Professional Documents
Culture Documents
Experiment:
Oscilloscope Rise Time Measurements: Measure the rise time of the Sync output signal
from the HP 33120A Function Generator, using various, mostly improper, connections
between the instruments:
Theoretical:
Background:
All experiments in the laboratory will be performed at a test bench which has several
basic electronic instruments. They include: (1) A DC power supply. (2) A function
generator. (3) A digital multi-meter (DMM). (4) An oscilloscope (analog/DSO). On one
hand, DC power supplies and function generators are power sources. It means that they
generate power. In other words, they supply us with the power needed to perform
experiments and tests. On the other hand, digital voltmeters and oscilloscopes are
measuring devices that measure different electrical quantities such as voltage,
resistance, current, etc. DC Power Supply: The DC power supply output consists of two
independent units which can be connected in different configurations such as series,
parallel, or used independently. Some DC power supplies found in the lab have a third
unit which gives a fixed 5 Volts 3 Amperes that is not adjustable. This dedication of 5-
volts unit is important since it is frequently needed in electronic experiments.
Function Generator:
A DMM is a universal meter which measure voltage and current (both DC and AC) and
resistance, with high precision. Some DMMs measure capacitance, inductance, frequency, and
temperature. A multi-meter or a multi-tester, also known as a VOM (Volt-Ohm meter), is an
electronic measuring instrument that combines several measurement functions in one unit. A
typical multi-meter would include basic features such as the ability to measure voltage, current,
and resistance. Analog multi-meters use a micro-ammeter whose pointer moves over a scale
calibrated for all the different measurements that can be made. Digital multi-meters (DMM,
DVOM) display the measured value in numerals, and may also display a bar of a length
proportional to the quantity being measured. When a DMM is used to measure an AC quantity,
it actually measures the RMS value of it. Different waveforms have different RMS values. The
commonly encountered sinusoidal wave has an RMS value of Vp 2 where Vp is the peak
voltage of the sine wave.
Oscilloscope:
STEP 1:
Set the function generator for a 5.000 MHz sine wave. For this part the amplitude is
immaterial. Connect the Sync output (upper BNC jack on function generator) to the
Analog Input 1 (A1) on the oscilloscope using BNC to Banana Plug adapters and two 3
foot Banana plug wires.
STEP 2 :
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
STEP 3:
Repeat the same measurement, but this time connect the Sync output to the A1 scope
input using a BNC cable
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
STEP 4:
Repeat the same measurement, but this time with a 50 termination on the BNC cable
at the A1 input to the scope. (Use a BNC tee)
STEP 5:
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
STEP 6:
Remove the BNC cable and connect the 10x probe to the A1 input on the HP54622D
scope. Remove the probe hook adapter (if it is installed) and place the center pin of the
probe into the function generator Sync output BNC, making contact with the BNC center
conductor. Do not connect the probe ground to anything for this measurement!
STEP 7:
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
What is the signal current return path for this measurement?
STEP 8:
Repeat the last measurement, but with the probe ground wire clipped onto the shell of
the function generator Output BNC (immediately below the Sync BNC).
STEP 9:
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
STEP 10:
Repeat the last measurement, but with the probe ground wire clipped onto the shell of
the Sync BNC on the function generator.
STEP 11:
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
STEP 12:
Finally, repeat the last measurement, but use the probe-tip-to-BNC adapter, which
grounds the probe ground ring directly to the BNC shell.
STEP 14:
Measure the rise time and overshoot of the rising edge of the Sync signal on the
oscilloscope. Record any observations about the waveform.
Objective:
To investigate the RMS value of an A.C sinusoidal waveform and the power
In electronics circuits, the signal is associated with an AC (Alternating current) and DC (Direct
Current). The sinusoidal wave or sine wave represents the periodic AC signal that varies with
time and fluctuates between positive and negative cycles. While DC signal/waveform represents
the magnitude of voltage and current values easily and stable. But the magnitude of the
current and voltage values of the AC signal should be calculated by using its instantaneous
values, a peak value of the signal, peak to peak value, average value, and RMS voltage. This
article describes the RMS voltage theory and its methods.
