PQ Test Protocol MPA Sample Compartment 20231211 234224
PQ Test Protocol MPA Sample Compartment 20231211 234224
Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.26
Energy Test
________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0
1 25438cb8-051d-41e0-aa9d-a75b362063c8
2 3f5a0a9f-bb9b-4cfd-95ba-a62e60a1ab72
3 16855cf4-0f11-49ee-bb73-cfd0edfb1ea0
4 1bf769da-043f-41b2-8673-d6e3bca02e62
5 4363a3e0-9094-46ad-aa1c-6085fd437828
________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.25
0.15
0.05
-0.05
-0.15
25438cb8-051d-41e0-aa9d-a75b362063c8
________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (36.1 Deg. C, Peak is T. corr.), Spectrum
10245.73
0.8
Absorbance Units
0.4 0.6
0.2
a7878810-3e10-4820-a69a-e5901b886b5a
________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 20
5
2d928d11-c44f-44db-8eea-1967766ddab6
________________
Page 5 of 5 Page Signature