Professional Documents
Culture Documents
Bacterial
● The accuracy of the thermometer used to monitor the incubation
Content
temperature of an instrument should be verified every 6 to 12
(maximum <10 10^2 NA
months.
colony forming
● Reference thermometers must be certified by the NIST
unit/mL)
● For the monitoring of enzymatic reactions, thermometer should
agree within 0.1°C with the NIST reference thermometer,
pH NA NA NA
thermometers used to check refrigerator and freezer temp
should agree within 1.0°C with the reference thermometer;
Resistivity
discarded if they differ from NIST standard
(meagohms/c 1.0 (in-line) 1.0 0.1
● The temperature of the incubation area should be checked DAILY,
m at 25 degC
and the thermometer itself should be observed for splits in the
mercury column.
Silicates (mg/L 0.05 0.1 1.0
SiO2
maximum) PREVENTIVE MAINTENANCE
Particulate 0.22 um NA NA
● It is imperative that preventive maintenance procedures be
matter (filter
performed, and the results recorded for all laboratory
in system)
instrumentation.
● This includes maintenance of analytical balances, refrigerators,
Organic Activated NA NA
freezers, centrifuges, ovens, water baths, automated analyzers,
contaminants Carbon in
and all other laboratory equipment used for analyzing
system
specimens.
● Type I RGW is used for procedures that require maximum water ● PM is performed at scheduled times such as per shift, daily,
purity such as: weekly, monthly, or yearly.
○ Preparations of standard solution, ○ Celsius (°C) or centigrade - commonly used
○ Ultramicro chemical analysis, ○ Fahrenheit (°F)
○ Measurement of nanogram or subnanogram ○ Kelvin (°K) - SI Designation
concentration;
○ and tissue or cell culture (or both).
COMMON TEMPERATURE CONVERSIONS
○ Type I RGW is discouraged for storage, as contamination
chances are increased (higher resistivity)
Celsius to Fahrenheit °C(9/5) + 32
■ Proper procedure and handling of RGW
(especially Type I) should be followed.
Fahrenheit to Celsius (°F - 32)5/9
● Type II RGW is used in most laboratory determination in
chemistry, hematology, microbiology, immunology, and other
chemical laboratory areas.
● Type III RGW is used in most qualitative GLASSWARE AND PLASTICWARE
measurement/examinations, most procedures in urinalysis,
parasitology and histology, washing glassware and procedures
not requiring Type I or Type II water GLASSWARE PLASTICWARE
● Carbon-dioxide-free water is obtained by boiling Type II water.
○ This is used when carbon dioxide, ammonia and oxygen ● Breakable ● Highly
may affect analysis. ● Reuseable shockproof/shatterproof
● Ideal for acidic solutions ● Disposable
● High thermal and corrosion ● Ideal for alkaline solutions
CLINICAL LABORATORY SUPPLIES resistance ● Relatively inert
● Good Optical properties ● Flexible
THERMOMETERS ● Suitable for Cryogenic
(Freezing) experiments
● Are either an integral part of an instrument or need to be placed
in the device for temperature maintenance. ● Most laboratory supplies must satisfy certain tolerances of
● Three major types of thermometers: accuracy and fall into two classes or precision tolerance
○ Liquid in-glass, (American Society for Testing and Materials)
○ Electronic thermometer or thermistor probe, and ○ Class A: preferred for laboratory application
○ Digital thermometer ○ Class B: have twice the tolerance limits of class A
○ All temp reading devices must be CALIBRATED to ascertain
accuracy.
● Analytic Balances
○ Used for the preparation of any primary standard. Weight DIALYSIS
range of 0.01mg to 160g ● Is another method for separating macromolecules from a solvent
■ Mechanical analytic balance aka substitution or smaller substances.
balance
● Electronic Balances
○ Single-pan, uses an electromagnetic force to
counterbalance the weighed sample’s mass.