You are on page 1of 10

4.

Combination of Basic Data and Patient Data

As pointed out in Section 3, a patient differs from curved surface of a patient. Two points are indicated,
an homogeneous phantom in size, shape and composi- A and B. Consider first point A. The surface location
tion. In this section, methods are discussed for adjust- (along a ray line) above it is unaltered and the pri-
ing the basic data to take account of these differences. mary absorbed dose contribution to it would be un-
changed. For moderate skin obliquity, the scattered
component also would not be changed appreciably.
4.1 Contour Shape This would be especially true for HOCO and high-en-
ergy radiation and the absorbed dose at point A,
In obtaining basic absorbed dose distribution data, therefore, can be considered as unaltered by surface
the beam is usually directed at a right angle to the shape. Point B, on the other hand, has a considerably
surface of the phantom. During treatment, however, reduced amount of material above it and both the
the beam is directed through a body surface that may primary and scattered components of the radiation
be at an angle differing appreciably from a right beam will be altered. The primary component would
angle and, in addition, the surface may be curved in a be increased and the scatter slightly decreased. A
complex way. Two consequences arise from this; good value for the correction factor required can be
points below the surface may have altered thick- obtained by the ratio of tissue-air ratios for depths (d-
nesses of absorbing material between them and the h) em and d cm.
source, and they may be near to different volumes of T(d - h, Wd )
scattering material. CF = T(d, Wd ) (4.1)
Alteration in the absorbed dose due to both of the
above factors may be avoided by the use of tissue With d = 7, h = 3, and Wd = 10, from Fig. 4.1, and
equivalent bolus as discussed in Section 6. This, in with data for 60CO radiation from Cohen et ai . (1972),
fact, is the preferred method for use with x rays this becomes
generated by potentials of 200-400 kV, but for higher
energy rays, the skin sparing advantage would be F _ T(4, 10) _ 0.938 _
C - T(7, 10) - 0.825 - 1.137
lost. For such radiations, it is frequently desirable to
alter the incident beam by the imposition of a com-
pensating filter which approximates the role of the 5
bolus and is also discussed in Section 6.
Under some circumstances it is also permissible, or
even advantageous, to accept the beam as it is and,
by calculation, to determine the resultant actual dose
distribution. The most convenient manual procedure
for doing this involves starting with the basic isodose
data and calculating correction factors. The following
three methods produce satisfactory results for angles
of incidence of up to about 45° for 6OCO gamma rays or
high-energy x rays, and of up to about 30° from the
surface normal for 200-400 k V x rays.

Fig. 4.1 Diagram showing correction of isodose curves under


4.1.1 The Tissue-Air Ratio Method a sloping and curved surface such as S-S. The solid isodose curves
are from an isodose chart for which the phantom surface was
This method can be explained with the help of Fig. located at S'-S'. They would apply if the space between S-S and
4.1. This diagram shows isodose curves taken in a S'-S' were filled in with tissue-like material but, because of the
plane containing the central ray of the beam. The air gap, the absorbed dose at points such as B is increased. The
line S"-S" and the dashed isodose lines represent the isodose chart
solid line isodose curves represent the basic data that moved down a distance h . The relative dose at B can be read off
were obtained in a phantom with its surface coincid- and multiplied by an inverse square factor to make allowance for
ing with the line S' -S'. The line S-S represents the the distance change h .
19

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
20 . . . 4. Basic Data and Patient Data

If the uncorrected relative absorbed dose at point B is s


68%, the corrected value becomes
s'- - - - - - - - - - s'
P'B = P B x CF = 68.0 x 1.137 = 77.3%
This method takes both the beam dimensions and the
depth of concern into account and is essentially an
application of a method known as the "effective atten-
uation coefficient method" (ICRU, 1963). It can be
applied equally well for fixed SSD and isocentric s
techniques.

