System Software Update Procedure for Ver.2.03
■ Table of Contents
1 AP TOOL..................................................................................... 4
1.2 Configuration of AP Tool Disk ..................................................4
1.3 Configuration Windows Installable with AP Tool......................5
2 Creating AP Tool Disk ................................................................ 6
2.1 Acquiring the AP Tool Software ................................................6
2.2 Creating a Biochemical Analysis Condition Configuration File 7
2.3 Creating the Reagent Information File .....................................10
2.4 Creating a Contamination Avoidance Configuration File ........11
2.5 Creating the Method Deletion List ...........................................12
3 Installing Settings Using AP Tool ............................................ 13
4 Operation Check ....................................................................... 18
W-CA800758-01
System Software Update Procedure for Ver.2.03
All changes regarding the product and the service provided belong strictly to JEOL.
The content of the service manual may be revised at any time without prior notice.
The version of the manual packaged is the version at the time of product shipment or the
time the manual was printed. In the event that revisions are made to the document being
packaged together with the product, the revised version will not be provided
automatically to the owner.
■ Disclaimer
All information in this document provided by JEOL is for the purpose of using the
product.
Please note that JEOL shall not hold any legal responsibility or liability, under any
circumstances, for any kind of damage resulting from the use of this information.
Use of any part of this document is exclusively for qualified employees of JEOL or
JEOL associated companies or by those who have permission from JEOL or JEOL
associated companies.
This document might be used only in accordance with the terms of the license given by
JEOL. No part of this document may be reproduced or retransmitted in any form or by
any means, whether electronically or mechanically, including, but not limited to
photocopying, recording, or information recording and retrieval systems, and any
deviancy shall also be subject to litigation.
W-CA800758-01 1
System Software Update Procedure for Ver.2.03
■ Revision History
2 W-CA800758-01
System Software Update Procedure for Ver.2.03
■ Overview
This document describes how to use the AP Tool for BM6070 (batch setting software
for various analysis conditions for BM6070/C).
■ Glossary
■ Required Item
W-CA800758-01 3
System Software Update Procedure for Ver.2.03
1 AP TOOL
AP Tool Software is utility software for field engineers to install settings for
biochemical analysis conditions, contamination avoidance, and reagent information on
the user's BM6070/C.
To use AP Tool Software, you need to export the various analysis condition setting files
created in advance and create an AP Tool Disk in conjunction with the exported files.
4 W-CA800758-01
System Software Update Procedure for Ver.2.03
Setting screens that can be installed using AP Tool are shown in Table 2.
Name Description
Analytical Parameters Configure the settings for biochemical analysis.
(Chemistry)
Contamination Settings Configure contamination avoidance settings.
Reagent Information Specify the reagent settings.
System Test List The R-Code registered in the Reagent Information window can
be used for the settings.
Table 2 Setting windows list
W-CA800758-01 5
System Software Update Procedure for Ver.2.03
■ Procedure
2. Insert the optical disk for AP Tool Disk into the computer.
6 W-CA800758-01
System Software Update Procedure for Ver.2.03
■ Procedure
1. Log on with the Service privilege, and open the Analytical Parameter
(Chemistry) window.
2. Display the biochemical analysis conditions for the item number that you
want to set.
W-CA800758-01 7
System Software Update Procedure for Ver.2.03
6. Set and save other items that you want to set using the same procedure.
7. Click Export.
8. On the Export window, select Current, ALL, and Increment the minor
version number.
• Current: Only exports the currently displayed biochemical analysis conditions.
• All: Exports closed items in all biochemical analysis conditions.
• Increment the minor version number: Automatically increments the minor
version of the biochemical analysis conditions to be exported.
8 W-CA800758-01
System Software Update Procedure for Ver.2.03
10. Export the configuration file for the other system units (SI or Common)
using the same procedure.
11. Copy the biochemical analysis condition configuration file created in the
C:\A002\EXPORT folder to the AP Tool Disk \EXPORT folder.
W-CA800758-01 9
System Software Update Procedure for Ver.2.03
■ Procedure
10 W-CA800758-01
System Software Update Procedure for Ver.2.03
■ Procedure
W-CA800758-01 11
System Software Update Procedure for Ver.2.03
■ Procedure
3. Save the text file and copy it to the AP Tool Disk \EXPORT folder.
12 W-CA800758-01
System Software Update Procedure for Ver.2.03
■ Procedure
W-CA800758-01 13
System Software Update Procedure for Ver.2.03
7. Configure the settings on the Unit and User defined parameter Selection
window.
Select the unit system for the biochemical analysis conditions that
you want to install.
• Current Units: Installs the units in accordance with the unit system of
biochemical analysis conditions already installed.
• SI Units: Installs biochemical analysis condition items using SI units.
• Common Units: Installs biochemical analysis condition items using common
units.
Select whether to overwrite the User defined fields.
• If you select the Overwrite User defined fields check box, user defined
parameters will be overwritten.
• If you do not select the Overwrite User defined fields check box, the current
user defined parameters will be kept.
14 W-CA800758-01
System Software Update Procedure for Ver.2.03
8. Click Next>.
W-CA800758-01 15
System Software Update Procedure for Ver.2.03
16 W-CA800758-01
System Software Update Procedure for Ver.2.03
11. The System Software Backup window and Analytical Condition Install
in process window will appear and the following processes will be
executed automatically.
• Creation of a C\:A002 folder backup file.
Note that if any errors occur during the process, the backup created will
automatically restore the state of the system to the state before AP Tool was
used.
• Deletion and installation of the biochemical analysis item settings selected on
the Full Selective Install window.
12. The Method Deletion Result window appears.
13. Click Next>.
14. The Install complete window appears.
15. Click Finish to exit the AP Tool.
W-CA800758-01 17
System Software Update Procedure for Ver.2.03
4 Operation Check