Professional Documents
Culture Documents
Technical Information
OCPF81D/OCPF82D and
OCPF81/OCPF82
pH/ORP sensors, with digital Memosens
technology or analog
For mining operations, industrial water and
wastewater treatment
Application
• Flotation
• Leaching
• Neutralization
• Outlet monitoring
Your benefits
• With patented KNO3 electrolyte bridge for improved protection against electrode
poisons such as S2- or CN- ions
• Optionally available with flat membrane for improved wear resistance
• Threaded connection NPT ¾" top and bottom for easy installation at user end
• Suitable for measurements in the 0 to 14 pH range and in the 0 to 110 °C
(32 to 230 °F) temperature range
• Digital pH sensors with integrated temperature sensor, analog pH sensors with or
without integrated temperature sensor
• Protection guard against damage
Other advantages provided by Memosens technology
• Maximum process safety
• Data security thanks to digital data transmission
• Very easy to use as sensor data are saved in the sensor
• Predictive maintenance can be performed by recording sensor load data in the
sensor
OCPF81D/OCPF82D and OCPF81/OCPF82
1
A0024721
2
OCPF81D/OCPF82D and OCPF81/OCPF82
Digital sensors can store measuring system data in the sensor. These include the following:
• Manufacturing data
• Serial number
• Order code
• Date of manufacture
• Calibration data
• Calibration date
• Calibrated slope at 25 °C (77 °F) (CPF81D)
• Calibrated zero point at 25 °C (77 °F) (CPF81D)
• Calibrated offset (OCPF82D, ORP mV measuring mode)
• Slope as % (OCPF82D, ORP % measuring mode)
• Temperature offset
• Number of calibrations
• Serial number of the transmitter used for the last calibration
• Calibration database (stores the last 8 calibrations in the Memosens head)
• Application data
• Temperature application range
• pH application range (OCPF81D)
• ORP application range (OCPF82D)
• Date of first commissioning
• Maximum temperature value
• Operating hours at temperatures above 80 °C (176 °F) and 100 °C (212 °F)
• Operating hours at very low and very high pH values (Nernst voltage below -300 mV, above
+300 mV)
Reliability
Dependability Easy handling
Sensors with Memosens technology have integrated electronics that store calibration data and other
information (e.g. total hours of operation or operating hours under extreme measuring conditions).
Once the sensor has been connected, the sensor data are transferred automatically to the transmitter
and used to calculate the current measured value. As the calibration data are stored in the sensor,
the sensor can be calibrated and adjusted independently of the measuring point. The result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point.
• Thanks to the availability of the sensor data, maintenance intervals can be accurately defined and
predictive maintenance is possible.
• The sensor history can be documented on external data carriers and in evaluation programs.
• Thus, the current application of the sensors can be made to depend on their previous history.
3
OCPF81D/OCPF82D and OCPF81/OCPF82
Input
Measured values OCPF81D and
OCPF81 pH value
Temperature
OCPF82D, OCPF82
-1500 mV to +1500 mV
Pay attention to the operating conditions in the process.
Power supply
Electrical connection OCPF81D and OCPF82D
The electrical connection to the transmitter is established using measuring cable CYK10.
GY GND
YE Com B
GN Com A
WH –
BN +
A0024019
PM Coax BK Ref.
WH pH
pH/ORP
+ Pt 100 GN Temp
WH Temp.
YE Temp.
BN PM
1
A0024668-EN
4
OCPF81D/OCPF82D and OCPF81/OCPF82
A BK BK B
Reference
BK BK
pH/ORP
GN
Temperature
WH GN
Temperature
YE
Temperature
GN/YE GN/YE WH
Outer screen YE
BN BN
PM / n. c.*
A0024680-EN
Performance characteristics
Resistance Glass impedance
150 MΩ at 25 °C (77 °F)
Installation
Installation instructions • Before screwing in the sensor, make sure the assembly thread, the O-rings and the sealing surface
are clean and undamaged and that the thread runs smoothly.
• Pay attention to the installation instructions provided in the Operating Instructions of the
assembly used.
‣ Screw in the sensor and tighten by hand with a torque of 3 Nm (2.21 lbf ft) (specifications only
apply if installing in Endress+Hauser assemblies).
A0024681
5 Flow installation
5
OCPF81D/OCPF82D and OCPF81/OCPF82
A0024690
6
OCPF81D/OCPF82D and OCPF81/OCPF82
Environment
Ambient temperature range NOTICE
Danger of frost damage
‣ The sensor must not be used at temperatures below 0 °C (32 °F).
Process
Process temperature range OCPF81D, OCPF81
• Version LH: 0 to 110 °C (32 to 230 °F)
• Version NN: 0 to 80 °C (32 to 176 °F)
OCPF82D, OCPF82
0 to 80 °C (32 to 176 °F)
Process pressure range 1 to 10 bar absolute, at 80 °C (15 to 145 psi absolute, at 176 °F)
Pressure-temperature
ratings psi bar
145 10
120 8
80 6
4
40
2
15 1
0 20 40 60 80 100 °C
32 60 100 140 180 220 °F
A0024703
7 Pressure-temperature ratings
7
OCPF81D/OCPF82D and OCPF81/OCPF82
Mechanical construction
Design, dimensions Integrated preamplifier (optional)
With preamplifier versions, the sensor check function (SCS) of the transmitter is ineffective and
should be turned off.
