2.1.4. Sieves EUROPEAN PHARMACOPOEIA 11.

The lamp used should be capable of revealing without doubt the material of the sieve and the substance being sifted.
a standard spot of sodium salicylate about 5 mm in diameter Degree of comminution is prescribed in the monograph
on a TLC silica gel G plate R or a standard band of sodium using the sieve number, which is the size of the mesh in
salicylate on a TLC or HPTLC plate, the spot or band being micrometres, given in parenthesis after the name of the
examined while in a position normal to the radiation. substance (Table 2.1.4.-1).
For this purpose apply 5 μL of a 0.4 g/L solution of sodium Maximum tolerance(4) for an aperture (+ X) : no aperture size
salicylate R in ethanol (96 per cent) R(3) for lamps of maximum shall exceed the nominal size by more than X, where :
output at 254 nm and 5 μL of a 2 g/L solution of sodium
salicylate R in ethanol (96 per cent) R(3) for lamps of maximum 2 ( w0.75)
output at 365/366. X= + 4 ( w0.25)
3
The distance between the lamp and the chromatographic
w = width of aperture.
plate under examination used in a pharmacopoeial test never
exceeds the distance used to carry out the above-mentioned Tolerance for mean aperture (± Y) : the average aperture size
performance test. shall not depart from the nominal size by more than ± Y,
where :

w0.98
Y= + 1.6
27
01/2008:20104 Intermediary tolerance (+ Z) : not more than 6 per cent of the
total number of apertures shall have sizes between “nominal
+ X” and “nominal + Z”, where :

X+Y
Z=
2
2.1.4. SIEVES Wire diameter d : the wire diameters given in Table 2.1.4.-1
apply to woven metal wire cloth mounted in a frame. The
Sieves are constructed of suitable materials with square nominal sizes of the wire diameters may depart from these
meshes. For purposes other than analytical procedures, sieves values within the limits dmax and dmin. The limits define a
with circular meshes may be used, the internal diameters of permissible range of choice ± 15 per cent of the recommended
which are 1.25 times the aperture of the square mesh of the nominal dimensions. The wires in a test sieve shall be of a
corresponding sieve size. There must be no reaction between similar diameter in warp and weft directions.
Table 2.1.4.-1 (values in micrometers)
Sieve Tolerances for apertures Wire diameters
numbers
Maximum Tolerance for Intermediary Recommended Admissible limits
(Nominal
tolerance for mean aperture tolerance nominal
dimensions of
an aperture dimensions
apertures)
+X ±Y +Z d dmax dmin
11 200 770 350 560 2500 2900 2100
8000 600 250 430 2000 2300 1700
5600 470 180 320 1600 1900 1300
4000 370 130 250 1400 1700 1200
2800 290 90 190 1120 1300 950
2000 230 70 150 900 1040 770
1400 180 50 110 710 820 600
1000 140 30 90 560 640 480
710 112 25 69 450 520 380
500 89 18 54 315 360 270
355 72 13 43 224 260 190
250 58 9.9 34 160 190 130
180 47 7.6 27 125 150 106
125 38 5.8 22 90 104 77
90 32 4.6 18 63 72 54
63 26 3.7 15 45 52 38
45 22 3.1 13 32 37 27
38 – – – 30 35 24

(3) The ethanol (96 per cent) R used must be free from fluorescence.
(4) See the International Standard ISO 3310/1 (1975).