The term RMS stands for “Root Mean Square value”. The RMS value represents the
voltage value of an AC signal equivalent to the DC voltage. The amount of power dissipated or
heating effect is the same in both AC and DC signals. The values of a sinusoidal signal vary with
time because the magnitude of the current and voltage values always change with time and it is
not applicable for DC circuits (magnitude constant).
In simple words, it can be defined as the square roots of the average values of the squares of all
the instantaneous values of an AC signal. It is denoted by Vrms or Irms
This can be obtained by squaring the input values and calculating the average value of an AC
signal. The square root of the average value obtained gives the root mean square voltage.
In general AC and DC signals are represented in the form of voltage or current waveforms. We
knew that the magnitude of the DC signal is constant and it is very easy to calculate. But the AC
signal fluctuates between negative and positive half cycles and varies with time. So, it is difficult
to find the magnitude of the voltage and current values of the alternating waveform. This theory
can be explained from the alternating waveform as shown below.
RMS Voltage Theory
The most effective method to find the magnitude of the voltage values of the sinusoidal
waveform is the calculation of RMS Voltage value. We can compare the amount of power of AC
and DC signals given to the load or circuit. This voltage refers to the DC equivalent voltage of an
AC signal because the amount of power dissipated or heating effect of both AC and DC
waveforms in the given circuit is the same.
That means the power delivered to the load of a DC is equivalent to the RMS voltages of a sine
wave (AC signal). This gives the effective voltage value and it is denoted by Veff or Ieff. If the
supply voltage is 220V-240V, then the RMS voltage values of an AC are also 220V-240V,
which is equivalent to the power of DC voltage.
If the RMS voltages of AC and DC signal are the same, then the amount of power dissipated in
the circuit is also the same. This is also known as an effective voltage which is equivalent to the
DC voltage in providing power to the circuit.
Materials:
Multimeter (V.O.M)
2- channel oscilloscope
Initial Settings:
Connect the power supply unit to be main supply line. DO Not switch in yet.
Procedure:
3. Switch on the power supply unit and set the source voltage to 10 VDC, as shown
on the voltmeter.
4. Transfer the voltmeter across the potentiometer (mid position and one leg)
5. Get an appropriate time base on the oscilloscope to have a good trace of the
waveform and draw the trace or waveform you see on another sheet of paper.
6. Now vary the settings of the potentiometer decreasing and then increasing its
value. Notice the waveform on the oscilloscope and its reading or the voltmeter
reading.
the oscilloscope and then draw the trace or waveform you see on another sheet
of paper
8. Switch off the power supply unit (both D.C and A.C source) connect the circuit
9. Set the potentiometer to its mid position and initially set the slide to the left.
10. Switch on the power supply unit and set the D.C source voltage to 5v as shown
on the voltmeter. The lamp should be on, but dim. Notice the intensity of the
lamp.
11. Switch the slide to the right and adjust the potentiometer until the intensity of the
12. To ensure that intensity of the lamp is same for both positions switch the slide
switch left and right adjusting the potentiometer at the same time, until the
13. On the table 1.1 record the readings of the D.C and A.C voltage and current as
√
❑ 2
❑ Yn
∑
❑ n
=
Conclusion:
Laboratory 3: Resistive, Inductive and Capacitive Circuits with a Sinusoidal
Excitation
Objective:
THEORY:
An RLC circuit is an electrical circuit consisting of a resistor (R), an inductor (L), and a capacitor
(C), connected in series or in parallel. The name of the circuit is derived from the letters that are
used to denote the constituent components of this circuit, where the sequence of the
components may vary from RLC. An important property of this circuit is its ability to resonate at
a specific frequency, the resonance frequency, f0. Frequencies are measured in units of hertz.