35 (1.38)
35 (1.38)
35 (1.38)
1 1 1
2
2 2
3 3
3
150 (5.91)
140 (5.51)
150 (5.91)
12.5 (0.49)
23 (0.91)
2 2
58.5 (2.3)
4 4
4
5
Ø 22 (0.87)
Ø 22 (0.87) Ø 22 (0.87)
A0024671
8 OCPF81D with long shaft. Dimensions: mm (in) 10 OCPF82D with short shaft.
Dimensions: mm (in) Dimensions: mm (in)
1 Memosens plug-in head
1 Memosens plug-in head 2 NPT ¾" thread 1 Memosens plug-in head
2 NPT ¾" thread 3 Across flats AF 26 2 NPT ¾" thread
3 Across flats AF 26 4 Flat membrane 3 Across flats AF 26
4 Protection guard 4 Platinum ring
5 Protection guard
6
8
OCPF81D/OCPF82D and OCPF81/OCPF82
OCPF81, OCPF82
1 1
1
2 2
2
3 3
3
150 (5.91)
150 (5.91)
140 (5.51)
2
12.5 (0.49)
23 (0.91)
2 2
58.5 (2.3)
4
4 4
Ø 22 (0.87)
Ø 22 (0.87) Ø 22 (0.87)
A0024676
head and short shaft. 12 OCPF82 fixed cable version 13 OCPF82 fixed cable version
Dimensions: mm (in) and short shaft. Dimensions: and long shaft. Dimensions:
mm (in) mm (in)
1 TOP68 plug-in head
2 NPT ¾" thread 1 Fixed cable 1 Fixed cable
3 Across flats AF 26 2 NPT ¾" thread 2 NPT ¾" thread
4 Flat membrane 3 Across flats AF 26 3 Across flats AF 26
4 Protection guard 4 Protection guard
Weight 0.12 to 0.15 kg (0.26 to 0.33 lb), depending on the version and excluding a cable
EAC The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which
apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.
Marine approvals A selection of the devices and sensors have type approval for marine applications, issued by the
following classification societies: ABS (American Bureau of Shipping), BV (Bureau Veritas), DNV-GL
(Det Norske Veritas-Germanischer Lloyd) and LR (Lloyd’s Register). Details of the order codes of the
approved devices and sensors, and the installation and ambient conditions, are provided in the
relevant certificates for marine applications on the product page on the Internet.
9
OCPF81D/OCPF82D and OCPF81/OCPF82
Ordering information
Product page
OCPF81D ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
Version:
7 Basic version - Memosens
Application Range:
LH 0-14pH, 0-110oC
NN 0-11pH, 0-80oC
Insertion Length:
1 23mm + electrode guard
2 13mm, flat membrane
3 58mm + electrode guard
___ Approval:
1 Non-hazardous area
___
Customer Version:
NNT Neutral test version
WTI Wedgewood Analytical
OPF82D ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
Version
7 Basic version - Memosens
Application Range:
PA ORP, Platinum, 0-80oC
Insertion Length:
1 23mm + electrode guard
3 58mm + electrode guard
___ Approval:
1 Non-hazardous area
___
Customer Version:
NNT Neutral test version
WTI Wedgewood Analytical
10
OCPF81D/OCPF82D and OCPF81/OCPF82
Order
root Extensions
OCPF81 ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
Application Range:
LH 0-14pH, 0…110oC
NN 0-11pH, 0…80oC
YY Special version, TSP-no. to be spec.
Insertion Length:
1 23mm (0.91") + electrode guard
2 13mm (0.51") + flat membrane
3 58.5mm (2.28") + electrode guard
4 13mm (0.51") + recessed glass
5 24.5mm (0.96") + flat membrane
9 Special version, TSP-no. to be spec.
Process Connection
1 Thread NPT3/4"
___ Equipment:
E Standard
F Pt100
H Pt1000
Y Special version, TSP-no. to
be spec.
Cable Length:
1 W/O
2 4.6m (15ft)
3 9.1m (30ft)
7 custom length in meters
8 custom length in feet
9 Special version, TSP-
no. to be spec.
Connection
1 BNC
2 Crimp Sleeves
3 Top68/ESA plu-in head
9 Special version, TSP-
no. to be spec.
Customer Version:
NNN No Name
Y Special version, TSP-
no. to be spec.
11
OCPF81D/OCPF82D and OCPF81/OCPF82
OCPF82 ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
Application Range:
PE ORP, Platinum, 0…80oC
YY Special version, TSP-no. to be spec.
Insertion Length:
1 23mm (0.91") + electrode guard
2 13mm (0.51") + flat membrane
3 58.5mm (2.28") + electrode guard
5 24.5mm (0.96") + flat membrane
9 Special version, TSP-no. to be spec.
Process Connection
1 Thread NPT3/4"
___ Equipment:
E Standard
Cable Length:
1 W/O
2 4.6m (15ft)
3 9.1m (30ft)
7 custom length in meters
8 custom length in feet
9 Special version, TSP-
no. to be spec.
Connection
1 BNC
2 Crimp Sleeves
3 Top68/ESA plu-in head
9 Special version, TSP-
no. to be spec.
Customer Version:
NNN No Name
Y Special version, TSP-
no. to be spec.
12
OCPF81D/OCPF82D and OCPF81/OCPF82
Accessories
The following are the most important accessories available at the time this documentation was
issued.
‣ For accessories not listed here, please contact your Service or Sales Center.
13
OCPF81D/OCPF82D and OCPF81/OCPF82
14