WATO EX-65 Pro

Anesthesia Workstation

Physical Specifications
Dimensions and Weight
Height 1370 mm
Width 780 mm (not including breathing system)
945 mm (including breathing system)
Depth 690 mm 0, 3~60 cmH2O (CPAP/PS)
Weight <145 kg Rate 4~100 bpm (increments of 1 bpm)
(without vaporizers and cylinders) I:E 4:1 - 1:8 (increments of 0.5)
Inspiratory pause (Tip:Ti) OFF, 5% - 60% (increments of 1%)
Top Shelf Inspiratory time (Tinsp) 0.2 - 10.0 s (increments of 0.1 s)
Weight limit 30 kg Trigger window 5% - 90% (increments of 5%)
Width 305 mm Flow trigger 0.2 ~ 15 L/min (increments of 0.1L/min)
Length 545 mm Pressure trigger -20~ -1 cmH2O (increments of 1 cmH2O)
Expiration termination level 5% - 80% (increments of 1%)
Work Surface Minimum Rate 2 - 60 bpm (increments of 1 bpm)
Height 850 mm Tslope 0.0 - 2.0 s (increments of 0.1 s)
Area 1635 cm2 Apnea I: E 4:1~1:8 (increments of 0.5)
∆Papnea 3 - 60 cmH2O (increments of 1 cmH2O)
Drawer (3Xdrawers, Internal Dimension)
Height 130 mm Positive End Expiratory Pressure (PEEP)
Width 415 mm Type Integrated, electronic controlled
Depth 320 mm Range OFF, 3~30 cmH2O
(increments of 1 cm H2O)
Bag Arm
Height 1150 mm Ventilator Performance
Length 312 mm Driving pressure 280 kPa to 600 kPa
Connection ISO 22mm OD, 15mm ID Peak gas flow 120 L/min + Fresh Gas Flow

Casters Monitoring Parameters


Diameter 125 mm Minute volume 0 ~ 100 L/min
Brakes Centre brake system with Lock / Unlock Tidal volume 0~3000 ml
icons Inspired oxygen (FiO2) 18% ~ 100%
Peak airway pressure -20 ~ 120 cmH2O
Ventilator Specifications Mean pressure -20 ~ 120 cmH2O
Modes of Ventilation Plateau pressure -20 ~ 120 cmH2O
Manual/Spontaneous Ventilation/Bypass I:E 4:1 ~ 1:10
Volume Control Ventilation (VCV) with PLV function Rate 0 ~120 bpm
Pressure Control Ventilation (PCV) with/without volume guarantee PEEP 0 ~ 70 cmH2O
(VG) Resistance (R) 0 ~ 600 cmH2O/(L/s)
Synchronized Intermittent Mandatory Ventilation Compliance (C) 0 ~ 300 ml/cmH2O
(SIMV-Volume Controlled and SIMV-Pressure Controlled)
Pressure Support Ventilation (PS) with apnea backup Control Accuracy
Synchronized Intermittent Mandatory Ventilation Volume Guarantee Volume delivery 20 mL to 60 mL: ±10 mL
(SIMV-VG) 60 mL to 210 mL: ±15 mL
Continuous Positive Airway Pressure/Pressure Support Ventilation 210 mL to 1500 mL: ± 7% of the set value
(CPAP/PS) Pinsp ± 2.5 cmH2O or ± 7% of the set value,
whichever is greater
Compensation Plimit ± 2.5 cmH2O or ± 7% of the set value,
Circuit gas leakage compensation and automatic compliance whichever is greater
compensation ∆Psupp ± 2.5 cmH2O or ± 7% of the set value,
whichever is greater
Ventilation Parameters Range ∆Papnea ± 2.5 cmH2O or ± 7% of the set value,
Patient Size Adult, Pediatric, Infant whichever is greater
Tidal volume 20~1500 mL ( (Volume Mode) PEEP OFF: ± 3.0 cmH2O
(increments of 1 mL) 3 to 30 cmH2O: ± 2.0 cmH2O, or ± 8% of the
5~1500 mL (Pressure Mode) set value, whichever is greater
Pinsp 5~70 cmH2O (increments of 1 cmH2O) Rate ± 1bpm or ± 10% of the set value,
Plimit 10~100 cmH2O (increments of 1 cmH2O) whichever is greater
∆Psupp 3~60 cmH2O (increments of 1 cmH2O) I:E 2:1 to 1:4: ± 10% of the set value
Other range: ± 25% of the set value Screen display configurable
Tip:Ti ± 8% Display parameters All setting and alarm parameters (including
Tinsp ± 0.2s Breath rate, I/E ratio, Tidal volume, Minute
Trigger Window ± 10% volume, PEEP, MEAN, PEAK, PLAT, and O2
Flow Trigger ± 1L/min concentration, EtCO2, N2O, Aesthesia gas
Pressure Trigger ± 2cmH2O concentration, BIS)
Exp% ± 10% Display waveforms P-T, F-T, V-T, CO2 , BIS, O2, Anesthetic gas, N2O
Spirometry loops P-V, F-V and F-P
Monitoring Accuracy Timer On screen timer
Volume monitoring 0 mL to 60 mL: ± 10 mL
60 mL to 210 mL: ± 15 mL Communication Ports
210 mL to 1500 mL: ± 7% of the set value One RS-232C connector and one DB9 connector
Pressure monitoring ± 2.0 cmH2O or ± 4% of the reading, Ethernet (RJ-45)
whichever is greater USB
Rate ± 1bpm or ± 5% of the reading, whichever is VGA
greater
I:E 2:1 to 1:4: ± 10% of the reading Vaporizers
Other range: ± 25% of the reading Vaporizer Mindray V60 Anesthetic Vaporizer or Penlon
MV ± 0.1L/min or ± 8% of the reading, Sigma Delta Anesthetic Vaporizer
whichever is greater Support agents Halothane, Enflurane, Isoflurane,
Sevoflurane
Trend Graph Position MAX.2
Continuous trend information with time discrete events for the latest Mounting mode Selectatec®, with interlocking function
48 hours Plug-in®, with interlocking function

