Professional Documents
Culture Documents
plug insert HCG 2011 plug insert HCG 4011 plug connector
1.3 Measure the electrical resistances between all wires and shield. All measurements has to be infinite Ohm.
1.4 If any of the cable connectors does not look like in proper condition or the resistance measuring of one
wire is not infinite, the cable has to be renewed.
2.2 Do not touch the membrane. This could cause a zero-shift of the device or a permanent damage.
Remove protection cover after installation.
2.3 Protect the sensor’s connector and the cable ends from humidity.
HOPPE Bordmesstechnik GmbH
Kieler Str. 318 • 22525 Hamburg •Tel:040/561949-0 •Fax:040/561949-99 • E-Mail: service@hoppe-bmt.de
3.2 Always send back sensor together with cable for valuation complete system inhouse. Sign the used
sensor (recommended with permanent marker) with name of tank and error (current).
Please make a note, if the connection box is not OK, see 1.2 sample of a polluted HNA connecting box.
3.3 Do not separate cable from sensor or sign also the cable with name of tank.
3.4 Do not clean the menbrane if it is soiled. If possible put plastic caps on the membrane and on the plug
connector.
3.5 In case of return shipment to Hoppe, please attach a shipping note with:
- Ship‘s name
- NB-No.
- Claim or Order No.
- Hoppe Reference no. (See Hoppe delivery note)
- Name of tank
The shipping note may be even handwritten.
Purpose of redelivery:
e.g. Return delivery Claim, credit note, repair, valuation