4.1.2 The Effective SSD Method

An alternative method, which also allows for beam Fig. 4.2. Diagram showing the "isodose shift" method of con-
dimensions and depth, is as follows. Consider again structing isodose curves under a sloping surface. The isodose
Fig. 4.1. and the two points A and B. The dose to chart is moved down a distance k x h along lines such as that
point A, as before, will not be altered. The air gap through point B and the intersections of the isodose curves with
(tissue-deficit) above point B is h cm and the tissue the lines noted to construct the new isodose curves. The beam is
the same as that of Fig. 4.1.
thickness above it is (d - h). The absorbed dose, at
the point of maximum, at depth d m , along the ray line taken to be 2/3 and this is used in the example of Fig.
through B, will be altered by an inverse square rela- 4.2. The relative dose estimated at point B by this
tion. If we also note that the percentage depth dose method is 78%, which again is in fairly good agree-
does not change rapidly with distance from the source ment with the above two methods.
(provided that this distance is not less than 50 cm or The factor k is different for each radiation quality
so), the relative dose at point B can be obtained by and, strictly speaking, for each beam size, depth of
sliding down the isodose chart the distance h, so that interest and SSD. Satisfactory values for a few radia-
its surface is at S' -S', reading off the new value and tion energies for commonly used conditions are
multiplying it by the inverse square law factor to give given in Table 4.1. A discussion of these values is
the corrected percentage depth dose value: given by van der Giessen (1973). The method is fast
and useful, but less accurate than either of the two
P'
B
= p' x ( s+dm
B s+h+dm
)2 (4.2) previously discussed. It should be used with great
caution for points near the surface.
With s = 80 cm (the distance from the source to the The method is equally applicable for both fixed
surface S'-S'), d m = 0.5 cm, h = 3 cm and pII B = 81 SSD and isocentric techniques although in this latter
(the relative dose read off the relocated isodose case the reference surface becomes a virtual surface
chart), the corrected relative dose value at point B is and will not, in general, coincide with that for which
80.5)2
the isodose curves were drawn. The procedure is il-
P'B = 81.0 x ( 83.5 = 75.3% lustrated in Fig. 4.3. The reference surface, S'-S', is
located at the point of intersection of the central ray
This compares fairly well with the previous method. of the beam with the patients contour. Doses at points
such as A on the central ray will not be changed since
they are normalized to be 100 at the isocenter, C.
4.1.3 The Isodose Shift Method Doses along a ray such as that through point B will
be altered. A good value for the corrected dose may be
The foregoing methods are convenient for making obtained by shifting the isodose chart along this ray a
point calculations. Frequently, however, one wishes distance k x h as described above. The direction of
to construct an entire isodose chart for the actual the shift may be either up or down, depending on the
contour shape, in which case the following procedure displacement between S'-S' and S-S.
can be used with quite satisfactory results. Consider Of the three methods, the tissue-air ratio method
the ray through point B in Fig. 4.2. If the isodose gives the most accurate results over a wide range of
chart is slid down along this ray by an amount k x h, depths since it does implicitly take both the beam size
where k is a factor less than 1, the altered intersec- and depth into consideration. Its use is recommended
tion of the isodose lines with this ray can be read off for the estimation of doses at points of special inter-
directly as shown. For 60CO radiation, k is usually est. The first two methods of correction lend them-

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
4.3 Effect of Body Inhomogeneities . • . 21
TABLE 4.1-Isodose shift factors for different radiation energies TABLE 4.2 -Correction factors for determination of absorbed dose
Type of radiation with lack of underlying tissue (HVL 1.5 to 2 .5 mm Cu; valid depth
5-15 em )
x rays from 150 k V to 1 MV 0.8
T hickness of Area of fi e ld /cm ~
xrays from 1 to 5 MV 0.7 underlying
0.6 t issue/em 25 50 100 150 200 400
x rays from 5 to 15 MV
x rays from 15 to 30 MV 0.5 0 0.81 0.76 0.72 0.70 0.68 0.66
x rays above 30 MV 0.4 1 .89 .86 .82 .80 .78 .75
2 .93 .91 .87 .85 .84 .82
3 .95 .93 .91 .89 .88 .85
5 .98 .97 .95 .94 .93 .92
8 1.00 1.00 1.00 .99 .99 .99

beam receives the fraction 0.91 of the dose predicted


by basic data and other corrections so far mentioned.
For 60CO gamma-ray energies and higher, the cor-
rections due to lack of backscatter can be estimated
from the reciprocal of the backscatter factor, but it is
now small, less than 5%. Frequently, the material of
the couch is sufficient to provide backscatter and, in
these cases, this correction should be ignored.

