Service Report

SR Number PPMC_2024_01 Quotation No


Service Request No Engineer Arnel C. Dictado Jr.
P.O Number Invoice Number
Account Pampanga Premier Medical Centerlti
Location Pampanga Premier Medical

Machine details
Working Equipment
Machine type Asset no Serial number hours Software code Service contract type
5008S N/A 8VSAHY48 V4.62 PH – No Contract
Test Equipment
Serial number Equipment model Last Calibrated Test Type
N/A N/A N/A N/A

Spare parts
Batch / Serial
Code Name Quantity Location number
N/A N/A N/A N/A N/A

Work details
Start Date and Time 01/03/2024 09:00:00 AM End Date and Time 01/04/2023 11:30:00 AM
Reason for service 1B – Breakdown
Performed service type 00 – Repair
Error description Battery defective (Replacement of
battery) Overtime claimed (hours) 0.00
Call Type
Travel (Hours) 0.00 Labour (Hours) 0.00
Miscellaneous □ 0.00 Contract Charge □ 0.00
Comments
Work Details:
 checked in-line fuse for functionality – functional/intact
 per HD unit, T1 test fails during battery functional test
 back-up battery not functioning whenever there is a power interruption
 battery voltage not sufficient / quickly discharges over short period of time during power interruption
Recommendation:
 per technical manual, it is recommended to replace battery pack (24 V/7.2 Ah) at base rear every second
TSC / MA interval (48 months)
Resolution:
 Done
Acknowledged by:
Staff on duty
Michael Brent Apatt

We confirm this work has been done:

Arnel C. Dictado Jr.


Technical Service Engineer