In this article, however, angular frequency, ω0, is used which is more mathematically
convenient. This is measured in radians per second. They are related to each other by a simple
proportion, Resonance occurs because energy is stored in two different ways: in an electric field
as the capacitor is charged and in a magnetic field as current flows through the inductor. Energy
can be transferred from one to the other within the circuit and this can be oscillatory. A
mechanical analogy is a weight suspended on a spring which will oscillate up and down when
released. This is no passing metaphor; a weight on a spring is described by exactly the same
second order differential equation as an RLC circuit and for all the properties of the one system
there will be found an analogous property of the other. The mechanical property answering to
the resistor in the circuit is friction in the spring–weight system. Friction will slowly bring any
oscillation to a halt if there is no external force driving it. Likewise, the resistance in an RLC
circuit will "damp" the oscillation, diminishing it with time if there is no driving AC power source
in the circuit. The resonance frequency is defined as the frequency at which the impedance of
the circuit is at a minimum. Equivalently, it can be defined as the frequency at which the
impedance is purely real (that is, purely resistive). This occurs because the impedances of the
inductor and capacitor at resonance are equal but of opposite sign and cancel out. Circuits
where L and C are in parallel rather than series actually have a maximum impedance rather
than a minimum impedance. For this reason they are often described as antiresonators, it is still
usual, however, to name the frequency at which this occurs as the resonance frequency.
Materials:
Function generator
Multimeter (VOM)
2 – channel oscilloscope
Procedure:
A Resistive circuit
2. Set the sine wave function generator to 10 Vac peak-to-peak at 250 Hz as shown
in the oscilloscope
3. Set an appropriate time base on the oscilloscope to have a good trace of the
waveform and draw the trace or waveform you see on another sheet of paper.
What is the phase difference or phase shift of the two traves or waveform?
_____________________________________________________
What did you observe about the time in which the two waveforms started and
B Inductive circuit
2. Set the sine wave function generator to 10 Vac peak-to-peak at 250 Hz as shown
in the oscilloscope
3. Set an appropriate time base on the oscilloscope to have a good trace of the
waveform and draw the trace or waveform you see on another sheet of paper.
Based on the trave or waveform you draw, what is the phase difference or phase
_________________________________________________________
Which of the two monitor reaches its positive peak value first?
_________________________________________________________
C Capacitive circuit
2. Set the sine wave function generator to 10 Vac peak-to-peak at 250 Hz as shown
in the oscilloscope
3. Set an appropriate time base on the oscilloscope to have a good trace of the
waveform and draw the trace or waveform you see on another sheet of paper.
Based on the trave or waveform you draw, what is the phase difference or phase
__________________________________________________________
Which waveform monitor is leading?
__________________________________________________________
What relationship can you established between the voltage and current
__________________________________________________________
Table 2.3
Conclusion:
Laboratory 4: Impedance of a Series RLC Circuit
Objective:
To investigate the impedance of a series RLC circuit and compare it with the
Theory:
The behavior of a series circuit comprising all three kinds of circuits elements
(resistor, inductor and capacitor) will depend upon the relative magnitudes of the
act opposite each other, it should be clear that the combination behaves (1) like a R-L
circuit or as an inductive circuit when X L is greater than X C , (2) like a R-C circuit or as a
capacitive circuit when X C is greater than X L and (3) like a pure resistance R circuit or
as a resistive circuit when X L equals X C . This latter condition has a special significance
Materials:
Function generator
Multimeter (V.O.M.)
AC Miliameter 0 – 250mA
Initial Settings:
Connect the power supply unit to the main supply line. DO NOT switch on yet.
Procedure:
2. Set the sine wave function generator to 100Hz, with constant output amplitude of
3. On the table 1, following the results table, record the readings of total current,
What did you observe about the magnitude of the current flowing to the resistor,
_____________________________________________________________
What is the phase relationship between V L and V c ? How do you combine the
_____________________________________________________________
From your phasor diagram, what is the phase relationship between the resultant
of V L and V c to V R?