4.3 Effect of Body Inhomogeneities

The presence of tissues of different compositions or


density will cause the absorbed-dose distribution
within the patient to differ from that obtained in the
Fig. 4.3. Diagram showing the "isodose shift" method of con- water-equivalent body. The deviations caused will
structing isodose curves under a sloping surface for an isocentric vary widely, depending on the amount and type of
setup. The procedure is the same as that used in Fig. 4.2, but the non-water-equivalent material present and on the
reference surface S'-S' must be taken at the intersection of the
contour with the central ray . The shift at the central ray will thus
radiation quality. The effects may be considered as
be zero and may have positive or negative values on either side of arising from alterations in the attenuation and ab-
it. sorption of the primary beam and from the associated
pattern of scattered photons and effects arising from
selves to dose data representation by ray line coordi- alterations in the fluence of secondary electrons.
nates, since all values along a ray line would be An alteration in the attenuation of the primary
modified by the same factor. This is particularly use- component ofthe radiation beam affects the absorbed
ful in computer applications. dose at points which lie beyond the non-water-equiv-
alent region, while an alteration in the pattern of
scattered photons affects the absorbed dose at points
4.2 Effect of Patient Size which are relatively close to the non-water-equiva-
lent material, but need not be directly beyond it.
Near the exit surface of the patient there may be Alterations in the fluence of secondary electrons
insufficient material to provide full scattering condi- affect the tissues within or very close to the non-
tions. At the exit surface, provided there is a com- water-equivalent region . For x rays generated by
plete absence of back scattering material, the ab- potentials of less than 400 kV, the increased fluence
sorbed dose, obtained from the basic data, will be high of secondary electrons in bone will result in an in-
by a factor which is approximately equal to the recip- crease in the absorbed dose received by soft tissues
rocal of the backscatter factor for the beam in use within the bone, or immediately adjacent to it. The
(Legare, 1964). For points in the patient remote from actual dose may indeed be several times the uncor-
the exit surface, the correction factor approaches un- rected dose. For high energy beams in which the
ity. Table 4.2. shows corrections which should be higher effective atomic number of bone is of less
applied for radiation having HVL in the range from significance, this effect is small or virtually absent.
1.2 to 2.5 mm Cu for various thicknesses of tissue (or However, in the case of high energy radiation, there
tissue-equivalent material) below the point of inter- may be a loss of electron equilibrium in very low
est and for different field sizes. Thus, for example, a density materials or in air cavities which results in a
point 3 em above the exit point in a 10 cm x 10 em reduction in the absorbed dose received by the tissues

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
22· . . 4. Basic Data and Patient Data

just beyond the cavity and this decreased dose may Three methods of arriving at a correction factor are
extend an appreciable distance into the tissues. presented. They are arranged in order of increasing
number of parameters that are taken into account.
There is considerable similarity with the corrections
4.3.1 Methods of Correction for Effects Arising for surface shape discussed in section 4.1.
from Alterations in the Absorption of the Primary (aJ Tissue-Air Ratio Method. A correction factor
Beam and the Associated Pattern of Scattered may be obtained by using a ratio of two tissue-air
Photons ratios.

For all methods of correction discussed, it is first (4.3)


necessary to obtain the standard or basic absorbed
dose values by the application of standard depth-dose where d' is the equivalent water thickness and d is
data, isodose charts, etc., valid for water-equivalent the actual thickness of material between the point of
bodies and corrected for body shape. The correction calculation and the skin along a ray passing through
factors for inhomogeneities are adjustments made to the point. W d is the dimension of the cross section of
these values. For each method discussed, it is neces- the beam at point P.
sary to determine the water-equivalent thickness of To obtain the correction factor for the above exam-
material lying between the point of calculation and ple for 60CO y rays, it is necessary to know the beam
the beam entrance point. This is taken to be the dimensions. As an example, assume the beam to be
thickness which would attenuate the primary beam square with side Wd = 10 cm and use tissue-air ratio
by the same amount as the actual tissues of the body. data from Cohen et al. (1972) for depths of 6.5 and 10
This may be illustrated by reference to Fig. 4.4, cm to give
where a very stylized arrangement is depicted. Cal-
culation is to be made at point P, a distance d = 10 cm CF = T(6.5, 10) = .847 = 1.19
below the surface. This distance is made up of3 cm of T(10, 10) .709
material with a relative density of 1, followed by 5 cm This correction factor takes account of both the beam
of low density material. The relative density is given size and the depth of the point of calculation. It does
as 0.3 and since lung tissue differs from water equiva- not, however, take account of the distance between
lence only in density, the equivalent water thickness point P and the material with density different from
of the lung traversed is 0.3 x 5 cm = 1.5 cm. This water, nor does it take account of the lateral extent of
layer is followed by another layer (2 cm) of unit the inhomogeneity.
relative density material. The total thickness, d, is 10 The so-called "effective attenuation method" is very
cm. The equivalent water thickness d', is 3 + 1.5 + 2 similar in principle to the tissue-air ratio method.
= 6.5 cm. d' is greater than d, when a high density
However, now the correction factor, CF, would be
material such as bone is involved. obtained from CF = e~d-d'), where again d is the
actual thickness of material and d' the equivalent
water thickness. Here [J., is the effective linear atten-
uation coefficient which is empirically determined.
For example, [J. could be taken as a standard value of
0.050 cm- I (5% per cm) or obtained from the rate of
change of the tissue-air ratio for the 10 cm x 10 cm
beam which would be [J. = 0.049 cm- I for 60CO radia-
tion (obtained by plotting tissue-air ratios with re-
spect to depth on semi-logarithmic paper). In this
case
CF = eO. 049 (10-6.S) = 1.19
To increase the accuracy of this type of correction, a
"position correction factor" CP, may be introduced
such that
CF' = CF x CP
This correction has been determined experimen-
Fig. 4.4. Schematic diagram showing an inhomogeneity of
relative density 0.3 in a water equivalent phantom. Its thickness tally by a number of authors (Dutreix et al., 1960;
is 5 cm and a large lateral extent is assumed. The alteration in Massey, 1962) for 60CO gamma radiation and 4 MV x
dose. due to the inhomogeneity. is to be calculated at point P. rays in the case of the lung and by O'Connor (1957)

