Windowss
RS-232 cable
Figure 1
4. Enter “WinCE Upgrading Package” folder.
WinCE software
Figure 2
5. Enter “Tftpd32” folder and install the Tftp tool.
Figure 3
6. Click “Next”.
Figure 4
7. Click “Install”.
Figure 5
8. Click “Close” to finish installation of Tftp.
Figure 6
Figure 7
10. Click “Browse” to change “Current Directory” to “C:\WinCE Upgrading Package”,
where the Windows CE software is.
Note: Do not close Tftp tool in this step.
Figure 8
11. Enter “Clinet” folder.
Figure 9
12. Click to run “Clinet” software and change “Baud Rate” as “115.2k”.
Figure 10
13. Click “Ok”, you will see below figure.
Figure 11
14. Power on the instrument and click “Enter” in keyboard when you see “Hit any key to
stop auto boot” within 2 seconds.
Note: This step is important. If you do not click “Enter” within 2 seconds when you see
“Hit any key to stop auto boot”, you need to close “cliNet” and power off the
instrument. 2 minutes later, go step 14 again.
Click “Enter” within 2 seconds
Figure 12
Figure 13
16. Then input “setenv ipaddr + instrument IP address”, such as “setenv ipaddr
192.168.1.20” when you see SJPG2410#.
Note: As long as instrument IP address is different to PC IP address, it is ok. It means
that “192.168.1” should be the same and 104 should be changed to other number ( 2-
255, “192.168.1.1” is forbidden), such as “192.168.1.20”. For example, “202.168.1.1”
and “192.168.22.1”is wrong.
Figure 14
17. Input “run tftp_wince”
Figure 15
18. Then you will see below figure.
Figure 16
19. If Figure shows “TTTTTT……” for all the time, please change net card as below
setting and repeat the procedure from step 9. (Properties of my computer →
Hardware → Device Manager → Net Adapter → Select the property of the internet
card → Advanced. Change the link speed to 10 half mode.)
20. When you see “SJPG2410#” again, input “run write_wince”
Figure 17
21. The writing of Windows CE software is beginning.
Figure 18
22. Wait for about 3 minutes, the upgrading is finished when you see “SJPG2410#”.
Figure 19
23. Close “cliNet”, “Tftp”, and power off the instrument.
24. 2 minutes later, you can power on the instrument. If it is working well, you can use it
now.
25. If you can not find the software, but only mouse point showed on the screen. Please
upgrading” folder.