_____________________________________________________________
5. On the table 2, following the results table, calculate the required parameters.
Optional Activity:
2. Set the sine wave function generator to 60Hz, with constant output amplitude of
3. On table 2a and table 2b, following the results table, record the readings of total
current, voltage across the resistor and then the voltage across the coil. And then
4. On another sheet of paper, solve the internal resistance (r) and inductance (L) of
Figure 1
Table 1
100
200
300
400
500
Table 2
Formula Frequency Hz
Ohm’s Law VR
(Measured Values) R=
IR
VL
X L=
IL
VC
XC=
IC
VT
¿ Z∨¿
IT
Equation R=R
Impedance
X L =2 πfL
1
XC=
2 πfC 1
Pythagorean X =X L− X C
Theorem
¿ Z∨¿ √ R + X
❑ 2 2
(Impedance
Triangle)
−1 X
θ Z =tan
R
(rectangular form)
Z=|Z|∠ θ Z
(polar form)
Table 2a
Measured Values
Frequency Total Voltage (V T Total Current ( I T ) Resistor Voltage ( Coil’s Voltage (V coil
) V R) )
Hz mA RMS
V RMS V RMS V RMS
60
Table 2b
Calculated Values
Total Current ( Total Impedance Coil’s Impedance ( Coil’s Internal Coil’s Inductance
I T) ( ZT ) Z coil) Resistance (r) (L)
mA RMS Ω Ω Ω mH
Discussion:
Conclusion:
Laboratory 5: Impedance and Admittance of a Parallel RLC Circuit
Objective:
Theory:
characteristic of the equivalent circuit will depend on the relative magnitudes of the
inductive reactance X L and the capacitive reactance X C . Take note that X L is directly
of R, L, and C, the circuit will be capacitive when X L is greater than X C . This latter case
Materials:
Function Generator
Multimeter (V.O.M.)
Connect the power supply unit to the main supply line. DO NOT switch on yet.
Procedure:
2. Set the sine wave function generator to 300 Hz, with constant output amplitude of
3. On table 1, following the results table, record the readings of total current, current
flowing in the resistor, inductor and capacitor respectively. This can be done by
What did you observe about the magnitude of the voltage across the resistor,
_______________________________________________________________
What is the phase relationship between I L and I c? How do you combine the
_______________________________________________________________
Figure 1
Table 1
300
400
500
600
700
Table 2
Formula Frequency Hz
Ohm’s Law IR
(Measured G=
VR
Values)
IL −3
β L= ×10
VL
IC −3
β C= ×10
VC
VT
¿ Z∨¿
IT
1
|Y |=
IT −3
¿ Z∨¿= × 10 ¿
VT
Equation 1
Impedance
G=
R
(for purely resistive
branch only)
1 −3
β L= ×10
2 πfL
−3
β C =2 πfC × 10
Pythagorean −3
β=β L−β C ×10
Theorem
¿ Y ∨¿ √ G + β ×10
❑ 2 2 −3
(Impedance
Triangle) −1 β
θY =tan
G
Complex Number 1 1 1 1
Y= = + +
Z R jX L − jX C
(rectangular form)
Y =¿Y ∨∠θY
(polar form)
1
Z=
Y
(polar form)
Discussion:
Conclusion:
Laboratory 6: Series Resonance
Objectives:
circuit.
To verify the bandwidth and cut-off frequencies of the series RLC circuit.
Theory:
characteristic changes with changes in frequency. In this tutorial we will look at the frequency
response of a series resonance circuit and see how to calculate its resonant and cut-off
frequencies.
Thus far we have analyzed the behavior of a series RLC circuit whose source
voltage is a fixed frequency steady state sinusoidal supply. We have also seen in our
tutorial about series RLC circuits that two or more sinusoidal signals can be combined
using phasors providing that they have the same frequency supply.
But what would happen to the characteristics of the circuit if a supply voltage of
fixed amplitude but of different frequencies was applied to the circuit. Also what would
the circuits “frequency response” behavior be upon the two reactive components due to
reactance of the inductor becomes equal in value to the capacitive reactance of the
capacitor. In other words, XL = XC. The point at which this occurs is called the Resonant
Frequency point, (ƒr) of the circuit, and as we are analyzing a series RLC circuit this
Materials:
Function Generator
Multimeter (V.O.M.)
Initial Settings:
Connect the power supply unit to the main supply line. DO NOT switch on yet.
Procedure:
2. Set the sine wave function generator initially to a minimum frequency (say 10Hz),
value (say 10Hz) to a maximum value (say 2000Hz). Find the frequency at which
4. In table 1, following the results table, record the readings of total current, voltage
across the resistor, inductor and capacitor respectively. Calculate the required
parameters.
In this circuit diagram, what do you call this frequency in which the circuit
_______________________________________________________________
At this frequency, how can you relate the magnitude of total impedance to the
_______________________________________________________________
At this frequency, based on table 1, what did you observe about the magnitude of
voltage across the inductor and the magnitude of voltage across the capacitor?