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
4.3 Effect of Body Inhomogeneities • . • 23
for x rays generated by potentials in the neighbor- tion, Pe is the ratio of the number of electrons per cm3
hood of 200 kV. CP is a function of the distance d l in the heterogeneity to the number of electrons per
from the point of calculation to the inhomogeneity. cm3 in soft tissue . The correction factor so derived is
The following values give an idea of the size of the independent of the thickness of water-equivalent ma-
correction and are recommended: terial overlying the non-water-equivalent material,
but it does depend on the thickness ofthe heterogene-
dt/cm o 2 5 > 10
ity and its position with respect to point P . This
CP 0.92 0.95 0.97 1.0
expression is derived theoretically assuming Comp-
In the example used here , the amended value CF' = ton interaction only. It is not valid in the buildup
1.19 x 0.95 = 1.13. For very high x ray energies, the region nor within the non-unit-density material.
correction factor becomes independent of the distance The calculated correction factor for the example of
dt . Fig. 4.4 would be
( b) The Isodose Shift Method . A method of correc-
tion, similar to the method of the same name dis- = (T(7, 10))<0.3-1) = ( .827 )-0.7 = 1.145
cussed in Section 4.1.3, was proposed by Greene and CF T(2, 10) 1.004
Stewart (1965) and Sundbom (1965c). These authors This value is slightly lower than those given by the
propose to move the isodose lines by an amount equal other methods except where the position correction
to n times the thickness of the inhomogeneity as factor was included. It can, in general , be considered
measured along a line parallel to the central axis and the most accurate , except for very high energy x rays
passing through the point of interest. The isodose where the simple tissue-air ratio method should be
lines should be moved toward the skin for bone and used.
away from the skin for lung or air cavities. The
values of n given below were determined experimen-
tally and are independent of the field size; they are 4.3.2 Effects Arising from Alterations in the Flu-
valid for 60CO radiation and 4 MV x rays and should ence of Secondary Electrons
be lower for higher energies.
Inhomogeneity
(a) Bone . For x rays generated at potentials less
than about 250 kV, the electron fluence in the soft
air cavity -0.6
lung tissue -0.4 tissues adjacent to or surrounded by bone is increased
hard bone 0.5 by the electrons arising from the photoelectric ab-
spongy bone 0.25 sorption in the mineral contents of bone. The increase
of the absorbed dose in these regions has been com-
In the previous example, a shift of - 0.4 x 5 cm = prehensively studied by Spiers (ICRU, 1963).
- 2 cm would be indicated. It would correspond to a For radiation from 60CO, the absorbed dose at the
correction factor of 64.1/55.6 = 1.15 (data taken from interface between bone and soft tissues varies be-
percentage depth dose tables, Cohen et at ., 1972, for tween 95% and 108% of the dose to soft tissues, de-
depths of 8 cm and 10 cm). pending on the direction of the photons. These varia-
(c) The Power Law Tissue-Air Ratio M ethod . tions are still smaller at higher energies.
Batho (1964) and Young and Gaylord (1970) have (b) Air Cavities . For high energy photons, elec-
shown, at least in the case of 60CO gamma rays, that tronic equilibrium may be partially lost at the sur-
tissue-air ratios may also be used to calculate a cor- faces of the cavity and the actual dose to tissue be-
rection factor which accounts for not only the magni- yond and even in front of the cavity may be apprecia-
tude and nature of the heterogeneity but also its bly lower than expected. Epp et al. (1958), for 60Co y
position, by raising a ratio of tissue-air ratios to a rays, and Nilsson and Schnell (1976), for 6OCO y rays,
power as follows; referring again to Fig. 4.4, the 6 MV and 42 MV x rays, have comprehensively stud-
correction factor for point P is ied the buildup of the absorbed dose for different
CF = (T(d 2, W d))(P,-1l
geometric conditions. The relative absorbed dose at
(4.4) the surface beyond the cavity is smaller for 60CO
T(dl> W d )
radiation than for higher energies, but the maximum
where d I is the distance from point P to the non- of the buildup curve occurs at greater depths for
water-equivalent material and d 2 is the distance from higher energies. The most important decrease of the
point P to the front (source side) of the non-water- absorbed dose at the surface is observed for the larg-
equivalent material. Wd is the beam dimension (ef- est cavities (4 cm deep) and the smallest fields (4 cm x
fective) at the depth ofP, although, since ratios only 4 cm) tested. For those conditions, the ratio of the
are used, little loss in accuracy results from taking a absorbed dose at the surface to the maximum ab-
constant beam size whatever be the depth of calcula- sorbed dose is as low as 0.64 for 6OCO y rays, 0.72 for 6

