WinCE Upgrading Instruction

1. Download the “WinCE Upgrading Package” folder and save it C disk.


2. Prepare PC, instrument, power supply cable, USB mouse, RS-232 serial cable (2-3,
3-2, 5-5), and crossover network line (The network line is crossover cable, not
straight cable).
3. Connect instrument with PC as below figure.

Crossover net cable, USB MOUSE


another end to PC

Power supply cable

RS-232 cable

Figure 1
4. Enter “WinCE Upgrading Package” folder.

WinCE software

Figure 2
5. Enter “Tftpd32” folder and install the Tftp tool.

Double click to install

Figure 3
6. Click “Next”.
Figure 4
7. Click “Install”.

Figure 5
8. Click “Close” to finish installation of Tftp.
Figure 6

9. Click Tftp icon in desk to run Tftp tool.

Figure 7
10. Click “Browse” to change “Current Directory” to “C:\WinCE Upgrading Package”,
where the Windows CE software is.
Note: Do not close Tftp tool in this step.
Figure 8
11. Enter “Clinet” folder.

Figure 9

12. Click to run “Clinet” software and change “Baud Rate” as “115.2k”.
Figure 10
13. Click “Ok”, you will see below figure.

Figure 11
14. Power on the instrument and click “Enter” in keyboard when you see “Hit any key to
stop auto boot” within 2 seconds.
Note: This step is important. If you do not click “Enter” within 2 seconds when you see
“Hit any key to stop auto boot”, you need to close “cliNet” and power off the
instrument. 2 minutes later, go step 14 again.
Click “Enter” within 2 seconds

Figure 12

15. Input “setenv serverip PC IP address”, such as “setenv serverip 192.168.1.104”


when you see SJPG2410#.
Note: “192.168.1.104” is just an example and it is IP address of my computer.
As a result, you should enter IP address of your computer here.

Figure 13
16. Then input “setenv ipaddr + instrument IP address”, such as “setenv ipaddr
192.168.1.20” when you see SJPG2410#.
Note: As long as instrument IP address is different to PC IP address, it is ok. It means
that “192.168.1” should be the same and 104 should be changed to other number ( 2-
255, “192.168.1.1” is forbidden), such as “192.168.1.20”. For example, “202.168.1.1”
and “192.168.22.1”is wrong.
Figure 14
17. Input “run tftp_wince”

Figure 15
18. Then you will see below figure.
Figure 16
19. If Figure shows “TTTTTT……” for all the time, please change net card as below
setting and repeat the procedure from step 9. (Properties of my computer →
Hardware → Device Manager → Net Adapter → Select the property of the internet
card → Advanced. Change the link speed to 10 half mode.)
20. When you see “SJPG2410#” again, input “run write_wince”
Figure 17
21. The writing of Windows CE software is beginning.
Figure 18
22. Wait for about 3 minutes, the upgrading is finished when you see “SJPG2410#”.

Figure 19
23. Close “cliNet”, “Tftp”, and power off the instrument.

24. 2 minutes later, you can power on the instrument. If it is working well, you can use it

now.

25. If you can not find the software, but only mouse point showed on the screen. Please

upgrade software following “Software Upgrading Instruction V1.1e.doc” in “software

upgrading” folder.