________________________________________________________________
At this frequency, based on table 1, what did you observe about the magnitude of
the total voltage and the magnitude of voltage across the resistor?
________________________________________________________________
Based on table 2 or table 3, what did you observe about the magnitude of total
impedance?
________________________________________________________________
Based on table 2 or table 3, what did you observe about the magnitude of
________________________________________________________________
Based on table 2 and table 3, is the calculated value of the magnitude of total
________________________________________________________________
6. Based on the given value of R, L and C, compute the resonant frequency using
the formula:
1
f o=
2 π √ LC
❑
Does your calculated value agree with the frequency found previously for
maximum current of the circuit registered in the ammeter? If not, cite possible
reason.
________________________________________________________________
8. Based on the given values of R, L and C, compute the cut-off frequencies using
the formula:
f 1=
2
RL [√
❑
1+
4L
CR 2
−1
] ;
f 2=
2
RL [√
❑
1+
4L
CR 2
+1
]
2π 2π
Note: this formula is intended only for the given circuit diagram of figure 1
9. Input the computed cut-off frequencies in table 1, following the results table,
record the readings of total current, voltage across the resistor, inductor and
capacitor respectively.
10. In table 1, with a suitable interval of frequency in between f1 and the frequency at
which the ammeter reading registered its maximum value and then to f2,
following the results table, record the readings of total current, voltage across the
________________________________________________________________
________________________________________________________________
resonant frequency? From this, the cut-off frequencies are also known as
________________________________________________________________
11. On another sheet of paper, at the same axes, draw the curves of V R, V L and V c
Based on the curves you draw (from the two trials), what is the relationship can
________________________________________________________________
Based on the curves you draw (from the two trials), what is the relationship can
________________________________________________________________
Wiring Diagram:
Figure 1
Preferred settings for figure 1:
Table 1
TRIAL 1: Resistance =
TRIAL 1: Resistance =
Table 2:
Trial 1 Trial 2
Resistance VR
R=
IR
Inductive Reactance VL
X L=
IL
Capacitive Reactance VC
XC=
IC
Total Impedance VT
¿ Z∨¿
IT
Quality Factor VL
Q=
VT
VC
Q=
VT
Table 2:
Trial 1 Trial 2
Resistance R=R
Capacitive Reactance 1
XC=
2 πfc
Resonant Frequency 1
f 0=
2 π √ LC
❑
f 0=❑√ f 1 f 2
[√ ]
Cut-off Frequencies (f1, f2)
2 ❑ 4L
1+ −1
RL CR 2
f 1=
2π
f 2=
2
RL [√❑
1+
4L
CR 2
+1
]
2π
Bandwidth BW = f 2−f 1
R
BW =
2 πL
Quality Factor f0
Q=
BW
2 πf 0 L
Q=
R
1
Q=
2 πf 0 CR
Q=
√
1❑ L
R C
Discussion:
Conclusion:
Laboratory 7: Parallel Resonance
Objectives:
circuit.
To verify the bandwidth and cut-off frequencies of the parallel RLC circuit.
Theory:
Parallel Resonance means when the circuit current is in phase with the applied
parallel.
Materials:
Function Generator
Multimeter (V.O.M.)
Initial Settings:
Connect the power supply unit to the main supply line. DO NOT switch on yet.
Procedure:
2. Set the sine wave function generator initially to a minimum frequency (say
3. Notice the ammeter reading, as you vary the source frequency from the minimum
value (say 10Hz) to a maximum value (say 2000Hz). Find the frequency at which
4. In table 1, following the results table, record the readings of total current and
5. Based on the given value of L and C, compute the resonant frequency and the
√ ()
L
2
1 ❑ 1 r |Z 0|= C
f 0= − ;
2 π LC L r
Based on the table 1 (from the two trials) and the relationship you establish
between the resistance and bandwidth, what must be done to the inductor
Wiring Diagram:
Table 1:
Trial 1:
R2 =
Trial 2:
R2 =
Discussion:
Conclusion:
Laboratory 8: Power in AC circuits
Objective:
To investigate the power dissipated by the resistive component and the power
associated with reactive components as well as the power associated with the
To investigate the power factor on non-linear loads and adjust its factor near
unity.
Theory:
Alternating Current and Voltages change their magnitude and direction with time.
