Servicing Instructions
- the electrical installation of the relevant room complies with the applicable
national and local requirements, and
* All rights are reserved for instruments, circuits, techniques, and names appearing
in the manual.
Contents
1 BLOCK DIAGRAM...................................................................................................................................... 5
2 FUNCTIONAL DESCRIPTION.................................................................................................................. 7
3 NAMES OF SIGNALS.................................................................................................................................. 9
4 PIN CONNECTIONS.................................................................................................................................. 11
6 INSTRUMENT VERSIONS....................................................................................................................... 15
9 ADJUSTMENT INSTRUCTIONS............................................................................................................. 25
9.1 INTERIOR VIEW ...................................................................................................................................... 26
10 JUMPER TABLE .................................................................................................................................... 27
13 PARTS LISTS.......................................................................................................................................... 33
14 APPENDIX: DRAWINGS........................................................................................................................ 35
GE Medical Systems TONOPORT V Page 4
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
Revision History
This manual is subject to the GE Medical Systems Information Technologies (GEMS IT)
change order service. The revision code, a letter that follows the document part number,
changes with every update of the manual. The initial version of the manual has the letter A.
1 Block Diagram
GE Medical Systems TONOPORT V Page 6
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 7
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
2 Functional Description
Measurement limits
- Max. cuff inflation time 100s
- Max. measurement time after measurement begins = 180s
GE Medical Systems TONOPORT V Page 8
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 9
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
3 Names of Signals
n/a
GE Medical Systems TONOPORT V Page 10
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 11
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
4 Pin Connections
See diagram
GE Medical Systems TONOPORT V Page 12
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 13
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
6 Instrument Versions
The instrument is available in one version only. However, the printout can be programmed in
5 different languages. These languages are:
old versions
- 2001589-001 TONOPORT V (GER) German
- 2001589-002 TONOPORT V (ENG) English
- 2001589-003 TONOPORT V (FRE) French
- 2001589-004 TONOPORT V (ITA) Italian
- 2001589-005 TONOPORT V (SPA) Spanish
new version
- 2001589-001 TONOPORT V
* Switch TONOPORT V on and wait until the time appears on the screen
* Press INFO five times: Display shows "H 5".
Press START: firmware version is now displayed, e.g.,
* From firmware version V 1.3 the language can be changed by pressing INFO until the
required language code is displayed. Pressing START finishes the language setting.
Honeywell-Sensor
GE Medical Systems TONOPORT V Page 17
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
Motorola-Sensor
GE Medical Systems TONOPORT V Page 18
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 19
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
- Start Hypertrm.exe
– Type in the name: TONOPORT V
– Type in any Phone Number and click to <OK>
- Go to FILE - PROPERTIES and make the same configurations as shown in the
pictures
- Go to "Transfer" - "Send Text File..." to transfer files to TONOPORT
GE Medical Systems TONOPORT V Page 20
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 21
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
• Extract all TONOPORT V, V1.3 Firmware files into one directory (e.g.
C:\TONOPORT\V1.3)
by starting: FIRMWARE FIRMWARE TONOPORT V FW13.exe
• Load all files from the Telix program into the same directory on the Hard Disk
• Start the Telix program (\telix.exe)..
Remarks:
Enter <ALT> + <p> to select the COM port
Enter <ALT> + <o> to select Ascii transfers. Select F - Line pacing ......... 2
Enter <ALT> + <s> and select ASCII to transfer files to TONOPORT
Delete screen with <ALT> + <c>
Telix can be terminated with <ALT> + <x>
1 Switch on TONOPORT V HH : MM
(time)
2 Press "INFO" 5 times H5
3 Press "START" 13 : 01
(13 = Version V 1.3
01 = language German)
4 Press "INFO" to select the correct language 13 : 01 (German)
(which was displayed at Step 1 of the Loading 13 : 02 (Enslish)
Procedure) 13 : 03 (French)
13 : 04 (Italian)
13 : 05 (Spanish)
5 Press "START" to save the language setting HH : MM
(time)
GE Medical Systems TONOPORT V Page 23
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
"E 02": Battery depleted. Code is displayed to indicate that the battery charge level is
insufficient for further measurements. The recorder differentiates between two states: the
memory has just been cleared (i.e., the battery test is performed with a higher power
consumption to ensure that a fresh battery will be inserted at the beginning of the
measurement) or measurements have already been taken.
"E 03": Measurement time over. Code is displayed when a measurement is not taken within
60 seconds (excluding cuff inflation time).
"E 06": Inflation time over. The max. inflation time of 60 s has elapsed. This condition
indicates a leak in the cuff or tubing, or a defective gasket.
"E 07": The pump will not increase the cuff pressure any further. Code is displayed when no
systolic reading could be found although the cuff pressure was already increased twice. The
measuring system waits until the next measurement is due.
