Procedure manual
for AP Tool
System Software Update Procedure for Ver.2.03

■ Table of Contents

1 AP TOOL..................................................................................... 4
1.2 Configuration of AP Tool Disk ..................................................4
1.3 Configuration Windows Installable with AP Tool......................5
2 Creating AP Tool Disk ................................................................ 6
2.1 Acquiring the AP Tool Software ................................................6
2.2 Creating a Biochemical Analysis Condition Configuration File 7
2.3 Creating the Reagent Information File .....................................10
2.4 Creating a Contamination Avoidance Configuration File ........11
2.5 Creating the Method Deletion List ...........................................12
3 Installing Settings Using AP Tool ............................................ 13
4 Operation Check ....................................................................... 18

W-CA800758-01
System Software Update Procedure for Ver.2.03

■ Service Manual Revisions

All changes regarding the product and the service provided belong strictly to JEOL.
The content of the service manual may be revised at any time without prior notice.
The version of the manual packaged is the version at the time of product shipment or the
time the manual was printed. In the event that revisions are made to the document being
packaged together with the product, the revised version will not be provided
automatically to the owner.

■ Disclaimer

All information in this document provided by JEOL is for the purpose of using the
product.
Please note that JEOL shall not hold any legal responsibility or liability, under any
circumstances, for any kind of damage resulting from the use of this information.
Use of any part of this document is exclusively for qualified employees of JEOL or
JEOL associated companies or by those who have permission from JEOL or JEOL
associated companies.
This document might be used only in accordance with the terms of the license given by
JEOL. No part of this document may be reproduced or retransmitted in any form or by
any means, whether electronically or mechanically, including, but not limited to
photocopying, recording, or information recording and retrieval systems, and any
deviancy shall also be subject to litigation.

W-CA800758-01 1
System Software Update Procedure for Ver.2.03

■ Revision History

Revisions Date Description


1 2022-08-10 Release

2 W-CA800758-01
System Software Update Procedure for Ver.2.03

■ Overview

This document describes how to use the AP Tool for BM6070 (batch setting software
for various analysis conditions for BM6070/C).

■ Glossary

No. Term Definition


1 JCA-BM6070/C Automated clinical chemistry analyzer made by JEOL
2 JCA-BM6070/C The operation system software for the automated clinical
system software chemistry analyzer made by JEOL
3
J-OPECON The computer for the automated clinical chemistry
analyzer made by JEOL
4
McAfee Solidifier The allowlisting (safelisting) security software installed
in the J-OPECON
* McAfee and its logo are registered trademarks or trademarks of McAfee, LLC and/or
its affiliates in the U.S. and/or other countries.

■ Required Item

No. Part Number Item Quantity


1 N/A PC operates under BM6070/C System Software 1
1 N/A Recordable optical disk 1

W-CA800758-01 3
System Software Update Procedure for Ver.2.03

1 AP TOOL
AP Tool Software is utility software for field engineers to install settings for
biochemical analysis conditions, contamination avoidance, and reagent information on
the user's BM6070/C.
To use AP Tool Software, you need to export the various analysis condition setting files
created in advance and create an AP Tool Disk in conjunction with the exported files.

1.2 Configuration of AP Tool Disk


The AP Tool Disk to be created must contain all the files listed in Table 1. However, the
DETERGNT.TBL file is not required for unsupported languages.

File path Description


\SETUP.exe AP Tool Software
\MDE6070.dll Software used to delete biochemical
analysis settings
\Update\Specify_AnalCondVer.ini AP Tool version definition file
\Update\Specify_AnalCondVer.txt AP Tool version confirmation file
\EXPORT\(analytical parameter file) Biochemical analysis condition
configuration file
Note: The respective files for SI units
and common units must be configured
in AP Tool Disk.
\ETC\ReagentInfo.csv Reagent information file
\ETC\CONTMPIP.TBL File containing the contamination
avoidance settings for the reagent
probe.
\ETC\CONTMRRV.TBL File containing the contamination
avoidance settings for the cuvettes
\ETC\CONTMCHK.TBL Probe contamination avoidance and
cuvette contamination avoidance
with/without configuration file
・English Detergent configuration file
\ETC\DETERGENT\E\DETERGNT.TBL Separate folders according to the
・Chinese language of the software to be
\ETC\DETERGENT\C\DETERGNT.TBL installed.
・Russian Note: Files for unsupported languages
\ETC\DETERGENT\R\DETERGNT.TBL are not required.
\EXPORT\MethodDeletionList.txt File specifying the biochemical
analysis condition number to be
deleted.
Table 1 Configuration of AP Tool Disk

4 W-CA800758-01
System Software Update Procedure for Ver.2.03

1.3 Configuration Windows Installable with AP Tool

Setting screens that can be installed using AP Tool are shown in Table 2.
Name Description
Analytical Parameters Configure the settings for biochemical analysis.
(Chemistry)
Contamination Settings Configure contamination avoidance settings.
Reagent Information Specify the reagent settings.
System Test List The R-Code registered in the Reagent Information window can
be used for the settings.
Table 2 Setting windows list

W-CA800758-01 5
System Software Update Procedure for Ver.2.03

2 Creating AP Tool Disk


Create AP Tool Disk according to the following procedure.

2.1 Acquiring the AP Tool Software


The AP Tool software files can be acquired by installing the BM6070/C system
software on your computer.

The file path is as follows:


C:\A002\TOOLS\InternalTools\APTool

Figure 1 AP Tool Software

■ Procedure

1. Prepare a computer to install the BM6070/C system software on.

2. Insert the optical disk for AP Tool Disk into the computer.

3. Copy all files under the path C:\A002\TOOLS\InternalTools\APTool to AP


Tool Disk.

6 W-CA800758-01
System Software Update Procedure for Ver.2.03

2.2 Creating a Biochemical Analysis Condition


Configuration File
The Analytical Parameter (Chemistry) window contains the settings of the
biochemical analysis condition configuration files.
The only item numbers that can be exported are closed items.
It is necessary to create a biochemical analysis condition configuration file for SI units
and common units, respectively.

■ Procedure

1. Log on with the Service privilege, and open the Analytical Parameter
(Chemistry) window.

2. Display the biochemical analysis conditions for the item number that you
want to set.

3. If Calc.mthd is MSTD, perform the following steps. Skip these steps if


Calc.mthd is not MSTD.
Click RBL Setting to display the Reagent Blank Setting window.
Click OK on the Reagent Blank Setting window.
Click Save on the Analytical Parameters (Chemistry) window.

Figure 2 Procedure if Calc.mthd is MSTD

W-CA800758-01 7
System Software Update Procedure for Ver.2.03

4. If necessary, change the minor version for the biochemical analysis


conditions.
Note: The version number is on the upper-right of the Analytical Parameters
(Chemistry) window.

Figure 3 Version of biochemical analysis conditions

No. Name Description


1 Major Indicates the Fixed. Cannot be changed manually.
version BM6070/C system
software
2 Minor Version number for Any number can be entered manually. It is also
version each item. possible to have the number automatically incremented
when exporting.
3 Build Number of Cannot be changed manually.
version parameter changes It is automatically incremented when the item setting
is changed. It is also reset to "0" when the minor
version is changed.
4 Creation Date the Cannot be changed manually.
date configuration file When biochemical analysis conditions are installed
was exported with AP Tool Software, the date Analytical Parameters
(Chemistry) configuration file is created is
automatically entered.
Table 3 Description of biochemical analysis conditions version number
5. Click Save to save the settings.

6. Set and save other items that you want to set using the same procedure.

7. Click Export.

8. On the Export window, select Current, ALL, and Increment the minor
version number.
• Current: Only exports the currently displayed biochemical analysis conditions.
• All: Exports closed items in all biochemical analysis conditions.
• Increment the minor version number: Automatically increments the minor
version of the biochemical analysis conditions to be exported.

8 W-CA800758-01
System Software Update Procedure for Ver.2.03

9. Click OK. The biochemical analysis condition configuration file is exported


to the C:\A002\EXPORT folder.