This changes the way calculations for power and other quantities are done in circuits.
Furthermore, with the introduction of capacitors and inductances, many other effects
come into play which alters the power calculation in these circuits from the usual ways
of calculating power with DC sources. It becomes essential to know these workings and
concepts so that one can apply them in situations in real life where circuits are complex
Materials:
Function generator
Multimeter (VOM)
Wattmeter
Initial settings:
Connect the power supply unit to the main supply line. DO Not switch on yet.
Procedure:
i. Power measurement
2. Set the sine wave function generator to 250 Hz, with constant output
3. Record the result, record the readings of total current , the voltage across
the resistor , the voltage across the inductor and the total power absorbed
by the circuit.
4. Transfer the potential leads of the wattmeter across the resistor (nodes B
and C ) and record the reading of the power absorbed by the resistor.
5. Transfer the potential leads of the wattmeter across the inductor (nodes A
and B ) and record the reading of the power absorbed by the inductor.
6. From the given values and measured values in table 8.1 calculate the
2. Set the sine wave function generator to 250 Hz, with constant output
the resistor , the voltage across the inductor and the total power absorbed
by the circuit.
7. From the given values and measured values of table 8.3 calculate the
required parameter
maximum power.
3. Set the sine wave function generator to 259 Hz with constant outpur
4. Record the result, record the readings of total current , the voltage across
the resistor , the voltage across the inductor and the total power absorbed
by the circuit.
Table 8.1
Power Measurement
Parameter Given value Parameter Given value
Frequency (Hz) Total voltage (Vr)
(Vrms)
Resistance (R) Resistor voltage
(Vr)
(Vrms)
Inductance (L) Inductor Voltage
(VL)
(Vrms)
Reactance (X) Total current
IXI = 2πfL(Ω) Total power
Impedance (Z) Resistor power
IZI = ❑√ R2 +¿ X 2 ¿ (Ω) Inductor power
Table 8.2
Power measurement
Parameter Calculated values
Impedance angle
Power factor
Reactive power
Real power
Reactive power
Apparent power
Table 8.3
Table 8.4
Interpretation:
Conclusion:
Laboratory 9: Three-phase Circuits
OBJECTIVE
There are two types of system available in electric circuit, single phase and three phase
system. In single phase circuit, there will be only one phase, i.e the current will flow through
only one wire and there will be one return path called neutral line to complete the circuit. So in
single phase minimum amount of power can be transported. Here the generating station and
load station will also be single phase. This is an old system using from previous time.
In 1882, new invention has been done on polyphase system, that more than one phase can be
used for generating, transmitting and for load system. Three phase circuit is the polyphase
system where three phases are send together from the generator to the load.
Each phase are having a phase difference of 120o, i.e 120o angle electrically. So from the total
of 360o, three phases are equally divided into 120o each. The power in three phase system is
continuous as all the three phases are involved in generating the total power. The sinusoidal
waves for 3 phase system is shown below-
The three phases can be used as single phase each. So if the load is single phase, then one
phase can be taken from the three phase circuit and the neutral can be used as ground to
complete the circuit.
EQUIPMENT REQUIRED
Refer to the Equipment Utilization Chart, in Appendix A of this manual, to obtain the list of
equipment required to perform this exercise.
PROCEDURE
CAUTION!
G 1. a. Using your Power Supply and AC Voltmeter connect the circuit shown
in Figure 45-1
Figure 45-1.
b. Turn on the power supply and adjust the line-to-neutral voltage (as
indicated by the power supply voltmeter) to exactly 120 V ac.
E
4 to 5 = V ac
E
5 to 6 = V ac
E
4 to 6 = V ac
d. Return the voltage to zero and turn off the power supply.
E = V ac
line-to-line
b. Turn on the power supply and adjust the line-to-neutral voltage (as
indicated by the power supply voltmeter) to exactly 120 V ac.
E =
4 to N V ac
E =
5 to N V ac
E =
V ac
6 to N
d. Return the voltage to zero and turn off the power supply.
E = V ac
line-to-neutral
E /E =
line-to-line line-to-neutral
E E
1 to 2 1 to N
= =
V ac V ac
E E
2 to 3 2 to N
= =
V ac V ac
E E
1 to 3 3 to N
= =
V ac V ac
b. Are the fixed line-to-line and the line-to-neutral voltages reasonably
equal?
