You are on page 1of 1

Service report

SR Number PH0020330 Quotation No


Service Request No SPH0010085 Engineer John Lester Largado
P.O Number Invoice Number 1015689477
Account MCU-FDT Medical Foundation Hospital
Location MCU-FDT Medical Foundation Hospital

Machine details
Working Equipment
Machine type Asset no Serial number hours Software code Service contract type
Local RO 2017035 0 001 000 PH - PPT Contract

Spare parts
Batch / Serial
Code Name Quantity Location number
SE29 - John
PHT0001 SERVICE CHARGE 1
Lester Largado

Work details
Start Date and Time 20/12/2022 1:00:00 PM End Date and Time 21/12/2022 12:00:00 PM
Reason for service 1B - Breakdown
Performed service type 00 - Repair
Error description High Conductivity
Call Type - Overtime claimed (hours) 0.00
Travel (Hours) 0.00 Labour (Hours) 0.00
Miscellaneous 0.00 Contract Charge 0.00
Comments Run the Local RO
check the condutivity by conductivity meter with Sir Darrly and Maam Denise
Permeate conductivity at Product sample port is 4ppm
Permeate conductivity at Loop Return Sample port is 4ppm
Permeate conductivity is in normal range
Unit is working and Safe for clinical Use

We confirm this work has been done:

John Lester Largado Denise Penaredondo


Technical Service Engineer Technician