Trend Table Modules


Continuous trend information together with time discrete events for Anesthesia Gas (AG) Module
the latest 48 hours Monitor gases CO2, N2O, Halothane, Enflurane, Isoflurane,
Sevoflurane, Desflurane, MAC,
Alarm Log Book Paramagnetic O2 (optional)
500 events storage, first in first out Warm-up time 45 s (ISO accuracy mode)
10min (full accuracy mode)
Alarm setting Sample rate Adu/Ped: 150, 180, 200 ml/min
Tidal volume Low: 0 ~ 1595 ml Neo: 100, 110, 120 ml/min
High: 5 ~ 1600 ml Accuracy ± 10 mL/min or ± 10% of the set value,
Minute volume Low: 0 ~ 99 L/min whichever is greater
High: 0.2 ~ 100 L/min Range CΟ2: 0% ~ 10%
Inspired oxygen Low: 18% ~ 98% Des: 0% ~ 18 %
High: OFF, 20% ~ 100% Sev: 0% ~ 8%
Apnea alarm VTe < 10ml measured in 20s Enf, Iso, Hal: 0% ~ 5%
Paw < (PEEP + 3) cmH2O in 20s O2/N2O: 0% ~ 100%
Airway pressure low 0 ~98 cmH2O
Airway pressure high 2 ~100 cmH2O Carbon Dioxide (CO2) Modules
Sustained airway pressure alarm: 15s Method Infrared absorption
Subatmospheric pressure alarm: Paw < -10 cmH2O Module type Mindray side-stream
Alarm silence countdown timer: 120 to 0 seconds Capnostat mainstream
Oridion micro-stream
Lung Recruitment Tool (optional)
Increasing PEEP progressively (with a maximum of 7 stages) Work mode Standby or measurement
Displayed numerics EtCO2 , FiCO2
Ventilator Components Waveform Capnography
Flow Sensor
Type Variable orifice flow sensor Side-Stream Carbon Dioxide (CO2) Module
Location Inspiratory and expiratory port Measurement range 0 ~ 99 mmHg
Accuracy ± 2 mmHg (0 ~ 40 mmHg)
Oxygen Sensor ± 5% of the reading (41 ~ 76 mmHg)
Type Galvanic fuel cell ± 10% of the reading (77 ~ 99 mmHg)
FiO2 displayed 18% to 100% Resolution 1 mmHg
Accuracy ± (volume fraction of 2.5 % +2.5 % gas level) Sampling rate Neonatal: 100 mL/min and 120 mL/min
Response Time ≤20 seconds optional
Adult/children: 120 mL/min and 150
Ventilator Screen mL/min optional
Display type Color active matrix TFT touch screen Sampling rate accuracy: ± 15% of the set value or ± 15 mL/min,
Display size 15 in diagonal whichever is greater
Pixel format 1024 x 768 Warming-up time < 1 min, enter the ISO accuracy mode
Brightness Adjustable After 1 min, enters the full accuracy mode
Response time <4.5 s@100 mL/min whichever is greater
<4.5 s@120 mL/min
Measured by using neonatal watertrap and Electrical Specifications
2.5 m neonatal sampling line Current Leakage
<5.5 s@120 mL/min 100 ~ 240V < 500 μA
<5 s@150 mL/min
Measured by using adult watertrap and Power And Battery Backup
2.5 m adult sampling line Power input 220-240 Vac, 50/60 Hz, 6A
100-120 Vac, 50/60 Hz, 7A
Capnostat Mainstream CO2 Module 100-240 Vac, 50/60 Hz, 7A
Measurement range 0 ~ 150 mmHg Auxiliary electrical outlets
Accuracy ± 2 mmHg (0 ~ 40 mmHg) Up to 4 outlets (3A for each, total 5A)
± 5% of the reading (41 ~ 70 mmHg) Battery backup 150 minutes for 2 standard pieces of battery
± 8% of the reading (71 ~ 100 mmHg) (powered by new fully-charged batteries
± 10% of the reading (101 ~ 150 mmHg) with 25°C ambient temperature)
Resolution 1 mmHg Battery type Build-in Li-ion battery, 11.1 VDC, 4500 mAh
Rise time <60 ms Safety feature In case of electricity and battery failure,
Response time <2 s manual ventilation, gas delivery and agent