G Yes G No
d. Measure and record the voltages across, and the currents through,
the three load resistances R1, R2 and R3.
I
E1 = V ac =1 A ac
E2 = V ac I2 = A ac
E3 = V ac I3 = A ac
e. Return the voltage to zero and turn off the power supply.
G Yes G No
Eload = V ac
h. What is the average value of the line-to-line voltage (from procedure
1 (e)):
E = ac
line-to-line
E /E =
line-to-line load
P1 = W
P2 = W
P3 = W
PT= W
d. Measure and record the voltages across, and the currents through,
the three load resistances R1, R2 and R3.
E1 = V ac I1 = A ac
E2 = V ac I2 = A ac
E3 = V ac I3 = A ac
e. Return the voltage to zero and turn off the power supply.
G Yes G No
Iload = A ac
h. Disconnect the three current meters and insert them in series with
power supply terminals 4, 5 and 6. Replace the removed current
meters with connection leads as shown in Figure 45-4.
Figure 45-4
i. Turn on he power supply and adjust for 120 V ac.
j. Measure and record the three line currents.
I
4 = A ac
I
=5 A ac
I6 = A ac
k. Return the voltage to zero and turn off the power supply.
Iline = ac
I /I =
line load
G Yes G No
P1 = W
P2 = W
P3 = W
Laboratory 10: Maximum power transfer and superposition
Objective:
To apply Thevenin’s and Norton’s theorems in finding the current flowing in a particular resistor
(variable load) in a particular network.
To verify the theorems by comparing the simulated values to those obtained by measurement.
The Maximum Power Transfer Theorem is not so much a means of analysis as it is an aid to
system design. Simply stated, the maximum amount of power will be dissipated by a load
resistance when that load resistance is equal to the Thevenin/Norton resistance of the network
supplying the power. If the load resistance is lower or higher than the Thevenin/Norton resistance
of the source network, its dissipated power will be less than the maximum.
This is essentially what is aimed for in radio transmitter design, where the antenna or
transmission line “impedance” is matched to final power amplifier “impedance” for maximum radio
frequency power output. Impedance, the overall opposition to AC and DC current, is very similar
to resistance and must be equal between source and load for the greatest amount of power to be
transferred to the load. A load impedance that is too high will result in low power output. A load
impedance that is too low will not only result in low power output but possibly overheating of the
amplifier due to the power dissipated in its internal (Thevenin or Norton) impedance.
The Maximum Power Transfer Theorem does not: Maximum power transfer does not coincide
with maximum efficiency. Application of The Maximum Power Transfer theorem to AC power
distribution will not result in maximum or even high efficiency. The goal of high efficiency is more
important for AC power distribution, which dictates a relatively low generator impedance
compared to the load impedance.
Similar to AC power distribution, high fidelity audio amplifiers are designed for a relatively low
output impedance and a relatively high speaker load impedance. As a ratio, “output impedance”:
“load impedance” is known as damping factor, typically in the range of 100 to 1000.
Maximum power transfer does not coincide with the goal of the lowest noise. For example, the
low-level radio frequency amplifier between the antenna and a radio receiver is often designed
for the lowest possible noise. This often requires a mismatch of the amplifier input impedance to
the antenna as compared with that dictated by the maximum power transfer theorem.
Materials:
DC supply (VS = 15V)
Digital multimeters
Resistors R1=1.8kΩ; R2=3.6kΩ; R3=820Ω; R4=R5=100Ω; RL=180Ω
Procedure:
Thevenin’s Theorem
Norton’s Theorem
Measured
values
V1 R1 R2 R3 R4 R5 RL
Table 3-1
Table 3-2
Table 3-3
Calculations
Thevenin’s resistance
= 2.63Ω
= 94.73Ω
= 94.73Ω
Percentage Error
Thevenin’s voltage
Percentage Error
Norton’s Resistance
RN = RTH =
Norton’s Current
Percentage Error
% of Relative Error of IN =
Current in RL
IL = 5.058/(616.91 + 180)
= 6.13mA
Percentage Error
% of Relative Error of IL =
Conclusion: