PQ Test Protocol MPA Sphere 20231211 201305
Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.46
Energy Test
________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0
1 9d63c847-9a9f-4fb2-991f-cd03133a089d
2 fd3b5b48-4a57-44f1-baee-99723b82658d
3 62de363e-5efd-4a72-b6a2-3cf491a5f7a0
4 1ff5b8de-60c6-4560-864d-bb1ed2ef0c46
5 1bb5e885-00c5-4455-b6c4-8f50490d1a56
________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.4
0.2
0.0
-0.2
-0.4
9d63c847-9a9f-4fb2-991f-cd03133a089d
________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (31.4 Deg. C, Peak is T. corr.), Spectrum
10245.85
0.8
Absorbance Units
0.4 0.6
0.2
63f6f757-9f79-43ca-b687-b2c4593dffbe
________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 520
600e07b3-3856-4960-b891-6616a202cae0
________________
Page 5 of 5 Page Signature