Content
Meter malfunction
Meter power malfunction
Error Message Meter power insufficient
Meter malfunction
Strip malfunction
Blood drop malfunction
Inaccuracy handling
2
Error Message
Meter malfunction
Error Message
Strip malfunction
Inaccuracy handling
Meter malfunction
Remove battery, wait for 5 mins and reload new battery
If problem not be solved, please contact Bionime CS
Temperature Error
• Please wait 30 mins before testing
• Operating Temp.: 10 – 40℃ (50 – 104℉)
4
Error Message - Meter
System malfunction
• Remove the test strip from the meter.
• Turn off the meter (press the main button for 3 seconds.)
• Press the main button to turn ON the meter.
Sampling error
• Repeat the test by using a new test strip
• After a blood drop symbol appears on screen, the blood sample
could be applied 5
Meter malfunction
Meter malfunction
Meter power malfunction
Error Message Meter power insufficient
Meter malfunction
Strip malfunction
Inaccuracy handling
6
Meter power malfunction
Meter unable Meter unable Meter is not
to power on to power off working
Remove battery
Contact with
Customer Service
7
Battery malfunction
To prevent battery shortage and
unqualified battery leaking situation.
CR2032 & CR2016 thin battery
• Please use qualified lithium battery.
• Please remove the battery when meter
not use for a month.
CR2016 thin
battery causes
bad connection!!
8
Meter Power insufficient
Meter auto-running Meter battery error
Remove battery
Function normally
Contact with agent
or Bionime
The End Auto internal check after power on
10
Meter Error Message
Inserting port
ER1
Yes
Inserting port User Error, contact with agent
polluted by blood
No
Take a new strip to test
11
Strip malfunction
Meter malfunction
Error Message
Strip malfunction
Blood drop malfunction
Inaccuracy handling
12
The strip unable to measure
13
Strip malfunction
No measurement
NO Blood sample
Blood sample filled with view window Retesting
insufficiently
Yes
Yes
insertion point polluted or damaged Meter send back to agent or Bionime
NO
Insert a new strip
Insert CS Inform
Data agent
strip test
“Blood drop” flash *Sharking well CS liquid *To check the data with *Record data
*Press main button for *CS liquid drop on top of a data which showing *Send back strip
5 sec. entry CS mode vial cap outside of vial– *Offer strip lot no.
*Strip close to CS liquid inaccurate
to absorb
https://goo.gl/Jh3BpP
https://www.youtube.com/watch?v=vfXpnUM_UDo&feature=youtu.be
15
Inaccuracy handling
Meter malfunction
Error Message
Strip malfunction
Inaccuracy handling
16
Strip Inaccuracy
Accurate confirming
Contact with agent
Using test strip and
or Bionime
Control Solution to test
17
18
Thank you