Figure 4 Biochemical analysis condition configuration file

No. Name Description


1 Item number Analy.Cond.no.1 to 200
2 Test name Name
3 Metric system S : SI
C : Common
4 Analyzer BM6070C
5 Version Major version: 02
Minor version: 01
Build version: 00
e.g. For build version, the description is “020100”.
6 Creation date yyyymmddhhmmss
Table 4 Description of biochemical analysis condition configuration file

10. Export the configuration file for the other system units (SI or Common)
using the same procedure.

Figure 5 Selecting metric system

11. Copy the biochemical analysis condition configuration file created in the
C:\A002\EXPORT folder to the AP Tool Disk \EXPORT folder.

W-CA800758-01 9
System Software Update Procedure for Ver.2.03

2.3 Creating the Reagent Information File


The contents of the Reagent Information file will be reflected in the reagent information
file.
The item number and name in the biochemical analysis condition configuration file
must match exactly with the number and name in the reagent information file.

■ Procedure

1. Log on with a service account.

2. Enter the reagent information to be configured on AP Tool Disk on the


Reagent Information window.

Figure 6 Reagent Information

3. Click Save to save the settings.


When the settings are saved, ReagentInfo.csv in the C:\A002\ETC folder will be
updated.

4. Copy the ReagentInfo.csv file in the C:\A002\ETC folder to the AP Tool


Disk \ETC folder.

10 W-CA800758-01
System Software Update Procedure for Ver.2.03

2.4 Creating a Contamination Avoidance Configuration


File
The contents of the Contamination Settings window are reflected in the contamination
avoidance configuration file.
DETERGNT.TBL is the detergent configuration file. Since it differs for each language,
it is necessary to separate the files into different folders. If you want to support English,
Chinese, and Russian versions of the JCA-BM6070/C system software with a single AP
Tool Disk, create a DETERGNT.TBL file for each language in a folder for each
language in AP Tool Disk.

■ Procedure

1. Log on with a service account.

2. Open the Communication Settings window.

3. Enter the contamination avoidance conditions.

4. Click Save to save the setting. Saving the settings updates


CONTMPIP.TBL, CONTMRRV.TBL, CONTMCHK.TBL, and
DETERGNT.TBL in the C:\A002\ETC folder.

5. Copy CONTMPIP.TBL, CONTMRRV.TBL, and CONTMCHK.TBL in the


C:\A002\ETC folder to the AP Tool Disk \ETC folder.

6. Copy DETERGNT.TBL in the C:\A002\ETC folder to a folder in AP Tool


Disk in accordance with the language of the JCA-BM6070/C system
software.
• English: \ETC\DETERGNT\E
• Chinese: \ETC\DETERGNT\ C
• Russian: \ETC\DETERGNT\ R

W-CA800758-01 11
System Software Update Procedure for Ver.2.03

2.5 Creating the Method Deletion List


This file specifies the biochemical analysis conditions to be deleted and the settings
associated with the items.
The file is named "MethodDeletionList.txt" and is located in AP Tool Disk.
If a biochemical analysis condition configuration file with the same number as the
biochemical analysis condition number specified in the Method Deletion List is
configured in the EXPORT folder in AP Tool Disk, the existing settings will be deleted
and the new biochemical analysis parameter settings will be installed after the existing
settings are deleted.

■ Procedure

1. Create a text file using a text editor and save it as "MethodDeletionList.txt".


* Specify ANSI as the encoding type.

2. Enter the Analy. Cond. No. that you want to delete.


The format of a single line of text should be <Analy. Cond. No.><Line Feed>.
A semicolon ";" is recognized as a comment, and subsequent characters on the line are
ignored.
Note: If there is no biochemical analysis condition setting that you want to delete, leave
it blank.