"E 14": Diastolic reading below 40 mmHg. Code appears when the cuff pressure has dropped
to 40 mmHg and diastole could not be identified (TONOPORT does not measure diastolic
pressures below 40 mmHg).
"E 21": Difference between systolic and diastolic pressure too small (10 mmHg or less).
"E 27": Systolic pressure above 260 mmHg. Max. measuring range exceeded.
GE Medical Systems TONOPORT V Page 24
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
"E 29": Insufficient number of oscillations detected: For a correct measurement, the system
must detect at least 8 oscillations. Tighten the cuff so that one finger, but not two, can be
inserted between the patient's arm and the cuff. At the same time the device switches to a
deflation rate of 4 mmHg/s. When it detects more than 13 oscillations later on, the rate is
changed to 6 mmHg/s.
GE Medical Systems TONOPORT V Page 25
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
9 Adjustment Instructions
Only the pressure amplifier needs to be calibrated. This is done by switching on the
TONOPORT V and waiting until the time is displayed on the LCD.
Measure the direct voltages at TP2 and TP3. This is done by turning the trimming
potentiometer R59 with a screwdriver until the voltages at TP2 and TP3 are 2.0 ±0.2 V and
1.8±0.1 V, respectively.
Press "INFO" four times, LCD indicates "H 4".
Press "START" once, a 2-digit figure is displayed on the LCD.
Use a screw driver to turn the trimming potentiometer R53 until the LCD indicates a value of
approx. 50.
Connect a reference pressure gauge, a rubber bulb and air container (1 l) to the TONOPORT
pressure connector and press "START".
Using the rubber bulb generate a pressure of 100 mmHg in the tubing system and watch the
LCD.
Using a screwdriver, turn the trimming potentiometer R40 until the LCD indicates a value of
100.
Also check the values indicated on the LCD at pressures of 50 mmHg, 150 mmHg and
200 mmHg. No deviations may exceed ±3 mm Hg.
Switch the TONOPORT off.
GE Medical Systems TONOPORT V Page 26
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
R 59
R 40 TP 3
R 53 TP 2
TP 1
U-Lith 3 V
CR 2032
0V
+Batt
GE Medical Systems TONOPORT V Page 27
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
10 Jumper Table
11 Technical Specifications
Microprocessor
P80C558EFB
Storage capacity
128 x 8 bit ( = 1 Megabit) EEPROM
512 x 8 bit ( = 4 Megabit) EEPROM
Measuring range
Systolic pressure: 60 ... 260 mmHg
Diastolic pressure: 40 ... 220 mmHg
Mean pressure: 50 ... 260 mmHg
Pulse rate (HR): 35 ... 240 min-1
Recording period
30 h or 200 measurements max.
Battery
2 Mignon (size AA) NiMH batteries, 1.2 V, >1500 mA (HR-3U SANYO) or
2 Mignon (size AA) alkaline batteries
Battery charging time
2 ... 3 h
Max. cuff pressure
300 mmHg
Measuring method
oscillometric
Ambient conditions
Operation
Temperature +10...+ 40 oC
Relative humidity 30...75 %, no condensation
Atmospheric pressure 700...1060 hPa
Transport and storage
Temperature -10...+ 70 oC
Relative humidity 10...90 %, no condensation
Atmospheric pressure 500...1060 hPa
Dimensions and weight
Height 99 mm
Width 80 mm
Depth 27 mm
Weight 215 g, incl. Batteries
GE Medical Systems TONOPORT V Page 30
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
GE Medical Systems TONOPORT V Page 31
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
12 Technical Inspections
13 Parts Lists
Manuals
2001589-023 Service Manual (only in English)
2001589-006 Users Manual Ger.
2001589-007 Users Manual Eng.
2001589-008 Users Manual Fre
2001589-009 Users Manual Ita.
2001589-010 Users Manual Spa.
Accessories
2001589-011 Cable RS232 TONOPORT V ———>
PC (with CARDIOSOFT V4.14 or higher Version)
2001589-012 Cable RS232 TONOPORT V ———> Printer (LQ-Emulation)
Spare Parts
2002857-001 Cover Battery Skip
2002858-001 Cover Bottom with Switch and Cable
2002859-001 Cover Top with Keypad
2002860-001 Battery Buttom Cell Lithium CR2016 PCB 3V, 90mAh
2002861-001 Keypad
2002862-001 Battery Contact
2002863-001 Switch On / Off
2002864-001 Hose Connector (Luer)
2002865-001 Pump
2002866-001 PCB with Honeywell Valve and LCD Display
2002866-002 PCB with Motorola Valve and LCD Display
2006591-001 Label for Calibration: “Nächste MTK” (only for Germany)
GE Medical Systems TONOPORT V Page 34
Information Technologies 2001589-023 Rev. E V 1.3 / R 1.1
Exchange Units
14 Appendix: Drawings