1. In the Service window, select TankSetup and then TGE Replacement... button and
log in.
4. Press the Set Offline button. The TGE Status will change to No Service.
Note! If the selected TGE is already set offline, the Unit ID will be displayed as “0”.
151
Service Manual Rosemount CMS: TGE Replacement
Ed 08 June 2011
5. Press OK to continue.
6. Disconnect the deck cable in the SCU for the TGE you are replacing.
Note! If your system also is equipped with an independent High Level or Overfill TankRa-
dar STaR system then you must disconnect the deck cables for the TGE in those
systems as well.
TGE replacement with SCD Service CMS ed 01.fm
152
Rosemount CMS: TGE Replacement Service Manual
June 2011 Ed 08
Note! Make sure that no water enters the Gauge housing during the replacement.
Note! If the wires to the wire terminal have not previously been tagged, tag them before
removing them.
9. Remove the wire terminal from the holder or unstrap the TGE (TGU58) .
153
Service Manual Rosemount CMS: TGE Replacement
Ed 08 June 2011
Note! Do not try to open the TGE. The entire TGE is to be replaced.
12. Press the Exit button to finish the TGE Remove procedure.
154
Rosemount CMS: TGE Replacement Service Manual
June 2011 Ed 08
1. In the Service window, press the TankSetup and then the TGE Replacement... button
and log in.
4. Enter Unit ID’s for the new TGE that you are mounting. Unit ID’s are found at the top of
the TGE or on the side of the TGE (TGU58). Press the Next button.
155
Service Manual Rosemount CMS: TGE Replacement
Ed 08 June 2011
5. Fit the TGE in place. Make sure that the pin on the TGE fits into the groove on the
antenna feeder. If mounting TGU58 then make sure that the pin on the antenna feeder
fits into the larger groove on the TGE.
Note! Make sure that no water gets into the Gauge housing during the replacement.
7. Fasten the wire terminal to the holder and connect the new TGE. TGE replacement with SCD Service CMS ed 01.fm
156
Rosemount CMS: TGE Replacement Service Manual
June 2011 Ed 08
Note! Make sure the interior of the Gauge housing is completely dry and clean before you
fasten the Cover. Encased water or damp can cause corrosion on the wire terminal
and wire leads causing contact fault. Wipe with a clean and dry cloth if necessary.
Note! Change the o-ring if necessary and make sure that the o-ring is in place under the
cover.
10. Check the checkbox New TGE mounted and select the Set Online button.
TGE replacement with SCD Service CMS ed 01.fm
157
Service Manual Rosemount CMS: TGE Replacement
Ed 08 June 2011
11. A dialogue window will be displayed on the status of the procedure. Press OK to con-
tinue.
12. Verify that the TGE Status is OK. If status not OK, contact a service agent at Rose-
mount Tank Radar AB. Press Exit to finish the Add TGE procedure.
Note! It is recommended to backup the database after the TGE Replacement procedure.
See “Backup the Database” on page 89.