You are on page 1of 4

Characteristics

Chapter 10
There are many excellent texts covering both age of the true value or, more commonly, as a per-
instrumentation and measurement. Some are centage of the span of the instrument according to
generic, one of the best being by Bentley (2004). Error
Others, which are specific to the process sector, are Accuracy = ± × 100%
Span
referred to in the next few chapters.
As seen in Figure 3.3 there is usually some A temperature of 85 ◦ C should give an output
combination of sensor, transducer and transmitter of 10.4 mA whereas the output corresponding to
associated with the measurement process. These 84.5 ◦ C is only 10.08 mA. Thus the accuracy of the
devices may be characterised in many ways, e.g. instrument is ±2.0%. It does not particularly mat-
according to principle of operation, performance, ter what basis is used for quoting accuracy, as long
physical design or commercial considerations. For as it is properly understood.
specification purposes the most important charac-
teristics are as follows: b. Precision, which is often confused with accu-
racy, is associated with analogue signals and is a
a. Accuracy. This is a measure of the closeness of function of the scale used for measurement. Sup-
the measurement of a variable to the true value pose the above temperature measurement was in-
of that variable. Consider, for example, an instru- dicated on a gauge whose scale was calibrated from
ment for temperature measurement as shown in 75 to 100 ◦ C in divisions of 0.5 ◦ C between which it
Figure 10.1. Its calibration is linear, with an input is possible to interpolate to within 0.1 ◦ C. The pre-
range of 75–100 ◦ C (span of 25 ◦ C) corresponding cision is 0.1 ◦ C or ±0.4% of scale. Thus the mea-
to an output range of 4–20 mA (span of 16 mA). surement which is only ±2.0% accurate can be read
with a precision of ±0.4%.

θ( o C ) i ( mA ) c. Resolution is, strictly speaking, the largest


Trans.
change in input signal that can occur without any
corresponding change in output. Resolution is as-
Fig. 10.1 Instrument for temperature measurement
sociated in particular with digital signals and is
a function of the number of bits used to repre-
Suppose a temperature of 85 ◦ C is measured as sent the measurement. Again, using the 75–100 ◦ C
84.5 ◦ C.In absolute terms the error is 0.5 ◦ C and the example, suppose the signal is stored in a 10-bit
instrument is said to have an accuracy of 0.5 ◦ C.It is register. The register has a range of 10232, each bit
more usual to express accuracy on a percentage ba- corresponding to 25/1023 ◦ C, giving a resolution of
sis. The error may be expressed either as a percent- approximately 0.1% of range.
58 10 Characteristics

d. Repeatability is a measure of an instrument’s i ( mA )


ability to produce the same output signal for dif- 20
ferent instances of a given input signal. Lack of
repeatability is usually due to random effects in idn
the instrument or its environment. Repeatability is
characterised by statistical metrics,such as average 4 iup θ( o C )
value and standard deviation. For process control 75 100
purposes, where consistency of operation is of the
essence,an instrument’s repeatability is often more Fig. 10.3 Characteristic with hysteresis
important than its accuracy.
Hysteresis is also quantified in terms of the max-
e. Linearity, as discussed in Chapter 4, pervades imum difference between the two outputs ex-
automation. It is highly desirable that instruments pressed as a percentage of the span, i.e.
have linear I/O characteristics. If any significant
(idn − iup )max
non-linearity exists, it is normal to quantify it in × 100
terms of the maximum difference between the 16
output and what it would be if the characteris- g. Rangeability. Instruments are designed to have
tic were linear, expressed as a percentage of the some specified I/O relationship, linear or other-
instrument’s span. wise. Normally this relationship is valid over the
For the example of Figure 10.1, suppose the whole range. Sometimes, however, the relationship
ideal (linear) and nonlinear outputs are given by does not hold at one end or other of the range,
iid = −44.0 + 0.64. inl = f() usually the bottom end. Rangeability is expressed
as the ratio of the proportion of the scale over
and are as depicted in Figure 10.2 which the relationship holds to the proportion
over which it doesn’t. For example, an instrument
that is linear over the top 96% of its range has a
i ( mA ) rangeability of 24.
20

ideal
h. Drift. This is a time dependent characteristic,
normally associated with analogue signals. It is the
extent to which an output signal varies with time
4 non linear θ( o C ) for a constant input. Drift is invariably due to heat-
75 100 ing effects. Many components take a while to heat
up after being switched on so a device’s output may
Fig. 10.2 Ideal and nonlinear characteristics
initially drift prior to stabilising. Drift can also be
an indication that a device is overheating.
The percentage nonlinearity is thus
i. Sensitivity, otherwise referred to as gain, is a
(−44.0 + 0.64. − f())max steady state measure of an instrument’s respon-
× 100
16 siveness.It is defined to be the change in output per
unit change in input.Assuming that its I/O charac-
f. Hysteresis is a particular form of non-linearity teristic is linear across the whole range, sensitivity
in which the output signal may be different for is given by
any given input signal according to whether the
input is increasing or decreasing, as illustrated in Output span
Figure 10.3. Gain =
Input span
10 Characteristics 59

