHASSIS GROUND
G
* CONNECTOR
LINE FILTER
5A
L L FUSE
5A
AB.74.085
N FUSE
N G