2A201-474E
MODEL TSX-101A
(MULTISLICE CT SCAN SYSTEM)
SYSTEM S/N_________
(2A201-474E)
IMPORTANT!
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.
2
No. 2A201-474E
REVISION RECORD
3
No. 2A201-474E
4
No. 2A201-474E
RECORD OF INSPECTION
2. Function test
(1) Gantry :
(6) Interlock :
(7) Emergency :
5
No. 2A201-474E
3. Image quality
Scan the following phantoms. For each of the two reconstructed images, perform ROI
calculation on the center area of 250 × 250 pixels.
* For the Q/A phantom, scan the image noise module and perform measurements using the
Q/A software.
Measured value
Position Material Reference value
1st image 2nd image
(a) Air -1030 ∼ -990
(b) Low-density polyethylene -115 ∼ -95
(c) Delrin 350 ∼ 370
(d) Teflon 960 ∼ 1000
6
No. 2A201-474E
(3) Uniformity
Measured value
Reference value
1st image 2nd image
≤4
Measured value
Reconstruction function Reference value
1st image 2nd image
FC 30 ≥ 9 lp/cm
FC 90 ≥ 14.5 lp/cm
Measured value
Reference value
1st image 2nd image
≤ 0.6
Measured value
Reference value
1st image 2nd image
≤ 1.2
7
No. 2A201-474E
Measured value
Slice thickness Reference value
1st image 2nd image 3rd image 4th image
8 mm × 4 7.2 ∼ 8.8 mm
6 mm × 4 5.4 ∼ 6.6 mm
4 mm × 4 3.6 ∼ 4.4 mm
3 mm × 4 2.7 ∼ 3.5 mm
2 mm × 4 1.8 ∼ 2.5 mm
1 mm × 4 0.8 ∼ 1.5 mm
0.5 mm × 4 0.4 ∼ 0.8 mm
Scan the 320 mm dia. TOS phantom and measure the CT number for an area of 25 x 25 pixels
at <a> to <f> for each of the two reconstructed images (see figure below).
8
No. 2A201-474E
4. System composition
9
No. 2A201-474E
5. Inspection judgement
This is to certify that all the descriptions given above are correct.
Date
Quality
Assurance Group
10 E
1385,SHIMOISHIGAMI,OTAWARA-SHI,TOCHIGI-KEN 324-8550,JAPAN