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
24 . . . 4. Basic Data and Patient Data

MV and 0.76 for 42 MV x rays. For a cavity depth of 9 ningham and Milan, 1969; and Holmes, 1970) or,
mm, this ratio is about 0.9 for the three beam quali- more recently, the "sonic digitizer". 6 These devices
ties. For larger field sizes, this ratio approaches un- could be used to take the contour directly from the
ity. A slight decrease of the dose at the surface of the patient, thus eliminating the step of producing the
cavity on the side of the source is also observed owing diagram on paper. The advisability of this, however,
to a lack of back scattered electrons. This decrease of is not established and it is likely that the operators
the dose is less important than that beyond the cavity interaction with the paper diagram during the proc-
and it occurs in a smaller thickness of tissue. ess of radiation treatment planning is valuable.
(c) Lung Tissue. In a similar way, on account of Tomograms and other radiographs can be read and
the low density of the lung, a loss of secondary elec- digitized for computer input with a curve follower on
trons is observed in the first layers of soft tissue a light table and the magnification corrections and
beyond a large thickness of lung, when it is irradi- other desired processing can be performed directly by
ated by high energy photons, especially with a nar- the computer. At the present time, however, such
row beam. In common clinical conditions, this de- equipment is not widespread and so, in many cases,
crease of the absorbed dose could be as great as five the shapes, the location and nature of internal struc-
percent, but is normally ignored (Leung et at., 1970). tures and inhomogeneities are entered first onto a
paper diagram and then digitized for the computer.
As for manual calculations, it is most commonly as-
4.4 Input of Data for Manual and Computer Calcu- sumed that the two-dimensional description of the
lations contour, in the plane of calculation, is sufficient. The
use of the computer, nevertheless, makes it possible
4.4.1 Manual Calculations to consider an extension to three dimensional treat-
ment planning.
The basic data required for manual calculation of a Computer calculations will, in general, be more
dose distribution are tables of depth-dose data and/or accurate than hand calculations. However, they
tissue-air ratios, a set of isodose charts for the treat- should always be at least spot-checked by a manual
ment machine being used and a tracing, on a piece of calculation. If there is significant disagreement, both
paper, of the contour of the patient. This contour is should be questioned.
normally taken in the plane that will contain the An important distinction should be made between
central rays of the radiation beams that will be used . computer programs which carry out absorbed dose
Methods for obtaining this contour are discussed in calculations and those which are called treatment
section 3.2.1. If corrections for tissue heterogeneities planning programs. The former carry out calcula-
are required, their locations and shapes must be tions after the beam positions have been decided.
added to the diagram of the contour. Manual calcula- They form a set of tools with which an operator can do
tions will normally be restricted to the plane contain- treatment planning. The latter attempts to take over
ing the central ray. It is also usually assumed that part of the decision-making process and either them-
the contour shape in other planes has no influence on selves arrive at values for beam settings, or more or
the dosage pattern in this plane. With this assump- less directly assist the operator in doing so. The core
tion, information on other planes is not required. of both are the absorbed dose calculations and it is
This is an assumption, however, that one should be essential that these be correct.
conscious of and one which, at times, should be made
with caution.
4.5 Use of Anatomical Phantoms for
Checking Procedures
4.4.2 Computer Calculations
Water tanks are routinely used as phantoms for
The isodose chart, used in manual calculations, is measurement of basic radiation data. It can also be
essentially an analogue device and is not particularly useful to measure radiation absorbed in or transmit-
suited to use with computers. It is more convenient ted through anatomical phantoms. In these phan-
either to store the beam data in digital form as a toms the lungs should have the same atomic composi-
matrix of numbers or to generate the beam data as it tion as water, but a lower density. Sawdust or cork,
is required. The diagram of the patients' contour is after a careful check of density, can be used. Phantom
also analogue and this too must be entered in digital material representing bone is much more difficult to
form. The digitization can take place manually, point
by point, or by digitizing devices such as the "rho- " An example is the GRAF/PEN produced by Science Accesso-
theta" device used in the Programmed Console (Cun- ries Corp. of Southport, Connecticut, U.S.A.