Alarm limit EtCO2 High: OFF, 2 ~ 150 mmHg delivery are possible
EtCO2 Low: OFF, 0 ~ 148 mmHg
FiCO2 High: OFF, 1 ~ 150 mmHg Pneumatic Specifications
ACGO (Auxiliary Common Gas Outlet, Integrated)
Micro-stream CO2 Module Connector ISO 22 mm OD and 15 mm ID
Measurement range 0 ~ 99 mmHg
Accuracy 0 ~ 38 mmHg: ± 2 mmHg Pipeline Supply
39 ~ 99 mmHg: ± (5 % of the reading + 0.08 % Gas type O2, N2O and Air
of (the reading minus 38 mmHg)) Pipeline input range 280 to 600 kPa
Sampling rate 50 ml/min Pipeline connections DISS or NIST
Sampling accuracy -7.5 ml/min ~ + 15 ml/min
Initialization time 30s Pipeline Supply Pressure Gauges
Response time 2.9s Display type Electronical or Mechanical
Rising time < 190 ms Ranges 0 to 1000kPa
Delay time 2.7s Accuracy ± (4% of the full scale reading + 8% of the
Alarm range EtCO2 High: OFF, 2 ~ 99 mmHg actual reading)
EtCO2 Low: OFF, 0 ~ 97 mmHg
FiCO2 High: OFF, 1 ~ 99 mmHg Cylinder Supply
Cylinder Supply E Cylinder (American style or UK style)
BIS Module O2 Input Range 6.9 to 20 MPa
Measured parameters EEG N2O Input Range 4.2 to 6 MPa
BIS/BIS L, BIS R 0 ~ 100 Air Input Range 6.9 to 20 MPa
Sweep speed 6.25 mm/s,12.5 mm/s, 25 mm/s or 50 mm/s Cylinder Connections Pin-Index Safety System (PISS)
Input impedance > 50 Mohm Yoke Configuration O2, N2O, Air
Noise (RTI) < 0.3 uV (0.25 ~ 50 Hz)
Input signal range ± 1 mv Cylinder Supply Pressure Gauges
EEG bandwidth 0.25 ~ 100 Hz Display type Electronical or Mechanical
Patient leakage current < 10 uA Air Range 0 to 25 MPa
Alarm limit BIS high: 2 ~ 100 O2 Range 0 to 25 MPa
BIS low: 0 ~ 98 N2O Range 0 to 10 MPa
Calculated parameters SQI/SQI L, SQI R; EMG/EMG L, EMG R; SR/SR Accuracy ± (4% of the full scale reading+8% of the
L, SR R; SEF/SEF L, SEF R; TP/TP L, TP R; BC/BC actual reading)
L, BC R; sBIS L, sBIS R; sEMG L, sEMG R; ASYM
Impedance range 0 ~ 999 Kohm O2 Controls
Method N2O shut off with loss of O2 pressure
Agent Consumption Calculation Supply failure alarm ≤ 220.6 kPa
Calculation range 0 to 3000 ml O2 Flush 25 ~ 75 L/min
Accuracy ± 2 mL, or ± 15% of the reading, whichever
is larger Auxiliary O2 Flowmeter
Range 0 ~ 15 L/min
Agent consumption speed Indicator Flow tube
Anesthetic agents Desflurane, Enflurane, Isoflurane,
Sevoflurane and Halothane Electronic Flow control system (Electronic Mixer)
Consumption speed Desflurane: 0 ~ 900 ml/h Direct Flow Control Mode
Sevoflurane: 0 ~ 450 ml/h O2 flow range 0 to 15 L/min
Enflurane, Isoflurane and Halothane: 0 ~ 250 Air flow range 0 to 15 L/min
ml/h N2O flow range 0 to 12 L/min
Accuracy ± 2ml/h or ±15% of the displayed value, O2 flow accuracy ± 50 ml/min or ± 5% of setting value,
whichever is greater actual reading)
Balance gas (Air/N2O) flow accuracy
± 50 ml/min or ±5% of setting value, Ports and Connectors
whichever is greater Exhalation 22 mm OD / 15 mm ID conical
Total Flow Control Mode Inhalation 22 mm OD /15 mm ID conical
Total flow range 0.2 to 18 L/min Manual bag port 22 mm OD /15 mm ID conical
Total flow accuracy ± 100 ml/min or ± 5% of setting value,
whichever is greater Bag-to-Ventilator Switch
Type Bi-stable
O2 concentration Control Switch between manual and mechanical
Range 21% to 100% (The balance gas is Air) or 26% ventilation