Figure 7 Method Deletion List

3. Save the text file and copy it to the AP Tool Disk \EXPORT folder.

12 W-CA800758-01
System Software Update Procedure for Ver.2.03

3 Installing Settings Using AP Tool


This section describes how to install various settings using AP Tool.
If McAfee is installed in J-OPECON, set McAfee to update mode before using AP Tool.
After using AP Tool, return McAfee to enable mode.

■ Procedure

1. Start up the OPECON.


2. Close the JCA-BM6070/C startup window.
3. Insert the AP Tool Disk in the optical drive.
4. Run SETUP.exe in AP Tool Disk.
5. Click Next>.
6. Select distributor or service privileges, enter the applicable password, and
then click OK.

Figure 8 AP Tool log on window


Note: If a standard sample or QC request data remains in the target JCA-BM6070/C
system software, an error message will appear and AP Tool will close.
Restart AP Tool after following the error message.

W-CA800758-01 13
System Software Update Procedure for Ver.2.03

7. Configure the settings on the Unit and User defined parameter Selection
window.
Select the unit system for the biochemical analysis conditions that
you want to install.
• Current Units: Installs the units in accordance with the unit system of
biochemical analysis conditions already installed.
• SI Units: Installs biochemical analysis condition items using SI units.
• Common Units: Installs biochemical analysis condition items using common
units.
Select whether to overwrite the User defined fields.
• If you select the Overwrite User defined fields check box, user defined
parameters will be overwritten.
• If you do not select the Overwrite User defined fields check box, the current
user defined parameters will be kept.

Figure 9 Unit and User defined parameter Selection

14 W-CA800758-01
System Software Update Procedure for Ver.2.03

No. Name Description


1 Sub-analyt.conditions Only the values of Digits, Unit and Qualit.judge are user-
defined parameters.
2 Qualitative Judgement set All parameters
3 Real-time correction All parameters
Formula
4 Reanalysis conditions All parameters
5 Reanalysis conditions All parameters
(Urine)
6 Reanalysis condition Set All parameters
7 Standards setting All parameters
8 One-point Calibration All parameters
Setting
9 Multi -Standards Setting All parameters
10 Reagent Blank Setting All parameters
11 Normal value Setting All parameters
12 Calculation method Only the Prozone value is a user defined parameter.
setting
Table 5 User defined parameters for the biochemical analysis conditions

8. Click Next>.

W-CA800758-01 15
System Software Update Procedure for Ver.2.03

9. Configure the settings on the Full/Selective Install window.


Select Full Install or Selective Install.
• If you select Full Install, installable biochemical analysis items are
automatically selected.
• If you select Selective Install, select the check boxes for the biochemical
analysis items that you want to install.

Figure 10 Full/Selective Install window


Click Next>.

16 W-CA800758-01
System Software Update Procedure for Ver.2.03

10. Configure the settings on the Reagent Information Install window.


Select Full Install or Selective Install.
• If you select Full Install, installable biochemical analysis items are
automatically selected.
• If you select Selective Install, select the desired biochemical analysis items.
Select whether to install R-Code.
• If you select Register the R-Codes in the System Test List window, the value
in the R-Code column on the Reagent Information Install window is
registered in the System Test List.

Figure 11 Reagent Information Install window


Click Next>.

11. The System Software Backup window and Analytical Condition Install
in process window will appear and the following processes will be
executed automatically.
• Creation of a C\:A002 folder backup file.
Note that if any errors occur during the process, the backup created will
automatically restore the state of the system to the state before AP Tool was
used.
• Deletion and installation of the biochemical analysis item settings selected on
the Full Selective Install window.
12. The Method Deletion Result window appears.
13. Click Next>.
14. The Install complete window appears.
15. Click Finish to exit the AP Tool.

W-CA800758-01 17
System Software Update Procedure for Ver.2.03

4 Operation Check

1. Log on to the JCA-BM6070/C system software with Distributor or higher


privileges.

2. Check that the setting screens installed by AP Tool are installed as


expected:
• Analytical Parameter (Chemistry)
• Reagent Information
• Contamination Settings
• System Test List

3. Perform calibration and QC sample measurement as necessary to confirm


that there are no problems with the measurement results.

18 W-CA800758-01