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
4.5 Anatomical Phantoms . . . 25
constitute although some plastics loaded with min- a treatment procedure or a new method of calcula-
eral compounds have been proposed (Surmont and tion, for instance to take account of body heterogenei-
Gest, 1957). The best bone phantoms are made of real ties. They can be used to check the accuracy of a new
bones, dried and injected with paraffin or plastic in technique such as use of simulators, transverse tom-
vacuum. Dry skeleton pieces cannot be used directly. ography or ultrasonic procedures. However, they can-
Realistic phantoms including skeleton and lung in- not be recommended for dosimetry of actual patients
serts are now commercially available although very because the variations in anatomy among patients
expensive. These are useful when one wants to check are far too great.

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
6.2 Beam Compensating Devices • • • 39
set-up, that is, positioning the patient on the couch No bolus Bolus Compensated
outside the treatment room, prior to each session, in - - -~-- ---t---
, high I
order to conserve time within the treatment room. density
This mode of use must be regarded as still in the
experimental stage.
I, I,
I I
6.2 Beam Compensating Devices

6.2.1 Bolus

The effect of the shape of the patient's surface in


distorting the dose distribution was discussed in sec-
tion 4.1. Fig. 6.1(a) shows, schematically, isodose
curves under a sloping surface. The dotted region (a) (b) (e) (d)

represents a target volume and it can be seen that the Fig. 6.1. Schematic diagrams illustrating the use of bolus
absorbed dose is not uniform across it. This distortion and compensating filters (From Johns, H. E. and Cunningham,
J. R. THE PHYSICS OF RADIOLOGY 3rd Ed., 1969. Courtesy
can be avoided by the use of tissue-equivalent bolus
of Charles C Thomas, Publisher, Springfield, Illinois.)
material placed on the surface so that it fills in the air
gap and presents a plane surface to the beam. This is lateral reduction and appreciable errors may result.
shown in Fig. 6.l(b). The use of such material is the A compensating filter may be used to allow for the
preferred method for use with x rays generated by effects of: (i) beam obliquity; (ii) curvature of the
potentials of 200-400 k V, but for higher energy rays entry surface of the body; (iii) body heterogeneity;
the skin sparing advantage would be lost. For such (iv) in the case of large, irregularly shaped beams,
radiations it is frequently desirable to alter the inci- variations in local scatter and SSD. A given filter
dent beam by the imposition of a compensating filter may be designed to allow for one or more of these
which approximates the role of the bolus. There are effects, the most common application being for a com-
some circumstances when the maximum dose is, in bination of beam obliquity and body curvature. Com-
fact, desired at the surface, even when high energy pensation may be thought of as an alternative to the
radiation is used. In these cases it is necessary to add calculation procedures (section 4.1) which correct the
a layer of bolus material to the surface. Its thickness dose distribution for obliquity and curvature. Insofar
should be equal to the electron buildup distance. The as correction procedures are readily available, partic-
use of bolus is easiest and most common with fixed ularly as part of an overall calculation method car-
SSD set-ups although it is also possible when isocen- ried out with the aid of a computer, the use of com-
tric techniques are used. pensators may be regarded as redundant. It must be
borne in mind, however, that most calculation meth-
ods are essentially two-dimensional, the body being
6.2.2 Compensating Filters assumed to be a cylinder or quasi-cylinder, although
in several regions of the body (e.g., the neck andjawl
In order to preserve the skin sparing properties of the surface curvature involves steep slopes in three
high energy radiation beams, and yet have the effect dimensions. Compensation may frequently be a bet-
produced by the bolus, it is convenient to place, away ter solution. On the other hand, exact compensation
from the patient's skin, a compensating filter so de- can be achieved only for the dose at a single depth,
signed that it duplicates, as nearly as possible, the whereas (in those situations where calculation is ap-
role of the bolus material. This is illustrated in Fig. plicable), the dose may be corrected at all depths. It
6.1(c). For convenience, the compensator is fre- must also be borne in mind that a dose distribution,
quently made of heavy material and it should be at although fully corrected, may still be unsatisfactory
least 15 cm or so away from the skin. Its dimensions and compensation may be the only method of provid-
therefore must be reduced (compared to the bolus) in ing an acceptable treatment plan. An individual case
the lateral directions to allow for the diverging rays may even require both calculated corrections and
and, in thickness, to allow for the higher attenuation compensation (Leung et al., 1974).
in the filter. This is shown in Fig. 6.1(d). The compensating filter method is constrained by
Compensators are not easily applicable to isocen- the fact that compensation can be made only for
tric techniques because the SSD is not fixed while the insufficient attenuation in body tissues. For this rea-
location for the compensator usually is. It is thus not son, when a compensating filter is used, the "nominal
possible to maintain a fixed geometrical ratio for the surface" of the body, for purposes of applying normal