to 100% (The balance gas is N2O)
Accuracy ± 5% V/V for flows < 1 L/min or 5% setting Integrated Adjustable Pressure Limiting (APL) Valve
for flows ≥ 1 L/min Range SP, 5 ~ 75 cmH2O
Tactile knob indication at above 30 cmH2O
Environmental Specifications Accuracy ± 3 cmH2O or ± 15% of the setting value,
Operating which is greater, but is not more than + 10
Temperature 10 ~ 40°C cmH2O
Relative humidity 15% ~ 95% (noncondensing)
Barometric (Kpa) 70 ~ 106 kPa Anesthetic Gas Scavenging System (AGSS)
Size (H x W x D) 430 x 132 x 114 mm
Storage Type of disposal system
Temperature -20 ~ 60°C for main unit, Active: High-flow or Low-flow
-20 ~ 50°C for O2 sensor Passive
Relative humidity 10% ~ 95% (noncondensing) Applicable standard ISO 80601-2-13
Barometric 50 ~ 106 kPa Pump rate 75 ~ 105 L/min (High-flow)
25 ~ 50 L/min (Low-flow)
Electromagnetic Compatibility Pressure relief device: Pressure compensation opening to the air
Immunity Complies with all requirements of IEC State indication of the disposal system: The float falls below the "MIN"
60601-1-2 mark on the sight glass when the disposal system does not work or
Emissions Complies with all requirements of IEC the pump rate is lower than 25 L/min (Low-flow) or 75 L/min
60601-1-2 (high-flow).
Connector of the disposal system: ISO 9170-2
Breathing System Specification
Breathing system volume (Pre-pak) Materials
Automatic ventilation 2850 ml All materials in contact with exhaled patient gases are autoclavable,
Manual ventilation 1800 ml except flow sensors (being not capable of being autoclaved), O2
Breathing system volume (Non Pre-pak) sensor, and mechanical pressure gauge.
Automatic ventilation 2600 ml All materials in contact with patient gas are latex free.
Manual ventilation 1800 ml
Suction Device
System Components Venturi Suction Regulator
Carbon dioxide absorbent canister Gas source Air, from system gas source
Absorbent capacity: 1500 mL Minimum flow 20 L/min
Integrated expiratory limb water trap Maximum vacuum ≥72 kPa at supply gas pressure of 280 kPa;
Capacity: 6 mL ≥73 kPa at supply gas pressure of 600 kPa
Continuous Suction Regulator
Breathing Circuit Parameters Supply Negative Pressure Suction
Compliance ≤4 mL/100Pa (bag mode) Maximum vacuum 517.5 mmHg to 540 mmHg (69 kPa to 72
Automatically compensates for kPa) with external vacuum applied of 540
compression losses within the breathing mmHg and 40 L/min free flow
circuit in mechanical mode Maximum flow 39 L/min to 40 L/min with external vacuum
Expiration resistance < 6.0 cm H2O @60 L/min applied of 540mmHg and 40 L/min free flow
Inspiration resistance < 6.0 cm H2O @60 L/min Minimum flow 20 L/min
--------------------------------------------------------------
System Pressure Gauge Please contact your local Mindray sales representative for the most
Range -20 ~ 100 cmH2O current information.
Accuracy ± (2% of the full scale reading + 4% of the

| Healthcare with reach are registered trademarks or trademarks owned by Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, © 2017 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China P/N: ENG-WATO EX-65 Pro datasheet-210285X4P-20180122
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: intl-market@mindray.com www.mindray.com

You might also like