The gain of the example of Figure 10.1 is 0.64 m. Process Interface. Those parts of an instrument
mA/◦ C. Sensitivity is determined by the specifi- that come into direct contact with the process have
cation of the I/O ranges and, for control purposes, to be capable of withstanding the temperatures and
is arguably the most important of an instrument’s pressures involved. They also have to be chemi-
metrics. cally resistant to the process medium, whether it
be acidic, caustic or otherwise. There is therefore a
j. Speed of response. An instrument’s dynamics need to specify appropriate materials of construc-
are of particular interest. The types of instrumen- tion.The process interface,surprisingly,is the most
tation used in process automation are relatively expensive part of an instrument. It usually consists
slow. It is therefore normal practice to approxi- of a casting of steel, or some other alloy, which
mate their response to that of a first order system houses the sensor. Its manufacture requires a sig-
and to characterise it in terms of a time constant nificant amount of machining and assembly work,
as explained in Chapter 69. If necessary, this can be all of which is labour intensive.
found empirically by carrying out a step response
test and measuring the time for a 63.2% change in n. Physical. Mundane, but of vital importance to
output. the successful installation of an instrument, are
some of its physical characteristics. For example,
k. Reliability. Process instrumentation has to be the correct fittings or flanges have to be speci-
designed to operate, with minimal maintenance, fied for the process interface, appropriate brackets
for 24 h/day throughout the life of the plant. High for panel or post mounting, suitable connectors
reliability is essential.For analogue devices,such as for wiring purposes, etc. The intent with regard to
sensors and transmitters, the most useful means of power supply must be made clear. There are many
quantifying reliability is either in terms of failure options: the supply may be local to the instrument
rates, e.g. number of failures/106 h, or in terms of or provided via the signal lines, the voltage may
the mean time between failures.MTBF is applicable be a.c. or d.c., the instrument may require special
to any type of instrument that can be repaired by earthing arrangements, screening may be speci-
the replacement of a faulty component or unit. For fied, etc.
discrete devices used in protection systems,such as
switches and solenoids, it is normal practice to ex- o. Availability. There is much variety in the supply
press reliability in terms of the probability that the of process instrumentation: different techniques of
device will fail on demand. Reliability is discussed measurement, alternative designs, choices in func-
in detail in Chapter 53. tionality, etc. There are many suppliers and the
market is both competitive and price sensitive. The
l. Intrinsic Safety. When an instrument has to be best way of identifying the principal suppliers of
situated in a hazardous area, i.e. one in which an instrument type is to consult a trade catalogue,
flammable gas/air mixtures could exist, it must be such as the InstMC Yearbook. Looking at the tech-
intrinsically safe. The extent of the IS requirement nical press on a regular basis helps to keep abreast
depends upon the zone number, the gas group of current developments. Some journals publish
and the temperature class. If the IS specification surveys which are particularly useful.
for an instrument cannot be met, perhaps on the
grounds of cost, then it is acceptable to install it p. Commercial. Most companies attempt to stan-
in a flameproof enclosure. Depending on what an dardise on a limited range of instrumentation and
instrument is connected up to, it may be necessary to maximise on its interchangeability. This reduces
to specify a barrier at the zone boundary. All of the inventory of spares required, which could oth-
these safety related issues are explained in detail in erwise be extensive, and makes for more effective
Chapter 52. maintenance and repair. This standardisation in-
60 10 Characteristics

evitably leads to a preferred, i.e. restricted, supplier tice because capital costs and operating costs come
list. Factors taken into account include instrument from separate budgets.
performance, product support, company stability
and pricing policy.
Comments
q. Cost. Whereas the purchase price of an instru- Specifying an instrument inevitably involves mak-
ment is relatively easy to establish, its true capital ing compromises between desirable technical char-
cost is not. This is because installation costs are acteristics and commercial criteria, subject to the
specific to the application and can vary widely. It is constraints of company policy. To do this properly
not uncommon for the installed cost, i.e. the pur- requires a good understanding of the application
chase price plus installation costs, to be more than and its requirements. It also requires alot of de-
double the purchase price. Strictly speaking, in tailed information. The extra time and effort spent
specifying an instrument, its life cycle costs should in getting the detail right at the specification stage
be taken into account, i.e. its installed cost plus saves much money and inconvenience correcting
maintenance costs. This seldom happens in prac- mistakes later on.

You might also like