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
40· . . 6. Planning and Delil/ery of Radiation Therapy

isodose curves, must be taken, as in Fig. 6.1, at the radiation and field size, it is impossible to design a
highest point of the surface. filter which will make absorbed dose compensation
An analogous problem occurs when compensation exact at all depths in the patient. The filter is there-
is attempted for body heterogeneity, since the atten- fore designed either to minimize the absorbed dose
uation of bone is greater than that of soft tissue to errors in the radiation field as a whole (van de Geijn,
which the normal isodose curves apply. Ellis and 1965) or to achieve exact compensation at a selected
Lescrenier (1973) designed compensators of minimum depth. Most authors have chosen the second ap-
thickness along those rays which include bone, the proach, and are prepared to accept the errors occur-
thickness along other rays being sufficient to equal- ing at depths other than the "compensation depth".
ize the transmission throughout the body. Thus the These errors are usually small for points within a few
compensator "matches" the body, not to a "standard" centimeters of the "compensation depth" but may be-
phantom (density 1 g/cm 3 ) but to a phantom of in- come appreciable at more distant points. For exam-
creased density (or of equivalent increased thickness) ple, Wilks and Casebow (1969) investigated the use of
corresponding to the pathway of maximum attenua- lead compensators in cobalt-60 therapy and showed
tion. The method therefore involves a considerable that, for a compensation depth of 5 cm below the
reduction in the useful beam intensity. nominal surface, the error is within ±3% for depths
A compensating filter is intended, in the first in- up to 8 cm but can be -10% at 15 cm depth (data for
stance, to provide the primary beam attenuation 12 cm x 12 cm field at 100 cm SSD).
which would otherwise occur in the "missing" tissue Compensating filters may be constructed of wax,
when the body surface is curved and/or the incident acrylic material, or metal. Wax compensators (Cohen
beam is oblique (Fig. 6.1). Such a filter must, there- et at., 1960) may be molded directly on the skin of a
fore, take account of: (a) the relative mass attenua- patient, or on a plaster cast, but this method makes
tion coefficients of the filter material and soft tissue no allowance for beam divergence. The errors result-
(or water), for the radiation concerned; (b) the geo- ing from the technique have been investigated by
metrical change in lateral dimensions needed when Jackson (1970), who concluded that the errors are
the attenuating material is moved nearer the source acceptable for small field sizes and retraction dis-
of a divergent radiation beam; (c) the reduction in tances, i.e., distance between the filter and the skin;
build-up in the skin if the filter is placed near to the however, the retraction distance is dictated by the
surface: this reduction is negligible for 60CO 'Y rays need to preserve skin-sparing and must be at least 15
and other high energy radiation, if the compensating cm if no reduction in skin sparing is to be seen. The
filter is separated from the skin by at least 15 cm. fabrication of wax or plastic compensators, with al-
A compensating filter which allows for primary lowance for beam divergence, has been described by
beam attenuation only will overcompensate, i.e., the Beck et at. (1971), Boge et at. (1974), and Khan et al.
dose at a given point in the patient will be lower than (1968, 1970). The latter authors used a mixture of an
that indicated by "standard" isodose curves, owing to acrylic plastic and glass bubble beads, the density of
the absence of scatter reaching the point from the the final solid being controlled by the proportions of
'missing' tissue. The attenuation of the filter must the mixture. In this way the effective attenuation
therefore be less than that required for primary ra- coefficient of the filter could be varied to suit the
diation only. This problem has been treated theoreti- needs of the individual case.
cally for plane parallel layers of "missing" tissue Compensating filters may also be fabricated in
(Sundbom, 1964; van de Geijn, 1965; Khan et al., metal. The original filters described by Ellis et at.
1970) but the more complex practical situation, in- (1959) were constructed of aluminum alloy blocks, the
volving body curvature and beam obliquity, is best cross-section of each block being such as to allow for
dealt with experimentally by direct measurement of beam divergence. Brass has been advocated by Khan
the relationship between the compensating and et at. (1970) since this material gives minimum elec-
'missing' material. Such experimental studies have tron contamination at the skin for a given distance.
been carried out by the authors already cited and by Probably the most useful material of all is lead (Wilks
several other workers (Ellis et at., 1959; Cohen et at., and Casebow, 1969); such filters are laminated and
1960; Hall and Oliver, 1961; Wilks and Casebow, sheet lead is well suited to semi-automatic mechani-
1969). cal design techniques.
The contribution of scattered radiation to the dose A user adopting a system for compensating filters
at any point in a phantom depends on the type and is strongly urged to make a few experimental com-
energy of the radiation, the field size and the depth parisons of predicted doses with careful measure-
and position of the point. It follows that a compensat- ments.
ing filter must take account of all these factors, in Compensating filters were originally developed for
addition to those mentioned above in relation to the fixed SSD techniques. They may, however, be used
primary radiation. Furthermore, even for a given for isocentric techniques provided that a separate

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.
6.4 Patient and Radiation Beams • • • 41

filter is constructed and used (if required) for each ning. The computer is given a series of constraints,
beam: since the SSD is likely to change with each specifying upper and lower limits of doses at chosen
field while the position of the filter remains fixed, locations and is asked to find the best beam arrange-
both the lateral correction for beam divergence and ment to meet these criteria. Hope and Orr (1965)
the "compensation depth," i.e., the depth of the iso- pioneered this subject and assigned "score functions"
center, will be individual to each beam. A design and or relative importance to such criteria as dose uni-
fabrication technique is therefore essential which al- formity and integral dose as well as dose maxima and
lows for variation in both the divergence and the minima at specified locations. A number of other
effective attenuation coefficient; several of the meth- workers have used linear programming techniques
ods described in the literature can provide this facil- (Bahr et aZ., 1970; Jameson and Trevelyan, 1969;
ity. Bourgat et aZ., 1974) to arrive at choices of beam
Other AppZications of Compensation Filters. As arrangements. The main difficulty encountered in
already indicated, filters can also be designed to com- satisfactory achievement of optimization is with the
pensate for the heterogeneity of the human body, in choice of input criteria. Certainly a consideration of
particular the presence of bone and lung. Ellis (1960) radiation dose alone is not enough as dose fractiona-
and Ellis et aZ. (1964) proposed a system of compensa- tion is known to be important. Calculation of doses
tion based on a knowledge of the cross-sectional com- with allowance for tissue inhomogeneities has al-
position of the body obtained by transaxial tomogra- ready been discussed (section 4.3) and shown to be
phy. More recently Ellis and Lescrenier (1973) have difficult in itself although such considerations should
described compensators which equalize the transmis- certainly be included in optimization procedures.
sion through a body section, on the basis of measure- Very large and well controlled statistical studies,
ments with photographic film. This method, which, especially for the purpose, would be required to eval-
corrects for body contours as well as for heterogene- uate the criteria systematically. Although there is
ity, is used in practical dosimetry only for irradia- much current interest in the topic of computer opti-
tions with parallel opposed beams. mization, it has not as yet been satisfactorily carried
A further application of compensating filters is to out.
allow for the increase in SSD which occurs towards
the edges of a large field. In addition, if the field is ir-
regular, with narrow "limbs" protruding from a cen- 6.4 Setting-Up of the Patient and the Radiation
tral area, there will be a local decrease in dose when- Beams
ever the scattered radiation is less than in the main
part of the field. These irregularities may also be The accurate planning of a radiation treatment
associated with dips or other changes in the body should be followed by corresponding accuracy in the
surface. An example of such a filter, for use in the setting-up of the patient and the radiation beam.
treatment of medulloblastoma by orthovoltage x Accurate calculation of the dose distribution is possi-
rays, was described by Bottrill et aZ. (1965). More ble and pertinent only if the patient is irradiated in a
recently, Leung et aZ. (1974) have discussed the de- well defined position. This is usually the supine posi-
sign of such filters for use in high energy radiother- tion since patient movement during irradiation is
apy, for example, the mantle technique for the treat- less probable when the patient is comfortable.
ment of Hodgkin's Disease. In this method the com- The treatment couch should allow irradiation from
pensator is designed on the basis of a calculated dose any direction around the patient. Most modern treat-
distribution in the absence of a compensator. ment couches are designed to make this possible
Svahn-Tapper and Landberg (1971) reported on an without also losing the skin sparing effect or causing
examination of dose uniformity in mantle treatments beam attenuation through heavy supporting struc-
of patients with Hodgkin's disease. They found that tures of the couch.
filters were not required to compensate for patient Various techniques have been developed to guar-
shape in the lateral direction but that they were antee accurate positioning of the patient on the couch
useful in the craniocaudal direction of the sagittal and to restrict movement as much as possible during
section. Dose calculations were verified with in vivo irradiation. The most convenient device is a cast
measurements (see Sec. 6.5). which is made to fit the individual patient. Casts are
made of plaster of Paris, plastics or other material
and can be supplied with beam alignment indicators
6.3 Methods of Optimization and portals to ensure the desired locations of the
entrance of the beams.
It should be possible, in principle, to program a When the beam is collimated by an adjustable
computer so as to optimize choices of radiation beam diaphragm, beam position is usually checked by
arrangements as part of radiation treatment plan- means of a light localizer. The accuracy of this system

ICRU 2015 -- All rights reserved.


AAPM Member Copy
Single use only, copying and networking prohibited.

You might also like